What Is HPLC?

Basic Principles

LAAQ-B-LC001B 1
Invention of Chromatography by
M. Tswett

Ether Chromatography

Colors
Chlorophyll

CaCO3

LAAQ-B-LC001B 2
Comparing Chromatography to the
Flow of a River...
Light leaf
Heavy stone Water flow

Base

LAAQ-B-LC001B 3
Mobile Phase / Stationary Phase

 A site in which a moving


phase (mobile phase) and
a non-moving phase
Mobile (stationary phase) make
phase contact via an interface
Strong Weak
that is set up.
 The affinity with the mobile
Stationary phase and stationary
phase phase varies with the
solute.  Separation
occurs due to differences
in the speed of motion.
LAAQ-B-LC001B 4
Chromato-graphy / -graph / -gram /
-grapher

 Chromatography: Analytical technique


 Chromatograph: Instrument
 Chromatogram: Obtained “picture”
 Chromatographer: Person

LAAQ-B-LC001B 5
Three States of Matter and
Chromatography Types

Mobile phase

Gas Liquid Solid

Gas

Stationary
phase Liquid
Gas Liquid
chromatography chromatography
Solid

LAAQ-B-LC001B 6
Liquid Chromatography

 Chromatography in which the mobile phase


is a liquid.
 The liquid used as the mobile phase is
called the “eluent”.
 The stationary phase is usually a solid or a
liquid.
 In general, it is possible to analyze any
substance that can be stably dissolved in
the mobile phase.
LAAQ-B-LC001B 7
Interaction Between Solutes, Stationary
Phase, and Mobile Phase

 Differences in the interactions between the solutes and


stationary and mobile phases enable separation.
Solute
Degree of adsorption,
solubility, ionicity, etc.

Stationary
Mobile phase
phase

LAAQ-B-LC001B 8
Column Chromatography and
Planar Chromatography

Separation column

Paper or a
substrate coated
with particles

Packing material

Column Chromatography Paper Chromatography


Thin Layer Chromatography (TLC)
LAAQ-B-LC001B 9
Separation Process and Chromatogram
concentration for Column Chromatography

Chromatogram
Output

LAAQ-B-LC001B Time 10
Intensity of detector signal Chromatogram

tR
Peak tR : Retention time
t0 t0 : Non-retention time
h
A A : Peak area
h : Peak height

Time

LAAQ-B-LC001B 11
From Liquid Chromatography to High
Performance Liquid Chromatography

 Higher degree of separation!


 Refinement of packing material (3 to 10 µm)
 Reduction of analysis time!
 Delivery of eluent by pump
 Demand for special equipment that can
withstand high pressures

The arrival of high performance liquid chromatography!

LAAQ-B-LC001B 12
Flow Channel Diagram for High
Performance Liquid Chromatograph

Detector

Column

Pump Column oven


(thermostatic
column chamber)
Eluent Sample injection unit Drain
(mobile phase) (injector)
Data processor
Degasser

LAAQ-B-LC001B 13
Advantages of High Performance
Liquid Chromatography

 High separation capacity, enabling the batch


analysis of multiple components
 Superior quantitative capability and reproducibility
 Moderate analytical conditions
 Unlike GC, the sample does not need to be vaporized.
 Generally high sensitivity
 Low sample consumption
 Easy preparative separation and purification of
samples
LAAQ-B-LC001B 14
Fields in Which High Performance
Liquid Chromatography Is Used

 Biogenic substances  Food products


 Sugars, lipids, nucleic  Vitamins, food additives,
acids, amino acids, sugars, organic acids,
amino acids, etc.
proteins, peptides, steroids,
amines, etc.  Environmental
 Medical products samples
 Inorganic ions
 Drugs, antibiotics, etc.
 Hazardous organic
substances, etc.
 Organic industrial
products
 Synthetic polymers,
additives, surfactants, etc.
LAAQ-B-LC001B 15
HPLC Hardware: Part 1

Solvent Delivery System,


Degasser, Sample Injection Unit,
Column Oven

LAAQ-B-LC001B 16
Flow Channel Diagram for HPLC

Detector

Column

Pump Column Oven


(thermostatic
column chamber)
Eluent Sample injection unit Drain
(mobile phase) (injector)
Data processor
Degasser

LAAQ-B-LC001B 17
Solvent Delivery Pump

 Performance Requirements
 Capacity to withstand high load pressures.
 Pulsations that accompany pressure
fluctuations are small.
 Flow rate does not fluctuate.
 Solvent replacement is easy.
 The flow rate setting range is wide and the
flow rate is accurate.

LAAQ-B-LC001B 18
Solvent Delivery Pump:
Representative Pumping Methods

 Syringe pump
 Plunger pump
 Diaphragm pump

LAAQ-B-LC001B 19
Solvent Delivery Pump:
Schematic Diagram of Plunger Pump

Pump head
Motor and cam

Check
valves

Plunger
Plunger seal 10 -100µL

LAAQ-B-LC001B 20
Solvent Delivery Pump:
Single Plunger Type

Check valves

Plunger head

LAAQ-B-LC001B 21
Solvent Delivery Pump:
Dual Plunger Type

Check valves

Plunger heads

Type Type
LAAQ-B-LC001B 22
Gradient System

 Isocratic system
 Constant eluent composition
 Gradient system
 Varying eluent composition
 HPGE (High Pressure Gradient)
 LPGE (Low Pressure Gradient)

LAAQ-B-LC001B 23
Aim of Gradient System (1)

 In isocratic mode

CH3OH / H2O = 6 / 4
Long analysis time!!

Poor CH3OH / H2O = 8 / 2


separation!!

LAAQ-B-LC001B
(Column: ODS type) 24
Aim of Gradient System (2)

 If the eluent composition is changed gradually during


analysis...
Concentration of methanol in eluent

95%

30%

LAAQ-B-LC001B 25
High- / Low-Pressure Gradient System

Low-pressure
gradient unit

Mixer
Mixer

High-pressure gradient Low-pressure gradient


LAAQ-B-LC001B 26
Advantages and Disadvantages of
High- / Low-Pressure Gradient Systems

 High-pressure gradient system


 Highgradient accuracy
 Complex system configuration (multiple
pumps required)
 Low-pressure gradient system
 Simplesystem configuration
 Degasser required

LAAQ-B-LC001B 27
Degasser

 Problems caused by dissolved air in the eluent


 Unstable delivery by pump
 More noise and large baseline drift in detector cell

In order to avoid these problems, the eluent


must be degassed.

LAAQ-B-LC001B 28
Online Degasser

Regulator Vacuum chamber


Helium Polymeric film tube
cylinder

To pump
To pump
To draft

Drain valve

Eluent container Eluent container

Helium purge method Gas-liquid separation membrane method

LAAQ-B-LC001B 29
Sample Injection Unit (Injector)

 Performance Requirements
 No sample remaining in unit
 Minimal broadening of sample band
 Free adjustment of injection volume
 Minimal loss
 Superior durability and pressure resistance

LAAQ-B-LC001B 30
Manual Injector

From pump

To column
LOAD position
From pump

To column
INJECT position
LAAQ-B-LC001B 31
Manual Injector:
Operating Principle of Sample Injection

From pump From pump


Loop

Loop
To column
To column

LOAD INJECT

LAAQ-B-LC001B 32
Manual Injector:
Injection Method

 Syringe measurement method


 Itis desirable that no more than half the loop
volume is injected.
 Loop measurement method
 Itis desirable that at least 3 times the loop
volume is injected.

LAAQ-B-LC001B 33
Autosampler
(Pressure Injection Method)

From pump To column From pump To column

Sample Loop

LAAQ-B-LC001B
LOAD INJECT 34
Autosampler
(Total-Volume Injection Method)

From pump To column From pump To column

Needle

Sample vial
LOAD INJECT
Measuring pump
LAAQ-B-LC001B 35
Column Oven

 Air circulation heating type


 Block heating type
 Aluminum block heater
 Insulated column jacket type
 Water bath

LAAQ-B-LC001B 36
Tubing and Preparation for
Solvent Delivery

Prior to Analysis

LAAQ-B-LC001B 37
Tubing

 Material  O.D. (outer diameter)


 Stainless steel (SUS)  1.6 mm
 PEEK (polyether  I.D. (inner diameter)
ether ketone)  0.1 mm
 Fluororesin
 0.3 mm
 0.5 mm
 0.8 mm etc.

LAAQ-B-LC001B 38
Connectors

 Male nut (SUS) Ferrule


Ferrule (SUS)
 Sealing possible up to 40
MPa Male nut

 Male nut (PEEK)


 Can be connected without
any tools Male nut (PEEK)
 Resists pressures of up to
approx. 25 MPa

LAAQ-B-LC001B 39
Dead Volume
(Extra-column volume)

 Dead volume can cause peaks broadening.

Male nut Dead volume

Tube

Excellent connection Poor connection


LAAQ-B-LC001B 40
Mobile Phase

 Water  Organic Solvent


 “Ultrapure water” can be  HPLC-grade solvent can
used with confidence. be used with confidence.
 Commercial “distilled  Special-grade solvent is
water for HPLC” is also acceptable depending on
acceptable. the detection conditions.
 Care is required regarding
solvents containing
stabilizers (e.g.,
tetrahydrofuran and
chloroform)

LAAQ-B-LC001B 41
Replacement of Eluent

 Mutually insoluble solvents  Aqueous solutions containing


must not be exchanged salt and organic solvents
directly. must not be exchanged
directly.
Water Buffer solution

2-Propanol Water

Hexane Water-soluble
LAAQ-B-LC001B organic solvent 42
Mixing, Filtration, and Offline
Degassing of the Eluent

Decompression Membrane filter with pore


by aspirator size of approx. 0.45 µm

Decompression
by aspirator

Ultrasonic
cleaning unit

LAAQ-B-LC001B 43
Reversed Phase Chromatography
Part 1

Basic Principles

LAAQ-B-LC001B 44
Polarity of Substances

 Polarity  Miscibility of solvents


 Property of a substance  Solvents of similar
whereby the positions of the polarities can be easily
electrons give rise to dissolved together.
positive and negative poles  Polar and nonpolar
 Water: Polar molecules have a similar
Methane: Nonpolar relationship to that of water
and oil.
H –
H O
O
C H C C –
H H O
+ H
H H
H
LAAQ-B-LC001B Methane Water Acetic acid 45
Nonpolar (Hydrophobic) Functional Groups
and Polar (Hydrophilic) Functional Groups

 Nonpolar Functional  Polar Functional


Groups Groups
 -(CH2)nCH3  -COOH
 Alkyl groups  Carboxyl groups
 -C6H5  -NH2
 Phenyl groups  Amino groups
 -OH
 Hydroxyl groups

LAAQ-B-LC001B 46
Partition Chromatography

 A liquid (or a substance regarded as a


liquid) is used as the stationary phase,
and the solute is separated according to
whether it dissolves more readily in the
stationary or mobile phase.
 Liquid-liquid chromatography

LAAQ-B-LC001B 47
Normal Phase / Reversed Phase

Stationary
Mobile phase
phase
Normal High polarity Low polarity
phase (hydrophilic) (hydrophobic)

Reversed Low polarity High polarity


phase (hydrophobic) (hydrophilic)

LAAQ-B-LC001B 48
Reversed Phase Chromatography

 Stationary phase: Low polarity


 Octadecyl group-bonded silical gel (ODS)
 Mobile phase: High polarity
 Water, methanol, acetonitrile
 Salt is sometimes added.

LAAQ-B-LC001B 49
Separation Column for Reversed
Phase Chromatography

 C18 (ODS) type  Phenyl type


 C8 (octyl) type  TMS type
 C4 (butyl) type  Cyano type

CH2 CH2 CH2 CH2 CH2 CH2 CH2 CH2 CH2


Si -O-Si
CH2 CH2 CH2 CH2 CH2 CH2 CH2 CH2 CH3

C18 (ODS)

LAAQ-B-LC001B 50
Effect of Chain Length of
Stationary Phase

C8

Medium
C18 (ODS)

Strong C4

Weak

LAAQ-B-LC001B 51
Hydrophobic Interaction

H2O H2O H2O


H2O
H2O
H2O Nonpolar solute H2O
H2O
H2O If a nonpolar H2O
H2O H2O substance is added... H2O H2O
Network of hydrogen bonds …the network is broken and...

H2O H2O
H2O
H2O H2O …the nonpolar substance
H2O H2O is pushed to a nonpolar
location.
Nonpolar solute

LAAQ-B-LC001B Nonpolar stationary phase 52


Relationship Between Retention
Time and Polarity

C18 (ODS) OH

Weak
Strong
CH3

LAAQ-B-LC001B 53
Basic Settings for Eluent Used in
Reversed Phase Mode

 Water (buffer solution) + water-soluble organic


solvent
 Water-soluble organic solvent: Methanol
Acetonitrile
Tetrahydrofuran etc.
 The mixing ratio of the water (buffer solution) and
organic solvent has the greatest influence on
separation.
 If a buffer solution is used, its pH value is an
important separation parameter.

LAAQ-B-LC001B 54
Difference in Solute Retention Strengths
for Water and Water-Soluble Organic
Solvents

Tightly packed network Loose network

H2O H2O CH3OH CH3OH


H2O H2O
H2O CH3OH
H2O H2O CH3OH CH3OH

CH3OH
Nonpolar solute CH3OH
Nonpolar solute

Nonpolar stationary phase


LAAQ-B-LC001B 55
Relationship between Polarity of Eluent and
Retention Time in Reversed Phase Mode

Eluent: Methanol / Water

60/40

70/30

80/20
LAAQ-B-LC001B 56
Chromatogram Parameters

Methods for Expressing Separation


and Column Performance

LAAQ-B-LC001B 57
Strength of detector signal Retention Factor, k

tR t R  t0
k
t0
t0
tR: Retention time
t0: Non-retention time

Time

LAAQ-B-LC001B 58
Theoretical Plate Number, N

2
tR
N  16
W
2
tR
 5.54
H W1/ 2
W1/2
tR  H
2
H1/2
 2
W Area
LAAQ-B-LC001B 59
Evaluation of Column Efficiency Based on
Theoretical Plate Number

 If the retention times are  If the peak width is the


the same, the peak width same, the retention time is
is smaller for the one with longer for the one with the
the larger theoretical plate larger theoretical plate
number. number.
N: Large N: Small
N: Large

N: Small

LAAQ-B-LC001B 60
Separation Factor, a

 Separation factor: Ratio of k’s of two peaks

k1 k2 k2

k1
( k 2  k1 )

LAAQ-B-LC001B 61
Resolution, RS

tR1 tR2

t R 2  t R1
RS 
1
(W1  W2 )
2
t R 2  t R1
 1.18 
W1/ 2 h,1  W1/ 2 h , 2
W1/2h,1 W1/2h,2 h1/2

W1 W2
LAAQ-B-LC001B 62
Resolution Required for Complete
Separation

(tR2 - tR1) (tR2 - tR1)

W1 W2 W1 W2
tR2 - tR1 = W1 = W2 tR2 - tR1 = W1 = W2
RS = 1 RS = 1
If the peaks are isosceles triangles, If the peaks are Gaussian distributions,
they are completely separated. RS > 1.5 is necessary for complete separation.
LAAQ-B-LC001B 63
Relationship Between Resolution
and Other Parameters

 The resolution is a
tR 2  tR1
function of the RS 
1
separation factor, the
(W1  W2 )
theoretical plate 2
number, and the
1  1 k’2
retention factor.  N
4  k’2  1
 The separation can be
improved by improving
these 3 parameters!

LAAQ-B-LC001B 64
Contribution of Capacity Factor to
Resolution

 Increasing the capacity 1.0

Contribution ratio for resolution


factor improves 0.8
separation!
0.6
 A capacity factor of
around 3 to 10 is 0.4

appropriate. Exceeding 0.2


this just increases the 0.0
analysis time. 0 5 10 15 20
Capacity factor

LAAQ-B-LC001B 65
Contribution of Theoretical Plate
Number to Resolution

 The resolution 2.0

Contribution factor for resolution


increases in
proportion to the
square root of the 1.0
theoretical plate
number.
0.0
0 10000 20000 30000
Theoretical plate number

LAAQ-B-LC001B 66
To Improve Separation...

Before
adjustment

k’ increased Eluent replaced with one


of lower elution strength.

Column replaced with one of


N increased superior performance.
Column lengthened.

Column (packing material) replaced.


 increased Eluent composition changed.
Column temperature changed.
LAAQ-B-LC001B 67
pH Buffer Solution Used for Eluent

Selection and Preparation of


Buffer Solution

LAAQ-B-LC001B 68
Acid Dissociation Equilibrium

H+ If an acid is added...

...the equilibrium shifts to


the left to offset the
increase in H+.

HA A- + H+
If an alkali is
added...
The equilibrium always shifts
…the equilibrium shifts in a way that offsets changes.
to the right to offset the OH-
decrease in H+.
LAAQ-B-LC001B 69
Acid Dissociation Constant and
pH-Based Abundance Ratio
1.0
HA A- + H+
CH3COOH CH3COO-
0.8
The acid dissociation constant, Ka,

Abundance ratio
is defined as follows:
0.6
 
[A ][ H ]
Ka  0.4
[HA ]
[A  ] 0.2
pH  pK a  log
[HA ] 0.0
1 2 3 4 5 6 7 8 9
pH
 pH   log[ H ]  
pKa
 
 pK   log K  Relationship Between Abundance Ratio
 a a  and pH Value of Acetic Acid and Acetic Acid Ions
LAAQ-B-LC001B 70
Preparing pH Buffer Solution

 Use a weak acid with a pKa value close to the


desired pH value.
 Example: Preparing a buffer solution for a pH value of
around 4.8.
 Use acetic acid, which has a pKa value of 4.8.
 Make the concentrations of HA and A- roughly
equal.
 Mix an acid with its salt.
 Example: Mix acetic acid and sodium acetate so that they
have the same molar concentration.
LAAQ-B-LC001B 71
Buffer Solutions Used for HPLC Eluent

 Requirements  Commonly Used Acids


 High buffering power at  Phosphoric acid
 pKa 2.1, 7.2, 12.3
prescribed pH.
 Acetic acid
 Does not adversely
 pKa 4.8
affect detection.
 Citric acid
 Does not damage
 pKa 3.1, 4.8, 6.4
column or equipment.
 Concentration
 Inexpensive.
 If only to adjust pH, 10
mmol/L is sufficient.

LAAQ-B-LC001B 72
Characteristics of Phosphate
Buffer Solution

 Advantages  Disadvantages
 Three dissociation  No volatility
states  Difficult to use for
(pKa 2.1, 7.2, 12.3) LCMS or evaporative
 Possible to prepare light scattering
buffer solutions of detection.
various pH values.
 No UV absorption
 Inexpensive

LAAQ-B-LC001B 73
Reversed Phase Chromatography
Part 2

Consideration of Analytical
Conditions

LAAQ-B-LC001B 74
Guidelines for Setting Mobile Phase
Conditions (1)
Neutral (Nonionic) Substances

 Eluent Composition
 Water / acetonitrile
 Water / methanol

 Separation Adjustment
 Changing the mixing ratio of the water and
organic solvent
 Changing the type of organic solvent

LAAQ-B-LC001B 75
pH of Eluent and Retention of Ionic
Solutes

COOH
Acidic
Increased
hydrophobicity
pH of eluent

COO
Alkaline
Increased
hydrophilicity
+
H
LAAQ-B-LC001B 76
Guidelines for Setting Mobile Phase
Conditions (2)
Acidic (Anionic) Substances

 Eluent Composition
 Acidic buffer solution / acetonitrile
 Acidic buffer solution / methanol

Increase retention strength by making


the eluent acidic and suppressing
ionization!

LAAQ-B-LC001B 77
Analysis of Basic Substances (1)
Problems Encountered with Alkaline Eluents

With alkaline eluents, although the


ionization of basic substances is
suppressed, and the retention
N+ N strength increases...
H

OH Si
O

OH Si OH
…silica gel dissolves in alkalis,
so the packing material
OH deteriorates rapidly.
OH
OH
LAAQ-B-LC001B 78
Analysis of Basic Substances (2)
Influence of Residual Silanol Groups

Basic substances interact with the


residual silanol groups, causing
delayed elution and tailing.

Si
O

Si -O-Si-O
Residual silanol group N+
O H
Si

LAAQ-B-LC001B 79
Analysis of Basic Substances (3)
Addition of Sodium Perchlorate

ClO4 Ion pair


N+
H

Si
O

Si Basic substances form ion pairs with perchlorate


ions, thereby balancing the charge and
increasing the retention strength.

LAAQ-B-LC001B 80
Guidelines for Setting Mobile Phase
Conditions (3)
Basic Substances (Cationic Substances)

 Eluent Composition
 Acidic buffer solution containing anions with a low
charge density (e.g., perchlorate ions) / acetonitrile
 As above / methanol

Making eluent acidic


 Suppresses dissociation of residual silanol groups
 Prevents tailing!

Adding perchlorate ions


 Forms ion pairs  Increases retention strength!
 Suppresses tailing!
LAAQ-B-LC001B 81
Reversed Phase Ion Pair
Chromatography

 Increase the retention strength by adding an ion pair


reagent with the opposite charge to the target
substance into the eluent.
Ion pair formation Ion pair formation

Ion exchange-like effect Ion exchange-like effect


LAAQ-B-LC001B
Basic Substance Acidic Substance 82
Representative Ion Pair Reagents

 Anionic Compounds
 Tetra-n-butylammonium hydroxide (TBA)
 Cationic Compounds
 Pentanesulfonic acid sodium salt (C5)
 Hexanesulfonic acid sodium salt (C6)
 Heptanesulfonic acid sodium salt (C7)
 Octanesulfonic acid sodium salt (C8)

LAAQ-B-LC001B 83
Points to Note Concerning the Use
of Ion Pairs

 Selection of Ion Pair Reagent


 In general, the retention strength increases with the length of
the alkyl chain.
 pH of Eluent
 The retention strength changes according to whether or not
ionization takes place.
 Concentration of Ion Pair Reagent
 In general, the retention strength increases with the ion pair
concentration, but there is an upper limit.
 Proportion of Organic Solvent in Eluent
 Optimize the separation conditions by considering the type and
concentration of the ion pair reagent.

LAAQ-B-LC001B 84
HPLC Separation Modes

Separation Modes Other Than


Reversed Phase Chromatography

LAAQ-B-LC001B 85
HPLC Separation Modes

 Adsorption (liquid-solid) chromatography


 Partition (liquid-liquid) chromatography
 Normal phase partition chromatography
 Reversed phase partition chromatography

 Ion exchange chromatography


 Size exclusion chromatography

LAAQ-B-LC001B 86
Adsorption Chromatography

 A solid such as silica gel is used as the


stationary phase, and differences, mainly
in the degree of adsorption to its surface,
are used to separate the solutes.
 Liquid-solid chromatography
 The retention strength increases with the
hydrophilicity of the solute.

LAAQ-B-LC001B 87
Partition Chromatography

 A liquid (or a substance regarded as a


liquid) is used as the stationary phase, and
the solute is separated according to
whether it dissolves more readily in the
stationary or mobile phase.
 Liquid-liquid chromatography

LAAQ-B-LC001B 88
Normal Phase and Reversed Phase

Solid phase Mobile phase

Normal High polarity Low polarity


phase (hydrophilic) (hydrophobic)
Reversed Low polarity High polarity
phase (hydrophobic) (hydrophilic)

LAAQ-B-LC001B 89
Normal Phase (Partition)
Chromatography

 Partition chromatography in which the


stationary phase has a high polarity
(hydrophilic) and the mobile phase has a
low polarity (hydrophobic)
 Essentially based on the same separation
mechanism as adsorption chromatography
in which the stationary phase has a
hydrophilic base, such as silica gel

LAAQ-B-LC001B 90
Invention of Chromatography by
M. Tswett
Ether
Chromatography

Chlorophyll Colors

CaCO3

LAAQ-B-LC001B 91
Stationary Phase and Mobile Phase Used
in Normal Phase Mode

 Stationary Phase
 Silica gel: -Si-OH
 Cyano type: -Si-CH2CH2CH2CN
 Amino type: -Si-CH2CH2CH2NH2
 Diol type: -Si-CH2CH2CH2OCH(OH)-CH2OH

 Mobile Phase
 Basic solvents: Aliphatic hydrocarbons,
aromatic hydrocarbons, etc.
 Additional solvents: Alcohols, ethers, etc.

LAAQ-B-LC001B 92
Relationship between Hydrogen Bonding
and Retention Time in Normal Phase Mode

SiOH HO
Strong
SiOH
Weak
Very weak
OH

Steric hindrance
LAAQ-B-LC001B 93
Relationship Between Eluent Polarity and
Retention Time in Normal Phase Mode

Eluent: Hexane/methanol

100/0

98/2

95/5
LAAQ-B-LC001B 94
Comparison of Normal Phase and
Reversed Phase

 Normal Phase  Reversed Phase


 Effective for separation  Wide range of applications
of structural isomers  Effective for separation of
 Offers separation homologs
selectivity not available  Stationary phase has long
with reversed phase service life
 Stabilizes slowly and is  Stabilizes quickly
prone to fluctuations in  Eluents are inexpensive
retention time and easy to use
 Eluents are expensive

LAAQ-B-LC001B 95
Ion Exchange Chromatography

R
Anion exchange N+ R
R

++++
Cation exchange SO3- + +
++++
Electrostatic interaction
(Coulomb force)
LAAQ-B-LC001B 96
Stationary Phase Used in Ion
Exchange Mode

 Base Material
 Resin is often used.
 Silica gel is also used.

 Cation Exchange Column


 Strong cation exchange (SCX) -SO3-
 Week cation exchange (WCX) -COO-
 Anion Exchange Column
 Strong anion exchange (SAX) -NR3+
 Week anion exchange (WAX) -NHR2+

LAAQ-B-LC001B 97
Dependence of Exchange Capacity of Ion
Exchanger on pH of Eluent

Strongly acidic Strongly basic


cation exchanger anion exchanger
Exchange capacity

Exchange capacity
Weakly acidic
cation exchanger Weakly basic
anion exchanger

0 7 14 0 7 14
pH pH
Cation exchange mode Anion exchange mode
LAAQ-B-LC001B 98
Relationship between Retention Time and
Salt Concentration of Eluent in Ion
Exchange Mode

Resin Resin Resin


The exchange groups An eluent ion is The solute ion is driven
are in equilibrium with driven away away by an eluent ion
anions in the eluent. and a solute ion and is adsorbed by the
is adsorbed. next exchange group.

Solute ions and eluent ions compete for ion exchange groups.

If the salt concentration of the eluent increases, the solutes are eluted sooner.
LAAQ-B-LC001B 99
Ion Exclusion Chromatography

H+

H+

H+

Strong acid ions are repelled by Depending on the level of dissociation,


charge and cannot enter the pore. some weak acid ions can enter the pore.

LAAQ-B-LC001B 100
Size Exclusion Chromatography

 Separation is based on the size (bulkiness)


of molecules.
 The name varies with the application field!
 Size Exclusion Chromatography (SEC)
 Gel Permeation Chromatography (GPC)
 Chemicalindustry fields, synthetic polymers,
nonaqueous systems
 Gel Filtration Chromatography (GFC)
 Biochemicalfields, biological macromolecules,
aqueous systems

LAAQ-B-LC001B 101
Principle of Size Exclusion Mode

The size of the solute molecules


determines whether or not they can
enter the pores.

Packing
material

LAAQ-B-LC001B 102
Relationship Between Molecular Weight and
Retention Time in Size Exclusion Mode

(logarithmic axis) Exclusion limit


Molecular weight

Permeability limit

Elution capacity

LAAQ-B-LC001B 103
Creating a Molecular Weight
Calibration Curve

For separation of large molecular weights

(logarithmic axis)
Molecular weight

For wide-range separation


(mix gel)

Elution capacity
For separation of small molecular weights
LAAQ-B-LC001B 104
Calculating Molecular Weights

Chromatogram  Various Average


Molecular Weights
 Mn: Number-average
molecular weight
 Mw: Weight-average
Calibration curve molecular weight
 Mz: Z-average molecular
weight, etc.
 Molecular weights and
molecular weight
distributions are calculated
Retention time using special calculation
LAAQ-B-LC001B software. 105
Guidelines for Selecting Separation Mode (1)
Required Information

 Soluble solvent
 Molecular weight
 Structural formula and chemical
properties
 Do the substances ionize?
 Is there UV absorption or fluorescence?
 Is derivatization possible? etc.

LAAQ-B-LC001B 106
Guidelines for Selecting Separation Mode (2)
Basic Policy

 Reversed phase mode using an ODS column


is the first choice!
 Exceptions
 Large molecular weight (> 2,000)  Size exclusion
 Optical isomers  Chiral column
 Stereoisomers, positional isomers  Normal phase /
adsorption
 Inorganic ions  Ion chromatography
 Sugars, amino acids, short-chain fatty acids
 Special column
LAAQ-B-LC001B 107
HPLC Hardware: Part 2

Detectors and Their Ranges of Application

LAAQ-B-LC001B 108
Detection Condition Requirements

 Sensitivity
 Thedetector must have the appropriate level of
sensitivity.
 Selectivity
 Thedetector must be able to detect the target
substance without, if possible, detecting other
substances.
 Adaptability to separation conditions
 Operability, etc.
LAAQ-B-LC001B 109
Representative HPLC Detectors

 UV-VIS absorbance detector


 Photodiode array-type UV-VIS absorbance
detector
 Fluorescence detector
 Refractive index detector
 Evaporative light scattering detector
 Electrical conductivity detector
 Electrochemical detector
 Mass spectrometer
LAAQ-B-LC001B 110
UV-VIS Absorbance Detector

C: Concentration
Detection cell

Ein Eout

A
l

A = e·C·l = –log (Eout / Ein) C


(A: absorbance, E: absorption coefficient)

LAAQ-B-LC001B 111
Optical System of UV-VIS
Absorbance Detector

Grating
Sample cell
l Ein Eout Photodiode

Ein Ein Photodiode

Reference cell

D2 / W lamp

LAAQ-B-LC001B 112
Spectrum and Selection of
Detection Wavelength

The longer wavelength


is more selective.

200 250 300 350


Wavelength [nm]
LAAQ-B-LC001B 113
Optical System of Photodiode
Array Detector

Sample cell
Grating

A single photodiode
D2 / W lamp measures the absorbance for
the corresponding wavelength
at a resolution of approx. 1 nm.

Photodiode array

LAAQ-B-LC001B 114
Data Obtained with a Photodiode
Array Detector

Spectrum

Chromatogram
Absorbance

LAAQ-B-LC001B
Retention time 115
Advantages of Photodiode Array
Detectors

 Peak Identification Using Spectra


 Complementation of identification based on
retention time
 Library searches
 Evaluation of Peak Purity
 Purityevaluation performed by comparison
of the shape of spectra from the peak
detection start point to the peak detection
end point

LAAQ-B-LC001B 116
Fluorescence Detector

Excitation wavelength

+ hv1 *

* hv2 +
Fluorescence wavelength
Excited state
Quasi-excited state
hv1
hv2
Fluorescence

LAAQ-B-LC001B
Ground state 117
Optical System of Fluorescence Detector

Xenon lamp
Fluorescence
grating
Photomultiplier tube

Fluorescence

Excitation
Excitation grating Sample cell
light

LAAQ-B-LC001B 118
Fluorescence Derivatization Reagents

 OPA Reagent (Reacts with Primary Amines)


S-R’
CHO
+ R-NH2 N-R
CHO R’-SH
o-phthalaldhyde
(OPA)

 ADAM Reagent (Reacts with Fatty Acids)


+ R-COOH

CHN2 CH2OCOR
9-anthryldiazomethane
(ADAM)
LAAQ-B-LC001B 119
Differential Refractive Index
Detector (Deflection-Type)

Light-receiving unit
Reference cell

Light

Sample cell

LAAQ-B-LC001B 120
Optical System of Differential Refractive
Index Detector (Deflection-Type)

Slit W lamp

Reference cell
Sample cell
The slit image moves if the
refractive index inside the
flow cell changes.

Photodiode

LAAQ-B-LC001B 121
Evaporative Light Scattering Detector

Light-receiving unit
Drift tube

Nebulizer

Column eluate

Nebulizer gas
Drain Assist gas

Light source

The column eluate is evaporated and the light scattered


by the particles of nonvolatile substances is detected.

LAAQ-B-LC001B 122
Electrical Conductivity Detector

Pure water NaCl aqueous


solution

Cl- Na+

The bulb does not light with water. The bulb lights if there are ions.

LAAQ-B-LC001B 123
Principle of Electrical Conductivity Detector

I A
V K   k
I
E L
L
k  K
A
A A K: Electrical conductivity [S]
I: Electric current [A]
E: Voltage [V]
A: Electrode surface area [cm2]
L Electrode L: Distance between electrodes [cm]
k: Specific electrical conductivity [S•cm-1]

LAAQ-B-LC001B 124
Limiting Equivalent Ion Conductance, l
[S•cm2/mol], in Aqueous Solution (25ºC)

Cation l Anion l
H+ 349.8 OH– 198.3
Li+ 38.6 F– 55.4
Na+ 50.1 Cl– 76.3
K+ 73.5 Br– 78.1
NH4+ 73.5 NO3– 71.4
(CH3)3NH+ 47.2 CH3COO– 40.9
Mg2+ 53.0 C6H5COO– 32.3
Ca2+ 59.5 SO42– 80.0

LAAQ-B-LC001B 125
Electrochemical Detector

Electrode
HO R

HO
2e-
O R
+ 2H+
O

LAAQ-B-LC001B 126
Cell Structure of Electrochemical
Detector (Amperometric Type)

Reference electrode Working electrode


(Ag/AgCl) (glassy carbon)

Eluent

Electrode couple
LAAQ-B-LC001B 127
Mass Spectrometer (LCMS)

Atmospheric
pressure High vacuum

Quadrupole MS analyzer
API probe

Electron
multiplier tube

RP TMP1 TMP2
(high vacuum pumps)
LAAQ-B-LC001B 128
Atmospheric Pressure Ionization

Electrospray Ionization (ESI)


+
+ +
++- - + -- + +
+- -+- +
+-+-+-
+ +- + - ++ +
- + +
+

3) n E
Liquid
Liquid Samples
Sample

1)
+

2) of S

Io
C va
C
High
High Voltage

ou p
Neburaizing
Nebulizing Gas

Ev o l
ha
Voltage

lo ora
ap ve
rg

n
Gas

ed

ol nt

E x tio
at
D

cl n
io
ro

us
n
pl

io
et

n
Atmospheric Pressure Chemical Ionization (APCI)

Molecular ion reaction


Liquid Sample
Liquid Samples Heater
Heater
Neburaizing Gas
Nebulizing Gas Colona Discharge
Corona Discharge
LAAQ-B-LC001B Nnndle
Needle 129
Advantages of LCMS (1)

 Quantitative analysis with excellent selectivity


m/z=100
A
TIC A:100 B
B:100
C:150 D:150

m/z=150
C D

LAAQ-B-LC001B 130
Advantages of LCMS (2)

 Peaks can be identified with MS spectra.

M/Z

M/Z

LAAQ-B-LC001B M/Z 131


Comparison of Detectors

Possibility of
Selectivity Sensitivity
Gradient System
Light-absorbing
Absorbance ng Possible
substances

Fluorescence Fluorescent substances pg Possible

Differential
None µg Impossible
refractive index
Evaporative light
Nonvolatile substances µg Possible
scattering
Electrical
Ionic substances ng Partially possible
conductivity
Oxidizing / reducing
Electrochemical pg Partially possible
substances
LAAQ-B-LC001B Note: The above table indicates general characteristics. There are exceptions. 132
Post-Column Derivatization

Reaction
chamber

Pump
Reaction
solution
LAAQ-B-LC001B 133
Application Examples of Post-
Column Methods

 Amino Acids  Bromate Ions


 Orthophthalic acid, OPA  Tribromide ionization
(fluorescence) (ultraviolet absorption)
 o-Dianisidine
 Ninhydrin (visible absorption)
(visible absorption)
 Reducing Sugars  Cyanide Ions
 Arginine (fluorescence)
 Chlorination - pyrazolone
 Carbamate Pesticides (visible absorption)
 Alkaline hydrolysis - OPA  Transition Metal Ions
(fluorescence)  4-(2-Pyridylazo)
resorcinol, PAR (visible
absorption)

LAAQ-B-LC001B 134
Quantitative Analysis

Absolute Calibration Curve Method


and Internal Standard Method

LAAQ-B-LC001B 135
Qualitative Analysis

 Identification based on retention time


 Acquisition of spectra with detector
 UV spectra
 MS spectra

 Transfer to other analytical instruments


after preparative separation

LAAQ-B-LC001B 136
Quantitative Analysis

 Quantitation performed with peak area or


height.
 Calibration curve created beforehand
using a standard.
 Absolute calibration curve method
 Internal standard method
 Standard addition method

LAAQ-B-LC001B 137
Calibration Curve for Absolute
Calibration Curve Method
Area
Concentration
A1 Calibration curve
C1
A4

A2

Peak area
A3
C2

A2
A3
C3
A1

A4
C4 C1 C2 C3 C4
LAAQ-B-LC001B
Concentration 138
Calibration Curve for Internal
Standard Method
Concentration Area

Area for target substance / Area for internal standard


Target Internal
substance standard A1 AIS Calibration curve
C1 CIS A4 /AIS

A2 AIS A3 /AIS
C2 CIS

A2 /AIS
A3 AIS
C3 CIS
A1/AIS

A4 AIS
C4 CIS C1/CIS C2 /CIS C3 /CIS C4 /CIS
Concentration of target substance /
LAAQ-B-LC001B 139
Concentration of internal standard
Advantages of Internal Standard
Method (1)

 Not affected by inconsistencies in injection volume.


IS
X AX / AIS
10 µL
injected

Same area
ratio
IS
X
9 µL
injected
CX / CIS
LAAQ-B-LC001B 140
Advantages of Internal Standard
Method (2)

 Not affected by the pretreatment recovery rate.


IS
X

AX / AIS
100%
recovery
rate

Same area
ratio
IS
90% X
recovery
rate
CX / CIS
LAAQ-B-LC001B 141
Selection Criteria for Internal Standard

 It must have similar chemical properties to the


target substance.
 Its peak must appear relatively near that of the
target substance.
 It must not already be contained in the actual
samples.
 Its peak must be completely separated from those
of other sample components.
 It must be chemically stable.
LAAQ-B-LC001B 142
Sample Pretreatment

Tasks Performed Before Injection

LAAQ-B-LC001B 143
Objectives of Pretreatment

 To improve the accuracy of quantitative


values
 To improve sensitivity and selectivity
 To protect and prevent the deterioration of
columns and analytical instruments
 To simplify measurement operations and
procedures
 To stabilize target substances
LAAQ-B-LC001B 144
Substances That Must Not Be
Injected into the Column

 Insoluble substances (e.g., microscopic


particles and precipitation)
 Substances that are precipitated in the
eluent
 Substances that irreversibly adsorb to the
packing material
 Substances that dissolve, or chemically
react, with the packing material
LAAQ-B-LC001B 145
Filtration and Centrifugal Separation

 In general, filter every


sample before injection!
 It is convenient to use a
disposable filter with a
pore diameter of approx.
0.45 µm.
 Centrifugal separation is
applicable for samples
Filter Syringe
that are difficult to filter.

LAAQ-B-LC001B 146
Deproteinization

 Precipitation
 Addition of organic solvent (e.g., acetonitrile)
 Addition of acid (e.g., trichloroacetic acid,
perchloric acid)
 Addition of heavy metal or neutral salt

 Ultrafiltration

LAAQ-B-LC001B 147
Solid Phase Extraction

(1) (2) (3) (4)


Conditioning Sample addition Rinsing Elution
Solvent with
low elution
strength

Solvent with
high elution
strength

Target
component
Unwanted
components

LAAQ-B-LC001B 148
Pre-Column Derivatization

 OPA Reagent (Reacts with Primary Amines)


S-R’
CHO
+ R-NH2 N-R
CHO R’-SH
o-phthalaldhyde
(OPA)

 2,4-DNPH (Reacts with Aldehydes and Ketones)


R
NHNH2 NHN=C
R
+ C=O R’
R’ H+
O2 N NO2 O2 N NO2
2,4-dinitrophenylhydrazine
(2,4-DNPH)
LAAQ-B-LC001B 149
Evaluation of the Reliability of
Analysis

Validation of Analytical Methods

LAAQ-B-LC001B 150
What Is “Validation of Analytical
Methods”?

 Scientifically  Validation
demonstrating that the characteristics
analytical methods  Accuracy / trueness
concur with the  Precision
intended purpose (i.e.,  Specificity
that errors are within a  Detection limit
permissible range)  Quantitation limit
 Evaluating required  Linearity
items from the  Range
validation  (Robustness)
characteristics
LAAQ-B-LC001B 151
Accuracy / Trueness

 Definition  Evaluation Method


 Degree of bias in  Comparison with
measurements obtained with theoretical values (or
analytical procedures authenticated values)
 Difference between true value
 Comparison with results
and grand mean of
measurements
obtained using other
analytical procedures for
True value which the accuracy
(trueness) is known
 Recovery test

Measurement

Average 95% confidence interval


LAAQ-B-LC001B 152
Precision

 Definition  Repeatability / Intra-


 Degree of coincidence of Assay Precision
series of measurements  Precision of

obtained by repeatedly measurements taken over


analyzing multiple a short time period under
the same conditions
samples taken from a
homogenous test  Intermediate Precision
substance  Reproducibility
 Variance, standard
deviation, or relative
standard deviation of
measurements
LAAQ-B-LC001B 153
Specificity

 Definition  Evaluation Method


 The ability to accurately  Confirmation that the target
analyze the target substance substance can be
discriminated (separated)
in the presence of other
from co-existing
expected substances components, related
 The discrimination capability substances, decomposition
of the analytical methods products, etc.
 Multiple analytical  If reference standards for
procedures may be impurities cannot be
combined in order to attain obtained, the measurement
results for samples thought
the required level of
to contain the impurities are
discrimination compared.

LAAQ-B-LC001B 154
Detection Limit

 Definition  Evaluation Method


 The minimum quantity of  Calculated from the
a target substance that standard deviation of
measurements and the
can be detected.
slope of the calibration
 Quantitation is not curve.
absolutely necessary.  DL = 3.3 /slope
(: Standard deviation of
measurements)
(Slope: Slope of calibration
curve)
 Calculated from the
signal-to-noise ratio.
 Concentration for which
S/N = 3 or 2
LAAQ-B-LC001B 155
Quantitation Limit

 Definition  Evaluation Method


 The minimum quantity of a  Calculated from the standard
target substance that can deviation of measurements
and the slope of the
be quantified
calibration curve.
 Quantitation with an  QL = 10 /slope
appropriate level of (: Standard deviation of
accuracy and precision measurements)
must be possible. (In (Slope: Slope of calibration
curve)
general, the relative
standard deviation must  Calculated from the signal-to-
noise ratio.
not exceed 10%.)
 Concentration for which S/N
= 10

LAAQ-B-LC001B 156
Linearity

 Definition  Evaluation Method


 The ability of the analytical  Samples containing
method to produce different quantities of the
measurements for the target substance (usually
quantity of a target 5 concentrations) are
substance that satisfy a analyzed repeatedly, and
linear relationship. regression equations and
 Values produced by correlation coefficients are
converting quantities or obtained.
measurements of the  Residuals obtained from
target substance using a the regression equations
precisely defined formula of the measurements are
may be used. plotted, and it is confirmed
that there is no specific
slope.
LAAQ-B-LC001B 157
Range

 Definition  Evaluation Method


 The region between  The accuracy,
the lower and upper precision, and linearity
limits of the quantity of are investigated for
a target substance that samples containing
gives appropriate quantities of a target
levels of accuracy and substance that
precision correspond to the
lower limit, upper limit,
and approximate
center of the range.
LAAQ-B-LC001B 158
Robustness

 Definition  Evaluation Method


 The ability of an  Some or all of the
analytical procedure variable factors (i.e.,
to remain unaffected the analytical
by small changes in conditions) are
analytical conditions. changed and the
effects are evaluated.

LAAQ-B-LC001B 159
Maintenance of Separation Column

Extending the Column’s Service Life

LAAQ-B-LC001B 160
Silica-Based Packing Materials and
Resin-Based Packing Materials

Silica-Based Resin-Based

Generally a wide
pH range 2 - 7.5
range
Organic Significant
No restrictions
solvent restrictions
Pressure Low pressure
25 MPa max.
resistance resistance
Depends on
Temperature 60ºC max.
packing material
LAAQ-B-LC001B 161
General Handling of Columns

 Observe restrictions  Use as low a load


related to solvents and pressure as possible.
the pH range.  Do not exceed the upper
 Never allow the pressure limit.
packing material to  Do not subject the column
dry. to sudden pressure
 Do not allow solids or changes.
microscopic particles  Do not subject the
to enter the column. column to intense
 Filter samples. shocks.

LAAQ-B-LC001B 162
Typical Problems (1)
Column Clogging

 Preventive Measures  Corrective Action


 Filter samples.  Check for clogging in
parts other than the
 Check that samples column.
dissolve in the eluent.
 Rinse with an appropriate
 Get in the habit of solvent.
observing pressure  Connect the column in
values. reverse and flush out the
insoluble substances at a
low flow rate.
 Open the column end
and perform ultrasonic
cleaning of the filter.
LAAQ-B-LC001B 163
Typical Problems (2)
Peak Deformation

Cause Corrective Action


Sample overload Reduce the sample injection volume or
concentration.
Inappropriate sample Replace the sample solvent with one of a
solvent low elution capacity.

Dirt Rinse the column.


Gap in column inlet Repair the column by supplementing it
with packing material.

Influence of secondary Rinse the column.


retention effects Replace the column with one that is only
minimally influenced.
LAAQ-B-LC001B 164
Typical Problems (3)
Decrease in Retention Time

 Check whether the  If the column is


cause of the problem identified as the
is not the column. cause...
 Eluent composition  Rinsing
 Eluent flow rate  Replacement
 Column temperature

LAAQ-B-LC001B 165
Typical Problems (4)
Baseline Drift

 Check whether the  If the column is


cause of the problem identified as the
is not the column. cause...
 If the problem persists  Rinsing
when the column is  Review of
removed, it is caused temperature control
by the eluent, the  Replacement
solvent delivery system
(pump or degasser), or
the detector.

LAAQ-B-LC001B 166
Guard Column and Pre-column

Guard column

Pre-column

LAAQ-B-LC001B 167
Column Rinsing

 Use an eluent with a high elution capacity


 Reversed phase mode: Solution with a high proportion of
organic solvent
 Ion exchange mode: Solution with a high salt
concentration
 Consider secondary retention effects
 To remove basic substances from a reversed phase
column  Use an acidic solution and add an ion pair
reagent.
 To remove hydrophobic substances from an ion
exchange column  Add an organic solvent.
LAAQ-B-LC001B 168
Checking Column Performance

2
tR
N  16
W
2
tR
 5.54
H W1/ 2
W1/2
tR  H
2
H1/2
 2
W Area
LAAQ-B-LC001B 169
Column Storage

 Storage Solution  Storage Conditions


 It is generally safe to use  Insert an airtight stopper in
the same storage solution the column end. Never allow
as used at shipment. the packing material to dry.
 In order to prevent  Make a record of the storage
putrefaction, alcohol or solution and final usage
some other preservative conditions and store it
substance may be added. together with the column.
 Store the column in a
location not subject to shocks
or sudden temperature
changes.

LAAQ-B-LC001B 170