Professional Documents
Culture Documents
Prepared by Michigan Department of Environmental Quality Operator Training and Certification Unit
Prepared by Michigan Department of Environmental Quality Operator Training and Certification Unit
Prepared By
Michigan Department of Environmental Quality
Operator Training and Certification Unit
Electrons
Orbits = Energy Level
wavelength - (lambda)
Wave
(time)
frequency - (gamma)
LIGHT
The Energy (E) of the Photon is
Related to the wavelength () and the
frequency () of the Wave
E = h = hc
Where:
h = Planck’s Constant
c = Velocity of Light
LIGHT
Constants
E = hc
Every wavelength ()
has a specific
Energy level.
Summary
700 nm
COLOR
RESULTS WHEN
RADIATION IS ABSORBED
BY AN ELEMENT OR
BY A COMPOUND FORMED
THROUGH A REACTION
RED
WHITE
YELLOW
GREEN
RED
YELLOW ABSORBED
BLUE
BLUE
TRANSMITTANCE (T)
RATIO
OF THE INTENSITY OF LIGHT
LEAVING SOLUTION (I)
TO THE INTENSITY OF LIGHT
ENTERING SOLUTION (IO)
TRANSMITTANCE
IO I
I
T=
IO
%T = T x 100
Comparing Light Transmittance to Cell Length
I0 1.0
.9
.8
Transmittance
.7
.6
.5
I1
.4
.3
.2 I2
.1 I3 I5 In
I4
0
0 1 2 3 4 5
Units of Optical Path
LAMBERT’S LAW
Relates the absorption of light to the depth or
thickness of the colored liquid
.9
.8
Transmittance
.7
.6
.5
I1
.4
.3
.2 I2
.1 I3 I5 In
I4
0
0 1 2 3 4 5
Units of Concentration
BEER’S LAW
T = 10 -abc
Where:
A = -log T = abc
ABSORBANCE (A)
A = - log T
A = abc
Where:
a = constant for particular solution
LIGHT ABSORBED
by the
COLOR DEVELOPED
in a sample
CONCENTRATION CAN BE COLORIMETRICALLY
DETERMINED IF:
1. Able to chemically develop a color with that substance and only that
substance
pH
Time
Temperature
Ionic Strength
COLORIMETRY
LIGHT ABSORBED
by the
COLOR DEVELOPED
in a sample
Color Measurement
Compare Sample Color to Known
Standards
“Color Comparators”
Spectrophotometer
“Calibration Curve”
(verified)
Colorimetric Instruments
Spectrophotometer
Sample
Cell
Monochromator Detector
Light
Source
Meter
Light Source
Constant
Controllable
Diaphragm Voltage Regulation
Fatigue
Voltage Adjustment
Color (wavelength) Band
Monochromator
APERATURE
OR
SLIT
PRISM
OR
Must be CAREFULLY Adjusted
DIFFRACTION
GRATING
Sample Cell
Must be
CAREFULLY
Aligned
Cuvette
PHOTOELECTRIC TUBE
“DETECTOR”
Differing
Response
for
Various
Wavelengths
Bausch & Lomb
33-29-71
340-600 nm
33-29-72 (w / filter)
600-950 nm
33-29-92 (w / filter)
400-700 nm
PHOTOELECTRIC TUBE
“DETECTOR”
Differing
Response
for
Various
Wavelengths
Must Use the
Correct
Combination
of Filter and
Phototube
For Wavelength
Of Analysis
INDICATING METER
Lenses
Mirrors
Apertures
Occluders
Optical System
Meter
COLORIMETRY
Instrument Operation:
Warm-up
Set Monochromator
Set ∞ Absorbance
Set Zero Absorbance w/Blank
Re-adjust as Needed
COLORIMETRY
Instrument Operation:
General Rule –
Absorbance Between 0.100 and 0.700
Best Readings –
Between Lowest and Highest Standards Used
In Calibration
Watch for Irregularities
COLORIMETER CALIBRATION
Calibration or Standardized
By Measuring Absorbance Readings
of a Series of Known Standards
Comparison of These Readings to
the Reading for a Sample
1. Computer Spreadsheet
2. Instrument with Internal Microprocessor
3. “Plotting” a Graph
COLORIMETER CALIBRATION
Calibration or Standardized
By Measuring Absorbance Readings
of a Series of Known Standards
Comparison of These Readings to
the Reading for a Sample
Verified Frequently
At Least One Standard
In Acceptable Range
Each Time Samples Are Analyzed
COLORIMETER CALIBRATION
Repeat Calibration:
1. Significant Change In Procedure,
Equipment, or Reagents
Calibration 0.4
0.6 0.312
Curve
(Using Phosphorus 0.3
Absorbance
Analysis Example)
0.2
0.1
Prepared By
Michigan Department of Environmental Quality
Operator Training and Certification Unit