BeneVision TM80 Service Training

 PMLS Global Technical Support Department


 Version: V1.0
 Number:MJH-19013-BeneVision TM80
Contents

 Features Introduction
 Operation

 Maintenance
 Upgrade

2 © 2018 Mindray Confidential


Features Introduction
Clinical Application Scene

3 © 2018 Mindray Confidential


Features Introduction
System Overview

4 © 2018 Mindray Confidential


Features Introduction
Appearance

Touch Screen

 LCD: 3.5” TFT color LCD  Size: Size:125.8 x 63.7 x 23 mm

 Screen: capacitive  Battery: a lithium-ion battery


or 3-AA batteries
 Pixels: 480 x 320

Power Off/On Main Menu


/Close screen

Nurse Call
5 © 2018 Mindray Confidential
Features Introduction
Rotatable

 ECG, HR, RESP, SpO2, PR,


NIBP
 For NIBP, you must use
BP10.

6 © 2018 Mindray Confidential


Features Introduction
Parameters and waveforms display

Customize the order of Drag up or down to read all


parameters and waveforms waveforms and parameters

7 © 2018 Mindray Confidential


Features Introduction
Alarm System

 Audio & Visual LED indicators (Red, Yellow, Cyan).


 Technical and physiological alarms on TM80 with messages
and LED indicators.
 You can change alarm settings on TM80.
 Battery status displayed on device and the BeneVision CMS.
 All alarms on CMS.

Notes:
 Physiological alarms and alarm settings only for the new
TM80(System software version 03.00.00).

8 © 2018 Mindray Confidential


Features Introduction
Pairing

 Real-time pairing between TM80 and N


series patient monitor using MPAN or
Wi-Fi.
 You can configure the pairing on the N
series and set the communication
mode to Auto/MPAN First/Wi-Fi only.

9 © 2018 Mindray Confidential


Features Introduction
ECG  3/5/6-lead ECG monitoring.
6-lead  When connecting with CMS 07.15.00 or
above, it supports 6 leads transmit to 12
leads function.
 25 types of arrhythmia analysis events,
including AF.
 Display ST and QT analysis. including ST
values, ST Graphic, ST and QT templates
analysis.

 High quality materials for ECG cable and


wires

PN Description
6-Lead leadset, New Telemetry,
009-004796-00
Disposable , AHA, Snap, 24″
5-Lead leadset, New Telemetry, AHA, Snap,
009-004782-00
24″
3-Lead leadset, New Telemetry, IEC, Snap,
009-004768-00
24″

10 © 2018 Mindray Confidential


Features Introduction
ST analysis and QT measurement

baseline Beyond the


In the normal
normal range
range

11 © 2018 Mindray Confidential


Features Introduction
SpO2

 Provide SpO2 value, PI numeric


and pleth waveform

512E IPX7 (Adult) Mindray SpO2 Module

PN Description PN Description

Reusable SpO2 Sensor , Mindray SpO2 module with


512E-30-21373 115-029488-00 extension cable
Adt,Finger-Tip

12 © 2018 Mindray Confidential


Features Introduction
NIBP-BP10

 Display: 2.4 inch


 Support ABPM
 Model: manual and auto
 Batteries: disposable and rechargeable
 Protection against: IP32
 Specification: 300 times measurement
 Data transfer to CMS
 Connect to TM80 with Bluetooth

support water soaking


CM1303/ and washing
CM1302

PN Description
040-000983-00 Bladderless Adult Cuff(CM1303)

040-000978-00 Bladderless Child Cuff(CM1302)

13 © 2018 Mindray Confidential


Features Introduction
ABPM

Analysis Software

 Measure Mode: day or night


 Interval: 5min, 10 min, 20 min, 30 min, 1 h,
1.5h, 3h
 Report: tabular and graphic trend, histogram,
pie
 Support Print report

14 © 2018 Mindray Confidential


Features Introduction
Disinfecting
Cleaning Agents Name
No. No. Cleaning Agents Name No. Cleaning Agents Name

Rely+OnTM Virkon® Powder Used as


1 1-Propanol 50% 18 35 Aniosurf ND Premium 0.25%
1% solution*
2 Alpet® D2 Surface Sanitizing Wipes 19 Sodium hypochlorite bleach 0.5% 36 Mikrobar Tissues
VIRAGUARD Hospital Surface
3 CIDEX® OPA Solution 20 37 Cleanisept wipes
Disinfectants
Clorox Dispatch® Hospital Cleaner Disinfectant
4 21 Virex® II 256 ( 1 : 256 ) 38 mikrozid PAA wipes
Towels with Bleach
5 Clorox Healthcare® Bleach Germicidal Wipes 22 Virex® TB 39 mikrozid sensitive wipes
Clorox Healthcare® Hydrogen Peroxide Cleaner HEALTH ESSENCE Disinfecting
6 23 40 Alcohol 70%
Disinfectant Wipe Effervescent Tablets
HEALTH ESSENCE Surface Spray
7 Diversey Oxivir® TB Wipes 24 41 Glutaraldehyde 2%
Disinfection
8 Hydrogen peroxide 3% 25 DIAN’ERKANG Surface Wipes 42 Ecolab Incidin Oxywipes S.

9 Isopropanol 70% 26 DIAN’ERKANG Spray Disinfectant 43 Descosept forte

10 Metrex CaviCide1™ 27 Clinell Universal Sanitising Wipes 44 Descosept AF

11 Metrex CaviWipes™ 28 Clinell Sporicidal Wipes 45 Dismozon® plus 0.4%

12 PDI Sani-Cloth® AF3 Germicidal Disposable Wipe 29 Tristel Duo 46 Mikrozid® AFwipes

13 PDI Sani-Cloth® Bleach Germicidal Disposable Wipe 30 Tristel Jet 47 Terralin® Liquid
HEALTH ESSENCE Double-chain
14 PDI Sani-Cloth® HB Germicidal Disposable Wipe 31 Tristel Fuse for surfaces 196ppm 48
quaternary ammonium salt disinfectant
15 PDI Sani-Cloth® Plus Germicidal Disposable Cloth 32 Surfanios Premium 0.25% 49 DIAN’ERKANG Surface Disinfectant

16 PDI Super Sani-Cloth® Germicidal Disposable Wipe 33 Surfa 'safe

17 Perform® Classic Concentrate OXY 0.5% 34 Wip' Anios premium

15 © 2018 Mindray Confidential


Features Introduction
Data Storage

 Only New TM80(System software 03.00.00 or above) can store data on local device. Old TM80 can only
store data on CMS.

Item TM80 CMS

event 200 720 hours

Trends 48 hours 240 hours

Full disclosure 2 hours 240 hours

16 © 2018 Mindray Confidential


Contents

 Features Introduction

 Operation
 Maintenance

 Upgrade

17 © 2018 Mindray Confidential


Operation
Display on/off and Screen lock

• Turn the display on/off

If alarm occurs on the TM80, you can not turn the display off.

• Unlock the screen


 If you set the screen lock, you need to input correct password to unlock the screen after the
display turns off.
 Enable the screen lock: Maintenance->Maintenance Password->Set the screen lock mode to
Locked or View Only->Enter the screen lock password.

18 © 2018 Mindray Confidential


Operation
Quick Key

The quick key is configurable.


->Maintenance->Quick Keys

 Discharge Patient: When prompt message “Is This A New Patient?”, select Yes to admit a new patient.
 Standby: Tap to enter Standby and press to exit Standby.
 Print: Tap to send a print request to central station.
 Alarm Reset: Reset the alarm both on the TM80 and CMS. Tap this menu to silence an alarm.

19 © 2018 Mindray Confidential


Contents

 Features Introduction

 Operation
 Maintenance

 Upgrade

20 © 2018 Mindray Confidential


Maintenance
Hardware Structure

21 © 2018 Mindray Confidential


Maintenance
Hardware

front component : 1 、 touch screen 2 、 TFT screen 3 、 front shell 4 、 control board

5 、 parameter 6 、 Anti-defibrillation protect


7 、 back component 8 、 battery
board board & ECG connector

22 © 2018 Mindray Confidential


Maintenance
No. Spare Part PN discriptions
1 Touch Screen
115-035605-00/115-048398-
2 TFT Screen
00
Front cover package

3 Front Shell
051-002450-00/051-002450-
4 Main Board
04/051-003297-00
0156 main board PCBA(5G CE)
051-002059-00/051-002059-
5 Parameter Board
01(Resp)
New Telemetry ECG Board(WiFi) PCBA batte
Anti-defibrillation 115-032159-00/115-047668- ry
6 protect board & ECG 00(Resp) ECG socket kit(FRU)/ECG socket kit(Resp) FRU

7 Back Component 115-032160-00 Back cover service kit(FRU)


8 Battery 115-030107-00(sales part) TP Li-ion battery Package

23 © 2018 Mindray Confidential


Maintenance
Compatibility
Compatibility
Item Changes Version
051-002450-00 051-002450-04 051-003297-00
(0156 main board (0156 main board (0156 main
PCBA(5G CE)) PCBA(5G CE)) board(silex)PCBA)
115-035605-00+115-
048398-00(Front cover 01.08.00
/ Y X X
package+0156 main or below
board PCBA(5G CE))

115-053698-00+115-
054726-00(Front cover
package+0156 main screen 02.00.00 X Y X
board PCBA(5G
CE)FRU)
115-060665-00(Front
WI-FI 03.00.00
cover assembly X X Y
module or above
FRU(silex/CE))

Notes:
 Software V03.00.00 can not downgrade to 02.00.00 or below.
 Software V02.00.00 or below can not upgrade to 03.00.00 or above.
 If customer wants to repair old TM80, please use the old spare part. If the old spare
part is out of stock, please change to the new hardware TM80.

24 © 2018 Mindray Confidential


Maintenance
Battery

Item Model 3AA Lithium


5-lead ECG ≧20 hours ≧40 hours
Display is off
5-lead ECG + Mindray SpO2 ≧16 hours ≧32 hours

5-lead ECG / ≧14 hours


Display is on
5-lead ECG + Mindray SpO2 / ≧12 hours

25 © 2018 Mindray Confidential


Maintenance
Central Charging

 Support maximum number of 10 slots


simultaneously used.
 Each slot has a separate charging status
indicator light.
 Support installing on the wall.
 It is recommended to change the Li-ion
battery every 2 years or 500 charge cycles.

26 © 2018 Mindray Confidential


Maintenance
Switch mode

 You can switch the mode of TM80, monitor or


telemetry.
 This feature is only for the new TM80 (system
software 03.00.00 or above).

Note:
 New TM80 is incompatible with BeneVision CMS 07.17.00 or below. If you want to
online the TM80 on CMS, you must set the mode to telemetry or upgrade the CMS to
07.19.00 or above

27 © 2018 Mindray Confidential


Maintenance
Configuration

 Saving the Current Settings

• In the main menu, tap [Maintenance]-> Input the password-> [Accept]-> [Defaults]-> [Save Departmental
Defaults]-> [Yes].

 Configuration Transmission

You can transfer the configuration from a TM80 to other TM80 via the Bluetooth.
• Main menu →[Maintenance] →input the password →[Accept] →[Defaults] →[Import Device Settings]
• Main menu →[Maintenance] →input the password →[Accept] →[Defaults] →[Export Device Settings]

28 © 2018 Mindray Confidential


Contents

 Features Introduction

 Operation
 Maintenance

 Upgrade

29 © 2018 Mindray Confidential


Upgrade
Software Upgrade

To TM80

TM80

Upgrade cable

USB Disk USB Port

PC or power
supply USB Power Adapter

Note:
 USB requirements: FAT/FAT32 、 storage 512M at least
 Power supply should be 5V, 500mA or higher.

30 © 2018 Mindray Confidential


Upgrade
Upgrade Process

 Create the folder as the following photo and copy the upgrade file to USB disk
 Connect USB disk, upgrade cable, power supply and TM80 together
 Go to upgrade screen in either of the following ways

1. Menu -> Maintenance -> Service -> Software Update

2. Press the menu button and power button simultaneously

31 © 2018 Mindray Confidential


Version Information

Version No. Editor Date Description

1.0 Jerry Yang 2019/3/22 First Edition