CP-6 DEPTH MICROMETER Go to CALIBRATION PROCEDURE HOME or INDEX for links to other procedures. DEPTH MICROMETERS AND GAGES All of the example procedures have a statement to inspect for obvious wear or damage. One such case is a bent rod. Many times, this condition will be visually obvious, but if not, the condition should be evident when turning the thimble of the micrometer. Our facility used depth micrometers with a resolution of .001 inch. Therefore, the required accuracy was stated as .001 inch. MIL-STD-120 states that depth micrometers are accurate only to .0003 inch even when graduated in .0001 inch increments and should be used only for tolerances .001 inch or greater.

CALIBRATION PROCEDURE HCJ MACHINE COMPANY

PROCEDURE NO. CP-6

SUBJECT:

DEPTH MICROMETERS, VERNIER & ELECTRONIC DEPTH GAGES

PROCEDURE

1.0 REFERENCE DOCUMENTS 1.1 FEDERAL STANDARD GGG-C-105 1.2 MIL-STD-120 1.3 CALIBRATION EQUIPMENT ACCURACY LIST 2.0 MATERIALS, EQUIPMENT REQUIRED 2.1 GAGE BLOCKS 3.0 PROCEDURE 3.1 VISUALLY INSPECT SUBJECT GAGE COMPONENTS FOR OBVIOUS DAMAGE OR EXCESSIVE WEAR THAT MIGHT AFFECT THE FUNCTION OF THE GAGE. 3.2 ASCERTAIN THAT THE IDENTIFICATION DATA IS CORRECT. 3.3 INSPECT THE DEPTH MICROMETER AT A MINIMUM OF FIVE (5) INTERVALS OVER ITS RANGE USING GAGE BLOCKS. INCLUDE POINTS AT APPROXIMATELY 10% AND

https://sites.google.com/site/calibrationprocedures/cp-6depthmicrometer

1/4/2013 11:52:30 AM

cp-6depthmicrometer - calibrationprocedures
90% OF FULL SCALE. ACCURACY REQUIREMENTS UNLESS OTHERWISE SPECIFIED ON GAGE RECORDS, ACCURACY REQUIREMENTS ARE AS FOLLOWS (REFERENCE MIL-STD-120 & GGG-C-105). .0001 GRADUATIONS & ELECTRONIC DIGITAL 0-3 3-6 6-9 .0003 .0004 .00045 .001 GRADUATIONS ALL SIZES .001

Page 2

3.4 ACCEPT/REJECT ACCORDINGLY. 3.5 UPDATE GAGE RECORDS. APPROVAL INITIALS XXX REV. DATE: X CP-6 X-XX-XX SHEET X OF X