Professional Documents
Culture Documents
I. Introduction
A. Spectroscopy is the study of the interaction of matter with the
electromagnetic spectrum
E = hn
I. Introduction
5. Because the speed of light, c, is constant, the frequency, n, (number of
cycles of the wave per second) can complete in the same time, must be
inversely proportional to how long the oscillation is, or wavelength:
c
___ hc
n= ___
E = hn =
l l
c = 3 x 1010 cm/s
7. Because the atomic particles in matter also exhibit wave and particle
properties (though opposite in how much) EM radiation can interact with
matter in two ways:
• Collision – particle-to-particle – energy is lost as heat and movement
I. Introduction
8. The entire electromagnetic spectrum is used by chemists:
Frequency, n in Hz
~1019 ~1017 ~1015 ~1013 ~1010 ~105
Wavelength, l
~.0001 nm ~0.01 nm 10 nm 1000 nm 0.01 cm 100 m
Energy (kcal/mol)
> 300 300-30 300-30 ~10-4 ~10-6
Visible
IR Spectroscopy
I. Introduction
C. The IR Spectroscopic Process
1. The quantum mechanical energy levels observed in IR spectroscopy are
those of molecular vibration
I. Introduction
C. The IR Spectroscopic Process
5. There are two types of bond vibration:
• Stretch – Vibration or oscillation along the line of the bond
H H
C C
H H
symmetric asymmetric
“coupled” wave
EM oscillating wave
from bond vibration
Infrared Spectroscopy
D. The IR Spectrum
1. Each stretching and bending vibration occurs with a
characteristic frequency as the atoms and charges
involved are different for different bonds
The y-axis on an IR
spectrum is in units of % In regions
transmittance where no osc.
bond is
In regions where the EM interacting with
field of an osc. bond IR light,
interacts with IR light of transmittance
the same n – nears 100%
transmittance is low (light
is absorbed)
IR Spectroscopy
D. The IR Spectrum
2. The x-axis of the IR spectrum is in units of wavenumbers, n, which is
the number of waves per centimeter in units of cm-1 (Remember E = hn
or E = hc/l)
IR Spectroscopy
D. The IR Spectrum
3. This unit is used rather than wavelength (microns) because
wavenumbers are directly proportional to the energy of transition
being observed –chemists like this, physicists hate it
4. This unit is used rather than frequency as the numbers are more
“real” than the exponential units of frequency
D. The IR Spectrum
7. In general:
Lighter atoms will allow the oscillation to be faster – higher energy
This is especially true of bonds to hydrogen – C-H, N-H and O-H
Energy/n of oscillation
Infrared Spectroscopy
Fingerprint
Region
4000 cm-1 2700 cm-1 2000 cm-1 1600 cm-1 400 cm-1
Vibrations
What is a vibration in a molecule?
Any change in shape of the molecule- stretching of
bonds, bending of bonds, or internal rotation
around single bonds
H2 I2
MM =2 g/mole
MM =254 g/mole
bending vibration
For C-C-C bond angle a
10 pm
change of 4o is typical.
4o
This moves a carbon atom
about 10 pm.
A little physics of electromagnetic radiation
Wavenumber (n’) n’ = 1/ l
◦ given in cm-1
Increasing
absorption of IR
radiation
C-C
bending
Increasing wavelength
Some results
Does the stretching energy have
any relationship to the strength of
the bond?
5000 Wavenumber vs. Bond Energy
Wavenumber (cm )
-1
4000
3000
2000
W = 6.3286BE + 401.38
1000
r2 = 0.7979
0
200 300 400 500 600
Bond energy (kJ/mole)
Let’s examine the carbonyl group on three
compounds
formaldehyde phosgene acetone
O-H
C=O C-O
alkenes
stretching
O-H aromatic
WAVENUMBER (cm-1)
Identification of functional groups on a
molecule – this is a very important tool in
organic chemistry
Use of IR spectra
FTIR Theory
The spectrometer described here is a modified
Bomem MB-100 FTIR.
The heart of the FTIR is a Michelson
interferometer (figure 2).
The mirror moves at a fixed rate. Its position is
determined accurately by counting the
interference fringes of a collocated Helium-Neon
laser.
The Michelson interferometer splits a beam of
radiation into two paths having different lengths,
and then recombines them.
A detector measures the intensity variations of
the exit beam as a function of path difference.
A monochromatic source would show a simple
sine wave of intensity at the detector due to
constructive and destructive interference as the
In the general case, a superposition of
wavelengths enter spectrometer, and the detector
indicates the sum of the sine waves added
together.
Figure 3 shows some idealized light sources, and
the interferograms that they would theoretically
produce.
The difference in path length for the radiation is
known as the retardation d (OM = OF + d) in
figure 1 and 2.
When the retardation is zero, the detector sees a
maximum because all wavenumbers of radiation
add constructively.
When the retardation is l/2, the detector sees a
minimum for the wavelength l. An interferogram
is the sum of all of the wavenumber intensities.
Figure 1.
Figure 2.