Professional Documents
Culture Documents
Pharamacuetical Analaysis II
(Phar 3122)
1
UV- Visible
spectrophotometry
2
Application of instrumental Technique in
Quality control Pharmaceuticals
Pharmaceutical analysis- deals with methods for determining the
chemical composition of pharmaceutical samples.
Two types
Classical (or so-called wet chemical methods) and
Instrumental
Instrumental methods involve studying the physical properties of
analytes.
Conductivity, electrode potential, light absorption or emission, mass-to-charge
ratios are properties often probed.
Qualitative - chromatography, electrophoresis and identification by measuring physical property
3
Types of Instrumental Methods
Property Example
Radiation emission Emission spectroscopy: fluorescence,
Radiation absorption Absorption spectroscopy -,
Spectrophotometry, photometry
Radiation scattering Turbidity, Raman
Radiation refraction Refractometry, interferometry
Radiation diffraction X-ray
Radiation rotation Polarimetry, circular dichroism
Electrical potential Potentiometry
Electrical charge Coulometry
Electrical current Voltammetry - polarography
Electrical resistance Conductometry
Mass-to-charge ratio Mass spectrometry
4
Introduction
5
Intro…
Spectroscopy
• Is the study of interaction between electromagnetic radiation and matter.
Spectrophotometry
It is more specific than the general term Electromagnetic spectroscopy in
6
Electromagnetic radiation
EMR is a form of energy whose behavior is described by the properties of
7
EMR…
8
EMR..
The human eye is only sensitive to a tiny proportion of the
total electromagnetic spectrum between approximately
380 and 780 nm and within this area we perceive the
colors of the rainbow from violet through to red.
If, however, only a portion of the light is absorbed and the
balance is reflected, the color of the sample is determined
by the reflected light.
Thus, if violet is absorbed, the sample appears yellow-
green and if yellow is absorbed, the sample appears blue.
The colors are described as complementary.
9
EMR…
However, many substances which appear colorless do have absorption spectra.
In this instance, the absorption will take place in the infra-red or ultraviolet and
not in the visible region
10
EMR……
Wave Properties of EMR
• EMR consists of oscillating electric and magnetic fields that
propagate
through space along a linear path and with a constant velocity.
• In a vacuum, EMR travels at the speed of light, c, which is 3 x
108 m/s
1
1
EMR……
An electromagnetic wave is characterized by several
fundamental properties, such as:
1-Wavelength (λ, lambda): which is the linear distance measured
along the line of propagation, between crest of one wave to that
of the next wave.
2-Amplitude: which is the vertical distance from midline of a
wave to the peak or trough.
12
EMR……
13
Example:
if we have a visible radiation of 500 nm, then:
14
EMR…
Light as energy
Light like any other matter consists of energy packets called photons.
The absorption and emission of light by compounds occur in these
packets (photons).
The energy (E) of a photon is directly proportional to the frequency
and inversely proportional to the wavelength.
It can be related to C, and by the following equation:
E = h = h C/
Where h is a constant called Planck’s constant , which equal to 6.625 x 10-27 erg.
sec.
16
How Light Interacts With Matter
18
How Light Interacts….
19
How Light Interacts….
Atomic absorption
In an atom there are a number of shells and of subshells
where e-’s can be found.
The energy level of each shell & subshell are different and
quantised.
The exact energy level of each shell and sub-shell varies with
sub.
Under normal situation an e- stays at the lowest possible shell
- the e- is said to be at its ground state.
Upon absorbing energy (excited), an e - can change its orbital
to a higher one - we say the e- is at its excited state.
20
How Light Interacts….
The excitation can occur at different degrees
Low E tends to excite the outmost e-’s first
An e- at its excited state is not stable and tends to return its ground state.
21
How Light Interacts….
ΔE transition = E1 - E0
330 nm
= hv = hc/
590 nm
ΔE transition =
E2 - E0 = hv = hc/
Absorption and emission for the sodium atom in the gas phase, Illustrates discrete energy transfer
22
How Light Interacts….
Molecular Absorption
More complex than atomic absorption because many more
potential transitions exist.
A molecule may absorb light energy in three ways:
By raising an electron to a higher energy level (electronic).
By increasing the vibration of constituent nuclei (vibrational).
By increasing the rotation of molecule about its axis
(rotational)
E total = E electr + E vibrat + E rotat
23
How Light Interacts….
Molecular spectrum
24
How Light Interacts….
When a molecule interacts with photons in UV-Vis region,
E = Es - Eg = h = h C/
The energy E associated with the absorption bands of a
molecule is given by:
E = E electronic + E vibrational + E rotational
cules)
expressed in terms of absorbance,
Line spec.(atoms)
(mole
transmission, Intensity.
spec.
It may be:
Band
a) line spectrum: occur with atomic
spectra such as sodium metal which
has a sharp line of at 589 nm.
26
Spectrum…
wavenumber or wavelength.
The wavenumber or λ at the tip of peak is the most important,
27
Spectrum
A broad peak may sometimes consist of several peaks
28
UV-Visible Spectrophotometry
What occur to a molecule when absorbing UV-visible photon?
A UV-visible photon (200-800nm) promotes a bonding or non-bonding electron
are:
Chromophores -functional groups each of which absorbs a characteristic
UV or visible radiation.
29
Types of electronic transitions
Absorption of radiation in the UV-VIS region depends upon the number
30
Types of electronic…
In excited state:
31
Types of electronic…
32
Types of electronic…
*- Antibonding
*- Antibonding
* Transtion
n-* Transtion
* Transtion
n-* Transtion
Energy
n- Nonbonding
Bonding
Bonding
35
Types of electronic…
near UV region.
They are useful as common solvents in near UV region.
36
Types of electronic…
Cut-off wavelengths of some common solvents:
Solvent , nm Solvent , nm
38
Some important…
Bathochromic (Red) shift:
Shift of absorption to longer wavelength due to substitution and
solvent effects.
Multiple conjugations -Reduce transition energy-high sensitivity
-Enhance probability of transition.
CH2=CHCH2CH2CH=CH2 λ max =185 nm &
CH2=CHCH=CH2 λmax =217
Hypsochromic (Blue) shift:
It is shift of absorption to shorter wavelength.
Hyperchromic & hypochromic effects:
It is the increase and decrease in absorption intensity respectively.
39
Absorption characteristics of
Chromophores
1. Ethylenic Chromophores:
Their bands are difficult to observe in near UV region, so they
are not useful analytically.
However, substitution and certain structural features may cause
red shift rendering the band observable in the near UV region.
Examples:
Alkyl substitution: cause red shift due to hyper-conjugation and
stabilization of excited state
Exocyclic nature: cause red shift due to relaxation of strain upon excitation.
Attachement to auxochromes: cause red shift and increased absorption
intensity due to extension of conjugation.
40
Absorption Characteristics…
C=S…
They exhibit some common characteristics; n- * band in
the
range of 275-300 nm., which is the most apparent band,
has low
intensity and long wavelength.
This band undergoes a blue shift on increasing the solvent
polarity due to increasing the energy of transition as a result of
H
bonding
41 Alkyl substitution; Cause red shift due to hyper-conjugation.
Absorption Characteristics…
3. Conjugated Chromophores
CH2 = CH – CH2 – CH = CH2
170-180 nm
CH2 = CH – CH = CH2
4. Aromatic Systems
(I) Benzene ring :
Benzene has three maxima at 184 nm ( the most intense),
204 nm and at 254 nm.
The first two bands have their origin in the Ethylenic
* transition, while the longest λ band is a specific
feature of benzenoid compounds.
This band abbreviated B-band, which is characterized by
vibrational fine structures
43
Absorption Characteristics…
In structure elucidation both the B-band and the 204-nm
ethylenic
band, (E-band) are useful while the far UV band (184 nm)
is
unsuitable for analytical purposes.
184 nm 254 nm
204 nm
44
Absorption Characteristics…
45
Absorption Characteristics…
CH3-CH2-CH=CH2 or CH3-CH2-CH=O
CH3-CH2COOH or CH3-CH2CH=O
CH2=CH-CH=CH2 or CH2=CH-CH2-CH=CH2
4
6
Effect of pH on absorption spectra
The spectra of compounds containing acidic or basic groups are dependent on the pH of
alkaline medium
Spectrum in alkaline medium exhibits bathochromic shift with hyperchromic effect.
The red shift is due to the participation of the pair of electrons in resonance with
the electrons of the benzene ring, thus increasing the delocalization of the
electrons.
47
Effect of pH on….
OH O O
-
OH
H+
48
Effect Of pH On….
On the other hand, UV spectrum of aniline in acid medium
shows
hypsochromic (blue) shift with hypochromic effect (decrease
in
absorption intensity).
This blue shift is due to the protonation of the amino group,
hence the
pair of electrons is no longer available and
+ the spectrum in
NH2 NH3
this case is
+ H+
similar to that of benzene (thus
- H+ called benzenoid spectrum).
50
Effect of Solvents…
n-* Transition:
Blue shift with increasing solvent polarity due to stabilization of
the
ground state by hydrogen bonding: R-C=O…….HOR
Hydrogen bonding lowers the energy of the ground state (i.e.)
increase
energy of transition and hence decrease wavelength, .
S1
S2
G1
G G2
- * bands of enones n - * bands of enones
51
Calculation of max of an organic compound
I- According to Woodward Feiser Rules:
Rules For conjugated dienes:
These rules specify a base value ( 214 nm ) for the parent diene which is 1,3-
butadiene.
R – CH = CH – CH = CH – R’
The value of 214 nm is red-shifted upon:
CH3
R CH2
gives a different value from that of heteroannular systems where the two
double
homoannular heteroannular
5
Extra double bond
3
OCH3
OH
SH OAc
Cl
Che ck tha t this che mica l compound conta ining
2-Extra double bonds
5- Auxochrome s a tta che me nts
5- Ring re s idue s
No a lkyl s ubs titutions
3- Exocyclic double bounds
One homoa nnula r na ture
5
4
Woodward Rules for Conjugated Dienes can be summarized as :
Component nm
Base value for heteroannular or opened-chain dienes 214
Base value for homoannular dienes 253
Add the following Values to the base value:
(a) Each extra double bond in conjugation 30
(b) Each Alkyl Substituent or ring residue 5
(c) Each exocyclic nature 5
(d) Each auxochrome has its corresponding value:
- OAc 0
- OR (including OH) 6
- SR (including SH) 30
- Cl or Br 5
- NR2 (including NH2 & NHR) 60
(e) Solvent Correction 0
5
5
The following examples illustrate the use of these rules:
OH OH
SH Cl
Cl
NH2 OR NH2 Cl OCH3
Cl Cl Cl
OCH3
NH2 OR
OH OH
OH OH Br
Cl CH2
OR
OAc
Cl SH NH2
Cl SH
OH
Br
Cl OR Br
Cl
OR OR Br
NH2
5
7
II- Simplified Kuhn and Hausser rules:
These rules can be used for calculating max for conjugated polyenes as follows:
max (nm) = 134 n + 31
where n is the number of conjugated double bonds.
Example : Calculate the max of the following compound :
CH2OH
C CH OH
CH
OH
CHOH
Calculate the approximate number of double bonds present in each compound , if you
gave the following λmax for each:
1- 420 nm , 2- 530 nm , 3- 485 nm
4- 565 nm 5- 612 nm 6- 710 nm
5
9
Instrumental design
Components
A spectrophotometer is an instrument for measuring the T or
A
of a sample as a function of the wavelength of EMR.
The key components of a spectrophotometer are:
1. Source that generates a broad band of EMR
2. Wavelength selectors
3. Sample holder
4. One or more detectors to measure the intensity of
radiation
5. Signal Processor
60
Instrum…
i- Light Sources. Two types:
1- Continuous Sources: which produce spectra over a broad range(e.g.):
Tungsten lamp (provides visible spectrum; 400-1200 nm)
Deuterium lamp (provides ultra-violet spectrum; 190-400 nm)
62
Instrum…
b) Tungsten-filament: consists of a
tungsten filament contained in a glass
envelope.
The life of the lamp is limited by the
evaporation of tungsten.
c) Tungsten-halogen lamp:
• The halogen gas prevents the evaporation of
tungsten and increases the lifetime of the lamp
to more than double that of the ordinary
tungsten lamp.
• yields good intensity over part of the UV tungsten-halogen lamp
spectrum and over the entire visible range.
This type of lamp has very low noise and low
drift and typically has a useful life of 10,000 h.
Most spectrophotometers used to measure the
UV-visible range contain both types of lamps.
63
Instrum…
Either a source selector is used to switch
between the lamps as appropriate, or the
light from the two sources is mixed to
yield a single broadband source.
An alternate light source is the xenon
lamp which yields a good continuum
over the entire UV and visible regions.
The noise from currently available
xenon lamps is significantly worse than
that from deuterium or tungsten lamps xenon lamps
Xenon lamps are used only for
applications in which high intensity is
the primary concern.
64
Instrum…
ii. Wavelength selectors
Narrower bandwidth tend to enhance the sensitivity and selectivity
of the absorbance measurements and give a more linear r/ship
between the optical signal and concentration of the substance to be
determined
i.e. narrower bandwidth representing better performance.
Ideally, the output from a wavelength selector would be radiation of
a single wavelength.
Two types of wavelength selectors are used:
Filters and
Monochromators.
6
5
Instrum…
A. Filters:
Either absorption or interference filters are used for wavelength selection:
1. Absorption filters:
They are inexpensive and widely used for band selection in the visible
region.
66
Instrum…
2. Interference filters:
As the name implies, an interference filter relies on optical
interference to provide a relatively narrow band of radiation.
It consists of a transparent material (calcium or magnesium
fluoride) sandwiched between two semitransparent metallic films
coated on the inside surface of two glass plates.
The thickness of the dielectric layer is carefully controlled and
determines the wavelength of the transmitted radiation.
When it is subjected to a perpendicular beam of light, a fraction
passes through the first metallic layer and the other is reflected.
67
Instrum…
Fraction that is passed undergoes a similar partitioning upon
passing
through the second metallic film, thus narrower bandwidths are
obtained.
White
Glass plates
radiation
Dielectric layer
Narrow Mealic films
band
radiation
Figure 16; Interference
Interference Filters filter
68
Instrum…
B) Monochromators:
an entrance slit,
a collimating lens or mirror to produce a parallel beam of light
a prism or grating to disperse the radiation into its component
wavelengths
a focusing element and exit slit
69
70
Instrum…
Dispersion devices
It cause different wavelengths of light to be dispersed at
different angles.
When combined with an appropriate exit slit, these devices can
be used to select a narrow waveband
Two types of dispersion devices, prisms and holographic
gratings, are commonly used in UV-visible spectrophotometers.
A prism generates a rainbow from sunlight.
This same principle is used in spectrophotometers.
Prisms are simple and inexpensive, but the resulting dispersion
is angularly nonlinear (see Figure).
Moreover, the angle of dispersion is temperature sensitive.
71
Instrum…
For these reasons, most modern
spectrophotometers contain holographic
gratings instead of prisms.
These devices are made from glass blanks,
onto which very narrow grooves are ruled.
The dimensions of the grooves are of the
same order as the wavelength of light to be
dispersed.
Finally, an aluminum coating is applied to
create a reflecting source.
Light falling on the grating is reflected at
different angles, depending on the
wavelength.
Holographic gratings yield a linear angular
dispersion with wavelength and are
temperature insensitive.
72
Instrum…
However, they reflect light in different orders, which overlap.
As a result, filters must be used to ensure that only the light from
the desired reflection order reaches the detector.
A concave grating disperses and focuses light simultaneously.
A monochromator consists of an entrance slit, a dispersion device,
focusing mirror and an exit slit.
Ideally, the output from a monochromator is monochromatic light.
In practice, however, the output is always a band(group), optimally
symmetrical in shape.
The width of the band at half its height is the instrumental
bandwidth (SBW).
73
Instrum…
Optics
Either lenses or concave mirrors are used to relay and
focus light through the instrument.
Simple lenses are inexpensive but suffer from chromatic
aberration(devation from what is normal or desirable), that
is, light of different wavelengths is not focused at exactly
the same point in space.
Achromatic lenses combine multiple lenses of different
glass with different refractive indices in a compound lens
that is largely free of chromatic aberration.
Such lenses are used in cameras.
74
Instrum…
They offer good performance but at relatively high
cost.
Concave mirrors- are less expensive to manufacture
than achromatic lenses and are completely free of
chromatic aberration.
However, the aluminum surface is easily corroded,
resulting in a loss of efficiency.
75
Instrum…
iii- Sample cells (sample holders):
For UV/Vis instrument, this is a light tight box in w/c the container
holding the sample so/n is placed.
The container is called cuvette.
For the UV region sample compartment is made of quartz since
quartz
will not absorbed in the UV region.
For the Visible region, compartment composed of simple glass or
plastic cells since they absorb in the UV but not absorb in the visible
region.
In UV any solvent that does not have any Ñ-bonding can be used including
Instrum…
The standard path-length of cells
77
Instrum…
iV) Detectors: Two Types of detectors are used in this respect:
1- Heat Sensitive Detectors
2- Photoelectric Detectors
Photoelectric detectors are the most frequently used for this purpose.
They give electrical signal, which is directly proportional to the
intensity of
the transmitted light.
The following types of photoelectric detectors are used:
1- Photovoltaic cells 2- Phototubes
3- Photomultiplier Tubes (PMT’s) (The most widely used)
4- Photoconductivity tubes and Silicon photodiodes
7 The main three types are illustrated in the following part:
8
Instrum…
80
Instrum…
This type of detector yields high
sensitivity at low light levels.
However, in analytical spectroscopy
applications, high sensitivity is
associated with low concentrations,
which result in low absorbances,
which in turn result in high intensity
levels.
To detect accurately small
differences
between blank and sample
measurements, the detector must
have
low noise at high intensity levels.
81
Instrum…
C) Photodiodes
Photodiode detectors have a wider dynamic range and, as
solid-state devices, are more robust (stronger) than
photomultiplier tube detectors
In a photodiode, light falling on the semiconductor material
allows e- s to flow through it, thereby depleting the charge
in a capacitor connected across the material.
The amount of charge needed to recharge the capacitor at
regular intervals is proportional to the intensity of the light.
Earlier photodiodes had low sensitivity in the low UV
range, but this problem has been corrected in modern
detectors.
83
Instrum…
Some modern spectrophotometers contain an array of
photodiode detectors instead of a single detector.
A diode array consists of a series of photodiode detectors
positioned side by side on a silicon crystal.
Each diode has a dedicated capacitor and is connected by a
solid-state switch to a common output line.
The amount of charge needed to recharge the capacitors is
proportional to the number of photons detected by each
diode, which in turn is proportional to the light intensity.
84
Instrum…
The absorption spectrum is obtained by measuring the
variation in light intensity over the entire wavelength
range.
The array typically comprises between 200 and 1000
elements, depending on the instrument and its intended
application.
Photodiode arrays are complex devices but, because
they are solid state, have high reliability.
85
Instrum…
a)The conventional spectrophotometer
86
Instrum…
The absorbance of a sample is determined by measuring the d/c b/n
intensity of light reaching the detector without the sample (the
blank) and with the sample.
This design is well-suited for measuring absorbance at a single
point in the spectrum.
It is less appropriate, however, for measuring different compounds
at different wavelengths or for obtaining spectra of samples.
To perform such tasks with a conventional spectrophotometer,
parts of the monochromator must be rotated
This introduces the problem of mechanical irreproducibility into
the measurements.
Moreover, serial data acquisition is an inherently slow process.
87
Instrum…
b)The diode array spectrophotometer
Polychromatic light from a source is passed through the sample
area and focused on the entrance slit of the polychromator.
88
Instrum…
The bandwidth of light detected by a diode is related to
the size of the polychromator entrance slit and to the size
of the diode.
Each diode in effect performs the same function as the
exit slit of a monochromator.
The polychromator disperses the light onto a diode array,
on which each diode measures a narrow band of the
spectrum.
The polychromator and the diode array are contained in
a unit known as a spectrograph.
This configuration often is referred to as reversed optics.
89
Instrum…
To minimize possible photochemical reactions, a shutter is
used.
When the measurement is initiated, the shutter is
automatically opened, and light passes through the sample
to the array of diodes.
The difference in the intensities of the light reaching the
detector with and without the sample is measured.
A diode array spectrophotometer :
inherently very fast owing to its parallel data acquisition and
electronic scanning capabilities
has excellent wavelength reproducibility, and is highly reliable.
90
Instrum…
v) Signal Processors/Readout
Signal Processing
Amplifying the signal coming from the detector
Integrators
9
1
Instrum…
Configuration
Various configurations of spectrophotometers are available.
i) Single-beam design
Both conventional and diode array spectrophotometers are
single beam.
The reference spectrophotometers used by national standards
institutions such as the NIST in the US and NPL in the UK
are single beam.
Diode array spectrophotometers in particular are well-suited
to single-beam configuration.
92
Instrum…
93
Instrum…
Dual-beam design
In a conventional single-beam spectrophotometer, Lamp drift
can result in significant errors over long time intervals.
The dual-beam spectrophotometer was developed to
compensate for these changes in lamp intensity between
measurements on blank and sample cuvettes.
In this configuration, a chopper is placed in the optical path,
near the light source.
The chopper switches the light path between a reference
optical path and a sample optical path to the detector.
It rotates at a speed such that the alternate measurements of
blank and sample occur several times per second.
94
Instrum…
Figure below shows a schematic of a dual-beam
spectrophotometer.
Compared with single-beam designs, dual-beam instruments
contain more optical components, which reduces throughput
and sensitivity.
95
Instrum…
In addition, the more complex mechanical design of the
dual-beam spectrophotometer may result in poorer
reliability.
Single-beam instruments offer higher sensitivity and greater
ease of use, with drift typically only a factor of two worse
than that of dual-beam instruments.
The first commercially available diode array
Spectrophotometer was a multibeam design (see Figure
below).
96
Instrum…
The beam director is used to shift the beam alternately
through the reference position and as many as four sample
positions (for clarity only one is shown in the figure).
97
Instrum…
Split-beam design
This configuration enables the blank and the sample to be
measured at the same time.
98
Instrum…
Although the split-beam design is mechanically
simpler than the true dual-beam instrument and
requires fewer optical elements, the use of two
independent detectors introduces another potential
source of drift.
This design provides high stability, although not as
high as a dual-beam instrument since two detectors
can drift independently, and good noise, although not
as good as a single-beam instrument since the light is
split so that less than 100 % passes through the
sample.
99
Applications for Spectrophotometry
It is one of the most useful tools available to the experts for
analysis.
Important advantages of spectrophotometric methods include:
1. Wide applicability; large number of organic and inorganic
species absorb light in the UV-Visible ranges.
2. High sensitivity; analysis for concentrations in the range from
10-4 to 106 M are ordinary in the Spectrophotometric
determinations.
3. Moderate to high selectivity; Due to selective reactions,
selective measurements and different mathematical treatments.
4. Good accuracy; Relative errors in concentration measurement
lie in the range of 0.1 to 2 %.
5. Ease and convenience; Easily and rapidly performed with
modern instruments.
100
I- Qualitative Analysis
101
Qualitative…
UV-visible spectra generally show only a few broad
absorbance bands.
UV-visible spectroscopy provides a limited amount of
qualitative information.
Most absorption by organic compounds results from the
presence of π bonds.
The presence of an absorbance band at a particular
wavelength often is a good indicator of the presence of a
chromophore.
102
Qualitative…
103
Qualitative…
1. Color Tests:
The color of matter is related to its absorptivity or reflectivity
a) Substance may have a characteristic color of its own.
Eg. Vitamin B12, indicators, Cr salts and KMnO4 etc…
b) Substance may be treated with a certain selected reagent to
give
characteristic colored product (e.g.):
104
Qualitative…
Red color
10
5
Qualitative…
10
6
Qualitative…
UV-Visible spectrum give useful information about substance
via
examination of its max and emax, which could be correlated
with the
structuralcharacteristics
Absorption features (See of the
somefollowing table).
common organic
chromophores:
Chromophore Example λmax (nm) єmax
+ H+
- H+
OH O O
-
OH
H+
10
8
II- Quantitative analysis
109
Quantitative …
Laws of light absorption
When a monochromatic light passes through a cell containing
an
absorbing substance, then effects occurring will include:
I o = Ia + I t
Where I0 is the total light entering.
110
Quantitative …
A beam of parallel radiation before and after it has passed
through a layer of solution with a thickness of b cm and a
conc. of c of an absorbing species.
As a consequence of interactions b/n the photons and
absorbing particles, the power of the beam is attenuated from
Po to P.
The T of the solution is the fraction of incident radiation
transmitted by the solution.
That is,
The absorbance of a solution is defined by the equation
111
Lambert’s law Quantitative …
According to this law, each layer of equal thickness absorbs
an equal
fraction of light passing through it (i.e.) the rate of decrease
in I0 with
the thickness of medium is proportional to I0 itself.
or -δI0/ δ t α I0
After mathematical treatment the equation become
ln I0/It = µ t
113
Quantitative …
Beer’s Law:
It stated that absorption is proportional to the number of
absorbant
molecules in the light path
log Io/It =K C , Where K is proportionality constant
corresponding and C is the concentration in g/L.
114
This is a combination of both laws ; Log I0/It = A = abC
a(1%,1cm) bC
Quantitative …
where a is a constant called absorptivity, b is the path-length in cm
and C is the
concentration in grams/Liter.
A(1%,1cm) is the absorbance of 1%W/V(1gm/100ml) sln in a 1cm cell, b is
the path length (1cm) C is the conc. in gm/100ml
The value of a will clearly depend upon the method of expression of the
conc.
116
Quantitative …
The extinction coefficient (ε) is characteristic of a given
substance under a precisely defined set of conditions, such as
wavelength, solvent, and temperature.
In practice, the measured extinction coefficient also depends
partially on the characteristics of the instrument used.
For these reasons, predetermined values for the extinction
coefficient usually are not used for quantitative analysis.
Instead, a calibration or working curve for the substance to be
analyzed is constructed using one or more standard solutions
with known concentrations of the analyte.
117
Q. Determine the concentration of glycogen-iodine
complex if the absorbance of light is 0.4. Also, the
absorption coefficient is 0.20 at 450 nm. The size of
the cuvette is 2 cm.
Answer:
Using Beer-Lambert’s law, we know that
A= 0.4 = (0.20 lit/mol.cm)(c)(2cm)
c = 1mole/lit
118
Q.2. What are the conc of the ff solns of drugs in
g/100ml mg/100 ml
a. carbimazole A(1%,1cm) value =557 at 291nm,
measured absorbance 0.557 at 291nm
b. hydrocortisone sodium phosphate A(1%,1cm) value
333 at 248nm, measured absorbance 0.666 at 248 nm
Answer: A=a(1%,1cm) bc; A=0.557 a=557 , b=1 C=?
C = 0.557/557*1 =0.001g/100ml=1mg/100ml
119
Quantitative …
Deviations from Beer-Lambert’s Law
As per the Beer’s law discussed above, there is a direct
proportionality between the absorbance and concentration.
A plot of absorbance versus concentration is expected to be
a straight line passing through origin.
However, this is not always true; there are certain
limitations.
The law does not hold for all species under every condition.
Many times instead of a straight line, a curvature in the plot
may be observed as shown in fig below.
120
Quantitative …
Presence of Electrolytes ti on
via
The presence of small amounts of e de
tiv
colourless electrolytes which do not e ga
n
react chemically with the coloured
components does not affect the light
absorption as a rule.
However, large amounts of
electrolytes may affect the
absorption spectrum qualitatively as
well as quantitatively.
121
Quantitative …
This is due to the physical interaction between the ions of the
electrolyte and the coloured ions or molecules.
This interaction results in a deformation of the coloured ions
or molecules
Thereby causing a change in its light absorption property.
H+ Concentration
There are a number of substances whose ionic state in
solution is greatly influenced by the presence of H+.
Yellow orange
122
Quantitative …
In some cases, two absorbing species are in equilibrium
and have a common value of absorptivity at a certain
wavelength.
For example, in case of bromothymol blue the absorption
spectra at different pH values are different.
However, at wavelength of 501 nm, we see that all species
have same molar absorptivity (see Fig).
123
Quantitative …
Such a wavelength is known as
isosbestic point.
At this wavelength the Beer’s
law holds, though the
measurements have low
sensitivity.
However, such wavelengths
should be avoided for the
quantitative work.
124
Quantitative …
Complexation, Association or Dissociation
Some salts have a tendency to form complexes whose colours are
different from those of the simple compounds.
For example, the colour of cobalt chloride changes from pink to
blue due to the following complex formation.
125
Quantitative …
Non-monochromatic Nature of the Radiation
In order for the Beer’s law to hold, it is necessary that
monochromatic light is used.
The wider the bandwidth of radiation passed by the filter or
other dispersing devices, the greater will be the apparent
deviation of a system from adherence to Beer’s law.
126
Quantitative …
127
Quantitative …
(1) Real deviation:
Valid only for dilute (< 10-3 M) solutions. Why?
Because interactions between absorbing species at higher
concentrations affect ability to absorb radiation due to crowding,
molecular interaction and association as well as charge distribution
Because also e depends on refractive index which depends on
concentration of the solution ( mostly at high concentration levels)
129
Quantitative …
2- Stray light
Stray light is light which falls on detecor within a UV instruments without
passing
. through the sample
It can arise either from light scattering within the instrument or by entry of
light into
the instrument from outside
It gives a false low absorbance reading for the sample since it appears as
though the
sample is absorbing less light than it actually is.
This is most serious where the sample has a high absorbance e.g. at an
1
3absorbance of 2
Quantitative …
c) Other errors:
131
Multicomponent analysis
Multicomponent analyses using UV-visible spectra were not
widely applied.
Modern instruments yield more precise data, and modern
curve-fitting techniques give more accurate results.
For these reasons, multicomponent UV-visible analyses are
becoming more popular.
According to Beer’s law absorbance is proportional to the
number of molecules that absorb radiation at the specified
wavelength.
This principle is true if more than one absorbing species is
present.
13
2
Multicomponent…
133
Multicomponent…
The absorbance of the mixture at each wavelength is the sum of
the absorbance of each component at that wavelength
Thus for two components x and y, the equations are:
where
A′ is absorbance at wavelength ′
A′′ is absorbance at wavelength ′′
e′ and e′′ is molar absorptivity at wavelength ′ and ′′
c is concentration, and b is path length.
134
Multicomponent…
In practice, however, measurement errors always occur.
Such errors can affect significantly the accuracy of results when
spectra
overlap significantly.
Figure below shows a simulated two-component mixture with no
overlap of the spectra at the absorbance maxima.
135
Multicomponent…
In contrast, Figure below shows a simulated two-component
mixture with significant overlap of the spectra at the absorbance
maxima.
be:
13
6
Multicomponent…
A 10 % error occurs in the measurement of A′(x + y) and A′′
(x + y), the quantitative calculation yields the results shown
in table
137
Spectrophotometric titration
In photometric method of equivalent point detection of titrations, the appearance
titrant & the resulting product (s). The usual procedure is to select some wave
length at w/c only one component absorb.
In spectrophotometric titration, the absorbance of the so/n at a specified λ is
graphically.
The point at w/c the two straight lines intersects the end point.
140
Spectrophotometric titra…
The graph is constructed on the basis of data obtained well before and well after the
end pt.
Some typical titration curves for the reaction X + Y → Z where X the component to be
determined, Y titrant & Z the product (s) of the reaction.
If X absorbs radiation energy at a specified λ and it is contaminated with other
absorbing sub, a spectrophotometric titration can be carried out if a titrant can be found
w/c react selectively with X.
If X & the contaminant are the only species that absorb, the absorbance of the so/n will
decrease as titrant is added.
If Y is the only species in so/n w/c absorbs, the so/n will not absorb until the end pt is
reached.
If Z is the only species in so/n w/c absorbs, the absorbance will increase as product is
formed.
141
Spectrophotometric titra…
The following are some of the spectrophotometric titration curves which can
be
observed in the normal conditions:
Where:
S = Sample , P = titration Product , T = Titrant ,
1
@ = Absorbing species , X = Non absorbing species.
4
Spectrophotometric titra…
Examples:
1- Titration of potassium permanganate against ferrous sulfate
MnO4- + Fe++ + H+ == Mn++ + Fe3+ + H2O
@ at 540 (X) at 540 (X) at 540 (X) at 540 nm
1
Cu-EDTA complex is absorbing at 745 nm
4
Spectrophotometric titra…
Advantages:
1. More accurate results than direct photometric analysis are obtained.
2. No interference from other absorbing substances because the end
point depends on the change in the absorbance curve and not on the
absorbance value
(affect only the curve shape and sharpness of end point).
3. Can be used for titration of very dilute solutions.
4. Not need favorable equilibrium constants as those required for
titration that depends upon observations near the end point.
5. Can be used for (applied to) all types of reactions (redox, acid-base,
complxometry, pptmetry ….etc.)
144
Derivative spectrophotometry
It involve the conversion of a normal spectrum to its first, second or
the band.
145
Derivative…
The second derivative ( D2 ) spectrum is a plot of the curvature of the D0
146
147
Derivative…
Determination of conc. of an analyte in presence of interfering species or
other
analytes can sometimes be made more easily and/or more accurately.
Ex; A mixture of two substances A and B gave a zero order spectrum as indicated
below which show no well-defined absorption bands, but the second order spectrum
deduced from this curve showed a better peaks resolution.
A
A d2A/d2 B
1
4
1
5
Colorimetry
o Colorimetry is the determination of the light absorptive capacity of a
system.
o Quantitative determination carried out by subjecting a colored so/n to
same color.
o It could be direct or indirect method
151
Colorime…
Direct method
The drug must be self colored. Some pharmaceutical important subs w/c
Indirect method
The drug must react with a reagent to produce a sub, w/c is colored.
152
Colorime…
General requirement for the colored sub
The so/n should be intensely colored.
The so/n should be stable. If the absorbance value changes with time, the so/n
temp.
The so/n should obey beer’s law. To check for compliance with beer’s law,
153
Colorime…
The chemistry of Colorimetry
Colored sub are formed by reacting the constituent with either inorganic or
organic reagent.
Rxn may classified on the basis of the type of reagent used & on the type of
rxn, w/c occurs when the constituent is reacted with the reagent.
Colorimetric method w/c is based on the chemical derivatization with the
154
Colorime…
A) Diazotization & coupling of primary aromatic amines
The amine functional group is first diazotized with aqueous so/n of nitrous acid at 0-5 0C .
H+
Ar-NH2 + NHO2 Ar-N + N + 2H2O
The colorless daizonium salt is very reactive and when treated with a suitable coupling
The azo cpds all absorb light in the visible region, but the wavelength and absorptivity
depend on
the structure of the aromatic ring (Ar & Ar’),
pH and solvent.
155
Colorime…
The most widely used coupling reagents N-(1-naphtayl)-ethane-1-2-
amine before the coupling stage to destroy the excess nitrous acid w/c inhibits the
coupling rxn.
The sensitivity & selectivity of the procedure permit the assay of low conc. of
impurities that contain a primary aromatic amine group in the presence of the
parent sub lacking amine function.
156
Colorime…
The BP (1988) employ this method for determination of free amine
impurities in
o Frusemide,
o Iothalemic acid,
o Meglutamine Iothamic inj.
o Meglumine diatrazoate inj. &
o the BP use this method in the determination of Folic acid
and its tablet.
157
Colorime…
B) Reaction of the α-ketal group with tetrazoluim salt
In the presence of a steroid α-ketal (21-hydroyl-20-keton) side chain
hydroxide) at 30-35 C for 1-2hrs and the absorbance of the red product is
measured around 488nm
The oxidation of the α-ketalgroup and the reduction of
158
Colorime…
Interference by reducing sugars is eliminated by alcohol extraction
CH2OH CH=O
C=O C=O N N
N N
OH OH + C H
+ C
N N N N
159
Applications for Spectrophotometry
Applications of spectrophotometric methods are so numerous and touch
every field
in which quantitative chemical information are required.
In general, about 90% of all the quantitative determinations are performed
by
spectroscopic techniques.
In the field of health alone , 95 % of all quantitative determinations are
performed by
UV-Visible spectrophotometry and similar techniques.( Over 3,000,000
daily tests
are carried out in USA only).
1
6
Spectrophotometric applications of pharmaceutical
interest
162
Clofazimine , Absorbed at 283 & 487 nm
163
b) Non-absorbing species :
Examples: Erythromycin
R'
NH.NH2
NH.N C R
NO 2 R' NO 2
+ R C O + H2O
NO 2
NO 2
Colourle s s & ha s coloure d product
no a bs orba nce Ora nge ( 435 nm)
164
To increase sensitivity and/or selectivity e.g. sulfa drugs and salicylic acid.
H2N SO2.NHR
OH OH
NH2 N N . HCl
R' R'
NaNO2
R R
HCl N N
R
Colourless & has Coloured Product
Absorbance O ( 480 nm)
COOH O
OH C Cu
Cu ++ O
coenzymes.
Example : Determination of ethanol content in blood via measuring
ADH
activity
CH3 ofCH2 OH + NAD+ CH3 CHO + NADH + H+
NADH
NAD+
260 nm 340 nm
169
Other Important…
(E) Monitoring drug degradation kinetics :
Simply done when the product has a different absorption spectrum than that of
un-degraded drug.
acid spectrum (as a function of time ) may be used to determine the rate
constant
170 for aspirin hydrolysis.
Other Important…
They are the most widely used detectors, because: Most drugs absorb UV-Visible radiation.
More sensitive and more selective than the bulk property detectors (e.g. R.I. detectors).
Some absorbance detectors have one or two fixed wavelengths (280 and/or 254 nm).
More modern HPLC instruments have variable wavelength detectors using the photodiodes.
171
Other Important…
(G) Determination of Equilibrium Constants :
Acid dissociation constants and metal ion-ligand stability constants can be determined
spectrophotometrically if the species involved have absorptivities which differ from one another.
Example : Determination of the pKa of Methyl red indicator ; Acidic (HMR) and basic (MR-) forms of
CO2- CO2-
+ HO-
(CH3)2N N NH (CH3)2N N=N
H+
Acid form, pH = 4, (HMR) Red, 520 nm Basic form, pH = 6, (MR-) Yellow 430 nm
172
Other Important…
174