You are on page 1of 22

Temperature

Training Des 2007

En dr
e s +H
a
u
s
Sicestherme
r
E+H Temperature sensors
Segmentation

S
M
Dwi S.....
T
E+H Temperature sensors
Segmentation

 Omnigrad "T"
 in future (2002?)
 no thermowell
 only clamps (and threads)
 without head (with plug) or
with "mignon" head
 without transmitter or with TMT "T"

 Omnigrad "M"
 sanitary line
 others

Dwi S.....
E+H Food & Pharma new line
TST "type a" (formerly TST14)

 Process connection
 ISO clamp (DN 12, 25, 40)
 Tri-clamp (1'', 1.5'', 2'')
 welding (cylidrical, 30x40)
 welding (cylidrical/spherical 30x40)
 SMS 114 (DN 25)
 G1'' (Liquiphant)
 dairy DIN 11581(DN 25, 32, 40,
50)
 aseptic DIN 11864-1/A
(DN 25, 40)
 Varivent (DN 50, 32, 15)
 flange 40x7
 Ingold (d=25 mm, l=?)
 special (Microclamp, Neumo...)

 material: same of tube

Dwi S.....
Sale structure
TST "type a" (formerly TST14)

 Neck
 D = 9 mm, L = 82 mm
 D = 20 mm, L = 45 mm
 special

 Pipe size and immersion length


 d = 9 mm, l = 30 mm
 d = 9 mm, l = 60 mm
 d = 9 mm, l = 100 mm
 d = 9 mm, l = 150 mm
 d = 9 mm, l = 210 mm
 d = 11 mm ...
 special

Dwi S.....
E+H Food & Pharma new line
TST "type a" (formerly TST14)

 Shape of tip
 straight
 reduced (stepped)
 tapered
 special

 Material and wetted part finishing


 AISI 316L (1.4404), 1.5 um
 AISI 316L (1.4404), 0.8 um
 AISI 316L (1.4404), 0.4 um
 AISI 316L (1.4435), 0.8 um
 AISI 316L (1.4435), 0.4 um
 AISI 316L (1.4435), 0.8 um +
electropolishing
 AISI 316L (1.4435), 0.4 um +
electropolishing
 special

Dwi S.....
E+H Food & Pharma new line
TST "type a" (formerly TST14)

 Terminal
 flying leads
 ceramic terminal block
 plastic terminal block

 RTD class and configuration


 1 Pt 100 class A, 3 wires
 1 Pt 100 class A, 4 wires
 2 Pt 100 class A, 3 wires
 2 Pt 100 class A, 4 wires
 1 Pt 100 class 1/3 DIN B, 3 wires
 1 Pt 100 class 1/3 DIN B, 4 wires
 special

Operating range: -200...400°C;


class A guaranteed from -50 to 250°C;
class B in the rest of the range
Dwi S.....
E+H Food & Pharma new line
TST "type a" (formerly TST14)

 Head transmitter
 TMT180
 TMT181
 TMT182
 TMT184
 TMT181 Atex
 TMT182 Atex
 TMT184 Atex

Dwi S.....
Transmitters
The iTEMP family

iTEMP PCP iTEMP PCP iTEMP HART iTEMP PA


TMT 180 TMT 181 TMT 182 TMT 184

Input RTD/TC Pt100


RTD -- / / /
Output 4...20 mA 4...20 mA 4...20 mA
L
O
C
O
T
O
R
P
S
N
I O
T

PCP PCP
A
I C
N
U

-PA
M
M
O
C
D
L
E
F I

Operation

Accuracy (Pt 100, -100...200°C)  0.1 K  0.2 K  0.2 K  0.15 K

Atex EEx ia -
FM / CSA - / / /

TMT 180: low cost PCP only for Pt 100;


accuracy 0.1 K (!) available as option in the range -50... 200°C
Dwi S.....
E+H Food & Pharma new line
TST "type b" (formerly TST44N)

 Same characteristics of TST "type a"


 No thermowell
 New internal construction
(max temp. 250°C)

 Tube diameters: 6 mm and 8 mm


 "American style" neck in addition

Dwi S.....
E+H Food & Pharma new line
TST "type c"

 Pipe DN 50 and 40 (32 or 25?)


 Maybe initially as "special product"
 Sale structure similar to TST "type a"
 Pipe connection: threaded (DIN 11851?) and
clamped

Dwi S.....
E+H Food & Pharma new line
TA "type 1" (formerly TA14)

 Sale structure similar


to TST "type a"

Dwi S.....
Temperature head transmitter
iTEMP PCP TMT 181

Terminal layout

Dwi S.....
Start the configuration software at the PC

Dwi S.....
Fault finding

Due to its construction the head transmitter cannot be repaired.

Dwi S.....
Fault finding

Dwi S.....
Dwi S.....
Dwi S.....
TMT 136, TMT 137

Calibration procedure
1 - Connect resistance decade as shown in figure 1.
2 - Use resistance/temperature table according to IEC 751.
3 - Apply resistance corresponding to 4 mA. Adjust zero
potentiometer to 4 mA output.
4 - Apply resistance corresponding to 20 mA. Adjust span
potentiometer to 20 mA output.
5 - Repeat procedure until readings are stable. Laquer
potentiometers to seal.

Calibration of the TMT 137


TMT 137 is a fixed range transmitter. The instrument is ordered
and delivered factory calibrated.

Dwi S.....
Electrical Connections

Dwi S.....
Calibration of the TMT 136

Calibration instruments
A) Resistance decade, inaccuracy <0.025%
B) Indicator 0-20 mA, inaccuracy <0.025%
C) Power supply 24 V DC, 40 mA
D) Copper wire for connections
E) Screw driver 3 mm

Dwi S.....
Dwi S.....