You are on page 1of 7

CHAPTER 38

DESIGNING FOR RADIATION PROTECTION

RADIOGRAPHIC PROTECTION FEATURES Beam Alignment


 Each radiographic tube should be provided
Protective X-ray Tube Housing with a mechanism
 It must be contained within protective o Purpose: to ensure proper alignment
housing of the x-ray beam & the IR
 It reduces leakage radiation
Filtration
Leakage radiation must be less than 100  Total Filtration:
mR/hr at a distance of 1 m from the protective o 2.5 mm Al – operated above 70 kVp
housing! o 1.5 mm Al – operated b/n 50-70 kVp
o 0.5 mm Al – operated below 50 kVp
Control Panel  HVL: it measures filtration
 It must indicate the condition of exposure
 It must positively indicate when the x-ray Mammography
tube is energized  Total Filtration: 30 μm Mo or 60 μm Rh
o kVp & mA indicators
Reproducibility
X-ray beam on must be positively & clearly  For any radiographic technique, the output
indicated to the radiologic technologist! radiation intensity should be constant form
one exposure to another
Source-to-Image Receptor Distance Indicator
 It must be provided The variation in x-ray intensity should not
o Tape measure attached to the tube exceed 5%!
housing
o Lasers Linearity
 Ability of a radiographic unit to produce
The SID indicator must be accurate to within constant radiation output for various
2% of the indicated SID! combinations of mA & exposure time
 Radiation Intensity: mR/mAs
Collimation
 Light-localized, variable-aperture The maximum acceptable variation in
rectangular collimators should be provided linearity is 10% from one mA station to an
 Cones & Diaphragms: for special adjacent mA station!
examination
Operator Shield
The x-ray beam & the light beam must  RT may be in the exposure room during
coincide to within 2% of the SID! exposure, but only if protective apparel is
worn
Positive-Beam Limitation (PBL)  Exposure Control: fixed, not a long cord
125
 Automatic, light-localized, variable-aperture
collimators Mobile X-ray Imaging System
 Lead Apron: should be assigned
Page

The PBL must be accurate to within 2% of  Exposure Switch: at least 2 m from x-ray
SID! tube during exposure

STEWART C. BUSHONG SUMMARIZED BY: MEYNARD Y. CASTRO


CHAPTER 38
DESIGNING FOR RADIATION PROTECTION

FLUOROSCOPIC PROTECTION FEATURES Cumulative Timer


 It produces an audible sound when the
Source-to-Skin Distance fluoroscopic time has exceeded 5 minutes
 Increased SSD: reduces entrance skin  Designed to ensure that the radiologist is
exposure (ESE) aware of the relative beam-on time during
 Stationary Fluoroscopes: not less than 38 each procedure
cm
 Mobile Fluoroscopes: less than 30 cm Dose Area Product (DAP)
 A quantity that reflects not only the dose but
Primary Protective Barrier also the volume of tissue irradiated
 Fluoroscopic IR Assembly:  Expressed in: R-cm2
o It serves as a primary protective  Better indicator of risk than dose
barrier  X-ray Beam Intensity
o It must be 2 mm Pb equivalent o Table Top: <2.1 R/min (each mA
o It must be coupled with the x-ray operation at 80 kVp)
tube & interlocked o No Optional High-Level Control:
<10 R/min
Filtration o With Optional High-Level Control:
 Total Filtration: 2.5 mm Al 20 R/min
 HVL should be measured when filtration is o Cineradiography/Videography: no
unknown limit
 Increased Field Size: increases DAP & risk
Collimation
 Collimators must be adjusted DESIGN OF PROTECTIVE BARRIERS
o Rationale: unexposed border is
visible on the image monitor Type of Radiation
 Automatic Collimators: unexposed border  Primary Radiation
should be visible at all heights above the  Secondary Radiation
tabletop
Primary Radiation
Exposure Control  The useful beam
 Dead man type  The most intense, hazardous & difficult to
 Controls: convetional foot pedal & pressure shield
switch
Primary Barrier
Bucky Slot Cover  Any wall to which the useful beam can be
 Bucky Tray: moved at the end of the table directed
during fluoroscopy  Examples: lead bonded to sheet rock (lb/ft2)
o Result: leaving a 5 cm opening in the or wood paneling
126
side of the table
 Covered with 0.25 mm Pb Secondary Radiation
 Scatter & leakage radiation
Page

Protective Curtain
 It should be position between fluoroscopist Scatter Radiation
& patient
 Equivalent: 0.25 mm Pb

STEWART C. BUSHONG SUMMARIZED BY: MEYNARD Y. CASTRO


CHAPTER 38
DESIGNING FOR RADIATION PROTECTION

 It results when the useful beam intercepts  Controlled Area: an area occupied primarily
any objects causing some x-rays to be by radiology personnel & patients
scattered o Limits: 100 mrem/week
o Based on the annual recommended
The intensity of scatter radiation 1 m from the occupational dose limit of 5000
patient is approximately 0.1% of the intensity mrem/yr or 50 mSv/yr
of the useful beam at the patient!  Uncontrolled Area: an area that can be
occupied by anyone
Patient o Limit: 2 mrem/week
 The single most important scattering object o Based on the annual recommended
dose limit for the public of 100
Leakage Radiation mrem/yr
 Radiation emitted from the x-ray tube  Protective Barrier: it should ensure that no
housing in all directions other than that of individual will receive more than 2.5
the useful beam mrem/hr
 Limit: 100 mR/hr at 1 m
LEVELS OF OCCUPANCY THAT MAY BE
Secondary Protective Barrier ADJACENT TO X-RAY ROOM AS
 Designed to shield areas from secondary SUGGESTED BY THE NCRP
radiation Occupancy Area
 Less thick than primary radiation Work areas, living
 Composition: gypsum board, glass or lead quarters, children’s play
Full
acrylic areas & occupied space
 Example: operating console barrier & in nearby buildings
ceiling Corridors, restrooms &
Frequent
patient rooms
Radiologic technologists receive most of their Waiting rooms,
occupational radiation exposure during stairways, unattended
Occasional
fluoroscopy! elevators, janitors’
closets & outside area
Factors That Affect Barrier Thickness
 Distance, Occupancy, Control, Workload, Workload (W)
Use Factor & kVp  Product of the maximum mA & the number
of x-ray examinations performed per week
Distance  Expressed in: mAmin/week
 It depends on the distance between the  Busy, General Purpose X-ray Room: 500
source of radiation & the barrier mAmin/week
 Private Office: <100 mAmin/week
Occupancy
127
 The use of the area that is being protected Use Factor (U)
 Rarely Occupied: less shielding required  The percentage of time during which the x-
ray beam is on & directed toward a
Page

Time Occupancy Factor (T) particular protective barrier


 Length of time that the area being protected  Walls: ¼
is used  Floor: 1
 Room for Chest Radiography: 1
Control
STEWART C. BUSHONG SUMMARIZED BY: MEYNARD Y. CASTRO
CHAPTER 38
DESIGNING FOR RADIATION PROTECTION

 Others: 0  Characteristics: limited range, sensitive &


energy dependent
The use factor for secondary barriers is
 Uses: personnel monitoring & emulsion
always 1!
kVp imaging
 It is used as a measure of penetrability Two Principal Applications of Film
 General Radiography: 100 kVp  The making of a radiograph
 Mammography: 30 kVp  The radiation monitoring of personnel (film
badge)
RADIATION DETECTION &
MEASUREMENT Gas-Filled Radiation Detector
 It is used widely as a device to measure
Radiation Detection Instrument radiation intensity & to detect radioactive
 Designed to detect or measure radiation contamination
 Operate in the pulse or rate mode  Three Types: ionization chamber,
o It is used to indicate the presence of proportional counter & Geiger-Muller
radiation counter

Pulse Mode The ionization of gas is the basis for gas-filled


 The presence of radiation is indicated by radiation detectors!
ticking, chirping or beeping sound
Ionization Chamber
Rate Mode  The instrument of choice for measuring
 The instrument response is expressed in radiation intensity
mR/hr or R/hr  Characteristics: wide range, accurate &
portable
Integrated Mode  Uses: survey for radiation levels 1 mR/hr
 It is used in instruments designed to measure
the intensity of radiation High sensitivity means that an instrument can
detect very low radiation intensities!
Dosimetry
 The practice of measuring the intensity of High Level of Accuracy
radiation  It means that an instrument can detect &
precisely measure the intensity of radiation
Dosimeters field
 The radiation-measuring devices
Instrument Accuracy
Radiation Detection & Measuring Device  Controlled by the overall electronic design
 Photographic, Ionization Chamber, of the device
Proportional Counter, Geiger-Muller
128
Counter, Thermoluminescence Dosimetry, Dose Calibrator
Optically-Stimulated Luminescence  Another application of precision ion
Dosimetry & Scintillation Detection chamber
Page

 It is used in nuclear medicine laboratories


Photographic Emulsion for the assay of radioactive material
 The earliest radiation detection device
Proportional Counter

STEWART C. BUSHONG SUMMARIZED BY: MEYNARD Y. CASTRO


CHAPTER 38
DESIGNING FOR RADIATION PROTECTION

 It has the ability to distinguish between  An imaging device used in nuclear imaging
alpha & beta radiation  It is also used in CT & DR imaging system
 Characteristics: laboratory equipment,
accurate & sensitive
 Uses: assay of small quantities of Liquid Scintillation Detectors
radionuclides  It is used frequently to detect low-energy
beta emission from carbon-14 & tritium
Geiger-Muller Counter
 It is used for contamination control in Types of Scintillation Phosphors
nuclear medicine laboratories  Thallium-activated sodium iodide (NaI: Tl)
 Characteristics: limited to 100 mR/hr & o Incorporated into gamma camera
portable  Thallium-activated cesium iodide (CsI: Tl)
 Uses: survey for low radiation levels & o Incorporated into image-intensifier
radioactive contamination tubes as the input phosphor & into
flat panel DR image receptors
Resolving Time
 The minimum time between ionizations that The Scintillation Detector Assembly
can be detected  Aluminum Seal, Window, Photocathode,
Dynodes, Glass Envelope, Collector & Base
Quenching Agent
 Added to the filling gas of the Geiger Aluminum Seal
counter to enable the chamber to return to its  It allows the light flash to be reflected
original condition internally to the window
 It also necessary to seal the crystal
Thermoluminescence Dosimetry (TLD) hermetically
 Characteristics: wide range, accurate &
sensitive Window
 Uses: personnel monitoring, stationary &  The portion of the glass envelope that is
area monitoring coupled to the scintillation crystal

Optically Stimulated Luminescence Dosimetry Hermetic Seal


 Characteristics: wide range, accurate &  One that prevents the crystal from coming
sensitive into contact with air or moisture
 Uses: newest personnel monitoring device
Hygroscopic
Scintillation Detection  It absorbs moisture
 Basis for the gamma camera
 It is used in the detectors arrays of CT Photomultiplier (PM) Tube
imaging system  It converts light flashes from the scintillator
129

 It is used as IR in digital imaging system into an electronic signal of pulses


 Characteristics: limited range, very
Glass Envelope
Page

sensitive, & stationary or portable


instruments  It provides structural support for the internal
 Uses: photon spectroscopy & imaging elements & maintains the vacuum inside the
tube
Gamma Camera
Photocathode
STEWART C. BUSHONG SUMMARIZED BY: MEYNARD Y. CASTRO
CHAPTER 38
DESIGNING FOR RADIATION PROTECTION

 A device that emits electrons when  The emission of light by a thermally


illuminated stimulated crystal following irradiation
 Composition: cesium, antimony & bismuth  Step Process:
o Exposure to ionizing radiation
Photoemission o Subsequent heating
 A process wherein electrons are emitted o Measurement of the intensity of
from the photocathode emitted light
 Materials: Lithium Fluoride, Calcium
Dynodes Fluoride, Lithium Borate & Calcium Sulfate
 The first series of plate-like elements  Advantages:
 Function: o Size
o Amplify the electron pulse through o Reusable
secondary electron emission o Responds proportionately to dose
o Rugged
Dynode Gain
 The ratio of secondary electrons to incident Lithium Fluoride
electrons  Commonly used
 Symbol: LiF
Photomultiplier Tube Gain  Density 103 (kg/m3): 2.64
 The dynode gain raised to the power of the  Effective Atomic Number: 8.2
number of dynodes  Temperature of Main Peak (oC): 195
 Formula: PM Tube = gn  Principal Use: patient & personnel dose
 Characteristic: sensitive
Collecting Electrode/Collector o as low as 5 mrad
 The last plate-like element of the PM tube o >10 rad
 Functions:  Accuracy: better than 5%
o Absorbs the electron pulse from the
last dynode Lithium fluoride is a nearly tissue-equivalent
o Conducts it to the preamplifier radiation dosimeter!

Preamplifier Calcium Fluoride Activated with Manganese


 It provides an initial state of pulse  Symbol: CaF2:Mn
amplification  Density 103 (kg/m3): 3.18
 Effective Atomic Number: 16.3
Base  Temperature of Main Peak (oC): 260
 A structure that provides support for the  Principal Use: environmental monitoring
glass envelope & internal structures  Characteristic: more sensitive
o < 1 mrad
The size of the electron pulse is proportional to
130
the energy absorbed by the crystal from the Lithium Borate
incident photon!  Symbol: Li2B4O7:Mn
Page

 Density 103 (kg/m3): 2.5


Gamma Spectrometry
 Effetive Atomic Number: 7.4
 It uses pulse height analysis
 Temperature of Main Peak (oC): 200
Thermoluminescence Dosimetry (TLD)  Principal Use: research

STEWART C. BUSHONG SUMMARIZED BY: MEYNARD Y. CASTRO


CHAPTER 38
DESIGNING FOR RADIATION PROTECTION

Calcium Sulfate
 Symbol: CaSO4:Dy
 Density 103 (kg/m3): 2.61
 Effetive Atomic Number: 15.3
 Temperature of Main Peak (oC): 220
 Principal Use: environmental monitoring

TLD Analyzer
 Electronic instruments that are designed to:
o Analyzed measure the height of the glow
curve
o The area under the curve
o Relate this to exposure or dose through a
convention factor

Glow Curve
 Graph that shows the relationship of light
output to temperature change

Optically Stimulated Luminescence Dosimetry


(OSL)
 Developed by Laundauer in late 1990s
 Material: Aluminum oxide (Al2O3)
 Step Process:
o Exposure to ionizing radiation
o Laser illumination
o Measurement of the intensity of
stimulated light emission
 Advantages Over TLD:
o More sensitive – 1 mrad
o Reanalysis
 Confirmation of dose
 Qualitative information about
exposure conditions
o Wide dynamic range
o Excellent long-term stability
131
Page

STEWART C. BUSHONG SUMMARIZED BY: MEYNARD Y. CASTRO

You might also like