Professional Documents
Culture Documents
Data Sheet
V2.00
Table of Contents
1 SPECIFICATION SUMMARY -------------------------------------------------------------- 1
2 GENERAL SPECIFICATION -------------------------------------------------------------- 3
2.1 Physical Specification ------------------------------------------------------------------ 3
2.2 Console Design ------------------------------------------------------------------ 3
2.3 Main Monitor ------------------------------------------------------------------ 3
2.4 Control Panel ------------------------------------------------------------------ 3
2.5 PC ------------------------------------------------------------------ 3
2.6 Electrical Specification ------------------------------------------------------------------ 3
3 System Specification -------------------------------------------------------------- 3
3.1 Applications ------------------------------------------------------------------ 3
3.2 Presets ------------------------------------------------------------------ 4
3.3 Operation Mode ------------------------------------------------------------------ 4
3.4 Display Mode ------------------------------------------------------------------ 4
3.5 Transducer Types ------------------------------------------------------------------ 5
3.6 System Standard Features ------------------------------------------------------------------ 5
3.7 System Options ------------------------------------------------------------------ 5
3.8 Display ------------------------------------------------------------------ 6
3.9 Language ------------------------------------------------------------------ 6
3.10 Operation Environment ------------------------------------------------------------------ 6
4 Processing -------------------------------------------------------------- 7
4.1 Data Processing ------------------------------------------------------------------ 7
4.2 Pre-Processing ------------------------------------------------------------------ 7
4.3 Post-Processing ------------------------------------------------------------------ 8
5 Connectivity -------------------------------------------------------------- 8
5.1 DICOM ------------------------------------------------------------------ 8
5.2 IHE ------------------------------------------------------------------ 8
5.3 Peripheral Interface ------------------------------------------------------------------ 9
6 Scanning Parameters -------------------------------------------------------------- 9
6.1 2D Mode ------------------------------------------------------------------ 9
6.2 M Mode ------------------------------------------------------------------ 10
6.3 Color Mode ------------------------------------------------------------------ 10
6.4 PWD Mode ------------------------------------------------------------------ 10
6.5 CWD Mode ------------------------------------------------------------------ 10
6.6 PD Mode ------------------------------------------------------------------ 11
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
www.samsungmedison.com
CONFIDENTIAL
THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION WHICH IS PROPRIETARY TO SAMSUNG MEDISON CO., Ltd. THE DOCUMENT AND ITS CONTENTS
(INFORAMTION) CANNOT BE COPIED OR DISCLOSED WITHOUT THE EXPRESSED WRITTEN CONSENT OF SAMSUNG MEDISON CO., Ltd.
www.samsungmedison.com
ECG
IMAGING MODES
M-Mode
IMAGE STORAGE
Color M-Mode
USER INTERFACE
ELECTRICAL PARAMETERS
PRESSURE LIMITS
HUMIDITY LIMITS
Operating: 30 % to 75 %
TEMPERATURE LIMITS
Operating: 10 °C to 35 °C
- Lift: 180 mm
6 User Keys
CONSOLE DESIGN
Tri-Status backlit
3 or 4 Active Probe Ports (Optional) 6 Probe Holders
4 Swivel Wheel Cart Based Type Height Adjustment: 180 mm
Articulated Monitor Arm Swivel : +/-30°
Built-in Printer Storages
PC
Ergonomic Operation Panel
Integrated PC Module
Frequency: 50/60 Hz
Integrated SSD
Voltage: 100 ~ 240 VAC
Windows Embedded Standard 7
Power Consumption: Max. 800 VA with Peripherals
ODD (Optional)
Heat Dissipation: 2,729.7 [BTU/h]
Gel Warmer (Optional)
System Noise: under 34.8 dB
ECG (Optional)
Built-in Equipotential Circuit
APPLICATIONS TCD
Uterus
Abdomen
Venous
Cardiac st
1 Trimester
Gynecology nd
2 Trimester
MSK rd
3 Trimester
Obstetrics
- Write Zoom / Read Zoom / Pen zoom/ Panning Image Reverse: Right/Left, Up/Down
SonoView™
Hybrid Full Digital Beam-forming
Data Backup / Restore
Frequency Range: 1 ~ 18MHz
Image Exporting and Importing
Displayed Imaging Depth (Probe dependent)
PW Velocity Range: 0.1cm/s ~ 8.8m/s
- Minimum Depth of Field: 2cm
CW Velocity Range: 1cm/s ~ 19.3m/s
- Maximum Depth of Field: 38cm
ClearVision ElastoScan
5D NT - Accession Number
XI STIC - Operator
DVD-RW - Indication
2D NT Focal Number
BatteryAssist Gain
Dynamic Range
Map
DISPLAY
Frame Average
Application Power
- 24 hours Pointer
Display Language
- Gain
OPERATING ENVIRONMENT
- Harmonic
- Reject
- Scan Area
PROCESSING - TGC
- Write Zoom
DATA PROCESSING - MultiVision (Probe Dependent)
- MagiCut™
POST-PROCESSING: Hậu xử lý
- VOCAL™
B-Mode - XI VOCAL™
DICOM Storage Commitment (SC) - Digital Color Video Printer: Sony UP-D25MD,
DICOM Worklist
Multi Frame
Single Frame
3D Volume Frame
Transfer Mode
- Send manually
Patient Information Reconciliation (PIR) Angle Steering: 0°, +/-7°, +/- 12°
Dual Live
Audio out L/R
Dynamic Range: 30 ~ 256
D-SUB output
Flip: L/R, U/D
S-Video output
Focus Number: 1 ~ 4
HDMI output
Frequency Compounding
USB 2.0 (6 ports)
Frequency: 3 ~ 5 steps (Probe Dependent)
Ethernet 10/100/1000BASE-T
- Pen, Gen1, Gen2, Res1, Res2
Foot Switch: USB 2.0 (IPX 8)
Gain: 0 ~ 100
DVD Recorder: Sony DVO-1000MD
Gray Map: 1 ~ 12
- S-video, NTSC/PAL
Harmonic: On, Off
- Recording only
Image Size: 70 ~ 100%
Printers
Line Density: Low, Med, High
- Digital BW Video Printer: Sony UP-D897,
Number of TGC Level: 8
Zoom Sensitivity: 0 ~ 5
Panning Velocity
- Size: 50/50, 70/30, 30/70 Display format: Up/down, Side by side, Doppler Only
Sound: 0 ~ 100
Balance: 0 ~ 16
Angle Correction: -80° ~ 80°
Baseline: -8 ~ 8
SV Position control
Color Map: 1 ~ 12
SV Size: 0.5 ~ 25mm
Line Density: Low, Med, High
Quick Angle: 0°, 60°, -60°
Dual Live: On, off
Smoothing: 0 ~ 5
Auto Calc.: Off, Live, Frozen
Steer Angle: 0°, +/-15°, +/-20°, +/-30°
Base line: -8 ~ 8
Filter: 1 ~ 4
Chroma Map: Off, 1 ~ 11
Doppler Map: 1 ~ 12 3D
Sound: 0 ~ 100 5D NT
- Layout
PD MODE - Play Mode: Loop, Yoyo
- Layout XI VOCAL™
- Bias - Start
- Init User1~3
- Ruler XI STIC™
- Solid / General / Prostate / Cystic / Sphere / Manual - Set color (Hue, Saturation, Lightness)
- Init
- Ref. Image: A / B / C ELASTOSCAN MODE
Line Density: Low, Med, High Application : Abdomen, MSK, Small Parts, Vascular, OB,
Frequency
CONVEX
Gain: 0 ~ 100
Vascular, Urology
Safety Class: BF
TRANSDUCERS
CA2-9AD
LINEAR Center frequency : 4.9MHz
Band Width: 2 ~ 9 MHz
LA3-14AD
Radius of curvature : 60.365 mm
Center frequency : 7.2MHz
Field of view : 58 °
Band Width : 3 ~ 14 MHz
Number of elements : 192
Radius of curvature : Flat
Biopsy Guide : Available
Field of view : 50 mm
Application : Abdomen, MSK, OB, GYN, Pediatric,
Number of elements : 256
Vascular, Urology
Biopsy Guide : Available
Safety Class: BF
Application : Abdomen, MSK, Small Parts, Vascular, OB,
GYN, Pediatric
CF4-9
Safety Class: BF
Center frequency : 5.65MHz
Band Width: 4 ~ 9 MHz
LA3-16A
Radius of curvature : 14 mm
Center frequency : 8.7MHz
Field of view : 92 °
Band Width : 3 ~ 16 MHz
Number of elements : 128
Radius of curvature : Flat
Biopsy Guide : Not available
Field of view : 38.4 mm
Application : Abdomen, MSK, OB, GYN, Pediatric,
Number of elements : 192
Vascular, Urology
Biopsy Guide : Available
Safety Class: BF
Vascular, Urology
Field of view : 90 °
Band Width : 4 ~ 9 MHz
Number of elements : 80
Radius of curvature : 10.073 mm
Biopsy Guide : Not Available
Field of view : 148.092 °
Application : Abdomen, Cardiac, Vascular, Pediatric
Number of elements : 128
Safety Class: BF
Biopsy Guide : Available
Safety Class: BF
PA3-8B
Center frequency : 4.7MHz
Band Width: 3 ~ 8 MHz
Vascular
PA4-12B Gynecology
Urology
Field of view : 90°
MSK
Number of elements : 96
Small Parts
Biopsy Guide : Not Available
Pediatric
Application : Cardiac, Pediatric
Safety Class: BF CALIPER
PENCIL 2D Distance
M Distance
CW6.0
2D Trace
Center frequency : 6.0MHz
2D Trace length
Application : Cardiac, Vascular
Doppler Manual Trace
Safety Class : BF
Doppler Limited Trace
2 Lines Angle
DP2B
3 Points Angle
Center frequency : 2.0MHz
Ellipse (Area / Circumference)
Application : Adult Echo, Ped Echo
Spline
Safety Class : BF
Open Spline
Closed Spline
MMPT3-7
%Stenosis (Diameter)
Phased Array (TEE Type)
%Stenosis (Area)
Radius of curvature 100mm
1 Distance Volume
Field of view 90°
2 Distance Volume
Number of elements 64
3 Distance Volume
Biopsy Guide Not Available
Ellipse Volume
Application : Cardiac
Ellipse + Distance Volume
Safety Class BF
Disk Volume
Slope
Velocity LV Mass
Acceleration RV (2D)
RI Aorta
Auto Trace RA
AV
ABDOMEN MV
TV
Gallbladder
PV
Pancreas
Shunt
Bowel
IVC
Kidney Vol. (Right / Light)
Tei Index
Liver
Plum. Vein
Spleen
Hepatic Vein
Aorta
Tissue Doppler
RA (Right / Left)
Qp/Qs
Seg. A (Right / Left)
LV (M)
Arc. A (Right / Left)
RV (M)
Celiac A
Splenic A
CAROTID
Hepatic A (C / R / L)
SMV
UE ARTERY
CARDIAC
Subclavian A (Right / Left)
GYNECOLOGY
LE ARTERY
Uterus
CIA (Left / Right)
Cervix
IIA (Left / Right)
Cyst (Right / Left)
EIA (Left / Right)
Ovary (Right / Left)
CFA (Left / Right)
Follicles (Right / Left / 1 ~ 12)
SFA (Left / Right)
Mass 1 ~ 3
DFA (Left / Right)
Ovarian A (Right / Left)
Popliteal A (Left / Right)
Uterine A (Right / Left)
ATA (Left / Right)
Pericystic Flow
PTA (Left / Right)
Endometrial Flow
Peroneal A (Left / Right)
Endo. Polyp
DPA (Left / Right)
Ovarian Mass (Right / Left)
MPA (Left / Right)
Uterine Tumor 1 ~ 3
LPA (Left / Right)
Cervical Tumor
Metatarsal A (Left / Right)
Ectopic Pregnancy
Digital A (Left / Right)
OBSTETRICS
LE VEIN
Fetal Biometry
CIV (Left / Right)
Fetal Cranium
IIV (Left / Right)
Fetal Long Bone
EIV (Left / Right)
Fetal others
CFV (Left / Right)
AFI
Maternal Others
Ratio BREAST
Umbilical Artery
Mass 1 ~ 10 (Right / Left)
Mid Cereb A
Breast Flow (Right / Left)
Uterine A (Right / Left)
Placenta A
MSK
Fetal Carotid (Right / Left)
Fetal Aorta
Shoulder (Right / Left)
Renal A (Right / Left)
Wrist (Right / Left)
Duct Venosus
Knee (Right / Left)
Fetal HR
Ankle (Right / Left)
PLI
THYROID
FETAL HEART
Thyroid Vol. (Right / Left)
LV Vol. (Simpson)
Thyroid Flow (Right / Left)
2D Echo
Mass 1 ~ 5 (Right / Left)
CTAR
MPA TESTICLE
Duct Artriosus
MV
TV SUPERFICIAL
PLI
Superficial Vol (Right / Left)
TEI
Superficial Flow (Right / Left)
Fetal HR
Mass 1 ~ 5 (Right / Left)
M Echo
PEDIATRIC
UROLOGY
Hip Angle (Right / Left)
WG Prostate
T-Zone Vol
Bladder Vol.
SAFETY / EMC
Residual Vol
dependent
SAFETY
Dedicated Output Display on the system monitor display
- Type of protection against electrical shock: Class I
of output acoustic
- Degree of protection against electrical shock (Patient
Power level, as well as thermal and mechanical indices:
connection): Type BF or CF Applied Part
PWR – Output Power level. Range: From 2 % of
EMC
maximum output
- RF Emission CISPR 11: Class A
Level is increased by 2% in each step.
Degree of protection against harmful ingress of water:
Mechanical Index (MI): 0.01~1.90 Range
Ordinary equipment (All of applied parts (IPX7) except
Thermal Index (TI): 0.01~6.00 Range
for ECG, and Foot switch (IPX8))
- TIC – Thermal Index, Bone at Surface
RoHS Compliant
- TIB – Thermal Index, Bone at Focus
WEEE Compliant
- TIS – Thermal Index, Soft Tissue
APPLIED STANDARDS
ANTI-VIRUS SOLUTION
Safety)
Disable USB Autorun Feature
- IEC/EN 60601-1: Ed. 3.0:2005 + A1:2012
- Executable applications in USB stick are never launched
- IEC/EN 60601-1: Ed. 2.0:1988 + A1:1991 + A2:1995
- Prevent autorun virus through USB stickDedicated
- IEC/EN 60601-1-1: Ed. 2.0:2000
Output Display on the system monitor display of output
- IEC/EN 60601-1-2: Ed. 2.0:2001 + A1:2004 + Ed.
acoustic
3.0:2007
Block Network Port (Except DICOM communication port)
- IEC/EN 60601-1-4: Ed. 1.1:1996 + A1:1999
- Ultrasound Machine allow only DICOM data through
- IEC/EN 60601-1-6: Ed. 2.0:2006 + Ed. 3.0:2010
DICOM port
- IEC/EN 60601-2-37: 2001 + A1:2004 + A2:2005 + Ed.
- The network data of other network ports are rejected
2.0:2007
by Windows firewall
- IEC/EN 60601-2-37: Ed. 1.0:2001 + A1:2004 + A2:2005
Prohibit user from accessing windows application (such
- IEC/EN 62366: Ed. 1.0:2007
as Explorer)
Biocompatibility
- Impossible to execute applications which is not allowed
- ISO/EN 10993-1: 2009
- Impossible to access internet web pages
Labeling
- EN 980: 2008
- EN 1041: 2008
NEMA/AIUM
- NEMA/AIUM UD-2: 2004
Phone: +82-2-2194-1226
E-Mail:mk.doo@samsungmedison.com