Features Introduction
Operation
Maintenance
Upgrade
Touch Screen
Nurse Call
5 © 2018 Mindray Confidential
Features Introduction
Rotatable
Notes:
Physiological alarms and alarm settings only for the new
TM80(System software version 03.00.00).
PN Description
6-Lead leadset, New Telemetry,
009-004796-00
Disposable , AHA, Snap, 24″
5-Lead leadset, New Telemetry, AHA, Snap,
009-004782-00
24″
3-Lead leadset, New Telemetry, IEC, Snap,
009-004768-00
24″
PN Description PN Description
PN Description
040-000983-00 Bladderless Adult Cuff(CM1303)
Analysis Software
12 PDI Sani-Cloth® AF3 Germicidal Disposable Wipe 29 Tristel Duo 46 Mikrozid® AFwipes
13 PDI Sani-Cloth® Bleach Germicidal Disposable Wipe 30 Tristel Jet 47 Terralin® Liquid
HEALTH ESSENCE Double-chain
14 PDI Sani-Cloth® HB Germicidal Disposable Wipe 31 Tristel Fuse for surfaces 196ppm 48
quaternary ammonium salt disinfectant
15 PDI Sani-Cloth® Plus Germicidal Disposable Cloth 32 Surfanios Premium 0.25% 49 DIAN’ERKANG Surface Disinfectant
Only New TM80(System software 03.00.00 or above) can store data on local device. Old TM80 can only
store data on CMS.
Features Introduction
Operation
Maintenance
Upgrade
If alarm occurs on the TM80, you can not turn the display off.
Discharge Patient: When prompt message “Is This A New Patient?”, select Yes to admit a new patient.
Standby: Tap to enter Standby and press to exit Standby.
Print: Tap to send a print request to central station.
Alarm Reset: Reset the alarm both on the TM80 and CMS. Tap this menu to silence an alarm.
Features Introduction
Operation
Maintenance
Upgrade
front component : 1 、 touch screen 2 、 TFT screen 3 、 front shell 4 、 control board
3 Front Shell
051-002450-00/051-002450-
4 Main Board
04/051-003297-00
0156 main board PCBA(5G CE)
051-002059-00/051-002059-
5 Parameter Board
01(Resp)
New Telemetry ECG Board(WiFi) PCBA batte
Anti-defibrillation 115-032159-00/115-047668- ry
6 protect board & ECG 00(Resp) ECG socket kit(FRU)/ECG socket kit(Resp) FRU
115-053698-00+115-
054726-00(Front cover
package+0156 main screen 02.00.00 X Y X
board PCBA(5G
CE)FRU)
115-060665-00(Front
WI-FI 03.00.00
cover assembly X X Y
module or above
FRU(silex/CE))
Notes:
Software V03.00.00 can not downgrade to 02.00.00 or below.
Software V02.00.00 or below can not upgrade to 03.00.00 or above.
If customer wants to repair old TM80, please use the old spare part. If the old spare
part is out of stock, please change to the new hardware TM80.
Note:
New TM80 is incompatible with BeneVision CMS 07.17.00 or below. If you want to
online the TM80 on CMS, you must set the mode to telemetry or upgrade the CMS to
07.19.00 or above
• In the main menu, tap [Maintenance]-> Input the password-> [Accept]-> [Defaults]-> [Save Departmental
Defaults]-> [Yes].
Configuration Transmission
You can transfer the configuration from a TM80 to other TM80 via the Bluetooth.
• Main menu →[Maintenance] →input the password →[Accept] →[Defaults] →[Import Device Settings]
• Main menu →[Maintenance] →input the password →[Accept] →[Defaults] →[Export Device Settings]
Features Introduction
Operation
Maintenance
Upgrade
To TM80
TM80
Upgrade cable
PC or power
supply USB Power Adapter
Note:
USB requirements: FAT/FAT32 、 storage 512M at least
Power supply should be 5V, 500mA or higher.
Create the folder as the following photo and copy the upgrade file to USB disk
Connect USB disk, upgrade cable, power supply and TM80 together
Go to upgrade screen in either of the following ways
2.0
32