Designation: E2941 − 21
1. Scope Sections
Nitric Acid Microwave Digestion 7 to 14
1.1 These practices cover the digestion of ores and related Four-Acid Total Digestion 15 to 21
metallurgical materials, such as mine soil, waste rock and 1.3 The values stated SI units are to be regarded as the
tailings, for subsequent determination of acid-extractable con- standard. No other units of measurements are included in this
tents of certain elements by such solution analytical techniques standard.
as atomic absorption spectrometry (AAS), inductively coupled
plasma atomic emission spectrometry (ICP-AES) (see Test 1.4 This standard does not purport to address all of the
Method D1976), and inductively coupled plasma mass spec- safety concerns, if any, associated with its use. It is the
trometry (ICP-MS) (see Test Method D5673). responsibility of the user of this standard to establish appro-
1.1.1 Contents of aluminum, antimony, arsenic, barium, priate safety, health, and environmental practices and deter-
beryllium, bismuth, boron, cadmium, calcium, chromium, mine the applicability of regulatory limitations prior to use.
cobalt, copper, gallium, iron, lead, lithium, magnesium, For specific hazard statements, see Sections 11 and 20.
manganese, mercury, molybdenum, nickel, phosphorus, 1.5 This international standard was developed in accor-
potassium, scandium, selenium, silver, sodium, strontium, dance with internationally recognized principles on standard-
thallium, tin, titanium, vanadium and zinc can be extracted ization established in the Decision on Principles for the
from ores and related metallurgical materials for determination Development of International Standards, Guides and Recom-
by analytical methods for elements in solution. Other elements mendations issued by the World Trade Organization Technical
may be determined from extracts produced using this practice. Barriers to Trade (TBT) Committee.
1.1.2 Actual element quantification in digested solutions
2. Referenced Documents
can be accomplished by following the various test methods
under other appropriate ASTM standards for element(s) of 2.1 ASTM Standards:2
interest in solution. D1193 Specification for Reagent Water
1.1.3 The detection limit and linear content range for each D1976 Test Method for Elements in Water by Inductively-
element is dependent on the atomic absorption, mass spectrom- Coupled Plasma Atomic Emission Spectroscopy
etry or emission spectrometric technique employed and may be D5258 Practice for Acid-Extraction of Elements from Sedi-
found in the manual accompanying the instrument used or ments Using Closed Vessel Microwave Heating
ASTM standard method for analysis of the solutions. Consider D5673 Test Method for Elements in Water by Inductively
the dilution factor in content calculations due to digestion and Coupled Plasma—Mass Spectrometry
dilution of solid samples. E50 Practices for Apparatus, Reagents, and Safety Consid-
1.1.4 The extent of extraction of elements from ores and erations for Chemical Analysis of Metals, Ores, and
related metallurgical materials by these practices is dependent Related Materials
upon the physical and mineralogical characteristics of the E135 Terminology Relating to Analytical Chemistry for
prepared sample and the digestion practice used. Metals, Ores, and Related Materials
E882 Guide for Accountability and Quality Control in the
1.2 The digestion practices appear in the following order: Chemical Analysis Laboratory
1
These practices are under the jurisdiction of ASTM Committee E01 on
Analytical Chemistry for Metals, Ores, and Related Materials and are the direct
responsibility of Subcommittee E01.02 on Ores, Concentrates, and Related Metal-
2
lurgical Materials. For referenced ASTM standards, visit the ASTM website, www.astm.org, or
Current edition approved Oct. 1, 2021. Published October 2021. Originally contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
approved in 2013. Last previous edition approved in 2014 as E2941 – 14. DOI: Standards volume information, refer to the standard’s Document Summary page on
10.1520/E2941-21. the ASTM website.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
15.2 This practice provides a sample suitable for analysis by 20.3 HClO4 is a strong oxidizer that poses an explosive
AAS,ICP-AES, or ICPMS. property on contact with organic substances. Only fume hoods
certified for the use of HClO4 should be used.
16. Significance and Use
20.4 HF is a known contact hazard and is highly toxic.
16.1 This digestion method is applicable to solid, pulverized
20.5 All operations should be performed in a laboratory
samples for determination of total metals by AAS or ICP-AES.
fume hood. A laboratory coat, safety glasses, and protective
Analytes that fall below the established detection limit of the
gloves should be worn at all times. As with all tests, the
ICP-AES can be easily determined by ICP-MS following a
primary responsibility for safety lies with the analyst. Consult
pre-dilution of the solution.
the appropriate SDS and the chemical hygiene plan, if
16.2 Alternative digestion practices are more appropriate available, for further details.
for determination of silicon, arsenic, antimony, and selenium,
due to volatile losses during acid fuming. 21. Procedure
17. Interferences 21.1 Weigh 0.5 g 6 0.01 g of sample into a 250-mL
fluoropolymer beaker.
17.1 No interferences to the digestion of ores and related
metallurgical materials using four-acid total digestion have 21.2 In a fume hood, dispense the acids in the following
been observed. order: 10 mL water, 10 mL HNO3, 10 mL HClO4, 5 mL HF.
17.2 Silica is lost during the volatilization of the acid fumes 21.3 Swirl to mix acids and place the beakers on the hot
and cannot be determined by this practice; arsenic, antimony, plate and cover with fluoropolymer watch glasses for 30 min.
and selenium may also be lost during acid fuming. 21.4 Rinse watch glasses with water into the beaker and
remove watch glasses. Continue to digest to dryness. Heavy
18. Apparatus white fumes will evolve from the mixture during the process.
18.1 Fluoropolymer Beakers, 250-ml capacity. Observe closely the samples as they reduce in volume. If the
18.2 Fluoropolymer Watch Glass, for 250-mL capacity beakers are left unsupervised after drying, they will melt.
beaker. 21.5 Remove the beakers from the hotplate and allow to
18.3 Hot Plate, capable of digesting samples in fluoropoly- cool.
mer beakers to dryness. 21.6 Add 10 mL water and then 10 mL HCl. Warning—
Water must be dispensed before the HCl or an explosion may
19. Reagents and Materials occur or acid may (violently) spatter due to an exothermic
19.1 d-Tartaric Acid. reaction with water.
19.2 d-Tartaric Acid Solution (10 % m/v)—In a 100-mL 21.7 Return the beakers to the hotplate and reheat uncov-
volumetric flask, dissolve 10.0 g d-tartaric acid crystals in ~30 ered for about 5 min, only to re-dissolve the residue.
mL water. Bring to volume and mix thoroughly. NOTE 1—Allowing the samples to reheat beyond the allotted time will
cause the volatilization of HCl and will result in loss of Ag through
19.3 Polypropylene Tubes, 10-mL and 50-mL capacity, with precipitation.
volume markings.
21.8 While still warm, transfer sample into a labeled 50 mL
20. Hazards polyethylene tube. Cool to ambient temperature and add
0.5 mL d-Tartaric Acid Solution (19.2) and bring to a final
20.1 For hazards to be observed in the use of reagents and volume of 50 mL with water. Cap tightly and shake to mix.
apparatus in this practice, refer to Practices E50. NOTE 2—d-Tartaric Acid Solution (19.2) is for stabilization of Sb and
20.2 This method involves the use of acids and strong Bi.
oxidizers at elevated temperatures. Concentrated HNO3 and 21.9 Allow samples to settle overnight, or, if analysis is
HCl are highly corrosive and both evolve irritating fumes. needed immediately, centrifuge an aliquot.
ANNEX
(Mandatory Information)
A1. PROCEDURE TO DETERMINE DELIVERED POWER OF MICROWAVE OVEN AT 100 % INSTRUMENT POWER
A1.1 Remove the turntable, drive lug, and all vessels from A1.9 Calculate the delivered power as follows:
the instrument cavity.
P 5 ~ T f 2 T i ! 3 35 (A1.1)
A1.2 Adjust the instrument cavity exhaust to minimum where:
airflow.
P = delivered power, watts,
A1.3 Program the instrument for 4 min time and 100 % Tf = final water temperature, °C, and
power. Ti = initial water temperature, °C.
and
A1.4 Transfer 2000 mL 6 2 mL of room temperature (19 °C
to 25 °C) water into a 2-L polypropylene beaker. 4.2 3 1.0 3 M
35 5 (A1.2)
t
A1.5 Measure and record the initial water temperature, Ti,
to the nearest 0.1 °C. where:
4.2 = conversion factor for thermochemical calories to watts,
A1.6 Place the beaker in the right front corner of the 1.0 = heat capacity of water, calories/g per °C,
instrument cavity. M = mass of water, g, and
t = heating time, s.
A1.7 Heat the water for the programmed time.
A1.8 When the heating cycle is complete, immediately
remove the beaker from the cavity, thoroughly stir the water to
ensure even heat distribution, and measure the final
temperature, Tf, to the nearest 0.1 °C.
ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.
This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org). Permission rights to photocopy the standard may also be secured from the Copyright Clearance Center, 222
Rosewood Drive, Danvers, MA 01923, Tel: (978) 646-2600; http://www.copyright.com/