You are on page 1of 128

SERVICE MANUAL

Electrocardiograph
FX-2111
4R2043

FUKUDA DEIMSHI CO., LTD.

Copyright 1997 by Fukuda Denshi Co., Ltd.


No part of this document may be copied
or transmitted in any form without

the prior written permission of Fukuda


Denshi Co., Ltd.

Manufactured in Japan

Foreword

We list up here the warning marks used in Fukuda operation and


service manuals.

When you sen/ice the FX-2111, read this service manual thoroughly and pay attention especially to instructions bearing the following marks:

Warning Marks
Warning marks used in operation and service manuals and labelled
on the instrument have the following meanings:
Read them carefully to understand the meanings and make sure of
the significance of each particular.

This mark is used to indicate the direct hazards

Danger which may lead to the death or serious injury of the


person, may wholly damage the instrument, or may
cause fire hazard, unless the instructions written
there are followed.

This mark is used to indicate the indirect potential

Warning hazards which may lead to the death or serious injury


of the person, may wholly damage the instrument
or may cause fire hazard, unless the instructions
written there are followed.

This mark is used to indicate the possible hazards

Caution which may lead to a mild or medium injury of the


person, may partially damage the instrument or may
erase data from the computer.

NOTE

"NOTE" is not warning instructions but offers information to prevent the person from doing erroneous
sen/icing.

Other Marks

Notice to indicate general unspecific prohibited


matters.

Notice to indicate general unspecific caution, warning or hazard.

This service manual describes technical information on FX-2111 to

aid the service engineer in troubleshooting.


The manual is intended to be used by service engineers of Fukuda
representatives and authorized technical staff concerned with medical electronic equipment. Description includes repairing and assembling methods
of each component unit of FX-2111. For parts lists and diagrams, refer to
the Part II of the service manual.

The service manual consists of the following nine chapters:


1

General Description
The outline of FX-2111, specifications, controls and indicators are
described.

Circuit Description
Circuit configuration and functions are explained.

Troubleshooting

Troubles vs. causes and countermeasures are described.


4 Maintenance
.

Procedures to replace the power fuse, ROM and battery and to perform
self-test are described.
5

Periodical Inspection
Inspection procedures to prevent troubles and ensure safe and complete operation of the instrument are described.

Circuit Diagrams
Assembly Diagrams

8 Electrical Parts Lists


.

Structural Diagrams

A Caution

. Never remodel Fukuda medical electronic equipment


. The service manual

is intended for the service

engineers of Fukuda representatives and the technical staff concerned with medical electronic equipment. Servicing, reassembling, and adjustment shall
be performed by authorized service engineers.
. Prepare proper facilities and tools when servicing
. Be sure to follow the instructions of operation manual
when operating the instrument. For operating precautions, refer to the operation manual.
.

ii

A Servicing Precautions
A Cautions listed below are the instructions of prohibit danger, warn,

ing, and caution described in this service manual. When taking the
procedure bearing the following mark, read the description thoroughly,
then start the task.

A Caution

(page 3-2)

When checking the power supply and related circuits for troubleshooting, take sufficient care to avoid a short circuit.

A Caution

(page 3-3)

When checking power fuses, be sure to turn the power off and disconnect the power plug from the wall outlet beforehand.

A Caution

(page 3-8)

The inserting part of key connector is made of carbon. Avoid frequent


repeated disconnection and connection.

A Caution

(page 4-1)

When replacing power fuses, be sure to turn the power off and disconnect the power plug from the wall outlet beforehand.

A Caution

(page 4-1)

When replacing the ROM, be sure to turn the power off. Also, take care
to install the ROM in correct position.

A Caution

(page 4-2)

When replacing the battery, be sure to turn the power off and disconnect the power plug from the wall outlet beforehand.

iii

A Caution

(page 4-6)

Disassembling/Reassembling Precautions
Be sure to disconnect the power cord and make sure the instrument
is turned off before disassembling or reassembling.
. Do not remove the battery before disconnect any PC board

Take care that repeated disconnection of the key panel and sensor
board may result in poor contact.
Use proper tools to loosen screws.

. When reassembling make sure that all screws are securely tightened and all connectors are completely inserted.
,

PC Board Handling Precautions


. PC boards are equipped with highly sensitive components to static
electricity.
. PC boards are highly sensitive electronic devices Put removed PC
boards in a proper protective bag or take appropriate measures to
protect them.
.

Handle PC boards carefully. A shock to them may damage the components.


Never insert a connector to the powered PC board nor remove the
powered PC board.

A Caution

(page 5-2)

If you find a value which exceeds the allowable level, be sure to let
the user avoid using the FX-2111. If the user operates the FX-2111 as
it is, he/she may receive an hazardous accident.

Equipment Classification
The FCP-2155 is classified into the following equipment:
1

Protection against electrical shock

Class I
2

Type against electrical shock


Applied part: Type CF

Degree of protection against harmful water invasion


Other equipment

Degree of safety in using under air-inflammable anesthetic gases or


oxygen/nitrous oxide-anesthetic gases
Equipment used under an environment containing no inflammable
anesthetic gases or no inflammable cleaning agent.

Running mode
Continuous running mode
iv

CHAPTER

General Information
1 Outline of the FX-2111
.

1-1

Specifications
2 1 Electrocardiograph Section

1-1
1-1

2 2 General

1-2

23

1-2

Environmental Conditions

3 Controls and Indicators


.

1-3

1 Top Panel
3 2 Side Panels, Left and Right

1-4
1-4

33

1-4

Bottom Panel

1 Outline of the FX-2111


.

The FX-2111 is the easy-to-use single-channel electrocardiograph


featuring a simple operation panel. The compact yet high-performance
design makes it suitable not only for use in the consultation room but
also for-carrying in a visit to the hospital ward or patient s home and
examination at an emergency site.
'

It has a high-density thermal dot printer incorporated to provide clear


ECG recording. The FX-2111 can operate on either AC line or rechargeable Ni-MH battery.

2
2

Specifications

1 Electrocardiograph Section
Input circuit:

Floating from the ground

Leads:

Standard and Cabrera 12 leads

Input impedance:
Input circuit current:
Calibration voltage:

20MQ min. (referred to 10Hz)


5x10"8 max.
1mV within 5%

Common mode rejection: 10mm or less per IEC 62DC06 test


method

Polarization voltage:

300mV min.

Time constant:

3 2 sec min.

Sensitivities:

1/2, 1 and 2 cm/mV

Frequency response:

0 05

AC filter:

50/60Hz, -20dB or lower

Muscle filter:

35 to 45Hz, -3dB (-6dB/oct)


10mm/0.5V, unbalanced, 100kfi min.
Liquid-crystal display, 20 chars, x 2 lines
(character: 5x7 dots)
Thermal dot printer, 8 dots/mm

DC input:
Display:
Recording system:
Paper speeds:
Chart papersRoll paper:
Z-fold paper:

to 150Hz (within -3dB)

25 and 50 mm/sec within 3%

63mm or 50mm wide x 30m long


63mm or 50mm wide x 20m long,
75mm/fold

AID conversion:

12 bits

Sampling rate:

1ms

1-1

2 2 General
.

Safety:

I EC 601-1 Class I

.Type CF
. Internally powered equipment Type
CF (IEC 601-1)
Power requirementsAC operation:
DC operation:

Dimensions:

Weight:

115V AC, 50/60H2; 19VA max.


230V AC, 50/60Hz; 19VA max.
9 6V, 7W max. (rechargeable Ni-MH
battery)
Continuous operation: approx. 120 min
at 20oC (according to IEC 62D Testing
Method)
Charging time: within 3 hours
26 (W) x 18.2(D) x 6.3(H) cm
Approx. 1.7kg (excluding battery)
.

2 3 Environmental Conditions
.

OperatingTemperature:
10 to 40oC
Humidity:
30 to 80%RH (no dew condensation)
Atmospheric pressure: 70 to 106kPa
Transportation & StorageTemperature:
-10 to 40oC
Humidity:
10 to 95%RH (no dew condensation)
Atmospheric pressure: 70 to 106kPa

1-2

3 Controls and Indicators


.

Top Panel

X Liquid-crystal Display

31

@ Paper Magazine

(D Operation Panel

Left Panel

Right Panel

Potential Equalization
Terminal
0

Power Connector

F3

Main Power Switch

@ DC Input Terminal

Lead Connector

Bottom Panel

o
{
*

i<

"

iii
0

Fuse Holders

)
~

1-3

Rechargeable feattery Room

1 Top Panel

Display:

Indicates current status such as recording mode, lead, sensitivity, heart rate, or program setting.

Operation Panel:
ON
OFF

Mode

Turns the FX-2111 on in DC operation.


Turns the FX-2111 off in DC operation.
Pressing this key during AC operation lets the FX-2111 enter the
charge mode.
Select a mode in the following order when pressed during cessation of recording:
|-> Automatic recording - Manual recording -> Programming -|
Holding this key during automatic recording lets the FX-2111 record
the current lead continuously until detached.
Holding this key during manual recording lets the FX-2111 record
an event mark until detached.

Sensitivity
Lead Select <
Reset

1mV
Start/Stop

Paper Magazine:
3

Selects a recording sensitivity.


Select a lead for recording. In programming mode, these keys allow
you to select a setting value.
Resets the measuring circuit while held during recording. In
programming mode, this key selects a setting parameter.
Applies a 1mV calibration waveform when pressed during recording. In programming mode, this key selects a setting parameter.
Starts the FX-2111 recording the ECG waveform. Another press stops
it from recording.
Accommodates a chart paper.

2 Side Panels, Left and Right

DC Input:

Inputs external DC signals with a sensitivity of 10mm/0.5V.

Lead Connector

Connects to the lead cable.

Power Connector

Connects to the power cable.

Main Power Switch:

Turns AC power on/off.

@ Potential Equalization
Terminal:

Makes the FX-2111 equipotential to another instrument used in combination with it. To that effect, connect both instruments to a

common grounding conductor using an optional grounding wire.


3 3 Bottom Panel
.

Battery Room:

Houses the rechargeable Ni-MH battery.

Fuse Holders:

Have power fuses inserted.

1-4

CHAPTER

Circuit Description
1

Introduction

2-1

Isolated Input Circuit


2 1 Buffer Amplifier and RF Driver

2-2
2-2

2 Lead Network and Lead Selector

2-3

3 Preamplifier and 1mV Generator

2-3

4 R-wave Detector and Overinput Detector


5 Amplifier Control
2 6 Signal Isolator

2-4
2-5
2-5

27

2-6

Power Isolator

Middle Amplifier and AID Converter


1 Middle Amplifier and DC Input

2-6
2-6

3 2 AID Converter

2-7
2-8
2-8

4 Motor Control
5 Sensor Circuit
.

1 Detection of Magazine Open Condition


5 2 Detection of Paper End and Paper Marks
5

6 CPU Circuit

2-9
2-9

6 2 CPU, ROM and RAM

2-9
2-9
2-9

63

2-9

61
.

Reset Circuit

Gate Array

LCD Control Circuit


8 Thermal Print Head Control Circuit

2-11
2-11

Memory Backup Circuit


10. Power Supply and Charging Circuit

2-12
2-13

10.1 Introduction

2-13

10.2 Rectifier/Smoothing Unit and ON/OFF Control

2-13

10.3 + 10V and +5Vd Power Generators


10.4 5Va Power Generator
10.5 +24V Power Generator

2-14
2-15
2-15

10.6 Charging Circuit

2-16

1 Introduction
.

The FX-2111 is composed of the following circuits:


(1) Main Board PCB-6409
-Isolated input circuit
-Middle amplifier and AID converter circuit
-Motor control circuit
-Sensor circuit
-Reset circuit

-CPU circuit (CPU ROM, RAM, gate array)


,

-LCD control circuit

-Thermal print head control circuit


-Memory backup circuit
-Power supply and charging circuit
(2) Sensor Board PCB-6239
(3) AC Inlet Board PCB-6410

ECG signals input through the lead connector are amplified by the
isolated input circuit and the middle amplifier circuit, then converted from
analog to digital signals by the AID converter circuit. The digitized signals
are digitally filtered, if the filter is set to ON, then sent through the gate
array to the thermal print head control so that they are recorded as ECG
waveform by the thermal print head.
During these processes, the heart rate detected by a hardware
technique and operation status such as filter ON/OFF setting are displayed
on the LCD.

The power supply provides circuits with necessary powers. The


charging circuit, if activated, charges the Ni-MH battery in AC operation
or transfers the power from the Ni-MH battery to the power supply. The
state of battery is indicated on the LCD.
:

PCB-6409
i5V

AC Inlet

I0V

"

re

Isolated Input Circuit

If

Power Supply

-24V
AC Power

VDO
+2 5V
.

ZEE
'

'

DC

Input

/Charging;-;

AyiCircuil ;--

Motor

Motor

Control

Thermal

Thermal ]
Print

Head

LCD

*
.

5 M mt

:'

J Gate Array

CPU Circuit

Print it
Head
[
Control

LCD
Control

PCBSensor
Board

Backup
Circuit

6239

Overall Block Diagram


2-1

NiMH

Battery

DC

:pcb-64io

Key Panel

Isolated Input Circuit


The isolated input circuit is composed of the following:
(1)
(2)
(3)
(4)
(5)
(6)
(7)

Buffer amplifier and RF driver


Lead network and lead selector
Preamplifier and 1mV generator
R-wave detector and overinput detector
Amplifier control
Signal isolator
Power isolator

L
F

CI

D
5

5 5
t a5

"

ECG
-

1
5

D
-

I 2

C6

Preamplifier

"

> Signal
Isolator

I
1mV

RESET

Amplifier

T
RF Drive <-

"

Control

Signal
Isolator

RF

Overinput/

(Photo

R-Wave Detector
Circuit

Coupler)

I
i
i

.e-

Power

Isolator <-

Block Diagram of Isolated Input Circuit

1 Buffer Amplifier and RF Driver

oac- Hit icr

I
0

9 ISO*

5
Since the buffer amplifier should input signals with high impedance
and output them with low impedance, it is configured as an impedance
conversion circuit using an operational amplifier. The circuit shown above
is individually applied to all leads except for RF (right foot). Also, a limiter
using a dual transistor is mounted for protection against overinput.
The RF driver feeds back the composite signal of limb leads to the

right foot for improved common mode rejection ratio.


2-2

2 2 Lead Network and Lead Selector


.

OK

211
OK

*a?

z 10

Ox

> -

304

OK

20 1

0*

Lead Select Signals


It:?
?

ft:

ior

Leads

BCD

STD

10

4;

Ok

II

10

III

110

Vr

200

0-<

Z0
OK

El

10*

Tan

m
l OX

10

aVL

10

10

aVF

VI

1110

V2

'

UNS2 L]

110
0

V3

10

V4

10

V5

110

V6

11

The lead network is formed with resistors and lead selector, with multi-

plexors IC36, IC38, IC41 and IC4. Input signals to multiplexors are selected
from RA, LA, LF, and CI to C6 according to four signals of A, B, C, and
D then synthesized to produce each ECG lead (see table above).
,

3 Preamplifier and 1mV Generator

C-1

fcTzOOHK

o.aou

r-II
"V

ci

cb>

in*

rc

.200

10m

JOK

MOK <

4= " [ F

.on

-*-

J frt5
i

|Slr4ii

(C J7I

cm. J
o

CCC

*.

OIL

1SS

. mi

117 1

1*3

1 . IK
t

3C on*

r mioi

5:

icia

41J

LIt004C$l-i

rc

1 Preamplifier
Signals output from the lead selector (multiplexors) are amplified 8
times by a differential amplifier, then 25 times by a noninverting amplifier,
thereby letting the isolated input circuit amplify the input signals by 200
times in total. Signals output at the first stage are sent to the overinput
23
.

detector.
2-3

The 3.5Mfi resistor and 1|jF capacitor set the time constant at 3.5
seconds.

2 3 2 1mV Generator
.

The 1mV generator divides the high-precision voltage reference output of 1.235V by a high-precision resistor, thereby applying a 8mV voltage
to the preamplifier when the 1mV application signal is at a low level.
2

4 R-wave Detector and Overinput Detector


era

Q.Qit

Li

C2

[C4 1C

J
*

in

*
.

>
10QN

4*2

U*f 111

I 4M

lOOH

nrvr-!>

0 23
.

JOCK
c 111

1J ..

CP
ft I

a*]

UMf I LI

r-

C 42
0 001 J
.

C12 J

"I

20OK
re

2 41
.

R-wave Detector

RA and LF signals of lead network are synthesized to II lead. The Rwave detector detects R wave by picking up R-wave component from the
II lead signal through a band-pass filter, then sending it to a comparator.
24
.

2 Overinput Detector
The signal output at the first stage of preamplifier is sent to a com-

parator and if the signal exceeds 360mV, the output of the comparator
becomes inverted.

Signals output from the R-wave detector and overinput detector are
adjusted in the pulse width by the multivibrator IC40 (TC4538), then input
into the photo coupler.
+2 9V
.

+0 31V
.

OV

ov

1T63

0 31V
.

2 9V
.

TP19

* 170S

TP20

ITOins

R-wave Detection Wavelorm

Overinput Detection Waveform


2-4

5 Amplifier Control

"

C S . IK

<5id*

2
n

of3

Si a

IT

. 2 17
IK

HJU

XL
lO.Ot
-

v 0

CO

This circuit picks up a reset, 1mV,or lead select signal from the serially
transferred amplifier control signals.
!<-% 1ms

;
I

1 KH 2
AMP

C L K

AM P

DATA

DO

D7

D6

D5

D4

D3

D2

Dl

DO

D7

D6

D5

D4

D3

D2

DL

DO

D7

DO

i
i

OLD DATA <-

> NEW DATA

i
i

8ms

Data wfitlen (rom CPU are latched.

0
a 0 8ms

Response
K
time: 32ms + a max.
_

16ms + a mm.

6 Signal Isolator
CT*

IC21 rjJiSJLp-Iilp

t|

o.c
0+?
.

am

<fl rCLK H

U U

-cTTirClK 1

C 73

lM

O
frwrr

ICJW y5)*3L.
l JLj-l1|

IK

mil

Ik

--W

c
-;

>r i

111 *
-

12

LTT-

T1 en

CE3d>

41

0 01
.

aiir

___

p y j- v -- -tret? >
O.Ol

o.ot

BIO-

rLji_.
a) i

Cr-

ISI3)

sr:

The signal isolator modulates input analog ECG signals with FETs
Q31 and Q32 and pulses TRF1 + and TRF1 -, then transfers the modulated

signals through the T1 isolation transformer (1:1). FETs Q22 and Q23
demodulate the signals. A low-pass filter formed with capacitors and
resistors in two stages eliminates noise in the demodulated signals.

Digital signals (AMPCLK, AMPDATA and HR) are transferred using a photo
coupler.
2-5

2 7 Power Isolator
.

-020
<ianz>

11

<3BJ-pW-i
0

ISV

*fsl\v

iSiiQl

-r-

30

'

JC 10 *
i

'2

JH

5p

TVrr-J ..7,11V

11

ISSIOJ

<2irD--f-1 J-

ilO

-r-

*a

lisjoj

0 . 1

C 10
200xhi

The power is transferred using 10V pulse at 100kHz. At the isolated


side, a 3-terminal regulator supplies 5V.

3
3

Middle Amplifier and AID Converter

1 Middle Amplifier and DC Input


C4 T

:v.\]

at
i

3;
*

imrraT>

1 10

.0.2

il-Lt
irrw

LA

21

>
"

10. ?

...

ECG signals demodulated at the isolated signal transmission section


are output to the middle amplifier just after passing through a low-pass
filter formed with capacitors and resistors in two stages. The middle amplifier amplifies the input signals 1.25 times to a total gain of 250 times
and sends them to the AID converter. Signals output from the AID converter
are digitally filtered with a software technique. The isolated input circuit
and the middle amplifier are designed so as to provide a general hardware frequency response of 150Hz (-3dB).
The DC input amplifies input signals 0.5 times, then sends the signals
to the AID converter through a buffer amplifier.

2-6

3 2 AID Converter
.

icu.

...

it:?
DC

Li

J .

100 T
* 0.022

'

[l 01 l

"

I -3

'

t<S302

Tin

100

Tl

4 JQ

1C4l4
I|l 010*
ittlQI

0 O22
.

1 C*1

t 1

HP

100

1300f

,1

El00"

vvv-r

0110

w-

-<}
a it

The A/D converter is a single slope type. It is composed of a triangular


wave generator, sample & hold circuit, and comparator. The comparator
compares input signals with 3.2\/ triangle wave generated at every 1ms.
The signals are then converted from the voltage to a pulse width and transferred to the gate array. In the gate array, the pulse width is counted, then
converted into a digital value.
520ps

<

5k

Si

480,1s

A/D CLK

--<?

>

96ps

;<

96)13

S/H

(TP35)

Bps
+3 2V
.

OV

Triangular
(TP7)

3 2V

Start ol counting
A/D Conversion Timing

Besides ECG signals, the AID converter digitizes DC input signals,


battery voltage, thermistor signals of thermal print head, and the reference voltage of 2.5V for calibration of A/D conversion.

2-7

4 Motor Control
.

</00

. LJJ
s nn?

>

(IT

lOQ-

LI

11
-

C*

VCOIm

- -

i
Li

C3

Li
11%
_

Li
11

Li

* t J*

10*
_

itsio:

. T jiv ;

\ Ten
14 1

I C 1

The motor to drive the recorder is the DC motor which has a photo
sensor built in for detection of motor speed. The PLL control IC25 compares a signal detected by the photo sensor with the reference frequency
in phase, thereby providing a motor control signal at pin 13 of IC25.
The motor control signal goes to the integrator circuit formed with
R140, R142, and C64 and is made by a proper motor drive voltage of the
PWM control IC24.
25mm/s

(256Hz)

SOmra/s (5L2Hz)
r i

r~i

-+5V

IC25 (pin 14)


SIN
-

ov

+5V

IC25 (pin 3)
CIN

.OV

.+5V

IC25 (pin 13)

. +3 IV
.

P2

Delayed

Early

+2 2V
.

\
Delayed

/
Delayed

OV
Early

Delayed

5 Sensor Circuit
.

The sensor circuit is provided to check for a magazine opening


condition and paper end as well as detecting paper marks.
1*4
0

fi2E2>

T
t la
7

1
lit

:
>

il

3
i &
<1

if*
-

llflOl

2-8

4 n*

re it

-H> -

t*ii

51
.

Detection of Magazine Open Condition

When the paper magazine is open, the microswitch SPVC2-1 is turned


off to send an interrupt signal to the gate array. Magazine opening is detected with the fall signal and magazie closed is detected with the rise signal.
52
.

Detection of Paper End and Paper Marks

A reflection type photo interrupter acknowledges a black paper mark


or paper end status (no paper remaining) if receiving no reflection from the
location of paper. It discriminates between paper end status and paper mark
based on a duration of the low level starting from a fall signal.

6 CPU Circuit
.

The CPU circuit is composed of a reset circuit, CPU, ROM, RAM, gate
array, and backup power circuit.
61
.

Reset Circuit
VOO

1 14

10K

ION

vsa

vACSEt
*

ViB

12
!2M

VSC

OUIC

QhO

Cf

C%3
-

k cj

Q-i

L'ltV

M O T
.

Using NJM2103, the reset circuit generates a reset pulse as well as


monitoring the battery voltage. It outputs the reset pulse when Vdd falls
to lower than 4.27V or when it receives a signal from WDT.
As for the battery voltage, the reset circuit monitors the power voltage
divided by resistor and when the voltage becomes lower than 7.5V referred
to the battery terminal, it outputs a pulse to the LOWS terminal to turn
the power off. To protect the output pulse against power fluctuation noise,
a filter formed with a capacitor and resistor in one stage and a Schmidt
trigger inverter are provided.
6 2 CPU, ROM and RAM
.

The CPU is HD63B09E. It controls the overall circuit of the instrument

through the gate array. The ROM is 64K-byte AM27C512-150DC and the
RAM is 8K-byte SRM2264LM12. By changing the jumper connection, the
ROM can be replaced with a 128K-byte ROM and the RAM, with a 32Kbyte RAM (with a data holding current of lower than 2\iA).
6

3 Gate Array

The gate array is FD88007-AC with a clock frequency of 16MHz. Its


control signals are as follows:
2-9

631
.

2MHz E and Q Clocks for CPU


'

-5 00 n s-->

<

500ns

V,

$->, 125ns

63
.

2 Interrupt Signals

There are three types of interrupt signals-NMI, FIRQ and IRQ. All
these signals, active at low level, are output to the CPU.
-NMI
Output at every 1ms.
-FIRQ
Output at every 10ms.
-IRQ
Output when any one of MAGAZIN, MARK, HR, LOWB, CHG
IRQ and AC/DC signals is input.
,

!<

S>

Ims

NM I

j* 500jis

1 0 0 Hz
1 Oras

FIRQ
. CLEAR

MARK

RUN

PAPER
END

a = 390ms (9.75mm) with a paper speed of 25mm/sec


190ms (9.75mm) with a paper speed of 50mm/sec
NOTE: ACDC, CHG IRQ, and MAGAZIN DOWN signals are detected at
their rise and others are at their fall.
63
.

3 Watchdog Timer (WDT)


.

10ms

'

<

1 0 0Hz

WDT
CLEAR

70ras .

WDT

a = Response time of IC
2-10

a i

634
.

Chip Select Signals

Chip select signals are all active at low level.


WE: Write enable signal
RD: Read enable signal
E

ROM ]
RAM [
KEY J

1
1

|
1

WE I

RD J

'

'

'

NOTE: LCD signal differs from the above since the memory is ready. (See
LCD control circuit.)

7 LCD Control Circuit


.

Since the LCD (NDM202A00) has a controller built in, it is interfaced directly with the CPU. However, E and Q clocks make the bus timing as shown below.
E

LCD

Data
Bus

Address
Bus

CM
ZX

$3040 ~ $30<tF

>
X

8 Thermal Print Head Control Circuit


.

A waveform is traced at every 1ms and alphanumerics are at every


6ms with a paper speed of 25mm/sec or every 3ms with a paper speed
of 50mm/sec. The RAM for the thermal print head control uses 4K-byte
for alphanumeric data. This circuit controls dot heating temperatures based
on previously printed data to provide a proper printing condition.

When the paper is not driven or when the magazine is open or there
remains no paper in the magazine, the circuit generates TOFF signal to
shut off power supply to the thermal print head at the FET Q24, thereby
protecting the thermal print head.
There are four control signals of CLOCK, DATAIN, LATCH, and STROBE.
The STROBE signal is varied in pulse width according to thermal print
head temperatures.
2-11

1ms

LATCH

e-

8,3

STROBE

us

!<

960vs

(Max)
895,is

CLOCK

(SOOKHz)

9
<

5us
S

DATA

2v s
CLOCK
.

Sks

DATA

Memory Backup Circuit


The battery-backed RAM SRM2264LM12 is installed to keep program

contents when the FX-2111 is turned off. The RAM should feature a data
holding current of lower than 2nA.

2 k otci 1 . o>

O . I

T ."

5 "

When the FX-2111 is turned on, Vdd is supplied to the circuit and
turns Q6 and Q7 on to send +5\/ Vdd to the RAM.

When the FX-2111 is turned off, Q6 and Q7 are turned off but the back-

up lithium battery supplies +3V to the RAM to let it keep the stored data.

2-12

10. Power Supply and Charging Circuit


10.1

Introduction

The power supply transforms the AC power by the transformer or


receives DC power from the battery, then regulates the power to stabilized
voltages required by respective circuits.
The FX-2111 is equipped with a charging circuit to permit the user
to charge the Ni-MH battery.

AC Power

Transformer

Rectifier

+ 1 0 VN

Regulated
Power

Supplies
+ I 0 V

ON/OFF
Control

DC Power

+ 5 VD

Battery

5 VA

Charging

+ 2 4 V

Circuit

Block Diagram

10.2 Rectifier/Smoothing Unit and ON/OFF Control


10.2.1 AC Operation
The AC power voltage is transformed by the transformer, then rectified
by D1 to D4 and smoothed by C136. D1 to D4 are Schottky diodes with
low Vpand form a bridge circuit. This rectifier circuit suppresses voltage
decrease and features less heat generation.
10.2.2 ON/OFF Control

In battery operation, a press of the ON key switches FET 04 on and


electrifies D7, thereby making the base of 05 driven. If the operator
detaches the finger from the ON key, 04 is kept on and the power is
supplied to the FX-2111.

The battery voltage is detected via D6. If the voltage between battery
terminals is lower than 7.5V or so, the gate array generates POWER OFF
signal to turn 03 on and cancel the bias voltage of 05, thereby switching
04 off. As the result, the FX-2111 receives no power supply and is turned off.
Similarly, a press of the OFF key cancels the bias voltage of 05, thereby
turning the FX-2111 off. In AC operation, 01 is on; therefore the base of
05 cannot be driven and 04 cannot be kept on, thereby making battery
operation impossible.
2-13

VDO

AC DETECT

D 1 -

rpBHogy

-(sczeo

r i

'

C t 36

if"

H ' - >fouR

{>-'

0?

tJNS2 1 2

OS

D3FSA

in

or 3-2P - 20SA

gnr Tr

fr T-Tr -rrTinr
"

Shis

It

E32EHZ>

2(>

-O-

25
tOOK

HI7

20OK

<

azcz-S.iv

"

12

UMS212

K2-

<\

LSS301

0-

UNS2 1 Z

08

T P48

UNS2 12

{>
2fc
20K

EKZO

ear t

-> 0 o T I

10.3

*
-

+10V and +5Vd Power Generators

10.3.1 10V Power Generator

The dropper type 4-terminal regulator IC22 is used to regulate unstabilized IOVnU) stabilized lOV. R118 and R119 are to set the output voltage.

The 10V power is supplied to the motor and the 5Vd power generator.
If the battery voltage becomes lower than 10V in battery operation,
IC25 drops the voltage to several ten millivolts.
10.3.2 5Vd Power Generator

C7 is a drop type low-loss 3-terminal regulator. It produces regulated

5V power from 10V. IC7 can operate if an input-output voltage difference


becomes up to 0.5V and thus it can output the stable 5V power if the input
voltage fluctuates due to lowered battery voltage.
The 5V power thus produced is supplied to the CPU and other digital
circuits.

IPS*

IC22

IC7

POIQIV11

iovn

->

TP27

U-

msz

11?
3

C20
O.Ol

CIS

I
3 119

JO

gno

>

2-14

*
TP38

OFT

10.4

5Va Power Generator

Using the unregulated 10Vn power from the AC power supply or the
battery, the 3-terminal regulator ICQ produces +5V. IC10 is a CMOS inverting
type switching regulator, in which the output voltage is fixed and which
inverts the input +5V and outputs -5V. A capacitor at the output is an
organic semiconductor capacitor which features superior characteristics
at low temperatures and least leakage current.

5V powers thus produced are supplied to analog circuits such as


operational amplifier.
ICS

GEC

.Not hj n

C I J
D Ol

ZZ.

VIN

-VOUI

oT*
ir 44*

ss

C2

10.5

ft

>

+24V Power Generator

Using the unregulated 10Vn power from the AC power supply or the
battery, the +24V power generator, a boosting type DC-DC converter,
produces 24V power for driving the thermal print head and charging the
battery.

IC6 is a PWM-controlled switching regulator. R22 and C18 determine


the oscillation frequency. The PWM control is performed at approximately
170kHz to drive Q9 at a high speed. While Q9 is on, the current flows to
LI which in turn stores the energy. When Q9 is turned off, the stored energy
becomes a counter-electromotive force and is stored in C137 via D13. By
repeating a series of these operations, the +24V power generator can
produce an output voltage larger than the input voltage. R41 and R42 set
the output voltage at 24V. R24 checks for an overcurrent and a characteristic is provided to gradually lower the output voltage when the load
current increases to over 0.4A.

The 24V power thus boosted is used to drive the thermal print head.
When the thermal print head is printing, a signal from pin 21 of the gate
array drives the base of Q15 to switch FET Q24 on so that the -I-24V is
supplied to the thermal print head. When the thermal print head is not
printing, Q15 and Q24 are off and the thermal print head is not supplied
with power.

Pin 16 of IC6 outputs 2.5V, which is used by the AID converter and
the thermistor temperature detector of thermal print head.
I
-

11

;;

j1UY

i 111 * 51 i
= LIL

2-15

UJ

;v

..

i-0-<nm

10.6 Charging Circuit


The charging circuit is equipped with the quick charging IC18 (bq2003),
which performs constant current control as a frequency modulation controller for switching regulation of the charging current and monitors the
battery temperature and voltage and the charging time to ensure proper
charging. It also checks whether or not the Ni-MH battery is connected
and enables charging only when the battery is connected.
In AC operation, the circuit is supplementarily charging the battery
even if the FX-2111 is not placed in charging mode.

10.6.1 Starting the Charging Circuit


When the OFF key is pressed in AC operation, Q11 is turned on to
supply +5V for operation of the IC. Since the base of Q12 is driven by
turning Q11 on, Q11 is kept on. When +5V is supplied to IC18, the charging circuit automatically starts operating.

10.6.2 Charging Temperature and Voltage Monitoring/Control


Charging is limited by battery temperature and voltage so that it is
made in a preset range.
To monitor the battery temperature, a thermistor (with negative temperature coefficient) is installed in the battery pack. The thermistor outputs
a voltage signal to pin 6 (TS) of IC18 and charging is enabled if the voltage
is within the preset limits in voltage converted from battery temperature.
The sensitivity of a battery temperature rising ratio (AT/At) is adjusted by
resistances of the thermistor and R59, R61, and R62 on the charging circuit.

The battery voltage is divided to a voltage per cell by R58 and R60
which are connected between battery terminals, then sent to pin 7 (BAT)
of IC18. Charging is enabled if the voltage is within the preset limits in
voltage per cell.
Divided +5V powers of R84 and R85 are used to determine the maximum cell voltage (MCV) of the battery and the maximum voltage (TCO)
for the battery temperature to stop charging, respectively. They are output
to pin 10 (TCO) and pin 11 (MCU) of IC18.
Preset battery temperature limits and cell voltage limits are as follows:
. Battery temperature limits
Minimum charge enable temperature limit LTF: Approx. -50C
(voltage level 2.0V)
Maximum charge enable temperature limit HTF: Approx. 60oC
(voltage level 1.13V)
High temperature to stop charging TCO: Approx. 650C
(voltage level 1.01V)
. Battery cell voltage limits to enable charging
Minimum: Approx. 1.0V
Maximum: Approx. 1.78V

That is, when the charging circuit is turned on, it starts charging the
battery if the battery temperatre is within a range of 50C to 60oC and
the cell voltage is within a range of 1.0V to 1.78V. If either one of the above

2-16

conditions is not satisfied, it does not start charging the battery but is
placed in a standby status until both conditions are satisfied.
Once the charging circuit starts charging the battery, the charge complete system described later will be effective. However, if the maximum
cell voltage of 1.78V is exceeded before the start of charging, the charging
circuit judges the battery abnormal and does not start charging.

10.6.3 Charge Current Control


The power required for charging is supplied from the 24V power
generator. The charging current to the battery is detected by the resistor
R18 then sent to pin 9 (SNS) of IC18. The output of pin 14 (MOD) of IC18
is switched based on the voltage value of R18 to switch Q17 on/off, thereby
making the current constant. The charging current is set at approximately
0 45C (480mA).
.

Q16, D22, and Q10 form a MOS FET driver and level shifter circuit to

heighten the switching frequency.

10.6.4 Charge Complete System


When the battery temperature and cell voltage are within the respective ranges, the charging circuit continues to charge the battery until one
of the following five conditions is satisfied:
. - AV detected

.
.
.
.

- AllAt (temperature rising ratio) detected


Maximum temperature of 650C to stop charging TCO reached
Maximum charging time of 3 hours reached
Maximum cell voltage MCV of 1 78V reached
.

Usually, the charging circuit stops charging by detecting -AV.


However, if charging is started at high ambient temperatures, the charging
circuit may detect the maximum temperature (TCO) to stop charging. Also,
as a safety measure against no - AV detection, the maximum charging
time of three hours is set by timer and TM1 and TM2 signals at pins 4
and 5 of IC18, respectively, stop the charging circuit from charging if the
time is reached.

......

jLL1Ji -

I
'

(tl

"

Hit

1
,

r s
net

. <1

U l.

ix

5
1

I
T..

2-17

T
-S

10.6.5 Other Functions

In AC operation, supplemental charging is made by R66. In AC operation, Q14 is on and the base of Q13 is driven to supply R66 with the 24V
power. The current of which the value is determined based on a voltage
difference at R66 flows for supplemental charging. However, if the battery
is not connected, Q8 is turned on and Q14 is kept on, thereby disabling
supplemental charging and at the same time cancelling the bias voltage
of Q12. Thus, the 5V power to operate the charging circuit is not held and
charging is impossible.
The charging circuit outputs CHG IRQ and CHG STATUS signals. When
the circuit is supplied with 5V, the CHG IRQ signal becomes high level
to indicate the instrument is placed in charge mode. The CHG STATUS
signal is output from pin 13 (CHG) of IC18 to indicate the current charging
status. If the battery is not connected, the CHG IRQ signal cannot become
higher than 0.7V since D15 is electrified. Thus, the instrument is not placed
in charge mode.
The CHG signal is output in the following manner:
CHG STATUS Output Level
Low

High
Continuous

Abnormal power voltage

Continuous (rise)

Start of charging

1 375s 0.225s

Standby

125ms 20ms

Continuous

Charging
125ms 20ms

Charged

125ms 20ms

CHG IRQ Signal


CHG swilch ON

Cha/ping Mode

2 0s
.

'

CHG STATUS Signal


in ordinary charging

Start of Charging Mode

Charolng In Progress

Churging ConipleloiJ

CHG STATUS Signal


with some pending of charging
2 0s
.

Rending

CHG STATUS Signal


with abnormal battery voltage

Charging in Progress

No charging made

CANT CHARGE" message appears aller 5 seconds Irom the start

"

2-18

CHAPTER

Troubleshooting
1 Introduction
.

1 Internal Power Supplies


1 2 Faulty Power Supplies vs. Troubles
Troubleshooting
2 1 All Specified Secondary Power Supplies Are Not Available
in AC Operation
2 2 All Specified Secondary Power Supplies Are Not Available
in Ni MH Battery Operation
2 3 +24V Is Not Output
2 4 + 10V Is Not Output
2 5 Vdd Is Not Output
2 6 Unisolated 5Va Are Not Output
2 7 Isolated 5V Are Not Output
2 8 Troubles Related to Charging
2 8 1 Charging Mode or "CHARGING" Status Is Not Effected.
2 8 2 Virtually No Charging Is Made Despite "CHARGING
COMPLETE" on Display
2 9 LCD Operation Is Abnormal
2 10 No Key Operation Is Possible
2 11 Paper Is Not Driven or Abnormally Driven
2 12 No Mark Detection on Z-fold Paper Is Available
2 13 No Paper End Detection Is Available Or "PAPER END"
Alarm Is Constantly Displayed
2 14 Program Contents Change
2 15 Buzzer Does Not Operate
2 16 No Waveform Recording Nor Alphanumeric Printout
1

3-1

3-1
3-2
3-3

3-3

3-3
3-4
3-4
3-4
3-5
3-5
3-6

3-6
3-7
3-7
3-8
3-8
3-9

3-9

3-9
3-9

17
18
2 19
2 20
2

Is Available

3-10

No DC Input Waveform Recording Is Available


Any Waveform of All 12 Leads Is Recorded As Baseline
Waveform of Some Specific Lead Is Recorded As Baseline...
Checking AID Converter

3-10
3-10
3-11
3-11

1 Introduction
.

11
.

Internal Power Supplies


First, understand internal power supplies and their functions.

10Vn

The IOVn power is output from the secondary side of transformer.


Since it is not regulated, it fluctuates depending on the load status
and AC input voltage. It is transformed to various regulated voltages
to be supplied to respective circuits.
+10V

Inputting 10Vn, the low-loss 4-terminal regulator outputs regulated 10V. However, if the input is lower than 10V, the output
voltage will be lower by several ten millivolts than 10V.
This voltage is supplied to the motor and the regulator which
produces -i-5Vd.

+ 5Yp
Inputting + 10V, the dropper type low-loss regulator outputs
regulated -(-5Vd.
This voltage is used as the power supply for digital circuits,
such as CPU, of the unisolated section.
+ 24V

Inputting 10Vn, the switching regulator outputs regulated


+24V. A limiter is applied at approximately 0.4A referred to 10V,
thereby lowering the output voltage.
The +24V voltage is supplied to the thermal print head. In the
case of FX-2111N, it is also used as a charging power supply.
+5Va (unisolated)
Inputting IOVn, the dropper type 3-terminal regulator outputs
regulated -i-5Va.
5Va (unisolated)

Inputting +5Va, the inverting type switching regulator


produces -5Va.
5Va are used as analog power supplies for DC input, ECG
circuit and AID converter.

5Vf (isolated)

Receiving 10Vn through the isolation transformer which


isolates the voltage from other circuits, the 3-terminal regulator
outputs regulated 5Vf.
The 5Vf are used as power supplies for digital circuits of the
isolated section.

5Vb (isolated)

5Vb are produced from 5Vf through diodes. They are


used as power supplies for analog circuits (input amplifiers)
of the isolated section.
3-1

2 Faulty Power Supplies vs. Troubles

1 21

If 10Vn is not output...


The FX-2111 does not operate at all since 5V, 24V and all other powers
are not produced.
.

122

If 10V only is not output...


The dropper type low-loss regulator cannot produce +5Vd, thereby disabling the CPU, etc. to operate. The FX-2111 does not apparently operate
.

at all while 24V and others are available.

3 If +5Vd only is not output...


The CPU and other components are not supplied with power. The
FX-2111 does not apparently operate at all while powers other than +5Vd

12
.

are available.
12
.

4 If +24V only is not output...


The LCD and keys operate normally but waveform recording and al-

phanumeric printout are not available at all.


12
.

5 If unisolated 5Va are not output...

The AID converter and comparator do not operate normally, thereby


making ECG waveform or DC input waveform a baseline with alphanumerics
printed out normally.

6 If isolated 5Vf and 5Vb are not output...


All 12 leads are displayed and recorded as baselines with alphanumerics
printed out normally.

12
.

A Caution
When checking the power supply and related circuits for troubleshooting, take sufficient care to avoid a short circuit.

3-2

2
2

Troubleshooting

1 All Specified Secondary Power Supplies Are Not


Available in AC Operation

(1) Check the FX-2111 is securely connected to the wall outlet using the
specified power cord.
(2) Check the AC inlet has two specified power fuses inserted and the fuses
are not blown out.

(3) Check AC voltage at the secondary side of power transformer. Disconnect the output jack, which comes from the transformer, from P1 connector and measure the voltage between pins 1 and 3 of the output
jack. The AC voltage should be as follows:
9 9V to 12.5V for 115V version
.

8 5V to 12.5V for 230V version


.

(4) Check supply voltages on PCB-6236.


1) Check voltages between the anode side of D2 and D3 and the cathode
side of D1 and D4. They should be in a range of 10 to 16V DC.
If DC power output is not available on the board while the transformer
output is found normal, the contact of PI connector may be faulty
or diodes D1 to D4 may be damaged.
2) Also verify that the voltage between the cathode side (TP22) of D5
and GND (TP21) is in a range of 10 to 16V DC.
If the voltage between the cathode side of D5 and GND is not available, the fuse F1 on the board may be blown out or D5 may be
damaged.

A Caution
When checking power fuses, be sure to turn the power off and disconnect the power plug from the wall outlet beforehand.
2

2 All Specified Secondary Power Voltages Are Not


Available in Ni-MH Battery Operation

(1) Open the rear cover of instrument and check the presence of battery.
Then check that the battery is securely connected to the FX-2111 via
the connector.

(2) Check supply voltages on PCB-6236.


1) Verify that the battery voltage is available between pins 1 and 4 of
P10 and between the plus side (TP25) and GND (TP21).
If the battery voltage is not available there, the contact of P10 may
be faulty or the fuse F2 may be blown out.
2) Press the ON key and verify that the battery voltage is available
between the drain of Q4 (anode side of D6) and GND.
If the battery voltage is not available there, the holding circuit around
Q4, D7 and Q5 may be faulty or the connection of key panel may
be inferior.

3) Verify that the battery voltage is available between the cathode side
(TP22) of D6 and GND.
If the battery voltage is not available there, D6 may be faulty.
3-3

3 +24V Is Not Output

(1) Check the presence of IOVn (8V to 16V). Then verify that the voltage
is aupplied to pins 10 and 13 of IC6.
(2) Check operations of components around IC6.
1) Using the oscilloscope, verify that pin 4 of IC6 outputs a triangular
wave at approximately 170kHz.
If not, IC6 may be faulty.
2) Using the oscilloscope, verify that the gate of Q9 outputs a rectangular wave at approximately 170kHz.
If not, IC6 may be faulty.
3) Using the oscilloscope, verify that the waveform at the drain of Q9
is switched at approximately 170kHz.
If not, Q9 or D13 may be faulty.
4) If +24V is supplied to the thermal print head only, conduct the
following:
Press the START/STOP key to place the instrument in recording condition. Then verify that -I-24V is available at the drain of Q24 and
pins 13, 14 and 15 of P5 connector.
If the + 24V is not available there, consider Q24 or Q15 may be faulty
or check whether or not signals are sent from the gate array (pin 21).
Signal level of Q15's base input: High-24V ON
Low- 24V OFF

If +24V is available there, check the cable of thermal print head.

4 + 10V Is Not Output

(1) Check the presence of IOVn at pin 1 of IC22. To make 10V output available, the input 10Vn should be higher than 10V.
If the input voltage is present but the output is not available, setting
resistors R119 and R118 or IC22 may be faulty.

5 Vdd Is Not Output

(1) Verify that 10V input is available at pin 1 of IC7. If the IOVn output from
the secondary side of power transformer is lower than 10V, however,
the 10V input at pin 1 of IC7 is lower by several ten millivolts than the
specified -I-10V.
If the input voltage is present but the output is not available, IC7 may
be faulty.

3-4

6 Unisolated 5Va Are Not Output

(1) Verify that the 10Vn input is available at pin 3 of ICQ.


If not, both 5Va should not be output. Conduct the steps described
in 2.1 and 2.2.

(2) Verify that +5Va is output at pin 1 of ICQ and pins 3 and 5 of IC10.
If not, ICQ may be faulty.
(3) Verify that -5Va is output at pin 4 of IC10 (TP2Q, TP53).
If not, IC10, L2 or D14 may be faulty.

7 Isolated 5V Are Not Output

(1) Verify that +10Vn (8V to 16V) is supplied to pins 6 and 7 of 12.
If not, check whether or not 10Vn is output from the secondary side of
power transformer.
(2) Check the circuit around 12.
1) Verify that pin 11 (TP49) of IC20 inputs a 200kHz rectangular wave as
clock signal.
If not, IC16 or IC1Q on the oscillator circuit may be faulty.
2) Verify that pins 8 and 9 of IC20 output rectangular waves at 100kHz
in reversed phases.
If not, IC20 may be faulty.
3) Verify that pin 3 (TP60) of D29 and pin pin 3 (TP61) of D30, each at
the isolated side, output rectangular waves at 100kHz.
If not, check Q20, Q21 and 12, which may be faulty.
4) Verify that rectified and smoothed voltages in plus and minus directions are input into pin 3 of IC33 and pin 2 of IC34, respectively.
If not, D29 or D30 may be faulty.
If all the abovementioned inputs and outputs are found normal but
isolated 5V are not output, IC33 or IC34 may be faulty.

3-5

8 Troubles Related to Charging

28
.

1 Charging Mode or "CHARGING" Status Is Not Effected

If charging mode is not effected, +5V (CHG IRQ) may not be output
at pin 9 of IC18 or +5V may not be supplied to the charging circuit. Conduct
steps (1) and (2)-1) below.
If charging mode is effected but "CHARGING" status is not initiated,
+ 5V may be available on the charging circuit and at pin 9 (CHG IRQ) of
IC18 but either the battery voltage or temperature may not satisfy the charging
condition. Also, if +5V is not correctly output, the voltage may vary con-

siderably and may be beyond the specified range. So conduct steps (1), (2)-1)
and (2)-2) below.

(1) Open the rear cover of instrument and check the presence of battery.
Then check the battery is securely connected to the instrument via the
connector and the battery cable is not disconnected.
(2) Check the circuit around IC18.
1) Press the OFF key in AC operation and verify that pin 16 (TP44) of
IC18 outputs + 5V. Also verify that pin 9 (CHG IRQ) of gate array IC8
outputs +5\/.
If not, check the operation of + 5V power supply.
. If +5V is output only when the OFF key is pressed Q12 may be
faulty.
. If +5V is not output at all even if the OFF key is pressed Q11, Q12,
or D8 may be faulty.
. If only +5V for CHG IRQ is not output check whether or not the
collector (TP48) of Q8 is at low level. If it is low, Q8 may be faulty.
2) Verify that voltages between pin 6 (TS) and pin 9 (SNS) and between
pin 7 (BAT) and pin 9 (SNS) of IC18 are in the specified ranges as
,

follows:

Pin 6 (TP42) 1.01V to 2.0V


Pin 7 (TP2)
1.0V to 1.78V
. If the voltage at pin 7 (BAT) is beyond the specified range check
the terminal voltage of battery. If the battery voltage is not higher
than 8.0V or so, the FX-2111 is not placed in "CHARGING" status.
Owing to this, supplemental charging is made in AC operation. Check
whether or not 24V output is available at the collector of Q13 in AC
operation.
. If the voltage at pin 6 (TS) is beyond the specified range remove
the battery and check the thermistor in the battery. Measure the
resistance between pin 2 (blue) and pin 4 (black) of the collector.
The resistance should be in a range of 3kft to 30kl2 (lOkft at 250C).
Ambient temperatures should be proper.
,

3-6

28
.

2 Virtually No Charging Is Made Despite "CHARGING


COMPLETE" on Display

(1) Check the circuit around IC18.


1) Verify that the source side of Q17 is supplied with +24V.
If not, conduct the steps described in "8.3 +24V Is Not Output:'
2) Verify that pin 14 (MOD) of IC18 outputs a rectangular wave at
approximately 140kHz; the drain (TP52) of Q17 outputs a rectangular
wave at approximately 140kHz; and pin 9 (SNS) of IC18 outputs a
triangular wave at approximately 140kHz.
If not, check the following:
. R18 and R82 including their mounting condition
,

. 017 Q10, 016, and D22


,

. If above components are found normal IC18 may be faulty.


NOTE: Charging current is detected as a voltage value at pin 9 (SNS),
thereby switching the output of pin 14 (MOD).
,

(2) Check the number of charge-discharge times and the number of operating times after charging.
Usually, the number of charge-discarge times is 200 max, though it may
decrease depending on the operating environment.
The larger the number of operating times after charging, the shorter
the service life of battery becomes.
Thus, if the charging circuit is found normal but virtually no charging
is made despite "CHARGING COMPLETE" on display, the battery may
exhaust the service life.

9 LCD Operation Is Abnormal

(1) Check sure connection between the P2 connector on the main board
and the connector of LCD.
Check the connector terminals for any possible inferior soldering.

(2) Verify the following voltages on the P2 connector:


GND level at pin 1 (Vss)
+5V at pin 2 (Vdd)
+ 0.238V at pin 3 (Vee)
(3) Check the continuity of the following signal lines referring to "4.2 LCD
Control Circuit!'
Rs

signal

R/W

E signal
Signals of DB0 to DB7
NOTE: The LCD module used for FX-2111 has a built-in control driver and

is directly accessed by the CPU, with MEMORY READY put.

3-7

2 10
.

No Key Operation Is Possible

(1) Verify sure connection between the P7 connector on the main board
and the flexible cable of key panel.
Check the P7 connector for any possible inferior soldering.
Check the flexible cable of key panel for any possible disconnection
and wear of the carbon part.

(2) Verify that key signal lines except for ON and OFF keys are normally
pulled up by checking TP120 to 125 and TP128.
(3) Check the voltage at pin 20 (power supply) of IC15.
Check signals at pin 1 (KEY) and pin 19 (AO) of IC15. Signal at pin 1
should be at low level at every one millisecond.

Verify that pins 2 to 9 and 11 to 18 of IC15 are at low level when a key
is pressed.

NOTE: If only some specific key is not effective, the key may have a poor
contact.

A Caution
The inserting part of key connector is made of carbon. Avoid frequent
repeated disconnection and connection.
2

11 Paper Is Not Driven or Abnormally Driven

(1>/erify sure connection between the P4 connector on the main board


and the motor cable.

Check the P4 connector for any possible inferior soldering.


Check the motor cable for any possible disconnection.

(2Verify that pins 3 and 5 (TP103) of P4 connector are supplied with +5V.

(3yerify that the following signal is available at pin 1 (TP101) of P4 connector.


!<

5|19. 5;s

;<

5*L9. 5,s

+ 10V

JUULJUU
Paper speed 25mm/sec

Paper speed 50mm/sec

Check the signal at pin 14 (TP59) of IC25. (Refer to "2.4 Motor Control:")
It should be 256Hz with a paper speed of 25mm/sec and 512Hz with
a paper speed of 50mm/sec.

Verify that the signal at pin 2 (JP5Q) of IC12 is at high level when the
paper is driven.

3-8

12 No Mark Detection on Z-fold Paper Is Available

(1) Verify that "RECORDING MODE" is set to "MARK" in the program.


(2) Check the mark sensor for any possible dirt deposit and inferior mounting condition.
(3) Verify sure connection between the P6 connector on the main board
and the flexible sensor cable.

Check the P6 connector for any possible inferior soldering.


Check the flexible sensor cable for any possible disconnection.
NOTE: The mark sensor is the one and only part in the FX-2111, which is
adjusted by a variable resistor. Before conducting the steps
described above, therefore, check the variable resistor for any
possible peeling-off of the paint lock and slippage.
2 13
.

No Paper End Detection Is Available Or "PAPER


END" Alarm Is Constantly Displayed

(1) Check the mark sensor for any possible dirt deposit and inferior mounting condition.
Verify that the magazine sensor switch is positioned in place and
pressed when the magazine is closed.
(2) Verify sure connection between the P6 connector on the main board
and the flexible sensor cable.

Check the P6 connector for any possible inferior soldering.


Check the flexible sensor cable for any possible disconnection.
NOTE: The mark sensor is the one and only part in the FX-2111, which is
adjusted by a variable resistor. Before conducting the steps
described above, therefore, check the variable resistor for any
possible peeling-off of the paint lock and slippage.
2 14
.

Program Contents Change

(1) Verify that the lithium battery voltage is +2.5V.


Check the lithium battery for any possible inferior soldering.
(2) Verify that when the FX-2111 is turned off, pin 28 of IC2 is supplied with
+3V from the lithium battery to back up the RAM.

NOTE: Program contents are stored in the RAM IC2 and backed by the
lithium battery when the FX-2111 is turned off.
If program contents change despite sufficient lithium battery
voltage, refer to 2 9 Memory Backup Circuit:'
"

2 15
.

Buzzer Does Not Operate

(1) Verify that pin 15 (or R73) of the gate array outputs a rectangular wave
at 1kHz under the condition where the buzzer should sound.

If the wave is normally output, R73, soldering of the buzzer or the buzzer
itself may be faulty.
In the case where no QRS-synchronized sound is generated, make sure
that "QRS BEEP" is set to "ON" in the program.
3-9

2 16
.

No Waveform Recording Nor Alphanumeric


Printout Is Available

(1) Verify sure connection between the P5 connector on the main board
and the thermal print head cable.
Verify sure connection between the thermal print head and the cable.
Check the P5 connector for any possible inferior soldering.

(2) Verify that +24V is available between pins 13 and 15 (TP112) of the P5
connector under the condition where a waveform should be recorded.

Verify that pin 9 of P5 connector is supplied with +5V.


(3) Verify that the TOFF signal at pin 21 of gate array or the base of Q15
is at high level under the condition where recording should be made.
Verify that signals of LATCH (pin 2 of P5 or TP107), CLOCK (pin 3 of
P5 or TP108), STROBE (pins 5, 6,and 7 of P5 or TP109) and DATA IN
(pin 1 of p5 or TP106) are available at the timings as described "2.8
Thermal Print Head Control Circuit;'

(4) To make sure of the correct operation of the RAM for thermal print head:
. Verify that pin 28 of IC5 is supplied with +5V
. Verify that JP3 is short-circuited by soldering
.

NOTE: If electrical signals are normal mechanical failure may make waveform recording and alphanumeric printout impossible. So check
the spring mounting condition and other mechanical status around
the thermal print head.
,

2 17
.

No DC Input Waveform Recording Is Available

(1) Make sure that "DC RECORDING" is set to "ON" in the program.
(2) Check the P3 connector on the main board for any possible inferior
soldering.
(3) Verify that pin 4 of IC21 is supplied with +5V and pin 11, with -5V.
If IC21 is not correctly supplied with power, the FX-2111 should not be
able to record ECG waveform too.

(4) Check the AID converter referring to "2.20 Checking AID Converter"
If the AID converter is faulty, the FX-2111 should not be able to record
ECG waveform too.

2 18
.

Any Waveform of All 12 Leads Is Recorded As Baseline

(1) Check isolated 5V and unisolated 5V referring to 2.6 and 2.7.


Verify that power supply pins of IC12,13,17, 21, 31 and 35-45 (operational amplifier, multiplexor, etc.) are supplied with power.
(2) To check the reset circuit and the control signal, verify that pin 13 of
IC44 is not fixed to high level.

3-10

19 Waveform of Some Specific Lead Is Recorded As


Baseline

(1) Check the lead cable for any possible internal disconnection and inferior
hookup.

(2) In relation to the lead which results in a baseline, check the buffer IC,
the resistance of resistance network, power supplies to lead selector
ICs (IC36,38,41 and 45) and their connection status such as soldering.
2

20 Checking AID Converter

Check the AID converter if the trouble occurs despite the normal condition detected by taking steps described in Section 2.16, 2.17 or 2.18.

(1) Verify that pin 7 (TP7) of IC13 outputs a triangular wave with an amplitude of 3.2V and a pulse width of 520ns. If the triangular wave is
normally output, proceed to step (4) below.

(2) Verify that power supply pins of IC17,21 and 13 are supplied with power.
Verify that the reference voltage of +2.5V is supplied.
If not, check the power supply.

(3) Verify that a rectangular wave ADCLK (L: 520(is, H: 480|js) of 1ms cycle
is output at 1ms cycles from the collector terminal of Q1.
(4) Verify that the low level S/H signal with a pulse width of 96ns is output
at every 1ms from pin 6 of IC17, referring to 2 3 AID Converter Circuit!'
"

(5) Check the output of comparator as follows.


*
Verify that pin 1 (ECG signal) of IC13, pin 8 (DC input signal) or pin 14
(battery/thermistor temperature) outputs a pulse which changes in the
pulse width corresponding to the changing input signal.
Verify that pins 4, 6 and 8 of IC12 output the abovementioned pulse
signal of which the amplitude is limited to a range of 0 to 5V.

3-11

CHAPTER

Maintenance
1

Replacing Power Fuses


Replacing ROM
Replacing Rechargeable Battery
Operation Check by Self-testing
4 1 Self-testing Procedures
4 2 Examples of Test Results
4 3 System Errors
Disassembling/Reassembling the FX-2111
5 1 Disassembling Procedure
.

4-1
4-1
4-2
4-3
4-3
4-4
4-5
4-6
4-7

Replacing Power Fuses

Power fuses are hardly blown out. But if they are blown out due to
some reason, turn the FX-2111 off and disconnect the power cord from
the wall outlet. Then pull fuse holders off the bottom panel as shown
below. Insert spare fuses and put fuse holders in original positions.

A Caution
When replacing power fuses, be sure to turn the power off and disconnect the power plug from the wall outlet beforehand.

Replacing ROM

(1) Remove the upper casing.

(2) The ROM will be accessible on the PC board. Remove the ROM.
(3) Mount the new ROM, taking care of the direction. The ROM's socket
has 36 pins.
If 1M-byte ROM is used, it will be mounted taking the full space for
it. If the ROM is other than "IM-byte ROM, put it aside the rear of the
socket (aside IC3).

(4) Check the ROM is not contacted with feet of R18 when mounted.
(5) If there is no problem, put the upper cover and fix it with screws.
(6) Turn the FX-2111 on. Verify that the control program version on the
standby display is the same as the label on the ROM.

A Caution
When replacing the ROM, be sure to turn the power off. Also, take
care to install the ROM in correct position.

tfC6 tft
i)

4-1

Replacing Rechargeable Battery

(1) Turn the FX-2111 off and disconnect the power cord from the wall
outlet.

(2) Put the FX-2111 upside down. Loosen two screws off the battery cover.

(3) Replace the battery with new one. Insert the connector of battery cable
securely in place.
(4) Put the battery cover in the original position and fix it with two screws.
If the battery cover is not put squarely, rearrange the battery cable,
etc. to neatly accommodate the battery in the room.
(5) Press the ON key to verify that the FX-2111 is powered.
(6) The new battery may have been stored for a long period of time and
lowered in capacity due to self-discharge. So be sure to charge the
battery after replacing.

A Caution
When replacing the battery, be sure to turn the power off and disconnect the power plug from the wall outlet beforehand.

"v

NOTE: It is recommended to charge the battery in the stock every six


months. Also to make full use of the battery in the FX-2111, follow instructions in the operation manual.

4-2

Operation Check by Self-testing


The self-testing function permits you to perform the following tests:
1 Total Test

6. LCD Test

Sensitivity Test

7. Key Test

3 Time Constant Test

8. Status Test

4 Print Test

9. Automatic Power-off Test

5 Recorder Test
.

41
.

Self-testing Procedures

(1) Press the MODE key to place the FX-2111 in the program mode. The
display will be as follows:
1

AC FILTER

(OFF)

(2) Press the ImV (A) key. The following message will appear:
MAINTENANCE

PUSH (START) KEY

(3) Press the START/STOP key. The display will be as follows:


M-1 AC FILTER

FREQUENCY

(50HZ)

(4) Press the ImV (A) key. The following message will appear:
SELF TEST

PUSH (START) KEY


(5) Press the START/STOP key. The display will be as follows:
1

TOTAL TEST

Press the START/STOP key once more. The total test will start to test
sensitivity, time constant, print, recorder, and LCD.
To proceed to the Key Test, Status Test or Auto Power-off Test, select
it using the < or
key.
9 AUTO POWER OFF

4-3

1/2

TEST
1

42
.

2 Print Test

'

n t a *

1 *

'

42
.

3 Recorder Test

....

;..>

Mm"

RECORDER 'EST

42
.

4 LCD Test

0 o
0 0 0 0 0
0 0 0 0 0
0
0
0
0
0 0 0
0 0 0 0
0 0 0 0 0
0 0 0 0 0
0 0 0 0 0

0 0

4-4

3 System Errors

System errors are caused by troubles with hardware or software.


Usually, they cannot be treated by users.

If a system error occurs, the following alarm message blinks and


the buzzer sounds:

(Blinks)
S

CODE

m R

If a system error occurs during recording, the operator should stop


the FX-2111 from recording (stop the motor). All the operation keys will
be ineffective. Also, the normal operation will not be recovered unless
the power is turned off once.
Error codes and causes are as follows:
Code
01

Type

Description

Overheat

Thermal print head is abnormally heated to over 60oC.

02

ROM error

Checksum error

03

RAM error

RAM's read/write error

04

Unused interrupt 1

Unused software interrupt

05

Unused interrupt 2 Unused software interrupt

06

RTM error 1

Task No. to RTM macro is illegal.

07

RTM error 2

Task status to RTM macro is improper.

08

RTM error 3

Task's queue area overflows.

RTM: Real-time monitor

4-5

Disassembling/Reassembling the FX-2111

For maintenance, inspection and replacement of ROM and other


components, you need to disassemble the instrument. Use the following steps to disassemble: The FX-2111 will be reassembled in the reverse
order. Also, refer to the Chapter 9 Structural Diagrams" when disassembling.
"

When disasssembling or reassembling the FX-2111, take the following cautions:

A Caution
Disassembling/Reassembling Precautions
. Be sure to disconnect the power cord and make sure the instrument is turned off before disasasembling or reassembling.

. Do not remove the battery before disconnecting any PC board

. Take care that repeated disconnection of the key panel and sensor board may result in poor contact.
. Use proper tools to loosen screws

. When reassembling make sure that all screws are securely tightened and all connectors are completely inserted.
,

PC Board Handling Precautions

. PC boards are equipped with highly sensitive components to static


electricity.
. PC boards are highly sensitive electronic devices Put removed PC
boards in a proper protective bag or take appropriate measures to
protect them.
.

Handle PC boards carefully. A shock to them may damage the components.

. Never insert a connector to the powered PC board nor remove the


powered PC board.

4-6

51
.

Diassemling Procedure

Take the following steps to disassemble the FX-2111: (The procedure described here
is for the model equipped with the battery.)

Doublt Semus Screw (S)


~

M3x8

e7)

g3

jo

Step 1: Remove the cover of paper magazine


and the paper shaft.

Step 2: Open the battery cover on the bottom


panel and remove the battery.

Irom Ttanslormer
2-Double

Semus Screw (S)

Irom RecodJer Motor

Thermal Array Head Cable

Sensor Cable
Double Semus Sciew (S)

fe

M3x8

1/

51
.

0*
4-Double Semus Sciew

(S) M3x8
.

Step 3: Loosen four screws on the bottom


panel and remove the upper case.

Step 4: Diconnect connectors from the PC


board and loosen two screws on the
PC board. Then, the PC board can be
removed.

4-7

Double Semus Screw (S), M3x8

3-Double Semus

Screw (I). M3xB

>

0
0

Step 5: Loosen three screws which fix the


recorder onto the botton panel and
remove the recorder.

Step 6: Loosen the screw which fixes the


power board PCB-6238 and the transformer onto the bottom panel. Disconnect

the

connector

from

the

transformer and remove the power


board.
2- Double Semus Screw

(S)

M38

m
8

8
el

Step 7: Loosen the screws whch fix the


transformer onto the bottom panel.
Then, remove the transformer and

grounding parts. Now, disassembling is complete.


4-8

CHAPTER

Periodical Inspection
1

Periodical Inspection
1 1 Checking lead cable and power cord
1 2 Visual inspection
1 3 Cleaning the FX-2111

5-1
5-1
5-1
5-1

1 4 Self-test and total function check

5-1

15

Battery check
Safety Inspection
2 1 Measuring leakage current
2 2 Measuring protective grounding resistance

5-2
5-2
5-2
5-5

2 3 Remarks

5-6

Periodical Inspection

In this chapter, we describe the inspection to prevent troubles and


let the FX-2111 keep the sufficient safety and complete operating condition.

Perform the following at least once a year:


Check the lead cable, power cord, and grounding wire for any possible
damage.
. Conduct visual inspection

. Clean the FX-2111

. Perform the self-test and total function check

. Check the battery


. Check the leakage current
Measure the protective grounding resistance.
.

As with simple inspection, recommend the user to make it every day


or week. But let the user refer a complete function check and troubleshooting to the service engineer authorized by Fukuda Denshi.
11
.

Checking lead cable and power cord

Visually check the lead cable and power cord for any possible damage.
Check also the connectors for any possible looseness. If a damage is
found, replace them as required. Merely applying a tape may not repair
the cable sufficiently.
Also measure the resistance of each electrode to check for internal

cable disconnection and damage.


1

2 Visual inspection

.
.
.
.

Check that mechanical parts such as screws are securely fitted


Check that connectors inside the FX-2111 are securely connected
Check that there is no trace of damage
If a loose part is found tighten it securely.

3 Cleaning the FX-2111

The recommended number of cleaning times depends on the operating frequency and environment.

For cleaning, wipe the casing with a properly wetted soft cloth. You
can use the cleanser for tableware by weakening it with water. However,
take care that the cleaning solution or water may not enter the inside

through openings. Finally, wipe off wetty substance with a dry cloth.
1 4 Self-test and total function check
.

Perform the self-test described in "4. Maintenance" and check the total

function of the FX-2111 to ensure that the equipment operates safely and
completely.

5-1

15
.

Battery check

Measure the voltage between plus and minus sides of battery to check
the capacity. If it is below 9V, charge the battery. Also, if the battery has
not been charged for six months, charge it. If the battery will not be used
for a long period of time, remove it from the FX-2111 and store in a cool
place.

If the battery is below 6V, it may be difficult or impossible to charge.


In such a case, replacement is required.

Safety Inspection

To ensure the safety of FX-2111, it is recommended to perform safety


inspection. The test methods and measuring instruments are stipulated
in the standard for safety test. It is considered extremely difficult to fully
follow the standard for the safety test at the health care site. In checking
for maintenance and inspection, therefore, measure each specified value
as a rough rule of thumb.
We describe here examples of simplified measurement of leakage
current and protective grounding resistance.

A Caution
If you find a value which exceeds the allowable level, be sure to let
the user avoid using the FX-2111. If the user operates the FX-2111 as
it is, he/she may receive an hazardous accident.
2

1 Measuring leakage current

Prepare an instrument to measure the leakage current (electronic or


digital voltmeter), impedance device and power switch box.
The impedance device has the following configuration:
R1: 10kn 5%
R2: 1kfi 1%
R 1

C1: 0.015tiF 5%
V:

R 2

C 1

-r

Voltmeter

You can obtain a leakage current by measuring voltages at both ends


of the impedance device.

5-2

The digital voltmeter shall indicate a true root mean square value to
a composite waveform of a frequency band from DC to 1MHz. If such a
voltmeter is not available in hand, you may use a commercially available
digital voltmeter for the purpose of simple maintenance and inspection.
However, make sure of the frequency band of the voltmeter and note that
the voltmeter will not indicate a leakage current at a frequency exceeding
the capacity.

(1) Example of checking ground leakage current (current flowing in protective grounding conductor)
Power Switch Box

Polarity Switches
oa

POWER

ob

Wall Outlet

Medical
Electronic
Instalment

GND
GND

Oft

'

A Point

8 Point

Power Cord
M D

MD=lmpedance Measurement Device

See the figure above. A gounding terminal and the power switch box
which allows you to switch the polarity facilitate measurement of
ground leakage current.
The procedure to measure ground leakage current using the power
switch box is as follows:

. Measurement of ground leakage current is made by measuring


voltages at B point (grounding conductor of power cord) and A point
(wall grounding terminal) under normal condition and single fault condition.

Normal condition

. Measurement of ground leakage current under normal condition is


made by switching the polarity switches to a" positions, then to
b" positions. Thus, measurement is made in two ways.
"

"

Single fault condition


. Measurement of ground leakage current under single fault condition is made with a power fuse removed by switching the polarity
switches as above. Since the same measurement is made with

another fuse removed, measurement for single fault condition is made


in four ways.

5-3

(2) Example of measuring enclosure leakage current (current flowing from


enclosure to the ground)
Power Swdch Box

Polanty Switches

POWER

Oj-

ob

Medical

Electronic
Insirument

Wall OuUel
GND
GND

T-O

A Point

B Point

Opsn lor ilngla lauli condillocyPower Cord

| M D \-

Metal Foil

Probe

MO Impedance Measuiement Device

Or

Apply a metal foil of 20 x 10 cm to the isolated enclosure. Put the probe


to the metal foil for measurement.

Measurement of enclosure leakage current is made by measuring


voltages at A point (wall grounding terminal) and B point (desired
part of enclosure).
If the enclosure is isolated, stick a metal foil (e.g. aluminium foil) onto
the enclosure with a conductive tape and apply the probe to that
metal foil.

Measurement of enclosure leakage current under normal condition


is made by switching the polarity switches to "a" positions, then to
b" positions. A value measured under normal condition will be a
very small value near zero.
Measurement of enclosure leakage current under single fault condition is made with the wall grounding terminal and the grounding
conductor of power cord removed and by switching the polarity of
"

power source.

In leakage current measurement, take a maximum value as the measurement result.

The following are maximum allowable ground leakage current and


enclosure leakage current: In inspection, it is important to compare measurement values with past ones to check for any outstanding difference.
Maximum allowable leakage currents (117/253V AC)

Ground leakage current


Enclosure leakage current

Normal condition

Single fault condition

0 5mA

1 0mA

0 1mA

0 5mA

5-4

22
.

Measuring protective grounding resistance

According to the standard, protective grounding resistance shall be


measured by letting the testing transformer flow an AC current of 10-20A
from a power source of which the voltage with no load does not exceed
6V. Practically, however, it is not easy to prepare such a testing device.
We show therefore a simplified method as follows:
V
O

Power Iranslormer shall be able lo let a curfent of 13A How at a voltage ol less than 6V with no load
.

Medical Eqjipmenl
VW

A Point

Power

30

Transformer

Or

C Point

EO

(SOW)
6

A C
6V

Metal Part

VI (Voltage between
A and C)

2 30 V

V2 (Voltage between

B and C)

R: Protective Grounding
Resistance

B Point
.

El

Power Cord

'

Simplified Measurement of Protective Grounding Resistance


The testing circuit above allows you to measure the protective grounding resistance of equipment using an AC current of 2A or so. Connect the
power cord of equipment to El and the metal part (such as equipotentialization terminal) to EO. Make the conductor to EO shortest possible to
suppress the contact resistance. Flow a current of 2A or so to the equipment and measure voltage V1 between A and C points and V2 between
B and C points.
Since the flowing current is the same, the subject protective grounding
resistance can be obtained through the relations between the voltage ratio
and resistance ratio as follows:
R/3 = V2/V1

"

R (protective grounding resistance) = 3 V2A/1 (fi)

Measurement current may be 1A or so.


With the standard, the resistance between an accessible conductive

part and the protective grounding terminal shall be 0.1 Q maximum. In measurement at the grounding conductor of power cord with the power cord
connected, the grounding resistance including that of the protective
grounding conductor of power cord shall be 0.2ft maximum.
Protective grounding resistance

Resistance between protective grounding terminal


and potential equalization terminal

0 10 maximum

Protective grounding conductor of power cord and


potential equalization terminal

0 20 maximum

5-5

2 3 Remarks
.

Abovementioned measurement of protective grounding resistance


requires the power transformer which can flow an AC current of several
amperes and a testing fixture. If you cannot prepare such a testing power
transformer, you may measure it with the following method to obtain a
rough rule of thumb:
Measure the resistance between an accessible conductive part and
the protective grounding terminal to verify that it is lower than 0.1ft. If you
measure the protective grounding resistance with the grounding conductor
of power cord as the protective grounding terminal, the resistance shall
be lower than 0.2ft including that of the protective grounding conductor.

You need also to adjust the resistance of the testing lead or probe
used for measurement.

Medical Ecjjipmenl

Reslsianca Measuring InsUument

Testing Lead
Accessible Conductive Part

< 0 .

(e.g. potential equalization terminal)

1 fi

Protective Grounding Terminal

Potential Equalization Terminal


po-m

nvj

Protective Grounding Terminal

Simplified Method of Resistance Measurement


This method considerably differs from that of the standard. Consider
the result as a rough rule of thumb.
Protective grounding resistance
Resistance between protective grounding terminal
and potential equalization terminal

0 1ft maximum

Protective grounding conductor of power cord and


potential equalization terminal

0 2ft maximum

5-6

CHECK SHEET FOR PERIODICAL INSPECTION


Inspection Date:

Inspected by:

Model Number

Installation Site

Serial Number

Years of Use

Check Items

Judgment/Measurement : Remarks (Repair Required)

VISUAL INSPECTION
1 Cords and Cables
.

Power cord and lead cable for damage

Pass/Fail

Connections of power cord and lead cable

Pass/Fail

Casing and Accessories

. Casing for damage such as crack


. Casing for dirt

Pass/Fail

. Labels for fading letters

Pass/Fail '

Pass/Fail

. Screws for loooseness and dropout

Pass/Fail

. Electrodes for dirt and rust

Pass/Fail

. Expendables and operation manual

Pass/Fail

3 Connectors
.

Pass/Fail

. Power and lead connectors for damage and looseness


r

4 Others
.

. Power fuses for breakdown

Pass/Fail

Protective grounding terminal for damage

Pass/Fail

Switch panel for damage and dirt

Pass/Fail

Thermal print head for damage and dirt

Pass/Fail

Paper magazine for damage and dirt

Pass/Fail

LCD module for damage and dirt

Pass/Fail

ELECTRICAL INSPECTION
Pass/Fail

. AC operation and battery operation


. Time constant test

Pass/Fail

. Recorder test (paper speed & sensitivity)

Pass/Fail

. Printout test (for missed dots)

Pass/Fail

. Key test

Pass/Fail

. LCD test

Pass/Fail

. Auto power-off test

Pass/Fail

V)

. Battery charge

Pass/Fail

SAFETY

|jA Pass/Fail

. Ground leakage current (normal condition)

pA Pass/Fail

(single fault condition)

|iA Pass/Fail

. Enclosure leakage current (normal condition)


(single fault condition)

|jA Pass/Fail
Q Pass/Fail

. Protective grounding resistance


OTHERS

5-7

Charge OK/NG

CHAPTER

Circuit Diagrams
Overall Block Diagram
Overall Block Diagram
Main Board Circuit Diagram, 1/9 to 9/9
Power Board Circuit Diagram

433-5125
6-1
433-4893
6-3
523-2020... 6-5 to 6-21
524-2021
6-23

Sensor Board Circuit Diagram


Motor Module Circuit Diagram

524-1959
524-1960

6-24
6-25

Battery Pack Circuit Diagram


Overall Connection Diagram
Overall Connection Diagram
LCD Connection Diagram
Motor Connection Diagram

524-1961
433-5126
433-4894
434-4895
434-4896

6-26
6-27
6-29
6-31
6-32

r .
.

DC Input

Main Board

Pr.B-6409

Bult.r

Amplifier'

, i

Power Supply ' :


i

A C 1 1 5 V

>

I i

5V

5
u

Isolated Section

rs

Butttr
Amplifier

Preamplilici

ConsUnt/

it

Reset

'

Circuit

n
ModulaiJon

Middle

Ampliner
i25

Tranjlormer

Circuit

! fi!

Time

Resislanct
NelworK lor
Lead Slectior

Demodulation
Circuit
I->

Amplifier

Control

Pustv

Power
Circuit

Isolation
Translormer

Pull

'

21

I .

Detector

Detector

Circuit

Circuit

PCB-410

i:

i . - - -

. I
*

Digital
Power

+ 5V

Rectifier
Circuit

Overinpul

AC Inlet Board

\\
\\

'

Supplies

Oscillator

R-wav*

'

l
I

Circuit

'

+VFB

: i

+5V

j-

Generator ~

+5V.

I . , AC Input
i

i US

Triangular |_

.
""

i
'

< IA

Amplilicr

r-Wav*

liaV

J
r-

f jlse
t
Generator
Circuit

Trans I or mc/

: i
; i
i

+ 10'/

Recorder

Photo

Photo

Coupler

Coupler

A
Sensor Board:

Sensor
Circuit

.
'

PCB-62 39

Gale Array

Motor

Control

ROM

AM27C512

54Kbyte

QS6-F
L

- . - - . .1

Charging
Circuit

| A V A T

CLK

SRH2254-12

16HHz

8Kbyte
lor Thermal Prinl Head

RAM

SRM2264-12

RAM

'

TEMP
Prinl Head

+ 24V

s
4 \

Thermal

2HHz

In

B
Motor Module

H063B09

T
FD88O07-AC

NF-26P1/25|*

KM

; i
1 i

CPU

7t

:i
:i

ji

SKbvte
Ni-MH

LCD
NOH202AOO

Battery
PacK

Key Panel
i

DRAWING NO.

TITLE

Overall Block Diagram


MODEL NO.

ASSEMBLY NO.

433-5125
DATE

96.10.9

FX-2111 (USA)
6-1

DC Inpul

Main Board .

pcb-6409

Butter
'

-.1

Power Supply 1 ;

Amplitief

AC

sv

i
o

|i

Isolated Section

rs

Time

Resistance
5 Network lor
Lead Seltctior

o
V

Preamplilicr

xB

Reset

'

Modulation

Amplifier

Circuit

125

Circuit

Middle

ConslanU

Amplilier

Isolation
Transformer'

Demodulation
Circuit

Amplifier

2S1

xl.25

Triangular

r-SWave

ItnV

Power
Circuit

1 ,

Isoiation k
Translormer

PushPull

Circuit
in

AT

Digital
Power

+ 5V

Supplies

Oscillator

Rectifier
Circuit

Oelector
Circuit

Detector
Circuit

Pulse
Generator
Circuit

t
I

Photo

Photo

Coupler

Coupler

CPU
HD63809
2MHz

Recorder

FET

Sensoi
Circuit

PCB-5239

Gate Array

Motor
Control

NF-26P1/25

<

ROM

AM27C512

64Kbyte

R AM

CLK

SRH2264-12

16MHz

8Kbyte

Charging
Circuit
I

> AVJT

>

TEMP
Thermal
Print Head
056-F

+2 V

i D

FD88007-AC
Motor Module

Transformer

+ 10V
v

Sensor Board:

pcB-e4io

11.
_

iVFB

Overinput

AC Input

Generator

i5V

Amplilier
Control

R-wave

AC Inlet Board

+ 5V

fc

230 V

RAM

SRM2264-12

8Kbyte

Ni-MH

LC 0

Battery

Pack

Key Panel

NDM202AOO
-

lor Thermal Prim Head


t

DRAWING NO.

TITLE

Overall Block Diagram


MODEL NO.

ASSEMBLY NO.

4334893
DATE

96.10.9

FX-2111 (CE)
6-3

noi!) i miss omtnsi smcifih


.

AU IEC1ST01S tit IN Q \%, I. ||


,

ALL cmcims tit IK U F

VBF-f VRF -

uPC4064G2

IC37

4
4

1039

<D tfl. tin IIS! PAH

IC31

uPC4054G

4
14

IC35

TC-4051BF

IC3Q

TC-4051BF

is
16

TC-4S3flRF
TC-40S1BF
IC44

SPICUl SYHtOlS UUCE

JUL
7

7,a
7,ft

16

16
3

7,6

16

7 ,a

VBF-

I1 <

C97 Jcxot, )<:

Ci 19

Cl IB

2.

TC-40g4BF

IC451TC-4051BF

<M iM 1!S1 PtD

FC

Fl
TCI? .TC-4Sfl4RF

F2
7

+RV
-

-RV
-

VDD
-14

IC13

uPC4074C

IC15

TC7W04F

TC17

TC-40S3RF

11
-

ICig

TC-4Ba4BF

IC20

TC-74PC74F

u PC40B4G

HEF4046BT

TC7W02F

4
4

IC1
TC2

QM27CS12-150DC
SRM2264-12

IC3
ICS.

HD53B09E

ICS

FDa6007-PiC

5RM2264-12

C93 jcil? |ci25 |ci28

14

L4_

C94 \ci Ifc |ci2t ]ct.27

IC26
IC27

JL4.
_

IC21

SVF

.SWF

IB

9
FC

FC

E 1

VDD

VDB

14

25

14

26

6
VDB

14
IB
40
65

'

C I

I Ci

\ CC

|ci3 jcT |cZ2 ] C30 j C31 | C

4>

JTTTTTTTT

26
3
26
63

| C3I joi Jclfc | C37 | COS jcfcl lct.6 ] C68 J_C7 3

jJTTTTTTTTT

90
TC11

TC74HC24BftF

IC14

TC74HC04nF
TC74HCB41ftF

IC15
IC23

TC74HC04RF

1
7

J 4
_

1A

22l
4.

TITLE

DRAWING NO.

Main Board Circuit Diagram, 1/9


MODEL NO.

FX-2111

ASSEMBLY NO.

523-2020
DATE

PCB-6409

6-5

95.01.09

co1

H3
o
o

o
o

<

2 OK

mil

i
3

. IIS
2 OK

JL

ao

rHI--

n 163

12

5 4

It

IC37B

. 11

X4

>

%7

"

>

rnuj

n tai

ni&i

caj

20K

3K

-VW-t

. 21 1
'C

19

lT

tlK

IJ IQH

>-fl.

,_

-L.
>
| C 10]

uPCIC

20K

iC3a,.

2,OK
-

ICS 10

rc

P'

1 3

20K

5=|

LI

LU.

I t11170 > 1-

100

WPC4044

434

3.5n >

J*

1 SS14

Jl J

rc

>]

> 30QK

LSS302

uPC40t*

If
<(
a?

20K

fC
ZuK

. 201

1
10M

n i hi

H 183

t 40K

20K

1 7 I

5VF

. IK

I NH

2r

032a
T 1SS301

C 104

HQ*

'

104 S

iBsr~

> 2 If

> JOK

522 <.i/jsv

xo

[CI 13

4 I

JL

<5

rc
t NH
.

'

UPC4044

1051

40

HOP

tg41
,

?- ai

R20

30K

XC

sm

i- 02

%- QJ

CLK

[ j 04

oc

3:

Cl 12
0

I cue

. 220

3t

LOK
UPC40ft4

I OK

....
I NH

1035

IK

0223

Cl IS

S IK

0 2

TOTi

[ C 120

ClOl

I 3io0i.r

4 0?S

I
12

Cl 14

'~

TWPPrtT A l

A2ao
LOK

IK

Paga

svr

1cjo

Q4Q
runs

oo**r

4093

Jl

mi'

Pa

UIIO04CSB-1.2

9 HOP
131

Xft2i4
? 9. IK

03
33
UN3

If

uf*C4044

11214 >
5 IK ?

A ,c"

It 1

> 113 19

C I

rc

4 . B

10* ?

i i
u

JK

13K0

] .4KB

rnua

t 1*1

22

IC370

n ia2

JL

40
X I

tC3

01S|

. 294

017

CB7

-"vw

'0"

. 210

I) HOP

<1K

[
c04 X
X
C04

140K <

VW

. IBM

7 OK

IC3JC

""

i-

JJ

. 1

rc:200H>

0.0013

Aiit

IC3T0

XL

aoac-isscicr >

I A

JJl

0 01

-svr

Ton-

V 1S0P

4093

CO

Li

44 1

IC400
C42C
41

<1K

C I

lOK

F _

. 342

1C430

vr

rC404

00.

UX5 212

*IK
V L0H

1C43C

044
-

rnM2

rs
"140

svr

JL

Li

11 2

C7B

LirC404

SIM

o.oca

[ C 123
1S0P

. 1 BO

941

rnu2

. 174

. It 7

-S\J~V-

1>-

IK i J 1 OK JL

p
9

17*
10K

PC
rc40*4

10K

in

-T-n <.pc404 4 o.jj ; i?i


5bi7S

/300K

} 140K

peo44

10K

rc

rc

<fl 1 73

II
. 24S

..

p.

04 3
UN

112

R210

uPC33 5

24
oloaii
HI-

l 30

PC33

e
. IK
_

zook

n2:e

C122

2S2
:

IJ

IC43n

IK

2 2 OOP

. 24*

Ji

--*/W-

O . 4 t

I C4 J

SiOP

. 20

24

rc

R 2SS

3or

? . IK

C 134

C 1 32

A224
49aK

C I I 1

fll4ft

C133

6ok

PC40%4

1124 1

'

1C43U

3 .003 3

. svr

svr

/W

HI-

0 22

rc

rc

'

CI3S

IC42n

"

iC3in

?wr

. 234

200K

1
. 230

R73

C79

rc

R7S1

. 169

0 22

5 . IK

M23T

uPH131

ic2o

0234

JJL

.301

,J lOM

4S30

. 23ar
13K

fl 2 3 Z

JL

"

CO

IC4aA
4S3a

? 3 f
/tiOK

. 247
-

IOK

S IK

. 244
.

CI g)

1-0
uFC 4064

25V|

C131
HO'

DRAWING NO.

TITLE

Main Board Circuit Diagram, 2/9


MODEL NO.

FX-2111

523-2020
DATE

ASSEMBLY NO.

PCB-6409

6-7

95.01.08

R12(.

Past

N 1 7 B

n bfi-

2K

v/V

>

Ri 28

>

<1I> I-

2K

C47

K 1 79

1' SI* 1

C22

T 1

5514793

zr.

30Kr

120KF
IO00P

Rlt6

rtu

nBD

p 10 '

5511391

Pa??

f c : 25

>
03 2

age

0.022

2 C21D

E j

>

R4 a

Li
1M

1O0OP

R1CS
1- r

1M
k 1 25

D O1
.

1 n

UPCb6(jt)

D . 0 J

0.01

IT

L3

z
uPC07a

* 1

63 1

02

5S1393

10O

SSI 4393

FC

Irc5e

0
11B

ISS302

V O 022
.

E J

zo

R56

c 1

6
1 1

in

Kb 2

I NH

1 C 13 C

2K

R57

.10

L 00

E2

D 19B

1SS302

3
.

P9*

I"

R 1 IS

UPC4064

B2K

1SS302

R5

RSO

I C 130

S-C8036-01

rc58
D20e

SSI 4393
IK

0 19

Rill

at

v 0 02 2

K 7B

vv,

fa

C J

2
R 109

R 106

Pasff

E 1

I C2 lO
P3

20K

] C3e
_

rc jjc

uPC407a
5V

n v

20K

T C23
_

IC 128

R43

iOOOP

J C 13

2K

in

OOP
5

ce5.
0 J

C5B

CS7
12

ce7

in

CfcO

C56

Dl 90

ZK

"ess

1SS3D2

1DO

ICIJO

R4

5 IP
vr-rr (2 . 5V >

II

SV

C5 4

E 1

2 C2 IB

78.7KB

R5

R4B

lOOKB

J0D T

330OP

R 114

R53

PC4074

JL.

1-I

20K

'

1SS302

022

1 C 13B

I C2 1 C
39KB

(EaIIiZ>

10

Rl 13

UPC4064

7F.7KB

R5 1

Rl 12
UPC4064

UPC4074

30K
t-3

76.7KB

C55

2 - *Z .2V J
Cfe.25 V LSB)

22

S60P

El

4.

E 1
C 1

Pgr

t C45e
D 12B

Z 1

. Z

R130
Par

R 1D3

Pave

018

5 . IK

UNS112

R 1 OB
B2K

< t enp ysm 1

1 p,r 6

I cizc
5/M

P9r

"

C45HC

nToi

DRAWING NO.

TITLE

Main Board Circuit Diagram, 3/9


MODEL NO.

FX-2111

VOL >

523-2020
DATE

ASSEMBLY NO.

PCB-6409

6-9

95.01.08

<-i

6vn I Pa9,' 7

-E13Z> p,*. 3
-(Tfti-

TP60

<Tftf 1<rBFi

PaF

Page

Page

TPS 1

gg0

IC33

Page

>

030

Njn78L0SUa
TP69

yj

i"

rl

42

1SS302

:T2

1 55302

<n2!D-

ySKT.T .

VDD

R123

WV-

C 107 .

CiOl

22Z

0 . 1

T P7C

13K

Ci02

<zsso-

"
a

f *3

ci3e
33QP

2P

I5_

I 34

1SS302

HJI"179l 0SUn

E2

? R 12 1

7P71

921

P
R

T P4 9

CLK<-

RL20 I
13K <

029

3 3/2S<y

11

U7'>C74
3

430
D23B
3

Rioa

D33C '

'W

T JSKA-

inrn ao

T o

IC20B

H1.31

155302

42 .2

1SS302

C93

Pfi'se 2
3C37,39.02
Pauer

"

Pin

(?) |

TP72

R10S

R79

W**

IK

034A

'm;

1SS302

IC16

IC16B

TC7W04F

TC7W04F

"vV

2 . 2K
Cqfe

100C",
1000
IC16C
IClt
S

IC19C

C10O

O. 1

4584

2O0KH2

FQ

C74

I C29

PS26S2L I

NC

Pag

TC7W04F

1|
O . 04 7
0152

430

TPS

ICi90

C72

tC28
NC

P jge

<ftt1PDaTO

fS26S2L |

6
_

1|
0 047
.

V-

R149
430

7P9

ic j.9e
4

Past

IC30

5VF

PS26S2L

2
.

NC

[hp~

TPli

IC1.9C

R222
TC4SB4

R102

3K

15K

E2

DRAWING NO.

TITLE

Main Board Circuit Diagram, 4/9


MODEL NO.

FX-2111

523-2020
DATE

ASSEMBLY NO.

PCB-6409

6-11

95.01.08

V/OD

6
Rise
20K

P6

08370061000800

CKYOCERO

ELCO

I C 12F

R 159

13

2K

r 04584
C77

To
Senior

PCe-6239

E2
2

R160

% IK

7
C76

0 3O

I 270

ft"
a

I C27B

N.

02
0270

R ISO

R iS6

IC19D

IK

4584

L3k

ozeo

IK

155
ISS302

UNS212

1S5302

IC7U02F

S2K

IC7W02F

C7 1

Q .022

R 15 X

C 75

240K

7
C2

E2

E2

Page

R1S7

C2

C2

C2

0 1

R 153
ISO

R154
R 132

ISO

929

/N/V

P4

1 K

iC2oa

HI 33

2SA14e3

027

JN5212
S

as

>

C2

>CLK

cJ>-i

Y 74C7<I

250

1SS302

E2

HZ?

baov

GEEEE

UM51 12

V0C

PA9e

IK

026

UNS212

Ra9

Dr3A-6P-20SO

925

>

CI N

PP

VCOUT

p i

SIN

Pas*

NC
R 145

CX
VCOIN
[MH

16tC

R 1
R2

OCnO
ZEN

I N 1 -

Rl 40 I

Z
10NC

D T C

CG5

PC35B

Lj

C70
4 7 X2SV
.

R14*.

L2

HCFOUI
QUICNI

L2

cc

L2

R T
7

R137

LS
L2

C i

20K

C64

4 . T/SSV

T 24K

R 138
24K

R 135

I 4-3K
.

7
C2

i-<>

C2

C2

R13

5 1./-0. 2S J

CNO

5
1SS302

IN2-

R 1 42

lOOK
NCTT

a 14 7

10 OK

8148

IC260
D26A

IN2R 1 41

&

I OK

I C26B

P2

2NC

c i

a i

C 1

C62
. .

228

UN51 12

uPC 4 9 4 C5
C69

2200

R 1 43

5 IK

7K

7
C2

C2

ce3

R13G

0 . 1

1K

R13?

3&K

1 C 1 2A

K 1 44
1 K

1 t <i5b4

DRAWING NO.

TITLE

Main Board Circuit Diagram, 5/9


MODEL NO.

FX-2111

ASSEMBLY NO.

523-2020
DATE

PCB-6409

6-13

95.01.08

2SC

R 14
22; Jfcv

2i
6C

P5t

55
I!
r,

"

A
as

O NWI

5-a

ib

-a r mo

ei i

11
1

AC
AC

it vis

C 1

Si

H-<

"1

A?

il...

P*9#

17

;3

B9<

26

ST RST
'
ST OF

Y2

VOC

C2

V7

< KE Yd

OI127C512-1S6DC
tic

"D* 1 t.

E2

D7

Old

DC

D93

fid

2!

SI

It- e

* R92-Bfc

iJ

6iC

NSt

it

at

. NOTE J

B9-

0D

- 12
2

IDC

< CMC

5 T O T u

20K

2QK

F*q.
"

MS

g...ml!5iii MSfEE
rifM
ass -ea
jfrlff
3 ft?

IE si
etT

vol

BIO 1
240k

Thro 12

TmmDO

-S-5

-" P

i xo?

cS:
C5

f 10
t
A 1 1
aii

CSI
we

ft 12

oE

flTTu

[CI 1
i

JF3

<4

8gi ;
s-11
?

DE7

rH

01 R

>

1*

TMno=

-S-i

LOC

h 01

ISSI il

I C 1

19

f P30

< AQCLK
Fa**

Fag*

NOT
I OK
OMFD

sHhiigfsJiI

ill

NDFCISSKZR

0 12

5 ES S

It23F "":C23C
I C23D
C2

IlOTCH

IC2SO
i

[SJ2__CiiN>

>

R251

W0TC2

. NO 1 SI

R01

in

U*
U*>

No.

C 3:S 12Kbjt

<F_orr

ici-ac

f
c
Ktei 1 RAW 1t-.n connect JF1, JP3
256Kb.it RAM t *n connrct JF2.JF4

< 3: Jlb it

5 r -

{> fe

CHC

JKC__>

JPfc

R7?
9

1Kr>
Us*

Soc-t

connect

JPfc

infen

co*-.n*c-t

JPfc. JPB

DRAWING NO.

TITLE

Main Board Circuit Diagram, 6/9


MODEL NO.

FX-2111

523-2020
DATE

ASSEMBLY NO.

PCB-6409

6-15

95.01.08

01 -*

03

nC PCTCCT

rpBBOO?

"

"

D3
3

l a

IP M T f R T>

XT

@
0 3

2*

OS
D3f Si

CND

024

2S J

O 1

C 12
2

R.*

2900

BL. 0 1 BM 1 -Ofe 2

R3

C16

I
T

09

i'3

lit.

2DOM

< mo

'

S-

> 20K

AH

R2
3 4BK

UM5212

o-

NMRCN/NM

1-

08
Tf>8

ON

CI37

086

3?

M * - >JW

o IS

272

IOO/2?v
UM5ti2

( *2*v J

I OOP

C9

>

i
[C22

2
Ion .i

>

D 1

013

/3SV

/1*V

BJ IB

UN5i12

>

C17 4*

ik r
9 11

251
2

1553D1

C29

"t?

20K

No t kl

IPS?

J OuM

C 19

W8 L

ISS3D2

UN3212

>

Q 17

ZSJ2>SS
Di6a

D21

D.DJ
vi n

3L

-wout

oen

f 70uM

KB J
100

I
o- i

a4

TP2a

<n

>

7 2

R71

Dirs *

021

j k? uiza 12

12t(

IP3

Rt2

3 . 4 ex

OVEN
%

BID

MOO

2SK

T Ml

S
R63

C?

E2V
*

lODw

C3

VS5

I
T

c i

lOD

6q2oo3

0-15S30J

D-1SS102

*JK7aLOSU*

NJMT9LOSUA

2SJ24S

T P45

f f
1

OS-CON

?ft

out
OUf

I
i

GWD

IN

CA'C

In

COW
COHBON
Out

2
3

[>R*IN
SOURCE

DRA1n

DRAWING NO.

TITLE

Main Board Circuit Diagram, 7/9


MODEL NO.

FX-2111

523-2020
DATE

ASSEMBLY NO.

PCB-6409

6-17

95.01.08

P7

0637015 1000600

CKY0CERO

ELC0>

nop E
-

SCNSI
-

10

1 0

11

1 1

12

12

NC

SOU

SZL

NC

(key?

>

4XEY3

~>

fKE Y4

>

ST CRT /STOP
O

O-

1 mV

O-

INST
O

ON
-

P9*

P*9 c

&

(KEYS

(OF F /CMC

1 >

to

_> >
_

13

L 3

11

11

IS

IS

fc

O-

OFF/CHG
-

P * S r

{X E Yfc

>

NC
NC

E2

rxr2A.U..KKT. poncu

KEY

PRNEL

DRAWING NO.

TITLE

Main Board Circuit Diagram, 8/9


MODEL NO.

FX-2111

523-2020
DATE

ASSEMBLY NO.

PCB-6409

6-19

95.01.08

nxi 2 1 1 90 I
.

P2
n xg j i4qqi
VDO

6
LCO-PCB

Vtt
VOO

VCC

2
3

0F3-1SS-2C
15

1 5

l->

13

1 3

<sr

12

1 2

<R/M

1 1

1 1

10

10

<vS5

R /w
_

ooo
oe i
002

086
DB7

<e:

8
9

8
9

JO

1 I

1 1

12

1 2

13

1 3

1 <

1 4

OB L

<i b r ch

-rCLOCK

Page

POT POUT

STROBES

STW08CI

THERMIST OR

DB 2

VDD

0B3

CNQ

0B4

CNO

PBS

GNO

P85
0B7

connoN

COMnON

ConnpN

1
-

LCD
MODULE

T ST Ro6E

'

lO

OPS

CLOCK

"

DB3

081

DOT ft r N

PS
0F3-15P-2DS

Cpgo-oBfl

Page?

S ISB-Ph-k.-S;

THERMAL

ph5

10

10

1 1

I 1

12

1 2

L3

L 3

14

1 4

IS

1 S

tirnp

P .gp

>

C467
0
1

E2

<COMnON

P a<jr

Page

iS

HEPD

MODULE

DRAWING NO.

TITLE

Main Board Circuit Diagram, 9/9


MODEL NO.

FX-2111

523-2020
DATE

ASSEMBLY NO.

PCB-6409

6-21

95.01.08

n
DD

CD

O
KJ

cn
i.

I?

cnn
Dm

I; > !;<

Z
r
m

ML/1

"

CP
o
D
XI
O

1
9

s2

n
B

I (") C")

TITLE

DRAWING NO.

Power Board Circuit Diagram


MODEL NO.

FX-2111

1
B

ASSEMBLY NO.

524-2021
DATE

PCB-6410

6-23

96.01.08

s r>

3 5
-

IN

DRAWING NO.
TITLE

Sensor Board Circuit Diagram


MODEL NO.

FX-2111

524-1959
DATE

ASSEMBLY NO.

PCB-6239
6-24

94.07.14

NF-26P . /2'i
.

WED
2

r: i

to

n72

Or3i>-SP-2D50

n; 3
MT 4

ELU

s
6

MOTOR
MODULE

DRAWING NO.

TITLE

Motor Module Circuit Diagram


MODEL NO.

ASSEMBLY NO.

524-1960
DATE

94.07.14

FX-2111

6-25

VHfl-4M
(-

VHfi-ON

2
9 6V
.

HOOmOh

3
4

-I H H H

Ni-MH

'

BATTERY

PRCK

6-HRflRFD

DRAWING NO.

TITLE

Battery Pack Circuit Diagram


MODEL NO.

524-1961
DATE

ASSEMBLY NO.

94.07.14

FX-2111

6-26

F C P-2 2 2 2 A T

INLET BOARD

PI

PZ

0 0

P10

LCD

Ni-MH
i

P2

BATTERY
PACK

MOTOR

P3

NF-26P1/25

DC [ N

P4

PC8-6410
MOTOR CABLE

MA I N
THE RMAL
HEAD

THEMAL CABLE

B 0 A RD i-

P5
PATIENT

P8

CABLE

/
Q56-F

P6
P7

S W

SENSOR

BOARD

PANEL

F X - 2 1 1 1

PCB-6239

DRAWING NO.

TITLE

Overall Connection Diagram


MODEL NO.

ASSEMBLY NO.

433-5126
DATE

96.10.09

FX-2111 (USA)
6-27

FCP-2222W
INLET

BOARD

PI

L CD

: pio
i

Ni-MH
BATTERY
PACK

P2

u
PCB-6410

v
.

P2

MOTOR

P3

NF-26P1/25

DC TN

P4

MOTOR CABLE
MA IN
THE RMAL

THEMAL CABLE

BOARD

P5

HEAD

PATIENT

P8

CABLE

Q56-F

P6

P7

SW

SENSOR

BOARD

PANEL

F X-2 1 1 1

PCB-6239

DRAWING NO.

TITLE

Overall Connection Diagram


MODEL NO.

ASSEMBLY NO.

4334894
DATE

96.01.08

FX-2111 (CE)
6-29

SOLDER

NDM202AOO
14

>

AXL214901

33

(7)
m

TO

F4

TITLE

DRAWING NO.

LCD Connection Diagram


MODEL NO.

ASSEMBLY NO.

434-4895
DATE

96.01.08

FX-2111

6-31

MOTOR

NF-26P1/2 5

r
10

SUMITUBE F2
3

xO. 25

SDMI TUBE F2
6x 0. 25

o
HIRING

(BRN)

1 0 0 mm

HIRING

SOLDER

HIRING

(RED)

SOLDER

UL1007

AHG28
6

HIRING

LENGTH

BRN

connector

RED
ORG

1 7 0

contact

YEL
BLU
PUR

DF3-6S-2C

3 pin
4 pin
Spin
Spin

DF3-24 28SC

DRAWING NO.

TITLE

Motor Connection Diagram


MODEL NO.

1 6 Omml pin
14 0
2 pin

434-4896
DATE

ASSEMBLY NO.

96.01.08

FX-2111
6-32

CHAPTER

laqram
Main Board Assembly Diagram, 1/2 & 2/2 433-4897
Inlet Board Assembly Diagram
434-4899

7-1 to 7-3
7-5

HOW 10 MOUNT IC22


(PQ30RV11)

COMPONENT SIDE
PCS-6X09

C35

SILK CHART

is FLUX AREA .

65

rfi

.:iiiiiiisiTnfria
fp

CI

ir

ZD

? o

su

=s----flm?ji0

a
z
a

u-l
1

1 c

IC4

EES)
I

lO"'" FUKUOA DEN5HI CO

..

I *

i-i

how TOMOmr ic socket f en

PCB-6409

lr_

LTD.

- "li-=nt-rJ rn0 31

0.5

(ZDc
_

ICIO

225

AT

D
CI

HOW TO MOUNT 04 .at

.2

(2SJ245)
M
n

B1

1 3
.

2J

B2<

= x x X 1
I I 1
-

s [v' Pi
LED
ys'
L

m AREA :

PC BOARD

OPEN

MAX

(MOUNT

icia
3

I .i-l

IC1 :512K

IC1 : 1M)

Li

HOW 7t?MQUNT Cft?


CECHEW105JZ)

LEAD CUT LIMIT

MAX 1 5 mm
.

St

IC31

DRAWING NO.

TITLE

Main Board Assembly Diagram, 1/2


PC BOARD

MODEL NO.

FX-2111

ASSEMBLY NO.

433-4897
DATE

PCB-6409

OMPOfCNT SIDE
7-1

95.12.15

is FLUX AREA .

tV

jr''

W AREA ARE NOT REQUI

TITLE

DRAWING NO.

Main Board Assembly Diagram, 2/2


MODEL NO.

FX-2111

DATE

ASSEMBLY NO.

PCB-6409

0
m

>
X)

>
CD

6
CO
o
33

5
CO

>

ft
s

t
CD

CD
4

LTI

CD

O
<
ro
30
-

O
o

o
o

cz

-i

mu

be

3 -I

CO
m

TITLE

DRAWING NO.

Inlet Board Assembly Diagram


MODEL NO.

FX-2111

434-4899
DATE

ASSEMBLY NO.

PCB-6410

7-5

96.01.08

CHAPTER

Electrical Parts Lists


1 FX-2111 Main Blocks

674-11338

8-1

2 Main Board, PCB-6409 SAS SMD for FX-2111

674-11339

8-2

3 Main Board, PCB-6409 THD for FX-2111

674-11340

8-9

AC Inlet Board, PCB-6410 THD for FX-2111 (115V)... 674-11664

8-11

AC Inlet Board, PCB-6410 THD for FX-2111 (230V)... 674-11340

8-12

LCD Assembly for FX-2111


Parts Supplied to Wiring/Assembling Process

674-11342

8-13

674-11343

8-14

for FX-2111

1 FX-2111 Main Blocks


.

674-11338

Parts No.

9E3041

Description

Symbol

Drawing No.

PC Board AS

Q'ty

Main Board
1

9F3419

PCB-6409 with ROM

9F3293

ROM Set

9F3424

PCB-6409 SAS

9G3739

PCB-6409 SMD

674-11339

9G3740

PCB-6409 THD

674-11340

674-10486

AC Inlet Board
9F3420

PCB-6410 Assembly

674-11341

9F3421

LCD Assembly

674-11342

9D0995

Set of Supplied Parts to Assembling Process (CE)

9E2793

Set of Assembled Electrical Parts (2307)

674-10513

9E3042

Set of Assembled Electrical Parts (CE)

674-11359

9D0996

Set of Supplied Parts to Wiring Process (CE)

674-11343

8-1

Remarks

2 Main Board, PCB-6409 SAS SMD for FX-2111


.

674-11339

Parts No.

Drawing No.

Description

Symbol

Q'ty

Resistors

1A7510A

R1, 15, 23,


75-77,
145, 163,
168, 176,
181, 192,
197, 200,
202, 207,
209, 212,
219, 220,
226, 227,
229, 230,
232, 236,
239, 242,

RK73H2AF, Square Chip

lOkfi

30

RK73H2AF, Square Chip

20kfi

51

245
1A7503A

R2~11, 14,

26, 29, 30,


34-40,
43, 46 57,
68-70,
86-92,
140, 158,
182, 183,
189-191,
,

195,
201,
210,
225,

196,
204,
211,
241,

254, 255

1A7685A

R12, 84, 110

RK73H2AF, Square Chip

12kG

1A9662A

R13, 41, 51,


72, 80,
199, 206,

RK73H2AF, Square Chip

30kn

RK73H2AF, Square Chip

560kn

RK73H2AF, Square Chip

200kn

RK73H2AF, Square Chip

3kil

RK73H2AF, Square Chip

1kfi

14

215
1A7641A

R16, 117,
148

1A9666A

R17, 32, 58,


116, 169,
173, 231,
252

1A9657A

R19, 59,
119, 164,
222

1A9655A

R20, 28, 64,


65, 67, 82,
93-100,
105, 109,
132, 133,
136, 144,
150, 156,
217, 248

8-2

Remarks

2 Main Board, PCB-6409 SAS SMD for FX-2111. continued


.

674-11339

Parts No.

Symbol

Description

Drawing No.

Q'ty

1A7512A

R85, 147

RK73H2AF, Square Chip

16kn

1A7502A

R22

RK73H2AF, Square Chip

6 8kG

1A9263A

R24

SR73K2EJ, Metal Film

0 1fi/0.5W

1A7522A

R25, 63, 81,

RK73H2AF, Square Chip

lOOkfi

RK73H2AF, Square Chip

5 1k0

10

141,142
1A7545A

R27, 61,
103, 171,
174, 214,
216, 228,

249, 250
1A7571A

R31

RK73H2AF, Square Chip

300fl

1A9631A

R33

RK73K2AF, Square Chip

100

1A7534A

R42, 62

RK73H2AF, Square Chip

3 48kU

1A7584A

R44, 45, 56,


78, 124,
125, 165,

RK73H2AF, Square Chip

1Mfi

11

RK73H2AF, Square Chip

1001]

RK73H2AF, Square Chip

2kU

14

166

218,

244, 246
1A9653A

R47, 48, 55,


74, 115,
161

1A9656A

R49, 50, 52,


71, 73,
126-129,
143, 159,
178, 179,
187

1A9396

R53

RN73F2ATD-B, Metal Film

1A9398

R54

RN73F24TD-B, Metal Film

1A9303A

R60

RH73H2AF, Square Chip

28.7k(]

1A9271

R66

RK73H3AF, Metal Film

IkO/IW

1A7681A

R79

RK73H2AF, Square Chip

2 2kQ

R7519A

R83, 146,
160, 193,

RK73H2AF, Square Chip

51 kl)

15

198,
205,
213,
223,
240,

203,
208,
221,
237,
243,

247

8-3

Remarks

2 Main Board, PCB6409 SAS SMD for FX-2111., continued


.

674-11339

Parts No.

Symbol

Drawing No.

Description

Q'ty

1A7547A

R85, 147

RK73H2AF, Square Chip

16kn

1A7656A

R101, 151,

RK73H2AF, Square Chip

240ki]

162
1A7558A

R104, 122

RK73K2AF, Square Chip

42.2fl

1A7689A

R106, 108,
111, 155,

RK73H2AF, Square Chip

82kn

167

1A9665A

R107

RK73H2AF, Square Chip

120kQ

1A9397

R112~114

RN73F2ATD-B, Metal Film

78.7ki]

1A7674A

R118, 121,
123, 149,

RK73H2AF, Square Chip

430n

RK73H2AF, Square Chip

13kn

152
1A7652A

R120, 131,
157, 238

1A7616A

R130

RK73H2AF, Square Chip

56kn

1A9270A

R134

SR73H2EF, Metal Film

0 51Q/0.25W

1A7604A

R135

RK73H2AF, Square Chip

1 43kl

1A7540A

R137, 138,

RK73H2AF, Square Chip

24kQ

233, 251
1A7554A

R139

RK73H2AF, Square Chip

36kfi

1A7578A

R153, 154

RK73H2AF, Square Chip

150fi

1A7202

R170

TRN60LG, Cylindrical Tantalum Alloy

1A7591A

R175, 180,

RK73H2AF, Square Chip

3 5Mn

160kfi

234, 253
1A7526A

R172, 177

RK73H2AF, Square Chip

1A9393

R184

RN73F2ATD-B, Metal Film

11011

1A7640A

R185, 186

RK73H2AF, Square Chip

140kfi

1A7587A

R188

RK73H2AF, Square Chip

3 4kQ

1A9395

R194

RN73F2ATD-B, Metal Film

13kn

1A7668A

R256~259

RK73H2AF, Square Chip

220n

1A7528A

R224

RK73H2AF, Square Chip

698kH

8-4

Remarks

2 Main Board, PCB-6409 SAS SMD for FX-2111, continued


.

674-11339
Parts No.

Description

Symbol

Drawing No.

Q'ty

Capacitors
1B5021A

01,2, 4, 6,
7 11, 13,
14, 22, 25,
26, 28-34,
36, 37, 39,
40, 42, 43,

GRM40F104Z50, Layered Ceramic Chip

0.1 pF

59

45, 49, 50,


52, 53,
61 -63,
66-68,
73, 76, 77,
85, 86, 89,

93, 94, 96,


97, 99,
100,

105-108,
116-119,
125-128
1B1701

C3

ECSTCY105, Tantalum Electrolytic 1.0(jF

1B5412A

C5, 27

EEV-H81CV220, Aluminum Electrolytic 22nF/16V

1B5408A

C8

EEV-HBIHV0010, Aluminum Electrolytic 1mF/50V

1B5073A

C9, 16, 41,


56, 58, 87,

GRM40CH1U2J50, Layered Ceramic

lOOOpF

88
1B0702

1B5055A

C10

C12, 51, 69,

25CV100GX, Aluminum Electrolytic Chip


100uF/25V
for High Frequency

GRM40B222K, Layered Ceramic

2200pF

134
1B0209

C15

16SM6R8M, OS

6.8mF/16V

1B0208

C17, 21

10SM22M, OS

22tiF/10V

1B5019A

C18

GRM40CH101J, Layered Ceramic

100pF

1B5022A

C18, 20, 57,


59, 60, 65,

GRM40B103K, Layered Ceramic

O.OIpF

MCH215C223K, Layered Ceramic

0.022mF

51pF

75, 115
1B5062

C23, 24, 38,


47, 71

1B5002A

C35

GRM40CH510J, Layered Ceramic

1B8501

C44, 90, 91

35CV47GX, Aluminum Electrolytic Chip


for High Frequency
47pF/35V

1B5815

C46

ECHU1H102G, Layered Film

8-5

1000pF

3
1

Remarks

2 Main Board, PCB-6409 SAS SMD for FX-2111


.

674-11339
Parts No.

1B5027

Description

Symbol

C48, 131,

Drawing No.

MCH215A561J, Layered Ceramic

3300pF

Q'ty

133
1B5816

C54

ECHU1C332G, Layered Film

1B5402A

C64, 70, 95

EEV-HB1 EV4R, Aluminum Electrolytic


4 7nF/25V

1B5513A

C55, 83

GRM42-6F224Z, Layered Ceramic

0.22nF

1B5011A

C72, 74

GRM42-6B473K, Layered Ceramic

0.047mF

1B5818

C78

ECHU1C683J, Layered Film

0.068mF

1B5261

C79, 80, 92

ECWU1C224J, Layered Film

0.22nF

1B5076A

C81, 100,
110, 112,

GRM40B332K, Layered Ceramic

3300pF

124, 135

1B5044A

C84

GRM40B153K, Layered Ceramic

0.015nF

1B5003A

C98, 103,
104, 113,
114, 120,
121, 123,

GRM40CH101J, Layered Ceramic

150pF

129
1B5403A

C101, 102

EEV-HB1EV330, Aluminum Electrolytic


4 7nF/16V

1B5808

C111,113

ECHU1H140J, Layered Film

O.lpF

1B5014C

C130

MCH325F274Z, Layered Ceramic

0.47nF

1B5001A

C132

GRM40CH300J, Layered Ceramic

30pF

1B5074A

C138

GRM40CH331J, Layered Ceramic

330pF

Diodes
3929

D1~6

D3FS4A (2.6A, 40V)

1D4200

D7

02CZ-5.1V, Zener

1B3885

D8, 1, 15, 32

1SS301

1D3886

D9, 11, 12,


16, 17, 19,
20, 23-31,

1SS302

18

33, 34
1D4113

D10

02CZ-3.0K, Zener

1D3928

D13, 14, 18,

D1FS4A

21

8-6

Remarks

2 Main Board, PCB-6409 SAS SMD for FX-2111, continued


.

674-11339
Parts No.

1D4552

Description

Symbol

D22

U1ZB12, Zener

Drawing No.

Q'ty

Transistors
1D1735

Q1~3, 5, 8,
12 14, 15,
27, 29, 30,

UN5212

13

33, 38

1D989

Q6

2SC-2713GR

1D0116

Q7

2SA-1163G

1D2385

Q9

2SK-739-2, FET

1D2392

Q10

2SK-1062, FET

1D0117

Q11

2SA-1162V

1D1734

Q13, 18, 26,


28, 42, 43

UN5112, Digital

1D0968

Q16

2SC-1623(L5/L6)

1D2394

Q17

2SJ245S, FET

1D2382

Q19, 22, 23,


31, 32, 34

SST4393, FET

1D2342

Q20, 21

2SK1920-FA, FET

1D0153

Q25

2SA-1463(1 K)

1D1713

Q35~37,
39-41,

FMW2

44-46

Integrated Circuits
1E5037A

IC2, 5

HM6264BLFP-10LT, RAM

1E2769

IC4

NJM2103M, Reset IC

1E6168

IC6

HA16120FP, Switching Regulator

1E6245

IC7

PQ05SZ5, Low-loss Regulator

IE2621

ICS

FD88007-AC, Gate Array

1E6235

IC9, 33

NJM78L05UA, 3-terminal Regulator

1E7205

IC10

S-8437AF

1E5390A

IC11

TC-74HC245AF, CMOS Logic

Inversion-type Switching Regulator

8-7

Remarks

2 Main Board, PCB-6409 SAS SMD for FX-2111, continued


.

674-11339

Parts No.

Drawing No.

Description

Symbol

Q'ty

1E4424

IC12, 19

TC-4584BF, CMOS Logic

1E0374

IC13

HPC4074G2, Low-noise FET-input OP Amp.

1E5397A

IC14, 23

TC-74HC04AF, CMOS Logic

1E5412A

IC15

TC-74HC541AF, CMOS Logic

1E6167

IC18

bq2003S> Quick Charging IC

1E5543

IC16

TC7W04F, L-MOS Logic

1E4169

IC17

TC-4053BF, CMOS Logic

1E5595

IC20

TC-74AC74F, CMOS Logic

1E0198

IC21, 31

pPC4064G2, Low Power CoSsumption J-FET


Input Operational Amplifier

1E0715

IC24

pPC4064G2(2), Low Power Oonsumption J-FET


Input Operational Amplifier

1E2613A

IC25

HEF4046BT, PLL

1E0266

IC26

MPC358G2, Comparator

1E5542

IC27

TC7W02F, L-MOS Logic

1E8961

IC28~30

PS2652L2-V(K), Photo Coupler

1E2083

IC32

LT1004CS8-1.2, Voltage Reference

1E7221

IC34

NJM79L05UA, 3-terminal Regulator

1E4177

IC35

TC-4093BF, CMOS Logic

1E4170

IC36, 38, 41,

TC-4051BF, CMOS Logic

45

1E0264

IC37, 39, 42

MPC4064G2(2), Low Power Oonsumption J-FET


Input Operational Amplifier

1E4423

IC40

TC-4535BF, CMOS Logic

1E0365

IC43

pPC339G2, Comparator

1E4182

IC44

TC-4094BF, CMOS Logic

Coil
1T6047

1H6409

L2

470mH

LPC4045TE-471K

Printed Circuit Board, PCG-6409

8-8

Remarks

3 Main Board, PCB-6409 THD for FX-2111


.

674-11340

Description

Symbol

Parts No.

Drawing No.

Q'ty

Resistor
1A7040

R18

RSSX3L20 0.01J/3W, Metal Film

Capacitors
1B4701

C82

ECHE1H105JZ, Metalized

1B0866

IC136

URS1E222MHZ 2200/25V, Aluminum Electrolytic

1B0713

CI37

URS1V102MHZ 1000/35V, Aluminum Electrolytic

Transistors
1D2393

Q4, 24

2SJ245L, FET

Crystal Oscillator
1D7620

XI

JX0-5S 16MHz (with stand)

Integrated Circuits
1E1957

IC3

HD63B09E, CPU

1E7021

IC22

PQ30RV1/11, Low-loss Regulator

1E9198

DIC?iG11S1 IC Socket

1E9199

DIC640G11S1, IC Socket

Lithium Battery
1U0231

BT1

CR-1/3-FT2-1

684-2428

Connectors
1F0215

P1

B3P-VH

1F1068

P2

AXB114001 BB

1F3208

P3

S-G8036 #01

1F0562

P4

DF3A-6P-2DSA

1F9376

P5

DF3A-15P-2DSA

1F0548

P6

008370061000800

1F1131

P7

008370151000800

1F1004

P8

RDAG-15SE1(F)

8-9

Remarks

3 Main Board, PCB-6409 THD, continued


.

674-11340
Parts No.

1F0365

P10

Drawing No.

Description

Symbol

B4P-VH

Q'ty

Arrestor

1L4807A

NE1

Y06S-230B

Fuses
1L4057

F1, 2

3 15A

TR-5 K19374

Coils & Ferrite Core


1T6242

L1

HK-05S035-2010 RBP

200mH

684-2109

1T6243

L3

HK-04D030-1510 RBP

150nH

684-2108

1L4584

L4

BL01RN1-A62, Ferrite Core

Isolation Transformers
1T6015A

T1

FT-2111

684-2153

1T6016A

T2

FT-2155

684-2154

Piezoelectric Buzzer
1S3200

BZ1

PKM13EPP-4002

8-10

Remarks

AC Inlet Board, PCB-6410 THD for FX-2111 (115V)


674-11664

Parts No.

Symbol

Description

Drawing No.

Q'ty

Fuses
1L4476

239001

1L4905

MF-561A, Fuse Holder

1A

Locker Switch

1G3508

SJ-2S4A-07BB

AC Inlet
1D2393

NC176-1.5

GND Terminal
1K1151

FX-2111GND (A)

Connector
1F0159

B2P-VH

Choke Coil
1T6342

TF1028S-801Y2R0-01

Spacer
1Z4311

1/2

LR-1.5

Printed Circuit Board


1H6410

323-3961

PCB-6410

8-11

Remarks

AC Inlet Board, PCB-6410 THD for FX-2111 (230V)


674-11340

Parts No.

Drawing No.

Description

Symbol

Q'ty

Fuses

1L4091

No. 19195

1L4905

MF-561A, Fuse Holder

500mA

2
2

Locker Switch

1G3508

SJ-2S4A-07BB

AC Inlet
1D0288

NC176-1.5

GND Terminal
1K1151A

314-3471

GND (A)

Connector

1F0159

B2P-VH

Choke Coil

1T6342

TF10283-801Y2R0-O1

Spacer
1Z4311

1/2

LR-1.5

Printed Circuit Board


1H6410

PCB-6410

8-12

Remarks

LCD Assembly for FX-2111


674-11342

Parts No.

Symbol

Description

Drawing No.

Q'ty

LCD

1 LI 709

NDM202A00

Connector
1F1085

AXL214901

Spacers
6B8308

SP-15

6B9216

Grounding Piece

Shielding Fingers
6B9217

(A)

6B9218

(B)

8-13

Remarks

Parts Supplied to Wiring/Assembling Process for FX-2111


674-11343

Parts No.

Description

Symbol

Drawing No.

Q'ty

434-4725

1M0385

Motor Assembly

1M0466

Inlet Board Grouding Cable 3 with Core

1M0482

Inlet Board Grounding Cable 3

1L4581

Ferrite Core, 5977-000601

6G0234

Heat Shrink Tube, F2 25x0.4, 30mm long

1K0031

Press-fit Terminal, TMEV1.25-3

8-14

684-2415

Remarks

CHAPTER

Spare Parts Lists


PCB-6238
2 PCB-6236

674-10554
674-10555

9-1
9-2

Paper Magazine

674-10556

9-3

Recorder

674-10557

9-4

Upper Case

674-10558

9-5

Lower Case
7 Screws

674-10559
674-10560

9-6
9-7

1 PCB-6238
.

674-10554
Parts No.

Symbol

Description

Drawing No.

Q'ty

9E2963

PCB-6238B (2) Assembly for 115V

9F3331

PCB-6238 Assembly for 115V

5T3111

Nuts, Type 3 M6

5T2009

Spring Washer, Nominal Dia. 6mm

9E2964

PCB-6238B (3) Assembly for 230V

9F3310

PCB-6238 Assembly for 230V

5T3111

Nut, Type 3 M6

5T2009

Spring Washer, Nominal Dia. 6mm

1L4476

Fuse, 239001 for 115V

1L4475

Fuse, 239500 for 230V

9-1

Remarks

2 PCB-6236
.

674-10555
Parts No.

Symbol

Description

Drawing No.

Q'ty

9E2965

PCB-6236 (1) Assembly

9F3307

PCB-6236 Assembly for FX-2111(N)

5R1604

Double Semus Screw, M3x6

9E2966

PCB-6236 (2) Assembly

9F3309

PCB-6236 (with ROM) for FX-2111 (N)

5R1604

Double Semus Screw, M3x6

1E8531

Programmed ROM for FX-2111 (36-202)

1L4057

Fuse, TR-5 K19374, 3.15A

1U0231

Lithium Battery, CR-1/3-FT2-1

684-2428

9F3305

LCD Assembly

634-4726

9F3305

Spacer, SP-15

9-2

Remarks

Paper Magazine
674-10556

Parts No.

Symbol

Description

Drawing No.

Q'ty

9H0402

Platen Roll Assembly

6B8324A

Platen Roll

6A5350

Bearing, LF-740ZZ

6B8325

Gear, 38A

5R3003

Countersunk Screw, M2x6

9F2648

Paper Magazine Assembly

6B8323

Paper Magazine

9H0402

Platen Roll Assembly

224-0890

224-0313

222-0889

9-3

Remarks

4 Recorder
.

674-10557
Parts No.

Description

Symbol

Drawing No.

Q'ty

9H0617

Gear 38B Assembly

6B8320

Gear, 38B

5R8001

Setscrew, M3x3

9H0812

Thermal Array Head Assembly

6B8316

Thermal Array Head, Q56-F

684-2092

6B8316A

Thermal Array Head Mounting Plate

213-3448

6B8394

Tension Coil Spring, DE542

5R0602

Semus Screw, M3x5

9F3354

Recorder Assembly

9H0812

Thermal Array Head Assembly

9H3215

Recorder Chassis SU

1M0385

Motor Assembly

9H0617

Gear 38B Assembly

9H3216

Grounding Piece (C) SU

6B8307

Grounding Flexible Conductor

6B8395

Sensor Presser Sponge

9F3289B

PCB-6295SMD Sensor Board

1M0381

Thermal Array Head Cable

5R1851

Double Semus Screw, M3x4

5R1604

Double Semus Screw, M3x8

9H3522

Paper Shaft SU

244-0314

313- 3461

6B8307

314- 3465

674-10490

684-2093

224-0891

9-4

Remarks

Upper Case
674-10558

Parts No.

Description

Symbol

Drawing No.

Q'ty

9F3355

Upper Case Assembly

9H3211C

Upper Case SU

111-6621

6B8305

LCD Filter

114-6625

6B9171

Lock Spring

114-6811

1G9340A

Key Panel

684-2146

5H9128

Rated Power Label (B) for 115V

154-4573

5H9130

Rated Power Label (C) for 230V

154-4575

9-5

Remarks

6 Lower Case
.

674-10559
Parts No.

Symbol

Description

Drawing No.

Q'ty

9F3356

Lower Case Assembly (1)

9H3212C

Lower Case SU

111-6622

6B8306

Insulation Sheet

314-3463

6B8327

Batter Presser Sponge

314-3466

9H3213A

Grounding Piece (A) SU

313-3459

1T1558

Transformer, FCP2222AT for 115V

684-2158

1T1559

Transformer, FCP-2222UT for 230V

684-2157

6B8326A

Battery Cover

112-6624

5H9127

Instructions Label (B)

154-4572

5H9138

Instructions Label (C) for 230V

154-4577

5H9129

Fuse Label (B) for 115V

154-4574

5H9131

Fuse Label (C) for 230V

154-4576

5H5639

Lithium Battery Label

154-3574

6B6796

TM-166 No. 12

9-6

Remarks

7 Screws
.

674-10560
Parts No.

Symbol

Description

Drawing No.

Q'ty

9H3238

Set of Screws for Ni-MH

5R1604

Double Semus Screw, M3x8

5R0602

Semus Screw, M3x5

5R1851

Double Semus Screw, M3x4

5R3003

Countersunk Screw, M2x6

5R8001

Setscrew, M3x3

9-7

1
24

Remarks

CHAPTER

10

Structural Diagrams
FX-2111 (USA)
External Appearance (1)
External Appearance (2)

413-0994
413-0995

10-1
10-3

424-3619

10-5

423-3620
423-3621
422-3622
422-3623
412-0996

10-7
10-9
10-11
10-13
10-15

413-0849
413-0927

10-17
10-19

AC Inlet Board

424-3443

10-21

Main Board*

423-3444

10-23

Paper Magazine Assembly*


Paper Shaft Assembly*
Recorder Assembly*
Upper Case Assembly
Lower Case Assembly (1)
Lower Case Assembly (2)
Lower Case Assembly (3)
Exploded View

423-3250
424-3251
423-3252
423-3445
423-3446
423-3447
422-3448
412-0928

10-25
10-27
10-29
10-31
10-33
10-35
10-37
10-39

AC Inlet Board

Upper Case
Lower Case Assembly (1)
Lower Case Assembly (2)
Lower Case Assembly (3)
Exploded View
FX-2111 (CE)
External Appearance (1)
External Appearance (2)

>

Asterisked diagrams are common to both FX-2111 (USA) and FX-2111 (CE).

3sg

II
8

.S5

ii
A

Of

as

TO

[J
a

DRAWING NO.

TITLE

External Appearance (1)


MODEL NO.

ASSEMBLY NO.

413-0994
DATE

96.09.30

FX-2111 (USA)
10-1

JUL

JUL

0)

Q
ClITIOI-Ti
II.t

r iin

iti(tci|

til
It

r i 11

if

dtclric ibid, ti

t*tIi(i d

if f i 111

ul

rii*Tc

tirtr.

firiooi. 1,

D
DASOBI-liil *f ciflniiB
flmtblt

<1

DA NG B K - 1 i ii it

o -

if ttti
Ii

'

pruiii* d lBtltkttilstt

51

10DE

1ISBT

<3

taV

ei

2jnv

Mil

IS

STOP

3J

dpUr*'

ilflilBikltl.

HAD

START
DFF

the grtietct tf

ruse

$1

C A R D I MA X

Dl

pn

HIJITIIltT

[ Vrjcuwdensh)
Pi- 2 111

ii

ittilkitici.

d'tiplntoi.

2 6 0

CA8T10R:FBD8RAl Ul SISTIICTS T8IS DtVICt TO SALS


BT 01 01 71! 0IDI1 01 A PB1S1CIA1.

Tizr

6 5mm
DRAWING NO.

TITLE

External Appearance (2)


MODEL NO.

ASSEMBLY NO.

413-0995
DATE

96.09.30

FX-2111 (USA)
10-3

\
a*
o

2-M8 Nut & Plain Washer

Key
No.

Parts No.

1L4476

1K1151A

Description

Drawing Na

314-3471

Fuse, 239001

GND Terminal (A)

AC inlet Board

FX-2111 (USA)

Remarks

DRAWING NO.

TITLE

MODEL NO.

Q'ty

424-3619
DATE

ASSEMBLY NO.

PC 6-6410

10-5

96.09.30

''

A
-

55

33

ti

Si'S
II

Of

S!|!Li

TO
8

Key
No.

Parts Na

Description

Drawing Na

Q'ty

Upper Case Assembly

9H3494

111-6862

Upper Case Assembly SU for USA

1G9340A

684-2146

Key Panel

6B8305

114-6625

LCD Filter

5H9128

154-4573

Rated Power Label (B)

6B9171

114-6811

Lock Spring

5H9237

154-4670

Caution Label

Remarks

DRAWING NO.

TITLE

Upper Case
MODEL NO.

ASSEMBLY NO.

423-3620
DATE

96.09.30

FX-2111 (USA)
10-7

4
2-Double Semus Screw M3x8
,

2-Taping

Pan-head Screw M2.6x6


,

6
A

A
C

3
Pasted

2
A

C
f_

TV

Key
Na

Parts Na

Description

Drawing Na

Q'ty

Lower Case Assembly (1)

9H3495

111-6863

Lower Case Assembly SU for USA

6B8327

314-3466

Battery Presser Sponge

6B8326

112-6624

Battery Cover

5H9374

154-4669

Fuse Label (D)

9E7329

Rating label

Remarks

DRAWING NO.

TITLE

Lower Case Assembly (1)


MODEL NO.

ASSEMBLY NO.

423-3621
DATE

96.09.30

FX-2111 (USA)
10-9

Key
No.

Drawing No.

Parts No.

Q'ty

Lower Case Assembly (1)

423- 3621

Lower Case Assembly (1)

9H3213A

313- 3459

Grounding Piece

684-2156

Transformer, FCP2222AT

Description

1T1558

9F3420

424- 3619

PCB-6410 Assembly

6B8306

314- 3463

Insulation Sheet

6B9198A

314-3601

Insulation Cover

Remarks

.v

3
B

3
o

0
6

DRAWING NO.

TITLE

Lower Case Assembly (2)


MODEL NO.

ASSEMBLY NO.

422-3622
DATE

96.09.30

FX-2111 (USA)
10-11

9-Double Semus Screw M3x8


,

IF
D

D
A

it

5
B

F
O

11

0
-1
p

6
V

Key
Na

Parts Na

Description

Drawing Na

Q'ty

422- 3622

Lower Case Assembly (2)

9F3354

423- 3252

Recorder Assembly

1M0381

684-2093

Thermal Array Head Cable

1T1558

423-3444

PCB-6409 Assembly

Remarks

DRAWING NO.

TITLE

5H9615

154-4836

Lead Label (C)

5H5639

154-3574

Lithium Battery Label

5H9417

154-4821

Patient Label

Lower Case Assembly (3)


MODEL NO.

ASSEMBLY NO.

422-3623
DATE

96.09.30

FX-2111 (USA)
10-13

5S>
'

II
D
At

3*

Li
6

Double Semus Screw

M3x8

ii
s"

2-D-jubl

e Semus Screw M3x8

11

Key
Na

Parts No.

Description

Drawing Na

Q'ty

Remarks
DRAWING NO.

TITLE

423-3623

Lower Case Assembly (3)

423-3620

Upper Case Assembly

423-3250

Paper Magazine Assembly

3
4

9F2648
9H3522

224-0891

Paper Shaft SU

Exploded View
MODEL NO.

ASSEMBLY NO.

412-0996
DATE

96.09.30

FX-2111 (USA)
10-15

2?5

>5-

DRAWING NO.

TITLE

External Appearance (1)


MODEL NO.

ASSEMBLY NO.

413-0849
DATE

94.10.07

FX-2111 (CE)
10-17

/Ft,

o
Cilil01-Ti

Itfir

TtAtEi

itttitin

tki

11

nil

ilttltit

tiflid

tkit

nrvlci

t* tit

rtiif*

ititt.

pt r 11 d i * 1.

u
DkHQtl'litl

If

tlfltiitt

f 11 mm t k It

if tie*

ii tk* frimci sf

(inlbitiei.

D A B C E S -1 i n *'h|Iiiii. It in tMplertf >


ptititu d inttl*Miii iifliaiiMn.

I 2TI

5i

oo

1DDI

UISITITITT

5 I

;
CA R D I MA X
FX-Z11 J

IBID
'

11
oil:?:
"

STAKT
PIT

OTf

A
STOP

3J
v

!=.

ii

si

2 6 0

nzr

DRAWING NO.

TITLE

External Appearance (2)


MODEL NO.

ASSEMBLY NO.

413-0927
DATE

96.01.24

FX-2111 (CE)
10-19

2_M8 Nut & Plain Washer

T
T

Key
No.

Parts Na

1L4475

1K1151A

Drawing Na

314-3471

Description
Fuse, 239500

GND Terminal (A)

TITLE

FX-2111 (CE)

Remarks

DRAWING NO.

AC Inlet Board
MODEL NO.

Q'ty

424-3443
DATE

ASSEMBLY NO.

PCB-6410

10-21

96.01.24

2-Binding

Screw M3x8
,

7
r

CD

.j
3
o
o

v.

6
y
y

y
y

4
y

y
y

y
y

0
y

2-Double Semus Screw M3x8


,

Key
No.

Parts No.

Drawing No.

Description

Q'ty

9F3424

PCB-6409 SAS

9F3419

PCB-6409 with ROM

1L4057

Fuse, TR-5 K1937, 3.15A

1U0231

Lithium Battery, CR-1/3N-FT-2-1

1E8631

Programmed ROM, FX-2111 (36-202)

9F3421

LCD Assembly

1L1709

LCD, NDM202A0O

6B8308

Spaeer, SP-15

6B9216

314-3602

Grounding Piece (D)

10

6B9217

314-3603

Shielding Finger (A)

684-2428

Remarks

TITLE

11

6B9218

314-3604

Shielding Finger (B)

DRAWING NO.

Main Board
MODEL NO.

ASSEMBLY NO.

FX-2111 (CE)

423-3444
DATE

96.01.24

10-23

5
Countersunk Screw M2x6
,

J6

2
3

Key

Parts No.

Drawing No.

6B8323

222-0889

9H0402

Na

6B8324A

6A5350

6B8325

Platen Roll

224-0313

Description
Paper Magazine

Platen Roll Assembly

Remarks

LF-740ZZ Bearing

Gear, 38A

DRAWING NO.

TITLE

Paper Magazine Assembly


MODEL NO.

Q'ty

ASSEMBLY NO.

423-3250
DATE

94.10.01

FX-2111

10-25

Key

Description

Q'ty

Parts Na

Drawing Na

6B8321

223-0892

Paper Shaft

6B8322

224-0893

Back Tension Spring

No.

DRAWING NO.

TITLE

Paper Shaft Assembly


MODEL NO.

Remarks

ASSEMBLY NO.

424-3251
DATE

94.10.07

FX-2111

10-27

2-Double Semus Screw M3x5


,

1
1

Setscrew, M3x3

)
a

1
v

Key
Na

1 1
9
Double Semus Screw M3x4
,

2
D
3-Double Semus Screw, M3x8

Parts Na

Drawing Na

Description

Q'ty

9H3215

211-3447

Recorder Chassis SU

1M0385

434-4725

Motor Assembly

9H0617

Gear 38B Assembly

6B8320

Gear 38B

9H0812

Thermal Array Head Assembly

6B8316A

Thermal Array Head Mounting Plate

6B8317

Tension Coil Spring, DE511

224-0314

213-3448

Thermal Array Head Q56-F

PCB-6239 SMD, Sensor Board

313- 3461

Grounding Piece C SU

6B8395

314- 3465

Sensor Presser Sponge

6B8307

314-3464

Flexible Grounding Conductor

1W3006

684-2092

9F3289B

674-10490

10

9H3216

11
12

TITLE

Remarks

DRAWING NO.

Recorder Assembly

423-3252

Double Semus Screw,


M3x8

MODEL NO.

ASSEMBLY NO.

FX-2111

DATE

94.10.07

10-29

II

II
It

.St

.4>

2
4

Key
Na

Parts Na

Drawing Na

9F3355

9H3211C

111-6621

Description

Q'ty

Upper Case Assembly

Upper Case Assembly SU

1G9340A

684-2146

Key Panel

6B8305

114-6625

LCD Filter

5H9130

154-4575

Rated Power Label (C)

6B9171

114-6811

Lock Spring

5H9428

154-4694

CE Marking Label

Remarks

TITLE

DRAWING NO.

Upper Case Assembly


MODEL NO.

ASSEMBLY NO.

FX-2111 (CE)

423-3445
DATE

96.01.24

10-31

b
2-Double Semus Screw

M3x8

rJ5

A
A

4
<S>

Pasted

<s>

1/

OS

Key
No.

Parts Na

Description

Drawing Na

'

Q ty

9F3356

Lower Case Assembly (1)

9H3212D

111-6622

Lower Case Assembly SU

6B8327

314-3466

Battery Presser Sponge

6B8326

112-6624

Battery Cover

5H9131

154-4576

Fuse Labell

9E7293

Remarks

TITLE

Rating Label

Lower Case Assembly (1)


MODEL NO.

DRAWING NO.

ASSEMBLY NO.

FX-2111 (CE)

423-3446
DATE

96.01.24

10-33

Key

Parts Na

No.

Drawing Na

9F3356

9H3213A

313- 3459

1T1559

684-2157

Transformer FCP2222UT

9F3420

424- 3443

PCB-6410 Assembly

6B8306

314- 3463

Insulation Sheet

Insulation Cover

TM-166 No 12

2
4

Q'ty

Lower Case Assembly (1)


Lower Case Assembly (1)
Grounding Piece (A) SU

B
A

Description

423- 3446

1
1

6B91108A

6B6796

314-3601

Remarks

1
1

IB

ft

8
o

TITLE

DRAWING NO.

Lower Case Assembly (2)


MODEL NO.

ASSEMBLY NO.

FX-2111 (CE)

422-3447
DATE

96.01.24

10-35

2
4-Double Semus Screw M3x&
,

F
O

si fi'

/I

ii

i"

41
V

Key
Na

Parts Na

9F3354

1M0381

Q ty

Lower Case Assembly (2)

Recorder Assembly

684-2093

Thermal Array Head Cable

423-3444

PCB-6409 Assembly

422-3271

Description

Drawing Na

'

Remarks

TITLE

5H9138

154-4577

Lead Label (C)

5H5639

154-3574

Lithium Battery Label

DRAWING NO.

Lower Case Assembly (3)


MODEL NO.

ASSEMBLY NO.

FX-2111 (CE)

422-3448
DATE

96.01.24

10-37

35

it
r

5S>
"J

ir.>
r

re

Double Semus Screw, M3x8

si

'

4-Doubte

t'

Semus Screw M3x8


,

Key
Na

Parts No.

Drawing Na

Description

Q'ty

Remarks
TITLE

Lower Case Assembly (3)

Upper Case Assembly

Paper Magazine Assembly

Paper Shaft SU

9F2648

9H3522

224-0891

DRAWING NO.

Exploded View
MODEL NO.

ASSEMBLY NO.

FX-2111 (CE)

412-0928
DATE

96.01.24

10-39

FUKUDA DENSHI CO., LTD.


39-4, Hongo 3-chonne, Bunkyo-ku, Tokyo 113, Japan
Phone:(03)3815-2121

Fax: (03)5684-3791

You might also like