MINISCAN IR VISION
OPERATION MANUAL
M-V1.09 SW-V2.4
Part 1 of 2
IMPRINT
Issued by
Updating This instruction manual can be changed without prior notice and without
liability on the part of AMETEK, Inc. or its subsidiaries.
Table of content
Part 1 of 2
1 General 5
1.1 Identification of the instrument ............................................................................................. 5
1.2 Purpose of the documentation ............................................................................................... 5
1.3 Care of the documentation ..................................................................................................... 5
1.4 Liability ................................................................................................................................... 5
1.5 Warranty ................................................................................................................................. 6
2 Safety 7
2.1 Explanation of the symbols and types of note ....................................................................... 8
2.1.1 Safety instructions ....................................................................................................................... 8
2.1.2 Important mandatory instructions .............................................................................................. 9
2.1.3 Information and notes on use ...................................................................................................... 9
2.2 Appropriate use .................................................................................................................... 10
2.3 Reasonably predictable incorrect use ................................................................................. 10
2.4 Safety and protective devices ............................................................................................... 11
2.4.1 Symbols on the instrument........................................................................................................ 11
2.5 Areas of responsibility .......................................................................................................... 12
2.5.1 Areas of responsibility on the part of the operator ................................................................... 12
2.5.2 Areas of responsibility on the part of personnel....................................................................... 12
2.5.3 Personal protective equipment ................................................................................................. 12
2.6 Safety instructions ................................................................................................................ 13
2.6.1 Setting up.................................................................................................................................... 13
2.6.2 Mains connection ....................................................................................................................... 13
2.6.3 Operation .................................................................................................................................... 14
3 Transportation 15
4 Unpacking 16
5 Description 19
5.1 Front ..................................................................................................................................... 19
5.2 Outlet side............................................................................................................................. 20
5.3 Rear side ............................................................................................................................... 21
6 Installation 22
6.1 DC/AC Power converter ....................................................................................................... 23
6.2 External keyboard ................................................................................................................ 23
6.3 Printer................................................................................................................................... 23
6.4 LAN interface........................................................................................................................ 23
MINISCAN IR VISION
Revision / Issue: M-V1.09 SW-V2.4
© 2020 AMETEK, Inc. All rights reserved.
General
8 Operation 25
8.1 Common Description............................................................................................................ 25
8.1.1 User Interface............................................................................................................................. 25
8.1.2 Login ........................................................................................................................................... 25
8.1.3 Structure of the User Interface ................................................................................................. 26
8.1.4 Status .......................................................................................................................................... 27
8.1.5 Important Buttons ...................................................................................................................... 28
8.1.6 Navigation Bar ............................................................................................................................ 29
8.1.7 Keyboard and Numpad............................................................................................................... 29
[Link] Chinese Keyboard .................................................................................................................. 30
8.2 Work with the IR VISION ....................................................................................................... 31
8.2.1 Measurement ............................................................................................................................. 31
[Link] Switch on the instrument ....................................................................................................... 31
[Link] Define measurement settings ............................................................................................... 31
[Link] Options ................................................................................................................................... 37
[Link] Measurement Warnings......................................................................................................... 42
8.2.2 Calibration: Analyzer Preparation for Measurements ............................................................. 43
[Link] Setting up the IR VISION for Calibration Measurements ....................................................... 43
8.2.3 Calibration of methods (Concentration) .................................................................................... 44
[Link] Switch on the instrument ....................................................................................................... 44
[Link] Preparation of calibration samples ....................................................................................... 44
[Link] Measure Calibration Samples................................................................................................ 44
[Link] Gain and Offset ....................................................................................................................... 45
[Link] Substance Calibration ............................................................................................................ 46
8.2.4 Calibration of new samples (Prediction) ................................................................................... 48
[Link] Switch on the instrument ....................................................................................................... 48
[Link] Calibration.............................................................................................................................. 48
8.2.5 Group Assignment for Methods ................................................................................................. 51
8.2.6 Methods Training........................................................................................................................ 53
[Link] Train Selected Methods ......................................................................................................... 53
[Link] Train All Methods ................................................................................................................... 54
8.2.7 How to improve results for predicted properties ..................................................................... 55
[Link] Step 1: Change a method's group assignment ...................................................................... 55
[Link] Step 2: Add new calibration data............................................................................................ 57
[Link] Step 3: Create a new method ................................................................................................. 58
8.2.8 Calibration of Density Meter ...................................................................................................... 59
9 Instrument Screens 63
9.1 Main Menu ............................................................................................................................ 63
9.2 Measurement Menu ............................................................................................................. 64
9.2.1 Parameter Menu ........................................................................................................................ 65
9.2.2 Measurement Settings ............................................................................................................... 66
9.3 Rinsing .................................................................................................................................. 68
9.4 Overview ............................................................................................................................... 69
9.4.1 Columns...................................................................................................................................... 70
9.4.2 Results Details ........................................................................................................................... 71
9.4.3 Density Units............................................................................................................................... 72
9.4.4 Diagrams .................................................................................................................................... 72
9.4.5 Goodness Parameter and Quality Parameter (for PLS methods and cluster analysis) .......... 73
9.4.6 List of Parameters ..................................................................................................................... 75
9.5 Graphics ................................................................................................................................ 79
MINISCAN IR VISION
Revision / Issue: M-V1.09 SW-V2.4
© 2020 AMETEK, Inc. All rights reserved.
General
10 Cleaning 107
10.1 Touch Screen ...................................................................................................................... 107
10.2 Housing ............................................................................................................................... 107
10.3 Rinsing the instrument....................................................................................................... 107
10.3.1 Rinsing with cleaning mixture ................................................................................................. 107
11 Maintenance 108
11.1 Oil filling.............................................................................................................................. 108
11.2 CO2 peak too high – spectrometer evacuation with pump................................................. 109
MINISCAN IR VISION
Revision / Issue: M-V1.09 SW-V2.4
© 2020 AMETEK, Inc. All rights reserved.
Dreckstext damit ers druckt der
word-scheissdreck
General
1 General
1.4 Liability
GRABNER INSTRUMENTS accepts no liability if:
The instrument is used for tasks other than the intended tasks
Damage is caused by incorrect operation
Damage is caused by inadequate or incorrect maintenance
Original spare parts approved by GRABNER INSTRUMENTS are not used
Unauthorized or improper modifications and maintenance operations are
carried out (without prior discussion with GRABNER INSTRUMENTS)
Damage is caused because documents are not kept up to date
5
General
1.5 Warranty
GRABNER INSTRUMENTS assumes no responsibility for damage caused if:
The instrument is used for tasks other than the intended tasks
Damage is caused by incorrect operation
Damage is caused by inadequate or incorrect maintenance
Original spare parts approved by GRABNER INSTRUMENTS are not used
Unauthorized or improper modifications and maintenance operations are
carried out (without prior discussion with GRABNER INSTRUMENTS)
Damage is caused because documents are not kept up to date
Unsuitable operating fluids are used by the owner
Damage is caused by incorrect set up of the instrument
Damage is caused by improper mounting of the instrument
Damage is caused by chemical, electrochemical or electrical influences
6
Safety
2 Safety
NOTICE
ALWAYS INFORM anyone who operates the instrument.
Make sure that this section has been carefully read and understood by anyone who
operates the instrument.
This section
defines the appropriate use of the instrument.
contains generally applicable and observable safety instructions and
regulations.
explains the meaning of the symbols and pictograms which are used in this
manual and on the signs on the instrument itself.
provides information on the required protective equipment and the
requirements relating to operating and maintenance personnel.
provides information on the dangers and residual risks which can occur even
when the instrument is used properly.
Special handling and situational safety instructions are provided in the following
sections of this instruction manual, as and when these are applicable.
Latest technology
The instrument conforms to the latest technology available at the time it was
supplied and was constructed in accordance with recognized safety regulations.
The standards adhered to in the construction of the instrument are listed in the
declaration of conformity.
Nevertheless, the instrument can represent a source of danger if the safety
instructions in this document are not followed and implemented.
7
Safety
SIGNAL WORD
Description of the hazardous situation, problem
Remedy and measures to be taken
General symbols:
8
Safety
Signal word
Instruction text for an important activity or measure to be taken.
Signal word
Information text or recommended usage
9
Safety
NOTICE
The instrument is not intended for use by persons (including children) with reduced
physical, sensory or mental capabilities, or lack of experience and knowledge, unless
they have been given supervision or instruction concerning use of the instrument by a
person responsible for their safety.
Ambient conditions
Air temperature in operation: 5°C to +40°C (+41°F to +113°F)
Relative humidity in operation: 0 to 90%, non-condensing
Always run the instrument under the ambient conditions and temperatures listed
above. The instrument may be damaged if it is operated in conditions other than
those as described, which will create unpredictable risks.
NOTICE
Only use the instrument appropriately and in a perfectly safe condition.
This is the only way to guarantee the operational safety of the instrument.
10
Safety
WARNING
Risks can occur due to inappropriate use. Inappropriate use includes:
Operation in explosive atmospheres
Exceeding the technical values defined for normal operation
Any working method that jeopardizes operational safety
Commissioning by personnel not authorized by the owner
Modifications or changes
Modifications or additions to the instrument are strictly prohibited and shall render
any liability and warranty of the manufacturer null and void.
NOTICE
Only operate the instrument with all safety devices fitted and in full working order.
Do not remove or bypass any safety devices.
Defective safety devices must only be replaced by identical components.
Check that the safety devices are in full working order at least once a week.
Symbol Description
CE marking
Identifies conformity with applicable EU directives that concern
the product and that require a CE marking.
Dispose of the instrument in an environmentally friendly
manner. The instrument must not be disposed of with
household waste. Dispose of the instrument at a public
collection point.
11
Safety
12
Safety
NOTICE
You are personally responsible for:
Wearing the proper personal protective equipment
Regular care and cleaning of the equipment
Timely replacement of damaged and unusable parts of the protective equipment
2.6.1 Setting up
Place the instrument on a stable, flat and even surface. If necessary, place an
underlay under the instrument in order to protect the surface beneath.
Ensure that the instrument is stable: it must be placed so that it does not
wobble or tilt.
Install the instrument where it has proper ventilation.
Do not operate this instrument in wet or damp conditions.
To avoid injury or fire hazard, do not operate this instrument in an explosive
atmosphere.
Do not operate the instrument near gas appliances.
Keep the instrument and the power cord away from heat sources.
Do not sit or stand on the instrument.
13
Safety
2.6.3 Operation
The instrument must only be operated by persons who are familiar with the
risks and features of the instrument.
Be careful with flammable liquids! Only qualified employees are allowed to
handle flammable liquids.
In the event of a fault, disconnect the instrument from the power supply
immediately.
Never leave the instrument to operate unsupervised.
Never immerse the cord and instrument in water.
Never touch an electrical instrument if it has fallen into water. Disconnect
the instrument from the power supply immediately.
Only operate the instrument using the accessories supplied. The use of other
parts may cause accidents.
In order to prevent the risk of electric shock, do not remove the covers and
panels under any circumstances. There are no parts inside that can be
repaired by the user. Repairs must only be carried out by an authorized
specialist.
Do not ignore warnings on the display or warning signals. Please read the
manual or call our representative if you are not sure what to do.
Do not cover the ventilation slots. Place the instrument so that proper
ventilation is guaranteed.
14
Transportation
3 Transportation
The instrument is originally shipped with a special designed shock proof box.
Please keep this box. You might need it some-day to ship the instrument back to us
for service.
15
Unpacking
4 Unpacking
1 MINISCAN IR VISION
16
Unpacking
17
Unpacking
OPTIONS
Vacuum Pump Assembly IRV-SUB017
18
Description
5 Description
5.1 Front
1
1 Handle
2 10" Color Touch Screen
3 Inlet with flexible filling tube or glass syringe
4 Sample container
5 On switch
19
Description
20
Description
10
15
11
12
13
14
16
9 DVI interface
10 2x LAN interface
11 2x USB interface (printer, keyboard, etc.)
12 Serial RS 232 port (inactive)
13 Main-switch
14 Power supply
15 Fan
16 Port for spectrometer evacuation pump
CAUTION
The maximum permitted length of the LAN cable is 30 m.
21
Installation
6 Installation
Connect to mains
1 Place the instrument on a vibration free solid surface.
2 Connect the power supply cable to the power inlet (14) on the rear panel
of the instrument and the mains voltage.
Install waste container (8)
22
Installation
NOTICE
If on start-up an error message “Touchscreen not found. Check wire connection.”
attach a mouse to operate the instrument.
6.3 Printer
Connect a postscript printer to an USB-Interface (11) on the rear panel of the
instrument.
CAUTION
The maximum permitted length of the LAN cable is 30 m.
23
Concept of the MINISCAN IR VISION
Corresponding examples for the above types are the following methods:
1. RON, MON, T50 ...
2. Ethanol, Toluene, ...
3. FAME, EHN ...
4. AKI = (RON+MON)/2 and more complex calculations
5. All
A good fuel group for Gasoline should contain 100+ samples of normal, super and
special finished in one group to get better statistics for the PLS.
NOTICE
For information on the calibration of specific parameters please review the Section
9.4.6.
24
Operation
8 Operation
CAUTION!
Be careful with flammable liquids!
Flammable liquids should only be used by qualified employees. These substances
should be used with constant regard to the danger they pose to life and property.
Under normal circumstances they should always be used in a fume-hood away from
possible sources of ignition. For further information read the safety data sheet of the
substance.
8.1.2 Login
To operate the analyzer, please login at the first menu page. There is no factory set
password, please read the password section to setup a user-specific password.
25
Operation
1 2
4 6
5
1 Menu-Heading
2 Return to Main Menu ("Home" button)
3 Actual GUI (Graphical User Interface)
4 Return to the previous menu page ("Back" button).
5 Status light:
Green Instrument is ready for measurement.
Yellow Instrument is not ready for measurement, but warming up /
stabilizing temperature / stabilizing IR-source
Red Instrument is not ready for measurement.
6 Time and Date
26
Operation
8.1.4 Status
27
Operation
Details
Save
Save data from this menu on the connected USB-Stick or the internal analyzer disk
(Path: /home/irv/ftp/results/). Please wait, depending on the amount of data
transferred, this may take a few minutes!
Print data from this menu. The HTML printer also prints to the internal instrument
folder /home/irv/ftp/reports/ so that users can access printed reports from the
analyzer.
Delete
OK
Cancel
Abort
28
Operation
1 2 4 5
1 Go to first page
2 Go to previous page
3 Actual page / Last page
4 Go to next page (press button longer for fast scrolling through the
pages)
5 Go to last page (press button longer for fast scrolling through the
pages)
29
Operation
Latin Characters
30
Operation
NOTICE
If the instruments hard drive is nearly full and only 50 additional measurements can
be stored, the IR VISION instrument will issue a warning. Please delete some
measurement results to free space for new measurements. If the instrument hard
drive is full, no measurements can be performed.
2
1
31
Operation
4a
3 Click on "Edit Sample Type" and choose the fuel type of the sample.
5 Select a profile.
32
Operation
4c
4d
4e
4f
4b
4g
4h
4f Delete a profile.
33
Operation
7b
7a 9
10
34
Operation
11
12
NOTICE
If a new / valid reference measurement needs to be started before the test, the
"Start" letters will turn italic. The reference measurement will be automatically
started before the next measurement is done. In case of a cell position issue, an error
message will be displayed and the measurement cannot be started.
35
Operation
36
Operation
[Link] Options
[Link].1 Dilution
A warning message appears if saturation is detected and the sample needs to be
diluted.
Press:
Click "OK" to stop the measurement and dilute the sample. No result is
displayed.
Click "IGNORE" to ignore the warning. A result of the measurement of the
undiluted sample is displayed. All results that are affected by saturation are
marked in red.
NOTICE
If dilution is required but ignored, measurement results affected by saturation may be
wrong and are marked in red.
NOTICE
For samples with a concentration exceeding 25% FAME content, a dilution warning is
very likely, because the FAME absorption band of the 0-7% FAME method is in
saturation. Values determined by the 7-40% FAME method are not affected, the
warning can be ignored.
37
Operation
Before starting a diluted measurement, turn “Diluted on”. The dilution button is only
visible, if a n-Dodecane reference spectrum is stored in the analyzer.
If the absorption spectrum is still in saturation after dilution, no further dilution can
be performed.
IMPORTANT NOTICE
If a sample is diluted, properties predicted from undiluted samples like RON/MON,
Cetane Number, Vapor Pressure or Distillation no longer yield a correct value. If you
plan to measure diluted samples more frequently, create a sample group with diluted
samples and choose this group for property prediction.
38
Operation
To start a measurement with the second cell, toggle the "IR"-button until it says
"Cell 100 µm" in the Measurement Menu and press "Start".
2
1
39
Operation
NOTICE
Please activate the 100 µm Cell in the Factory Calibration Menu before use.
NOTICE
If a two cell measurement for Aniline measurement has been selected, only the
results from the 100 µm cell are shown, whereas results from the IR cell are
[Link] perform an Aniline measurement through the regular cell, please
reactivate the Aniline methods via the Parameter Menu.
1 2
2 Results for GB/T from the 100 µm cell are displayed with the suffix (GB)
40
Operation
To provide a VOC calculation, some sample characteristics must be set prior to the
sample measurement.
1
2
3 If all parameters were filled in, the VOC value is available after the
measurement if not the value is zero.
41
Operation
42
Operation
NOTICE
For information on the calibration of specific parameters please review the
Section 9.4.6.
[Link].1 Procedure
To ensure that a measurement meets the requirement for the PLS calibration we
recommend going through the following steps prior to your database build-up.
1 Let the Instrument warm up for at least 10 minutes and max. 20 min.
2 Go to Main Menu → Instrument Setup → Maintenance → Auto Adjust Mirror.
3 Press the "Start Adjust" button.
43
Operation
2 Use this fuel to mix your calibration matrix. A calibration matrix must
have at least five different concentrations and shall cover the expected
concentration range. Concentrations should be measured in w%.
44
Operation
Make sure that in this method the values are: Quad = 0, Gain = 1 and Offset = 0.
Note down the current Min and Max values.
Then set Min = -10 and Max = 100.
Press "Save" to confirm the new method settings.
Return to the Measurement Menu and record the calibration samples like
regular samples. If needed set your pyrosensor stabilization cycles and sample
scans to 4 and the rinsing cycles to 3.
45
Operation
1 2 5
3 4
Delete the current calibration. Gasoline and Diesel calibrations can be deleted
separately.
Record all calibration samples at this point only then continue with step (3)
3a
46
Operation
4 Press the "Calibrate" button and enter the required data as described below.
The shown table must be filled with the needed data. Depending on the color of the
field the following data is required:
Grey text: Measurement or value must be entered. A click on the pre-
defined Reagecon sample name opens a dialog to select the
corresponding measurement. A click on the value opens a
numpad to enter the reference value.
Black text: The value is pre-defined and should not (but can be) changed.
Blue text: Caption of the Reagecon sample. In the right cell, the
corresponding measurement must be selected.
Green BG: Validation measurements.
IMPORTANT NOTICE
Concentration values are always displayed in w% and must be entered in w%.
Check the Log and Plot tab for the result of a calibration.
Press "Create Certificate" (5) to save the calibration certificate as HTML file in the
FTP download folder.
To download the certificate, an FTP connection must be established. The file name of
the certificate is "Certificate_serialnumber.html".
If plugged, the certificate is also saved to the USB-flash drive.
47
Operation
[Link] Calibration
48
Operation
4a
4
3
7a
49
Operation
7a If a property has no specific unit like RON, the blank unit ("<none>" ) must be
selected.
You also can skip step 6 and 7(a) and enter the known properties' values more
comfortable afterwards with the desktop software IRDB Tool.
8
8 Enter the Measurement Menu.
9 Click on "Edit Sample Name" and define a sample name so you can easily
find this measurement in the database. Type a name with the virtual
keyboard and confirm with the button "ENTER".
10
10
11
50
Operation
After all samples are added, re-training the affected methods in the Training Menu,
to improve the prediction.
2 Choose the fuel type of the sample and the cell type.
3 Click on "Change Group Assignment and Train".
51
Operation
4 5
NOTICE
Repeat this procedure for every method, which you want to train with new groups or
groups to which calibration samples have been added.
52
Operation
2 Choose the fuel type of the sample and the cell type.
3 Select "Train Selected Methods".
53
Operation
5
6
4 Select methods for training.
5 Confirm the selected methods with "OK".
6 Start the training
NOTICE
Note that the training can take a few minutes. After starting, do not press any other
buttons or disconnect the instrument from power until the training is completed. An
incomplete training may corrupt the database.
54
Operation
NOTICE
Note that the analysis can take some time. After starting, do not press any other
buttons or disconnect the instrument from power until the training is completed. An
incomplete training may corrupt the database.
55
Operation
Group China
Group China
56
Operation
Please make sure that your sample set uses roughly the same range you want to
test in e.g., if you plan to test RON 95 samples, it will not work with RON 92-93
calibration samples. Extrapolation is not allowed for PLS methods. If the calculated
result value is outside of the training set´s range, then the result is undefined.
Create a new group for your calibration samples (refer to the section [Link].1 for
creation of a new calibration group). A new group is per default unlocked and
ensures the user can always add calibration samples to this group.
Sample #1 #2 #3 #4
57
Operation
NOTICE
Some groups have been optimized by the factory.
These groups are locked to protect the quality of the group.
If the prediction cannot be improved by adding new samples (Step 2), you may
consider creating a new method. Use the IRDB tool to create a new method.
NOTICE
New methods should only be created by experts.
58
Operation
1 Manual Filling, use this button to open or close the filling system.
2 Check the current temperature. If it is lower or higher than 20°C, set it
to 20°C.
3 Press the "Cyclohexane Correction" button
59
Dreckstext damit ers druckt der
word-scheissdreck
Dreckstext damit ers druckt der
word-scheissdreck
PORTABLE FUEL ANALYZER
MINISCAN IR VISION
OPERATION MANUAL
M-V1.09 SW-V2.4
Part 2 of 2
IMPRINT
Issued by
Updating This instruction manual can be changed without prior notice and without
liability on the part of AMETEK, Inc. or its subsidiaries.
Table of content
Part 1 of 2
1 General 5
1.1 Identification of the instrument ............................................................................................. 5
1.2 Purpose of the documentation ............................................................................................... 5
1.3 Care of the documentation ..................................................................................................... 5
1.4 Liability ................................................................................................................................... 5
1.5 Warranty ................................................................................................................................. 6
2 Safety 7
2.1 Explanation of the symbols and types of note ....................................................................... 8
2.1.1 Safety instructions ....................................................................................................................... 8
2.1.2 Important mandatory instructions .............................................................................................. 9
2.1.3 Information and notes on use ...................................................................................................... 9
2.2 Appropriate use .................................................................................................................... 10
2.3 Reasonably predictable incorrect use ................................................................................. 10
2.4 Safety and protective devices ............................................................................................... 11
2.4.1 Symbols on the instrument........................................................................................................ 11
2.5 Areas of responsibility .......................................................................................................... 12
2.5.1 Areas of responsibility on the part of the operator ................................................................... 12
2.5.2 Areas of responsibility on the part of personnel....................................................................... 12
2.5.3 Personal protective equipment ................................................................................................. 12
2.6 Safety instructions ................................................................................................................ 13
2.6.1 Setting up.................................................................................................................................... 13
2.6.2 Mains connection ....................................................................................................................... 13
2.6.3 Operation .................................................................................................................................... 14
3 Transportation 15
4 Unpacking 16
5 Description 19
5.1 Front ..................................................................................................................................... 19
5.2 Outlet side............................................................................................................................. 20
5.3 Rear side ............................................................................................................................... 21
6 Installation 22
6.1 DC/AC Power converter ....................................................................................................... 23
6.2 External keyboard ................................................................................................................ 23
6.3 Printer................................................................................................................................... 23
6.4 LAN interface........................................................................................................................ 23
MINISCAN IR VISION
Revision / Issue: M-V1.09 SW-V2.4
© 2020 AMETEK, Inc. All rights reserved.
General
8 Operation 25
8.1 Common Description............................................................................................................ 25
8.1.1 User Interface............................................................................................................................. 25
8.1.2 Login ........................................................................................................................................... 25
8.1.3 Structure of the User Interface ................................................................................................. 26
8.1.4 Status .......................................................................................................................................... 27
8.1.5 Important Buttons ...................................................................................................................... 28
8.1.6 Navigation Bar ............................................................................................................................ 29
8.1.7 Keyboard and Numpad............................................................................................................... 29
[Link] Chinese Keyboard .................................................................................................................. 30
8.2 Work with the IR VISION ....................................................................................................... 31
8.2.1 Measurement ............................................................................................................................. 31
[Link] Switch on the instrument ....................................................................................................... 31
[Link] Define measurement settings ............................................................................................... 31
[Link] Options ................................................................................................................................... 37
[Link] Measurement Warnings......................................................................................................... 42
8.2.2 Calibration: Analyzer Preparation for Measurements ............................................................. 43
[Link] Setting up the IR VISION for Calibration Measurements ....................................................... 43
8.2.3 Calibration of methods (Concentration) .................................................................................... 44
[Link] Switch on the instrument ....................................................................................................... 44
[Link] Preparation of calibration samples ....................................................................................... 44
[Link] Measure Calibration Samples................................................................................................ 44
[Link] Gain and Offset ....................................................................................................................... 45
[Link] Substance Calibration ............................................................................................................ 46
8.2.4 Calibration of new samples (Prediction) ................................................................................... 48
[Link] Switch on the instrument ....................................................................................................... 48
[Link] Calibration.............................................................................................................................. 48
8.2.5 Group Assignment for Methods ................................................................................................. 51
8.2.6 Methods Training........................................................................................................................ 53
[Link] Train Selected Methods ......................................................................................................... 53
[Link] Train All Methods ................................................................................................................... 54
8.2.7 How to improve results for predicted properties ..................................................................... 55
[Link] Step 1: Change a method's group assignment ...................................................................... 55
[Link] Step 2: Add new calibration data............................................................................................ 57
[Link] Step 3: Create a new method ................................................................................................. 58
8.2.8 Calibration of Density Meter ...................................................................................................... 59
9 Instrument Screens 63
9.1 Main Menu ............................................................................................................................ 63
9.2 Measurement Menu ............................................................................................................. 64
9.2.1 Parameter Menu ........................................................................................................................ 65
9.2.2 Measurement Settings ............................................................................................................... 66
9.3 Rinsing .................................................................................................................................. 68
9.4 Overview ............................................................................................................................... 69
9.4.1 Columns...................................................................................................................................... 70
9.4.2 Results Details ........................................................................................................................... 71
9.4.3 Density Units............................................................................................................................... 72
9.4.4 Diagrams .................................................................................................................................... 72
9.4.5 Goodness Parameter and Quality Parameter (for PLS methods and cluster analysis) .......... 73
9.4.6 List of Parameters ..................................................................................................................... 75
9.5 Graphics ................................................................................................................................ 79
MINISCAN IR VISION
Revision / Issue: M-V1.09 SW-V2.4
© 2020 AMETEK, Inc. All rights reserved.
General
10 Cleaning 107
10.1 Touch Screen ...................................................................................................................... 107
10.2 Housing ............................................................................................................................... 107
10.3 Rinsing the instrument....................................................................................................... 107
10.3.1 Rinsing with cleaning mixture ................................................................................................. 107
11 Maintenance 108
11.1 Oil filling.............................................................................................................................. 108
11.2 CO2 peak too high – spectrometer evacuation with pump................................................. 109
MINISCAN IR VISION
Revision / Issue: M-V1.09 SW-V2.4
© 2020 AMETEK, Inc. All rights reserved.
Dreckstext damit ers druckt der
word-scheissdreck
Instrument Screens
9 Instrument Screens
Measurement
Rinsing
Correlation Calibration
Results
Instrument Setup
Shutdown
63
Instrument Screens
Diluted On / Off
Select, if you want to perform a standard measurement (1) or use the 100 µm (2) or
run a combination measurement of IR and 100 µm cell (3).
Profiles/Parameter
Select a profile with a click or go to the Parameter Menu with a long click.
Measurement Settings
Repeats
64
Instrument Screens
Start
Start the measurement. If the Start button is displayed italic a new reference is
recorded prior to the sample measurement.
1 2
1 Select the parameters to be shown in the result details menu. Use the
buttons below to navigate between the pages.
2 Sort parameters by fuel type, selection, name, unit and sort key.
3 Select the displayed result unit.
4 Edit the sort key. Parameters in the Result Details Menu are sorted acc.
to the sort key
5 Name of the current profile
Load
65
Instrument Screens
Save
Delete
Delete a profile.
Apply
NOTICE
Density units can only be selected in the Results Details Menu.
Temperature Control
Activate or deactivate the temperature control during the measurement. Note that in
the Measuring Menu the nominal temperature is displayed. The real temperature is
displayed in the Results Details Menu.
66
Instrument Screens
Fill method
Choose the fill method: Manual with glass syringe or automatic with filling tube and
Luer connector.
A new reference measurement will be automatically started before the next run.
Cluster Analysis
Activate or deactivate the cluster analysis feature. With the button you can
change the group assignment.
Density Temperature
Select, which density temperature shall be shown in the result. The density will be
recalculated using the given fuel specific expansion coefficient.
Expansion Coefficient
67
Instrument Screens
9.3 Rinsing
Rinse the instrument if a switch from Diesel /Jet to Gasoline samples or vice versa is
performed. Or in case samples containing a high amount of FAME have been
measured.
Use a new sample or solvents to rinse the instrument.
Fill Method
Change the rinsing and fill method from Automatic to Manual (Syringe) or from
Manual to Automatic.
Empty
Start emptying the cells in the instrument. Ensure that a waste container is
installed. If the process is finished, "Finished emptying cells" is displayed.
Rinse
Start rinsing the cells in the instrument. If the process is finished, "Finished rinsing
cells" is displayed. In manual filling mode, the instrument valves are opened to allow
users to perform a manual rinsing using a syringe.
68
Instrument Screens
9.4 Overview
The results menu lists the results, details and spectra from fuel measurements. If a
GB/T two cell measurement has been performed, a (GB) suffix will appear after the
sample name (1).
Details
Select, which two properties should be shown in the measurement overview list.
Customer / Calibration-Measurements
Save
69
Instrument Screens
Columns
Select, which two properties should be shown in the measurement overview list.
Delete
NOTICE
Do not delete calibration data in this menu, unless it is required. Deleting calibration
data will severely affect the precision of property prediction.
9.4.1 Columns
Main Menu → Results → Columns
70
Instrument Screens
On the top you see the measurement's metadata that is date, sample name, fuel
type, temperature, density and user that was used for testing.
The individual properties are listed in the table below. Use the navigation buttons to
see all properties.
Graphics
Save
Save data from this menu on the connected USB-drive and the internal hard disk
(Path: /home/irv/ftp/export/ ). Please wait, depending on the amount of data
transferred, this may take a few minutes.
Print data from this menu. The HTML printer also prints to the internal instrument
folder /home/irv/ftp/reports/ so that users can access printed reports from the
analyzer.
Next Measurement
Recalculate Properties
71
Instrument Screens
1
2
3
To change the displayed density unit
1 Click on the small button near the density result
2 Select a unit.
3 Press "OK"
9.4.4 Diagrams
Bar diagram
The bar diagram shows the value in the defined range (e.g. RON: 70 - 105) at a
glance.
If values are out of the measuring range, the following symbols appear: <, >, n.d.
n.d. not detected
< below lower limit
> above upper limit
72
Instrument Screens
NOTICE
You can change the upper and the lower limit of a method in the Method Edit Menu.
9.4.5 Goodness Parameter and Quality Parameter (for PLS methods and
cluster analysis)
The Goodness Parameter (%) is a measure for the similarity of the measured
sample, when compared to the samples that are used for the prediction. It
determines how good a new sample fits to the samples in the groups assigned to a
method (e.g. RON).
Thus, the Goodness Parameter indicates, how good a result fits to the underlying
prediction model. 100% indicates the maximum distance of a spectrum, when
compared to all samples of a group assignment. A result that shows a Goodness
Parameter higher than 100% is outside of the current group assignment, and the
result must be treated with care. A result with a Goodness Parameter lower than
100% indicates that the spectrum is inside the current group assignment.
Let us say there are 3 groups (United States, China, Russia) assigned to the method
RON and you have recorded two unknown gasoline samples. Sample A has a
Goodness values of 36% that means it fits to the three assigned groups and you can
assume that the prediction will work. Sample B has a Goodness value of 127% which
means that it is further away from the center than the furthest sample of the
73
Instrument Screens
assigned groups thus outside of the prediction model. The predicted RON value of
this sample must be treated with care.
For cluster analysis results the Quality Parameter (CA) is used. The CAx value is
scaled from the best (CA1) to the worst (CA5) match of the database. Depending on
the fuel type, the Quality Parameter can be:
CA-Distance
CA5
CA4
CA3
CA2
CA1
Perfect Match
Let us say there are 3 groups (United States, China, Russia) assigned to the cluster
analysis and you have recorded two unknown gasoline samples. Sample P has the
Quality Parameter CA1 that means that a very close spectrum was found in the
database thus the fuel properties must also be very similar and are assigned to
sample P. Sample Q has no matching spectrum within the maximum allowed
gasoline distance of CA3 thus no fuel properties can be assigned and the result
values are calculated with the measuring methods of type 1-4 (see 7.1).
NOTICE
The Goodness Parameter is only available for parameters calculated with PLS. Other
parameters and parameters that yield physical impossible figures will show as "-".
NOTICE
The Quality Parameter is only available for parameters found by the cluster analysis.
NOTICE
Recalculation of results will take longer for IR VISION versions 2.2.0 and above
because the Goodness Parameter needs serious calculation power.
74
Instrument Screens
75
Instrument Screens
76
Instrument Screens
77
Instrument Screens
78
Instrument Screens
9.5 Graphics
9.5.1 Absorption Spectrum
Main Menu → Results → Details → Graphics
Press the buttons to move left and right in the spectrum screen.
1:1
Press the button 1:1 to fit the whole spectrum in the screen.
Add spectra
Press the button to add additional spectra to the diagram for comparison.
79
Instrument Screens
An unlimited number of spectra can be chosen for direct comparison in the diagram.
2 3
80
Instrument Screens
The distance between two spectra is calculated over the complete spectral range.
1 You may drag and drop the distance information field.
2 You may activate or deactivate the distance field.
3 The button opens a measurement selection dialog.
Export to USB:
1 Attach a USB device
4 If more than one spectrum is shown on the graphics menu, all spectra are
exported to /serialnumber/[Link]. In this file, data of different spectra
are listed in separate columns.
2 If more than one spectrum is shown on the graphics menu, all spectra are
exported to the directory /home/irv/ftp/spectrum/[Link].
81
Instrument Screens
NOTICE:
Sample names may not contain the following characters: \ / : * ? “ < > |
These characters are automatically replaced during the export.
Maintenance
Calibration
Settings
82
Instrument Screens
9.6.1 Maintenance
Main Menu → Instrument Setup → Maintenance
[Link] Self-test
Main Menu → Instrument Setup → Maintenance → Selftest
Before the self-test can be started, a flask containing 50ml cyclohexane must be
connected to the inlet.
A self-test can be run as a full test or a spectral test.
Actuators displayed italic do not have any controlling mechanism like a light barrier
or a Hall sensor. In this case only the software and the electrical part can be tested.
83
Instrument Screens
To be sure that their work properly one must watch them moving / hear them
clicking.
Some test must complete successfully before the next test can start. If this is not the
case, the self-test stops at the failed step.
Follow the quick guide to calibrate the instrument. See chapter [Link] for details.
84
Instrument Screens
Check
Start Adjust
To adjust the intensity, select "Start Adjust". Intensity values will change their
color to help to judge, if the intensity is good enough for measurements.
85
Instrument Screens
Start / Abort
Start / Abort the cell check. Follow the instructions shown on the screen. For cell
flushing Grabner Instruments recommends using a [Link] mixture of Acetone,
Toluene and Ethanol. Repeat flushing until the required cleanliness is achieved.
NOTICE:
This procedure requires manual filling from syringe with a cleaning solution and dry
air. "OK" indicates that the reference test passed and the cells are clean.
86
Instrument Screens
1 2
Full Backup
A full backup will also backup spectral data, which will take a while.
A full backup from the instrument with the serial number 23-530-0008 will create a
folder with a time stamp: IR3_EXPORT_23-530-0008_2013_6_27_15_11_52.
It is recommended to use a USB 3.0 flash drive.
Core Backup - Service work only
To restore a full backup, rename the folder from "EXPORT" to "IMPORT" and remove
the time stamp from the folder´s name. When working with the IRDB Tool the folder
is automatically renamed.
Example: "IR3_EXPORT_23-530-0008_2019_xxxx" to "IR3_IMPORT_23-530-0008".
Load the import file on a USB-device, attach it, select "Restore" and follow the
instructions.
NOTICE
A full backup is required for service and the IRDB Tool.
87
Instrument Screens
NOTICE
If the full backup restore procedure crashes do not try to repeat it. The previous state
is stored internally and can be restored by the Grabner service team.
[Link] Update
Main Menu → Instrument Setup → Maintenance → More → Software Update
Copy the update file into the root directory of your USB device
Attach the USB device to the IR VISION
Update
NOTICE
A number of methods is affected from the version 2.1 method update and need to be
recalibrated after the update. If you do not plan to recalibrate your analyzer, please
do not update to version 2.1!
After update to 2.1, please follow this procedure before you use the analyzer
1 It is required to retrain everything after update (all methods, all fuel types)
88
Instrument Screens
2 In contrast to previous updates, after the update from version 2.0 to 2.1 some
methods must be recalibrated (see table below in red)
3 New methods are installed, these need to be calibrated as well (see table
below in red)
4 The changed methods also affect several methods that are depending on
these changed methods (see table below in black). Results of these
"dependent" methods will change after the update.
5 The training of all methods will also train PLS methods that are not trained
by default and results of these methods will change (see table below in black).
Table: Results that are affected by updating to software version 2.1 and the
following "train-all-methods" step that is required.
NOTICE
Methods that have to be recalibrated are listed red in the table above. Please
recalibrated before performing the all methods training.
89
Instrument Screens
This menu lists various export and import functions (Service and Support only).
USB-Export
Export
Import
File System
Meas.
Methods
Full
90
Instrument Screens
Log
This function allows to connect the instrument with Grabner Instruments Service
and Support Center via Internet. The instruments must be connected to the internet
(see chapter Network Settings).
91
Instrument Screens
9.6.2 Calibration
Main Menu → Instrument Setup → Calibration
NOTICE
Analyzers that are updated from SW 2.0 to 2.1 must be recalibrated. Please refer to
the “Update” section for further information.
92
Instrument Screens
Cyclohexane Correction
Fill
Automatic filling.
Empty
Automatic emptying.
Open / Close
93
Instrument Screens
Add a sample from the region with known properties to your custom group or to a
preinstalled group
Train All
Start
94
Instrument Screens
Name
Select the group you want to add the new calibration sample to.
Count
Add
Create a new group, e.g. for your region. Type a name with the virtual keyboard and
confirm with the button "Enter".
NOTICE
You cannot create two or more groups with the same name.
95
Instrument Screens
Type known values for the properties in the respective lines. Always add a unit to any
property.
You can also use the IRDB Tool desktop software to enter the values later.
Name
List of all properties. Use the buttons below to navigate between the pages.
Value
Type a known value for the property with the virtual numpad.
Unit
Always choose a unit for any property. For unitless values like RON the "empty" unit
must be selected.
Start
IMPORTANT NOTICE
If wrong reference values are entered, the MINISCAN IR VISION uses these values for
calibration. A calibration with wrong reference values inevitably leads to wrong result
values.
96
Instrument Screens
Name
Select one or more groups (e.g. regions) to train the selected method.
Lock
Count
NOTICE
Note that the analysis can take a few minutes!
After starting, do not press any other buttons or disconnect the instrument from
power until the analysis is completed.
97
Instrument Screens
NOTICE
Training may take some time, depending on the number of methods selected.
98
Instrument Screens
[Link] Methods
Main Menu → Instrument Setup → Calibration → Methods
In this menu, you may review methods and change some method settings.
Min / Max
Define the lower (Min) and the upper (Max) display limit of the method.
# Dec
Groups
Save
Select method
Update the calibration if needed. These buttons are inactive (greyed) for arithmetic
methods.
99
Instrument Screens
9.6.3 Settings
Main Menu → Instrument Setup → Settings
Define common settings like network, language, date / time, data export and
settings for the measurement. Activate instrument upgrades. Protect the instrument
with a password.
Plug the network cable on the rear side of the instrument and connect to the
network.
1 Set the DHCP-Server to <Yes>
100
Instrument Screens
2 Click on the "Connect" icon. Once DHCP is activated the instrument will try
to connect to the network until DHCP is deactivated again.
The analyzer IP address is shown after successful connection (e.g. [Link])
Open your browser and establish an FTP-Connection (e.g. [Link]
With Windows 7 and higher you can also use the Windows® Explorer.
If the IR VISION is unable to connect to your network, please consult your IT.
If the IR VISION is unable to connect to your computer, please consult your IT.
101
Instrument Screens
In this menu you select, if all measurements or only the last measurement shall be
stored on the FTP export location of the instrument.
Historical Data
102
Instrument Screens
Choose your language. Click on "Layout" to choose the layout of the virtual keyboard.
103
Instrument Screens
[Link] Printer
Main Menu → Instrument Setup → Settings → Printer
Click on a number and change it with the arrow buttons. Confirm with the "OK"
button and the set time and date are displayed at the bottom right.
104
Instrument Screens
5
1
To change a password, enter the current password (1) and the new password twice
(2)(3). If both new passwords match each other press the "Change Password" button
(4). To disables the password protection set an empty password.
On the top right corner of the keyboard buttons the current layout is displayed (5).
The layout of the virtual keyboard can be changed by a long click on the keyboard
button.
9.7 Shutdown
Confirm the shutdown with the button "Shutdown".
105
Instrument Screens
After a license key for the IRDB Tool has been purchased, the software is available
for download at [Link]
NOTICE
Please register your instrument via our website [Link].
Databases collected with the IRXpert can be transferred to the IR Vision analyzer.
Download a full backup of your IRXpert and your IR VISION analyzer. Use the IRDB
Tool to merge the IRXpert full backup into the IR VISION full backup.
NOTICE
For more information and about do's and don'ts of moving databases between
analyzers please contact support at [Link]@[Link].
106
Cleaning
10 Cleaning
CAUTION!
Do not use cleaning agents that could cause a hazard as a result of reaction with the
used probe material; the manufacturer or his agent is to be consulted if there is any
doubt about the compatibility of cleaning agents!
10.2 Housing
Clean the housing with a damp cloth.
CAUTION!
Do not use cleaning agents that could cause a hazard as a result of reaction with the
used probe material; the manufacturer or his agent is to be consulted if there is any
doubt about the compatibility of cleaning agents!
107
Maintenance
11 Maintenance
Set the piston position to "0" (you can either use the spin box buttons or a USB
keyboard) and press the "Goto Pos" button. If the analyzer's filling method is set to
Automatic you can also use the "Rezero" button.
Use the 2 mL syringe to fill oil up until the bottom reservoir of the oiler is filled (1).
108
Maintenance
109
Customer Support and Information
110
Technical Data
13 Technical Data
111