Professional Documents
Culture Documents
Structure of Tenofovir
From the IR spectra of Tenofovir different function groups were found and these are
given in following table 3.1.
The proper wavelength was needed to determine maximum detector response. The first
step was to run a UV-VIS spectrum (from 190-400 nm) using an HPLC system equipped
with the Photo Diode Array Detector, from the spectrum it is clear that Tenofovir absorbs
maximum light between 269.2 nm to 260 nm.
The wavelength of 260 nm was selected since it produces less noise, which minimizes
problems that may exhibit around the active ingredient when attempting to quantify
Tenofovir. The spectra of diluted solutions of the Tenofovir in mobile phase (at 10 ppm
concentration) were recorded separately on UV spectrophotometer. The spectra of
Tenofovir, from both the concentrations showed absorption maximum at 260 nm. Also,
the solvents in mobile phase/ diluents was shows no interference at the same wavelength
and hence it is selected as detection wavelength.
Here the aim was to develop simple RP-HPLC method which can be used to monitor
degradation pattern of Tenofovir under different stress conditions.
Various trials were carried out for developing a method and results of respective trial are
depicted in following chromatograms.
In first trial, mobile phase ratio was set to 30:70 v/v i.e 10 mM Ammonium Formate pH:
Methanol adjusted to 3.5 with o- phosphoric acid. Result of the trial is depicted in
following figure (Figure 3.3).
In this trial, retention time of analyte was found to be 2.744 min with tailing factor 1.188.
Next trial was carried out by changing mobile phase composition. In this trial, solvent A
(Ammonium acetate) was changed to sodium acetate and mobile phase ratio was set to
70:30 Methanol: 10mM Sodium Acetate Buffer pH adjusted to 3.5 with Acetic acid.
Result of the trial is depicted in following figure. (Figure 3.4)
In the next trial Mobile phase ratio was set to (70:30 v/v) Methanol: 10 mM Ammonium
Acetate Buffer pH adjusted to 8.5 with Acetic acid. Result of the trial is depicted in
following figure. (Figure 3.5)
Stability of the stock solution as well as working standard solution was checked and it
was found that Tenofovir is stable if kept at 8-16°C for several weeks.
Results of analysis of samples kept under different stress conditions are as follows
* To get desired degradation initially degradation was carried out at room temperature
and if necessary samples were subjected to higher temperature or strength of stress
reagent was increased.
Results of analysis of samples kept under different stress conditions are as follows.
To achieve desired % degradation, drug solution was treated with 0.005N HCl and kept at
room temperature for 1 h 12.99%. The drug underwent acid hydrolysis and produced
three degradation products which were eluted at 2.54, 2.93, 3.13 and 4.93.
(Mobile Phase: Methanol: 10mM Ammonium Acetate Buffer pH - 8.5; 60:40 v/v, Flow
rate 1 mL/min, Wavelength: 260 nm; Column: Kromasil C18 (150 X 4.6 mm, 5µm).
The diode array detector (DAD) spectra provided information in terms of wavelength
maxima and peak purity (from fig. 3.9 to fig 3.11) for degradation product and Tenofovir.
The details are depicted as follows.
Figure 3.9: UV spectra, wavelength maxima and peak purity of Acid degradation
product 1
Figure 3.10: UV spectra, wavelength maxima and peak purity of Acid degradation
product 2
To achieve desired % degradation, drug solution was treated with 0.001N NaOH and kept
at room temperature for 1 h 17.37%drug degradtion observed along with three degradants
at 2.56, 2.94 and 3.76.
The spectra recorded by diode array detector (DAD) in terms of wavelength maxima and
peak purity for alkali degradation products and Tenofovir are depicted as follows. (Figure
3.14 to 3.17)
Figure 3.14: UV spectra, wavelength maxima and peak purity of Alkali degradation
product 1
Figure 3.15: UV spectra, wavelength maxima and peak purity of Alkali degradation
product 2
Figure 3.16: UV spectra, wavelength maxima and peak purity of Alkali degradation
product 3
To obtain desired range of degradation drug was allowed to react with 3% hydrogen
peroxide at room temperature for 1 h and subjected for HPLC analysis. One degradant is
observed under this stress condition at 2.939
(Mobile Phase: Methanol: 10mM Ammonium Acetate Buffer pH - 8.5; 60:40 v/v, Flow
rate 1 ml/min, Wavelength: 260 nm; Column: Kromasil C18 (150 X 4.6 mm, 5µm)
The spectra recorded by diode array detector (DAD) in terms of wavelength maxima and
peak purity for alkali degradation products and Tenofovir are depicted as follows. (Figure
3.20 and 3.21)
The spectra recorded by diode array detector (DAD) in terms of wavelength maxima and
peak purity for Neutral degradation products and Tenofovir are depicted as follows.
(Figure 3.23 and 3.24)
To check photo stability of the drug in solution form, drug solution prepared in methanol
in volumetric flask was exposed to direct sun light up to 5 days. Same way control was
also maintained by wrapping another volumetric solution with aluminum foil. Tenfovir
exposed to ICH recommended dose of light( 1.2million lux h and 200 W h/m2 ) in photo
stability chamber. No additional peak emerged in control sample. Drug was stable to
photolytic condition.
In present study, stress studies were carried out as per trials suggested by software and %
degradation obtained in various stress conditions are as follows:
For acid stress, software suggested following trials. They were practically performed and
% degradation is observed as below.
%Degradation = 86.39+0.87*A+31.57*B-1.44*AB-5.27*A2-34.78*B2
From above equation, factor A (Strength) and factor B (Time) are having positive sign,
factor AB (strength*Time), factor A2 Strength2 and factor B2 Time2
Model Quadratic
P-value 0.0004
F- value 251.73
r2 0.9976
Values of "Prob > F" (p- value) less than 0.0500 indicate model terms are significant. The
Model F-value of 251.73 implies the model is significant. There is only a 0.04% chance
that an F-value this large could occur due to noise. The "Pred R-Squared" of 0.9493 is in
reasonable agreement with the "Adj R-Squared" of 0.9937; i.e. the difference is less than
0.2.
Design-Expert® Software
Factor Coding: Actual
%Degradation (%)
Design points above predicted value
Design points below predicted value
80.11
12.99
100
%Degradation (%)
X1 = A: Strength
X2 = B: Time
80
60
40
20
24 0.1
19.4 0.081
14.8 0.062
10.2 0.043
B: Time (h) 5.6 0.024 A: Strength (N)
1 0.005
Desirability
In the stress study, the response ie; % degradation was desired to be in the range of 5-30
%, and it was inserted in the software as desirability range, A list of solutions was
displayed by software indicating various strength and exposure time and the selected trial
(trail 9) was further practically performed.
Design-Expert® Software
Factor Coding: Actual
Desirability
1.000
1.000
0.000
Desirability = 1.000
Std # 1 Run # 7
X1 = A: Strength = 0.005
X2 = B: Time = 1
1.200
Desirability
1.000 1
0.800
0.600 5.6
0.400
0.200
0.000 10.2
0.1
14.8 B: Time (h)
0.081
0.062 19.4
0.043
0.024
0.005 24
A: Strength (N)
For alkali stress, software suggested following trials. They were practically performed on
HPLC and % degradation in different conditions is obtained as below.
After inserting the above data, following equation was generated by the software
% Degradation = 66.56+4.64*A+28.89*B+0.66*AB-11.65*A2-5.47*B2
From above equation, factor A (Strength) and factor B (Time) are having positive sign,
factor AB (strength*Time), factor A2 Strength2 and factor B2 Time2
Design-Expert® Software
Factor Coding: Actual %degradation (%)
%degradation (%) 24
Design Points
83.46
17.37
19.4
80
X1 = A: Strength
X2 = B: Time
B: Time
14.8
60
10.2
5.6
40
20
1
0.001 0.012 0.023 0.034 0.045 0.056 0.067 0.078 0.089 0.1
A: Strength
Model Quadratic
P-value 0.0001
2
r 0.9996
F value 1687.09
The Model F-value of 1687.09 implies the model is significant. There is only a 0.01%
chance that an F-value this large could occur due to noise. Values of "P > F" less than
0.0500 indicate model terms are significant.
Design-Expert® Software
Factor Coding: Actual
%degradation (%)
Design points above predicted value
Design points below predicted value
83.46
17.37
%degradation (%)
100
X1 = A: Strength
X2 = B: Time
80
60
40
20
24 0.1
0.089
19.4 0.078
14.8 0.067
0.056
10.2 0.045
0.034
B: Time 5.6 0.023
0.012 A: Strength
1 0.001
DESIRABILITY:
In order to get the desired response (%degradation), the desired range (5-30%) is inserted
in the software then the list of solutions is displayed as below
The selected trial was further practically performed and % prediction error is
calculated.
Design-Expert® Software
Factor Coding: Actual
Desirability
1.000
0.000
X1 = A: Strength
X2 = B: Time
1.200
Desirability
1.000
0.800
0.600
0.400
0.200
0.000
1.000 0.1
24 0.089
0.078
19.4 0.067
0.056
14.8
0.045
10.2 0.034
0.023 A: Strength
5.6
B: Time 0.012
1 0.001
c) Oxidation
For oxidation stress, software suggested following trials were practically performed and
% degradation in different conditions is obtained as below.
After inserting the above data following equation was obtained along with the 3D graph.
% Degradtion = 42.94+14.94*A+10.35*B-2.65*AB+5.39*A2-7.20*B2
From above equation, factor A (Strength) and factor B (Time) are having positive sign,
factor AB (strength*Time), factor A2 Strength2 and factor B2 Time2 , while interaction
between A&B has antagonistic effect . These effects can be well interpreted with the
help of response surface graphs.
Design-Expert® Software
Factor Coding: Actual % degradation (%)
% degradation (%) 24
Design Points
63
12.97
19.4
X1 = A: Strength
X2 = B: Time
60
B: Time
14.8
40 50
10.2
30
5.6
20
1
3 6 9 12 15 18 21 24 27 30
A: Strength (%)
Values of "Prob > F" (p- value) less than 0.0500 indicate model terms are significant
Model Quadratic
P-value 0.0007
r2 0.9964
F value 165.25
The Model F-value of 165.25 implies the model is significant. There is only a 0.07%
chance that an F-value this large could occur due to noise. Values of "Prob > F" less than
0.0500 indicate model terms are significant.
Design-Expert® Software
Factor Coding: Actual
% degradation (%)
Design points above predicted value
Design points below predicted value
% degradation (%)
63
12.97
70
X1 = A: Strength
X2 = B: Time 60
50
40
30
20
10
24 30
27
19.4 24
14.8 21
18
10.2 15
12
B: Time 5.6 9 A: Strength (%)
6
1 3
DESIRABILITY:
The selected solution was trialed on RP-HPLC and following values were observed,
along with graphical representation.
Design-Expert® Software
Factor Coding: Actual
Desirability
1.000
0.000
Desirability = 0.000
Std # 1 Run # 5
X1 = A: Strength = 3
X2 = B: Time = 1
0.000
Desirability
0.200
0.400
0.600
0.800
1.000 24
30 19.4
27 24 14.8
21 18 10.2
15
1.000 12 9 6 5.6 B: Time
A: Strength (%) 3 1
Linearity of Tenofovir
350000
300000
f(x) = 1472.85571428571 x + 135353.666666667
250000 R² = 0.99955125075308
Peak Area
200000
150000
100000
50000
0
0 20 40 60 80 100 120 140
Concentrations ( µ/mL)
Linearity range of Tenofovir was observed at 20, 40, 60, 80, 100 and 120 µ /mL. The
equation of linearity for Tenofovir y = 1472x+135354.Correlation coefficient (r2) was
found to be 0.999.
3.7.2 Precision
It was found that the precision results were found satisfactory with respect to % Relative
standard deviation (% RSD) for all levels which were within limit (NMT than 2). The
developed method was precise for determination of Tenofovir.
3.7.3 Recovery
The mixture of stressed sample spiked with the drug concentrations of 80, 100 1nd 120
µg/mL. The % Recovery of Tenofovir was found to be 98.65 %. The result was found
satisfactory % mean recovery was within accepted range (98% to 120%).
3.7.4 Robustness
As per ICH guideline, small but deliberate variations, by altering the pH, wavelength
range or flow rate were made to check the capacity of the method. The method was found
to be unaffected by changing the pH. It also does not show any significant changes due to
alteration of flow rate and wavelength range.
258 nm 283170
Wavelength 260 nm 286452
287026 1.45
262 nm
291456
0.9 mL/min 286458
8.4 283654
pH 8.5
286351 284553 0.54
8.6 283654
The method was found to be unaffected by changing the wavelength, flow rate and pH.
The % RSD within limit (NMT 2.0). The Developed method was found to be robust.
Limit of detection and limit of quantification were calculated by using slope method (3.3
x SD/S and 10 x SD/S respectively) and found to be 0.463 μg/mLand 1.405 μg/mL
respectively.
Parameter Tenofovir
Peak area (100µg/mL) 283170
No. of Theoretical plate 4066.886
Tailing Factor 1.212