Equipment Applications

MICROLAB 200®

® TRADEMARKS REVISION 00 - APR/2006


CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil.
Site: www.biotecnica.ind.br / e-mail: sac@biotecnicaltda.com.br
Equipment Applications
MICROLAB 200®
PRODUCT ALBUMIN
CATÁLOG BT 10.002.00
METODOLOGY COLORIMETRIC
DETERMINATIONS 500
TEST NAME ALBUMIN
- The number of determinations was calculated NOT TYPE FINAL POINT
taking into account the “dead volume” of the equipment FILTER 620
flask. TEMPERATURE 25
UNIT g/dL
- It´s recommended to read the Instructions before REAG BLANK SI
using this product. ABS - LOW 0.000
ABS - HIGH 2.300
Some complementary information for the realization of CALIB - NAME
the test are described joint to the Automation Protocol STANDARD
below: ............- NR 1
............- CONC @
REAGENT PREPARATION ............- REP 1
Reagents ready to use. ............- ABS LOW 0.000
CALIBRATOR CONTROLS ............- ABS HIGH 2.300
To evaluate the performance of FACTOR 1
Autocal H the test it´s sugested the use of SAMPLE BLANK No
CAT BT the BioTécnica Controls: RETARD 5
13.002.00 READING TIME 2
- Quantinorm CAT BT REACTION VOL 500
13.003.00 REPLICATES 1
REF. LOW *
- Quantialt CAT BT 13.004.00 REF. HIGH *
SPECIAL OBSERVATIONS ABS LOW 0.000
None ABS HIGH 2.300
.
CONTROL - 1
# Parameter defined by the user. LOT *
TARGET VALUE *
@ Insert the concentration of the calibrator LOW LIMIT *
according to the indication in the label of Autocal H
HIGH LIMIT *
- CAT BT 13.002.00

® TRADEMARKS REVISION 00 - APR/2006


CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil.
Site: www.biotecnica.ind.br / e-mail: sac@biotecnicaltda.com.br
Equipment Applications
MICROLAB 200®
PRODUCT ALPHA-AMYLASE
CATÁLOG BT 11.001.00
METODOLOGY KINETIC COLORIMETRIC
DETERMINATIONS 60

TEST NAME AMY


- The number of determinations was calculated NOT TYPE KINETIC
taking into account the “dead volume” of the equipment FILTER 405
flask. TEMPERATURE 37
UNIT U/L
- It´s recommended to read the Instructions before REAG BLANK NO
using this product.
CALIB NAME NO
Some complementary information for the realization of FACTOR 6950
the test are described joint to the Automation Protocol
SAMPLE BLANK NO
below:
RETARD 5
READING TIME 30
REAGENT PREPARATION REACTION VOL. 500
Reagents ready to use. REPET. TEST 1
CALIBRATOR CONTROLS REF.LOW *
To evaluate the performance of REF. HIGH *
Autocal H the test it´s sugested the use of ABS LOW 0.800
CAT BT the BioTécnica Controls: ABS HIGH 2.300
13.002.00 .
- Quantinorm CAT BT CONTROL - 1
13.003.00 LOT *
TARGET VALUE *
- Quantialt CAT BT 13.004.00
LOW LIMIT *
SPECIAL OBSERVATIONS
HIGH LIMIT *
None

# Parameter defined by the user.

@ Insert the concentration of the calibrator


according to the indication in the label of Autocal H
- CAT BT 13.002.00

® TRADEMARKS REVISION 00 - APR/2006


CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil.
Site: www.biotecnica.ind.br / e-mail: sac@biotecnicaltda.com.br
Equipment Applications
MICROLAB 200®
PRODUCT ALKALINE
PHOSPHATASE
CATÁLOG BT 11.005.00
METODOLOGY KINETIC COLORIMETRIC
DETERMINATIONS 100

- The number of determinations was calculated NOT TEST NAME ALP


taking into account the “dead volume” of the equipment TYPE KINETIC
flask. FILTER 405
TEMPERATURE 37
- It´s recommended to read the Instructions before UNIT U/L
using this product. REAG BLANK NO
Some complementary information for the realization of CALIB NAME NO
the test are described joint to the Automation Protocol FACTOR 2757
below:
SAMPLE BLANK NO
RETARD 60
REAGENT PREPARATION READING TIME 60
Mix 4 parts of Reagent A with 1 part of Reagent B REACTION VOL. 500
and homogenize it gently. REPET. TEST 1
CALIBRATOR CONTROLS REF.LOW *
To evaluate the performance of REF. HIGH *
Autocal H the test it´s sugested the use of ABS LOW 0.000
CAT BT the BioTécnica Controls: ABS HIGH 2.300
13.002.00 LOT *
- Quantinorm CAT BT TARGET VALUE *
13.003.00 LOW LIMIT *
HIGH LIMIT *
- Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS
None

# Parameter defined by the user.

@ Insert the concentration of the calibrator


according to the indication in the label of Autocal H
- CAT BT 13.002.00

® TRADEMARKS REVISION 00 - APR/2006


CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil.
Site: www.biotecnica.ind.br / e-mail: sac@biotecnicaltda.com.br
Equipment Applications
MICROLAB 200®
PRODUCT ALT/TGP
CATÁLOG BT 11.008.00
METODOLOGY KINETIC UV
DETERMINATIONS 500

- The number of determinations was calculated NOT TEST NAME ALT/TGP


taking into account the “dead volume” of the equipment TYPE KINETIC
flask. FILTER 340
TEMPERATURE 37
- It´s recommended to read the Instructions before UNIT U/L
using this product. REAG BLANK NO
Some complementary information for the realization of CALIB NAME NO
the test are described joint to the Automation Protocol FACTOR 1746
below:
SAMPLE BLANK NO
RETARD 5
REAGENT PREPARATION READING TIME 30
Mix 4 parts of Reagent A with 1 part of Reagent B REACTION VOL. 500
and homogenize it gently. REPET. TEST 1
CALIBRATOR CONTROLS REF.LOW *
To evaluate the performance of REF. HIGH *
Autocal H the test it´s sugested the use of ABS LOW 0.800
CAT BT the BioTécnica Controls: ABS HIGH 2.300
13.002.00 LOT *
- Quantinorm CAT BT TARGET VALUE *
13.003.00 LOW LIMIT *
HIGH LIMIT *
- Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS
None

# Parameter defined by the user.

@ Insert the concentration of the calibrator


according to the indication in the label of Autocal H
- CAT BT 13.002.00

® TRADEMARKS REVISION 00 - APR/2006


CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil.
Site: www.biotecnica.ind.br / e-mail: sac@biotecnicaltda.com.br
Equipment Applications
MICROLAB 200®
PRODUCT AST/TGO
CATÁLOG BT 11.007.00
METODOLOGY KINETIC UV
DETERMINATIONS 500

TEST NAME AST/TGO


- The number of determinations was calculated NOT TYPE KINETIC
taking into account the “dead volume” of the equipment FILTER 340
flask. TEMPERATURE 37
UNIT U/L
- It´s recommended to read the Instructions before REAG BLANK NO
using this product.
CALIB NAME NO
Some complementary information for the realization of FACTOR 1746
the test are described joint to the Automation Protocol
SAMPLE BLANK NO
below:
RETARD 5
READING TIME 30
REAGENT PREPARATION REACTION VOL. 500
Mix 4 parts of Reagent A with 1 part of Reagent B REPET. TEST 1
and homogenize it gently. REF.LOW *
CALIBRATOR CONTROLS REF. HIGH *
To evaluate the performance of ABS LOW 0.800
Autocal H the test it´s sugested the use of ABS HIGH 2.300
CAT BT the BioTécnica Controls: LOT *
13.002.00 TARGET VALUE *
- Quantinorm CAT BT LOW LIMIT *
13.003.00 HIGH LIMIT *
- Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS
None

# Parameter defined by the user.

@ Insert the concentration of the calibrator


according to the indication in the label of Autocal H
- CAT BT 13.002.00

® TRADEMARKS REVISION 00 - APR/2006


CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil.
Site: www.biotecnica.ind.br / e-mail: sac@biotecnicaltda.com.br
Equipment Applications
MICROLAB 200®
PRODUCT BILIRUBIN DIRECT
CATÁLOG BT 10.003.00
METODOLOGY COLORIMETRIC
DETERMINATIONS 250 TEST NAME BDI
TYPE FINAL POINT
FILTER 546
- The number of determinations was calculated NOT TEMPERATURE 25
taking into account the “dead volume” of the equipment UNIT mg/dL
flask. REAG BLANK NO
ABS - LOW 0.000
- It´s recommended to read the Instructions before
ABS - HIGH 2.300
using this product.
CALIB - NAME
STANDARD
Some complementary information for the realization of
............- NR 1
the test are described joint to the Automation Protocol
below: ............- CONC @
............- REP 1
............- ABS LOW 0.220
REAGENT PREPARATION ............- ABS HIGH 0.330
Reagents ready to use. FACTOR 1.0
CALIBRATOR CONTROLS SAMPLE BLANK SI
To evaluate the performance of RETARD 5
Autocal H the test it´s sugested the use of READING TIME 2
CAT BT the BioTécnica Controls: REACTION VOL 900
13.002.00 REPLICATES 1
- Quantinorm CAT BT REF. LOW *
13.003.00 REF. HIGH *
ABS LOW 0.000
- Quantialt CAT BT 13.004.00
ABS HIGH 2.300
SPECIAL OBSERVATIONS
LOT *
None
TARGET VALUE *
LOW LIMIT *
HIGH LIMIT *
# Parameter defined by the user.

@ Insert the concentration of the calibrator


according to the indication in the label of Autocal H
- CAT BT 13.002.00

® TRADEMARKS REVISION 00 - APR/2006


CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil.
Site: www.biotecnica.ind.br / e-mail: sac@biotecnicaltda.com.br
Equipment Applications
MICROLAB 200®
PRODUCT BILIRUBIN TOTAL
CATÁLOG BT 10.003.00
METODOLOGY COLORIMETRIC
DETERMINATIONS 250

- The number of determinations was calculated NOT


taking into account the “dead volume” of the equipment TEST NAME BTO
flask. TYPE FINAL POINT
FILTER 546
- It´s recommended to read the Instructions before
TEMPERATURE 25
using this product.
UNIT mg/dL
Some complementary information for the realization of REAG BLANK NO
the test are described joint to the Automation Protocol ABS - LOW 0.000
below: ABS - HIGH 2.300
CALIB - NAME
STANDARD
REAGENT PREPARATION ............- NR 1
Reagents ready to use. ............- CONC @
CALIBRATOR CONTROLS ............- REP 1
To evaluate the performance of ............- ABS LOW 0.220
Autocal H the test it´s sugested the use of ............- ABS HIGH 0.330
CAT BT the BioTécnica Controls: FACTOR 1.0
13.002.00 SAMPLE BLANK SI
- Quantinorm CAT BT RETARD 5
13.003.00 READING TIME 2
REACTION VOL 900
- Quantialt CAT BT 13.004.00
REPLICATES 1
SPECIAL OBSERVATIONS
REF. LOW *
None
REF. HIGH *
ABS LOW 0.000
ABS HIGH 2.300
# Parameter defined by the user.
LOT *
TARGET VALUE *
@ Insert the concentration of the calibrator
according to the indication in the label of Autocal H LOW LIMIT *
- CAT BT 13.002.00 HIGH LIMIT *

® TRADEMARKS REVISION 00 - APR/2006


CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil.
Site: www.biotecnica.ind.br / e-mail: sac@biotecnicaltda.com.br
Equipment Applications
MICROLAB 200®
PRODUCT CÁLCIUM ASX
CATÁLOG BT 12.002.00
METODOLOGY ARSENAZO III
DETERMINATIONS 200

- The number of determinations was calculated NOT TEST NAME CALCIUM


taking into account the “dead volume” of the equipment TYPE FINAL POINT
flask. FILTER 620
TEMPERATURE 25
- It´s recommended to read the Instructions before UNIT mg/dL
using this product. REAG BLANK Si
ABS - LOW 0.000
Some complementary information for the realization of
ABS - HIGH 2.300
the test are described joint to the Automation Protocol
below: CALIB - NAME
STANDARD
............- NR 1
REAGENT PREPARATION ............- CONC @
Reagents ready to use. ............- REP 1
CALIBRATOR CONTROLS ............- ABS LOW 0.220
To evaluate the performance of ............- ABS HIGH 0.330
Autocal H the test it´s sugested the use of FACTOR 1.0
CAT BT the BioTécnica Controls: SAMPLE BLANK NO
13.002.00 RETARD 5
- Quantinorm CAT BT READING TIME 2
13.003.00 REACTION VOL 900
REPLICATES 1
- Quantialt CAT BT 13.004.00 REF. LOW *
SPECIAL OBSERVATIONS REF. HIGH *
None ABS LOW 0.000
ABS HIGH 2.300
LOT *
# Parameter defined by the user. TARGET VALUE *
LOW LIMIT *
@ Insert the concentration of the calibrator
HIGH LIMIT *
according to the indication in the label of Autocal H
- CAT BT 13.002.00

® TRADEMARKS REVISION 00 - APR/2006


CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil.
Site: www.biotecnica.ind.br / e-mail: sac@biotecnicaltda.com.br
Equipment Applications
MICROLAB 200®
PRODUCT CÁLCIUM
CATÁLOG BT 12.001.00
METODOLOGY CPC
DETERMINATIONS 200

- The number of determinations was calculated NOT TEST NAME CALCIUM


taking into account the “dead volume” of the equipment TYPE FINAL POINT
flask. FILTER 578
TEMPERATURE 25
- It´s recommended to read the Instructions before UNIT mg/dL
using this product. REAG BLANK Si
ABS - LOW
Some complementary information for the realization of STANDARD
the test are described joint to the Automation Protocol ABS - HIGH 1
below:
CALIB - NAME @
............- NR 1
REAGENT PREPARATION ............- CONC 0.000
Mix the Reagent A + Reagent B in a proportion 1:1 ............- REP 2.300
and homogenize it gently. ............- ABS LOW 1.00
CALIBRATOR CONTROLS ............- ABS HIGH No
To evaluate the performance of FACTOR 10
Autocal H the test it´s sugested the use of SAMPLE BLANK 2
CAT BT the BioTécnica Controls: RETARD 500
13.002.00 READING TIME 1
- Quantinorm CAT BT REACTION VOL *
13.003.00 REPLICATES *
REF. LOW 0.000
- Quantialt CAT BT 13.004.00 REF. HIGH 2.300
SPECIAL OBSERVATIONS . *
None CONTROL - 1 *
LOT *
TARGET VALUE *
# Parameter defined by the user. LOW LIMIT
HIGH LIMIT
@ Insert the concentration of the calibrator
according to the indication in the label of Autocal H
- CAT BT 13.002.00

® TRADEMARKS REVISION 00 - APR/2006


CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil.
Site: www.biotecnica.ind.br / e-mail: sac@biotecnicaltda.com.br
Equipment Applications
MICROLAB 200®
PRODUCT CK NAC
CATÁLOG BT 11.002.00
METODOLOGY KINETIC UV
DETERMINATIONS 40

- The number of determinations was calculated NOT


taking into account the “dead volume” of the equipment
flask.
TEST NAME CK NAC
- It´s recommended to read the Instructions before TYPE KINETIC
using this product. FILTER 340
TEMPERATURE 37
Some complementary information for the realization of UNIT U/L
the test are described joint to the Automation Protocol REAG BLANK NO
below:
CALIB NAME NO

REAGENT PREPARATION FACTOR 4127


Mix 4 parts of Reagent A with 1 part of Reagent B SAMPLE BLANK NO
and homogenize it gently. RETARD 5
CALIBRATOR CONTROLS READING TIME 30
To evaluate the performance of REACTION VOL. 500
Autocal H the test it´s sugested the use of REPET. TEST 1
CAT BT the BioTécnica Controls: REF.LOW *
13.002.00 REF. HIGH *
- Quantinorm CAT BT ABS LOW 0.000
13.003.00 ABS HIGH 2.300
LOT *
- Quantialt CAT BT 13.004.00 TARGET VALUE *
SPECIAL OBSERVATIONS LOW LIMIT *
None HIGH LIMIT *

# Parameter defined by the user.

@ Insert the concentration of the calibrator


according to the indication in the label of Autocal H
- CAT BT 13.002.00

® TRADEMARKS REVISION 00 - APR/2006


CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil.
Site: www.biotecnica.ind.br / e-mail: sac@biotecnicaltda.com.br
Equipment Applications
MICROLAB 200®
PRODUCT CK MB
CATÁLOG BT 11.003.00
METODOLOGY KINETIC UV
DETERMINATIONS 40

- The number of determinations was calculated NOT TEST NAME CK MB


taking into account the “dead volume” of the equipment TYPE KINETIC
flask. FILTER 340
TEMPERATURE 37
- It´s recommended to read the Instructions before UNIT U/L
using this product. REAG BLANK NO
Some complementary information for the realization of CALIB NAME NO
the test are described joint to the Automation Protocol FACTOR 8254
below:
SAMPLE BLANK NO
RETARD 5
REAGENT PREPARATION READING TIME 120
Mix 4 parts of Reagent A with 1 part of Reagent B REACTION VOL. 500
and homogenize it gently. REPET. TEST 1
CALIBRATOR CONTROLS REF.LOW *
To evaluate the performance of REF. HIGH *
Autocal H the test it´s sugested the use of ABS LOW 0.000
CAT BT the BioTécnica Controls: ABS HIGH 2.300
13.002.00 LOT *
- Quantinorm CAT BT TARGET VALUE *
13.003.00 LOW LIMIT *
HIGH LIMIT *
- Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS
None

# Parameter defined by the user.

@ Insert the concentration of the calibrator


according to the indication in the label of Autocal H
- CAT BT 13.002.00

® TRADEMARKS REVISION 00 - APR/2006


CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil.
Site: www.biotecnica.ind.br / e-mail: sac@biotecnicaltda.com.br
Equipment Applications
MICROLAB 200®
PRODUCT CHLORIDE
CATÁLOG BT 12.003.00
METODOLOGY COLORIMÉTRIC
DETERMINATIONS 100
TEST NAME CHLORIDE
- The number of determinations was calculated NOT TYPE FINAL POINT
taking into account the “dead volume” of the equipment FILTER 505
flask. TEMPERATURE 25
UNIT mEq/L
- It´s recommended to read the Instructions before REAG BLANK Si
using this product. ABS - LOW 0.000
ABS - HIGH 2.300
Some complementary information for the realization of CALIB - NAME
the test are described joint to the Automation Protocol STANDARD
below: ............- NR 1
............- CONC @
REAGENT PREPARATION ............- REP 1
Reagents ready to use. ............- ABS LOW 0.000
CALIBRATOR CONTROLS ............- ABS HIGH 2.300
To evaluate the performance of FACTOR 1.00
Autocal H the test it´s sugested the use of SAMPLE BLANK NO
CAT BT the BioTécnica Controls: RETARD 5
13.002.00 READING TIME 2
- Quantinorm CAT BT REACTION VOL 500
13.003.00 REPLICATES 1
REF. LOW *
- Quantialt CAT BT 13.004.00 REF. HIGH *
SPECIAL OBSERVATIONS ABS LOW 0.000
None ABS HIGH 2.300
LOT *
TARGET VALUE *
# Parameter defined by the user. LOW LIMIT *
HIGH LIMIT *
@ Insert the concentration of the calibrator
according to the indication in the label of Autocal H
- CAT BT 13.002.00

® TRADEMARKS REVISION 00 - APR/2006


CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil.
Site: www.biotecnica.ind.br / e-mail: sac@biotecnicaltda.com.br
Equipment Applications
MICROLAB 200®
PRODUCT CHOLESTEROL
CATÁLOG BT 10.004.00
METODOLOGY ENZIMÁTIC
COLORIMÉTRIC
DETERMINATIONS 1000
TEST NAME CHOL
TYPE FINAL POINT
- The number of determinations was calculated NOT FILTER 505
taking into account the “dead volume” of the equipment TEMPERATURE 25
flask. UNIT mg/dL
REAG BLANK SI
- It´s recommended to read the Instructions before
ABS - LOW 0.000
using this product.
ABS - HIGH 2.300
Some complementary information for the realization of CALIB - NAME
STANDARD
the test are described joint to the Automation Protocol
............- NR 1
below:
............- CONC @
............- REP 1
REAGENT PREPARATION ............- ABS LOW 0.000
Reagents ready to use. ............- ABS HIGH 2.300
CALIBRATOR CONTROLS FACTOR 1
To evaluate the performance of SAMPLE BLANK No
Autocal H the test it´s sugested the use of RETARD 5
CAT BT the BioTécnica Controls: READING TIME 2
13.002.00 REACTION VOL 500
- Quantinorm CAT BT REPLICATES 1
13.003.00 REF. LOW *
REF. HIGH *
- Quantialt CAT BT 13.004.00
ABS LOW 0.000
SPECIAL OBSERVATIONS
ABS HIGH 2.300
None
LOT *
TARGET VALUE *
LOW LIMIT *
# Parameter defined by the user.
HIGH LIMIT *
@ Insert the concentration of the calibrator
according to the indication in the label of Autocal H
- CAT BT 13.002.00

® TRADEMARKS REVISION 00 - APR/2006


CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil.
Site: www.biotecnica.ind.br / e-mail: sac@biotecnicaltda.com.br
Equipment Applications
MICROLAB 200®
PRODUCT CHOLESTEROL HDL
CATÁLOG BT 10.005.00
METODOLOGY PRECIPITATION +
ENZIMATIC
COLORIMÉTRIC
DETERMINATIONS 100

- The number of determinations was calculated NOT


taking into account the “dead volume” of the equipment
flask.

- It´s recommended to read the Instructions before TEST NAME HDL CHOL
using this product. TYPE FINAL POINT
FILTER 505
Some complementary information for the realization of TEMPERATURE 25
the test are described joint to the Automation Protocol UNIT mg/dL
below: REAG BLANK SI
ABS - LOW 0.000
REAGENT PREPARATION ABS - HIGH 2.300
Reagents ready to use. CALIB - NAME
STANDARD
CALIBRATOR CONTROLS
To evaluate the performance of ............- NR 1
Autocal H the test it´s sugested the use of ............- CONC @
CAT BT the BioTécnica Controls: ............- REP 1
13.002.00 ............- ABS LOW 0.000
- Quantinorm CAT BT ............- ABS HIGH 2.300
13.003.00 FACTOR 1
SAMPLE BLANK No
- Quantialt CAT BT 13.004.00 RETARD 5
SPECIAL OBSERVATIONS READING TIME 2
None REACTION VOL 500
REPLICATES 1
REF. LOW *
# Parameter defined by the user. REF. HIGH *
ABS LOW 0.000
@ Insert the concentration of the calibrator ABS HIGH 2.300
according to the indication in the label of Autocal H LOT *
- CAT BT 13.002.00 TARGET VALUE *
LOW LIMIT *
HIGH LIMIT *

® TRADEMARKS REVISION 00 - APR/2006


CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil.
Site: www.biotecnica.ind.br / e-mail: sac@biotecnicaltda.com.br
Equipment Applications
MICROLAB 200®
PRODUCT CREATININE
CATÁLOG BT 10.007.00
METODOLOGY COLORIMÉTRIC
DETERMINATIONS 500

- The number of determinations was calculated NOT TEST NAME CRE


taking into account the “dead volume” of the equipment TYPE TWO POINTS
flask. FILTER 505
TEMPERATURE 37
- It´s recommended to read the Instructions before UNIT mg/dL
using this product. REAG BLANK SI
ABS - LOW 0.000
Some complementary information for the realization of
ABS - HIGH 2.300
the test are described joint to the Automation Protocol
below: CALIB - NAME
STANDARD
............- NR 1
REAGENT PREPARATION ............- CONC @
Reagents ready to use. ............- REP 1
CALIBRATOR CONTROLS ............- ABS LOW 0.000
To evaluate the performance of ............- ABS HIGH 2.300
Autocal H the test it´s sugested the use of FACTOR 1
CAT BT the BioTécnica Controls: SAMPLE BLANK No
13.002.00 RETARD 12
- Quantinorm CAT BT READING TIME 60
13.003.00 REACTION VOL 500
REPLICATES 1
- Quantialt CAT BT 13.004.00 REF. LOW *
SPECIAL OBSERVATIONS REF. HIGH *
None ABS LOW 0.000
ABS HIGH 2.300
LOT *
# Parameter defined by the user. TARGET VALUE *
LOW LIMIT *
@ Insert the concentration of the calibrator
HIGH LIMIT *
according to the indication in the label of Autocal H
- CAT BT 13.002.00

® TRADEMARKS REVISION 00 - APR/2006


CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil.
Site: www.biotecnica.ind.br / e-mail: sac@biotecnicaltda.com.br
Equipment Applications
MICROLAB 200®
PRODUCT GAMMA GT
CATÁLOG BT 11.006.00
METODOLOGY KINÉTIC COLORIMÉTRIC
DETERMINATIONS 100

- The number of determinations was calculated NOT


taking into account the “dead volume” of the equipment TEST NAME GGT
flask. TYPE KINETIC
FILTER 405
- It´s recommended to read the Instructions before TEMPERATURE 37
using this product. UNIT U/L
REAG BLANK NO
Some complementary information for the realization of
the test are described joint to the Automation Protocol CALIB NAME NO
below: FACTOR 2757
SAMPLE BLANK NO
REAGENT PREPARATION RETARD 60
Mix 4 parts of Reagent A with 1 part of Reagent B READING TIME 60
and homogenize it gently. REACTION VOL. 500
CALIBRATOR CONTROLS REPET. TEST 1
To evaluate the performance of REF.LOW *
Autocal H the test it´s sugested the use of REF. HIGH *
CAT BT the BioTécnica Controls: ABS LOW 0.000
13.002.00 ABS HIGH 2.300
- Quantinorm CAT BT LOT *
13.003.00 TARGET VALUE *
LOW LIMIT *
- Quantialt CAT BT 13.004.00 HIGH LIMIT *
SPECIAL OBSERVATIONS
None

# Parameter defined by the user.

@ Insert the concentration of the calibrator


according to the indication in the label of Autocal H
- CAT BT 13.002.00

® TRADEMARKS REVISION 00 - APR/2006


CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil.
Site: www.biotecnica.ind.br / e-mail: sac@biotecnicaltda.com.br
Equipment Applications
MICROLAB 200®
PRODUCT GLUCOSE
CATÁLOG BT 10.008.00
METODOLOGY ENZIMÁTIC
COLORIMÉTRIC
DETERMINATIONS 1000
TEST NAME GLU
TYPE FINAL POINT
- The number of determinations was calculated NOT FILTER 505
taking into account the “dead volume” of the equipment TEMPERATURE 25
flask. UNIT mg/dL
REAG BLANK SI
- It´s recommended to read the Instructions before
ABS - LOW 0.000
using this product.
ABS - HIGH 2.300
Some complementary information for the realization of CALIB - NAME
STANDARD
the test are described joint to the Automation Protocol
............- NR 1
below:
............- CONC @
............- REP 1
REAGENT PREPARATION ............- ABS LOW 0.000
Reagents ready to use. ............- ABS HIGH 2.300
CALIBRATOR CONTROLS FACTOR 1
To evaluate the performance of SAMPLE BLANK No
Autocal H the test it´s sugested the use of RETARD 5
CAT BT the BioTécnica Controls: READING TIME 2
13.002.00 REACTION VOL 500
- Quantinorm CAT BT REPLICATES 1
13.003.00 REF. LOW *
REF. HIGH *
- Quantialt CAT BT 13.004.00
ABS LOW 0.000
SPECIAL OBSERVATIONS
ABS HIGH 2.300
None
LOT *
TARGET VALUE *
LOW LIMIT *
# Parameter defined by the user.
HIGH LIMIT *
@ Insert the concentration of the calibrator
according to the indication in the label of Autocal H
- CAT BT 13.002.00

® TRADEMARKS REVISION 00 - APR/2006


CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil.
Site: www.biotecnica.ind.br / e-mail: sac@biotecnicaltda.com.br
Equipment Applications
MICROLAB 200®
PRODUCT IRON CRX
CATÁLOG BT 12.004.00
METODOLOGY COLORIMETRIC
DETERMINATIONS 200
TEST NAME IRON
TYPE FINAL POINT
- The number of determinations was calculated NOT FILTER 620
taking into account the “dead volume” of the equipment TEMPERATURE 25
flask. UNIT µg/dL
REAG BLANK SI
- It´s recommended to read the Instructions before
ABS - LOW 0.000
using this product.
ABS - HIGH 2.300
Some complementary information for the realization of CALIB - NAME
STANDARD
the test are described joint to the Automation Protocol
............- NR 1
below:
............- CONC @
............- REP 1
REAGENT PREPARATION ............- ABS LOW 0.000
Reagents ready to use. ............- ABS HIGH 2.300
CALIBRATOR CONTROLS FACTOR 1
To evaluate the performance of SAMPLE BLANK No
Autocal H the test it´s sugested the use of RETARD 5
CAT BT the BioTécnica Controls: READING TIME 2
13.002.00 REACTION VOL 500
- Quantinorm CAT BT REPLICATES 1
13.003.00 REF. LOW *
REF. HIGH *
- Quantialt CAT BT 13.004.00
ABS LOW 0.000
SPECIAL OBSERVATIONS
ABS HIGH 2.300
None
LOT *
TARGET VALUE *
LOW LIMIT *
# Parameter defined by the user.
HIGH LIMIT *
@ Insert the concentration of the calibrator
according to the indication in the label of Autocal H
- CAT BT 13.002.00

® TRADEMARKS REVISION 00 - APR/2006


CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil.
Site: www.biotecnica.ind.br / e-mail: sac@biotecnicaltda.com.br
Equipment Applications
MICROLAB 200®
PRODUCT LACTATE
DEHYDROGENASE
CATÁLOG BT 11.004.00
METODOLOGY KINÉTICO UV
DETERMINATIONS 100

- The number of determinations was calculated NOT TEST NAME LDH


taking into account the “dead volume” of the equipment TYPE KINETIC
flask. FILTER 340
TEMPERATURE 37
- It´s recommended to read the Instructions before UNIT U/L
using this product. REAG BLANK NO
Some complementary information for the realization of CALIB NAME NO
the test are described joint to the Automation Protocol FACTOR 8095
below:
SAMPLE BLANK NO
RETARD 60
REAGENT PREPARATION READING TIME 60
Mix 4 parts of Reagent A with 1 part of Reagent B REACTION VOL. 500
and homogenize it gently. REPET. TEST 1
CALIBRATOR CONTROLS REF.LOW *
To evaluate the performance of REF. HIGH *
Autocal H the test it´s sugested the use of ABS LOW 0.000
CAT BT the BioTécnica Controls: ABS HIGH 2.300
13.002.00 LOT *
- Quantinorm CAT BT TARGET VALUE *
13.003.00 LOW LIMIT *
HIGH LIMIT *
- Quantialt CAT BT 13.004.00
SPECIAL OBSERVATIONS
None

# Parameter defined by the user.

@ Insert the concentration of the calibrator


according to the indication in the label of Autocal H
- CAT BT 13.002.00

® TRADEMARKS REVISION 00 - APR/2006


CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil.
Site: www.biotecnica.ind.br / e-mail: sac@biotecnicaltda.com.br
Equipment Applications
MICROLAB 200®
PRODUCT MAGNÉSIUM
CATÁLOG BT 12.007.00
METODOLOGY MAGON
DETERMINATIONS 100
TEST NAME MAG
TYPE FINAL POINT
- The number of determinations was calculated NOT FILTER 505
taking into account the “dead volume” of the equipment TEMPERATURE 25
flask. UNIT mg/dL
REAG BLANK SI
- It´s recommended to read the Instructions before
ABS - LOW 0.000
using this product.
ABS - HIGH 2.300
Some complementary information for the realization of CALIB - NAME
STANDARD
the test are described joint to the Automation Protocol
............- NR 1
below:
............- CONC @
............- REP 1
REAGENT PREPARATION ............- ABS LOW 0.000
Reagents ready to use. ............- ABS HIGH 2.300
CALIBRATOR CONTROLS FACTOR 1
To evaluate the performance of SAMPLE BLANK No
Autocal H the test it´s sugested the use of RETARD 5
CAT BT the BioTécnica Controls: READING TIME 2
13.002.00 REACTION VOL 500
- Quantinorm CAT BT REPLICATES 1
13.003.00 REF. LOW *
REF. HIGH *
- Quantialt CAT BT 13.004.00
ABS LOW 0.000
SPECIAL OBSERVATIONS
ABS HIGH 2.300
None
LOT *
TARGET VALUE *
LOW LIMIT *
# Parameter defined by the user.
HIGH LIMIT *
@ Insert the concentration of the calibrator
according to the indication in the label of Autocal H
- CAT BT 13.002.00

® TRADEMARKS REVISION 00 - APR/2006


CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil.
Site: www.biotecnica.ind.br / e-mail: sac@biotecnicaltda.com.br
Equipment Applications
MICROLAB 200®
PRODUCT PHÓSPHORUS UV
CATÁLOG BT 12.006.00
METODOLOGY Molibidate UV
DETERMINATIONS 100

TEST NAME PUV


- The number of determinations was calculated NOT TYPE FINAL POINT
taking into account the “dead volume” of the equipment FILTER 340
flask. TEMPERATURE 25
UNIT mg/dL
- It´s recommended to read the Instructions before REAG BLANK SI
using this product.
ABS - LOW 0.000
ABS - HIGH 2.300
Some complementary information for the realization of
the test are described joint to the Automation Protocol CALIB - NAME
STANDARD
below:
............- NR 1
............- CONC @
REAGENT PREPARATION ............- REP 1
Reagents ready to use. ............- ABS LOW 0.000
CALIBRATOR CONTROLS ............- ABS HIGH 2.300
To evaluate the performance of FACTOR 1
Autocal H the test it´s sugested the use of SAMPLE BLANK No
CAT BT the BioTécnica Controls: RETARD 5
13.002.00 READING TIME 2
- Quantinorm CAT BT REACTION VOL 500
13.003.00 REPLICATES 1
REF. LOW *
- Quantialt CAT BT 13.004.00 REF. HIGH *
SPECIAL OBSERVATIONS ABS LOW 0.000
None ABS HIGH 2.300
LOT *
TARGET VALUE *
# Parameter defined by the user. LOW LIMIT *
HIGH LIMIT *
@ Insert the concentration of the calibrator
according to the indication in the label of Autocal H
- CAT BT 13.002.00

® TRADEMARKS REVISION 00 - APR/2006


CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil.
Site: www.biotecnica.ind.br / e-mail: sac@biotecnicaltda.com.br
Equipment Applications
MICROLAB 200®
PRODUCT PROTEÍN TOTAL
CATÁLOG BT 10.009.00
METODOLOGY COLORIMÉTRIC
DETERMINATIONS 500
TEST NAME PRO
TYPE FINAL POINT
- The number of determinations was calculated NOT FILTER 546
taking into account the “dead volume” of the equipment TEMPERATURE 25
flask. UNIT g/dL
REAG BLANK SI
- It´s recommended to read the Instructions before
ABS - LOW 0.000
using this product.
ABS - HIGH 2.300
Some complementary information for the realization of CALIB - NAME
STANDARD
the test are described joint to the Automation Protocol
............- NR 1
below:
............- CONC @
............- REP 1
REAGENT PREPARATION ............- ABS LOW 0.000
Reagents ready to use. ............- ABS HIGH 2.300
CALIBRATOR CONTROLS FACTOR 1
To evaluate the performance of SAMPLE BLANK No
Autocal H the test it´s sugested the use of RETARD 5
CAT BT the BioTécnica Controls: READING TIME 2
13.002.00 REACTION VOL 500
- Quantinorm CAT BT REPLICATES 1
13.003.00 REF. LOW *
REF. HIGH *
- Quantialt CAT BT 13.004.00
ABS LOW 0.000
SPECIAL OBSERVATIONS
ABS HIGH 2.300
None
LOT *
TARGET VALUE *
LOW LIMIT *
# Parameter defined by the user.
HIGH LIMIT *
@ Insert the concentration of the calibrator
according to the indication in the label of Autocal H
- CAT BT 13.002.00

® TRADEMARKS REVISION 00 - APR/2006


CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil.
Site: www.biotecnica.ind.br / e-mail: sac@biotecnicaltda.com.br
Equipment Applications
MICROLAB 200®
PRODUCT PROTEÍN URINÁRY
CATÁLOG BT 10.016.00
METODOLOGY COLORIMÉTRIC TEST NAME PRO-U
DETERMINATIONS 200 TYPE FINAL POINT
FILTER 600
TEMPERATURE 25
- The number of determinations was calculated NOT UNIT g/dL
taking into account the “dead volume” of the equipment REAG BLANK SI
flask.
ABS - LOW 0.000
ABS - HIGH 2.300
- It´s recommended to read the Instructions before
using this product. CALIB - NAME
STANDARD
............- NR 1
Some complementary information for the realization of
the test are described joint to the Automation Protocol ............- CONC @
below: ............- REP 1
............- ABS LOW 0.000
............- ABS HIGH 2.300
REAGENT PREPARATION FACTOR 1
Reagents ready to use. SAMPLE BLANK No
CALIBRATOR CONTROLS RETARD 5
To evaluate the performance of READING TIME 2
Autocal H the test it´s sugested the use of REACTION VOL 500
CAT BT the BioTécnica Controls: REPLICATES 1
13.002.00 REF. LOW *
- Quantinorm CAT BT REF. HIGH *
13.003.00
ABS LOW 0.000
ABS HIGH 2.300
- Quantialt CAT BT 13.004.00
LOT *
SPECIAL OBSERVATIONS
TARGET VALUE *
None
LOW LIMIT *
HIGH LIMIT *
# Parameter defined by the user.

@ Insert the concentration of the calibrator


according to the indication in the label of Autocal H
- CAT BT 13.002.00

® TRADEMARKS REVISION 00 - APR/2006


CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil.
Site: www.biotecnica.ind.br / e-mail: sac@biotecnicaltda.com.br
Equipment Applications
MICROLAB 200®
PRODUCT TRIGLICÉRIDES
CATÁLOG BT 10.010.00
METODOLOGY ENZIMÁTIC
COLORIMÉTRIC
DETERMINATIONS 500
TEST NAME TGR
TYPE FINAL POINT
- The number of determinations was calculated NOT FILTER 505
taking into account the “dead volume” of the equipment TEMPERATURE 25
flask. UNIT mg/dL
REAG BLANK SI
- It´s recommended to read the Instructions before
ABS - LOW 0.000
using this product.
ABS - HIGH 2.300
Some complementary information for the realization of CALIB - NAME
STANDARD
the test are described joint to the Automation Protocol
............- NR 1
below:
............- CONC @
............- REP 1
REAGENT PREPARATION ............- ABS LOW 0.000
Reagents ready to use. ............- ABS HIGH 2.300
CALIBRATOR CONTROLS FACTOR 1
To evaluate the performance of SAMPLE BLANK No
Autocal H the test it´s sugested the use of RETARD 5
CAT BT the BioTécnica Controls: READING TIME 2
13.002.00 REACTION VOL 500
- Quantinorm CAT BT REPLICATES 1
13.003.00 REF. LOW *
REF. HIGH *
- Quantialt CAT BT 13.004.00
ABS LOW 0.000
SPECIAL OBSERVATIONS
ABS HIGH 2.300
None
LOT *
TARGET VALUE *
LOW LIMIT *
# Parameter defined by the user.
HIGH LIMIT *
@ Insert the concentration of the calibrator
according to the indication in the label of Autocal H
- CAT BT 13.002.00

® TRADEMARKS REVISION 00 - APR/2006


CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil.
Site: www.biotecnica.ind.br / e-mail: sac@biotecnicaltda.com.br
Equipment Applications
MICROLAB 200®
PRODUCT URÉA UV
CATÁLOG BT 10.012.00
METODOLOGY KINÉTICO UV
DETERMINATIONS 250

- The number of determinations was calculated NOT TEST NAME UUV


taking into account the “dead volume” of the equipment TYPE TWO POINTS
flask. FILTER 340
TEMPERATURE 37
- It´s recommended to read the Instructions before UNIT mg/dL
using this product. REAG BLANK SI
ABS - LOW 0.000
Some complementary information for the realization of
ABS - HIGH 2.300
the test are described joint to the Automation Protocol
below: CALIB - NAME
STANDARD
............- NR 1
REAGENT PREPARATION ............- CONC @
Mix 4 parts of Reagent A with 1 part of Reagent B ............- REP 1
and homogenize it gently. ............- ABS LOW 0.000
CALIBRATOR CONTROLS ............- ABS HIGH 2.300
To evaluate the performance of FACTOR 1
Autocal H the test it´s sugested the use of SAMPLE BLANK No
CAT BT the BioTécnica Controls: RETARD 20
13.002.00 READING TIME 60
- Quantinorm CAT BT REACTION VOL 500
13.003.00 REPLICATES 1
REF. LOW *
- Quantialt CAT BT 13.004.00 REF. HIGH *
SPECIAL OBSERVATIONS ABS LOW 0.600
None ABS HIGH 2.300
LOT *
TARGET VALUE *
# Parameter defined by the user. LOW LIMIT *
HIGH LIMIT *
@ Insert the concentration of the calibrator
according to the indication in the label of Autocal H
- CAT BT 13.002.00

® TRADEMARKS REVISION 00 - APR/2006


CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil.
Site: www.biotecnica.ind.br / e-mail: sac@biotecnicaltda.com.br
Equipment Applications
MICROLAB 200®
PRODUCT URIC ACID -
MONOREAGENT
CATÁLOG BT 10.001.00
METODOLOGY ENZYMATIC
COLORIMETRIC
DETERMINATIONS 500 TEST NAME UAC
TYPE FINAL POINT
FILTER 505
- The number of determinations was calculated NOT TEMPERATURE 25
taking into account the “dead volume” of the equipment UNIT mg/dL
flask. REAG BLANK SI
ABS - LOW 0.000
- It´s recommended to read the Instructions before
ABS - HIGH 2.300
using this product.
CALIB - NAME
STANDARD
Some complementary information for the realization of
............- NR 1
the test are described joint to the Automation Protocol
below: ............- CONC @
............- REP 1
............- ABS LOW 0.000
REAGENT PREPARATION ............- ABS HIGH 2.300
Reagents ready to use. FACTOR 1
CALIBRATOR CONTROLS SAMPLE BLANK No
To evaluate the performance of RETARD 5
Autocal H the test it´s sugested the use of READING TIME 2
CAT BT the BioTécnica Controls: REACTION VOL 500
13.002.00 REPLICATES 1
- Quantinorm CAT BT REF. LOW *
13.003.00 REF. HIGH *
ABS LOW 0.000
- Quantialt CAT BT 13.004.00
ABS HIGH 2.300
SPECIAL OBSERVATIONS
LOT *
None
TARGET VALUE *
LOW LIMIT *
HIGH LIMIT *
# Parameter defined by the user.

@ Insert the concentration of the calibrator


according to the indication in the label of Autocal H
- CAT BT 13.002.00

® TRADEMARKS REVISION 00 - APR/2006


CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil.
Site: www.biotecnica.ind.br / e-mail: sac@biotecnicaltda.com.br
Equipment Applications
MICROLAB 200®
PRODUCT URIC ACID -
TWOREAGENT
CATÁLOG BT 10.014.00
METODOLOGY ENZYMATIC
COLORIMETRIC
DETERMINATIONS 250

- The number of determinations was calculated NOT


taking into account the “dead volume” of the equipment
flask.
TEST NAME UAC
- It´s recommended to read the Instructions before TYPE FINAL POINT
using this product. FILTER 505
TEMPERATURE 25
Some complementary information for the realization of UNIT mg/dL
the test are described joint to the Automation Protocol REAG BLANK SI
below:
ABS - LOW 0.000
ABS - HIGH 2.300
REAGENT PREPARATION CALIB - NAME
STANDARD
Mix 4 parts of Reagent A with 1 part of Reagent B
and homogenize it gently. ............- NR 1
CALIBRATOR CONTROLS ............- CONC @
To evaluate the performance of ............- REP 1
Autocal H the test it´s sugested the use of ............- ABS LOW 0.000
CAT BT the BioTécnica Controls: ............- ABS HIGH 2.300
13.002.00 FACTOR 1
- Quantinorm CAT BT SAMPLE BLANK No
13.003.00 RETARD 5
READING TIME 2
- Quantialt CAT BT 13.004.00 REACTION VOL 500
SPECIAL OBSERVATIONS REPLICATES 1
None REF. LOW *
REF. HIGH *
ABS LOW 0.000
# Parameter defined by the user. ABS HIGH 2.300
LOT *
@ Insert the concentration of the calibrator TARGET VALUE *
according to the indication in the label of Autocal H LOW LIMIT *
- CAT BT 13.002.00
HIGH LIMIT *

® TRADEMARKS REVISION 00 - APR/2006


CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil.
Site: www.biotecnica.ind.br / e-mail: sac@biotecnicaltda.com.br
Equipment Applications
MICROLAB 200®
PRODUTO UREA
CATÁLOGO BT 10.013.00
METODOLOGIA ENZYMATIC
COLORIMETRIC
Nº DETERMINAÇÕES 500

- The number of determinations was calculated NOT


taking into account the “dead volume” of the equipment
flask.

- It´s recommended to read the Instructions before


using this product.

Some complementary information for the realization of TEST NAME URE


the test are described joint to the Automation Protocol TYPE FINAL POINT
below: FILTER 578
TEMPERATURE 25
REAGENT PREPARATION UNIT mg/dL
Please follow the Instructions of the kit. REAG BLANK SI
CALIBRATOR CONTROLS ABS - LOW 0.000
To evaluate the performance of ABS - HIGH 2.300
Autocal H the test it´s sugested the use of CALIB - NAME
STANDARD
CAT BT the BioTécnica Controls:
13.002.00 ............- NR 1
- Quantinorm CAT BT ............- CONC @
13.003.00 ............- REP 1
............- ABS LOW 0.000
- Quantialt CAT BT 13.004.00 ............- ABS HIGH 2.300
SPECIAL OBSERVATIONS FACTOR 1
None SAMPLE BLANK No
RETARD 5
READING TIME 2
# Parameter defined by the user. REACTION VOL 500
REPLICATES 1
@ Insert the concentration of the calibrator REF. LOW *
according to the indication in the label of Autocal H REF. HIGH *
- CAT BT 13.002.00 ABS LOW 0.000
ABS HIGH 2.300
LOT *
TARGET VALUE *
LOW LIMIT *
HIGH LIMIT *

® TRADEMARKS REVISION 00 - APR/2006


CUSTOMER TECHNICAL SERVICE – BioTécnica Ind e Com. Ltda.
Tel / Fax: +55 35 3214-4646 – Varginha – MG – Brasil.
Site: www.biotecnica.ind.br / e-mail: sac@biotecnicaltda.com.br