Professional Documents
Culture Documents
Extruder
Applications
Extruder
Feed
Compression
Metering
Winder
Quantitative analysis of the flow in the extrusion process Residence time is the time resins are in exposed to heat in the extruder Residence time is an average value
Concept
Introduction of a marker into the extruder along with the resin. The concentration of the marker in the plastics is monitored as a function of time
Choice of Markers
Similar flow properties as the resin Easily detected Safe to use Economical
Example of marker
TiO2 Carbon black Radioactive traces Glass microsphere Fillers Dyes Functionalized tracers
Methods
Step Input
Continuous feed of the marker when the process is at steady state conditions
Marker concentration
Time
Marker concentration
Pulse Input
Introduce small amount of the marker into the extruder system
Time
Chi-square distribution
Application
Understanding the effect of the processing equipments Characterize purging and mixing Determine expected exposure time
Procedure: Extruder
Warm up (1 hour)
Plug in the three power cords Open the air supply valve (from the ceiling above the machine) Press run on the GDP 333 blowing motor Turn the take up roll on (Red-yellow knob) Turn the winder on by pulling start and run switch up)
Procedure: Extruder
Check that the screw speed is 0 (if not adjust to zero) Press start and cold start Slowly increase the speed of the screw by the gradual speed knob Feed LDPE resin through the hopper
It takes about 5 minute for the extrudate to come out from the die
Procedure: Extruder
Drag the extrudate up to the top nip Pass it through idle rolls to the winder
Procedure: Extruder
Adjust the blow up ratio by introducing the air by slowly turning the air valve. After the desired size is obtained, close the valve
When the mark reaches the winder, start the timer Mark the film every one minute until the color disappear
(Now, its time to turn on the UV spec.)
Purge with LDPE resin to clean Follow the Shut down procedure in the lab manual except #3. It should be changed to
3. Turn off the screw speed by turning the gradual speed knob to 0 rpm. Then, push stop button
High concentration high light absorption Low concentration low light absorption
Cut sample from the mark area 1.5 1 Measure light absorption at 500 nm
Results
Time 1 2 Absorption . .
Absorption
. . 10
. . .
Time
Calculation
A t A
i i i
i i
Standard deviation
A t
Tavg
i
Tmax = Tavg + 3
Report includes
Twag, , and Tmax Residence time distribution with the mark shows Twag and Tmax Factors affecting residence time and residence time distribution
Introduction, discussion, and conclusion Tabulated summary of the result Identification of unknown sample
Justification of your identification from each results Which result did not agree with identification? What is the reason you discard that result?