Professional Documents
Culture Documents
Saqlain Tahir
Potentiometry Head of Quality Assurance
Highnoon Laboratories Limited
Electrochemistry
The relationship between chemical reactions and electricity.
• Certain chemical reactions can create electricity.
• Electricity can make certain chemical reactions happen that wouldn’t
happen otherwise.
Electrochemical Cell
It is a device for the conversion of electrical energy into chemical
energy and vice versa. It consists of two electrodes dipping into an ionic
solution and connected by an external metallic conductor.
Types
• Voltaic or Galvanic Cell
• Electrolytic Cell
Electrochemical Cells
A galvanic cell is a device where the chemical energy is converted into
electrical energy, such as dry cell or lead storage battery. In an
electrolytic cell, the electrical energy from an external source is used to
bring about a physical or a chemical change. Charging of a lead storage
battery or electrolytic purification of metals are examples where
electrical energy brings about chemical changes.
Electrode Potential
The potential difference that exists across the metal-solution interface is
known as electrode potential. If the ions in the solution are at unit
activity, the electrode potential is termed as the standard electrode
potential.
The Nernst temperature compensation corrects only for the electrode millivolt output change with temperature, not the actual
pH change of the buffer solution with temperature which is unique for each buffer.]
Potentiometry
Potentiometry is based on the measurement of the potential of an electrode
system (e.g. electrochemical cell). Potentiometric measurement system consists of
two electrodes called reference and indicator electrode, potentiometer and a
solution of analyte (figure 2).
Actions
Calibrate at least once per day, more if high temperature fluctuations
Make sure conditions at calibration and measurement (temperature, stirring etc.) are equal
Always use fresh buffers – if buffers not accurate, pH calibration is not accurate and the
measurement will not be accurate
Perform at least a two point calibration
Make sure the calibration points frame the expected sample pH
Electrode Calibration
Significance of pH Measurement
• Large number of qualitative and quantitative analyses are carried out
a definite pH values .
• pH has a great importance in agriculture. Soils are often tested to
determine whether acidic or basic fertilizers are required for a
particular crop.
• Food preservation needs definite pH.
• It has also got a great importance in biochemical reactions such as
digestion of food, working of enzymes etc.
Titration Errors
Potentiometric Titration Curve
Potentiometric Titrator
Electrode calibration – buffer handling
• Buffers have expiry date – don’t order in bulk
• Store well sealed at room temperature
• Take out needed amount and close bottle immediately again
• Never calibrate electrode directly in the bottle
• Never re-use already used buffer solution
• Take single-use buffer sachets
→ Buffers are “testing equipment” !
Electrode calibration – indicating electrode
condition
• “Offset” value (mV) – indicates the age of electrode and provides an estimation
when the electrode need to be changed.
• Calibration Slope (%) – indicates the sensitivity of the glass membrane
Recommended offset range at pH 7.00 is ± 30mV.
Recommended calibration slope range is 95% - 105%.
(DIN 19263 requirements: Offset 0 ± 30 mV; Zero Point: pH0 = 7 ± 0.5 pH)
In brief
If electrode is not performing well
Membrane dehydrated, contaminated or damaged
Electrolyte contaminated, has crystallized, hasn’t been filled enough or is missing,
has too low ion concentration
Air bubbles behind the ceramic junction
Diaphragm (junction) contaminated or blocked
Electrode worn out (out of specs), is too old
Simply the wrong way
Filling hole closed during measurement
Storage in deionized water
Wipe to clean glass membrane
Calibration once a week
Temperature has no influence
Never replace electrolyte
Re-use buffers for calibration
Stir extremely fast during measurement and …