INSTRUCTIONS FOR USE

VIRUZYME & VIRUSOLVE+ EDS


(Routine Procedure)
1. Dilute 5 to 20ml of Viruzyme Enzymatic Cleaner with 1 Liter tap water.
2. CLEANING STAGE: Soak surgical and dental instruments to diluted Viruzyme
solution for at least 5-10 minutes.
3. After soaking, lightly brush or scrub to remove gross debris.
4. Rinse off thoroughly with tap water.
5. Pat dry to remove excess water.
Note: Viruzyme Solution could be used for the entire day, but would require
immediate disposal once heavy soilage is observed.
6. Transfer to the next stage or High Level Disinfection with Virusolve+ EDS
7. HIGH LEVEL DISINFECTION STAGE: Dilute 25ml Virusolve+ EDS with 1 liter
tap water.
8. Soak instruments for 5 minutes to achieve bactericidal, tubercolocidal, virucidal,
fungicidal and SPORICIDAL activities.
9. Final rinse with clean and sterile water. Must be done twice.
10.Dry with clean cloth.

Note: Virusolve+ EDS shelf life is 14 days upon dilution.


Efficacy of solution can be monitor red with Virusolve+ EDS MEC Strips.

Undiluted Virusolve + EDS solution shelf life is 4 years.

Undiluted Viruzyme Enzymatic Cleaner shelf life is 2 years.