USP 36 General Information / 1058 Analytical Instrument Qualification 1
1058 ANALYTICAL INSTRUMENT
QUALIFICATION INTRODUCTION A large variety of laboratory equipment, instruments, and computerized analytical systems, ranging from simple nitro- gen evaporators to complex multiple-function technologies (see Instrument Categories), are used in the pharmaceutical industry to acquire data to help ensure that products are suitable for their intended use. An analysts objective is to consistently obtain reliable and valid data suitable for the intended purpose. Depending on the applications, users val- idate their procedures, calibrate their instruments, and per- form additional instrument checks, such as system suitability Figure 1. Components of data quality. tests and analysis of in-process quality control check samples to help ensure that the acquired data are reliable. With the increasing sophistication and automation of analytical instru- ments, an increasing demand has been placed on users to qualify their instruments. Analytical Instrument Qualification Unlike method validation and system suitability activities, analytical instrument qualification (AIQ) currently has no AIQ is the collection of documented evidence that an in- specific guidance or procedures. Competing opinions exist strument performs suitably for its intended purpose. Use of regarding instrument qualification and validation procedures a qualified instrument in analyses contributes to confidence and the roles and responsibilities of those who perform in the validity of generated data. them. Consequently, various approaches have been used for instrument qualification, approaches that require varying Analytical Method Validation amounts of resources and generate widely differing amounts of documentation. This chapter provides a scientific ap- Analytical method validation is the collection of docu- proach to AIQ and considers AIQ as one of the major com- mented evidence that an analytical procedure is suitable for ponents required for generating reliable and consistent data. its intended use. Use of a validated procedure with qualified Note that the amount of rigor applied to the qualification analytical instruments provides confidence that the proce- process will depend on the complexity and intended use of dure will generate test data of acceptable quality. Additional the instrumentation. This approach emphasizes AIQs place guidance on validation of compendial procedures may be in the overall process of obtaining reliable data from analyti- found in the general information chapter Validation of Com- cal instruments. pendial Procedures 1225. Validation versus Qualification System Suitability Tests In this chapter, the term validation is used for manufac- System suitability tests verify that the system will perform turing processes, analytical procedures, and software proce- in accordance with the criteria set forth in the procedure. dures and the term qualification is used for instruments. These tests are performed along with the sample analyses to Thus, the phrase analytical instrument qualification (AIQ) ensure that the systems performance is acceptable at the is used for the process of ensuring that an instrument is time of the test. USP general chapter Chromatography 621 suitable for its intended application. presents a more detailed discussion of system suitability tests as related to chromatographic systems. COMPONENTS OF DATA QUALITY Quality Control Check Samples There are four critical components involved in the genera- tion of reliable and consistent data (quality data). Figure 1 Many analysts carry out their tests on instruments stan- shows these components as layered activities within a qual- dardized using reference materials and/or calibration stan- ity triangle. Each layer adds to the overall quality. Analytical dards. Some analyses also require the inclusion of quality instrument qualification forms the base for generating qual- control check samples to provide an in-process or ongoing ity data. The other components essential for generating assurance of the tests suitable performance. In this manner, quality data are analytical method validation, system suita- AIQ and analytical method validation contribute to the qual- bility tests, and quality control check samples. These quality ity of analysis before analysts conduct the tests. System suita- components are described below. bility tests and quality control checks help ensure the quality of analytical results immediately before or during sample analysis. ANALYTICAL INSTRUMENT QUALIFICATION PROCESS The following sections address in detail the AIQ process. The other three components of building quality into analyti- cal dataanalytical method validation, system suitability 2 1058 Analytical Instrument Qualification / General Information USP 36 Table 1. Timing, Applicability, and Activities for Each Phase of Analytical Instrument Qualification* Operational Performance Design Qualification Installation Qualification Qualification Qualification Timing and Applicability Prior to purchase of a new At installation of each instrument After installation or Periodically at specified intervals model of instrument (new, old, or existing major repair of for each instrument unqualified) each instrument Activities Assurance of manufacturers Description Fixed parameters Preventive maintenance and DQ repairs Assurance of adequate support Instrument delivery Establish practices to address availability from manufacturer operation, calibration, maintenance, and change control Utilities/facility Environment Instruments fitness for use in laboratory Assembly and installation Network and data storage Secure data storage, backup, and archive Installation verification Instrument function Performance checks tests * Activities under each phase are usually performed as given in the table. However, in some cases, it may be more appropriate to perform or combine a given activity with another phase. Such activities spanning more than one qualification phase are shown as connected by double arrows. If an activity listed under a given phase is performed under another phase, it is not necessary to repeat the activity under the phase where the activity is listed. Performing the activity is far more important than the phase under which the activity is performed. tests, and quality control check samplesare not within the the selected environment, and that this environment is suit- scope of this chapter. able for the instrument. IQ applies to an instrument that is new or was pre-owned, or to any instrument that exists on site but has not been previously qualified. Relevant parts of Qualification Phases IQ would also apply to a qualified instrument that has been transported to another location or is being reinstalled for Instrument qualification is not a single continuous pro- other reasons, such as prolonged storage. The activities and cess, but instead results from several discrete activities. For documentation typically associated with IQ are as follows. convenience, these activities can be grouped into four DescriptionProvide a description of the instrument or phases: design qualification (DQ), installation qualification the collection of instrument components, including its man- (IQ), operational qualification (OQ), and performance quali- ufacturer, model, serial number, software version, and loca- fication (PQ). tion. Use drawings and flow charts where appropriate. Some AIQ activities cover more than one qualification Instrument DeliveryEnsure that the instrument, phase, and analysts potentially could perform them during software, manuals, supplies, and any other instrument ac- more than one of the phases (see Table 1). However, in cessories arrive as specified in the purchase order and that many instances there is need for specific order to the AIQ they are undamaged. For a pre-owned or existing instru- activities; for example, installation qualification must occur ment, manuals and documentation should be obtained. first in order to initiate other qualification activities. The AIQ activities will be defined and documented. Utilities/Facility/EnvironmentVerify that the installa- tion site satisfactorily meets manufacturer-specified environ- mental requirements. DESIGN QUALIFICATION Assembly and InstallationAssemble and install the in- strument, and perform any preliminary diagnostics and test- Design qualification (DQ) is the documented collection of ing. Assembly and installation may be done by the manu- activities that define the functional and operational specifica- facturer, vendor, specialized engineers, or qualified in-house tions of the instrument and criteria for selection of the ven- personnel. Manufacturer-established installation tests and dor, based on the intended purpose of the instrument. De- guides provide a valuable baseline reference for determining sign qualification (DQ) may be performed not only by the instrument acceptance. Any abnormal event observed dur- instrument developer or manufacturer but also may be per- ing assembly and installation merits documenting. Installa- formed by the user. The manufacturer is generally responsi- tion packages purchased from the manufacturer or the ven- ble for robust design and maintaining information describ- dor may, however, need to be supplemented with user- ing how the analytical instrument is manufactured (design specific criteria. specifications, functional requirements, etc.) and tested Network and Data StorageSome analytical systems before shipment to users. Nonetheless, the user should en- require users to provide network connections and data stor- sure that commercial off-the-shelf (COTS) instruments are age capabilities at the installation site. When required, con- suitable for their intended application and that the manufac- nect the instrument to the network, and check its function- turer has adopted a quality system that provides for reliable ality. equipment. Users should also determine the manufacturers capability for support installation, services, and training. This Installation VerificationPerform the initial diagnostics determination might be aided by the users previous interac- and testing of the instrument after installation. tion with the manufacturer. OPERATIONAL QUALIFICATION INSTALLATION QUALIFICATION After a successful IQ, the instrument is ready for OQ test- Installation qualification (IQ) is the documented collection ing. Operational qualification (OQ) is the documented col- of activities necessary to establish that an instrument is de- lection of activities necessary to demonstrate that an instru- livered as designed and specified, and is properly installed in ment will function according to its operational specification USP 36 General Information / 1058 Analytical Instrument Qualification 3 in the selected environment. Testing activities in the OQ ment is used. It may also be scheduled for regular intervals. phase may consist of these test parameters. Experience with the instrument can influence this decision. It may be useful to repeat the same PQ tests each time the Fixed ParametersThese tests measure the instruments instrument is used so that a history of the instruments per- nonchanging parameters such as length, height, weight, formance can be compiled. Alternatively, the instrument voltage inputs, acceptable pressures, and loads. If the manu- may be incorporated into an integrated support system to facturer-supplied specifications for these parameters satisfy assure that it remains continually qualified. Some system the user, the test requirements may be waived. However, if suitability tests or quality control checks that are performed the user wants to confirm the parameters, testing can be concurrently with the test samples also imply that the in- performed at the users site. Fixed parameters do not strument is performing suitably. change over the life of the instrument, and therefore never need redetermination. [NOTEThese tests could also be per- Preventive Maintenance and RepairsWhen an instru- formed during the IQ phase (see Table 1); if so, fixed pa- ment fails to meet PQ test specifications, it requires mainte- rameters need not be redetermined as part of OQ testing.] nance or repair. A periodic preventive maintenance may also be recommended for many instruments. The relevant PQ Secure Data Storage, Backup, and ArchivingWhen test(s) should be repeated after the needed maintenance or applicable, test secure data handling such as storage, repair to ensure that the instrument remains qualified. backup, audit trails, and archiving at the users site accord- ing to written procedures. Practices for Operation, Calibration, Maintenance, and Change ControlEstablish practices to maintain and cali- Instrument Function TestsInstrument functions re- brate the instrument. Each maintenance and calibration ac- quired by the user should be tested to verify that the instru- tivity should be documented. ment operates as intended by the manufacturer. Manufac- turer-supplied information is useful in identifying specifications for these parameters and in designing tests to ROLES AND RESPONSIBILITIES evaluate the identified parameters. Users, or their qualified designees, should perform these tests to verify that the in- strument meets manufacturer or user specifications in the users environment. Users The extent of OQ testing that an instrument undergoes depends on its intended applications. Therefore, no specific Users are ultimately responsible for instrument operations OQ tests for any instrument or application are offered in and data quality. The users group encompasses analysts, this chapter. their supervisors, instrument specialists, and organization Routine analytical tests do not constitute OQ testing. OQ management. Users should be adequately trained in the in- tests are specifically designed to verify the instruments op- struments use, and their training records should be main- eration according to specifications in the users environ- tained as required by the regulations. ment, and repeating the testing at regular intervals may not Users should also be responsible for qualifying their instru- be required. However, when the instrument undergoes ma- ments because their training and expertise in the use of jor repairs or modifications, relevant OQ and/or PQ tests instruments make them the best-qualified group to design should be repeated to verify whether the instrument contin- the instrument test(s) and specification(s) necessary for suc- ues to operate satisfactorily. If an instrument is moved to cessful AIQ. Consultants, equipment manufacturer or ven- another location, an assessment should be made of what, if dors, validation specialists, and quality assurance (QA) per- any, OQ test should be repeated. sonnel can advise and assist as needed, but the final OQ tests can be modular or holistic. Modular testing of responsibility for qualifying instruments lies with the users. individual components of a system may facilitate interchang- The users must also maintain the instrument in a qualified ing of such components without requalification. Holistic state by routinely performing PQ. tests, which involve the entire system, are also acceptable. Quality Unit PERFORMANCE QUALIFICATION The role of the Quality Unit in AIQ remains the same as for any other regulated activity. Quality personnel are re- Performance qualification (PQ) is the documented collec- sponsible for assuring that the AIQ process meets compli- tion of activities necessary to demonstrate that an instru- ance requirements, that processes are being followed, and ment consistently performs according to the specifications that the intended use of the equipment is supported by defined by the user, and is appropriate for the intended use. valid and documented data. After IQ and OQ have been performed, the instruments continued suitability for its intended use is demonstrated through performance qualification. The PQ phase may in- Manufacturers clude the following parameters. Performance ChecksSet up a test or series of tests to Manufacturers and developers are responsible for DQ verify the acceptable performance of the instrument for its when designing the instrument. They are also responsible intended use. PQ tests are usually based on the instruments for validation of relevant processes used in manufacturing typical on-site applications and may consist of analyzing and assembly of the instrument. Manufacturers should test known components or standards. The tests should be based the assembled instruments before shipping them to users. on good science and reflect the general intended use of the Finally, it is desirable that manufacturers and vendors instrument. Some system suitability tests or quality control should notify all known users about hardware defects dis- checks that are performed concurrently with the test sam- covered after a products release; offer user training, service, ples can be used to demonstrate that the instrument is per- repair, and installation support; and invite user audits as forming suitably. PQ tests may resemble those performed necessary. during OQ, but the specifications for their results may be set differently if required. Nevertheless, user specifications for PQ tests should demonstrate trouble-free instrument op- SOFTWARE VALIDATION eration for the intended applications. As is the case with OQ testing, PQ tests may be modular or holistic. Software used for analytical work can be classified into Testing frequency depends on the ruggedness of the in- three categories: firmware; instrument control, data acquisi- strument and the criticality of the tests performed. Testing tion, and processing software; and stand-alone software. Al- may be unscheduledfor example, each time the instru- though software validation is not the primary focus of this 4 1058 Analytical Instrument Qualification / General Information USP 36 chapter, the following sections describe in which cases this addition as a result of the installed change. Where the activity is under the scope of the analytical instrument change calls for additions, deletions, or revisions to the OQ qualification. or PQ tests, follow the procedure outlined below. Operational QualificationRevise OQ tests as necessi- tated by the change. Perform the relevant tests affected by Firmware the change. This ensures the instruments effective operation after the change is installed. Computerized analytical instruments contain integrated Performance QualificationRevise PQ tests as necessi- chips with low-level software (firmware). Such instruments tated by the change. Perform the PQ testing after installa- will not function without properly operating firmware, and tion of the change if similar testing is not already performed users generally cannot alter firmware design or function. during OQ. In the future, perform the revised PQ testing. Firmware is therefore considered a component of the instru- For changes to firmware and to software for instrument ment itself. Indeed, the qualification of hardware is not pos- control, data acquisition, and processing, change control is sible without operating it via its firmware. Thus, when the performed through DQ/IQ/OQ/PQ of the affected instru- hardware (that is, the analytical instrument) is qualified at ment. Change control for stand-alone software requires the users site, the integrated firmware is also essentially user-site testing of changed functionality. qualified. No separate on-site qualification of the firmware is needed. Whenever possible, the firmware version should be recorded as part of the IQ activities. Any changes made to AIQ DOCUMENTATION firmware versions should be tracked through change control of the instrument (see Change Control, below). Documents obtained during instrument qualification should be retained in an accessible manner. Where multiple Instrument Control, Data Acquisition, and instruments of one kind exist, documents common to all instruments and documents specific to an instrument may Processing Software be stored separately. During change control, additional doc- uments may supplement those obtained during the qualifi- Software for instrument control, data acquisition, and pro- cation process, and both sets of documents should be re- cessing for many of todays computerized instruments is tained and maintained in a suitable manner that allows for loaded on a computer connected to the instrument. Opera- appropriate protection and access. tion of the instrument is then controlled via the software, leaving fewer operating controls on the instrument. Also, the software is needed for data acquisition and postacquisi- INSTRUMENT CATEGORIES tion calculations. Thus, both hardware and software, their functions inextricably intertwined, are critical to providing Modern laboratories typically include a suite of instru- analytical results. ments and equipment varying from simple nitrogen evapo- The manufacturer should perform DQ, validate this rators to complex automated instruments. Therefore, apply- software, and provide users with a summary of validation. ing a single set of principles to qualifying such dissimilar At the user site, holistic qualification, which involves the en- instruments would be scientifically inappropriate. Users are tire instrument and software system, is more efficient than most capable of establishing the level of qualification modular validation of the software alone. Thus, the user needed for an instrument. On the basis of the level needed, qualifies the instrument control, data acquisition, and pro- it is convenient to categorize instruments into three groups: cessing software by qualifying the instrument according to A, B, and C, as defined below. Examples of instruments in the AIQ process. each group are provided. Note that the list of instruments provided here is for illustration only and is not meant to be Stand-Alone Software exhaustive. It does not provide the exact category for an instrument at a user site. That category should be deter- An authoritative guide for validating stand-alone software, mined by users for their specific instruments or applications. such as LIMS, is available. 1 The validation process is adminis- The exact grouping of an instrument must be determined tered by the software developer, who also specifies the de- by users for their specific requirements. Depending on indi- velopment model appropriate for the software. Validation vidual user requirements, the same instrument may appro- takes place in a series of activities planned and executed priately fall into one group for one user and another group through various stages of the development cycle. for another user. Therefore, a careful selection of groups by users is highly encouraged. CHANGE CONTROL Group A Changes to instruments, including software, become inev- Group A includes standard equipment with no measure- itable as manufacturers add new features and correct known ment capability or usual requirement for calibration, where defects. However, implementing all such changes may not the manufacturers specification of basic functionality is ac- always benefit users. Users should therefore adopt changes cepted as user requirements. Conformance of Group A they deem useful or necessary and should also assess the equipment with user requirements may be verified and doc- effects of changes to determine what, if any, requalification umented through visual observation of its operation. Exam- is required. The change control process enables them to do ples of equipment in this group are nitrogen evaporators, this. magnetic stirrers, vortex mixers, and centrifuges. Change control may follow the DQ/IQ/OQ/PQ classifica- tion process. For DQ, evaluate the changed parameters, and determine whether need for the change warrants imple- Group B menting it. If implementation of the change is needed, in- stall the changes to the system during IQ. Evaluate which of Group B includes standard equipment and instruments the existing OQ and PQ tests need revision, deletion, or providing measured values as well as equipment controlling 1 General Principles of Software Validation: Final Guidance for Industry and FDA physical parameters (such as temperature, pressure, or flow) Staff, U.S. Department of Health and Human Services, Food and Drug Admin- that need calibration, where the user requirements are typi- istration, Rockville, MD, January 11, 2002. http://www.fda.gov/Medi- cally the same as the manufacturers specification of func- calDevices/DeviceRegulationandGuidance/GuidanceDocuments/ucm085281. tionality and operational limits. Conformance of Group B htm (accessed December 2012. instruments or equipment to user requirements is deter- USP 36 General Information / 1058 Analytical Instrument Qualification 5 mined according to the standard operating procedures for atomic absorption spectrometers the instrument or equipment, and documented during IQ differential scanning calorimeters and OQ. Examples of instruments in this group are bal- dissolution apparatus ances, melting point apparatus, light microscopes, pH me- electron microscopes ters, variable pipets, refractometers, thermometers, titrators, flame absorption spectrometers and viscometers. Examples of equipment in this group are high-pressure liquid chromatographs muffle furnaces, ovens, refrigerator-freezers, water baths, mass spectrometers pumps, and dilutors. microplate readers thermal gravimetric analyzers X-ray fluorescence spectrometers Group C X-ray powder diffractometers densitometers Group C includes instruments and computerized analytical diode-array detectors systems, where user requirements for functionality, opera- elemental analyzers tional, and performance limits are specific for the analytical gas chromatographs application. Conformance of Group C instruments to user IR spectrometers requirements is determined by specific function tests and near-IR spectrometers performance tests. Installing these instruments can be a Raman spectrometers complicated undertaking and may require the assistance of UV/Vis spectrometers specialists. A full qualification process, as outlined in this inductively coupled plasma-emission spectrometers document, should apply to these instruments. Examples of instruments in this group include the following: