Professional Documents
Culture Documents
Temperature Training Des 2007: H Auser E Ndress
Temperature Training Des 2007: H Auser E Ndress
En dr
e s +H
a
u
s
Sicestherme
r
E+H Temperature sensors
Segmentation
S
M
Dwi S.....
T
E+H Temperature sensors
Segmentation
Omnigrad "T"
in future (2002?)
no thermowell
only clamps (and threads)
without head (with plug) or
with "mignon" head
without transmitter or with TMT "T"
Omnigrad "M"
sanitary line
others
Dwi S.....
E+H Food & Pharma new line
TST "type a" (formerly TST14)
Process connection
ISO clamp (DN 12, 25, 40)
Tri-clamp (1'', 1.5'', 2'')
welding (cylidrical, 30x40)
welding (cylidrical/spherical 30x40)
SMS 114 (DN 25)
G1'' (Liquiphant)
dairy DIN 11581(DN 25, 32, 40,
50)
aseptic DIN 11864-1/A
(DN 25, 40)
Varivent (DN 50, 32, 15)
flange 40x7
Ingold (d=25 mm, l=?)
special (Microclamp, Neumo...)
Dwi S.....
Sale structure
TST "type a" (formerly TST14)
Neck
D = 9 mm, L = 82 mm
D = 20 mm, L = 45 mm
special
Dwi S.....
E+H Food & Pharma new line
TST "type a" (formerly TST14)
Shape of tip
straight
reduced (stepped)
tapered
special
Dwi S.....
E+H Food & Pharma new line
TST "type a" (formerly TST14)
Terminal
flying leads
ceramic terminal block
plastic terminal block
Head transmitter
TMT180
TMT181
TMT182
TMT184
TMT181 Atex
TMT182 Atex
TMT184 Atex
Dwi S.....
Transmitters
The iTEMP family
PCP PCP
A
I C
N
U
-PA
M
M
O
C
D
L
E
F I
Operation
Atex EEx ia -
FM / CSA - / / /
Dwi S.....
E+H Food & Pharma new line
TST "type c"
Dwi S.....
E+H Food & Pharma new line
TA "type 1" (formerly TA14)
Dwi S.....
Temperature head transmitter
iTEMP PCP TMT 181
Terminal layout
Dwi S.....
Start the configuration software at the PC
Dwi S.....
Fault finding
Dwi S.....
Fault finding
Dwi S.....
Dwi S.....
Dwi S.....
TMT 136, TMT 137
Calibration procedure
1 - Connect resistance decade as shown in figure 1.
2 - Use resistance/temperature table according to IEC 751.
3 - Apply resistance corresponding to 4 mA. Adjust zero
potentiometer to 4 mA output.
4 - Apply resistance corresponding to 20 mA. Adjust span
potentiometer to 20 mA output.
5 - Repeat procedure until readings are stable. Laquer
potentiometers to seal.
Dwi S.....
Electrical Connections
Dwi S.....
Calibration of the TMT 136
Calibration instruments
A) Resistance decade, inaccuracy <0.025%
B) Indicator 0-20 mA, inaccuracy <0.025%
C) Power supply 24 V DC, 40 mA
D) Copper wire for connections
E) Screw driver 3 mm
Dwi S.....
Dwi S.....