MICROLAB 200®
- It´s recommended to read the Instructions before TEST NAME HDL CHOL
using this product. TYPE FINAL POINT
FILTER 505
Some complementary information for the realization of TEMPERATURE 25
the test are described joint to the Automation Protocol UNIT mg/dL
below: REAG BLANK SI
ABS - LOW 0.000
REAGENT PREPARATION ABS - HIGH 2.300
Reagents ready to use. CALIB - NAME
STANDARD
CALIBRATOR CONTROLS
To evaluate the performance of ............- NR 1
Autocal H the test it´s sugested the use of ............- CONC @
CAT BT the BioTécnica Controls: ............- REP 1
13.002.00 ............- ABS LOW 0.000
- Quantinorm CAT BT ............- ABS HIGH 2.300
13.003.00 FACTOR 1
SAMPLE BLANK No
- Quantialt CAT BT 13.004.00 RETARD 5
SPECIAL OBSERVATIONS READING TIME 2
None REACTION VOL 500
REPLICATES 1
REF. LOW *
# Parameter defined by the user. REF. HIGH *
ABS LOW 0.000
@ Insert the concentration of the calibrator ABS HIGH 2.300
according to the indication in the label of Autocal H LOT *
- CAT BT 13.002.00 TARGET VALUE *
LOW LIMIT *
HIGH LIMIT *