Professional Documents
Culture Documents
ASHSV Manual
ASHSV Manual
qVisc 4320 AX
Autosampled High Speed Viscometer Service Manual
•Follow the procedure in this manual carefully. Use safe handling practices and use proper personal
protection equipment to protect yourself and your instrument. Ensure that the vent/exhaust tubing from
the vacuum system is routed to an outside location as directly as possible. The vacuum system vent
will contain high levels of vapor and may also have small amounts of liquid solvent. The exit points of
this vent must NOT be located near any ignition source, such as a furnace, HVAC, exhaust fan, etc.
This is not a complete list, if you are unsure, consult with someone who is familiar with your suggested
exhaust location.
•The qVisc 4320 AX does not need to have a fume hood or any special ventilation besides the above.
Under normal operation very little solvent vapor escapes the instrument. However, the qVisc 4320 AX
should be located in an air space with adequate air exchange to prevent the small amount of vapor from
building up.
•Never use anything but an inert gas to pressurize the vacuum system during emptying. Failure to do
so may cause a fire or explosion.
•Never use the instrument if you see solvent has leaked or suspect a solvent leak or you can smell or
suspect high levels of solvent vapor near the instrument. Investigate immediately.
•To prevent the vacuum waste trap tank from overflowing, always empty the waste tank when the
solvent is filled. If you are not using the vacuum system designed for your qVisc 4320 AX, your waste
tank should be larger than the solvent container.
•The qVisc 4320 AX has over temperature cutout safety switches and will automatically turn off the
heating system(s) in the event of an over temperature. However, if you are going to leave the
instrument unattended for longer than 2 days, we recommend you turn the instrument off.
• Unless otherwise indicated, the instrument can left on for many tasks. If you chose to do so, or the
instrument has not been turned off for at least 8 hours, be careful as parts in and around any bath
set to 100°C will be hot and parts in and around any bath set to 40°C will be noticeably warm.
• When reattaching parts, always start all fasteners loosely and only tighten when all are in place.
•When you need to slide the instrument out from under the autosampler for easier access, you must first
exit the software and then disconnect all inject lines from the autosampler.
Table of Contents
Table of Contents..............................................................................................................................................3
Instrument Setup............................................................................................................................................... 7
Stand assembly..............................................................................................................................................7
Stand height adjustment..............................................................................................................................11
Unpacking.......................................................................................................................................................11
Autosampler................................................................................................................................................11
Instrument................................................................................................................................................... 14
Instrument setup..............................................................................................................................................16
Glass install.................................................................................................................................................16
Instrument Prep.......................................................................................................................................16
Tube installation..................................................................................................................................... 19
Tube solvent line connections.................................................................................................................26
Tube inject line connections................................................................................................................... 28
Instrument reassembly............................................................................................................................ 29
Fill...........................................................................................................................................................29
Pc connections........................................................................................................................................ 31
Power up................................................................................................................................................. 32
Autosampler setup...................................................................................................................................... 32
Probe tip..................................................................................................................................................32
Drain line................................................................................................................................................ 37
Solvent / waste containers.......................................................................................................................38
Pc connections........................................................................................................................................ 40
Vacuum setup............................................................................................................................................. 41
Exhaust....................................................................................................................................................41
Instrument connections........................................................................................................................... 42
Auxiliary Vacuum line............................................................................................................................42
Waste spout.............................................................................................................................................43
Nitrogen.................................................................................................................................................. 43
Software ..................................................................................................................................................... 43
Install...................................................................................................................................................... 43
Com / basic settings................................................................................................................................ 44
Tray setup............................................................................................................................................... 44
Deck mapping......................................................................................................................................... 45
General................................................................................................................................................45
Trays................................................................................................................................................... 46
Samples...............................................................................................................................................48
Waste and Cleaning Ports................................................................................................................... 49
Inject Ports..........................................................................................................................................49
Calibration.................................................................................................................................................. 50
Temperature Adjustments.......................................................................................................................50
Autosampler Connections.......................................................................................................................52
Cut-out Adjustments............................................................................................................................... 53
Volume Adjustments.............................................................................................................................. 54
Instrument test.........................................................................................................................................56
Calibration.............................................................................................................................................. 59
Verification............................................................................................................................................. 60
Removal & Installation Of Parts.....................................................................................................................60
Case.............................................................................................................................................................60
Front Access Door.................................................................................................................................. 60
Side Access Doors.................................................................................................................................. 61
Case front................................................................................................................................................61
Case sides................................................................................................................................................63
Page 3 of 196 June 14 /2006 V1.0
qVisc 4320 AX Autosampled Viscometer
Case Skirt................................................................................................................................................64
Case top front..........................................................................................................................................65
Case top rear........................................................................................................................................... 65
Case rear................................................................................................................................................. 67
Utility panel............................................................................................................................................ 70
............................................................................................................................................................... 72
Tanks...........................................................................................................................................................72
Tank cover top........................................................................................................................................ 72
Tank cover.............................................................................................................................................. 72
Heater assembly......................................................................................................................................74
Tank........................................................................................................................................................ 76
Tubes.......................................................................................................................................................... 78
Tubes...................................................................................................................................................... 78
Tube Inject Line......................................................................................................................................80
Tube Solvent Line...................................................................................................................................82
Valves......................................................................................................................................................... 83
Connecting Tube.....................................................................................................................................83
Lower Solenoid Valve Assembly........................................................................................................... 84
Replace Coils (Lower Valves)................................................................................................................86
Replace Coils (Solvent Valves).............................................................................................................. 88
Valve Rebuild Kit (Lower Valves)......................................................................................................... 88
Valve Rebuild Kit (Tube Vent/Solvent Valves)..................................................................................... 90
Inject valves............................................................................................................................................ 92
Top Vent / Solvent Valves Assembly..................................................................................................... 93
Vent Tubes..............................................................................................................................................94
Electronics.................................................................................................................................................. 96
Cpu board............................................................................................................................................... 96
Solenoid board........................................................................................................................................97
Power supply...........................................................................................................................................98
Bath Temperature Controller..................................................................................................................99
Eprom................................................................................................................................................... 100
Fan (bottom rear).................................................................................................................................. 100
Fan (top)............................................................................................................................................... 101
Motor Run Cap..................................................................................................................................... 101
Replacement Temp Sensor................................................................................................................... 103
SSR....................................................................................................................................................... 104
Timing Sensor.......................................................................................................................................105
Wiring....................................................................................................................................................... 109
Instrument to PC serial cable................................................................................................................ 109
Motor Run Cap Wiring Harness........................................................................................................... 110
Optic Sensor Harness............................................................................................................................110
Power Receptacle Harness....................................................................................................................111
Power Switch........................................................................................................................................ 112
Power Switch Harness.......................................................................................................................... 114
Serial port harness.................................................................................................................................114
Solenoid board ribbon cable................................................................................................................. 114
SSR Wiring Harness............................................................................................................................. 115
Switch Supply Harness......................................................................................................................... 116
Plumbing...................................................................................................................................................117
Instrument Solvent feed Lines.............................................................................................................. 117
Internal Vacuum Plumbing................................................................................................................... 118
Replacement Vacuum Gauge................................................................................................................119
Vacuum Block...................................................................................................................................... 121
Vacuum Block Distribution Lines........................................................................................................ 123
Vacuum Pump Rebuild Kit...................................................................................................................124
Mounting...................................................................................................................................................125
Page 4 of 196 June 14 /2006 V1.0
qVisc 4320 AX Autosampled Viscometer
Cpu mount.............................................................................................................................................125
Inject Valve Mount............................................................................................................................... 126
Top vent valve mount........................................................................................................................... 127
Solenoid board mount...........................................................................................................................128
Lower valve mount............................................................................................................................... 128
Autosampler Parts.....................................................................................................................................129
Syringe.................................................................................................................................................. 129
Syringe Seal.......................................................................................................................................... 131
3 Port Valve Head.................................................................................................................................131
Pump Drive 3 Port................................................................................................................................ 132
Pump Drive 8+ Port..............................................................................................................................135
8+ Port Valve Head.............................................................................................................................. 137
Probe Line Tube Support......................................................................................................................138
Smart Valve Head.................................................................................................................................139
Z-Rack.................................................................................................................................................. 140
Autosampler Plumbing............................................................................................................................. 141
Clean Lines........................................................................................................................................... 141
Inject Lines........................................................................................................................................... 142
Smart Valve Lines.................................................................................................................................143
Solvent Feed Lines................................................................................................................................144
Autosampler Drain Line....................................................................................................................... 145
Inject Block...........................................................................................................................................146
Inject Ports............................................................................................................................................148
Insulation Block....................................................................................................................................150
Probe & Line.........................................................................................................................................151
Inject Block Drain Line........................................................................................................................ 154
............................................................................................................................................................. 155
Procedures.................................................................................................................................................... 155
Tubes........................................................................................................................................................ 155
Setting the Height................................................................................................................................. 155
Setting Oil Volume............................................................................................................................... 156
Tanks.........................................................................................................................................................158
Filling....................................................................................................................................................158
Draining................................................................................................................................................ 160
Optic Sensors............................................................................................................................................ 161
Adjusting...............................................................................................................................................161
Testing.................................................................................................................................................. 163
Autosampler..............................................................................................................................................164
Deck Mapping...................................................................................................................................... 164
General..............................................................................................................................................164
Trays................................................................................................................................................. 165
Samples.............................................................................................................................................167
Waste and Cleaning Ports................................................................................................................. 168
Inject Ports........................................................................................................................................168
Instrument................................................................................................................................................. 169
Temperature Calibration.......................................................................................................................169
Over Temperature Cutout Adjustment..................................................................................................171
Inject Volume Adjustments.................................................................................................................. 172
Repeatability Testing............................................................................................................................ 174
Instrument Testing................................................................................................................................ 177
Calibration............................................................................................................................................ 178
Vacuum Sensor Adjustment..................................................................................................................179
Sliding Out for Access..........................................................................................................................180
Draining Solvent................................................................................................................................... 181
Vacuum System........................................................................................................................................ 181
Startup...................................................................................................................................................181
Page 5 of 196 June 14 /2006 V1.0
qVisc 4320 AX Autosampled Viscometer
Shutdown.............................................................................................................................................. 181
Draining................................................................................................................................................ 181
Maintenance..................................................................................................................................................181
Recommended Schedule...........................................................................................................................181
Tubes........................................................................................................................................................ 181
Glass Cleaning...................................................................................................................................... 181
Back Flushing....................................................................................................................................... 182
Removing a Blockage with Cleaning Wire...........................................................................................184
Lower Valves............................................................................................................................................185
Cleaning................................................................................................................................................ 185
Operational Testing.............................................................................................................................. 187
Top Vent/Solvent Valves..........................................................................................................................189
Cleaning................................................................................................................................................ 189
Operational Testing.............................................................................................................................. 190
Autosampler..............................................................................................................................................191
Drive Lubrication..................................................................................................................................191
Troubleshooting............................................................................................................................................191
Symptom................................................................................................................................................... 192
Recommended Course of Action.............................................................................................................. 192
Instrument Setup
Stand assembly
• Place the top of the stand on the floor with the slot to the right side and facing up so the bottom is
visible. The top will be easy to tip over so have someone hold it in place or lean it against a wall.
• Prepare a screw with a lock washer. All stand bolts require a lock washer. Start all bolts only until
instructed to tighten.
• Place one of the sides next to the top with the bolt holes for the rear brace at the bottom, and
loosely bolt together with four bolts as shown.
• Place the rear brace and loosely bolt together with two bolts as shown.
• Place the other side next to the top and loosely bolt together with four bolts.
• Loosely bolt the rear brace to the new side with two bolts.
• Tighten all bolts until the lock washer is flattened using a 7/16 wrench.
• Tip the stand to the vertical position from the back.
• Bolt one of the four legs with casters on near the top. Note that there are two left and two right
legs. Each one requires four bolts. Loosely start all bolts, then tighten.
• Note that the deck has a front and back. The back of the deck does not have an obstruction on the
slide.
• Extend the slide from the stand. As the leveling feet are fully in for shipping you may need to raise
the stand to align the slide.
• Extend the slide over the corresponding slide on the deck. Duplicate on the other side.
• Slide the deck fully into the stand. As the stand has not been adjusted and has no weight on it, the
stand may slide backwards. To prevent this place the stand against the wall or simply hold onto
the stand while pushing the deck in. The deck slide will be stiff as you are sliding against the ball
bearing action, ensure the castors are unlocked.
• You should be able to rotate each castor with noticeable drag/slide on the floor.
• Adjust each foot until all castors have the same noticeable drag. This provides the best weight
spread between the stand and the deck when loaded.
Unpacking
Autosampler
Note: the autosampler box contains the fragile glass capillary tubes and accessories, handle with care.
• Lift the autosampler off of the box and place on the top of the stand. The autosampler is heavy and
awkward, two people are required.
• Position the autosampler so that the right edge is aligned with the slot in the top of the stand as
shown.
• Position the autosampler so that the inject block on the deck is centered on the slot from front to
back as shown.
Instrument
• Open the top of the instrument box.
• Remove the top wood cover.
• Remove the top pieces of foam.
• Slide the box off the top.
Vacuum system
Instrument setup
Glass install
Instrument Prep
• Slide the instrument deck fully out.
• Remove the top of the instrument by pulling up on in (it pops off).
• Unscrew the two screws holding the front panel access door and remove the door.
• Unscrew the six screws (three on each side) holding the front panel on.
• Remove the front panel by pulling the bottom out until it is past the front of the deck. You may
need to slide the instrument slightly forward. Be careful not to slide the front leveling feet off of
the deck.
• Disconnect the four optic sensors from the cpu board. Place the loose connectors into the top of
the tank cover.
• Remove the tank cover (10 screws). Do not remove the screw holding the cpu board mounting
bracket (marked with an X).
Tube installation
• Carefully open the box containing the glass capillary tubes.
Page 19 of 196 June 14 /2006 V1.0
qVisc 4320 AX Autosampled Viscometer
• The tubes are packaged in two layers and are labeled with their tube number (1-8). The tube
positions in the instrument go from left to right and then rear to front (see table below).
1 2
3 4
5 6
7 8
• Remove the tube to be installed from the packaging, start each bath with the rear-most tube (1 or
2). Locate the correctly numbered tube inject line and tube solvent line with fittings from the
accessories.
• Verify the correct placement of the o-rings with the optic heighting tool. The bubble joining the
two different sizes of capillary should be just below (not visible) the hole for optic two (shown by
the arrow).
• Unscrew the nut from the tube inject line fitting. The inject line can be identified by the white
compression fitting on the other end.
• Slide the nut and then the three o-rings over the inject tube of the glass as shown. The inject tube
is the glass tube that connects to the horizontal section from below, it also has the smaller inside
diameter.
• Using a slight amount of pressure, hold the fitting on the end of the glass and screw the nut on as
tightly as comfortable with your fingers. Do not tighten with a wrench or the tube will crack
when heated.
• Attach the solvent line to the other top glass tube in the same manner.
• Open the side door of the instrument.
• Slide the lower valves down by loosening the two wing nuts at the back of the valves.
• Loosen the vertical vent tube so it points down to the left.
• Unscrew the black and brass connecting tubes and place aside.
• Loosen the nut for the tube at the bottom of the tank.
• Insert the tube from the top of the instrument into the hole in the top of the optical sensor as
shown.
• Gently slide the tube down and through the nut at the bottom. You may have to remove the nut
and manually slide the two o-rings over the glass and retread the nut. Do not place any horizontal
force on the glass as it will break.
• Slide the tube down slowly, until the three o-rings positioned earlier just touch the top of the
optical sensor. If you slide the tube down too far and shift the o-rings, you we need to remove the
tube and re-height it.
• Orient the first tube with the fittings towards the front as shown below.
• Finger tighten the nut at the bottom of the tank while holding the top of the tube to prevent it from
turning and moving vertically.
• Replace the tank top (without the tank top cover). Do not screw into place at this time.
• Reconnect the four optical sensors to the cpu board. Verify that the wires are not crossed and are
connected to the correct jack.
• Using a ½ inch wrench tighten the nuts holding the tubes in place at the bottom of the tank by
another half turn. Hold the tubes from the top to prevent them from rotating or moving vertically.
• The tube solvent lines connect to the steel tube protruding from the bottom of the block at the rear
by simply sliding over it. The numbers are indicated on the solenoid control board just below the
block. For reference they are numbered 1, 3, 5, 7 – 8, 6, 4, 2 from left to right.
• Tighten finger tight. Using one index finger from each hand makes this easier.
• Repeat for each tube.
Instrument reassembly
• Verify that none of the tubes or their fittings are in contact with each other or the sides of the tank
top.
• Reattach the tank top.
• Reattach the black and brass connecting tubes top the lower valves.
• Slide the lower valves back up, ensure the black connecting tube slides over the protruding glass
tube. Slide the lower valves gently up until you meet resistance of the rubber stop inside the
connecting tube, be careful not to move the glass itself up (if this happens, lower the tube back
down to the 3 o-rings that set the height are correctly positioned on the optical sensor. Tighen the
two wing nuts at the back of the valves.
• Position the vertical vent tube so it points up.
• Orient the optical wires and lines neatly through the slots provided.
• Reinstall the front panel and access door in the reverse order of removal.
• Reinstallation of the tank top cover and instrument top are not done until the baths are filled.
Fill
• Verify that the drain valve is closed and has the drain plug tight.
• Verify that all tubes are inserted or plugged and that that all tube fittings are tight.
• Place rags around the tank cover to catch any accidental spills.
• Position another person so they can see the top bolts holding the tank window on, or pour slowly
and check periodically for correct fill level (check frequently near the correct level).
• Pour the silicon bath oil into the tank through the tank cover slowly being careful not to spill. A
container with an air hole prevents the oil from glugging, if one is not available, pour with the
spout at the top of the container to allow the air to get in easier.
• The correct fill levels are half way across the screws for the 40°C side and just under them for the
100°C side (the oil will expand slightly when heated).
• Orient the optical wires and lines neatly through the slots provided.
• Reinstall the tank cover tops.
Pc connections
• Connect the instrument RS-232 jack to one of the computers com ports (default is com2) using the
cable provided.
Power up
• Plug the instrument in.
• Turn on the instrument.
• The bath controllers will self test and then display the target and actual temperature.
• Check that all three cooling fans are running by simply placing your hand behind each one.
• Check that the bath circulator motors are running by the movement of the bubbles in the oil baths.
• Use the time needed to warm the baths up to continue with the rest of the setup.
Autosampler setup
Probe tip
• Your probe tip, line, z-rack, tube support and insulation block come as one assembly.
• To install, disconnect the fitting from the end of the largest line.
• Orient the z-rack with the threads on the left and insert into the hole from below.
• Position the z-rack so that the bottom is almost into the z-rack hole.
• Slide the tube through the z-rack until the insulation block and probe are correctly mounted. The
z-rack has a flat spot on the left side, the set screw of the insulation block should be over this flat
spot.
• Tighten the set screw ½ turn past finger tight.
• Slide the line back through the tube support into the long end.
• When the tube is visible in the access slot near the end of the tube support, use a small screwdriver
to pull the end out of the slot and continue sliding the tube through until enough slack is through to
allow the line to reach the entire deck but not to sag.
• Pass the line through the large hole in the tube support bracket on the right side of the autosampler,
insert the tube support into the small hole.
Drain line
• Insert the drain line into the drain hole on the autosampler deck.
• Push and rotate the drain line until you can grab it and pull it out from under the autosampler.
• Pull the drain line through the hole, and slide the end over the drain fitting on the inject block.
• Work the drain line up onto the drain fitting.
• Pass the other end of the drain line over the back of the stand.
• For heptane, place the solvent container on the stand next to the autosampler on the right.
• For toluene, place the solvent container in front of the waste container.
• Place the autosampler and instrument solvent feed lines into the solvent container until they reach
the bottom of the container. The instrument solvent lines should exit the instrument through the
top. Pass the lines from the front access door to the top just behind the front panel.
Pc connections
• Connect the autosampler to one of the computers serial ports (default com1) using the supplied
cable.
Page 40 of 196 June 14 /2006 V1.0
qVisc 4320 AX Autosampled Viscometer
Vacuum setup
Exhaust
• Connect the ½ inch nylon line to the exhaust vent connection.
• When using compression fittings, always check that you have both parts of the ferrule and nut over
the line before tightening. Tighten the fitting 1 turn past finger tight.
•Route the line to an outside location. Ensure that the vent/exhaust tubing from the vacuum system is
routed to an outside location as directly as possible. The vacuum system vent will contain high levels
of vapor and may also have small amounts of liquid solvent. The exit points of this vent must NOT be
located near any ignition source, such as a furnace, HVAC, exhaust fan, etc. This is not a complete list,
if you are unsure, consult with someone who is familiar with your suggested exhaust location.
• Alternately, you can route the exhaust line to a fume hood provided you place the line in a
container to catch any liquid solvent and ensure that the fume hood is always on when the vacuum
system is running. This should be checked whenever someone uses the vacuum system and
periodically while running to prevent solvent vapor build up if the fume hood fan fails.
Instrument connections
• Connect the 3/8 nylon line from the ‘viscometer’ connection on the vacuum system to the
instrument vacuum fitting. The shorter this run, the better.
Waste spout
• Connect the copper waste spout to the drain fitting, orient the spout facing down.
Nitrogen
• Connect the ¼ nylon line from a 10 PSI nitrogen (or other inert gas) to the nitrogen fitting. This is
not required to run the instrument, but will be needed to periodically drain the vacuum trap.
Software
Install
• Insert the cd into the cd drive. The installer should start automatically, if not, open the cd and run
setup.exe.
• Follow the on screen instructions to install the software.
Tray setup
• Place the racks on the deck as needed for the rack configuration selected by the customer (a typical
rack configuration is shown below).
• The software can have many rack configurations and allows the users to switch between them.
You need to map the deck for each configuration (you can cut and paste the inject block mapping
to all maps).
Deck mapping
• Setup the configuration to be mapped.
General
• Select Diagnostics from the setup menu.
• Go to the Arm Control tab.
• From here you can see the current layout, move and measure the current arm position.
• The arrow keys on the keyboard move the probe in the x and y directions. The page up and down
keys move the probe up and down. Tapping the keys once moves the arm exactly one count.
Holding the keys down causes the arm to continue moving in the direction request with slowly
increasing speed.
• Open the default deck configuration file (asdeck.asdc) in the HSV_DATA sub directory in
notepad. As objects on the deck are located this file will be updated with that information. The
software will allow the user to select from any *.asdc file in the HSV_DATA directory as a deck
layout.
• Use the keys to move the probe least ½ inch above the highest object on the deck.
• Read the current Z position and enter this as ZTravel, ZDispense, and ZStart in the general section.
Be careful not to accidentally read the coordinates in the ‘Move To’ editable boxes below the
current position data.
[General]
Trays = 2
Samples = 1
ZTravel = 500
ZDispense = 500
ZStart = 500
ZMax = 1615
InjectDia = 12.5
XRange = 1745
YRange = 2150
ZRange = 1670
• You can define sample positions in two ways, tray and samples. Samples define one position. A
tray defines any number of sample positions in rows and columns with only two corners setup.
The example deck shown above will have 2 trays (one for the small round racks and one for the
larger square ones, and 1 sample position for the large plastic container.
• To allow optimal sample pickup the software will only submerge the probe as far under the surface
as needed. To allow this to work the software needs information on the volume of the containers
used as well as the top and bottom levels. The simplest way to get volumes is to tare the container
on a balance and fill to the low side of the expected fill level and weigh the water, the weight in
grams is the same as the volume in ml’s. The number entered into the software should be slightly
smaller then the measured value in case some samples are short on volume. If you find that the
software submerges correctly near the top of the container but not as the container empties, then
your volume is too high. This feature can be turned off entirely or skipped in the deck layout by
simply entering 0 as the volume.
Trays
• Enter the number of trays you are defining in the general section.
• For each tray, place a container in the upper left position.
• Move the probe to just above the maximum liquid level and centered over the container.
• Find the section for the tray number you are defining. The tray section will be [Tray#] where # is
the tray number (eg ‘[Tray1] is tray number 1).
• If there is no section for the tray, create one.
• Enter the current x position as StartX.
• Enter the current y position as StartY.
• Enter the current z position as ZStart.
• Move the probe to just above the bottom of the container. Moving the container up to touch the
probe as the probe is moved slowly down in steps will allow you to check the space remaining on a
container you can’t see through.
• Enter the current z position as ZMax.
• Move the probe up and move the container to the lower right position.
• Move probe to just above the container and centered over the container.
Samples
• Enter the number of samples you are defining in the general section. Do not include sample
positions defined in trays.
• For each sample, place a container the correct position.
• Move the probe to just above the maximum liquid level and centered over the container.
• Find the section for the sample number you are defining. The sample section will [Sample#]
where # is the sample number (eg ‘[Sample1] is sample number 1).
• If there is no section for the sample, create one.
• Enter the current x position as x.
• Enter the current y position as y.
• Enter the current z position as ZStart.
• Move the probe to just above the bottom of the container. Moving the container up to touch the
probe as the probe is moved slowly down in steps will allow you to check the space remaining on a
container you can’t see through.
• Enter the current z position as z.
• Enter the container volume in ml’s as Vol
• If the container is round, enter the diameter in mm as Dia. This is only used to present a
representative display of the deck and does not need to be accurate.
• If the container is rectangular, enter the width in mm as SizeX and the depth as SizeY. This is only
used to present a representative display of the deck and does not need to be accurate.
• An example of a round container sample is shown below.
[Sample1]
ZStart = 550
z = 1615
x = 1231
y = 309
Dia = 40
Vol = 220
• Enter the current x and y position in the [Waste] section. Set the z to the same as the clean section
z.
Inject Ports
• Move the probe over the inject port to be mapped and center it.
• Move the probe down until it’s approximately 1 inch over the white inject port at the bottom of the
hole.
Page 49 of 196 June 14 /2006 V1.0
qVisc 4320 AX Autosampled Viscometer
• Using a stool, look down on the port from above. As the probe is in the way, look from both
forward of the z-rack and behind. You should see light coming from the center part of the inject
port. Move the probe up or down until the crescent shape of the inject port is seen.
• Based on the shape of the crescent(s), adjust the probe position as shown below until the crescents
are equal and centered as shown with the check mark. This will center the probe over the inject
port.
• Slowly move the probe down until you hear or see that it has made contact with the inject port.
Verify the probe is centered by carefully moving the probe slightly to the right and left, if the
probe was not centered it should fall into the inject port. Adjust the x or y position as required to
re-center the probe over its new position.
• Using the exercise x 1 button, listen to the sound of the contact with the inject port. Increase the
depth by small steps until the sound becomes noticeably louder. If a ‘z-axis error’ occurs, the
probe has been moved down too far, re-initialize the autosampler and start again.
• View the probe from the front and exercise the probe, watch that the probe does not shift to the left
or right as it makes contact with the port, adjust the x axis until this no longer occurs.
• Perform the same with the y axis by looking from the side.
• Enter the current x, y, and z position in the correct inject section of the deck file, the inject ports
are numbered from back to front.
Calibration
Temperature Adjustments
• It is highly recommended to calibrate the temperature controllers, if you are using heptane, you
must calibrate at least the 100°C side to prevent the solvent from boiling over. As long as the bath
Page 50 of 196 June 14 /2006 V1.0
qVisc 4320 AX Autosampled Viscometer
temperatures are close to the required set points the error will be calibrated out by the calibration
samples. Calibration also allows you to reset the controllers to adjust for the temperature sensors
drifting with time instead of recalibrating the entire instrument.
• If required, wait for the bath temperature to stabilize, it should read the same as the set point.
• Slide the instrument out.
• Insert an ASTM thermometer into the access port at the back of the tank cover. The ASTM
thermometer are required to get the correct accuracy (they only cover +/- 1 degree around the
target set point).
• Slide the thermometer down the access port until the bulb(s) are submerged. You can place a
grommet on the thermometer to allow you to release it.
• To correct the temperature controller, press the button to access the menu system.
• The controller will display ‘Ac,Cd’ in the lower half of the display alternating with the menu
number. Use the up and down arrow keys to select menu 4.
• If the temperature reading on the thermometer is high, press the down arrow ( ) until the
temperature reading on the controller reads the same as the thermometer.
• If the temperature reading on the thermometer is low, press the up arrow ( ) until the
temperature reading on the controller reads the same as the thermometer.
Autosampler Connections
• Slide the instrument fully back on the stand.
• Connect the eight inject lines from the autosampler to the tops of the inject valves. Note that the
lines are numbered from back to front.
• Pass the inject block drain line through the instrument to exit through the top.
Cut-out Adjustments
• Turn the over temperature cut outs down slowly, one at a time, until the cut out trips. This will
turn off the heater system and the temperature controller. You may not be able to cause baths set
for 40°C to trip, in that case simply leave the dial turned all the way down and skip the next step
for that cut out only.
• Turn the dial back up 1/8 of a turn and press the red reset button to turn the heater system back on.
Volume Adjustments
• You are now ready to start testing the instrument, startup the vacuum system.
• Flush the probe tip three times to prime the pump by pressing the flush button in the manual
control window.
• Quick clean all tubes. Observe the cleaning cycle to ensure correct operation.
• Have a quantity of new oil poured into one of the calibration containers. 100 – 200ml works well.
The new oil should be around an ISO100 hydraulic oil or a 5W30 or 10W30 engine oil. This oil
will be used instead of calibration fluid to adjust the instrument setup and test the instrument
before calibration.
• If you are running in Dual Temperature Mode, create four samples. Otherwise create eight
samples.
• Create the samples in the 'sample info' window of the software shown below.
• Select the correct autosampler location. Click on the drop down button , or press F4 to see the
graphical location selection window where you can verify the correct location, or select one by
clicking on it. Set all samples from the same location.
• Enter the typical viscosity for your test oil for all samples. As you are using a new oil, a number
from the spec book, or a typical value for the oil will be fine.
• Go to the Auto Analysis window:
Start of
Timing 1
Start of
Cleaning
Instrument test
• Once all of the volumes look correct, refill the test oil container.
• Enter the oil as a calibration oil in the calibration oils window (from the setup menu).
• To define a calibration oil you need to enter two lines on this window, one for the 40°C
measurement and one for the 100°C measurement.
• Ensure you enter the correct viscosities in the correct columns. The column labeled simply
viscosity is the viscosity of the oil at the temp in the next column. The column labeled ‘Visc @
40C’ is always the viscosity of the oil at 40°C, regardless of the temperature in the previous
column.
• Select the correct location for the oil.
• Click ok to save your changes.
• Select Calibration Curve from the setup menu.
• Set the minimum number of samples to 10 and the maximum RSD to 9999. This will cause the
system to run 10 samples and stop when calibrating.
Calibration
• Go to the calibration oils window.
• Set the minimum number of samples to 3 and the maximum RSD to 1.2. This will cause the
system to run 3 samples and stop when calibrating if the repeatability is good.
• Go to the manual control window.
• Quick clean all tubes.
• Flush the probe.
Page 59 of 196 June 14 /2006 V1.0
qVisc 4320 AX Autosampled Viscometer
• When all tubes have returned to idle, select calibrate all. Select all oils if no oils share locations,
otherwise select the ones placed on the deck.
• The instrument will now run at least three samples of each oil and generate a calibration curve.
• If you need multiple batches due to oils that share locations, run the additional oils after the system
finishes the first set in the same manner.
Verification
• Go to sample information window.
• Enter one sample for each tube or tube pair of one of the calibration fluids (or a qc oil with a
known viscosity).
• Go to the automated analysis window
• Ensure the 'Samples Filled on Start' option is selected.
• Clear the 'Quick Clean all Tubes on Start' option.
• Press start.
• Verify that the measured results match the known value to within +/- 2-3%.
• The instrument is now ready to run.
Notes:
• Unless otherwise indicated, the instrument can be left on for many tasks. If you chose to do so, or
the instrument has not been turned off for at least 8 hours, be careful as parts in and around any
bath set to 100°C will be hot and parts in and around any bath set to 40°C will be noticeably warm.
• When reattaching parts, always start all fasteners loosely and only tighten when all are in place.
• When you need to slide the instrument out from under the autosampler for easier access, you must
first exit the software and then disconnect all inject lines from the autosampler, disconnect the
inject block drain, and remove the waste line from the waste container (place a rag to catch any
drips from the waste line).
• When handling electronic parts, ensure you are grounded or using a grounded wrist strap.
Case
Front Access Door
Removal:
• Unscrew the two screws holding the front panel access door and remove the door.
Installation:
• Reattach the door using the screws, washers and lock washers removed.
• Loosen the screws and adjust the door left to right and front to back to maintain even seems with
the front panel.
Installation:
• Reattach the door using the screws, washers and lock washers removed.
• Loosen the screws and adjust the door left to right and front to back to maintain even seems with
the side panel.
Case front
Removal:
• Remove the front access door.
• Unscrew the six screws (three on each side) holding the front panel on. If you have removed the
three screws at the back of the side panel, the side panel will come off. Either reattach at least the
top screw in the back of the side panel to hold it in place, or remove the side panel first.
• Remove the front panel by pulling the bottom out until it is past the front of the deck. You may
need to slide the instrument slightly forward. Be careful not to slide the front leveling feet off of
the deck.
Case sides
Removal:
• Remove the side access door.
• If you are removing both side panels, remove the front panel first, otherwise the other side panel
screws will hold the front panel in place for you.
• Unscrew the three screws at the front of the side panel if not already removed.
• Unscrew the three rear screws holding the side panel on and remove the side panel. Be prepared to
support the side panel.
Case Skirt
Removal:
• Turn the instrument off.
• Remove the case top front panel.
• Remove the case top rear panel.
• Unscrew the eight ball studs and six nuts holding the skirt pieces to the frame. Do not remove the
screws, or the nuts and washers underneath the skirt (these have lock-tight on them to make
accidental removal difficult).
• Disconnect the top rear case fan harnesses (2) from the cpu board.
Case rear
Removal:
• Turn the instrument off.
• Place an index mark for the current location of both over temperature cutout dials.
• Remove both over temperature cutout dials by unscrewing the small flat head screw in the dial.
• Remove the six screws in the rear of both side panels. If you have removed the front screws in the
side panels either replace the screws, or fully remove both side panels.
• Unscrew the four screws holding the rear panel on. Be prepared to hold the rear panel in place.
Do not let the rear panel rest on or move the over temperature cutout dial posts or reset button as
they may break off.
• Tilt the back panel out at the bottom and slide down to remove.
• Place the back panel a short distance away from the rear of the instrument and disconnect the rear
fan harness.
Utility panel
Removal:
• Dry all tubes 4 times to remove all liquids from the internal vacuum lines.
• Turn instrument off.
• Turn off the vacuum system.
• Remove rear case panel.
• Remove over temperature cutout switch by unscrewing the two screws holding it in place. Move
the switch out of the way.
• Disconnect the internal vacuum line from the bulkhead fitting by reaching down from the top.
Tanks
Tank cover top
Removal:
• Remove the case top front.
• Unscrew and remove, the top of the tank cover (6 screws).
Tank cover
Removal:
• Turn the instrument off.
• Remove the case top front.
• Follow the instructions to disconnect the inject and solvent lines from the instrument to the glass
capillary tubes (ensure that you leave the lines connected to the tubes themselves).
• Disconnect the four optics harnesses from the cpu board.
• Remove the tank cover (10 screws). Do not remove the screw holding the cpu board mounting
bracket.
Heater assembly
Removal:
• Turn the instrument off.
• Wait for at least 10 minutes to allow the heaters to cool. Note, the 100°C side will still be hot.
• Remove the top rear case panel.
• Remove the rear case panel.
• Disconnect the motor harness, the heater harness, and the over temperature cutout harness.
• Unscrew the six screws holding the heater assembly to the tank.
• Gently remove the heater assembly by sliding it straight up. Be careful not to touch the
temperature sensor located on the front of the stainless steel baffle plate.
• As you pull up you may need to guide the over temperature cutout switch past the rear of the
instrument frame.
Tank
Removal:
• Turn the instrument off.
• Drain the tank.
• Remove the top front and top rear case panels.
• Remove all tubes from the tank.
• Remove the front, side, and rear case panels.
• Remove the tank cover.
• Remove the heater assembly.
• Remove the cpu board mount.
• Unbolt the four bolts holding the tank on from below. You can also do this by loosening the four
screws shown.
• Before tightening the tank bolts and screws, loosely reattach the cpu mount tank connecting rails.
Tubes
Tubes
Removal:
• With care you can remove and replace a tube while the instrument is full of bath oil. You may
drain the instrument if you wish. If the instrument has been running recently, the tubes will be hot,
use work gloves.
• Remove the case top front panel.
• Remove the case front panel.
• Remove the tank cover top.
• If you are removing tubes 1 or 2, remove the tank cover.
• Remove the solvent feed lines from the solvent container. Place into a pail on the floor to catch
outflow.
• Quick clean all tubes twice to drain the lines of any solvent.
• Exit the instrument software, and turn the instrument off.
• Locate and disconnect the top vent / solvent line from the top vent / solvent valve block. The
numbers are indicated on the solenoid control board just below the block. For reference they are
numbered 1, 3, 5, 7 – 8, 6, 4, 2 from left to right.
• Locate and disconnect the inject line from the inject valve. The inject valves are numbered from 1
to 8 going from back to front. You may unmount the inject valve if required to more easily access
the connections.
• Carefully slide the tube up until the bottom of the tube is close to the retaining nut.
• Use the tube plug to push the tube up by aligning the plug with the glass. You may need to gently
rotate the tube plug slightly back and forth or loosen the nut slightly to allow the plug to replace
the tube in the seal.
• Once the tube is free, remove it from the top of the instrument and set in a safe place.
• Tighten the retaining nut to prevent any bath oil from leaking.
Installation:
• Slide the nut and then the three o-rings over the inject tube of the glass as shown. The inject tube
is the glass tube that connects to the horizontal section from below; it also has the smaller inside
diameter.
• Using a slight amount of pressure, hold the fitting on the end of the glass and screw the nut on as
tightly as comfortable with your fingers. Do not tighten with a wrench or the tube will crack
when heated.
Installation:
• Slide the nut and then the three o-rings over the inject tube of the glass as shown. The inject tube
is the glass tube that connects to the horizontal section from below; it also has the smaller inside
diameter.
• Using a slight amount of pressure, hold the fitting on the end of the glass and screw the nut on as
tightly as comfortable with your fingers. Do not tighten with a wrench or the tube will crack
when heated.
Valves
Connecting Tube
Removal:
• If the instrument software is running, wait for the tube to be idle and then disable the tube.
• Open the side access door.
•Loosen the two wing nuts holding the solenoid valve to the black bracket and carefully slide the valve
down so the black rubber tube slides off the glass capillary tube.
•Unscrew the brass nut to remove the black rubber connecting tube.
•This step removes the vacuum hose. You will either need to shut off your vacuum system or have an
object to plug the 1/8” ID of the tubing. Slide off the black rubber vacuum tubing and brass ring
located at the bottom of the solenoid valve.
•Fully remove the two wing nuts at the back of the solenoid valve and remove the whole assembly.
• Disconnect the solenoid coil harness from the solenoid control board.
• When reinstalling the solenoid coils, the one with the blue dot sticker, or blue dot on the face of
the coil (where the nut goes) must be placed on the upper valve.
• Disconnect the solenoid coil harness from the solenoid control board.
• Carefully remove the armature by wiggling it gently back and forth. There is an o-ring seal at the
bottom of the armature that can be easily damaged.
• Remove the poppit from the armature.
• At this point the internals of the valve block should be cleaned with Q-tips and solvent.
• Clean the armature and poppit if new parts are not being used in reassembly.
• After assembly adjust the orientation of the vent tube if required by loosening the two armature
plate screws and attaching the vent tube. Rotate the vent so that it points up and slightly to the left
when tightened by pushing on the tube about 1 to 1.5 inches from the fitting, do not over tighten or
you will damage the seal. Tighten the screws when finished.
• Carefully remove the armature by wiggling it gently back and forth. There is an o-ring seal at the
bottom of the armature that can be easily damaged.
• At this point the internals of the valve block should be cleaned with Q-tips and solvent.
• Clean the armature and poppit if new parts are not being used in reassembly.
• When reinstalling the armature into the valve block, place a thin coat of new oil on the o-ring seal
to help insert it without cutting/pinching the o-ring. Gently rock the armature while applying a
small amount of pressure to reseat the seal.
Inject valves
Removal:
• Exit the instrument software.
• Turn off the instrument.
• Remove the case top front panel, and case front panel.
• Locate and disconnect the tube inject line from the bottom of the correct inject valve. The inject
valves are numbered 1 to 8 from back to front.
• Disconnect the inject valve harness from the solenoid control board.
• Unscrew the two screws holding the inject valve to the mounting bracket. There are two washers
on each screw, do not loose the washers as the screws will prevent the valve from operating if they
are missing.
• Remove all solvent valve coils and disconnect from the solenoid control board.
• Unscrew the two screws holding the solvent valve assembly to the mounting bracket.
Vent Tubes
Removal:
• Open the side access door.
• Disable the tube you wish to work on.
• Locate the vent tube you wish to remove.
• Unscrew the vent tube counter clockwise.
Installation:
• Fold the vent tube in half to make starting the threads easier.
• Loosely attach the vent tube until the seal starts to grab. The threads on the vent tube are plastic,
be careful not to cross thread.
• Tighten by pushing on the tube about 1 to 1.5 inches from the fitting. The vent tube should point
up (slightly to the right or left is ok). If not, adjust the orientation of the vent tube by removing the
solenoid coils and loosening the two armature plate screws. Then rotate the vent so that it points
up and slightly to the left when tightened by pushing on the tube about 1 to 1.5 inches from the
fitting. Tighten the screws.
Electronics
Cpu board
Removal:
• Turn the instrument off.
• Turn off the vacuum system.
• Remove the case front top panel.
• Remove the case rear top panel.
• Remove the inject valve mount and place the valves on one of the tank covers.
• Remove the plastic protection sheet.
• Disconnect all harnesses from the cpu board.
• Disconnect the vacuum sensor line from cpu board.
• Unscrew the 8 nuts and two standoffs holding the cpu board to the mounting bracket.
Solenoid board
Removal:
• Turn off the instrument.
• Remove the case front panel.
• Disconnect all harnesses from the solenoid board.
• Slide the solenoid board out to the side past the solvent feed lines (not shown).
Power supply
Removal:
• Turn off the instrument.
• Wait 15 minutes to allow all capacitors to discharge.
• Remove the case rear panel.
• Disconnect the power harnesses, and switching supply harness from the power supply.
• Rotate the instrument so the back is towards the front of the stand.
• Carefully move one foot of the instrument off the front of the stand so you have access to of the
four screws holding the power supply on (on the same side). Be careful to fully support the
instrument as it may wish to tip toward the foot you have slid off.
• Remove the two screws from below.
• Slide the controller out of the front of the case rear top.
Eprom
Removal:
• Turn the instrument off.
• Remove the case top front panel.
• Slide the protective plastic sheet out of the way.
• Using a small flat screwdriver, gently pry the eprom loose by alternately prying from both sides.
Installation:
• Line up the eprom with the pin one indicator on the lower left.
• Ensure all pins are correctly aligned, then push the eprom down.
• Unscrew the four screws and nuts holding the fan and finger guard to the case rear panel. Be
careful not to loose the nuts and washers from the other side.
Fan (top)
Removal:
• Turn the instrument off.
• Remove the case rear top panel.
• Unscrew the four screws and nuts holding the fan and finger guard to the case rear panel. Be
careful not to loose the nuts and washers from the other side.
• Locate and disconnect the ground wire that comes from the motor harness from the frame star
ground.
• Do not over tighten the wire retainers as they could cut into the wiring.
SSR
Removal:
• Turn instrument off.
• Remove rear case panel.
• Unscrew the two screws holding the ssr to the utility panel.
Page 104 of 196 June 14 /2006 V1.0
qVisc 4320 AX Autosampled Viscometer
Timing Sensor
Removal:
• Turn the instrument off.
• Remove the case front top panel.
• Remove the tank cover top.
• Prepare a clean rag to wipe the sensor and another to place the sensor on after removal.
• Use gloves as the sensor and tools used will become hot.
• Remove the capillary tube for the timing sensor you are removing.
• Disconnect the timing sensor harness from the cpu board.
• Using a 5/8 socket on an extension handle, reach down and unscrew the timing sensor. Hold onto
the wiring harness and allow it to spiral around the extension handle.
• Once the timing sensor has been unscrewed, use the extension handle and wiring harness to pull
the timing sensor out of the oil bath.
• Use the wiring harness to hold the sensor to the extension handle while installing the sensor as
shown.
Wiring
Instrument to PC serial cable
Removal:
• Exit the instrument software.
• Disconnect the cable from both the instrument and the computer.
• Disconnect the four harness wires from the cap. Note the correct grouping of the wires by color.
• Disconnect the two ground wires from the star ground post.
Power Switch
Removal:
• Turn the instrument off.
Page 112 of 196 June 14 /2006 V1.0
qVisc 4320 AX Autosampled Viscometer
• Unscrew the two screws that hold the harness to the utility panel. Be careful not to loose the nuts
and washers on the other side.
Plumbing
Instrument Solvent feed Lines
Removal:
• Exit the software.
• Disconnect the internal vacuum line from the bulkhead fitting on the utility panel by reaching
down from the top.
• Remove vacuum plumbing through the front access door. You will need to flex the plastic line.
• Remove the orange lock pin from the top seal of the gauge by sliding it out.
• Using a small flat head screw driver, remove the top seal.
Vacuum Block
Removal:
• Dry all tubes 4 times to remove all liquids from the internal vacuum lines.
• Exit instrument software.
• Turn off the vacuum system.
• Note: you can perform this task through the front and side access doors. Pictures were taken with
one of the tanks removed for clarity.
• Remove the inject block drain line from the vacuum block.
• Disconnect the vacuum block distribution lines from the vacuum block.
• Unscrew the two screws holding the vacuum block to the frame.
Page 122 of 196 June 14 /2006 V1.0
qVisc 4320 AX Autosampled Viscometer
• You can now follow the vacuum pump rebuild kit instructions to rebuild the pump in place.
• Reconnect all lines after installing the rebuilt kit in the reverse order of removal.
Mounting
Cpu mount
Removal:
• Turn the instrument off.
• Remove the case top front panel.
• Remove the case top rear panel.
• Remove the inject valve mount.
• Remove the cpu board.
• Unscrew the four screws holding the cpu mount plastic to the tank connecting rails.
• Unscrew the four screws holding the tank connecting rails to the tank. These screws have spacers
on the bottom, be careful not to drop the spacers.
• Place the inject valves and mount on one of the tank cover tops.
Autosampler Parts
Syringe
Removal:
•Wait until all tubes are idle.
•As the syringes contain solvent use proper protection equipment.
•Select Diagnostic Tools from the setup menu.
•Go to the maintenance tab and click the change syringe button.
•Select the syringe you wish to replace and click ok.
•The software will instruct you to remove the solvent line, this is the line that goes to the solvent
container.
Pump 1 Pump2
•The software will next lower the syringe and tell you to change it now (do not click ok).
•Unscrew the nut holding the bottom of the syringe to the drive.
•Unscrew the small Phillips screw on the black syringe clamp near the top of the syringe.
Installation:
•Install the syringe by reversing the previous steps.
•Press Ok, the syringe will be moved back up.
•Reattach the solvent line.
Syringe Seal
Removal:
•Remove the syringe.
•Remove the plunger from the syringe.
•Using pliers, gently remove the white syringe seal, do not scratch the plunger.
•Remove and replace the o-ring under the seal and press a new seal onto the plunger.
Installation:
•Install the syringe by reversing the previous steps.
•Press Ok, the syringe will be moved back up.
•Reattach the solvent line.
• Unscrew the four screws holding the pump drive to the autosampler. Remove the drive by sliding
it out the front.
• Unscrew the four screws holding the pump drive to the autosampler. Remove the drive by sliding
it out the front.
• Disconnect the clean lines and smart valve line from the valve head.
Z-Rack
Removal:
• Exit the instrument software.
• Turn off the autosampler.
• Remove the probe and line.
• Slide the z-rack vertically out of the top of the autosampler.
Autosampler Plumbing
Clean Lines
Removal:
• Remove the autosampler solvent feed lines from the solvent.
• Quick clean all tubes to remove most of the solvent from the lines.
• Exit the instrument software.
• The clean lines will still have some solvent in them, use appropriate personal protection.
• Disconnect the clean lines from the inject block, note the ordering of the numbers on the lines.
• Disconnect the clean lines from the valves, note the ordering of the numbers on the lines.
• Feed the lines through the exit hole on the right side of the autosampler.
Inject Lines
Removal:
• Exit instrument software.
• Disconnect the eight inject lines from the bottom of the inject block.
• Feed the lines through the exit hole on the right side of the autosampler.
• Unmount the inject block by loosening the two fittings clamping it down, and slide the block
vertically up.
Inject Block
Removal:
• Exit the instrument software.
• Disconnect the clean lines from the inject block.
• Disconnect the inject block drain line from the inject block.
• Unmount the inject block by loosening the two fittings clamping it down, and slide the block
vertically up.
Inject Ports
Removal:
• Exit the instrument software.
• Disconnect the inject line from the bottom of the inject block for the port you are removing.
• Using a 5/16 socket on an extension, unscrew the inject port through the top as shown.
• When the inject port is loose, tilt the socket and pull the inject port out vertically. If the inject port
falls into the hole, use the thick cleaning wire to extract the port by passing the wire through the
center of the port.
Insulation Block
Removal:
• Exit the instrument software.
• Turn off the autosampler.
• Unscrew the set screw holding the insulation block to the z-rack.
• Remove any remaining tubing from the insulation block (there is a short piece of tubing that goes
from the thick probe line to the probe itself).
• Slide the probe out of the bottom on the block (it will be tight).
• Disconnect the insulation block harness.
• Slide the probe line out of the hole in the right of the autosampler.
• Disconnect inject block drain line from inject block by sliding it off the metal tube.
Procedures
Tubes
Setting the Height
• Remove the tube from the instrument.
• Slide the three o-rings down the tube until they are just above the black line on the tube. If the
black line is missing or you wish to reset the height from scratch, move the o-rings down close to
the transition to the smaller capillary tube.
• Verify the correct placement of the o-rings with the optic heighting tool. The bubble joining the
two different sizes of capillary should be just below (not visible in) the hole for optic two (shown
by the arrow).
• Select the correct autosampler location. Click on the drop down button , or press F4 to see the
graphical location selection window where you can verify the correct location, or select one by
clicking on it. Set all samples from the same location.
• Enter the typical viscosity for your test oil for all samples. As you are using a new oil, a number
from the spec book, or a typical value for the oil will be fine.
• Go to the Auto Analysis window:
Start of
Timing 1
Start of
Cleaning
Tanks
Filling
• Remove the case top front panel.
• Remove the tank cover tops.
• Verify that the drain valve is closed and has the drain plug tight.
• Verify that all tubes are inserted or plugged and that that all tube fittings are tight.
• Place rags around the tank cover to catch any accidental spills.
• Position another person so they can see the top bolts holding the tank window on, or pour slowly
and check periodically for correct fill level (check frequently near the correct level).
• Pour the silicon bath oil into the tank through the tank cover slowly being careful not to spill. A
container with an air hole prevents the oil from glugging, if one is not available, pour with the
spout at the top of the container to allow the air to get in easier.
• The correct fill levels are half way across the screws for the 40°C side and just under them for the
100°C side (the oil will expand slightly when heated).
• Orient the optical wires and lines neatly through the slots provided.
• Reinstall the tank cover tops.
Draining
• Turn the instrument off.
• Open the side access door.
• Ensure the drain handle is closed and remove the drain cover plug.
• Connect the 3/8 nylon drain line to the drain port. Careful, metal parts will be hot.
• Place the other end in a container to catch the oil. The oil is expensive and can be reused if it is
kept clean.
• Open the drain valve. Warning: the oil will be hot, be careful. It is best to drain the oil while hot
to speed up the process.
• When the drain is complete, close the drain valve.
• Remove the drain line and re-plug the drain port.
• Close the side access door.
Optic Sensors
Adjusting
• Perform the instrument test procedure first.
• Using the information form the instrument test, adjustments can be made to improve the
repeatability as well as diagnose several instrument problems.
• Do not adjust optics unless indicated from the instrument test section, ensure tubes are clean,
flowing with no blockages and solvent clean correctly first.
• Note: the optics sensors have been adjusted at the factory and the settings included on the
instrument disk. You may need to perform an adjustment if tubes are placed in the wrong
positions, an optic sensor is replaced, or things have changed during shipping.
• The instrument performs two independent measurements of the oil flowing calling timings. In
each sensor housing there are four pairs of detectors (numbered 1 to 4 from top to bottom). Each
pair consists of a transmitter and receiver. When the oil redirects the light from the transmitter, the
drop in intensity triggers the receiver. The four detectors are used as stop and start marks for the
timings as shown below.
Page 161 of 196 June 14 /2006 V1.0
qVisc 4320 AX Autosampled Viscometer
Oil
Large Glass
Capillary Tube
Optic 1
Optic 2
Optic 3
Optic 4
Small Glass
Capillary Tube
• Normally, the optics start as slightly too sensitive (with offset 0) and will sometimes pick up the
leading edge of the liquid meniscus. As the meniscus shape is not constant from run to run this
will introduce variability in the timings.
• The purpose of adjusting the optics is to set the sensitivity so that the detectors trigger on the
middle of the meniscus and give the most repeatable results.
• If the start optics (1 and 3) are too sensitive and trigger early, the timing they start will be long.
• If the stop optics (2 and 4) are too sensitive and trigger early, the timing they stop will be short.
• You can adjust the optics by selecting instruments from the setup menu. Then go to the optics
offset tab. The window below will be displayed.
• If you see that one of the timings from the instrument test has 1-3 high outliers that do not
correspond to high outliers in the other timing, adjust the start optic offset by adding 1. Repeat the
test, if you run out of range, or the results get noticeably worse after 3 tries, consider adjusting the
stop optics by subtracting 1. It is unusual for optics to have sensitivities set below –1.
• If you see that one of the timings from the instrument test has 1-3 low outliers that do not
correspond to low outliers in the other timing, adjust the stop optic offset by adding 1. Repeat the
test, if you run out of range, or the results get noticeably worse after 3 tries, consider adjusting the
start optics by subtracting 1. It is unusual for optics to have sensitivities set below –1.
Testing
• Tests are only valid on a clean tube, tubes that have fluid in them will not read correctly.
• Select the diagnostics option from the setup menu.
• Go to the Maintenance tab.
• To test an optic simply select the optic in the ‘select a cell’ box.
• The instrument will then continually probe the optic and report back all intensities received in the
lower white section of the window.
• The results are displayed as five numbers. The first is an error code, a code of zero means no
error. A code of 8x indicates an error and the second digit indicates which optic(s) set the error
code (one bit is set for each sensor).
• The next four numbers indicate the amount of light needed to send light through the tube from 0-
255. Normally these numbers are in the 90-150 range for clean glass tubes, and 120-200 for tubes
darkened by used oils. A sensor with 0 or 255 has indicated an error.
• A 255 means that the system cannot generate enough light. This can be caused by fluid in the tube,
a dirty tube, or a optics, optics harness failure, or cpu board failure. Verify the tube is clean, then
swap the harness on the cpu board with the adjacent working optic, if the failure is gone, the optic
or harness has failed, replace them.
• A 0 means that the system could generate enough light, but that the optic is false triggering. This
could be because the sensitivity has been adjusted too high, turn down the sensitivity by placing a
7 in the optics offset and retest, if the problem does not go away the optic is bad, replace it.
Otherwise, increase the original offset by 1 until the problem goes away.
Page 163 of 196 June 14 /2006 V1.0
qVisc 4320 AX Autosampled Viscometer
Autosampler
Deck Mapping
• Setup the configuration to be mapped.
General
• Select Diagnostics from the setup menu.
• Go to the Arm Control tab.
• From here you can see the current layout, move and measure the current arm position.
• The arrow keys on the keyboard move the probe in the x and y directions. The page up and down
keys move the probe up and down. Tapping the keys once moves the arm exactly one count.
Holding the keys down causes the arm to continue moving in the direction request with slowly
increasing speed.
• Open the default deck configuration file (asdeck.asdc) in the HSV_DATA sub directory in
notepad. As objects on the deck are located this file will be updated with that information. The
software will allow the user to select from any *.asdc file in the HSV_DATA directory as a deck
layout.
• Use the keys to move the probe least ½ inch above the highest object on the deck.
• Read the current Z position and enter this as ZTravel, ZDispense, and ZStart in the general section.
Be careful not to accidentally read the coordinates in the ‘Move To’ editable boxes below the
current position data.
[General]
Trays = 2
Samples = 1
ZTravel = 500
ZDispense = 500
ZStart = 500
ZMax = 1615
InjectDia = 12.5
XRange = 1745
YRange = 2150
ZRange = 1670
• You can define sample positions in two ways, tray and samples. Samples define one position. A
tray defines any number of sample positions in rows and columns with only two corners setup.
The example deck shown above will have 2 trays (one for the small round racks and one for the
larger square ones, and 1 sample position for the large plastic container.
• To allow optimal sample pickup the software will only submerge the probe as far under the surface
as needed. To allow this to work the software needs information on the volume of the containers
used as well as the top and bottom levels. The simplest way to get volumes is to tare the container
on a balance and fill to the low side of the expected fill level and weigh the water, the weight in
grams is the same as the volume in ml’s. The number entered into the software should be slightly
smaller then the measured value in case some samples are short on volume. If you find that the
software submerges correctly near the top of the container but not as the container empties, then
your volume is too high. This feature can be turned off entirely or skipped in the deck layout by
simply entering 0 as the volume.
Trays
• Enter the number of trays you are defining in the general section.
• For each tray, place a container in the upper left position.
• Move the probe to just above the maximum liquid level and centered over the container.
• Find the section for the tray number you are defining. The tray section will be [Tray#] where # is
the tray number (eg ‘[Tray1] is tray number 1).
• If there is no section for the tray, create one.
• Enter the current x position as StartX.
• Enter the current y position as StartY.
• Enter the current z position as ZStart.
• Move the probe to just above the bottom of the container. Moving the container up to touch the
probe as the probe is moved slowly down in steps will allow you to check the space remaining on a
container you can’t see through.
• Enter the current z position as ZMax.
• Move the probe up and move the container to the lower right position.
• Move probe to just above the container and centered over the container.
Samples
• Enter the number of samples you are defining in the general section. Do not include sample
positions defined in trays.
• For each sample, place a container the correct position.
• Move the probe to just above the maximum liquid level and centered over the container.
• Find the section for the sample number you are defining. The sample section will [Sample#]
where # is the sample number (eg ‘[Sample1] is sample number 1).
• If there is no section for the sample, create one.
• Enter the current x position as x.
• Enter the current y position as y.
• Enter the current z position as ZStart.
• Move the probe to just above the bottom of the container. Moving the container up to touch the
probe as the probe is moved slowly down in steps will allow you to check the space remaining on a
container you can’t see through.
• Enter the current z position as z.
• Enter the container volume in ml’s as Vol
• If the container is round, enter the diameter in mm as Dia. This is only used to present a
representative display of the deck and does not need to be accurate.
• If the container is rectangular, enter the width in mm as SizeX and the depth as SizeY. This is only
used to present a representative display of the deck and does not need to be accurate.
• An example of a round container sample is shown below.
[Sample1]
ZStart = 550
z = 1615
x = 1231
y = 309
Dia = 40
Vol = 220
• Enter the current x and y position in the [Waste] section. Set the z to the same as the clean section
z.
Inject Ports
• Move the probe over the inject port to be mapped and center it.
• Move the probe down until it’s approximately 1 inch over the while inject port at the bottom of the
hole.
Page 168 of 196 June 14 /2006 V1.0
qVisc 4320 AX Autosampled Viscometer
• Using a stool, look down on the port from above. As the probe is in the way, look from both
forward of the z-rack and behind. You should see light coming from the center part of the inject
port. Move the probe up or down until the crescent shape of the inject port is seen.
• Based on the shape of the crescent(s), adjust the probe position as shown below until the crescents
are equal and centered as shown with the check mark. This will center the probe over the inject
port.
• Slowly move the probe down until you hear or see that it has made contact with the inject port.
Verify the probe is centered by carefully moving the probe slightly to the right and left, if the
probe was not centered it should fall into the inject port. Adjust the x or y position as required to
re-center the probe over its new position.
• Using the exercise x 1 button, listen to the sound of the contact with the inject port. Increase the
depth by small steps until the sound becomes noticeably louder.
• View the probe from the front and exercise the probe, watch that the probe does not shift to the left
or right as it makes contact with the port, adjust the x axis until this no longer occurs.
• Perform the same with the y axis by looking from the side.
• Enter the current x, y, and z position in the correct inject section of the deck file, the inject ports
are numbered from back to front.
Instrument
Temperature Calibration
• It is highly recommended to calibrate the temperature controllers, if you are using heptane, you
must calibrate at least the 100°C side to prevent the solvent from boiling over. As long as the bath
temperatures are close to the required set points the error will be calibrated out by the calibration
Page 169 of 196 June 14 /2006 V1.0
qVisc 4320 AX Autosampled Viscometer
samples. Calibration also allows you to reset the controllers to adjust for the temperature sensors
drifting with time instead of recalibrating the entire instrument.
• If required, wait for the bath temperature to stabilize, it should read the same as the set point.
• Slide the instrument out.
• Insert an ASTM thermometer into the access port at the back of the tank cover. The ASTM
thermometer are required to get the correct accuracy (they only cover +/- 1 degree around the
target set point).
• Slide the thermometer down the access port until the bulb(s) are submerged. You can place a
grommet on the thermometer to allow you to release it.
• To correct the temperature controller, press the button to access the menu system.
• The controller will display ‘Ac,Cd’ in the lower half of the display alternating with the menu
number. Use the up and down arrow keys to select menu 4.
• If the temperature reading on the thermometer is high, press the down arrow ( ) until the
temperature reading on the controller reads the same as the thermometer.
• If the temperature reading on the thermometer is low, press the up arrow ( ) until the
temperature reading on the controller reads the same as the thermometer.
• Turn the dial back up 1/8 of a turn and press the red reset button to turn the heater system back on.
• Select the correct autosampler location. Click on the drop down button , or press F4 to see the
graphical location selection window where you can verify the correct location, or select one by
clicking on it. Set all samples from the same location.
• Enter the typical viscosity for your test oil for all samples. As you are using a new oil, a number
from the spec book, or a typical value for the oil will be fine.
• Go to the Auto Analysis window:
Start of
Timing 1
Start of
Cleaning
Repeatability Testing
• Enter the test oil as a calibration oil in the calibration oils window (from the setup menu).
• To define a calibration oil you need to enter two lines on this window, one for the 40°C
measurement and one for the 100°C measurement.
• Ensure you enter the correct viscosities in the correct columns. The column labeled simply
viscosity is the viscosity of the oil at the temp in the next column. The column labeled ‘Visc @
40C’ is always the viscosity of the oil at 40°C, even if the temperature in the previous is 40.
• Select the correct location for the oil.
• Click ok to save your changes.
• Select Calibration Curve from the setup menu.
• Set the minimum number of samples to 10 and the maximum RSD to 1000. This will cause the
system to run 10 samples and stop when calibrating.
Instrument Testing
• Perform or have available the results of a repeatability test.
• If all percentages are below 1.00%, the instrument is functioning correctly, otherwise continue.
• For each timing that is not below 1.00%, move the outlier farthest from the mean to the side.
Continue until the rsd is below 1.00%.
• Consult the chart of examples below to decide what action to take.
Action: Action:
Single high outlier timing 1, Single high outlier timing 2,
adjust optic 1 by +1. adjust optic 3 by +1.
Action: Action:
Increasingly lower timings, Single low outlier timing 2,
check for solvent in inject port. adjust optic 4 by +1.
Action: Action:
First timing good, second timing Single low outlier timing 1,
noisy, backflush and remove adjust optic 2 by + 1. Two high
any stuck particles, adjust the outliers timing 2, adjust optic 3
tube height. by +1.
Action: Action:
Random low measurements in Random high and low
both timings simultaneously, measurements, check for good
check for good drying, low oil flow and correct oil volume.
vacuum, solvent in inject block.
• Take the suggested action and then perform the repeatability test again.
Calibration
• Go to the calibration oils window.
• Set the minimum number of samples to 3 and the maximum RSD to 1.2. This will cause the
system to run 3 samples and stop when calibrating if the repeatability is good.
• Go to the manual control window.
• Quick clean all tubes.
• Flush the probe.
• When all tubes have returned to idle, select calibrated all. Select all oils if no oils share locations,
otherwise select the ones placed on the deck.
• The instrument will now run at least three samples of each oil and generate a calibration curve.
• If you need multiple batches due to oils that share locations, run the additional oils after the system
finishes the first set in the same manner.
• Turn the pot on the cpu labeled vac set (just above the vacuum sensor) down slowly until the
software displays ‘Low Vacuum’ in red on the status bar.
• Turn the pot 1/8 of a turn back up. The software should indicate ‘Vacuum Ok’.
• Disconnect the eight inject lines from the autosampler to the tops of the inject valves.
Draining Solvent
• Remove the solvent feed lines from the solvent container. Place into a pail on the floor to catch
outflow.
• Quick clean all tubes twice to drain solvent from the solvent feed lines.
• Flush the probe tip four times to remove solvent from the autosampler pump and probe line.
• Dry all tubes 4 times to remove all liquids from the internal vacuum lines.
Vacuum System
Startup
• Check trap level; follow the instructions on the system to empty if required.
• Check autosampler waste container and empty if required.
• Check that all valves are closed.
• Turn on vacuum pump using switch on bottom of pump (located on the right of the system).
• Open both blue valves (labeled tank and instrument vacuum).
• Wait approximately one minute for vacuum level to build.
Shutdown
• Wait for all measurements and cleaning to finish and tubes to read 'Idle'.
• Turn off vacuum pump.
• Close both blue valves.
• Slowly open the AUX Vacuum valve and wait until the tank pressure reads zero.
• Slowly open the tank drain valve to let the oil flow back into the trap tank.
• The liquid level in the trap tank can now be seen on in drain line.
Draining
• Follow the shut down instructions above.
• Follow the drain instructions on the vacuum system (upper left).
Maintenance
Recommended Schedule
Maintenance Frequency
Glass Cleaning Once a week, or every 1000 samples.
Clean Lower Valves Once every six weeks, or every 7500 samples.
Back Flushing As required
Drive Lubrication Once every 3 months, or every 25000 samples.
Tubes
Glass Cleaning
• Wait until all tubes are Idle.
• Turn down the 100°C temperature to 90°C, by pressing the down arrow button until 90 is
displayed.
• Fill container with a 50/50 mixture of FL70 (available from Fisher) glass cleaner and distilled
water, place on designated location (marked on the autosampler deck).
• Fill another container with distilled water and place on designated location.
• Fill the IPA container on the autosampler deck.
• Wait until temperature has reached 91°C or lower.
• Select the 'Clean Glass' command from the tube menu to display the glass cleaning window:
• Select the tubes to be cleaned by clicking on them, or press 'Select All' to quickly select all tubes.
• Press the Start button.
• Cleaning tubes takes around one to one and half hours.
• Close the window when cleaning is complete, all tubes will be cleaned with solvent automatically.
• Return the 100°C side to 100 degrees by pressing the up arrow until 100 is displayed.
• Wait until the 100°C side is stable before running samples.
Back Flushing
•Quick clean the tube.
•Take the tube offline by pressing the disable button in the manual control window.
•Prepare by filling a 50ml beaker with solvent, use appropriate personal protection.
•Locate the solenoid valve for the tube. Loosen the two wing nuts holding the solenoid valve to the
black bracket and carefully slide the valve down so the black rubber tube slide off the glass capillary
tube.
•Unscrew the inject line from the top of the inject valve for the tube that is blocked. The inject line
goes from the autosampler to the instrument.
•Place the beaker under the glass tube at the bottom of the instrument and allow the vacuum to draw up
multiple short shots of solvent from the beaker.
•To return to normal operation, reassemble the instrument by reversing the above steps.
•Run a quick-clean cycle on the tube before testing to see it the blockage has been removed.
• If this will not dislodge the object, some pieces of fine wire are included with the instrument for
this purpose. See the section below.
• Note: there is a thicker wire for the larger tubes and a thin wire for the narrow ones. Ensure there
are no kinks in the wire that could cause it to bind in the capillary tube.
•Follow the steps above for back flushing up to and including connecting the auxiliary vacuum line to
the adapter tube.
•Very carefully slide the wire up until you can see it inside the glass, above the timing sensors (never
allow the wire to reach the bends in the glass as this will scratch the tube). NEVER force the wire as
this can cause the tube to break. If you reach an obstruction that you cannot simply push through,
apply gentle pressure and rotate the wire, this will slowly dig through the object.
•Slowly remove the wire, which will have dislodged the debris.
•Place the beaker under the glass tube at the bottom of the instrument and allow the vacuum to draw up
multiple short shots of solvent from the beaker.
•To return to normal operation, reassemble the instrument by reversing the above steps.
•Run a quick-clean cycle on the tube before testing to see it the blockage has been removed.
Lower Valves
Cleaning
• Note: there may be some solvent and/or oil present in the valves. Use appropriate personal
protection.
• Remove the lower valve assembly.
• Unscrew the two armature screws holding the retaining plate on for the upper or lower solenoid.
• Carefully remove the armature by wiggling it gently back and forth. There is an o-ring seal at the
bottom of the armature that can be easily damaged.
• Remove the poppit from the armature.
• The internals of the valve block should be cleaned with Q-tips and solvent.
• Clean the armature and poppit with Q-tips and solvent.
• After assembly adjust the orientation of the vent tube if required by loosening the two armature
plate screws and attaching the vent tube. Rotate the vent so that it points up and slightly to the left
when tightened by pushing on the tube about 1 to 1.5 inches from the fitting. Tighten the screws
when finished.
Operational Testing
• Ensure that the tube is idle and the vacuum system is on.
• Lower the valve but do not unscrew the connecting tube or vent tube.
• Go to the diagnostics window from the setup menu.
• Select the valve control tab.
• Click on the tube button to turn on the tube valve (the upper of the two).
• Verify that you can hear or feel the vacuum on the connecting tube.
• Close the tube valve and open the vent valve.
• Verify that you can hear or feel the vacuum on the connecting tube and the vent tube.
• Close the vent valve.
• Place a vacuum gauge on the vent tube and block the connecting tube with a rubber stopper.
• Wait 1 minute, the gauge should not move, if the gauge moves, the valve is leaking vacuum and
will cause measurement problems. Clean the valve and or replace or rebuild if required.
• Open the tube valve.
• Remove and replace the rubber stopper.
• The gauge should take longer then 15 seconds to reach 10 inHg. If the valve fails this test, the
instrument may have trouble cleaning out thick oils. Clean the valve and or replace or rebuild if
required.
• Close the tube valve.
• Return the valve to the raised position.
• Carefully remove the armature by wiggling it gently back and forth. There is an o-ring seal at the
bottom of the armature that can be easily damaged.
• At this point the internals of the valve block should be cleaned with Q-tips and solvent.
• Clean the armature and poppit.
Operational Testing
• Remove the solvent feed lines from the solvent container. Place into a pail on the floor to catch
outflow.
• Ensure that the tube is idle and the vacuum system is on.
• Go to the diagnostics window from the setup menu.
• Select the valve control tab.
• Click on the Solv button to turn on the solvent valve (the lower of the two).
• Verify that you can hear the valve click as is opens
• Close the solvent valve and open the top vent valve.
• Verify that you can hear the valve click as is opens
• Close the vent valve.
• Exit the diagnostics window.
• Return the lines to the solvent container.
• Quick clean the tube.
Autosampler
Drive Lubrication
• Periodically you need to lubricate the drive screw.
• Using the syringe of silicon grease supplied, apply a thin bead of grease to the entire length of the
driver screw (accessible through the from of the pump drive).
• Normal usage will blend the new grease in.
Troubleshooting
This section will describe common problems encountered, and techniques for troubleshooting them. It is
highly recommended that the customer routinely run an oil of known viscosity to check the system is
working correctly. Diagnosing problems is easier and problems are discovered sooner, which makes the
maintenance or repair also easier. A calibration oil is not needed for these QC samples, a container of bulk
oil that has been checked using another instrument or averaged over all tubes bracketed by calibration
checks will save costs. The recommended frequency is before and after each tray (if trays are run back to
back, you can start the first tray with a manual QC check and have the software include the end of tray
check to eliminate back to back QC samples). If QC checks are not being routinely run, you may need to
run three samples of a known viscosity to get the data needed for troubleshooting.
A list of symptoms and suggested actions are provided below. Find the symptom description that best
matches the instruments symptoms and follow the recommend actions until the problem is resolved.
quick calibration.
5. Perform the instrument test procedure on the
tube. If repeatability is good, recalibrate or
quick calibrate the tube.
QC fails high on both 1. Make sure the injection and the volume look Refer to maintenance schedule for
timings good for the injected sample. recommended intervals of running the
2. Back-flush the tube. Glass Clean
3. Glass clean.
4. Make sure the timings are consistent (over a
QC run of 3 or more) and that the value
which is ‘Outside Range’ is relatively close to
the expected QC value. If so, perform a
quick calibration.
5. Perform the instrument test procedure on the
tube. If repeatability is good, recalibrate or
quick calibrate the tube.
Tube continually fails 1. Do a clean on the tube (not a quick clean).
RSD 2. Run at least 3 QC injections on the tube.
3. Make sure the injection and the volume look
good for the injected sample.
4. Proceed to diagnose the problem based on the
timings from step 2.
5. Back-flush the tube.
6. Glass clean.
7. Perform the instrument test procedure on the
tube. If repeatability is good, recalibrate or
quick calibrate the tube.
Timings for injections 1. Check bath temperature with ASTM This will happen if the bath
on all tubes are thermometer. temperature is significantly lower
extremely high and are 2. Correct if possible with the temperature than the set point
failing rsd , outside controller’s calibration.
range, or > CAL 3. Check to see if stirrer motor is operating and Replacing the RTD sensor and/or
the bath liquid is indeed moving. motor assembly should be a last
(Or all 4 Tubes in the 4. Verify temperature controller settings. resort if the other steps have failed
same bath) 5. Verify temperature controller wiring (make
sure the connections are secure).
6. Replace RTD sensor.
7. Replace the motor assembly.
The sample/QC injection 1. Do a clean on the tube (not a quick clean). It is important to adhere to the
is stuck in the tube and 2. Clean out the lower valve body. maintenance schedule for the lower
does not flow down 3. Back-flush the tube. valve bodies
4. Clean out the upper valve body.
Oil is stuttering when 1. Do a clean on the tube (not a quick clean).
flowing down the tube 2. Make sure the injection and the volume look
good for the injected sample.
3. Back-flush the tube.
4. Glass clean.
5. Clean the lower valve body.
6. Clean the upper valve body.
Injection does not back 1. The top vent may be blocked. In #3, it is possible that the solvent
up in the tube at all, it 2. Test the top vent valves with the valve and/or top vent valves become
just flows straight down diagnostics in the software. physically inoperable if soap (i.e. the
the tube 3. Clean out top vent valve and poppets. glass clean procedure) is run at 100
4. Check for a lower valve vacuum leak. degrees Celsius. This is why it is
important to drop the bath
temperature to 90 degrees before
running the Glass Clean procedure.
Injection is being sucked 1. Check for vacuum leak on the lower valve Possible vacuum leak on the tube
down the tube – results body. poppet in the lower valve body
in RSD failure and/or ‘< 2. Clean lower valve body.
CAL’ 3. Replace poppets and/or valve body if It is important to adhere to the
necessary. maintenance schedule for the lower
valve bodies.
Low volume in the tube 1. Make sure QC oil/samples are filled to the
proper level.
2. Make sure probe line is properly flushed to
eliminate any air in the line.
3. Make sure the probe is making good contact
with the inject port – re-map inject block if
necessary.
4. Check probe line for leaks.
5. Check to see if there is air in the syringe
when aspirating a sample - Replace syringe
seal and/or syringe if necessary.
6. Check for blockages in the tube – piano-wire
and back-flush as necessary.
7. Clean out lower valve body.
8. Adjust volume offset.
Injection volume of QC 1. Adjust the volume offset. Adjust the volume offset if the
is too high volume is too high
Otherwise, if to low, refer to the ‘low
volume’ procedure above.
Tube stuck on ‘Waiting’ 1. Check to see if the oil has flowed down the For #1, if there hasn’t been enough
tube or not. volume injected to flow down the
2. Back-Flush. tube, the tube will be ’stuck’ on
3. Glass clean. waiting until it times out.
4. Adjust the optics intensities.
Optics adjustment are extremely rare
and should only be done as a last
resort
Tube stuck on ‘Optic 1’ 1. Back-Flush. Optics adjustment are extremely rare
2. Glass clean. and should only be done as a last
3. Adjust the optics intensities. resort
Tube stuck on ‘Timing 1. Back-Flush. Optics adjustment are extremely rare
1’ 2. Glass clean. and should only be done as a last
3. Adjust the optics intensities. resort
Tube stuck on ‘Optic 3’ 1. Back-Flush. Optics adjustment are extremely rare
2. Glass clean. and should only be done as a last
3. Adjust the optics intensities. resort
Tube is ‘< CAL’ with 1. Back-Flush. Optics adjustment are extremely rare
timings of 0 and/or 1 2. Glass clean. and should only be done as a last
3. Adjust the optics intensities. resort
Optic reads ‘ERROR’ 1. Follow the Optics Test Procedure.
Error message: Step-loss 1. Re-initialize autosampler and use the Occurs when probe tip had descended
on Z-axis diagnostic tools to check the alignment of the to somewhere it shouldn’t have and
inject block. hits something with force
2. Re-map inject block if necessary.
Heptane is coming out 1. Check the 100 degree bath temperature with In #3, it is possible that the solvent
of the top vent an ASTM thermometer and calibrate the and/or top vent valves become
controller if necessary. physically inoperable if soap (i.e. the
2. Check the boiling point of the heptane itself, glass clean procedure) is run at 100
this is to make sure the heptane is not degrees Celsius. This is why it is
contaminated with another solvent which important to drop the bath
could lower the boiling point significantly. temperature to 90 degrees before
3. Check and make sure the solvent and top vent running the Glass Clean procedure.
valves are operating properly (electrically
using the diagnostic valve controls).
4. Clean upper valves.
Software is frozen 1. Exit the Software – (by force if required).
2. Turn off instrument.
3. Turn off Autosampler.
4. Wait 5 minutes.
5. Re-start in reverse order.
Watchdog on instrument 1. Exit the Software – (by force if required). The watchdog indicator light is
is lit up 2. Turn off instrument. located under the case top front panel
3. Turn off Autosampler. on the cpu board near the vacuum
4. Wait 5 minutes. sensor.
5. Re-start in reverse order.
Temperature reading is 1. Check to see if stirrer motor is operating and Replacing the RTD sensor and/or
erratic the bath liquid is indeed moving. motor assembly should be a last
2. Verify temperature controller settings. resort if the other steps have failed
3. Try adjusting the temperature controller’s
calibration to match that of an ASTM
thermometer.
4. Verify temperature controller wiring (make
sure the connections are secure).
5. Replace RTD sensor.
6. Replace the motor assembly.
Temperature controller 1. Verify temperature controller settings. Replacing the RTD sensor and/or
is outputting a error 2. Check to see if stirrer motor is operating and motor assembly should be a last
message the bath liquid is indeed moving. resort if the other steps have failed
3. Verify temperature controller wiring (make
sure the connections are secure).
4. Try adjusting the temperature controller’s
calibration to match that of an ASTM
thermometer.
5. Replace the RTD sensor.
6. Replace the motor assembly.
Temperature controller 1. Adjust the over temperature cutout reset.
is shut off
Temperature cutout reset 1. Check the bath temperature with a
keeps triggering thermometer and calibrate the controller
accordingly.
2. Make sure the bath oil is moving and is at a
proper level.
3. Adjust temperature cutout.