Professional Documents
Culture Documents
medicine
Chapter 8
The first application of transmitting light through optical fibers in medicine was the
development of endoscopes that were based on optical fibers.
The fiberoptic endoscope is a medical system that makes use of the fiberscope for imaging
inside the body.
In addition, this system contains several ancillary channels that are used for the
introduction of thin mechanical tools or for the introduction of liquids inside the body.
3) Endoscopy
An endoscopy is a special test that involves
examining the inside of a person's body using
an endoscope.
The name endoscope is based on two Greek
words : endon (within) and skopein (view).
It's being used to describe optical instruments
that facilitate visual inspection and imaging of
internal organs.
In addition, an endoscope may include ancillary
channels through which the physician can
perform other tasks.
The endoscopic system consists of several
subsystems: the endoscope itself; subsystems
that provide, irrigation, and suction; and auxiliary
medical subsystems such as electrocautery and
biopsy forceps
Endoscopes may be inserted into the body
through natural openings (mouth, rectum, etc.)
or through small incision in the skin.
flexible
shaft
proximal
end
Example 8-1: An optical fiber has core with an index of refraction ncore=1.40 and a diameter d= 80 μm.
It is surrounded by a cladding with an index of refraction ncladding=1.34.
Light is sent into the fiber along its axis. Calculate the smallest outside radius R min permitted for a bent
in the fiber if no light is to escape.
The distal end of the endoscope houses the distal tips of several of the optical and
mechanical components.
These are the tip of the light guides, which provide illumination; the distal tip of the imaging
bundle; and a miniature objective lens (or a complex lens system), which is attached to the tip
of this bundle.
A thin, flat protective window that is easy to clean may cover the tip to protect it from
contamination by blood or debris.
The distal end of the endoscope is often immersed in fluids, secretions, or blood, which tend to
hinder viewing. A flushing port is supplied through which a stream of water can be sprayed
on the distal tip of the imaging system to keep it clean.
The distal end may also contain one or several outlets that are used for irrigation, aspiration,
suction, or insertion of thin instruments.
The flexible shaft holds the light guide, imaging bundle, and ancillary tubes together.
The shaft is connected at one side to the distal end and at the other side to the proximal end
of the endoscope.
This shaft is often constructed of steel mesh and can be strengthened further by a steel spiral.
This construction serves two purposes. It stiffens the endoscope, giving it mechanical
support, and protects the delicate optical and mechanical components inside from being
crushed or kinked.
The flexible shaft is sheathed by a plastic jacket that is biologically inert and that forms a
hermetic seal. It protects the inner parts of the endoscope from interaction with water, blood, or
other biological fluids.
The proximal end of the endoscope includes viewing optics, controls, and several ports.
These are often housed in a handpiece held by the physician. All the control knobs, buttons,
levers, and the photographic equipment are connected to this handpiece
For large enough endoscopes, a tiny video camera can be attached to the tip of the
endoscope for what is called direct imaging.
The same integrated circuit chips as the imaging elements in commercial cameras are used
in these endoscopes.
The resulting image can be viewed directly on a monitor and stored for future reference.
New endoscopes have digital capabilities for manipulating and enhancing the video images.
Ultrathin endoscopes
In some applications it is desirable to have endoscopes whose diameter is smaller than that
of thin endoscopes.
Ultrathin endoscopes have been developed with outer diameters of less than 1 mm.
The imaging bundle of such an endoscope consists of less than 3000 fibers each of diameter
of roughly 2-5 μm.
The outer diameter of the imaging bundle is often less than 0.3 mm and the optical resolution
is lower than that of a thin endoscope. This endoscope is too thin to include a practical
ancillary channel.
Ultrathin endoscopes have been used clinically for imaging of peripheral and coronary
arteries.
Because the body's organs naturally settle compactly together, leaving no clear air-filled
spaces, a gas channel is usually included and used to inflate the region to be examined
with inert carbon dioxide gas to lift and separate the tissues.
Similarly, another insert is used for injecting water and suctioning off fluid and debris.
Other common tools include forceps for manipulating and removing tissue for
examination, tiny scalpels, miniaturized staple guns, tools for making sutures, and snares
for both cutting and heating tissue.
The snare works by being looped around a region of tissue while an electric current is
passed through the wire.
This heats the wire, causing the tissue to heat up as well, and destroy tissue and seal off
blood vessels at the same time, an effect called cauterization.
Medical endoscopes go by specific names that are determined by which part of the body
each scope is used to image.
In principle, the various endoscopes are similar to each other, although distinct differences
were introduced to make the endoscopes more suited to each particular discipline.
Endoscopes, like other optical systems, are designed and built to perform specific goals.
The first image shows, at left, a polyp growing within the colon, as seen through an
endoscope.
In the middle two images, a snare is used to hook and remove the polyp.
Diagnostic medical techniques can be divided into two classes: invasive and noninvasive.
The noninvasive methods, do not cause pain to the patient and are potentially safe.
The invasive methods, which require penetration into the body, are still necessary because
they provideinformation that cannot be obtained otherwise.
Optical fibers are the basis for a relatively new family of "least invasive" diagnostic tools
called optical fiber sensors.
Optical fiber sensors make use of electronic devices that perform chemical/physical
analysis inside the body (They collect measurement information over fiber optics).
They rely on the fact that alterations in a specific physical/chemical property of the medium
being probed would cause a predictable change in light transmission characteristics of the
optical fiber.
Specificity: The measurement of only one parameter at a time without being affected by
other parameters.
Sensitivity: The ability to measure small changes in a given parameter.
Accuracy: The ability to give an accurate and stable measurement over a long period.
Cost: A low cost would enable the sensor to be disposable.
In principle, fiberoptic biomedical sensors have all these attributes, including the following
advantages:
Miniaturization: very thin fibers can easily be inserted into various places in the body (in
vivo and in real time measurements).
Biocompatibility: All the materials from which the fibers and transducers are made are
nontoxic and compatible with biological tissue.
Accuracy: Fiberoptic sensors may be in principle fast, stable, and accurate.
Safety: No electrical wires and circuits are introduced into the body.
At present, blood samples are generally sent to a laboratory for a chemical analysis of one or
more of their constituents.
An alternative is the use of miniaturized sensors that can be inserted into the body of the
patient.
It is possible to make a sensor small enough to fit into blood vessels without interfering too
much with the blood flow.
The sensor should be able to perform chemical analysis in real time and should be
disposable to avoid transmissible diseases.
In a typical sensor, a thin optical fiber is part of a complex optical system. The fiber is placed
in a location and is used to measure some parameter such as temperature, or pressure.
A typical fiber optic sensor is shown schematically in the figures bellow.
The fiber optic sensors fall into two categories : direct and indirect.
In direct sensors, the end of the fiber is bare simply inserted into the sampling region.
Indirect sensors incorporate some transducer (called optrode) at the end of the fiber.
In both cases, the optical fibers are the vehicles for transmitting the light into the body and
back again. The major advantage of optical fibers is that they enable measurements to be
made inside the human body (even inside the heart or other major organs).
Fiber optic sensors can be broadly divided into physical and chemical.
Physical sensors respond to some physical change in the sample such as temperature,
pressure, or flow
Chemical sensors respond to changes that are chemical in nature such as pH or
pCO2,pO2, or glucose content in the blood
4-1) Direct sensors
Photometry
Monochromatic light is directed onto tissue or blood and the absorption or reflection at this
wavelength is measured.
If this is performed at many wavelengths, the spectral absorption or the spectral reflection of
the sample is obtained.
Such spectra are "fingerprints" of specific compounds in tissue (or in blood).
Fluorometric method
The fiberoptic radiometry system consists of an infrared detector and a flexible IR fiber.
Cardiologists need to measure blood supply to the heart in order to know if the coronary
arteries are blocked and to what degree.
Flow: volume of fluid that passes through the end of the tube per unit time. (cm 3/s).
According to the Doppler effect, the change in frequency of the reflected radiation is a
function of the targeted object's relative velocity.
the velocity of the object can be obtained by measuring the change in frequency of the
reflected laser light.
LDV can be used with fiberoptic systems.
The only difference is that the incident and reflected laser beams are sent through optical
fibers, rather than through air.
The method is then called fiberoptic LDV or FOLDV.
A fiber is inserted into a blood vessel.
The laser beam interacts with a small volume of blood near the distal end of the fibers.
The light scattered back from the flowing blood cells is collected and transmitted back.
This scattered light is shifted in frequency (with respect to the regular frequency of the
laser) because of the Doppler effect.
The frequency shift is proportional to the velocity.
If a cardiologist is interested in determining the blood flow in a diseased coronary artery, an
ultrathin angioscope can be threaded into the artery and positioned next to the blockage
under observation.
Direct chemical sensors
Oximetry
Oxygen in the blood is carried by the hemoglobin in red blood cells.
Oxygen saturation is the ratio (in percent) between the oxygen content in a blood sample
and the maximum carrying capacity.
The blood in the arteries should be more than 95% saturated and that in veins about 75%
saturated.
Optical methods can be used to monitor the oxygen saturation, based on different optical
properties of saturated and desaturated hemoglobin.
The optical absorption spectrum and the optical reflection spectrum are different for each.
Cancer:
The auto-fluorescence emitted by healthy tissue is different from that of cancerous tissue in
both the spectral emission and the temporal behavior.
Such differences may be used to detect cancer endoscopically.
Cardiology:
In order to deal with plaque blockages inside the arteries, the plaque must be identified
during endoscopic imaging.
The intrinsic auto-fluorescence spectrum of plaque can be compared to that of the normal
blood vessel wall.
Dentistry:
The luminescent emission from a healthy tooth is different from that emitted by a tooth with
caries.
When excited at 480 nm, the carious areas absorb more than non-carious areas and
also emit more red light.
Analysis of the emitted light signal can thus reveal whether the tooth is healthy or not.
Similar fiberoptic methods have been applied inside the tooth to determine the state of the
root canals
4-2) Indirect sensors
For indirect sensors, a miniature transducer is attached to the distal tip of one or two
fibers (a miniature chemical/physics laboratory).
The optical transducer is called an optrode
The input light sent through the fiber actuates the transducer, which in turn interacts with
the sample.
The output light sent back from the transducer is analyzed by an optical processing
system.
If the optrode is physical in nature, the parameters that can be measured are physical,
such as temperature or pressure.
If it is chemical, the parameters are chemical, such as pH or glucose content in the
blood.
Fiberoptic indirect temperature sensors are based on optrodes in which some physical
property changes as a function of temperature.
Liquid crystals (LSs) show a dramatic change in color in response to a temperature change.
LCs are used in inexpensive (or disposable) thermometers, in which the temperature is
shown by colored digits.
These LCs were attached to the distal end of an optical fiber and used in the same manner
for measuring temperatures.
Luminescent probe
pH Sensors
In fiberoptic pH probes, an optrode is attached to the distal tip of the fiber and is coated
with a thin protective coating.
The coating permits small ions, such as hydrogen ions, to penetrate and interact with the
optrode but keeps larger ions out.
When the whole tip is inserted into blood, the interaction with the optrode, through the
coating, gives rise to optical changes which are measured through the fiber.
A common technique for determining pH is based on substances called indicators, which
are used to indicate the acidity or alkalinity of a solution by changing color.
The indicator changes color depending on the pH, in the physiological range of interest
(pH between 7.0 and 7.4).
This effect can be quantified by measuring the optical absorption of the indicator (through
an optical fiber).
Light sent through the input fiber is scattered by
the indicator.
Part is absorbed by the indicator and the rest is
transmitted back through the output fiber.
The amount of light transmitted back depends
on the absorption in the indicator.
Two irradiating colors, red (600 nm) and green
(550 nm), from two LEDs are used alternately.
The green light is absorbed by the indicator and the amount transmitted is measured by a
small semiconductor photodetector.
This amount depends on the pH.
The red light is not pH dependent, serving only to calibrate the system.
The ratio (green)/(red) is a direct measure of the pH.