Professional Documents
Culture Documents
Review
•1
EMG Basics
• Electromyography (EMG)
• Evoked Potential (EP)
•2
Introduction to EMG
• What is EMG?
− Assessing electrical activity of muscles
− Normally performed with another test that measures the conducting function of
nerves - nerve conduction study (NCS)
− Both tests performed at the same office visit and by the same physician – NCS
often performed by EDx tech
− Muscular movement involves action of muscles and nerves and needs an
electrical current (this electrical current is much weaker than the one in your
household wiring)
− In some medical conditions the electrical activity of the muscles or nerves is not
normal
− Finding and describing these electrical properties in muscle or nerve helps
diagnose condition
− The instrument used is called an electromyograph
•3
Introduction to EMG
• History of EMG
− 1666 - Francesco Redi documented highly specialized muscle of the
electric ray fish generates electricity
− 1773 - Walsh demonstrates Eel fish’s muscle tissue could generate a
spark of electricity
− 1792, - Luigi Galvani documents that electricity could initiate muscle
contractions
− 1922 - Gasser and Erlanger use an oscilloscope to show electrical
signals from muscles
− 1930 – 1950 – Research continues
− 1980s - advanced techniques allow production of the required small
and lightweight instrumentation and amplifiers
•4
Introduction to EMG
• When is EMG testing performed?
− Find the cause of weakness, paralysis, or muscle twitching
− Diseases that damage muscle tissue, nerves, or the junctions between nerve
and muscle (neuromuscular junctions). These include amyotrophic lateral
sclerosis (ALS), or myasthenia gravis (MG), muscular dsytrophy, etc.
• When is NCS testing performed?
− Identify the cause of abnormal sensations, such as numbness, tingling, or pain
− Find damage to the peripheral nervous system, which includes all the nerves
that lead away from the spinal cord and the smaller nerves that branch out from
those nerves
− Includes nerve disorders, such as carpal tunnel syndrome or Guillain-Barré
syndrome, etc.
•5
Introduction to EMG
• How is EMG testing performed?
− Small pins or needles are inserted into muscles to measure electrical activity
− The needles are small and solid, not hollow like hypodermic needles
− Patient asked to contract muscle by moving a small amount
− Muscle activity is monitored on a display screen and through a speaker - Sounds
like a popping or soft roaring noise.
•6
Electromyography
Electromyography
EMG Recruitment
MUP Analysis
•7
Electromyography (EMG)
Electromyography is widely used to
Concentric needle insertion
distinguishEMG Needles
weakness due to muscle Fasciculus
Recording area Active
disease, neuropathy or other causes.
Bipolar concentric
Muscle fibers Reference
The EMG is recorded by means aof
Active
15°
special concentric or monopolar EMG
Concentric EMG needle
needle inserted in the muscle.
Reference
Motor units, in a given muscle, vary
Orientation marker
with respect to territory and fiber
density, and with respect to physiologic Peripheral
motor nerve
properties such as contraction time, Motoneuron
Muscle fibers
fatigability,Monopolar
rise time of the spike and Orientation marker
order of recruitment. Active Motor unit Neuromuscular junction
Baseline crossing
Amplitude
Satellite
Phase
Criteria
Single potential: # of phases ≤ 4 Rising time
Just a reminder
Neuromuscular
Axon junctions 2
Moderate
The Electromyography EMG needle contraction
test records action 4 MUPs
EMG needle
Myopathy
•11
EMG Tests vs Pathologies
CENTRAL
In WEAKNESS
percentage of patients
Spinal cord - Brain
(simplified model)
T/A MUP
90% 50% 70%
MYOPATHY NEUROPATHY
Muscle fibers EMG Recruitment Axon - Myelin or both
MAXIMUM
Settings EFFORT
Peak/Peak
Filters amplitude
High pass 5 – 20 Hz
Low pass 10 – 20 KHz
At rest:
EMG Needles: Normally no potentials “Flat recording”
If fibrillation potentials:
- Disposable concentric • Partial denervation
- Disposable monopolar • Myopathy
Amplitude: 100 - 200 µV
• Polymyositis
•13
Next: Turns / Amplitude
Turns / Amplitude (T/A) Analysis 1 second
Force 1
Turns / Amplitude
Force 2
Force 1 second recording Myopathy
Force 3
Pull Amplitude
Turns
Force 4
100 µV
Amplitude
segment 1
S1 S2
Neuropathy
Up to 20 measurements
Concentric or monopolar needle
Mean Turns Amplitude
Neuropathy
at different forces
Amplitude
Settings:
A turn issee EMG recruitment
a change in signal direction ( ) Turns
1second
A segment is delimited
of EMG signalby two valid turns
A turn is valid if segment amplitude is >100 µV,
Few sites in the
represented bysame muscle
This MUP has 3 turns and 2 segments (S1 & S2)
Graphical representation of normal Myopathy
values are established for the most
commonly investigated muscles. Number of Turns / s
•14
Next: Number of Small Segments
Motor Unit Potential (MUP)
The interference pattern is a sum of motor unit potentials.
The number of recruited MUPs depends on the contraction.
Each motor nerve innervates
Identification of MUPs one or more muscle fibers,
and generates simple or polyphasic compound motor action potentials
Slope (CMAP).
> 30µV /0.1 ms
From the same MUP, the maximum
firing
Example: 1 second pattern recorded with moderated
> 30µV contraction.
frequency criteria is 400 Hz (2.5 ms)
To be a MUP, the negative amplitude1 second
should be >30 µV ( >50 µV pp) and
First slope (depolarization) is >300 µV/1 ms
or 30 µV/0.1 ms > 2.5 ms
(see Neurophysiology – Peripheral Nerve)
10000
Analysis
All potentials in thisAsimple
graphical depiction can
recording of easily be identified
2 SD
and classified.
normal values is established 1000 1 SD
Amplitude (µV)
3 main criteria: for each main investigated
muscle. 100
1stAmplitude
Main MUP 2nd Main MUP Isolated MUP Isolated MUP
Duration Regular
• Green points: within the Irregular firing
Regular firing firing 10
Irregular
firing
Constant amplitude normal range
Constant amplitude 3 phases 2 phases
Number of phases
and duration • Red points: outside 2 SD
and duration 1
5 phases 2 phases
• Cross: mean of all 0 5 10 15
Duration (ms)
20
•15
MUP Recording
3 methods to collect MUPs
Triggering selection
Averaging (AVG) &
Trigger level > 2.5 ms
Analysis of selection
2 – Classification of MUPs Amp. 612 µV
Dur. 19.8 ms
(Limited to high amplitude MUP)
To be in the same MUP classification, Phases 5
potentials must have Central slope
Multi-MUP,Amplitude
automatic recognition and AVG (the fastest method)
µV/µs
Same amplitude µV
Recording method:
Same central slope
5 seconds analysis
Same
Settings: see EMG #recruitment
of phases
Moderate contraction AVG
Few sites in the same muscle
Amplitude, Duration, # of Phases, etc… and Firing frequency
•16
Next: Power Spectrum
FFT – Power Spectrum
In needle EMG, the frequency distribution
shifts slightly from high to low frequency Normal Fatigue
with increasing force. 0dB 0dB
•18
Introduction to EMG
• Nerve Conduction Studies (NCS)
•19
Nerve Stimulation
Nerve Stimulation
Electrical Current
Nerve Stimulation
A nerve is a Chain of Polarized Membrane Segments (Myelin Action).
Stimulation in one point generates a depolarization.
Myelin Axon
Peripheral nerve Muscle
Antidromic Orthodromic
Motor latency
Motor response
The time before repolarization is called the Refractory Period (see Neurophysiology – Peripheral Nerve).
Motor Latency is the propagation time from stimulus site to
During this time a nerve cannot be depolarized.
muscle contraction (motor response). Motor Latency is expressed in ms
•21
Next: Electrical Current
Electrical Constant Current
A current stimulator is a voltage generator (V) with
a predefined constant current (I).
I known
I Z Impedance
Reference unknown
set by user
V Variable
Constant Current
The reference automatically adjusts the voltage to reach the preset
A constant current offers
current the advantage
through to always
the tissues. Thekeep the sameofreference
impedance of stimulation
the tissues (Z) variesfor any patient.
from site to site and from patient to patient.
If the impedance of the tissues is too high, you will be informed by an alarm. Then you have to clean and
sometimes brush the skin.
V
Ohm Law I =
Z
•22
Cross section of the wrist
The
Current Stimulation
medianImpulse
Stimulus nerve at the wrist, as an example.
Median nerve
Ulnar nerve
The
The current has to go through the skin (incl. fat,
stimulation is performed Intensity
tendon and other tissues) before QE is
mA the nerve
with a quantity of electricity.
actually stimulated.
Duration µs
+ -
And, the surface impedance of the skin, between Skin, Tissues
Fat, Tendon
10 KΩ
•23
Rheobase and Chronaxy
Rheobase is the property whereby a current value
Nerve potential
mV
action potential
must reach a minimum value before depolarization ms
will occur, no matter how long the current may flow.
Depolarization
threshold level
•24
Motor Response
Stimulation
Let us consider a Electrodes
stimulation of the median nerve at the wrist.
Hand-Held Digital Ring
Click to start motion
Surface
Felt tips 2 3
Surface electrodes are used to stimulate
Velcroa nerve
1 “Blue dot”
Springs 4
from theActive
surface
handel of the skin (see Acc. blue dot).
(Sensory)
DIN connector 1.5 mm TPC
Current stimulator mA
with surface electrodes
Nerve Stimulation and Stimulus strength
Motor Response
mV
+ -Threshold
+ Minimal
-
The motor threshold corresponds to Skin, Tissues
the minimal intensity to reach a Fat, Tendon
beginning of response. Median nerve
•25
Introduction to EMG
• How is NCS testing performed?
− Motor NCS performed by electrical stimulation of a peripheral nerve and
recording from a muscle supplied by this nerve.
− The time it takes for the electrical impulse to travel from the stimulation to the
recording site is measured. This value is called the latency and is measured in
milliseconds (ms).
− The size of the response - called the amplitude - is also measured. Motor
amplitudes are measured in millivolts (mV).
− By stimulating in two or more different locations along the same nerve, the
velocity (NCV) across different segments can be determined.
•26
Introduction to EMG
• How is NCS testing performed?
− Sensory NCS are performed by electrical stimulation
of a peripheral nerve and recording from a purely-
sensory portion of the nerve, such as on a finger.
− Like the motor studies, sensory latencies are also
measured in ms. Sensory amplitudes are measures
in microvolts (ųV). The sensory NCV is calculated
based upon the latency and the distance between
the stimulating and recording electrode.
•27
Introduction to EMG
• How is NCS Testing performed?
− F-wave study uses stimulation of a motor nerve and recording of action
potentials from a muscle supplied by the nerve
− This is not a reflex, potential travels from the site of the stimulating electrode in
the limb to the spinal cord and back to the limb in the same nerve that was
stimulated
− The F-wave study evaluates conduction velocity of nerves between the limb and
spine, whereas the motor and sensory nerve conduction studies evaluate
conduction in the limb itself
•28
Introduction to EMG
• How is NCS Testing performed?
− Repetitive Stimulation study uses stimulation of a motor nerve and recording of
action potentials from a muscle supplied by the nerve
− Often called “Decrement” study. Used to evaluate changes in muscle response to
series of stimulations due to neuromuscular junction disease – Myasthenia Gravis,
Eaton-Lambert
•29
Introduction to EMG
• How is NCS Testing performed?
− H-reflex study uses stimulation of a nerve and recording the reflex electrical
discharge from a muscle in the limb. This also evaluates conduction between the
limb and the spinal cord, but in this case, the afferent impulses (those going
towards the spinal cord) are in sensory nerves while the efferent impulses
(those coming from the spinal cord) are in motor nerves.
•30
Nerve Conductions
Nerve Conduction Reflexes
Motor Nerve
Sensory Nerve
Carpal Tunnel
F-Waves
Exit
Nerve Conduction
A nerve is a chain of polarized membrane segments (myelin action).
Reminder
Stimulation at one point generates a depolarization.
Myelin Axon
Peripheral nerve Muscle
Antidromic Orthodromic
Motor latency
Once
The successive
a nerve isdepolarization
depolarized ofateach
somesegment
point, ais wave
followed
of by a repolarization.
depolarization
passes inbefore
The time both directions from
repolarization that point.
is called the Refractory Period (see Neurophysiology – Peripheral Nerve).
During this time
Antidromic (to aspinal
nerve cord)
cannotand
be depolarized.
Orthodromic (to muscle).
Temperature Effect
Motor Latency is the propagation time from stimulus site to
Cool temperature slows nerve conduction and interferes with the size of responses.
Not onlycontraction
muscle that, they can affect response).
(motor nerves in anMotor
asymmetrical
Latencyfashion, sometimes
is expressed in mimicking
ms individual
nerve injuries.
•32
Nerve Conduction Studies
Nerve Conduction Studies (NCS) are an essential part of an EMG examination.
Nerve Conduction Study
Distance
Recording responses
The findings reflect the functional state of:
Neuromuscular junction
• the myelinated motor nerves, Stim. 1
• the neuromuscular
Conduction velocity transmission,
• thethe
of muscle fibers.
median nerve Stim. 2
Muscle fibers
EMG equipment
•33
Motor Response
Four parameters to characterize a motor response: Latency, Amplitude, Duration and Area
Negative peak
Be coherent in measurements !
Negative Negative
Latency is measured at the onset. Measurement at peak is sometimes necessary in
deflection
presence of large stimulus artifact, generally due to a wrong technique.
Onset Total
Amplitude, the negative deflection is the most important, some schools use peakPeak to peak
to peak
value – then careful with reference values.
Duration of negative deflection is measured in coherence with the negative deflection of
the amplitude, if not, the total duration is measured.
Area is measured in coherence with duration (negative deflection or total).
Negative
Total
Stimulation
artifact Duration
•34
Pathologic Responses
Pathology Latency Amplitude Duration
Just a reminder
Normal Normal Normal Normal
Axon MyelinConduction
Schwann cell Node of Ranvier
block
Severe
AXON and MYELIN are main demyelination
players in the transfer of NERVE IMPULSES
•35
Stimulation & Recording Rules
By convention,
Black lead is the cathode of the stimulator and the active electrode for recording.
Red lead is the anode of the stimulator and the reference electrode for recording.
Organization: “Black to Black”, stimulation polarity: Negative
•36
Stimulation Points Extremities
Nerve
Site
Plexus Sciatic
Axillaries
Stimulation point 3 Stimulation point 2 Stimulation point 1 Recording responses
Elbow Wrist Palm
Median nerve Radial
Ulnar
Tibial Stim. 1
Above Elbow
Median Popliteal Fossa
Ulnar Elbow Stim. 2
Below Elbow Peroneal
Peroneal Stim. 3
Ulnar
Wrist
APB muscle EMG equipment
Median
Wrist
Sural
Stimulation points are located on the- most
Median Ulnar accessible pathway of each nerve,
above and below joints. Palm
Posterior
Tibial
•37
Stimulation Points Head & Trunk
Erb’s Point
Higher
F Thora-
a Medium codorsalis
c
i
Lower
a
l
Long Thoracic
Trunk
Auricular Posterior
Crural
Spinal XI
Phrenic
Erb’s Point
•38
Next: Motor Nerve Conduction
Motor Nerve Conduction
Technique
With a normal patient and well performed technique,
Median nerve
these two responses would be superimposable.
Recorder
Stimulator
Technique
Direct response Distance measurement
Digit IV Lat. 2.3 ms
Amplitude 20 µV to 100 µV between Rec. and Stim.
Normally biphasic Wrist- Dist.130 mm
Duration 1 to 3 ms. Wrist Ulnar CV 57 m/s
Latency depends on stimulation site
Stimulation current depends on nerve and site,
between
Ulnar 5 - 10 mA, Median
with 200 µs stim. duration.
Digit IV Lat. 2.2 ms
Note Wrist- Dist.130 mm
Note Median CV 59 m/s
A too high current stimulation will stimulate motor
Response
fibers to a
which add stim.
motorat Elbow cantobethe
response recorded.
sensory
Latencies will be little be higher with digital
response.
rings – Be careful with normal values.
Peroneal nerve
Recorder
Motor Conduction
Stimulation 2
14 cm
Sural nerve
Sensory Conduction
Recorder
Stimulation
•42
Some Normal Values
The most sensitive normal value for a particular NCS is usually obtained by performing the
same NCS on the corresponding nerve in the contralateral limb (assuming it is normal).
A decrease of more than 50% in amplitude is considered abnormal.
The results of NCS - Motor, Sensory must be compared to some normal values.
These must be age-related as well :
Infants and young children: Quite slow conduction rate.
Adults: Amplitudes and conduction rate decrease with age.
•43
Next: Carpal Tunnel Syndrome
Carpal Tunnel Syndrome
Carpal Tunnel Syndrome (CTS) is a condition, in which the
median nerve is compressed within the carpal tunnel, where it Median nerve is
Carpal
passes underTunnel Syndrome
the trans-carpal ligament (flexor retinaculum). compressed at the
Symptoms are numbness in the first, second, third and the Median nerve
wrist, resulting in
radial aspect of the fourth digit, andpopulation
sometimes pain in the hand numbness or pain.
Approximately 0.2% of the
and arm.
undergo surgery every year.
The condition is well-known, and has been diagnosed since the
1960ies by neuro-physiological methods (Thomas & al., 1967).
Transverse carpal
Old method ligament released
Patients •with
with various devices to immobilize the wrist.
CTS constitute the largest single patient group investigated
• with the injection of steroids into the carpal tunnel
by neurophysiologists (up to 70% in some countries).
Therefore many labs try to shorten the time to diagnose this condition by various
methods.
•45
Kimura Jun MD. - Kyoto University-Japan
Reflexes
Reflex studies are performed to detect a
H - Reflex
lesion involving the efferent arc of reflex.
(see Neurophysiology – Somatic System)
Blink Reflex
Spinal cord
Sensory nerve
Tendon
InterneuronReflex (T-Waves)
Stimulation
Motion M-HLL = Leg Length, A = Age, PH = Patient Height in cm
Cathode Stimulation level higher the Motor Threshold.
Technique M response appears. H amplitude decreased because the partial
Popliteal Fossa Other Methods
collision takes effect between efferent M and sensory reflex.
1. Find
Stimulator the stimulation point with M response.
Motion M diagnose an axonal degeneration (dying-back
2. Start from current level 0 mA. Amplitude ratio M/H (for Soleus) arguments to
3. Increase current by steps of 0.02 mA until Stimulation level increased to maximal M response.
the H “sensory threshold” and continue until phenomenon).
Motion Total collision – H response disappears.
H
Motion
H Max.
increase current to Latencies: M M
4. After H-Max you can Recorder
“H index” formula (
≈ 10 ms H ≈ 30 ms (depends
Max
(Guilhéneuc-France)
PH
) 2
x 2 collision
HLat - MLaton patient
Total height)
M-H 0.5-1 mA until M Max response.
For practical purposes, H-Responses are of major value only in
Motion detecting generalizedAdult ≈ 100
polyneuropathies R/L Diff.
and <7
lumbosacral radi-
M culopathies involving the S1 roots.
Record
MLat HLatbutton
Click on “Record”
•47
Blink Reflex Blink responses are performed to assess the 5th and the 7th cranial
Stim. left
nerves. They are abnormal, for example, with unilateral facial paralysis,
such as Bell’s palsy.
Settings (eachFacial
channel)
Motor nerve
Normal Values
Indirect Reflex (Contra)
Indirect Reflex (Ipsi)
Ch1
7th µV/Div.
Sensitivity : 100 Cranial nerve Stimulator Latencies
Ipsi
Stimulation
High Frequency Filter : 2-5 KHz R1 10.6 ms
5th Cranial nerve
±2.5
TrigeminalAbnormal >13.0 ms
Low Frequency Filter
Supraorbital nerve: 20 Hz
R1 11.5
R2 31.1 ms
Sweep Sensory
Speed :nerve
10 ms/Div. R /msL diff. < 1.2 ms Supraorbital nerve
R2 Brain
Ch2 ipsi Stem 31 ms ±10 Abnormal >40 ms
Current stimulation Contra
R2 contra 32 ms Direct±11
Resp. (Ipsi)
Abnormal >41 msmotor nerve
Facial
Stim. Right • Duration 0.1
Stim. ms
Left 7th Cranial nerve
Stim. Frequency : Manual R2i / R2c R2
diff.31.7
< ms
5 ms
Orbicularis Oculi
Amplitude
Blink responses are action potentials (AP) muscle
recorded from theR1orbicularis oculi
Stim. right
0.4 muscle.
mV ±0.2
Technique Rec. Ch1
These facial muscle
R2 AP result 0.55
from mV
stimulation
±0.2 of the supraorbital nerve,
Repeat stimulation at least 3 times per site. one of the sensory nerves derived from the 5 th cranial.
Blink responses are reflex studies.
Ground Ch1
Stimulus alternative
Orbicularis Oculi muscle Rec.
ContraElectrodes alternativeIpsilateral side
Contalateral side
A 0.1 ms duration is painless, but for some patients Instead of surface electrodes, concentric needles
it is difficult to obtain responses. offer the advantage of sharpR2recordings.
32.6 ms
•48
Next: Tendon Reflex
Tendon Reflex (T-Wave) Recorder
The T-Wave is a compound action potential,
evoked from a muscle by rapid stretch of its
tendon.
Latency measurement
Motion
•49
Next: MUNE
Introduction to EP
• What is EP?
− Tests that measure electrical activity in certain areas of the brain and spinal cord
− Electrical activity is produced by stimulation of specific sensory nerve pathways
− Types of EP’s
• Brainstem Auditory Evoked Potential (BAEP) - Checks the pathway from the ear to the
brain. The BAEP test may help uncover the cause of hearing and balance problems, and
other symptoms.
• Visual Evoked Potential (VEP) - Checks the pathway from the eyes to the brain. May
help find the cause of certain vision problems and other conditions.
• Somatosensory Evoked Potential (SEP or SSEP) - Checks the pathway from the nerves in
the limbs to the brain. It is a way to study the function of the nerves, the spinal cord
and brain.
•50
Introduction to EP
• BAEP – Why and How?
− Used to diagnose hearing losses, assess high freq hearing, determine brain
death, monitor brainstem function during surgery
− Can distinguish damage to the acoustic nerve pathways within the brainstem
− Headphones used to deliver series of clicks to one ear at a time. A masking or
static sound is played into the other ear. Each ear is usually tested twice, and the
entire procedure takes approximately 30-45 minutes
•51
Introduction to EP
• VEP – Why and How?
− Used to evaluate optic neuritis,
optic tumors, retinal disorders,
and demyelenating diseases such
as multiple sclerosis
− The patient sits before a screen on
which an alternating checkerboard
pattern is displayed
− Each eye tested separately
− Each eye is usually tested twice, and the entire
procedure takes approximately 30-45 minutes.
•52
Introduction to EP
• SSEP – Why and How?
− Evaluate the nerve pathway from the arms and legs through
the spinal cord to the brain
− Identify spinal cord injuries/diseases, neuromuscular and
demyelineating diseases. Monitor patients during surgery
on the spine
− Electrodes are attached to the wrist, the back of the knee
− Mild electrical stimulus is applied through the electrodes
− Electrodes on the scalp determine the time it takes current
to travel along the nerve to the brain
− Procedure takes approximately 30 minutes
•53
•54
Evoked Potentials (EP)
General Information
Dawson’s averager, first
demonstrated in 1951,
Somatosensory EP formed a new area in
clinical neurophysiology, -
the area of Evoked
Auditory EP
Potentials.
Links between Evoked
Visual EP
Potentials and clinical
practice are firmly
Motor EP established in neurology,
neurosurgery,
Event Related Potentials ophthalmology, otology,
pediatrics, psychiatry and
urology.
Pitfalls
General Information : Averager (AVG)
Direct signal
Averaged signal
Amplified
•56
General Information : Recording &
Form and latency depend on the type of the EP and
the recording site.
Latencies SEP Tibial nerve
Recording sites are identified by
anatomic abbreviations Cpz-FPz
• Positive P..
Oz-Cz
Normal value
N75 Negative at around 75 ms
P100 Positive at around 100 ms P100
•57
20%
Cz Fz
20%
20%
Cz
20%
C3 C4
20% Pz 20%
Pz Fpz 10%
20%
Nasion T3 Oz T4
Oz
A : Auricular 10% 10%
C : Central 10%
A2 Pg2 A1 A2
Inion
F : Frontal
Fp : Frontal-polar
O : Occipital
P : Parietal
T : Temporal 10%
Nasion Nasion
Pg1 Pg2 20%
-z : Midline 25%
Odd no. : Left side Fp1 Fp2 Fp1 Fp2
Even no.: Right side 20%
F7 F3 Fz F4 F8 F7 Fz F8
25%
25% P3 Pz P4 Pz
T5 T6 T5 T6 20%
O1 O2 O1 O2
25%
Inion Inion
20%
10%
•58
Next: Somatosensory EP
Somatosensory Evoked Potentials (SEPs)
Thalamo-cortical potentials Somatosensory EPs offer a look at
“physiologic anatomy”. They provide a
Cervico-medullary potentials
sensitive tool for assessment of spinal
Upper Brachial plexus potentials cord and brainstem posterior columns,
extremities
as well as medial lemniscal tracts and
nearby structures.
SEPs are performed on both sides.
Cauda equina
Lower cord potentials
•59
Recording SEP – Lower Extremities
Ref. Ch1 & Ch2
Left Tibial nerve
L FPz R
AVERAGING
Act. Ch3
Settings Fz
Accessories Ch1
Cortex Other Scalp needlesCPi-Fpz
5 cm
C5S Cz
Filters Ref. Ch3 Disposable
2 cm (iron)
High pass 0.5 Hz 20 Hz Ø 0.3 mm length 10 mm
CPi CPz CPc Reusable (platinum) P37
Low passGround 1 KHz 3 KHz Ch2
Act. Ch2
Sensitivity 5 µV/Div. 5 µV/Div. Scalp surfaceCPz-Fpz
Sweep Speed Act. Ch4 10 ms/Div. Act. Ch1
length 100 cm
Nuprep
TPC 1.5mm P37
T12S AVG sensitivity 0.5 µV/Div. 72cm
µV/Div.
7 cm
N34
No of Epochs 200-500 - Ten20 for fixing and contact
Ch3
L4 Stim. Duration Placement
0.1-0.2 ms of electrodes
- NuPrep for skin preparation
Fpz-C5S
IC Current Level 1.5 x Check
Motor threshold
impedancesDisposable surface P31
Stim. Frequency 3 Hz
Ref. Ch4 (ICi + ICc) Length 50 cm
Blue dot LP Area 9 x 6 mm
Ch4TPC 1.5mm LP: Lombar Potential
Electrodes T12S-IC
Recording Stimulation Stimulation surface
Cortex EP needles Frequency : 3 Hz
or EP surface
Level : 1.5 time motor threshold
- Handheld electrode
20 40 60 ms
Other Disposable surface
- Elastic strap for fixing
Stimulating Averaging : 200-500 epochs Normal Latency ms
--i ipsilateral Surface Values Mean SD Range
--c contra lateral LP 10.8 ±0.9 8.6-13.1
N34 33.5 ±1.5 30.3-41.3
Stimulation P37 36.3 ±2.4 30.5-41.7
•60
Recording SEP – Upper Extremities Left Median nerve
L FPz R AVERAGING
N19
Settings Fz Accessories Ch1
C5S Act. Ch3 5 cm CPc-Cpi
Cortex Other
Cz Scalp needles
Epc Epi 2 cm
Filters Disposable (iron)
High pass
Act. Ch4 0.5 HzCPi 20
CPzHzCPc Ø 0.3 mm length 10 mm
Low pass 1 KHz 3 KHz Reusable
Act. Ch1 (platinum) N18
Ch2
Ref. Ch2, Ch3 & Ch4
Sensitivity 5 µV/Div. 5 µV/Div. Scalp surface CPi-Epc
Ref. Ch1
Sweep Speed 5 ms/Div. length 100 cm
Nuprep
7 cm 7 cm TPC 1.5mm P14
AVG sensitivity 0.5 µV/Div. 2 µV/Div.
No of Epochs 200-500 -Act.
Ten20
Ch2 for fixing and contact N13
Ch3
Stim. Duration 0.1-0.2 ms of electrodes - NuPrep for skin preparation
Placement C5S-Epc
Current Level 1.5 x Motor threshold
Disposable surface
Stim. Frequency Check
3 Hz impedances
EP
Length 50 cm
Stimulation Blue dot Area 9 x 6 mm
Ch41.5mm
TPC
Electrodes Epi-Epc
Stimulation Stimulation surface
Recording
Frequency : 3 Hz
Cortex EP needles or EP surface
Level : 1.5 time motor threshold
- Handheld electrode
Other Disposable surface 10 20 30 ms
Averaging : 200-500- Elastic
epochsstrap for fixing
Normal Latency ms Amplitude µV
Stimulating
--i ipsilateral Values Mean SD Mean SD
--c contra lateral Surface
EP 9.6 ±0.7 5.4 ±2.5
N13 13.2 ±0.8 2.9 ±1.3
N19 18.9 ±1.0 2.8 ±1.6
•61
Next: Auditory EP
Brain Stem Auditory Evoked Responses
BAERs evaluate the function of several levels
(BAERs) of the peripheral and brain stem auditory
Primary auditory cortex
system, including auditory nerves, the pons,
and mid-brain.
VII
V
IV
VI
III
II
Mid-Brain I VI
V VII
Brain Stem
IV
Pons
III I Distal spiral ganglia and auditory nerve
Ear
II II Proximal auditory nerve (cochlea nucleus)
I III Ventral cochlear nucleus in pons
Auditory Nerves IV Inferior lateral lemiscus in pons
V Lateral lemiscus in lower mid-brain
Auditory Nervous System VI Medial geniculate nucleus (upper mid-brain)
(see Neurophysiology – Brain Functions)
VII Primary auditory cortex
•62
Hearing Sensitivity
Absolute silence 0 dB peSPL Pressure Pascal
Wegel’s Curves (1922)
Decibel dB SPL
10 5 194
(peak-equivalent Sound Pressure Level) Atmospheric Pressure
equal 20.10-6 Pa. 104 174
103 154
Hearing Level depends on sound B
102 l 134
frequency (A) – This is the eve
sL
rou
normal hearing level (nHL). 10 nge 114
Da
1 94
•63
Next: BAER testing
BAER Testing Cz (dB peSPL)
Stimulating left
V AVERAGING
Nuprep
Act. Ch1 & Ch2
TPC 1.5mm
AVG sens. 0.1-0.2 µV/Div.
No of Epochs 1000-4000 - Ten20 for fixing and contact
Stimulus unfiltered click - NuPrep for skin preparation
Ch2
Duration
Two runs must Normalon
be performed
Infant
100each
µs ear.
50 µs
(see Nerve & Brain Stimulation – Pulsed Sound)
A white noise is applied on the contralateral
Polarity Rarefaction
ear (40-80 (condensation
dB), to mask oftenthe effect
enlarges the of bone
wave I)
Normal Values 5 10 ms
conduction (contralateral to ipsilateral).
Electrodes placement (Click 90 dB SPL) Latencies Amplitudes Inter Peaks
In some difficult cases,impedances.
electrodes are Peaks ms SD µV SD ms
1. Place electrodes and check
placed on mastoid (M1 and M2) instead Shielded I Standard
1.7 0.15
Tubal 0.28 0.14
Bone conductor I-III 2.2
2. Headset in place (Red/Right).
of A1 and A2. Look for the headset II headset
2.8 insert
0.17 0.23 0.12 III-V 1.8
patient hearing level (normally around 35 dB). III 3.9 0.19
Red: right ear Blue: left ear0.25 0.12
IV 5.1 0.24 0.40 (Bone0.13
vibrator)
Also R-L
3. Add 50-60 dB to the hearing level value. V 5.7 0.25 0.47 0.16 difference
Wait min. 30 s. before the recording is started. VI 7.3 0.29 0.43 0.16
•64
Next: OHL testing
OHL Testing
Objective Hearing
Latency ms
(dB OHL)
Level (OHL) Normative data (60 dB OHL) 80 dB
11
Latency/Intensity Function (LIF)
(see Nerve 5
2 & Brain Stimulation – Pulsed Sound) 0 5 10 ms dB OHL 20 40
Wave I 60 80 100
•66
VEP Testing Cz Distance to patient and check size:
see Nerve & Brain Stimulation
Flashed Light – Visual Field Angle
1. Place electrodes and check impedances.
2. Stimulation rate 1 Hz – Maxi. 2 Hz.
Fpz
Settings Accessories
O’1 O’2 Stimulation right eye AVERAGING
2 cm Ch1, 2, 3 & 4
Scalp needles
Filters
O1 Oz O2
High pass 0.5 Hz Disposable
Ch1(iron) Run 1
A1 Ref. Ch1, Ch2 & Ch3 Ø 0.3 mm length 10 mm
Low pass A2 1 KHz O1-Cz
Reusable (platinum) Run 2
Act. Ch4
Sensitivity 5 µV/Div. Act. Ch3
N145
Sweep Speed 30 ms/Div. Scalp surface N75
Act. Ch2
Ch2 length 100 cm
AVGneedles,
sens. 1-2 µV/Div.
Nuprep
With scalp Act. Ch1 Oz-Cz
TPC 1.5mm
Noless
O’ line is of painful
Epochs 200-500 Ground
Ref. Ch4 - Ten20 for fixing and contact
Pattern Stimulation P100
The VEP isStimulus:
performed with checkerboard
monocular stimulation - NuPrep for
Ch3 skin preparation
reversing pattern
O2-Cz
and recording
Size with
of CRTa minimum
Monitor of
17”four channels.
(Diag. 43 cm)
Number
Two runs must beofperformed
Checks 48 onxeach
64 eye. Stimulation:
StimulationDistance to patient
with a reversing 70 cm
checkerboard pattern.(see Nerve & Brain Stimulation – Flashed Light)
Stim. Frequency 1 Hz Ch4
Use flash or LED goggles for patients who cannot Cz-A1 N100
focus on screen
Screen(see Nerve & Brain
W angle 26.3°Stimulation).
The distance between
Screen H anglethe subject
20.3° and screen, and
the check size influence
Checks angle visual27’responses. 100 200 ms
Normal value: Full-field, screen size 9°, check size 26’
Smaller checks stimulate central vision more Checkerboard LED
Flash
Ambient luminosity Dark room pattern Latency
goggles ms Amplitude µV
selectively than do larger checks.
Mean SD Range Mean SD Range
Luminosity directly influences amplitude and P100 102.3 ±5.1 89-114 10.1 ±4.2 3-21
latency of VEP, and thus is kept constant. R/L diff 1.3 ±2.0 0-6 1.6 ±1.4 0-5.5
•67
Next: Motor EP
Motor Evoked Potentials (MEPs) In clinical medicine, single-pulse Magnetic
stimulator
transcranial magnetic stimulation (TMS) is
primarily used to evoke motor responses in
slightly activated target muscles. Coil
CNS
Efferent
motor pathway
Motor EPs are used to test the
Motor nerve efferent motor pathway:
PNS (nerves and roots)
Root
Motor tracts in the CNS
PNS
Reminder:
Motor Nervous System Left side extremities are controlled by
(see Neurophysiology – Nervous System)
PNS: Peripheral Nervous System
the right side cortex and vice versa.
CNS: Central Nervous System
•68
Cortical Magnetic Stimulation Stimulation of a specific area
Coils and Coil orientation Power and Motor response
of motor cortex
Knee
Trunk
Hip
Shoulder
Elbow
Upper limb: over the left
Wr
Ha
ist
5
40% hemisphere for recording on
nd
Cortical stimulation 3
4 F
Ankle
2
the right limb.
in
Th
ge
Ne umb Toes
rs
c
Ey k ( b
eb a
50%
Eye row ck) Lower limb: over the vertex
s
Fac
e
Motor cortex
1/2 Lips Motor cortex
Here, tibial area…
45° Jaw 60%
organization
Tongue
Swallowing
70%
•69
MEP Considerations Amplitude consideration
Amplitude Large variations ! Latency Strongly influenced ! Area Generally enlarged !
The amplitude ratio MEP / Distal response is frequently calculated.
A ratio comprised between 20 and 30% is considered as normal!
Latency consideration
Muscles Latencies ms ± SD
•71
Next: ERPs
Event-Related Potentials (ERPs)
Right Brain ERPs
The right side is mainly attributed to learn allow
actions, to qualify
spatial projections and the
musicrelationship
art.
between somatosensory, auditory (P80),
Somatosensory cortex visual, frontal cortex (N100 and P200) and
Parietal lobe
Frontal lobe parietal cortex (cognitive potentials – P300).
Auditory cortex
ERPs (Event-Related
Potentials) are performed with
Visual cortex
various stimuli like pictures,
sounds and/or phrases.
•72
Active P300
Active P300 is one way to diagnose alterations P80: Middle latency
of cognitive functions. N1, P2: Long latency responses
Clinical case P3: P300 Cognitive response
Cz Ground N1
Myotonic
FPzDystrophy
BAEP
Ref. Ch1 N2
Cz-A12 Action
N1 Cz
Action
A1 A2
Left Right P80
N2 M1 M2
Cz-A12 Act. Ch1 P2
Reaction time
P2
L P3 R 5 µV
P3
-100 0 500 ms
Binaural stimulation - Rarefaction – 70 dB SPL
Stim. Rate -100
0.5 Hz0 500 ms
A1 Cz A2
Left Right
M1 M2
N1
Act. Ch1
Oddball
Cz-A12 N2
L R
P2
Monoral stimulation – Rarefaction – 80 dB SPL 3 µV
Rate 0.5 Hz P3
Standard: 1000 Hz Burst (10/100/10 ms), occurrence 80%
Randomized oddball: 2000 Hz Burst (10/100/10 ms), -100 0 250 500 750 1000 ms
occurrence 20%
2 channel recording – alternated Standard/Oddball
Record P3: P300 cognitive response
Settings (2 channels) N1 and P3 presence can evaluate the
Filters
degree of chance for re-awakening of
High pass 0.05 - 0.1 Hz
Low pass 0.1 - 0.05 KHz the patient.
Sensitivity 5 µV/Div. AVG sensitivity 1 - 2 µV/Div.
Sweep Speed 100 ms/Div. No of Epochs 30 – 50
•74
Contingent Negative Variation (CNV)
CNV is aPsychiatric
large negative change
clinical cases
type of event-related potential, believed to indicate attention, consisting of a
in potential
Normal young voltage across the cerebral cortex, especially in the
subject (30-50)
µV
frontal lobe,
- 20 that develops slowly in a person who is actively expecting the occurrence
of some- significant
10 stimulus requiring a response.
0
Action
+ 10
µV
- 20
+ 20 Cz Ground
FPz
0 2000 4000 ms - 10
Ref. Ch1
µV
Non-demented old subject
Cz 0
Left- 20 Right (70-80) A1 A2
M1 M2
- 10 + 10
Act. Ch1 S1 S2
0
+ 10 L R + 20
0 1000 2000 ms
+ 20
0 2000 4000 ms
Binaural stimulation - Rarefaction – 80 dB SPL Settings
Stim. Oldstimulus
subject with primary Filters
RandomizedµVpaired (manually 5 to 60 seconds)
- 20 degenerative senile dementia
1000 Hz click (S1) followed 1” later by long 500 Hz burst (S2). High pass 0.05 - 0.1 Hz
- 10 Low pass 0.1 KHz
Technique:0 Sensitivity 5 µV/Div.
Patient+ 10has to perform an action (handheld switch) Sweep Speed 200 ms/Div.
while the burst stimulus appears. AVG sensitivity 2 - 5 µV/Div.
+ 20
Averaging on 30 epochs are normally sufficient. No of Epochs 30 – 50
0 2000 4000 ms
•75
Motor Related Potential (MRP)
MRP is a variance of CNV test. “Hit”
Ref. Ch1 RP
10 µV
A1 Cz A2 Cz-A12
Left Right P200
M1 M2
Act. Ch1
L R SPP
100 µV
EMG
Oddball
Binaural stimulation - Rarefaction – 70 dB SPL
Stim.
Randomized Rate 3 – 10” manually
-1000 -500 0 500 1000 ms
Randomized paired standard/oddball stimulus (0.8-1”)
Standard: 1000 Hz Burst (10/100/10 ms), occurrence 80% Settings Cortex EMG
Oddball: 2000 Hz Burst (10/100/10 ms), occurrence 20% Filters Integrated
High pass 0.05 - 0.1 Hz 20 Hz
Technique: Low pass 0.1 KHz 2 KHz
Trigger recording is only performed on “Hit” stimuli. Sensitivity 5 µV/Div. 100 µV/Div.
Patient “hit” an handheld switch when the oddball Sweep Speed 200 ms/Div. 200 ms/Div.
stimulus appears.
Averaging on 30 epochs are normally sufficient. AVG sensitivity 2 - 5 µV/Div.
No of Epochs 30 – 50
•76
Pitfalls
Stimulation frequency too high (SEP, AEP & VEP)
Stimulus intensity too high (SEP, AEP & MEP)
Electrode impedance too high
Patient not relaxed (muscle activity)
Eyes movement
Too much light in the room (VEP)
Too much noise in the room (AEP)
Confusion between dBSPL & dBOHL (AEP)
Interference with other instruments (artifact)
See also “Techniques”
•77
Introduction to EMG
• Thank you!
• Questions?
•78