UGEO WS80A

Data Sheet
Version 1.00.00

September, 2013

Rev. 1.2

Samsung Medison Bldg., 42, Teheran-ro 108-gil, Gangnam-gu, Seoul 135-851, Korea

TEL:(82)2-2194-1400 FAX:(82)2-556-9537, www.samsungmedison.com

Strategy Marketing Team, Samsung Medison


www.samsungmedison.com
CONFIDENTIAL

Table of Contents

1 GENERAL SPECIFICATION -------------------------------------------------------------- 1


1.1 Physical Specification ------------------------------------------------------------------ 1

1.2 Electrical Power ------------------------------------------------------------------ 1

1.3 Console Design ------------------------------------------------------------------ 1

1.4 Control Panel ------------------------------------------------------------------ 1

1.5 Monitor ------------------------------------------------------------------ 1

1.6 Touch Screen ------------------------------------------------------------------ 2

2 System Specification -------------------------------------------------------------- 2


2.1 Application ------------------------------------------------------------------ 2

2.2 Transducer Types ------------------------------------------------------------------ 2

2.3 Operation Modes ------------------------------------------------------------------ 2

2.4 Display Modes ------------------------------------------------------------------ 2

2.5 System Standard Features ------------------------------------------------------------------ 3

2.6 System Options ------------------------------------------------------------------ 4

2.7 Peripheral Options ------------------------------------------------------------------ 4

2.8 Documentation Capability ------------------------------------------------------------------ 4

2.9 Peripheral Signals ------------------------------------------------------------------ 4

2.10 Operating Environment ------------------------------------------------------------------ 5

2.11 Acoustic Output Management ------------------------------------------------------------------ 5

2.12 Language ------------------------------------------------------------------ 5

2.13 Scan Mode ------------------------------------------------------------------ 5

2.14 Image Parameters ------------------------------------------------------------------ 6

3 Processing -------------------------------------------------------------- 6
3.1 Pre-Processing ------------------------------------------------------------------ 6

3.2 Post-Processing ------------------------------------------------------------------ 6

CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
www.samsungmedison.com
CONFIDENTIAL

3.3 Image Processing ------------------------------------------------------------------ 6

3.4 Image Archive ------------------------------------------------------------------ 7

3.5 Cine Memory/Image Memory ------------------------------------------------------------------ 7

4 Connectivity -------------------------------------------------------------- 7
4.1 DICOM(Optional) ------------------------------------------------------------------ 7

5 Scanning Parameters -------------------------------------------------------------- 8


5.1 2D Mode ------------------------------------------------------------------ 8

5.2 M Mode ------------------------------------------------------------------ 8

5.3 Color Doppler Mode ------------------------------------------------------------------ 9

5.4 Power Doppler Mode ------------------------------------------------------------------ 10

5.5 PWD Mode ------------------------------------------------------------------ 11

5.6 Volume Mode ------------------------------------------------------------------ 12

5.7 Calipers & General Measurements ------------------------------------------------------------------ 13

5.8 Doppler General Measurements ------------------------------------------------------------------ 13

6 Probes -------------------------------------------------------------- 14
6.1 Linear Array ------------------------------------------------------------------ 14

6.2 Convex Array ------------------------------------------------------------------ 14

6.3 Endocavity Array ------------------------------------------------------------------ 14

6.4 Volume Probe ------------------------------------------------------------------ 15

7 Measurement -------------------------------------------------------------- 15
7.1 OB Measurement ------------------------------------------------------------------ 15

7.2 OB Table Reference ------------------------------------------------------------------ 16

7.2 Gynecology Measurement ------------------------------------------------------------------ 17

7.3 Cardiac Measurement ------------------------------------------------------------------ 17

7.4 Carotid Measurement ------------------------------------------------------------------ 18

7.6 LE Artery Measurement ------------------------------------------------------------------ 18

7.7 UE Artery Measurement ------------------------------------------------------------------ 18

7.8 LE Vein Measurement ------------------------------------------------------------------ 18

7.9 UE Vein Measurement ------------------------------------------------------------------ 19

7.10 Fetal Echo Measurement ------------------------------------------------------------------ 19

CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
www.samsungmedison.com
CONFIDENTIAL

7.11 Urology Measurement ------------------------------------------------------------------ 19

7.12 Abdomen Measurement ------------------------------------------------------------------ 20

7.13 Thyroid Measurement ------------------------------------------------------------------ 20

7.14 Breast Measurement ------------------------------------------------------------------ 20

7.15 Testicle Measurement ------------------------------------------------------------------ 20

7.16 Superficial Measurement ------------------------------------------------------------------ 20

7.17 MSK Measurement ------------------------------------------------------------------ 20

7.18 Pediatric Hip Measurement ------------------------------------------------------------------ 21

8 Safety / EMC -------------------------------------------------------------- 21


8.1 Classifications ------------------------------------------------------------------ 21

CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
GENERAL SPECIFICATION  Integrated cable management

 Integrated high-fidelity stereo speakers


PHYSICAL SPECIFICATION
 Prevention of noise of the system

 Height: 1,350 ~ 1,800 mm  Easily removable air filter

 Width: 560 mm  Integrated SSD 512GB

 Depth: 760 ~ 860 mm  Integrated DVD Multi Recordable Driver

 Weight with monitor: approx. 108 kg (238 lb.)  EMC Class A

 Gel Warmer
ELECTRICAL POWER
CONTROL PANEL
 Voltage: 100- 240 V AC

 Frequency: 50/60 Hz • Height: adjustable 185 mm

 Power Consumption • Rotation: adjustable +/- 30° from center

714.7 VA with Peripherals • Ergonomic hard key operations

564.7 VA without Peripherals • Hard key layout for OB GYN and General Imaging

 Nominal Current: • Functional grouping of keys

110V: 8.2A • Interactive back-lighting

220V: 4.1A • Functional grouping of keys on 10.1” Wide LED

 Heat Dissipation: 3207.8BTU/h Touch Panel (Resolution 1280 x 800 x 24 bits)

 Safety Class • Integrated alphanumeric QWERTY backlit keyboard

Ultrasound system: A and Virtual Screen alphanumeric QWERTY

Transducer: Depend on transducer specification keyboard on Touch Panel

• Integrated recording keys for remote control of


CONSOLE DESIGN
peripherals or DICOM devices

 4 active Probe Ports • Trackball


 Height/Rotate adjustable control panel • 8 TGC Slide (Touch Slide Controls)
 Height/Tilt/Rotate adjustable monitor
MONITOR
 On-board storage for peripherals

 6 probe holders • 21.5 inch Wide Color LED Flat Display

 4 Swivel/Lock wheels (wheel diameter: 125 mm) • Resolution: 1920 x 1080 x 24 bits

 Front and rear handles • High-Resolution monitor Non Interlace Scan

 Rear storage cover for peripheral connectors and • Height: adjustable ~ 270 mm

cables • Forth/Back: adjustable ~ 160 mm

UGEO WS80A Datasheet V1.0_r1.2 Page 1


• Tilt: adjustable +20° / - 15° from center C2-6, SC1-6

• Rotate: adjustable +/- 40° from center,  Volume Probes

others +/- 50° from center High brightness & Convex Volume

contrast V4-8

• Digital brightness & contrast adjustment Endocavity Volume

V5-9
TOUCH SCREEN
 Linear Array
• 10.1” high resolution LED Monitor L5-13, L3-12A
• Display Size: 1280 x 800  Endocavity
• Capacitive touch type E3-12A(To Be Updated), VR5-9
• Virtual Alpha KBD
OPERATION MODES
• Interactive dynamic soft key menu display

• 2D-Mode

SYSTEM SPECIFICATION • M-Mode

• Color Doppler Imaging


APPLICATIONS
• Power Doppler Imaging

 General • S-Flow™ Mode

 Abdominal • Pulsed Wave (PW) Spectral Doppler

 Gynecology • Directional Power Doppler Imaging (DPDI)

 Obstetrical • Color M-Mode

 OB Early • Harmonic Imaging

 Renal • Free Angle M-Mode (Arbitrary M mode)

 Fetal Heart • Single/Dual/Quad mode

 Urology • Volume-mode(3D/4D)

 Vascular
DISPLAY MODES
 Small Parts

 Muskuloskeletal
• Simultaneous Capability:
 Breast
B+M, B+PW, B+C, B+PD, B+DPD, B+SFlow,
 Pediatric
B+SFlow+PW, B+PPI, B B+C+PW, B+PD+PW,

TRANSDUCER TYPES B+DPD+PW, B+PPI+PW, B+C+M, Dual B, Dual B+C,

Dual B+PD, B+E


 Convex Array
UGEO WS80A Datasheet V1.0_r1.2 Page 2
• •
TM
Selectable Alternative Modes: Volume NT


TM
B/M, B/PW, B+C/PW, B+PD/PW, B+DPD/PW, XI STIC


TM TM
B+C/M VOCAL , XI VOCAL

• Colorized Image: • 3D/4D


TM
Colorized B, Colorized M, Colorized D, Colorized 3D Multi Slice View

• •
TM
Time Line Display: Oblique View


TM
Format: Top/Bottom or Side/Side selectable Volume CT


TM TM
Timeline Method: Scrolling or Moving Bar OVIX , Multi-OVIX

• •
TM
Multi Image Split Screen: Volume Slice View


TM
Live and/or Frozen, B+B/C or PD Mirror View
Dual (B, B/C, B/PD, B/TD, B/PPI, B/PD) • Inversion 3D


TM
Quad (B, B/C, B/PD, B/TD, B/PPI, B/PD) FAD


TM
• Zoom: Write / Read (Magnification) / Pan SFVI


TM
VSI
SYSTEM STANDARD FEATURES •
TM
Smooth cut

• MagiCut
• Full Spectrum Imaging (3 bands)
• Raw Data Analysis
• Tissue Harmonic Imaging
• Post-image optimization
• Pulse Inversion Harmonic Imaging
• Patient Information Database
• Power Pulse Inversion Harmonic Imaging
• Image Archive integrated on CD/DVD and HDD
• Trapezoidal Imaging
drive
• Quick Scan™ (Automatic Optimization)
• Support for external USB 2.0 HDD drive
• SDMR™evo
• Cine for 12,700(Max Condition) frames and Loop

TM
Spatial Compounding Imaging
Review for 8,192 lines
• Elastoscan
• Auto Calc (Real-Time Automatic Doppler
L5-13: Breast, Thyroid
Calculation)
L3-12A: Breast, Thyroid
• Doppler Auto Trace
E3-12A(To Be Updated)
• User Configurable Measurement Menu
: Gynecology, Adnexa, Prostate
• Customizable Measurement Menu
VR5-9: Gynecology, Adnexa, Prostate
• Customizable Body Maker

TM
Panoramic Imaging
• Customizable User Keys

TM
AutoIMT+
• Post-Measurement

UGEO WS80A Datasheet V1.0_r1.2 Page 3


• On-board electronic documentation

• Fully Digital Real Time Recording


PERIPHERAL OPTIONS

TM
EZ exam

• Screen Keyboard • Integrated options for Digital B/W Printer, Digital

• MPI Color Printer, Digital Color Report Printer

• 5D Cine • External USB Printer Connections

• DVI-I Output Available for compatible devices

• Digital ECG
SYSTEM OPTIONS
• Foot switch (USB type)
 4D

 3D XI™
DOCUMENTATION CAPABILITY
 3D MXI™


TM
 Convex SCI* ADVR

*Countries which buy convex SCI as option only • On-board printing device control

applied. • Selective printing on two connected printers

 Cardiac Measurement • SonoView II (Image Filing Package)

 DICOM • Export Media: CD/DVD+R/-R/RW, USB Flash,

 Spatial Compound Imaging™ (SCI™) External USB HDD

 XI STIC • Export Format: JPEG,BMP,TIFF,DICOM

 Auto IMT+™ • Print Function

 ElastoScan • Patient list and data search

 Panoramic Imaging • Report save available


TM
 HDVI Post image processing available

 Volume NT
TM
• Caliper measurement available

 ADVR
TM
(Agile DVR) • DICOM 3.0
Service: Storage/Printer/Worklist/PPS/SC
 Mobile Export
DICOM SR Structured Reports
 ElastoScan (ECI)

 FRV  Mobile Export

 5D NT

 5D LB PERIPHERAL SIGNALS
 2D NT
 Audio port (I/O)
 e-DEB
 LAN (I/O) : Ethernet, 10/100 BASE-T
UGEO WS80A Datasheet V1.0_r1.2 Page 4
 B/W Printer port (USB type) (O)  Thermal Index (TI): 0.1~5.0 Range

 Patient Monitor TIC – Thermal Index, Bone at Surface

VHS TIB – Thermal Index, Bone at Focus

SVHS TIS – Thermal Index, Soft Tissue

d-Sub (1280 x 1024)

DVI-D (1920 x 1080)


LANGUAGE
DVI-I (1280 x 1024)

 Patient Monitor Power (O)  Display Language

 DVI port (O) for digital signals to the monitor English

 ECG AUX input (optional) German

 Foot switch port (USB type) French

 USB port (I/O, USB 2.0) : 8 ports(front2 , rear4, USB Spanish

Printer 2 ) Italian

 Microphone Russian

Chinese
OPERATING ENVIRONMENT
 Input Language

 Ambient temperature: 10°C–35°C (50°F–104°F) English

 Relative humidity: Up to 75% non-condensing German

 Pressure: 700~1060hPa French

 Audible noise: 34dB Spanish

Italian

Russian
ACOUSTIC OUTPUT MANAGEMENT
Korean

 User selectable, transducer and scanning mode Chinese

dependent Japanese

 Dedicated Output Display on the system monitor

display of output acoustic power level, as well as


SCAN MODE
thermal and mechanical indices:

 PWR – Output Power level. Range: From 10 % of  Simultaneous Mode : 3 Type off, Allow B/PWD,
maximum output, output level is increased by 5% Allow B/C/PWD
in each step.  Auto Unfreeze While Dual Operation
 Mechanical Index (MI): 0.1~3.0 Range  Change Window at Dual Mode

UGEO WS80A Datasheet V1.0_r1.2 Page 5


 2D/C Live Mode PROCESSING
Left/ Up, Right/ Down

2D/C Live left-Right Dual Only PRE-PROCESSING

 Freeze Action : 4 Type


• 2D / M-Mode
None
Gain, TGC, Dynamic Range, Transmission Focus
Body Marker
Position, Transmission Focus Number, Transmission
Caliper
Frequency, Edge Enhancement, Sweep Speed for
Measure
M-Mode
 End Exam Action : End exam Only, End exam and
• Color-Mode
Patient
Gain, Velocity Range, Wall Filter, Ensemble,

Spatial Filter, Frame Averaging, Baseline Shift,


IMAGE PARAMETERS
Smoothing Filter
• 2D Gain: 0 ~ 100 • D-Mode
• Power: 10 ~ 100 Gain, Dynamic Range, Transmission Frequency,
• Edge Enhance: -3 ~ 3 Velocity Scale/PRF, Wall Filter, Baseline Shift,
• Frame Average: 0 ~ 15 Sweep Speed
• 2D Filter

• View area Control

• View area Steering POST-PROCESSING

• Dynamic Range: 50~200 dB, User selectable in 1 dB


 2D Quick Scan (Automatic Tissue Optimization)
increments
 Doppler Quick Scan (Automatic Spectrum
• Reject Level: 1 ~ 32
Optimization)
• Gray map: 13 types
 Gray Maps (2D Mode: 13 types / M-Mode: 13 types)
• Chroma map: 13 types + 3 User types
 Chroma map (2D Mode : 16 types / M Mode : 16
• Trapezoidal Image
types)
• U/D Flip, L/R Flip
 Reject Function from 1 to 32 in 32 digit steps
• Frame Rate: Slow / Normal / Fast

• Max. Frame Rate: 1177Hz


(L5-13, Triple-beam, Depth 2cm) IMAGE PROCESSING
• Frequency optimized setting: Pen(penetration) /
Gen(general) / Res(resolution) • Digital Beamformer

• Imaging Depth: 2 ~ 30 cm

UGEO WS80A Datasheet V1.0_r1.2 Page 6


• Dynamic Receive Focus • After freezing immediate scrolling through Cine

• Dynamic Receive Aperture memory with the Track ball

• Tx Apodization • CINE memory: Max. 12700 frames

• Tx Pulse Shaping • Loop memory: Max. 8192 lines

• Adjustable Dynamic Range

• Adjustable Field of View


CONNECTIVITY
• Image Reverse: Left / Right, Up/Down

• Transmission Focus: DICOM (OPTIONAL)


1 ~ 8 Focal Points Selectable; Focus Position: 8
 Ethernet Network Connection
steps (probe independent)
 Archiving Format: DICOM
• 256 shades of gray, 8 bits
 DICOM with ultrasound raw data
• 16,777,216 colors, 8 bits for each RGB component
 Capture Area: pre-settable for each print key

 Video Area

IMAGE ARCHIVE  Application Window

Whole Screen
• Archiving Format: DICOM  Configurable Examination List Window, Patient
• Possibly exported to other format Information Window, and Search/Create Patient
• Archive Devices: Window
SSD Image Storage
 Extended search dialog, auto search for patient in
CD -RW Storage, DVD-RW Storage
Search/Create Pts window
USB Memory Stick
 Automatic generation of patient ID

 Request acknowledge of End Exam action

 Go directly screen from search


CINE MEMORY/IMAGE MEMORY
 Verify DICOM directory on removable media

• CINE Gauge and Cine Image number display  Format removable media (rewritable DVD)

• CINE Review Loop  The currently selected dataflow

• Selectable CINE Sequence for CINE Review (by Start  All configured data flows

Frame and End Frame)  The network structure tree

• Measurements/Calculations & Annotations on CINE  The configured buttons data flows

• Available in all modes(include loop)  AVI and JPEG Export : DICOM Support

• Imaging Cine, for real-time acquisition and review  DICOM Storage Commitment Push Model User

of 2-D  DICOM Modality Performed Procedure Step User


UGEO WS80A Datasheet V1.0_r1.2 Page 7
 DICOM Printer User  SDMR™evo 5 index

 DICOM Store User  Dual Live

 DICOM Modality Worklist User  Spatial Compound (High, Medium, low, Off)

 DICOM Structured Report Cardiology, Vascular and 4 step

OB/GYN User  Gray Map 1 ~ 13 step

 View Angle -20° ~ 20°


SCANNING PARAMETERS
 2D Image Size 50~100

2D MODE  Full Spectrum Image 1 ~ 3 step

 Store Method(Time, Manual) 2 Mode


 Frame Average 0 ~ 15 step
Store method time mode only 1~50 sec
 Dynamic Range 50 ~ 200 dB
Store method ECG Beat mode only 1~8 beat
 Edge Enhance -3 ~ +3 step
 Zoom 120 ~ 400 %
 View area 40 ~ 100 %
 Chroma Map
 Density(High, Middle, Low) 3 step
Type 1 ~ 13 step
 Tissue(1440~1620m/s) 10 step
User 1 ~ 3 step
 Trapezoidal (Linear)
 Elastoscan
 Apex (U/D Flip)
 Panoramic
 Direction (L/T Flip)
 Body marker
 Frame Rate(Slow, Normal, Fast) 3 step
 Quick-Scan
 Post Curve 1 ~ 13 step
 Moving Line (M Line)
 Harmonic
 Moving Line position(Up, Down, Left, Right)
 Pulse Inversion (Probe dependency)
 ECG (Probe dependency)
 Time Gain Control 8 slide
 Gamma(Hard, Medium, Weak, Off) 4 mode
 Gain 0 ~ 100 %

 Power 10 ~ 100 %

 Depth (Probe dependency) 2 ~ 30 cm M MODE


 Focal point 8
 M Line
 Annotation
 Body marker
 Single/Dual/Quad Control
 Annotation
 Top-Bottom Dual
 Single/Dual/Quad Control
 Frequency(Pen, Gen, Res) 3 mode
 Top-Bottom Dual(Dual mode only)
 SRF 3 index
 Dynamic range 50 ~ 200 dB
UGEO WS80A Datasheet V1.0_r1.2 Page 8
 Reject level 1 ~ 32 step  Frame Average 0 ~ 15 step

 Density(High, Middle, Low) 3 step  View Angle -20° ~ 20°

 Edge enhance -3 ~ 3 step  Scan Area 40 ~ 100

 Tissue(1440~1620m/s) 10 step  Full Spectrum Imaging 1 ~ 3 step

 Chroma map  Zoom 120 ~ 400%

Type 1 ~ 13 step  Quick Scan

User 1 ~ 3 step  ECG (Probe dependency)

 Gain control 0 ~ 100 %

 Time gain compensation 8 slice

 Depth (Probe dependency) 2 ~ 30 cm


COLOR DOPPLER MODE

 Frequency(Pen, Gen, Res) 3 step


 Color scale 256 colors
 Sweep speed 60, 120, 180, 240, 300, 360 Hz
 Frame average 0 ~ 15 step
 Post Curve
 Balance 1 ~ 16 Level
Type 1 ~9 step
 PRF 0.1 ~ 18.4KHz
User 1 ~ 2 step
 Velocity
 Frame Rate(Slow, Normal, Fast) 3 step
Minimum 2.25cm/sec
 Gamma(Hard, Medium, Weak, Off) 4 mode
Maximum 5.6 m/sec
 Display mode
 Wall filter 0 ~ 3 level
Top/Bottom (30/70, 40/60, 50/50, 60/40, 70/30)
 Color gain 0 ~ 100%
Left/Right
 Frame rate 2 ~ 30HZ
 Power 10 ~ 100
 Read Zoom 120 ~400%
 Gray Map 1 ~ 13 step
 Dual mode(Top/Down, Left/Right)
 Store Method
 Quad mode
Time (Store method time mode only) 1 ~50 sec
 Frequency(Pen, Gen)
Manual
 Steer(Linear, convex)
 Harmonic
 Color Invert
 Trapezoidal (Linear probe only)
 Single/Dual/Quad control
 Dual Live
 M Line
 SRF 1 ~ 3 Step
 M line position (Up/Down/Left/Right)
 Elastoscan
 M line Angle 0°~360°
 Up/Down Flip
 Frequency (Pen, Gen, Res)
 Left/Right Flip
 Steer(Left, Right, None)
UGEO WS80A Datasheet V1.0_r1.2 Page 9
 Filter 0~3 POWER DOPPLER MODE
 Scale 0.20 ~ 12.00 KHz
 Color scale 256 colors
 Baseline 0 ~ 8 step
 PRF 0.1 ~ 18.4KHz
 Display mode (Color + BW, BW only)
 Minimum velocity 2.25cm/sec
 Color mode (Velocity, Power, Variance, Velocity +
 Maximum velocity 5.6m/s
Variance)
 Wall filter 0 ~ 3 Level
 Color map 1 ~ 15 type
 Color gain 0 ~ 100%
 Sensitivity 2 ~ 31 step
 Frame rate 2 ~ 30Hz
 Density(High, Middle1, Middle2, Low)
 Read zoom 120 ~ 400
 Power 10 ~100%
 Dual mode Top/Down, Left/Right
 Harmonic
 Quad mode
 Trapezoidal
 Frequency(Pen, Gen)
 SRF 1 ~ 3 Step
 Steer (Linear, Convex)
 Elastoscan
 Single/Dual/Quad Control
 Up/Down, Left/Right
 M Line
 Dynamic Range 50 ~ 200 dB
 M Line position (Up/Down/Left/Right)
 Reject level 1 ~ 32 step
 M Line Angle 0°~360°
 Gray Map 1 ~ 13 step
 Frequency (Pen, Gen, Res)
 View Angle -20°~20°
 Steer (Left, Right, None)
 Scan Area 40~100
 Filter 0 ~ 3 Step
 2D Image Size 50 ~ 100
 Scale 0.80 ~ 5.00 KHz
 Full Spectrum Image 1 ~ 3 step
 Display mode
 Edge Enhance -3 ~ 3 step
Top/Bottom 4:6, 6:4, 5:5
 Chroma Map
Side by Side
Type 1 ~ 13 step
 PD mode (DPDI mode/PD mode)
User 1 ~ 3 step
 Color map 1 ~ 8 type
 Tissue(1440 ~ 1620 m/s) 10 step
 Balance 1 ~ 16 level
 Gain 0 ~ 100
 Sensitivity 6 ~ 31 step
 Zoom 120 ~ 400
 Store Method(Time, Manual)
 Depth 2 ~ 30cm
 Density (High, Middle1, Middle2, Low)
 Time Gain Control 1 ~ 8 Slide
 Frame Average 0 ~ 15 step
 Quick Scan
 Power 10 ~ 100
UGEO WS80A Datasheet V1.0_r1.2 Page 10
 Harmonic  Wall filter 0 ~ 3 step

 Trapezoidal (Linear probe only)  Doppler average 0 ~ 6 level

 SRF 1 ~ 3 step  Minimum velocity 5.0cm/sec

 Elastoscan  Maximum velocity 3.05m/sec

 Up/Down, Left/Right  Display : Sweep, Top/Down

 Dynamic range 50~200dB  D line update rate (120, 180, 240, 300, 360,420Hz)

 Reject level 1 ~ 32 Step  D mode display size (30/70, 40/60, 50/50, 60/40,

 Gray map 1 ~ 13 Step 70/30)

 View angle -20° ~ 20°  D mode display format(Top/Bottom, Side by Side)

 Scan Area 40 ~ 100  Loop Size 8192 loop

 2D Image Size 50 ~ 100  Steer

 Full Spectrum Image 1 ~ 3 step  Auto calculation

 Edge Enhance -3 ~ 3  Multi frequency

 Chroma Map  Doppler Invert

Type 1 ~ 13 step  Simultaneous (B,C,D/B,D)

User 1 ~ 3 step  Annotation

 Tissue (1440 ~ 1620m/s) 10 step  Single/Dual/Quad Control

 Gain 0 ~ 100%  M Line

 Zoom 120 ~ 400  M line position

 Depth 2 ~ 30cm  M Line Angle 0°~360°

 Time Gain Control 1 ~ 8 slide  AutoCalc (PSV, DEV, TAP, RI, PI, S/D)

 Quick scan  AutoCalc direction (All, Up, Down)

 Mean Trace

 Sweep speed(60, 120, 180, 240, 300, 360, 420Hz)


PWD MODE  Frequency (Pen, Gen, Rea)

 Steer (Left, Right, None)


 PRF 1 ~ 23 KHz
 Filter 0 ~ 3 step
 Display height 9 step
 Scale 0.6 ~ 14KHz
(70/30, 65/35, 60/40, 55/45, 50/50, 45/55, 40/60,
 Sound 0 ~ 100
35/65, 30/70)
 Baseline 0 ~ 8 step
 Sample Volume 0.5 ~ 15.0mm
 Sample volume 0.5 ~ 10mm
 Wall filter 0 ~ 3 level
 Gray map 1 ~ 13 step
 Spectrum types(type1, type2 ,type3)
UGEO WS80A Datasheet V1.0_r1.2 Page 11
 Spectrum Enhance 1 ~ 4 step  Time gain control

 Display Mode  Quick scan

Top/Bottom 4:6, 6:4, 5:5  HPRF

Side by Side Type 1 ~ 13 step

 Spectrum Type (Type1 ~ Type3)

 Dynamic range 50 ~ 200dB

 Store Method (Time, Manual)


VOLUME MODE

 Chroma map
 Live 3D, Static 3D
Type 1 ~ 13 step
 MPR (Multi-Planar Rendering) mode
User 1 ~ 3 step
 2D mode
 Power 10 ~ 100
 Full 3D mode
 Harmonic
 2D/3D mode
 Trapezoidal (Linear probe only)
 ROI 3D mode
 Dual live
 3D Accept ROI mode
 SRF 1 ~ 3 TM
 3DXI (3D eXtended Imaging)
 Elastoscan TM
 MSV (Multi-Slice View) display
 Panoramic TM
 Oblique View
 Spatial Compound (High, Medium, Low, Off)
 Static Line Oblique view
 Up/Down, Left/Right flip
 Dynamic Line Oblique view
 Top-Bottom Dual(Dual mode only)
 Contour Oblique view
 Frame Average 0 ~ 15 step
 Multi Oblique View
 Reject level 1 ~ 32 step TM
 OVIX
 Gray map 1 ~ 13 step
(Oblique View eXtended Imaging)
 View angel -20°~20° TM
 Multi-OVIX
 Scan area 40 ~ 100 TM
 Volume CT (VCT)
 Full Spectrum image 1 ~ 3 step
 Cube Volume CT
 Line Density (High, Middle, Low)
 Cross Volume CT
 Edge Enhance -3 ~ 3 TM
 3D-MXI (3D Multi-eXtended Imaging)
 Tissue (1440 ~ 1620m/s) TM
 Volume Slice
 Gain 0 ~ 100 TM
 Mirror View
 Zoom 120 ~ 400
 3D OH
 Depth 2 ~ 30 cm
 ROI single control
UGEO WS80A Datasheet V1.0_r1.2 Page 12
TM
 VOCAL • Ellipse function: Up to 4 pairs of calipers
 SHELL Histogram Distance between calipers
TM
 XI VOCAL Ellipse circumference
TM
 XI STIC Ellipse area

 Gray & Color STIC • Trace function. Displays:


 STIC + MSV Trace circumference

 Inversion 3D Traced area


TM
 Face Auto Detection • Minimum distance between calipers:
TM
 Volume Shade Imaging Transducer type, depth and HRZ box setting
TM
 Smart Filter Volume Imaging dependent
TM
 Smooth Cut
TM
 Volume NT/IT
DOPPLER GENERAL MEASUREMENTS
 Optimal volume resolution: Low, Mid, High

 3D rendering mode: Surface, Surface Smooth, • Velocity

Maximum, Minimum, X-ray • Frequency

 SDMR™evo • Time

 HDVI (HD Volume Imaging)


TM
• Acceleration

 Volume Contrast Enhancement(VCE)


TM
• Acceleration Time

 3D Compound Imaging (3D CI) • Ratios

 SeeThru mode: gray + color Doppler mode • A/B Ratio (Velocities/Frequency Ratio)

 MagiCut Plus: Available in both 3D and 4D • Peak Systole/End Diastole (PS/ED Ratio)

 Volume CINE (128MB) • End Diastole/Peak Systole (ED/PS Ratio)

 5D Cine • Heart Rate

 5D NT • AutoCalc

 5D LB Auto Doppler Trace function with automatic

calculations

• TAMAX (Time Average Maximum Velocity)


CALIPERS & GENERAL MEASUREMENTS
• Volume Flow (TAMEAN and Vessel Area)

• 4 pairs of 2-D calipers available. Screen display: • PI (Pulsatility Index)


Distance between calipers for each pair • RI (Resistivity Index)
Manual tracing in 2D distance

UGEO WS80A Datasheet V1.0_r1.2 Page 13


PROBES  Application : Abdomen, OB, Gynecology

 Safety Class: BF
LINEAR ARRAY

L5-13 SC1-6

 Linear Array  Curved Linear Array

 Center frequency : 8.0MHz  Center frequency : 3.2MHz

 Radius of curvature : Flat  Radius of curvature : 60mm

 Field of view : 38.4mm  Field of view : 60.5°

 Number of elements : 192  Number of elements : 192

 Biopsy Guide : Available  Biopsy Guide : Available

 Application : Musculoskeletal, Small Parts, Vascular  Application : Abdomen, OB, Gynecology

 Safety Class: BF  Safety Class: BF

L3-12A
ENDOCAVITY ARRAY
 Linear Array

 Center frequency : 7.2MHz E3-12A (To Be Updated)

 Radius of curvature : Flat  Micro Curved Linear Array (Endo Cavity Type)

 Field of view : 50mm  Center frequency : 8.5

 Number of elements : 256  Radius of curvature : 9.4

 Biopsy Guide : Available  Field of view : 210°

 Application : Small Parts, Vascular, MSK  Number of elements : 192

 Safety Class: BF  Biopsy Guide : Available

 Application : OB, Gynecology, Urology

 Safety Class: BF
CONVEX ARRAY

C2-6

 Curved Linear Array VR5-9

 Center frequency : 4.0MHz  Micro Curved Linear Array (Endo Cavity Type)

 Radius of curvature : 60mm  Center frequency : 5.95MHz

 Field of view : 58.12°  Radius of curvature : 10mm

 Number of elements : 192  Field of view : 150.3°

 Biopsy Guide : Available  Number of elements : 192

UGEO WS80A Datasheet V1.0_r1.2 Page 14


 Biopsy Guide : Available  Fetal Echo Measurement

 Application : Gynecology, OB, Urology  LE Artery Measurement

 Safety Class: BF  UE Artery Measurement

 LE Vein Measurement
VOLUME PROBE  UE Vein Measurement

 Abdomen Measurement
V4-8
 Thyroid Measurement
 3D Curved Linear Array
 Testicle Measurement
 Center frequency : 4.4 MHz
 Superficial Measurement
 Radius of curvature : 39.1 mm
 Pediatric Hips Measurement
 Field of view : 85°
 Musculoskeletal Measurement
 Number of elements : 192

 Biopsy Guide : Available

 Application : Abdomen, OB, Gynecology


OB MEASUREMENT
 Safety Class: BF

 Fetal Biometry

V5-9  Fetal Long Bones

 3D Curved Linear Array (EC-straight type)  Fetal Cranium

 Center frequency : 6.5 MHz  Fetal Others

 Radius of curvature : 10.1 mm  AFI

 Field of view : 150°  Maternal others

 Number of elements : 192  CTAR

 Biopsy Guide : Available  PLI

 Application : OB, Gynecology, Urology  Umb. Artery

 Safety Class: BF  MCA

 Lt. Uterine Artery

 Rt. Uterine Artery


MEASUREMENT
 Plac. Artery

 OB Measurement  Lt. Fetal Carotids

 Gynecology Measurement  Rt. Fetal Carotids

 Carotid Measurement  Fetal Aorta

 Cardiac Measurement  Ductus Venosus

 Urology Measurement  Rt Renal A


UGEO WS80A Datasheet V1.0_r1.2 Page 15
 Lt Renal A HUM

 Volume Flow ULNA

 Fetal HR TIB

 Fetal Description RAD

 Fetal Heart FIB

 Fetal Brain CLAV

 Fetal Abdomen Vertebral

 Biophysical Profile  Fetal Cranium

 Maternal Survey CEREB

 Comment OOD

IOD

CM
OB TABLE REFERENCE NF

NT
 Fetal Biometry
IT
GS
Lat Vent
CRL
NB
YS
HW
BPD
 Fetal Others
OFD
Foot
HC
Ear
APD
MP
TAD
Lt. Renal L
MAD
Rt. Renal L
AC
Lt. Renal AP
FTA
Rt. Renal AP
ThC
Pelvis
FL
EFW
SL
 Ratio Calculations
TTD
RL/BPD
APTD
FL/HC
APTD x TTD
FL/AC
 Fetal Long Bones
UGEO WS80A Datasheet V1.0_r1.2 Page 16
FL/Foot  Ectopic (Ectopic Pregnancy)

CI(BPD/OFD0  Abnormalities of Uterus

HC/AC  Environment

ThC/AC  Comment

Lat. Vent/HW

 Fetal Doppler trend graph


CARDIAC MEASUREMENTS

Mid Cerebral Artery


 Lt. Ventricle(2D)
Umbilical Artery
 Lt. Ventricle(M)
 Trend graph
 LV Vol.(Simpson)
AF Index
 LV Vol.(A/L)
HC/AC
 LV Vol.(Bullet)
FL/HC
 LV Mass
Lat.Vent/HW
 Rt. Ventricle(2D)

 Rt. Atrium
GYN MEASUREMENTS
 LVOT

 Uterus  RVOT

 Cyst.  Aortic Valve

 Rt. / Lt. Ovary  Right Ventricle(M)

 Rt. / Lt. Follicles  Mitral Valve(M)

 Mass 1  Mitral Valve

 Mass 2  Tricuspid Valve

 Mass 3  Pulmonic Valve

 Rt. /Lt. Ovarian A  Tei Index

 Rt. / Lt. Uterine A  Pulm. Veins

 Pericystic  Hepatic Veins

 Endometrial  Tissue Doppler

 Endo. Polyp  Qpilm:Qsys

 Rt. / Lt. Ovarian Mass  HR

 Uterine Tumor 1  Aorta

 Uterine Tumor 2  LA

 Uterine Tumor 3  LA Vol. (Simpson)

 Cervical Tumor  Shunt


UGEO WS80A Datasheet V1.0_r1.2 Page 17
 Comment  Dorsalis Pedis Artery

 Medial Plantar Artery


CAROTID MEASUREMENT  Lateral plantar Artery

 Metatarsal Artery
 Rt. / Lt. Subclavian Artery
 Digital Artery
 Rt. / Lt. Proximal Common Carotid Artery
 General
 Rt. / Lt. Mid Commom Carotid Artery
 Volume Flow
 Rt. / Lt. Distal Common Carotid Artery
 Heart Rate
 Rt. / Lt. Bulb
 Comment
 Rt. / Lt. Proximal Internal Carotid Artery

 Rt. / Lt. Mid Internal Carotid Artery


UE ARTERY MEASUREMENT
 Rt. / Lt. Distal Internal Carotid Artery

 Rt. / Lt. External Carotid Artery  Rt. / Lt. Subclavian Artery

 Rt. / Lt. Vertebral Artery  Rt. / Lt. Axillary Artery

 General  Rt. / Lt. Brachial Artery

 Volume Flow  Rt. Lt. Radial Artery

 HR  Rt. / Lt. Ulnar Artery

 Vertebral Artery  Rt. / Lt. Superficial Palmar Arches

 Rt. / Lt. ICA/CCA  General

 Rt. / Lt. A/B  Volume Flow

 Comment  Heart Rate

 Comment
LE ARTERY MEASUREMENT
LE VEIN MEASUREMENT
 Common Iliac Artery

 Internal Iliac Artery  Rt. / Lt. Comment Iliac Vein

 External Iliac Artery  Rt. / Lt. Internal Iliac Vein

 Common Femoral Artery  Rt. / Lt. External Iliac Vein

 Superficial Femoral Artery  Rt. / Lt. Common Femoral Vein

 Deep Femoral Artery  Rt. / Lt. PFV

 Popliteal Artery  Rt. / Lt. Superficial Femoral Vein

 Anterior Tibial Artery  Rt. / Lt. Great Saphenous Vein

 Posterior Tibial Artery  Rt. / Lt. Popliteal Vein

 Perineal Artery  Rt. / Lt. LSV


UGEO WS80A Datasheet V1.0_r1.2 Page 18
 Rt. / Lt. Anterior Tibial Vein  Ascending Aorta

 Rt. / Lt. Posterior Tibial Vein  Descending Aorta

 Rt. / Lt. Peroneal V  Mitral Valve Inflow

 Rt. / Lt. Metatarsal Vein  Mitral Valve Regurgitation

 Rt. / Lt. Digital Vein  Tricuspid Inflow

 Volume Flow  Tricuspid Regurgitation

 General  Preload Index

 Comment  Tei Index

 Fetal Heart Rate


UE VEIN MEASUREMENT  Environment

 Comment
 Rt. / Lt. Internal Jugular Vein

 Rt. / Lt. Innominate Vein


UROLOGY MEASUREMENT
 Rt. / Lt. Subclavian Vein

 Rt. / Lt. Axillary Vein  General

 Rt. / Lt. Brachial Vein Auto Trace

 Rt. / Lt. Cephalic Vein Limited Trace

 Rt. / Lt. Basilic Vein Manual Trace

 Rt. / Lt. Radial Vein PSV

 Rt. / Lt. Ulnar Vein EDV

 Volume Flow %StA

 General %StD

 Comment Vesl. Dist

Vesl.Area
FETAL ECHO MEASUREMENT  Bladder Volume

 Whole Gland prostate Volume


 LV Vol.(Simpson)
 Transitional Zone Volume
 2D Echo
 Residual Volume
 Cardio-Thorax Area Ratio
 Renal Volume
 Fetal M-mode
 Predicted PSA by WG
 Main Pulmonary Artery
 Prostate Specific Antigen
 Ductus Arteriosus
 Predicted PSA by T-Zone
 Inferior Vena Cava
 Digital Rectal Exam
 Ductus Venosus
UGEO WS80A Datasheet V1.0_r1.2 Page 19
 Transrectal US Prostate  Comment

 Trans rectal US Seminal Vesicles

 Comment
THYROID MEASUREMENTS

 Rt. / Lt. Volume


ABDOMEN MEASUREMENTS
 Rt. / Lt. Flow

 Gallbladder  Rt. / Lt. Mass1~Mass5

 Pancreas  Comment

 Bowel

 Liver
BREAST MEASUREMENT

 Prox / Mid / Distal Aorta


 Rt. / Lt. Mass1~Mass10
 Celiac Artery
 Rt. / Lt. Flow
 Splenic Artery
 Comment
 Common / R / L Hepatic Artery

 Prox / Mid / Distal Superior Mesenteric Artery TESTICLE MEASUREMENT


 Inferior Mesenteric Artery
 Rt. / Lt. Volume
 Prox / Mid /Distal Inferior Vena Cava
 Rt. / Lt. Flow
 Kidney
 Rt. / Lt. Mass1~Mass5
 Rt. / Lt. Renal Artery
 Comment
 Rt. / Lt. Arcuate Artery

 Rt. / Lt. Iliac Artery


SUPERFICIAL MEASUREMENT
 Rt. / Lt. Renal Volume

 Rt. / Lt. Portal Volume  Rt. / Lt. Volume

 L / M / R Portal Volume  Rt. / Lt. Flow

 L / M / R Hepatic Volume  Rt. /Lt. Mass1~Mass5

 SMV  Comment

 IMV
MUSCULOSKELETAL MEASUREMENT
 Spleen

 Splenic Volume  Rt. / Lt. Shoulder


 General  Rt. / Lt. Wrist
 HR  Rt. / Lt. Knee
 Renal Aortic Ratio  Rt. / Lt. Ankle
 Indications  Comment
UGEO WS80A Datasheet V1.0_r1.2 Page 20
PEDIATRIC HIPS MEASUREMENT - CAN/CSA 22.2 No.601.1-M90:1990, with R2003,

with R2005
 Rt. / Lt. Hips Angle
 EN(European Standard)
 Comment
- EN 980: 2008

- EN 1041: 2008
 NEMA/AIUM
SAFETY / EMC
- NEMA/AIUM UD-2: 2004

- NEMA/AIUM UD-3: 2004


CLASSIFICATIONS
 SAFETY
User selectable, transducer and scanning mode - Type of protection against electrical shock:
 IEC (International Electro technical Commission) Class I

- IEC/EN 60601-1: Ed. 3.0:2005 - Degree of protection against electrical shock


(Patient connection): Type BF or CF Applied
- IEC/EN 60601-1: Ed. 2.0:1988 + A1:1991 +
Part
A2:1995
 EMC
- IEC/EN 60601-1-1: Ed. 2.0:2000
- RF Emission CISPR 11: Class A
- IEC/EN 60601-1-2: Ed. 2.0:2001 + A1:2004 + Ed.  Probe IPX: 7
3.0:2007  RoHS

- IEC/EN 60601-1-4: Ed. 1.1:1996 + A1:1999  WEEE

- IEC/EN 60601-1-6: Ed. 2.0:2006 + Ed. 3.0:2010

- IEC 60601-1-9: Ed. 1.0:2007

- IEC/EN 60601-2-37: 2001 + A1:2004 + A2:2005 +

Ed. 2.0:2007

- IEC/EN 60601-2-37: Ed. 1.0:2001 + A1:2004 + For more information, please don’t hesitate to contact

A2:2005 SAMSUNG MEDISON

- IEC 61157: Ed. 2.0:2007

- IEC/EN 62304: Ed. 1.0:2006

- IEC/EN 62366: Ed. 1.0:2007


 ISO(International Organization of Standards)
- EN/ISO 10993-1: 2009

- ISO 14971: 2007


 UL(Underwriters Laboratories)
- UL 60601-1 / CAN/CSA 22.2 No.601.1-M90:1990
 CSA(Canadian Standards Association)

UGEO WS80A Datasheet V1.0_r1.2 Page 21