TEST REPORT
IEC 61010-1
Safety requirements for electrical equipment for measurement,
control, and laboratory use
Part 1: General requirements
General disclaimer:
1) The test report is only valid for the sample delivered by client.
2) The test results presented in this report relate only to the object tested.
3) Any erasure or without specific testing seal the report is invalid.
4) This report shall not be reproduced, except in full, without the written approval of EPIN Suzhou Ltd.
5) Please apply for rechecking within 15 days after receiving the report if there is any objection.
Attachment 1 of MEE202002154-02
Responsible Testing Laboratory (as applicable), testing procedure and testing location(s):
TRFNo. IEC61010 1L
Attachment 1 of MEE202002154-02
[>3 The product fulfils the requirements of IEC 61010-1:2010+A1:2016 except subclause 12.3 which was not
evaluated in this report.
MC1001000 C€
Zhejiang Push Kang Biotechnology Co., Ltd.
Add: C408. Science and Technology Innovation Park NO.398. Mahuan Road, Binhai
new Area. 312366 Shaoxing. Zhejiang, PEOPLE’S REPUBLIC OF CHINA
Warning label
H I!
Packing label
TRFNo. IEC61010 1L
Attachment 1 of MEE202002154-02
The equipment is supplied by AC power source and mainly consists of PSU, MCU board, temperature
control system, detection system, brushless DC motor, cooling fan, display unit, printer, etc. See figure below
© Carcasa de plástico, ©Impresora, ©Cubierta superior, ©Pantalla táctil, ©Carcasa metálica inferior, ©Panel metálico
trasero, ©Interruptor de encendido, ©Interfaz LAN, ©Interfaz USB, ®Interfaz RS232, O Entrada del aparato
El equipo proporciona aislamiento básico desde los circuitos principales hasta la carcasa metálica conectada a tierra,
y aislamiento doble/reforzado desde los circuitos principales hasta los circuitos secundarios de voltaje no peligroso.
La fuente de alimentación fue certificada según EN 60601-1, DI/RI se proporciona entre el circuito primario y el circuito
secundario. La condición de funcionamiento de la fuente de alimentación dentro del equipo no excede la condición
certificada; no se considera necesaria ninguna prueba adicional.
El equipo está diseñado con protección de enclavamiento para evitar que el disco giratorio de alta velocidad salga
volando cuando la cubierta superior se abre manualmente durante la prueba. El principio básico de la protección de
enclavamiento es que cuando se abre la cubierta superior del módulo de visualización, el sensor Hall devuelve la
señal de apertura al software. Después de que el sistema de software reciba la señal, finalizará la prueba y mostrará
información de alarma en la pantalla táctil. Una vez activado el enclavamiento, la prueba no continuará incluso si se
vuelve a cerrar la tapa.
EPiniOK
Report No. MEE202002154-01
Page 8 of 91
IEC 61010-1
Clause Requirement + Test Result - Remark Verdict
4 TESTS p
4.4 Testing in SINGLE FAULT CONDITIONS
IEC 61010-1
Clause Requirement + Test Result - Remark Verdict
IEC 61010-1
Clause Requirement + Test Result - Remark Verdict
IEC 61010-1
Clause Requirement + Test Result - Remark Verdict
5.4 Documentation p
5.4.1 General p
Equipment is accompanied by documentation for p
safety purposes for OPERATOR or RESPONSIBLE BODY
Safety documentation for service personnel p
authorized by the manufacturer
Documentation necessary for safe operation is p
provided in printed media or
in electronic media if available at any time N/A
Documentation includes: —
IEC 61010-1
Clause Requirement + Test Result - Remark Verdict
a) Intended use p
b) Technical specification p
c) Name and address of manufacturer or supplier p
d) Information specified in 5.4.2 to 5.4.6 p
e) Information to mitigate residual RISK N/A
(see also subclause 17)
f) Accessories for safe operation of the equipment P
specified
g) Guidance provided to check correct function of p
the equipment, if incorrect reading may cause a
HAZARD from harmful or corrosive substances of
HAZARDoUS live parts
TRFNo. IEC61010 1L
Attachment 1 of MEE202002154-02
IEC 61010-1
Clause Requirement + Test Result - Remark Verdict
IEC 61010-1
Clause Requirement + Test Result - Remark Verdict
IEC 61010-1
Clause Requirement + Test Result - Remark Verdict
TRFNo. IEC61010 1L
Attachment 1 of MEE202002154-02
IEC 61010-1
Clause Requirement + Test Result - Remark Verdict
6.5.1 General P
ACCESSIBLE parts are prevented from becoming
HAZARDOUS live by the primary means of protection
and supplemented by one of:
a) PROTECTIVE BONDING (see 6.5.2) p
b) SUPPLEMENTARY INSULATION (see 6.5.3) p
c) automatic disconnection of the supply (see 6.5.5) No such protective means N/A
d) current- or voltage-limiting device (see 6.5.6) No such protective means N/A
Alternatively one of the single means of protection is —
used:
IEC 61010-1
Clause Requirement + Test Result - Remark Verdict
TRFNo. IEC61010 1L
Attachment 1 of MEE202002154-02
IEC 61010-1
Clause Requirement + Test Result - Remark Verdict
IEC 61010-1
Clause Requirement + Test Result - Remark Verdict
TRFNo. IEC61010 1L
Attachment 1 of MEE202002154-02
IEC 61010-1
Clause Requirement + Test Result - Remark Verdict
IEC 61010-1
Clause Requirement + Test Result - Remark Verdict
Equipment rated for operating altitude greater than Rated operating altitude: up to N/A
2000 m correction factor of Table 3 of 61010-1 2000 m
applied
6.7.1.3 CREEPAGE DISTANCES p
Required CREEPAGE DISTANCES reflecting factors of (see Forms A.14 and A.15) p
6.7.1.1 a) tod)
CTI material group reflected by requirements Material group lllb is assumed p
CTI test performed N/A
6.7.1.4 Solid insulation P
Required solid insulation reflecting factors of (see Forms A.14 and A.15) p
6.7.1.1 a) tod)
6.7.1.5 Requirements for insulation according to type of (see Forms A.14 and A.15) p
circuit
a) 6.7.2 MAINS circuits of OVERVOLTAGE CATEGORY II p
up to nominal supply voltage of 300 V
b) 6.7.3 secondary circuits separated from circuits Secondary circuits separated N/A
defined in a) by transformer from mains circuits by PSU
c) K.1 MAINS circuits of OVERVOLTAGE CATEGORY III No such category N/A
and IV or OVERVOLTAGE CATEGORY II over 300 V
d) K.2 secondary circuits separated from circuits No such category N/A
defined in c) by transformer
e) K.3 circuits having one or more of: —
1) maximum TRANSIENT OVERVOLTAGE is limited to N/A
known level below the level of MAINS CIRCUIT
2) maximum TRANSIENT OVERVOLTAGE above the N/A
level of MAINS CIRCUIT
3) WORKING VOLTAGE is the sum of more than one N/A
circuit or a mixed voltage
4) WORKING VOLTAGE includes recurring peak N/A
voltage, may include non-sinusoidal or
non-periodic waveform
5) WORKING VOLTAGE with a frequency above N/A
30 kHz
6.7.2 Insulation for MAINS CIRCUITS of OVERVOLTAGE P
CATEGORY II with a nominal supply voltage up to
300 V
6.7.2.1 CLEARANCES and CREEPAGE DISTANCES (see Forms A.14 and A.15) —
Values for MAINS CIRCUITS of Table 4 are met p
Coatings to achieve reduction to POLLUTION DEGREE 1 No such coatings N/A
comply with requirements of Annex H
6.7.2.2 Solid insulation P
6.7.2.2.1 General P
IEC 61010-1
Clause Requirement + Test Result - Remark Verdict
67.3 Insulation for secondary circuits derived from MAINS No hazardous voltage in N/A
CIRCUITS Of OVERVOLTAGE CATEGORY II up to 300 V secondary circuits, no
insulation considered
necessary
67.3.1 General N/A
IEC 61010-1
Clause Requirement + Test Result - Remark Verdict
Complies as applicable: —
IEC 61010-1
Clause Requirement + Test Result - Remark Verdict
IEC 61010-1
Clause Requirement + Test Result - Remark Verdict
IEC 61010-1
Clause Requirement + Test Result - Remark Verdict
IEC 61010-1
Clause Requirement + Test Result - Remark Verdict
c) separable plug (without locking device) Separable plug is not used as N/A
disconnecting device
6.11.4 Disconnecting devices Appliance coupler is used as P
disconnecting device
6.11.4.1 General p
Disconnecting device part of equipment p
Electrically close to the SUPPLY p
Power-consuming components not electrically p
located between the supply source and the
disconnecting device
Except electromagnetic interference suppression p
circuits permitted to be located on the supply side of
the disconnecting device
6.11.4.2 Switches and circuit-breakers The switch is used only for N/A
functional purpose
When used as disconnection device: —
Circuit breaker meets the relevant requirements IEC No circuit breaker used N/A
60947-2 and is suitable for the application
Switch meets the relevant requirements IEC 60947-3 The switch is not used as N/A
and is suitable for the application disconnecting device
Marked to indicate function........................................... —
Not incorporated in MAINS cord P
Does not interrupt PROTECTIVE EARTH CONDUCTOR P
6.11.4.3 Appliance couplers and plugs P
Where an appliance coupler or separable plug is —
used as the disconnecting device (see 6.11.3.2):
Readily identifiable and easily reached by the P
operator
Single-phase portable equipment cord length not p
more than 3 m
PROTECTIVE EARTH CONDUCTOR connected first and p
disconnected last
TRFNo. IEC61010 1L
Attachment 1 of MEE202002154-02
IEC 61010-1
Clause Requirement + Test Result - Remark Verdict
IEC 61010-1
Clause Requirement + Test Result - Remark Verdict
7.4 Stability P
Equipment not secured to building structure is P
physical stable
Stability maintained after opening of drawers etc. by No drawers N/A
automatic means, or
warning marking requires the application of means N/A
Compliance checked by following tests as applicable: (see Form A.20A) —
a) 10° tilt test for other than handheld equipment p
b) multi-directional force test for equipment exceeds N/A
height of 1 m and mass of 25 kg
c) downward force test for floor-standing equipment N/A
d) overload test with 4 times maximum load for N/A
castor or support foot that supports greatest load,
or
e) castor or support foot that supports greatest load N/A
removed from equipment
7.5 Provisions for lifting and carrying P
7.5.1 General N/A
Equipment more than 18 kg.......................................... Not more than 18 kg N/A
Has means for lifting or carrying; or N/A
Directions are given in documentation N/A
7.5.2 Handles and grips N/A
Handles or grips withstand four times weight N/A
7.5.3 Lifting devices and supporting parts N/A
RATED for maximum load; or N/A
Tested with four times maximum static load N/A
7.6 Wall mounting Not for wall mounting N/A
Mounting brackets withstand four times weight N/A
One fastner removed and test repeated with two N/A
times weight
7.7 Expelled parts Rotating plate is considered P
Equipment contains or limits the energy P
Protection not removable without the aid of a tool N/A
IEC 61010-1
Clause Requirement + Test Result - Remark Verdict
TRFNo. IEC61010 1L
Attachment 1 of MEE202002154-02
IEC 61010-1
Clause Requirement + Test Result - Remark Verdict
IEC 61010-1
Clause Requirement + Test Result - Remark Verdict
IEC 61010-1
Clause Requirement + Test Result - Remark Verdict
IEC 61010-1
Clause Requirement + Test Result - Remark Verdict
IEC 61010-1
Clause Requirement + Test Result - Remark Verdict
11.1 General P
Protection to oPERAToRS and surrounding area P
provided by EQUIPMENT
All fluids specified by manufacturer considered P
11.2 Cleaning (see Form A.30) P
11.3 Spillage N/A
11.4 Overflow N/A
11.5 Battery electrolyte N/A
Battery electrolyte leakage presents no HAZARD Lithium battery is used on the N/A
display control board
11.6 Equipment RATED with a degree of ingress (see Form A.30) N/A
protection (IP code)
11.6.1 General N/A
Equipment marked with IP code................................... —
Conditions specified in the documentation N/A
11.6.2 Conditions for testing N/A
Equipment in clean and new condition, all parts in N/A
place and mounted as specified by manufacturer
Complete equipment tested, or N/A
representative parts tested N/A
HAND-HELD EQUIPMENT and PORTABLE EQUIPMENT N/A
placed in least favourable position of NoRMAL use
Other equipment positioned or installed as specified N/A
TERMINALS provided with protective cap or cover, are N/A
installed as specified by manufacturer
The equipment is operating (energized) during the —
treatment except:
a) If manufacturer specifies degrees of protection N/A
for non-operating (de-energized) equipment, or
b) Equipment is operating or non-operating during N/A
the treatment with does not affect the test results
11.6.3 Protection against solid foreign objects (including N/A
dust)
Applicable test of IEC 60529 for protection against N/A
solid foreign objects conducted
Additionally inspection of equipment resulted: —
a) No deposit on insulation parts that could lead to N/A
a HAZARD
b) No created accumulations that have the N/A
potential to cause spread of fire
11.6.4 Protection against water N/A
IEC 61010-1
Clause Requirement + Test Result - Remark Verdict
IEC 61010-1
Clause Requirement + Test Result - Remark Verdict
IEC 61010-1
Clause Requirement + Test Result - Remark Verdict
IEC 61010-1
Clause Requirement + Test Result - Remark Verdict
TRF No IEC61010 1L
Attachment 1 of MEE202002154-02
IEC 61010-1
Clause Requirement + Test Result - Remark Verdict
15 PROTECTION BY INTERLOCKS P
TRFNo. IEC61010 1L
Attachment 1 of MEE202002154-02
IEC 61010-1
Clause Requirement + Test Result - Remark Verdict
15.1 General P
Interlocks are designed to remove a HAZARD See general product P
before OPERATOR exposed information
15.2 Prevention of reactivation p
15.3 Reliability p
Single fault unlikely to occur; or N/A
Cannot cause a HAZARD The hall sensor was cycled to p
switch till twice the maximum
number that likely to occur
during the expected life had
completed. The protection was
not impaired after the test
17 RISK ASSESSMENT P
RISK assessment conducted, if HAZARD might arise P
and not covered by Clauses 6 to 16
TOLERABLE RISK achieved by iterative documented —
process covering the following:
a) RISK analysis P
Identifies HAZARDS and estimates RISK P
b) RISK evaluation P
Plan to judge acceptability of resulting RISK level P
based on the estimated severity and likelihood of
a RISK
c) RISK reduction P
Initial RISK reduced by counter measures; P
Repeated RISK evaluation without new RISKS p
introduced
IEC 61010-1
Clause Requirement + Test Result - Remark Verdict
IEC 61010-1
Clause Requirement — Test
Result — Remark Verdict
IEC 61010-1
Clause Requirement — Test Result — Remark Verdict
IEC 61010-1
Clause Requirement — Test Result — Remark Verdict
Method Test agent Remains legible Label loose Curled edges Comments
Verdict Verdict Verdict
1). 2) A,B P P P
Supplementary information:
TRFNo. IEC61010 1L
Attachment 1 of MEE202002154-02
IEC 61010-1
Clause Requirement — Test Result — Remark Verdict
IEC 61010-1
Clause Requirement — Test
Result — Remark Verdict
IEC 61010-1
Clause [Requirement —Test | Result — Remark | Verdict
IEC 61010-1
Clause Requirement — Test Result — Remark Verdict
TRFNo. IEC61010 1L
Attachment 1 of MEE202002154-02
IEC 61010-1
Clause Requirement — Test Result — Remark Verdict
6.5.2.5 TABLE: BONDING impedance of PERMANENTLY CONNECTED EQUIPMENT Form A.10 N/A
ACCESSIBLE part under test Test Voltage attained after 1 Verdict
current min (maximum 10 V)
[A] M
— — — —
Supplementary information:
IEC 61010-1
Clause I Requirement — Test I Result — Remark I Verdict
IEC 61010-1
IEC 61010-1
Clause Requirement — Test Result — Remark Verdict
L
220-240 Vac
GND
TRFNo. IEC61010 1L
Attachment 1 of MEE202002154-02
IEC 61010-1
Clause Requirement — Test Result — Remark Verdict
D Mains circuits to plastic DI/RI 240 340 0.05 3.0 >3.9 6.0 >7.8 lllb P
enclosure
E Between Pri. circuits and DI/RI 240 340 0.05 3.0 — 6.0 — lllb P Separately approved unit
Sec. circuits of PSU D
F Mains circuits to non- DI/RI 240 340 0.05 3.0 >3.9 6.0 >7.8 lllb P
hazardous voltage
secondary circuits
NOTE 1 - refer to Form A. 14 for type of insulation shown in the insulation diagram NOTE 2 - to be used for definition of required insulation (see Form A. 14)
IEC 61010-1
Clause | Requirement — Test | Result — Remark | Verdict
IEC 61010-1
Clause Requirement-Test Result — Remark Verdict
Manufacturer...................................................
Type.................................................................
Construction.....................................................
Potting compound............................................
CREEPAGE DISTANCES measured......................
CLEARANCES measured....................................
Thickness through insulation...........................
Adhesive test Pass/Fail...................................
Test temperature T °C.....................................
Cycles at U= AC 500 V Leakage current (at AC 500 V)
mA
Number of cycles Date 68 h / 1h/ 2h/ 1 h/
125 °C 25 °C 0°C 25 °C
1. Cycle from to
2. Cycle from to
3. Cycle from to
4. Cycle from to
5. Cycle from to
6. Cycle from to
7. Cycle from to
8. Cycle from to
9. Cycle from to
10. Cycle from to
After Cycling Test:
Humidity conditioning 48 h
Requirements for dielectric strength (s. insulation diagram) Test voltage V r.m.s. Verdict
Basic insulation V r.m.s.
Supplementary insulation V r.m.s.
Reinforced insulation V r.m.s.
NOTE - to be used for evaluation of components containing insulation through solid insulation, when the component standard require
thermal cycling test. Ref Clause 14.1 and Figure 15, option b)
Supplementary information:
IEC 61010-1
Clause Requirement — Test Result — Remark Verdict
TRFNo. IEC61010 1L
Attachment 1 of MEE202002154-02
IEC 61010-1
Clause Requirement — Test Result — Remark Verdict
IEC 61010-1
Supplementary information:
IEC 61010-1
Clause Requirement-Test Result - Remark Verdict
TRFNo. IEC61010 1L
Attachment 1 of MEE202002154-02
IEC 61010-1
Clause Requirement - Test Result - Remark Verdict
IEC 61010-1
Clause Requirement-Test Result - Remark Verdict
TRFNo. IEC61010 1L
Attachment 1 of MEE202002154-02
IEC 61010-1
Clause | Requirement — Test
| Result — Remark | Verdict
IEC 61010-1
Clause Requirement — Test Result — Remark Verdict
Material tested............................................................ —
Generic name............................................................. —
Material manufacturer................................................ —
Type............................................................................ —
Colour......................................................................... : —
Conditioning details.................................................... —
Sample
1 2 3 4 5 6
Thickness of specimen mm
Duration of flaming after first Application s
Duration of flaming plus glowing s
After second application
Specimen burns to holding clamp Yes/No
Cotton ignited Yes/No
Sample result Pass/Fail
Supplementary information:
IEC 61010-1
Clause | Requirement — Test~ Result — Remark Verdict
TRFNo. IEC61010_1L
Attachment 1 of MEE202002154-02
IEC 61010-1
Clause | Requirement — Test | Result — Remark
| Verdict
IEC 61010-1
Clause Requirement — Test Result — Remark Verdict
TRFNo. IEC61010 1L
Attachment 1 of MEE202002154-02
IEC 61010-1
Clause Requirement — Test Result — Remark Verdict
TRFNo. IEC61010 1L
Attachment 1 of MEE202002154-02
IEC 61010-1
IEC 61010-1
Clause Requirement — Test Result — Remark Verdict
TRFNo. IEC61010 1L
Attachment 1 of MEE202002154-02
IEC 61010-1
Clause Requirement — Test Result — Remark Verdict
IEC 61010-1
Clause Requirement — Test Result — Remark Verdict
IEC 61010-1
Clause Requirement — Test Result — Remark Verdict
IEC 61010-1
Clause Requirement — Test Result — Remark Verdict
IEC 61010-1
Clause Requirement — Test Result — Remark Verdict
IEC 61010-1
Clause [Requirement — Test |Result—Remark | Verdict"
IEC 61010-1
Clause Requirement — Test Result — Remark Verdict
11.7.2 TABLE: Leakage and rupture at high pressure Form A.31 N/A
Part Maximum Test Leakage Deformation Burst Comments
permissible pressure
working
pressure
[MPa] [MPa] Yes / No Yes / No Yes / No
— — — — — — —
NOTE - see also Annex G with requirements for USA and Canada.
Supplementary information:
Supplementary information:
IEC 61010-1
Clause Requirement — Test Result — Remark Verdict
TRFNo. IEC61010 1L
Attachment 1 of MEE202002154-02
IEC 61010-1
Clause Requirement — Test Result — Remark Verdict
IEC 61010-1
Clause Requirement — Test Result — Remark Verdict
I
[1,7] VRTCl >
inr
o>
2
J25 —
CR2032 SMD
TRFNo. IEC61010 1L
Attachment 1 of MEE202002154-02
IEC 61010-1
Clause Requirement — Test Result — Remark Verdict
IEC 61010-1
Clause Requirement —Test Result — Remark Verdict
IEC 61010-1
Clause Requirement — Test Result — Remark Verdict
IEC 61010-1
Clause | Requirement —Test | Result — Remark | VerdicT
14.8 | TABLE: Circuits used to limit TRANSIENT OVERVOLTAGES Form A.411 N/A
Circuit / Designation Overvoltage MAINS voltage Test voltage fcn fc fmax Ignited Safely Properly Verdict Comments
Category [V r.m s ] M [°C] rcj [°C] Yes / No suppressed functional
Yes/No Yes / No
— — — — — — — — — — — —
Test room ambient temperature:||CC
NOTE - tm = measured temperature
k = tr corrected (U-f.+ 40 °C or max RATED ambient)
UM = maximum permitted temperature
Conformity is checked by applying 5 positive and 5 negative impulses with the applicable impulse withstand voltage, spaced up to 1 min apart, from a hybrid impulse generator (see IEC 61180-1).
Supplementary information:
IEC 61010-1
Clause Requirement-Test Result — Remark Verdict
Supplementary information:
IEC 61010-1
Clause Requirement-Test Result — Remark Verdict
IEC 61010-1
Clause [Requirement — Test [Result— Remari? | Verdict
IEC 61010-1
Clause [Requirement— Test |Result—Remark | Verdict
IEC 61010-1
Clause [Requirement —Test [Result—Remark | Verdict
Photo Document
Figure 6. Top view of Coagulation Analyzer (with protective barrier of printer opened)
Attachment 1 of MEE202002154-02
ACCREDITED'
Testing Laboratory
Tl-700
EPini&K
TEST REPORT
IEC 61010-2-101
Safety requirements for electrical equipment for measurement, control
and laboratory use
Part 2-101: Particular requirements for in vitro diagnostic
(IVD) medical equipment
Testing Laboratory...........................
Address............................................ Building 4xNa55Q Fenhy Avenue, Wujiang District, Suzhou, China
Applicant’s name.............................. thejfang PushKangBiotechnology Co., Ltd.
Address............................................ C408, Science and Technology Innovation Park NO.398, Mahuan
Roaa>Binhaf>B^rea, 312366 Shaoxing, Zhejiang, PEOPLE'S
REPUBLIC75FCHINA
Test specification:
Standard........................................... EN 61010-2-101:2017
Test procedure................................. Testing
Non-standard test method............... None
Test Report Form IEC61010 2 101B
No........................ Test Report VDE Testing and Certification Institute
Form(s) Originator........ Master 2015-03
TRF...................................... onformity Assessment Schemes for Electrotechnical Equipment
All rights reserved.
Copyright © 2015 IEC System of C
and Components (IECEE System). jrCEE
in part for non-commercial purposes as long as the IECEE is acknowledged as
takes no responsibility for and will not assume liability for damages resulting from
This publication may be reproduced in whole laterial due to its placement and context.
copyright owner and source of the material. IE
the reader’s interpretation of the reproduced n
General disclaimer:
1) The test report is only valid for the sample delivered by client.
2) The test results presented in this report relate only to the object tested.
3) Any erasure or without specific testing seal the report is invalid.
4) This report shall not be reproduced, except in full, without the written approval of EPIN Suzhou Ltd.
5) Please apply for rechecking within 15 days after receiving the report if there is any objection.
Attachment 1 of MEE202002154-02
EPinifiK
Test item description.......................
Page 2 of 11 Report No. MEE202002154-02
Coagulation Analyzer
Trade Mark........................................ :
PUSHING
Manufacturer.....................................: Zhejiang PushKang Biotechnology Co., Ltd.
C408, Science and Technology Innovation Park NO.398, Mahuan
Road, Binhai new Area, 312366 Shaoxing, Zhejiang, PEOPLE’S
REPUBLIC OF CHINA
Model/Type reference...................... MC100
Ratings.............................................. : 220-240 V, 50/60 Hz, 60 VA
Responsible Testing Laboratory (as applicable), testing procedure and testing location(s):
□ Testing procedure:
SMT/CTF Stage 3 or 4:
Testing location/ address.............................:
IEC61010_2_101B
Attachment 1 of MEE202002154-02
Summary of testing:
IEC61010 2 101B
Attachment 1 of MEE202002154-02
20XX-X
SN MC1001000
IVD C€
Warning label
Label on package
IEC61010 2 101B
Attachment 1 of MEE202002154-02
EPiniOK Page 5 of 11
Report No. MEE202002154-02
Test item particulars...............................................
Classification of installation and use.................... Desk-top, professional use
Supply Connection.................................................. Detachable cord set
Possible test case verdicts:
- test case does not apply to the test object............N/A
- test object does meet the requirement...................P (Pass)
- test object does not meet the requirement...........F (Fail)
Testing...................................................................... :
Date of receipt of test item...........................................March 16, 2020
Date (s) of performance of tests.................................March 16, 2020 to April 30, 2020
General remarks:
’’(See Enclosure #)" refers to additional information appended to the report.
"(See appended table)" refers to a table appended to the report.
Throughout this report a comma / [X] point is used as the decimal separator.
When differences exist; they shall be identified in the General product information section.
Name and address of factory (ies)..............................Zhejiang PushKang Biotechnology Co., Ltd.
C408, Science and Technology Innovation Park
NO.398, Mahuan Road, Binhai new Area, 312366
Shaoxing, Zhejiang, PEOPLE’S REPUBLIC OF
CHINA
IEC61010 2 101B
Attachment 1 of MEE202002154-02
IEC61010 2 101B
Attachment 1 of MEE202002154-02
EPini&x Page 7 of 11
IEC 61010-2-101
Report No. MEE202002154-02
IEC61010_2_101B
Attachment 1 of MEE202002154-02
5.4.4.101 Instructions for use, self-test IVD medical Not self-test IVD equipment N/A
equipment
Instructions for use for self-test equipment comply N/A
with ISO 18113-5
5.4.101 Removal of equipment from use for repair or P
disposal
Instructions for the RESPONSIBLE BODY for —
eliminating or reducing HAZARDS includes:
Removal from use p
Transportation or disposal, or p
Appropriate contact information provided p
5.4.102 Transport and storage p
IEC61010 2 101B
Attachment 1 of MEE202002154-02
EPinie-K Page 9 of 11
IEC 61010-2-101
Report No. MEE202002154-02
IEC61010 2 101B
Attachment 1 of MEE202002154-02
17 RISK ASSESSMENT P
RISK assessment carried out and documented in p
acc. to the requirements of ISO 14971 for HAZARDS
which not addressed in this standard and Part 1.
IEC61010 2 101B
Attachment 1 of MEE202002154-02
EPiniax Page 11 of 11
IEC 61010-2-101
Report No. MEE202002154-02
NOTE - Where applicable include print method, label material, ink or paint type,
fixing method, adhesive and surface to which marking is fixed.
Method Test agent Remains legible Label loose Curled edges Comments
Verdict Verdict Verdict
1 A.C P P P —
Supplementary information:
None
IEC61010 2 101B
Attachment 1 of MEE202002154-02
1. Test purpose
By simulating the packaging and transportation environment, verify that the
packaging materials designed by Zhejiang Pushkang Biotechnology Co., Ltd. for
MC100 coagulation analyzer meet the requirements for transportation and storage.
2. Description of sample
Product name: Coagulation analyzer
Models: MC100
Sample quantity: 1
Size: 270mm x 180mm x 155mm
Weight: 3kg
Standard: ISTA 2A
4. Vibration test
Laboratory environment: temperature 25°C, humidity 56%.
4.1 Test method
Place the packaged instrument on the vibration test bench, fix the positioning device,
test according to the experimental parameters of Table 2 and Table 3, remove the
instrument after the test is completed, check whether the packaging is intact, and test
whether the instrument is working properly.
Table2 Test free uency and PSD
Frequency (Hz) PSD (g2/Hz)
1.0 0.0001
4.0 0.01
100.0 0.01
200.0 0.0001
Table3 Test time
No. Test time (min) Contact surface with vibration plane
1 90 Bottom surface
2 30 Top surface
3 30 Lateral surface
4 30 Frontal surface
4.2 Evaluation Criteria
4.2.1 Instrument packaging and appearance are in good condition with boot normal
after the test.
4.2.2 Withstand voltage test:
• Basic insulation: 1500V sine wave voltage test, lasting 5s, should not appear
breakdown or repeated arc
• Double insulation: 3000V sine wave voltage test, lasting 5 s, should not appear
breakdown or repeated arc
• Leakage current: test voltage AC 240V, normal working state (start-up),
0.5mA.
test voltage AC 240V, single fault state ( off ), < 3.5 mA.
Grounding Impedance between a protective conductor terminal and a
resistance: contactable component required for a protective connection
shall not exceed 0.IQ
4.3 Test process
The test situation is shown in Figure 3.
Item Result
Withstand voltage test AC withstand voltage BI 1500V, DI / RI 3000V with in
good condition after 5 s
Leakage current test Leakage current 0.33mA
Grounding resistance test Grounding resistance 80.5mQ
5. Test Conclusion
The test results show that the packaging material of the product meets the
requirements of transportation.