The document provides a detailed description of the iSYS-4001 GUI interface for configuring and displaying measurements from the iSYS-4001 radar sensor. It outlines the start-up sequence, connection process, and various configuration options available within the GUI, including live data streaming, output signal settings, and target filtering. Additionally, it includes contact information for support and technical assistance.

Uploaded by

wyortiz75
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
33 views12 pages

ISYS-4001 GUI Interface V1.1

The document provides a detailed description of the iSYS-4001 GUI interface for configuring and displaying measurements from the iSYS-4001 radar sensor. It outlines the start-up sequence, connection process, and various configuration options available within the GUI, including live data streaming, output signal settings, and target filtering. Additionally, it includes contact information for support and technical assistance.

Uploaded by

wyortiz75
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd

www.InnoSenT.

de

iSYS-4001
GUI interface - V1.1

Power up

LED blinking

Sensor selftest

Selftest YES LED flashes Measurement


passed yellow mode starts

NO

LED blinking red

object YES
LED flashes
approaching red

NO

YES
object LED flashes
receding green

NO

LED flashes yellow


iSYS-4001 speed and distance sensor | GUI description
www.InnoSenT.de

content
start-up sequence 4

1. connecting the iSYS-4001 to your PC 5


2. installing iSYS-4001 GUI 5
3. start screen 6
4. „Live Stream“ 6
5. “Save” - “Load” configuration 6
6. frequency channels 7
7. option „Raw Signal 7
8. tab „Detection“ 7
9. automatic amplification 8
10. output signals 1-3 8
11. target filter options 9
12. output configuration 9
13. manual amplification 10
14. angle correction 10

contact information 11

2 2 www.InnoSenT.de - Experience and Reliability in Radar-Technology


GUI description | iSYS-4001 speed and distance sensor
www.InnoSenT.de

iSYS-4001
GUI - description
The iSYS-4001 GUI can be used to configure and display the measured signals of an iSYS-4001 radar sensor. It
displays the target list, FFT magnitude with threshold and the signal raw data.

Experience and Reliability in Radar-Technology - www.InnoSenT.de 3 3


iSYS-4001 speed and distance sensor | GUI description
www.InnoSenT.de

start-up sequence
The integrated LED indicates the status of the sensor on power up.

Power up

LED blinking

Sensor selftest

YES
Selftest passed LED flashes yellow Measurement mode starts

NO

LED blinking red

object
YES
LED flashes red
approaching

NO

object
YES
LED flashes green
receding

NO

LED flashes yellow

4 4 www.InnoSenT.de - Experience and Reliability in Radar-Technology


GUI description | iSYS-4001 speed and distance sensor
www.InnoSenT.de

1. connecting the iSYS-4001 to your PC


To connect the iSYS-4001 with your PC use the following configuration (complete cables can be purchased from InnoSenT).

9 5
M1 2

M1 2
4
RS232-Connector

Input Pin # Signal Pin # Output


Buchse M12x8 1 OUT1 1 Stecker M12x8 2 1
Buchse M12x8 2 OUT2 2 Stecker M12x8 8
3 7
Buchse M12x8 3 OUT3 3 Stecker M12x8

Buchse M12x8 5 VCC 5 Stecker M12x8 4 6


Buchse M12x8 6 GND 6 Stecker M12x8 5

Buchse M12x8 4 not connected 4 Buchse RS232

Buchse M12x8 6 GND 5 Buchse RS232

Buchse M12x8 7 RS232_Rx 2 Buchse RS232

Buchse M12x8 8 RS232_Tx 3 Buchse RS232

Connector M12x8 A-type compatible to SACC-M12-SCO PLUG (PN: 1551493) from Phoenix Contacts

The sensor has a RS232 or RS485 interface (depending on sensor-type). An iSYS-4001 with RS232 interface can be
connected directly to the COM port of the PC otherwise there can be a RS232 to USB adapter used. The Adapter installs
a virtual COM port which can be selected in the GUI.

2. installing iSYS-4001 GUI


Software is available @ www.innosent.de
(Service -> Downloads)

• Copy installer „iSYS4001_GUI_Setup“ to your harddisk.


• Double-click on „iSYS4001_GUI_Setup“ to start Setup-
Wizard.
• Follow instructions.

• Click „Finish“ to close setup and to launch GUI.

Experience and Reliability in Radar-Technology - www.InnoSenT.de 5 5


iSYS-4001 speed and distance sensor | GUI description
www.InnoSenT.de

3. start screen
After launching the GUI the start screen apears.

To connect the iSYS-4001 sensor to the GUI

• click „COM Port Config“ click to configure COM Port


• select your COM-Port
• set baud rate to 115000
• click „Connect“
• click „Close“

During the connection process the GUI reads some parameters select COM-Port
from the Sensor. The red lamp in the connection window
set baud rate to 115000
switches to green if successfull.

click to connect

4. „Live Stream“
A click on „Start Live Stream“ starts displaying the live data in
the opened tab.

4.1 Freeze/Run
For an improved analysis of the sensor data, the display can be
frozen by clicking on “Freeze…”. To enable the live data again
click on “Run…” .

4.2. Write all Params tp EEProm


This command writes all data into the EEPROM. This command
is mandatory after setting all your parameters, otherwise all
configurations made are lost. click to start „Live View“

5. „Save“ - „Load“ configuration Short Range.ipf

To configure more than one sensor with the same settings all output1 approaching & receding
active low, if object exist
parameters can be saved in an *.ipf file. active high, if NO object exist
output 2 approaching
5.1 Save configuration to file active high, if object exist
By clicking on “File” => “Save configuration to file” the actual active low, if NO object exist
parameters will be saved in an *.ipf file. output 3 approaching
active high, if object exist
5.2. Load configuration from file active low, if NO object exist
By clicking on “File” => “Load configuration from file” parameter rising delay 375ms
sets can be loaded. To write these parameters to the sensor click falling delay 75ms
on “Write all parameters to EEProm” otherwise the settings are filter type highest amplitude
lost after shutdown the sensor. The Standard configuration is
single target filter 20% for all outputs
the “Short Range” parameter set (compare table beside). Other
amplification 20dB
parameters sets are available @ www.innosent.de
distance range 0....50m
detection level 0dB

6 6 www.InnoSenT.de - Experience and Reliability in Radar-Technology


GUI description | iSYS-4001 speed and distance sensor
www.InnoSenT.de

6. frequency channels
To avoid interferences if two sensors are mounted close to each
other 2 different frequency channels can be selected.

• click „Config“
• select „use channel 1“ or „use channel 2“ for using the
sensor in Europe
• select „use channel 3“ or „use channel 4“ for using the
sensor in the US

Channel 1 is corresponding to a transmit frequency of 24.190GHz


Channel 2 is corresponding to a transmit frequency of 24.210GHz
Channel 3 is corresponding to a transmit frequency of 24.155GHz
Channel 4 is corresponding to a transmit frequency of 24.135GHz

7. option „Raw Signal“


The Option „Raw Signals“ can be used to view the analog
received signals.

The raw signals tab shows the signal raw data which sampled by
the analog to digital converter. Alternatively the user can switch
to the FFT signal (Tab “Detection”). The FFT and threshold are
calculated by the sensor.

8. tab „Detection“
This tab shows the FFT and the threshold calculated by the radar
sensor.

The threshold consists of three parts:


• adaptive value, height of the threshold depending on the
noise level
• low-pass filter, for the frequency selection of the different
objects
• user defined threshold value, will be added to the threshold

8.1. User defined threshold


Changing this value, allow the user to set the sensitivity of
the detection. There can be set the height of the threshold by
adjusting the slider or by setting the level in dB (function is only
available if „Live View“ is stoped).

8.2. Linefilter
By activating the Line Filters several frequencies can be
cancelled. This can be used for example to prevent interference
from fluorescent lamps.

Experience and Reliability in Radar-Technology - www.InnoSenT.de 7 7


iSYS-4001 speed and distance sensor | GUI description
www.InnoSenT.de

9. automatic amplification
An automatic amplification can be activated in „Config“ =>
„Amplification“.

If the radar signals does not exceed the inner threshold (orange
area), the signal will be increased.
If the radar signal exceeds the inner threshold, but does not
exceed the outer threshold (green area), nothing happens to the
signal amplification (signal is inside hysteresis).
If the radar signal exceeds the outer thershold, the signal
amplification will be decreased.

10. output signals 1-3


The iSYS-4001 offers 3 output signals that can be activated and
configured indepentendly of each other.

The tabs “Output 1”, “Output 2” and “Output 3” shows the


different target lists. Each target list has separate filter parameters
which can be changed in the specific output tab.

During a live stream a coloured button shows the actual status


of the detected object:

• Green – object is receding


• Red – object is approaching
• Orange – no object detected

The output can configured by the „Filter Type“ Drop-Down menu.

Possible settings are:

• inactive (no targets)


• none (show all targets)
• highest Amplitude
• mean Value (range)
• median Value (range)
• mean Value (velocity)
• median Value (velocity)

The selection of the wanted option has a direct influence to the


signals in the GUI as well as on the Output Configuration.

Show Histogram
By activating the Histogram option you can follow the time
dependent trend of changes in speed and range of the object(s).

8 8 www.InnoSenT.de - Experience and Reliability in Radar-Technology


GUI description | iSYS-4001 speed and distance sensor
www.InnoSenT.de

11. target filter options


The target filter options declare the velocity-, range- and signal-
area within the sensor is able to detect objects. Velocity has an
additional option to set the sign of the velocity to detect only
receeding or approaching objects.

The filters can be set in the following range:

Speed range: 0....250km/h


Distance range 1: 0....50m
Distance range 2: 0....150m
Signal Level: 0....250dB

12. output configuration


To configure the output signals go to menu „iSYS“ => „Output
Configuration“

The output PINs can be activated and configured seperately


from each other.

Possible setting of output type are:

• digital
• pwm (velocity)
• pwm (range)

In case of „pwm“ a pulse-width modulated signal will be send


to the output PINs corresponding to the speed or range of the
objects (depending on which option is selected)

Output PIN configuration:

configuration description target detected output pin state


active low - normally open low side switch no high impedance

active low - normally open low side switch yes connected to GND

active low - normally closed low side switch no connected to GND

active low - normally closed low side switch yes high impedance

active high - normally open high side switch no high impedance

active high - normally open high side switch yes connected to VCC

active high - normally closed high side switch no connected to VCC

active high - normally closed high side switch yes high impednace

totem pole - normally open push-pull no connected to GND

totem pole - normally open push-pull yes connected to VCC

totem pole - normally closed push-pull no connected to VCC

totem pole - normally closed push-pull yes connected to GND

Experience and Reliability in Radar-Technology - www.InnoSenT.de 9 9


iSYS-4001 speed and distance sensor | GUI description
www.InnoSenT.de

13. manual amplification


The iSYS-4001 sensor features a variable IF amplification with an
8 Bit digital potentiometer. For changing the amplification, open
the slider window in the “Config” menu under “Amplification”.

The button “Store” stores all actual set parameters including the
amplification into the devices EEPROM.

The figures beside are showing the amplification for the different
potentiometer positions.

14. angle correction


The iSYS-4001 radar sensor features an angle correction
depending on the mounting of the sensor. The correction for the
elevation position can be set by mounting angle or by mounting
height.

• Correction with angle: In this case the system assumes iSYS-4001


that the detected object is measured in the main beam
direction. The correction is calculated with the cosine of the measured range to object
angle α. This angle is also called the mounting angle of the
mounting
sensor. height
α
• Correction with height: In this case the system calculates direction of moving object
the angle α for the correction with the measured range of
the detected object and the given mounting height. The
correction is calculated with the cosine of the angle α. This
angle is also called the mounting angle of the sensor.

10 10 www.InnoSenT.de - Experience and Reliability in Radar-Technology


GUI description | iSYS-4001 speed and distance sensor
www.InnoSenT.de

contact information
If you have any questions please contact us!

Corinna Mergenthal
Tel.: +49-9528-9518-84
E-Mail: corinna.mergenthal@innosent.de

Martin Maidhof
Tel.: +49-9528-9518-18
E-Mail: martin.maidhofl@innosent.de

part numbers

P/N: iSYS-4001 P/N: iSYS-prog_adap P/N: iSYS-pow_adap P/N: iSYS-pow_sup

Experience and Reliability in Radar-Technology - www.InnoSenT.de 11 11


Contact Details: