1. Scope
1.1 This practice2 covers the design, material grouping
classification, and manufacture of hole-type image quality
indicators (IQI) used to indicate the quality of radiologic
images.
1.2 This practice is applicable to X-ray and gamma-ray
radiology.
1.3 The values stated in inch-pound units are to be regarded
as standard.
1.4 This standard does not purport to address all of the
safety concerns, if any, associated with its use. It is the
responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
2. Referenced Documents
2.1 ASTM Standards:
B 139 Specification for Phosphor Bronze Rod, Bar, and
Shapes3
B 150 Specification for Aluminum Bronze Rod, Bar, and
Shapes3
B 161 Specification for Nickel Seamless Pipe and Tube4
B 164 Specification for Nickel-Copper Alloy Rod, Bar, and
Wire4
B 166 Specification for Nickel-Chromium-Iron Alloys
(UNS N06600, N06601, and N06690) and NickelChromium-Cobalt-Molybdenum Alloy (UNS N06617)
Rod, Bar, and Wire4
E 1316 Terminology for Nondestructive Examinations5
3. Terminology
3.1 DefinitionsThe definitions of terms relating to gamma
and x-radiology in Terminology E 1316, Section D, shall apply
to the terms used in this practice.
5. IQI Procurement
5.1 When selecting IQIs for procurement, the following
factors should be considered:
5.1.1 Determine the alloy group(s) of the material to be
examined.
5.1.2 Determine the thickness or thickness range of the
material(s) to be examined.
5.1.3 Select the applicable IQIs that represent the required
IQI thickness and alloy(s).
1
This practice is under the jurisdiction of ASTM Committee E-7 on Nondestructive Testing and is the direct responsibility of Subcommittee E07.01 on
Radiographic Practice and Penetrameters.
Current edition approved July 10, 1998. Published December 1998. Originally
published as E 1025 84. Last previous edition E 1025 95.
2
For ASME Boiler and Pressure Vessel Code applications see related Practice
SE-1025 in Section II of that Code.
3
Annual Book of ASTM Standards, Vol 02.01.
4
Annual Book of ASTM Standards, Vol 02.04.
5
Annual Book of ASTM Standards, Vol 03.03.
E 1025
NOTE
NOTE
NOTE
NOTE
NOTE
NOTE
1.500
60.015
1.500
60.015
...
2.250
60.030
1.330T
60.005
0.750
60.015
0.750
60.015
...
1.375
60.030
0.830T
60.005
0.438
60.015
0.438
60.015
...
0.750
60.030
...
0.250
60.015
0.250
60.015
...
0.375
60.030
...
0.500
60.015
0.500
60.015
...
1.000
60.030
...
0.250
60.030
0.250
60.030
...
0.375
60.030
...
Tolerances
(Note 2)
610%
60.0005
60.0025
60.005
60.010
NOTE 1This practice does not recommend or suggest specific IQI sets
to be procured. Section 5 is an aid in selecting IQIs based on specific
needs.
E 1025
TABLE 1 Typical Image Quality Levels
Group 03 IQI
for Magnesium
Group 02 IQI
for Aluminum
Group 01 IQI
for Titanium
A
Equivalent IQI sensitivity is that thickness of the IQI, expressed as a percentage of the part thickness, in which the 2T hole would be visible under the same
conditions.
B
For Level 2-2T RadiologicThe 2T hole in an IQI, 150(2 %) of the specimen
thickness, is visible.
Group 1 IQI
for Carbon Alloy and
Stainless Steel
Group 2 IQI
for Aluminum Bronze
Group 3 IQI
for Nickel-Chromium-Iron
Group 4 IQI
for Nickel-Copper (70-30)
Group 5 IQI
for Tin Bronze
Detail of Notch
NOTCH TOLERANCES
Width +15
0
Depth +116 in. (1.588mm)
132 in. (.794mm)
E 1025
bronze. Group 5 IQIs may be used on bronze of higher lead
content since leaded bronze increases in attenuation with
increase in lead content. This would be equivalent to using a
lower group IQI.
9
All Haynes alloys are registered trademarks of Union Carbide Corp., New
York, NY.
10
All Hastelloys and Haynes Stellite alloys are registered trademarks of Cabot
Corp., Boston, MA.
E 1025
APPENDIX
(Nonmandatory Information)
X1. EQUIVALENT IQI (PENETRAMETER) SENSITIVITY (EPS)11
5
5
5
5
Example:
Given:
X 5 0.5 in.,
(X1.1)
TH
2,
Solution:
100T 100 3 0.005
A5 X 5
5 1.0 %
0.5
B5
(X1.2)
(X1.3)
E 1025
Definitions:
A equals the visible IQI (penetrameter) plaque thickness (T) expressed as a percentage of the section (object) thickness to be radiographed in (inches).
B equals the diameter of the smallest IQI (penetrameter) hole (H) for which the image is visibly expressed as a percentage of the section (object) thickness to be
radiographed in (inches).
NOTE 1The nomograph is used for computing equivalent IQI sensitivity from T (T equals penetrameter thickness) inches and H (H equals hole
diameter) inches. Draw a straight line joining the values on any two scales, and look on the third scale where the line crosses and read the answer. Due
to normal reproduction methods in producing the nomograph, some small error (that is, less than 5 %) may occur. If more accurate results are required,
the formula in Appendix X1 should be used.
FIG. X1.1 Equivalent I.Q.I (Penetrameter) Sensitivity Nomograph
The American Society for Testing and Materials takes no position respecting the validity of any patent rights asserted in connection
with any item mentioned in this standard. Users of this standard are expressly advised that determination of the validity of any such
patent rights, and the risk of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM Headquarters. Your comments will receive careful consideration at a meeting of the responsible
technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should make your
views known to the ASTM Committee on Standards, 100 Barr Harbor Drive, West Conshohocken, PA 19428.