Professional Documents
Culture Documents
IN CT SCAN
RUMBIDZAI
DEWERE
SIMBARASHE
NYENGERE
TO
ADDR
• DEFINITION OF TERMS
ESS
• PHYSICS BASED ARTEFACTS
• PATIENT PROPERTIES BASED
ARTEFACTS
• SCANNER BASED ARTEFACTS
• HELICAL AND MULTISCLICE
ARTEFACTS
INTROD
UCTION
• Artifacts are any errors in the perception or representation
of any information introduced by modalities or techniques
• Ct artifacts are common and can occur for various reasons
• CT artifacts can seriously degrade CT images and make
images diagnostically unstable
• Knowledge of these artifacts is important because they can
mimic pathology or can degrade image to non-diagnostic
levels
INTROD
UCTION
Artefact refers to any systematic
discrepancy between the CT
numbers in the reconstructed image
and the true attenuation coefficients
of the object
CLASSIFICATION
• Ct artifacts can be classified according to the underlying cause of the artifacts
• Physics –based artifacts- result from physical processes involved in acquisition
data
• Patient based artifacts-these are caused due to patient movement or presence of
metallic materials in or in patients
• Scanner -based artifacts-these are due to imperfections in scanner function
• Helical and multi-section technique artifacts- thee are produced by image
reconstruction process
1. Patient based
• These are
• Presence of metallic materials
• Patient motion
• Incomplete projections
Metallic materials
• The presence of metal objects in the scan field can lead to severe
streaking artifacts.
• The metal produces a beam hardening and photon starvation artefact
• This can also happen with other high attenuation materials such as iv
contrast
• They occur because the density of the metal is beyond the normal
range that can be handled by the computer ,resulting in incomplete
attenuation profiles
Metallic materials
Avoidance
• Patients are normally asked to take off removable metal objects such as
jewelry before scanning commences
• For non-removable items such as dental fillings ,prosthetics
devices ,and surgical clips ,it is sometimes possible to use gantry
angulation to exclude the metal inserts from scans of nearby anatomy
• When it is impossible to scan the required anatomy without including
metal objects ,increasing technique ,especially kilovoltage ,may help
penetrate some objects ,and using thin sections will reduce the
contribution due to partial volume artifact
Software correction
• Streaking caused by overranging can be greatly reduced by means of
special software corrections.
• Manufacturers use a variety of interpolation techniques to substitute
the overrange values in attenuation profiles.
• The usefulness of metal artifact reduction software is sometimes
limited because, although streaking distant from the metal implants is
removed, there still remains a loss of detail around the metal-tissue
interface, which is often the main area of diagnostic interest.
Patient motion
• Motion artefact can be caused by patient swallowing ,breathing ,pulsating of
heart and vessels and patient moving
• If a patient or structure moves as the gantry rotates ,the object will be detected
being in several positions and represented as such
• Patient motion can cause misregistration artifacts, which usually appear as
shading or streaking in the reconstructed image .
• Steps can be taken to prevent voluntary motion, but some involuntary motion
may be unavoidable during body scanning
• However ,there are special features on some scanners designed to minimize the
resulting artifacts .
Patient motion
Avoidance by operator
• The use of positioning aids is sufficient to prevent voluntary movement in most patients.
• However, in some cases (eg, pediatric patients), it may be necessary to immobilize the patient by means of s
• Using as short a scan time as possible helps minimize artifacts when scanning regions prone to movement.
• Respiratory motion can be minimized if patients are able to hold their breath for the duration of the scan.
• The sensitivity of the image to motion artifacts depends on the orientation of the motion. Therefore, it is pref
start and end position of the tube is aligned with the primary direction of motion, for example, vertically abo
patient undergoing a chest scan.
BUILT IN FEATURES FOR MINIMIZING
MOTION ARTIFACTS
• Manufacturers minimize motion artifacts by using overscan and underscan modes, software
correction, and cardiac gating.
• Overscan and underscan modes: The maximum discrepancy in detector readings occurs between
views obtained toward the beginning and end of a 360° scan.
• Some scanner models use over scan mode for axial body scans, whereby an extra 10% or so is
added to the standard 360° rotation.
• The repeated projections are averaged, which helps reduce the severity of motion artifacts.
• The use of partial scan mode can also reduce motion artifacts, but this may be at the expense of
poorer resolution.
• Software correction: Most scanners, when used in body scan mode, automatically apply reduced
weighting to the beginning and end views to suppress their contribution to the final image.
However, this may lead to more noise in the vertical direction of the resultant image, depending
on the shape of the patient.
• Additional, specialized motion correction is available on some scanners.
• Cardiac gating –the rapid motion of the heart can lead to severe artifacts
in images of the heart and to artifacts that can mimic disease in
associated structures eg dissected aorta .
• To overcome these difficulties, techniques have been developed to
produce images by using data from just a fraction of the cardiac cycle,
when there is least cardiac motion.
• This is achieved by combining electrocardiographic gating techniques
with specialized methods of image reconstruction (4
Incomplete projections
• If any portion of the patient lies outside the scan field of view, the
computer will have incomplete information relating to this portion
and streaking or shading artifacts are likely to be generated.
• As the arms are outside the scan field, they are not present in the
image, but their presence in some views during scanning has led to
such severe artifacts throughout the image as to significantly degrade
its usefulness.
• Similar effects can be caused by dense objects such as an intravenous
tube containing contrast medium lying outside the scan field.
Ct image of the body
obtained with the
patient ,s arm down but
outside the scanning field
shows streaking artifacts
AVOIDANCE of INCOMPLETE
PROJECTIONS ARTIFACTS
• To avoid artifacts due to incomplete projections, it is essential to position
the patient so that no parts lie outside the scan field.
• Scanners designed specifically for radiation therapy planning have wider
bores and larger scan fields of view than standard scanners and permit
greater versatility in patient positioning.
• They also allow scanning of exceptionally large patients who would not fit
within the field of view of standard scanners.
• As an alternative, the CT system may be designed with reference detectors
on the tube side or with ray paths within the gantry to eliminate possible
interference with reference detectors on the tube side or with ray paths
within the gantry to eliminate possible interference with reference data
PHYSIC
S BASED
ARTEFA
Physics-based artifacts result
CT
from the physical processes
involved in the acquisition of CT
data.
1. BEAM HARDENING
• X-ray photons coming from a tube high are made up of a full spectrum
of x-ray energies, not just the voltage that the tube is set as
• The energy of a photon determines its level of attenuation when
passing through a sample, with low-energy photons attenuating much
faster than high-energy ones
• Therefore, when an x-ray beam begins to penetrate a sample material,
the lower energy x-rays preferentially attenuate, resulting in an overall
higher energy of the x-ray beam- Beam Hardening
1.BEAM HARDENING
• The effect of beam hardening on a single material can
usually be managed by adjusting CT reconstruction
algorithms.
• However, if the sample contains multiple materialsto
within a single scan volume, imaging differing densities
simultaneously tends to result in atifact
BEAM
0 HARDENING
CUPPING ARTEFACT
1 ARTEFACTS
• It happens when hardening is more prone to the center than the
0
periphery, therefore resembling a cup
• The resultant attenuation profile differs from the ideal profile that
2
would be obtained without beam hardening
• A profile of the CT numbers across the phantom displays a
characteristic cupped shape
BEAM
0 HARDENING
STREAKS AND DARK BAND
1 ARTEFACTS
• In very heterogeneous cross sections, dark bands or streaks
0
can appear between two dense objects in an image
• If a high density material severely reduces transmission, the