Professional Documents
Culture Documents
DOSING PUMP
BACKGROUND
P115 A/B are positive displacement rotary type helical rotor pumps also called Progressive
cavity pumps. The pumps were supplied by “prorites equipment pvt. ltd. and installed at offsite-
2[ Offsite-2 is the water treatment facility used for canal water treatment.] facility to pump Alum
solution (coagulant) into the flash mixing chamber where it is mixed with water in specific
quantity (in ppm) to carryout coagulation and flocculation.
Pump performance:
The P115 A/B is motor driven with the capacity and speed ranges from 400-1950 l/h and 125-
635 rpm, respectively. The suction side of the pump is flooded (filled with pumping fluid) which
avoid any type of cavitation and design pressure of pump is 5 bar. The pump has the power
requirement of 0.4 MW and its duty cycle is 12 hours each day.
Pump Metallurgy:
P115 A/B belongs to the k-range series of pc pumps manufactured by Prorites. Pump
casing is cast iron GG20, rotor is made of duplex steel and stator is made of nitrile. The
base plate of pump mild steel base plate.
Pump Driver:
P115 A/B is motor driven pump. The motor is provided by the pump “SIEMENS” and has
the power requirement of 0.5 KW. Gear based speed reduction mechanism is used in the
pump with speed reduction of 1:5. The drive service factor is 1.3.
PROGRESSIVE CAVITY PUMP
The progressive cavity pump is a positive displacement rotary pump. The pump has an
internal threaded rotor with double thread and externally threaded rotor with single
.
threads. The meshing of two forms a series of cavities 180 degrees apart, which progress
along the axis of the rotor as it rotates. As one cavity increase in volume the opposing
cavity decrease in volume at the same speed. Therefore, the sum of the discharge at the
constant volume. The result is a continuous flow from the pump.
The P115 A/B flow rate data is shown in the table below :
FAULT TREE ANALYSIS
There can be several reasons that can cause low or no flow rate problems in the progressive
cavity pumps we will do our FTA analysis and see which reasons are more plausible than
others. Now the pump manufacture data shows that the flow problem can occur from the
following reasons shown below:
1)Wrong rotor rotation direction
2)Pump and system characteristics does not match
3)Suction Lift too high
4)Leaked inlet or Suction line
5)Back Pressure too high
6)Rotor or Stator Wear
7)Clogged inlet, Suction Line or Cavity
Now there is no reason to believe that these are all the problems that can happen with the
pump but the most common one has been listed above. Now we will see these problems one by
one.
Wrong rotor rotation direction
• Wrong rotor rotation can be the reason for no flow rate. The reason for wrong rotor direction
can be related to electrical circuit of the pump. But the problem is that flow problem is
repetitive and the electrical problem cannot be reason for failure of the pump again and
again. So, we can rule this reason safely without any doubt.
System and pump are incompatible
• It is possible that the pump speed and flow doesn’t match with system characteristic and the
operating point of the pump is not at the optimum location but from the manufacture data
available as shown above in table 1 and 3 the system and pump are quite compatible and
there is no reason to believe that pump is failing because of this reason.
High back pressure
• High back or discharge pressure can also cause slip in the pump and result in the backward
flow of the fluid. Now there is not enough data to safely rule out this possibility but later
evidence show that the root cause of the problem is not high discharge pressure as we will
see as we further go on.
• The progressive cavity pumps do not create pressure across its ends It only transfer the fluid
from one end to the other and that is why they have low or no lift at all. Most of progressive
cavity pumps has a suction side that is always filled with fluid to keep the cavities with liquid.
Also, the system for alum dosing is the head system not a lift system .as you can see from
figure the suction side in the head system is flooded so the problem of high suction lift cannot
occur. So, we can also rule out this possibility too.
Leakage at inlet section
• Leakage at the inlet is can also be the reason for pump failure as high suction pressure can
cause leakage through unions or joints but the pressure at the suction is not very high is not
high enough to cause such failure at least no repetitively.
i) High Temperature
High temperature can cause the swelling of the stator as it is an elastomeric material which
results in decrease in internal diameter of the stator and when the pump is running it will result
in high wear of stator. But the operating temperature of the pump is well within materials
working range.
It has been my good fortune to get an opportunity to undergo training at a Prorites Equipment
Pvt. Ltd..
It has been a very enriching journey, helping me realize my hidden potential to work under tiring
circumstances.
The application of theoretical learning to practical experiences, cannot be taught in a classroom
but can be learnt only through actual work, and this company has provided an ideal platform for
the best practical experience one can get.
Working under a group of seniors who don’t give the solution directly but rather leading you to
it, thereby assisting you in finding your way through the obstacles, helped develop my thinking
actual experience one can get.
In-plant training has made an invaluable contribution to starting my career.
BIBLIOGRAPHY