Professional Documents
Culture Documents
AND EVALUATION
OSPAR SENIOR INTERNS
RECORDED DETAIL
The degree of clarity or sharpness with which structures are recorded on a radiographic
image.
Quick Review
The actual focal spot is the actually bombarded by electrons from the
filament. The effective focal spot is the foreshortened size of the focus as it is
projected down toward the image receptor; that is, as it would be seen
looking up into the emerging x-ray beam.
The angle of the anode can have a significant effect on recorded detail.
Differences in anode angle, all other factors remaining constant, will affect
the size of the effective focal spot.
FOCAL SPOT SIZE
If the use of a smaller focal spot size provides us with better recorded
detail, why is the smallest focal spot available not always used?
Quick Review
If the use of a smaller focal spot size provides us with better recorded
detail, why is the smallest focal spot available not always used?
Simply because focal spot size is also associated with the buildup of
heat within the x-ray tube. Large quantities of heat delivered to the x-ray
tube, especially in a short period of time, can be very damaging to the tube
and can shorten its life span.
Summary - FSS
1. Focal spot size affects detail by influencing the degree of blur or unsharpness: increased focal spot
size = increased blur = decreased detail.
2. Unsharpness or blur is directly related to focal spot size and OID, and inversely related to SID.
3. Using a small focal spot improves recorded detail, but generates more heat at the anode.
4. According to the line focus principle, the effective or projected focal spot size is always smaller than
the actual focal spot.
5. Effective focal spot size varies along the longitudinal axis of the image receptor, being largest at the
cathode end and smallest at the anode end of the X-ray beam.
6. Smaller anode angles can permit larger actual focal spot sizes while maintaining small effective focal
spot sizes—at the expense of accentuating the anode heel effect.
MOTION
PIXEL 1 PIXEL 2
PIXEL PITCH
The lower the pixel pitch number, the closer the pixels, and the sharper
the image.
PIXEL PITCH IN MONITORS
SPECS:
Pixel Matrix 3440 x 1440 pixels = 4,953,600 million pixels
Pixel Pitch 0.23175 x 0.23175 mm = 53 µm
PIXEL PITCH
ValenceINElectrons
DETECTORS
PIXEL PITCH
ValenceINElectrons
DETECTORS
Capture area/ Fill Factor = The area of the detector that is sensitive to
X-rays
GE HealthCare P600FP 17 x 17 radiography flat Angell Technology – 14 x 17 Multipurpose
panel detector radiography flat panel detector
DISTORT
ION
Feat. Kilabot ng OSPAR-1
SIZE DISTORTION
Size distortion or magnification, is defined as the misrepresentation in the image of the true size
of the object, particularly in comparison to other structures in the patient’s body.
The best relationship to produce the least magnification is to have the object as close to the
image receptor as possible and to have the x-ray tube as far away from the image receptor as
possible. In technical distance terms, it is best to have a short object-image distance (OID) and
a long source-image distance (SID).
OBJECT-IMAGE DISTANCE (OID) AND SIZE DISTORTION
If the SID is 72 inches, the SOD is 68 inches, and the object length is 6 inches,
what is the image length?
Image size SID
EXAMPLE =
Object size SOD
If the SID is 72 inches, the SOD is 68 inches, and the object length is 6 inches,
what is the image length?
summary of size distortion.
• Size distortion is the misrepresentation in the image of the true size of the object, particularly
in comparison to other structures in the patient’s body.
• For the least magnification use a short OID and a long SID.
• Always try to put the object or structure of interest close to the image receptor to avoid
magnification.
• The ratio of the image size to the object size is proportional to the ratio of the SID to the SOD.
When referring to “size” this is either the length of the image or object or the width of the
image or object.
summary of size distortion.
• The magnification factor tells the radiographer (and the physicians) how much longer and
wider the image will be compared to the object’s actual length and width.
• In some clinical situations, it is important for the radiologist (or orthopedic surgeon) to know
the amount of magnification produced on a radiograph. This is especially important during
angiograms and mammography. The image size, object size, magnification factor, and percent
of magnification can all be calculated.
SHAPE DISTORTION
Shape distortion is the misrepresentation by the image of the true shape of the object.
3 TYPES OF DISTORTION
• Foreshortening
• Elongation
• Spatial Distortion
FORESHORTENING
Foreshortening – the image appears shorter and fatter than the object.
E L O N G AT I O N
Elongation – the image appears longer and skinnier than the object.
SPATIAL DISTORTION
Spatial Distortion – the three-dimensional relationships of the object are misrepresented, for
example, appearing dislocated when the joint is actually in correct alignment.
AEC
AEC (AUTOMATIC EXPOSURE
CONTROL)
The AEC is a device that measures the quantity
of radiation that reaches the image receptor. It
automatically terminates the exposure when the
image receptor has received the required
radiation intensity
Phototimer
Phototimer AEC devices are considered exit-type
devices because the detectors are positioned
behind the image receptor.
Ionization or ion chamber
is a hollow cell that contains air and is connected
to the timer circuit via an electrical wire.
Ionization-chamber AEC devices are considered
entrance-type devices because the detectors are
positioned in front of the image receptor
RADIOGRA
PHIC GRIDS
is an accessory used solely for the
purpose of improving image quality.
Grids are designed to absorb scatter
radiation before it reaches the
recording medium.
RADIOGRA
Types of Grid Cutoff Errors:
UPSIDE DOWN
PHIC GRIDS FOCUSED
is an accessory used solely for the
OFF-LEVEL
purpose of improving image quality.
Grids are designed to absorb scatter
radiation before it reaches the
recording medium.
UPSIDE DOWN
FOCUSED