Professional Documents
Culture Documents
Advantage Disadvantage
• Simple • require gas supplies, appropriate analysers
• Requires a long period of rebreathing, which
• Easy can be demanding for patients.
• Requires less effort to perform • The most important limitation is the fact that
tracer gases (usually helium and nitrogen)
• Acceptable method for may not reach the more poorly ventilated
regions of the lung in patients with more
measuring lung volumes severe chronic airway obstruction, and this
• Useful in detection of airtrapping results in underestimation of lung volumes.
So it can underestimate FRC in obstructive
• Requires only tidal breathing so lung disease so can be interpreted as mixed
pattern (obstructive and restrictive) as not
minimal patient effort include areas with airtrapping
• Instrumentation is simple and • Commercial systems frequently limit
inexpensive equilibration time to no more than 4 to 5
minutes.
Helium dilution method
principle
• “Equilibriation of gas in the lung with a
known value of gas containing helium”
• If a gas with known He concentration is
breathed in, the He will be diluted by the
He-free gas within the lungs
• If the expired He concentration is monitored
the volume of gas within the lungs can then
be calculated from the dilution effect.
• Once the He reaches equilibrium between
the spirometer and the patient, the final
concentration of He is recorded
• The FRC can then be calculated
• Also should be corrected to BTPS conditions
Closed circuit helium dilution test procedure
• Patient preparation
• Checked for a perforated eardrum (if so, an earplug should be used)
• Seated comfortably, with no need to remove dentures
• Procedure is explained, emphasising the need to avoid leaks around the
mouthpiece during the test and to use a nose clip
• Test
• Circuit is flushed with air
• Oxygen is subsequently added to raise the final oxygen concentration to about 25-
30% • Helium meter reading adjusted to zero, helium is added to raise the helium
concentration to nearly full scale deflection (10%) on the analyser
• The patient breathes for 30-60 s on the mouthpiece to become accustomed to the
apparatus, and to ensure a stable end-tidal expiratory level
• The patient is turned ‘‘in’’ (i.e. connected to the test gas) at the end of a normal
tidal expiration and instructed to breathe regular tidal breaths
• Once the helium equilibration is complete, the patient is turned ‘‘out’’ (i.e.
disconnected from the test gas) of the system
Closed circuit helium dilution
test procedure
As the patient rebreathes from the closed circuit, the blower circulates the gas mixture.
The CO2 is absorbed by soda lime (CO2 absorber), while O2 is added through a valve and
flowmeter at a rate corresponding to the subject’s O2 consumption. As the helium, which
was at first contained entirely within the apparatus, mixes with air contained in the lungs,
its concentration, as monitored by the helium analyzer, falls. Stabilization of the helium
concentration, indicated by a rate of change in concentration of less than 0.02% over a 30-
second interval, signals the point at which the helium concentration has equilibrated
throughout the lung-breathing circuit system; equilibration, the end-point of the test,
occurs within 7 minutes in normal persons. However, in patients in whom the distribution
of ventilation is abnormal – for example, those with chronic obstructive pulmonary
disease (COPD)– equilibration may take much longer. Upon equilibration, the following
equation, based on the law of conservation of mass, is applied :
Closed circuit helium dilution
test procedure….continue
The initial volume of the system is
the volume of the spirometer and
circuit tubing, whereas the final
volume consists of the initial
volume plus FRC. The latter value is
the only unknown in the preceding
equation. Corrections are usually
made for the small amount of
helium dissolved in body tissues
during the test
Closed-Circuit Helium Dilution test
• Closing Volume
• Closing Capacity
• Deadspace
Single-Breath Nitrogen Washout (SBN2)
Procedure
• Patient exhales to RV
• Inspires a VC breath of 100% O2
• Patient exhales slowly and evenly
(0.3-0.5L/s) until feels lung is
empty
• The exhaled gas passes through
N2 analyzer that measures the
change in concentration of
nitrogen
• N2 concentration is plotted
against volume
Single-Breath Nitrogen Washout
(SBN2)
• Phase I: upper airway gas from
anatomical dead space (VDanat),
consisting of 100% O2
• Phase II: mixed airway gas in which the
relative concentrations of O2 and N2
change abrubtly as VDanat volume is
expired
• Phase III: a plateau caused by the
exhalation of alveolar gas in which
relative O2 and N2 concentrations
change slowly and evenly
% N2 750 – 1250
• Acceptability Criteria
Advantages Disadvantages
• The most widely used is • washout takes a
the N2 SBW test long time and,
• assess ventilation • it does not estimate
distribution inhomogeneity the volume of poorly
at differing lung volumes ventilated regions of
• Simple the lung.
• Inexpensive
• Less patient cooperation
Example of a typical single-breath washout (SBW) trace. Nitrogen gas (N 2) expirogram
showing calculation of phase III slope (SIII) in a vital capacity SBW test in a 60-yr-old
smoker. SIII is calculated between 25% and 75% of the expired volume (S III 4.4%·L−1),. The
four phases of the expirogram are also demonstrated: phase I (absolute dead space), phase
II (bronchial phase), phase III (alveolar phase) and phase IV (fast rising phase at end of
expiration). Closing volume (CV) is the expired volume (L) from the start of the upward
deflection where phase IV starts, to the end of the breath. If residual volume (RV) is
known, closing capacity (CC) can be calculated: CC = CV+RV.