Professional Documents
Culture Documents
VIDAS BLUE
PRESENTATION
VIDAS BLUE
VIDAS BLUE:
Removing
strip preparation tray
Mini VIDAS BLUE
Mini VIDAS BLUE
New screen /
keyboard Holder
Internal changes
VIDAS BLUE
On mini VIDAS:
Addition of a long screw to fix the right
trim
VIDAS BLUE
On mini VIDAS:
Screen/keyboard holder 3 fixing points:
VIDAS BLUE
VIDAS
I DAS
Mini V
Be carefull
Respect the correct connector!
VIDAS BLUE
New Power Supply Board:
-Air Fan
•Fixed by 4 plastic retainers
•Spare part
•If Air Fan out of service:
Software T° error will be
generated
Automatic Power Supply
cut off
VIDAS BLUE
i n g
a r l
New tray ball bearings: be
m a
al l o r ng
c b a n ri
t ri by a sp
en d n
xc ce o
E pla ed .
x
re e fi nis m
on cha
me
Tower screw:
Addition of a spacer
Incubator door:
- New door plunger ball nut
- Addition of a safety latch
VIDAS BLUE
Incubator Boards
Scanner Board
VIDAS BLUE
New Scanner Board: New Power Supply 3.3V
LED
Flash ROM U5 ( V2.51)
(replace the EEPROM) Sensors LEDs
Sensor LEDs
VIDAS BLUE
SCANNER:
SCANNER new parts
• New mini VIDAS/ VIDAS scanner cables
SCANNER:
SCANNER new parts
• New Scanner PRE-AMP board
The LEDs (DS9 to DS12 from the new scanner board) are on
When a cable is broken.
( off = cables OK)
The new Scanner head is compatible with the old Scanner head
But don’t forget to replace the new scanner cable by an old
scanner cable kit in order to be compatible with the old scanner
board
VIDAS BLUE
Changes impact
Decontamination: a new procedure has been defined to take
Into account the « Strip preparation tray ».
E-mail : gcsimmunomarcy@eu.biomerieux.com