Haskris chiller specifications 105830 Feedback
TC: WARNING: paper copies are uncontrolled Date: 4‐Dec‐2014 Rev: A

Haskris chiller specifications


Model R175EA-BHKL. (R5) (187/253V-1Ø-60Hz); (180/242V-1Ø-50Hz)
FEI Part Number: 9432.909.94311
DEPTH = 26" (66 cm)
TECHNICAL SPECIFICATIONS

65 F
1. Power Requirements: 187/253V-1Ø-60Hz ON
SET 12 OFF
2. F.L.A. Rating: 14.9A *
3. Recommended Fuse Size: 20A (time delay)
4. Maximum Ambient Air Temp: +90ºF
5. Minimum Ambient Air Temp: +55ºF

TECHNICAL SPECIFICATIONS
34" (86 cm)
1. Power Requirements: 180/242V-1Ø-50Hz
2. F.L.A. Rating: 15.9A
3. Recommended Fuse Size: 20A (time delay)
4. Maximum Ambient Air Temp: +27ºC
5. Minimum Ambient Air Temp: +12ºC Air Cooled Condenser
NOT DRAWN
(Air Discharge) TO SCALE

FRONT
VIEW Levelling Legs, Adjustable
From 1/2" To 2"
27" (68 cm)

DEPTH = 26" (66 cm)


REAR Supply and return connections
VIEW are ˇ” hose barbs

2.25" (6 cm)

SUPPLY 1/2” hose barb

RETURN 1/2 ‘ hose barb

25" (64 cm)


23" (58 cm)
Air Cooled Condenser NOT DRAWN
TO SCALE
(Air Intake)

R175EA-BHKL(R5).vsd

Figure 1: Model air 175EA BHKL(R5)

Page 1 of 4 FEI ‐ confidential and proprietary information


Doc. Number: Provide Field
Haskris chiller specifications 105830 Feedback
TC: WARNING: paper copies are uncontrolled Date: 4‐Dec‐2014 Rev: A

HASKRIS CO. MODEL R175EA


FEI Part Number: 9432 909 94311
Installed Options

Option (B): High Temperature Pump Cutout


Option (H): Electronic Close Temperature Control
Option (K): Remote Interlock Relay (24 VDC) & OFF Delay Timer
Option (L): 5 Micron Filter
Option (R5): Turbine Pump (Reduced Pulsation)

Electrical Requirements

Power: 187/253V-1 phase 60Hz + or - 10% or 180/242V-1 phase 50Hz + or - 10%


Full Load Amps: 14.9A or 15.9A
Recommended Fusing: 20A, time delay

Technical Features

Cooling Capacity (at 65ºF & 60Hz): 18.750 Btu/hr 5.500 Watts
Cooling Capacity (at 18ºC & 50Hz): 16.100 Btu/hr 4.700 Watts
Water Pump Capacity (60 Hz): 5.2 GPM (1181 LPH) @ 58 psi
Water Pump Capacity (50 Hz): 3.2 GPM (727 LPH) @ 4 Bar
Water Reservoir Volume: 9.0 Gallons (34 Litres)

Temperature Stability: Plus or minus 0.1ºC


Refrigerant: R404a
Unit Weight: 435 lbs (200 Kg)
Unit Dimensions: 27" (68 cm) W x 26" (66 cm) D x 34" (86 cm) H.
Coolant Required: Clean, potable, distilled water

Condenser Air Requirements

Acceptable Ambient Air Temperatures: +55×F To +80×F (+12×C To +27×C)

Condenser Air Flow: 772 Cu ft/min (21.9 Cu m/min)

Heat Rejection Into Room*: 24.375 Btu/hr 7.150 Watts


* Model R175EA with air cooled condenser will dissipate heat into the surrounding air. The
amount of heat equals the heat load of the instrument plus 30% for the motor heat of the Haskris.

Location

Must be installed in a clean, indoor environment.

We suggest access to the: TOP for routine maintenance, such as changing water in the reservoir.
FRONT for visibility of controls and readouts.
ONE SIDE for convenient servicing should a spare part be required.

Clearance Required (for adequate air flow across condenser): Rear: 6 in (.15 m)
Front: 3 ft (1 m)

R175EA-BHKL(R5).vsd

Figure 2: Haskris Model air R175EA BHKL(R5)

Page 2 of 4 FEI ‐ confidential and proprietary information