SU MPC SERIES
SERVICE MANUAL
INDEX
PREFACE .......................................................................................................................................................... 2
EQUIPMENT NEEDED ...................................................................................................................................... 2
INTRODUCTION................................................................................................................................................ 2
PRELIMINARY ................................................................................................................................................... 3
SETTING PROCEDURE.................................................................................................................................... 3
+5V AND +15V LOW POWER SUPPLY CHECK.............................................................................................. 3
D/A CONVERTER VOLTAGE REFERENCE CHECK....................................................................................... 3
SETTING OF NEUTRAL PLATE ALARM CIRCUIT .......................................................................................... 3
HF OUTPUT POWER FREQUENCY CHECK................................................................................................... 4
SETTING OF RAMP DRIVING SIGNAL ............................................................................................................ 4
HF OUTPUT MOSFETS DRIVING SIGNAL CHECK ........................................................................................ 4
SETTING OF SIGNAL TO ENABLE UNIT FROM HANDSWITCH WITH PUSHBUTTONS............................ 4
HF SECTION POWER SUPPLY CHECK.......................................................................................................... 5
SETTING OF MINIMUM AND MAXIMUM HF OUTPUT POWER LEVELS ...................................................... 5
SETTING OF I_LEAK SIGNAL .......................................................................................................................... 5
SETTING OF CURRENT LIMIT......................................................................................................................... 5
HF MAXIMUM OUTPUT POWER LEVELS AND WAVEFORMS CHECK IN ALL MODES ............................. 6
RICERCA GUASTI TROUBLESHOOTING ....................................................................................................... 8
SCHEMATIC BLOCK, ELECTRIC DIAGRAMS AND COMPONENTS LAYOUT ............................................. a
BILLS OF MATERIALS ...................................................................................................................................... b
801336 RF POWER SECTION & HANDLE CONTROL (SU 140/D MPC SU 140 MPC)............................ b
801337 uC SECTION & DRIVER SECTION (SU 140/D MPC SU 140 MPC) ............................................. c
801345 RF POWER SECTION & HANDLE CONTROL (SU 100 MPC SU 50 MPC) ................................. d
801346 uC SECTION & DRIVER SECTION (SU 100 MPC SU 50 MPC)................................................... e
801367 RF POWER SECTION & HANDLE CONTROL (SU 140/BD MPC) ................................................... f
801368 uC SECTION & DRIVER SECTION (SU 140/BD MPC).................................................................... g
LAYOUT OF INTERNAL ITEMS: BOARDS, OUTPUT SOCKETS, POWER ENTRY, MAINS SWITCH.......... h
PREFACE
Purpose of this manual is to supply instructions to test features and adjust the unit for a properly trained
technical staff, allowing to solve most common failures.
EQUIPMENT NEEDED
DESCRIPTION
Dual trace oscilloscope
True RMS digital
multimeter
Current probe
H.F. power and leakage
current tester for
electrosurgery
Supplementary display
FEATURES
BandWidth: > 40 MHz
EXAMPLE
Tektronix mod. TDS210
Precision Class: 2%
Precision Class: 5%
Fluke 454A
(*)
(*) Supplementary display is needed to test SU 100 MPC and SU 50 MPC, since they cant show setting
output power.
INTRODUCTION
ALSATOM SU MPC series electrosurgical units are made up of the following sections:
Microcontroller section;
HF output power section;
Power display section (only for SU 140, 140/D, SU 140/BD);
Handle control section (only for SU 140 e 140/D).
Microcontroller section performs main system functions: according to signals coming from HF output power
section, setting commands and monitoring circuits, it enables HF output power delivery, blocking it as soon
as control parameters are out of normal safety conditions. Driving signals for HF output power section are
delivered in this section as well.
HF output power section is composed by a mosfet push-pull configuration: throughout the HF output
transformer, it supplies HF power available in output connections. Some relays drive the waveform either to
monopolar or bipolar output, blocking HF power in not used output receptacle.
Since HF output power is higher than 50 Watts in all models, neutral plate alarm is supplied: this blocks HF
power delivery as soon as a neutral electrode failure occurs. Failed condition is displayed with a red light and
error code PPP (in 140/D and 140 only) on front panel together with an acoustic signal
SU 140/D and SU 140 models are provided with two 3-digit seven-segment displays to allow user to set HF
power in both cut and coagulation modes. Furthermore user can uses pencils with pushbuttons to activate
these models, due to a circuit to allow activation for these accessories.
Switching the unit ON, SU MPC performs a selftest phase to check if all parameters are in safety range.
During the selftest, displays show 8.8.8. 8.8.8., cut and coagulation activation lights is on. Next, CUT display
shows L with software release for few seconds. At the same time COAGULATION display shows a digit
from 1 to 5: this means what model is (refers to the following table).
Model
SU 140/D
SU 140
SU 100
SU 50
SU 140/BD
Schematics to perform maintenance on ALSATOM SU MPC are the same for the following models: SU
140/D MPC, SU 140 MPC, SU 100 MPC, SU 50 MPC. In the last two models some components are not
used. Refers to bills of materials of every model on the last page of this manual.
SU 140/BD MPC only has different schematics and bills of materials.
PRELIMINARY
PAY ATTENTION!! BEFORE TO CONNECT THE UNIT TO MAINS, CHECK MAINS VALUE ON REAR
SIDE ELECTRICAL DATA LABEL.
Remove F3 fuse connected to power supply for HF output placed on main board (alsa code 801336).
SETTING PROCEDURE
J11 connected to GND is needed to switch ON the unit in setting mode. To do this, a 100 ohms/ watts
resistors must be used. This avoids alarms due to setting not yet carried out or in case of troubleshooting.
After the unit is switched on, connection from J11 and GND can be removed, until the next switching on.
With the unit in setting mode, display shows different parameters according to what mode is selected to let to
monitoring the whole unit. Refers to the following table for shown parameters according to the mode selector.
Mode selector
Cut display
Coagulation display
Pure Cut
Blend Cut
Macro Coag
Micro Coag
Bipolar Coag
Check the neutral plate alarm circuit using various resistors: with values higher than 100-200 ohms alarm
occurs. With values lower than 100 ohms the circuit doesnt give alarm and the unit can be used.
HF OUTPUT POWER FREQUENCY CHECK
Connect the oscilloscopi probe to TPC test point on microcontroller board and check a frequency of 938
kHz ( 5 kHz); waveform must be like the following:
To check correct pencil activation mode using a standard pencil. If no activation occurs, turn R75 pot slightly
counter clockwise. Moreover, pushing both pencil buttons at the same time unit must give alarm.
Switch the unit OFF and place F3 fuse in main board.
HF SECTION POWER SUPPLY CHECK
Switch ON the unit in setting mode.
Connect the multimeter to R18 placed on main board; check a +81V ( 5 V) DC voltage. This voltage cant
be set.
SETTING OF MINIMUM AND MAXIMUM HF OUTPUT POWER LEVELS
Set 80 in CUT mode, connect a 500 ohms/200 Watts non-inductive load to monopolar output (in SU 140/BD
MPC load must be 400 ohms).
Start the unit: HF output current flowing in the load must be 400 mAmps (i.e. 80 Watts). In SU 140/BD MPC
current must be 440 mAmps.
If HF output current is different from the above mentioned value, turn R31 pot on microcontroller board.
Set 10 in CUT mode and start the unit.
Turn R39 pot on microcontroller board to obtain a HF output current of 145 mAmps (i.e. 10 Watts).
Since maximum and minimum settings are interactives, above mentioned steps must be performed two or
three times in order to obtain best setting.
SETTING OF I_LEAK SIGNAL
Start the unit in CUT mode setting up values in the following table and check corresponding values on COAG
section display.
CUTsetting
5
10
50
80
100
All models
COAG Display
5 7
SU 140/BD only
COAG Display
3839
5863
7982
6 7
2930
5657
7377
SU140/D SU 140
SU 100
HF OUTPUT
POWER (W)
Moreover, check HF output waveforms of every mode fit with the following pictures.
SU 50 MPC
hor: 1 us
vert: 500 mA
hor: 20 us
vert: 500 mA
hor: 20 us
vert: 500 mA
hor: 20 us
vert: 500 mA
hor: 20 us
vert: 500 mA
hor: 20 us
vert: 500 mA
hor: 20 us
vert: 500 mA
hor: 1 us
vert: 500 mA
hor: 1 us
vert: 500 mA
8
9
10
11
12
13
-
DETECTED ERROR
FAILURE PLACED IN
Microcontroller board
Microcontroller board
Microcontroller board
Microcontroller board
Main board
Microcontroller board / Main board
Main board
Microcontroller board
Main board
Main board
Microcontroller board
Main board
-
NOTE: last three errors have no alarm code because are recoverable failures.
Neutral plate
a red light on panel advises about it; in models with displays, unit
shows PPP on CUT display.
Cross activation
if a conflict exists from user activation and selector mode position.
The unit advises with blinking yellow and blue lights for CUT and
COAGULATION and a intermittent acoustic signal as well. In
models with displays unit shows EEE on CUT display. To solve this
problem, simply stop the activation.
Simultaneous activation
as above, when two pedals or two pencil pushbuttons or a
pushbuttons and a pedal are activated at the same time.
1
2
3
4
POSITION
FUNCTION - ACTIVATION
R.F. CONTROL
COAG / BIPOLAR
R.F. CONTROL
CUT / BLEND
DISPLAY
FUNCTION SELECTOR
MAINS
MAINS
SWITCH
VREF
SO1
ILIMIT
COAG / BIP
I_LEAK
V_PWR
8-bit
DAC
FOOT SWITCHES
CUT / BLND
CGP1_AN
CGP2_AN
CTP1_AN
CTP2_AN
PWM
COMP
PWM REGULATOR
INPUT RECTIFIER
AND FILTER
VOLTAGE
REGULATOR
VOLTAGE
REGULATOR
DRIVER
LINE
TRANSFORMER
COGP_COMD
INT7
INT6
INT5
INT4
CUTP_COMD
AN2
VREF
AN_G12
AN7
SO1
CKWDT
\RES
Z OUT
SELECT
SETUP
TTL-RX
TTL-TX
RES
VTEST
FVPP
CUTH_COMD
COGH_COMD
LED_CUT
LED_COG
AUDIO
N_PLT
I_LEAK
VALPWR
EN_MON
EN_BIP
RF POWER
TRANSFORMER
uC
MOD_MEAN
I_OUT FEED
TRANSFORMER
WDT
EN_RF
KZ938
MODLT
AN5
+5V
RF POWER
STAGE
+5V
+15V
Z_SLCT
AUDIO
FILTER
EXTERNAL
PROGRAMMING
FUNCTION
SWITCH
COAG_ON
CUT_ON
AUDIO
AMPLIFIER
R.F. I OUT
TRANSFORMER
HANDLE
CIRCUIT
HANDSWITCHES
Drawn by:
Approval:
apparecchi medicali
bologna
MONOPOLAR
23/07/2001
BIPOLAR
NEUTRAL PLATE
alsa code:------
N/P_ON
AUDIO
N/P
CIRCUIT
AUTOTEST
LOAD
s.r.l.
1
2
3
4
5
POSITION
FUNCTION - ACTIVATION
R.F. CONTROL
COAG / BIPOLAR
R.F. CONTROL
CUT / BLEND
DISPLAY
FUNCTION SELECTOR
MAINS
MAINS
SWITCH
VREF
SO1
CGP1_AN
CGP2_AN
CTP1_AN
CTP2_AN
PWM
COMP
PWM REGULATOR
INPUT RECTIFIER
AND FILTER
VOLTAGE
REGULATOR
VOLTAGE
REGULATOR
DRIVER
LINE
TRANSFORMER
COAG / BIP
FOOT SWITCH
8-bit
DAC
ILIMIT
I_LEAK
V_PWR
COGP_COMD
INT7
INT6
INT5
INT4
CUTP_COMD
AN2
VREF
AN_G12
AN7
SO1
CKWDT
\RES
Z OUT
SELECT
SETUP
TTL-RX
TTL-TX
RES
VTEST
FVPP
CUTH_COMD
COGH_COMD
LED_CUT
LED_COG
AUDIO
N_PLT
I_LEAK
VALPWR
EN_MON
EN_BIP
RF POWER
TRANSFORMER
uC
MOD_MEAN
I_OUT FEED
TRANSFORMER
WDT
EN_RF
KZ938
MODLT
AN5
+5V
RF POWER
STAGE
+5V
+15V
Z_SLCT
AUDIO
FILTER
EXTERNAL
PROGRAMMING
FUNCTION
SWITCH
COAG_ON CUT_ON
AUDIO
AMPLIFIER
R.F. I OUT
TRANSFORMER
HANDLE
CIRCUIT
HANDSWITCH
Drawn by:
Rev.:P4
apparecchi medicali
bologna
MONOPOLAR
23/07/2001
Approval:
BIPOLAR
NEUTRAL PLATE
N/P_ON
AUDIO
N/P
CIRCUIT
AUTOTEST
LOAD
s.r.l.
1
2
3
4
5
POSITION
CUT (footsw.)
BLEND (footsw.)
COAG (footsw.)
COAG MICRO (footsw.)
BIPOLAR (footsw.)
FUNCTION - ACTIVATION
R.F. CONTROL
COAG / BIPOLAR
R.F. CONTROL
CUT / BLEND
FUNCTION SELECTOR
TO EXTERNAL DISPLAY
MAINS
MAINS
SWITCH
VREF
SO1
CGP1_AN
CGP2_AN
CTP1_AN
CTP2_AN
PWM
COMP
PWM REGULATOR
INPUT RECTIFIER
AND FILTER
VOLTAGE
REGULATOR
VOLTAGE
REGULATOR
DRIVER
LINE
TRANSFORMER
COAG / BIP
FOOT SWITCH
8-bit
DAC
ILIMIT
I_LEAK
V_PWR
SO1
COGP_COMD
INT7
INT6
INT5
INT4
CUTP_COMD
AN2
VREF
AN_G12
AN7
\RES
Z OUT
SELECT
SETUP
TTL-RX
TTL-TX
RES
VTEST
FVPP
LED_CUT
LED_COG
AUDIO
N_PLT
I_LEAK
VALPWR
EN_MON
EN_BIP
RF POWER
TRANSFORMER
uC
MOD_MEAN
CKWDT
WDT
EN_RF
KZ938
MODLT
AN5
+5V
I_OUT FEED
TRANSFORMER
RF POWER
STAGE
+5V
+15V
Z_SLCT
AUDIO
FILTER
EXTERNAL
PROGRAMMING
FUNCTION
SWITCH
COAG_ON CUT_ON
AUDIO
AMPLIFIER
R.F. I OUT
TRANSFORMER
Drawn by:
alsa code:------
Rev.:P4
Approval:
23/07/2001
apparecchi medicali
bologna
MONOPOLAR
NEUTRAL PLATE
BIPOLAR
N/P_ON
AUDIO
N/P
CIRCUIT
AUTOTEST
LOAD
s.r.l.
1
2
POSITION
FUNCTION - ACTIVATION
R.F. CONTROL
COAG / BIPOLAR
R.F. CONTROL
CUT / BLEND
DISPLAY
FUNCTION SELECTOR
MAINS
VREF
SO1
ILIMIT
COAG / BIP
I_LEAK
V_PWR
8-bit
DAC
FOOT SWITCHES
CUT / BLND
CGP1_AN
CGP2_AN
CTP1_AN
CTP2_AN
COMP
PWM
MAINS
SWITCH
PWM REGULATOR
INPUT RECTIFIER
AND FILTER
VOLTAGE
REGULATOR
VOLTAGE
REGULATOR
DRIVER
LINE
TRANSFORMER
COGP_COMD
INT7
INT6
INT5
INT4
CUTP_COMD
AN2
VREF
AN_G12
AN7
SO1
CKWDT
\RES
Z OUT
SELECT
SETUP
TTL-RX
TTL-TX
RES
VTEST
FVPP
LED_CUT
LED_COG
AUDIO
I_LEAK
VALPWR
EN_MON
EN_BIP
RF POWER
TRANSFORMER
uC
MOD_MEAN
I_OUT FEED
TRANSFORMER
WDT
EN_RF
KZ938
MODLT
AN5
+5V
RF POWER
STAGE
+5V
+15V
Z_SLCT
AUDIO
FILTER
EXTERNAL
PROGRAMMING
FUNCTION
SWITCH
COAG_ON CUT_ON
AUDIO
AMPLIFIER
R.F. I OUT
TRANSFORMER
AUDIO
BIPOLAR
Drawn by:
Rev.:P4
Approval:
15/01/2002
AUTOTEST
LOAD
apparecchi medicali
bologna
s.r.l.
R2A
R7B
+5V
R8
R6
R3
R1
J2
J3
J4
low voltage
VREF
VREF
F2
F1
L1
+15V
R10
R7A
R9
R5
R2B
R4
C4
C1
C5
C2
J8
1
2
3
4
VI
VI
U1
U2
D8
OUT_B
D5
OUT_A
VO
VREF
CTP1_AN
CGP1_AN
CTP2_AN
CGP2_AN
I_LIMIT
NEU_P
OSC_NP
AUDIO
EN_BIP
EN_MON
+5VF
1
3
5
7
9
11
13
15
17
19
21
23
25
J18
2
4
6
8
10
12
14
16
18
20
22
24
26
R50
CGP1_AN
D27
+15V
+5VF
Z_SLCT
I_PEAK
RF_VOUT
I_LEAK
VAL_PWR
COGP_COMD
CUTP_COMD
COGH_COMD
CUTH_COMD
AL1
AL2
VO
CGP2_AN
CTP2_AN
RS2
RS1
15_NS
GND
GND
J1
R22
D28
+5VF
TP1
Q2
AL4 *
R19
Q1
AL3 *
C6
D9
D6
GND
+5V
C3
+15V
TP2
J5
R23
R20
C14
R24
C13
R21
D7
4
3
5
R15
D13
D10
Z_SLCT
R11
R12
D2
D1
15_NS
COGP_COMD
CUTP_COMD
COGH_COMD
D3
C34
R54
C33
R53
C32
R52
C31
R51
R30
R29
RL2
RL1
R17
+15V
J21
J20
J19
COGH_COM
CUTH_COM
I_PEAK
C17
I_LIMIT
R28
Q4
D15
4
2
5
7
7
5
2
4
D4
VAL_PWR
Q3
D14
1
2
6
R26
R27
D12
C15
R25
T1
R16
R14
CUTH_COMD
C16
D11
J9
C7
R13
T2
11
7
12
C9
Z_SLCT
C10
R35
C11
COAG
ACTIVE
CUT
+5V
GND
+15V
CUTH_COM
COGH_COM
EN_BIP
R33
R31
+15V
EN_MON
+15V
3
4
C8
C12
+5V
GND
+15V
CUTH_COM
COGH_COM
man_j99.sch
COAG
ACTIVE
CUT
4
2
5
7
J11
R32
RS3
D17
RL4
6
J10
Q6
C18
R36
4
2
RL3
5
7
D16
Q5
R34
R18
81V
R37
F3
R55
C30
I_LEAK
C23
C22
R56
1
2
6
R57
D20
D18
C25
C24
C50
high voltage
+15V
C21
C20
AUDIO
BIP2
C19
J7
J6
R42
R58
C35
D26
R41
R40
R39
R38
T3
D21
C38
C39
R60
C40
R48
NEU_P
C49
D24
16
10
C29
C28
D25
J22
J23
T4
ALSATOM SU MPC
Drawn by:
13/3/2002
Approval:
Rev.:P4.2
J17
J16
J14
J13
J12
to loudspeaker
BIP2
BIP1
NP1 1
NP2 2
COAG
CUT
ACTIVE
U3
C37
C27
D23
+15V
code: 801336-45
R59
C36
R47
R45
C26
D22
R44
+15V
OSC_NP
Q7
R43
C51
BIP1
D19
R46
4
3
5
6
1
8
4
CTP1_AN
J15
COGH_COM
CUTH_COM
+15V
GND
+5V
COGH_COM
CUTH_COM
GND
C42
VSS
VDD
COGH_COM
CUTH_COM
R61
U4A
+15V
+5V
C41
L2
+5V
12
U4D
U4B
R72
13
10
R87
U4E
U4C
OSC_MAN
OSC_MAN
U4F
100 kHz
TP3
R77
R73
R88
VDD
D35
11
D29
R62
VDD
R63
R66
+15V
R75
VDD
R82
Q9
C43
R81
C46
R68
C47
R84
R83
C44
R69
Trigger threshold
adjust
800 mV 2.8 V
R65
+15V
R78 R79 R80
R76
R74
Q8
R67
R64
R85
D36
30/40
R70
D30
30/40
T6
T5
D39
D37
D33
D31
CUT
ACTIVE
code: 801336-45
23/07/2001
Approval:
Rev.: P4.2
Drawn by:
COAG
ACTIVE
CUT
ACTIVE
C48
COAG
ACTIVE
C45
ALSATOM MPC
R86
D40
D38
R71
D34
D32
N_PLT
DISPLAY SECTION
DOUT
CS_DSPL
SCLK
DOUT
CS_DSPL
SCLK
+5VF
GND
display driver.sch
LED_CUT
LED_COG
GND
LED_CUT
LED_COG
MOD_MEAN
I_LIMIT
I_LEAK
I_PEAK
OSC_NP
NEU_P
N_PLT
AUX_3
+15V
+5VF
GND
VREF
OUT_DAC
EN_RF
KZ938
MODLT
SCLK
Upj99_smd.sch
OUT_B
OUT_A
R.F.DRIVER SECTION
+15V
+5VF
GND
VREF
OUT_DAC
EN_RF
KZ938
MODLT
SCLK
SCLK
MOD_MEAN
I_LIMIT
I_LEAK
I_PEAK
OSC_NP
NEU_P
N_PLT
AUX_3
Drivj99_smd.sch
To
Drawn by:
code: 801337-46
21/05/2002
Approval:
Rev.: 3.0
ALSATOM SU MPC
MAIN BOARD
J6
2
1
VREF
CTP1_AN
CGP1_AN
CTP2_AN
CGP2_AN
I_LIMIT
NEU_P
OSC_NP
AUDIO
EN_BIP
EN_MON
+5VF
+15V
1
2
3
4
5
6
VDD
TTL-TX
RES
VTEST
FVPP
TTL-RX
1
3
5
7
9
11
13
15
17
19
21
23
25
J5
2
4
6
8
10
12
14
16
18
20
22
24
26
U1
C17
+5VF
+15V
R13
R12
R11
R10
C16
C15
C14
C13
U2F
RES
R2
AN_G12 13
RS
Z_SLCT
I_PEAK
RF_VOUT
I_LEAK
VAL_PWR
COGP_COMD
CUTP_COMD
COGH_COMD
CUTH_COMD
TO POWER RF SECTION
J4
GND
+5VF
+5VF
GND
+15V
+15V
C47
R66
GND
+5VuP
PROGM
1
2
CGP1_AN
WDT
AN3
AN6
AN4
15
10
2
U3A
U3D
U3C
U3B
N_PLT
NEU_P
OSC_NP
I_PEAK
I_LIMIT
I_LEAK
14
11
C2
C11
VDD
N_PLT
NEU_P
OSC_NP
I_PEAK
I_LIMIT
I_LEAK
R14
DOUT
VREF
SCLK
DIN
OUT
SCLK
SYNC
U5
C12
SCLK
KZ938
MODLT
N_PLT
R4
J8
1
2
TPA
5 CS_DA
VREF
R17
R15
C18
OSC2
OSC1
RESET
P3.0/SCK1
P3.1/SI1
P3.2/SO1
P2.0/SCK3
P2.1/RXD
P2.2/TXD
+15V
R18
R16
U7
C19
J9
R21
R19
1 VTEST
2
+5VF
C6
U6
C20
R20
P9.1
P9.2
P9.3
P9.4
FTCI/P8.0
FTOA/P8.1
FTOB/P8.2
FTIA/P8.3
FTIB/P8.4
FTIC/P8.5
FTID/P8.6
P8.7
P7.3
TMRIV/P7.4
TMCIV/P7.5
TMOV/P7.6
P7.7
P6.0
P6.1
P6.2
P6.3
P6.4
P6.5
P6.6
P6.7
AUX_3
AUX_4
AUX_5
CS_DSPL
CS_DA
J10
12
13
14
15
39
40
41
42
43
44
45
46
COGP_COMD
COGH_COMD
2
3
4
5
6
7
8
9
10
EN_RF
LED_COG
LED_CUT
R27
+5VF
AUX_3
R25
R24
CS_DSPL
+5VF
AUX_3
C27
C28
R23
C26
Drawn by:
code:801337-46
Rel.:3.0
J11
Approval:
J3A
13
SET UP
+5VF
6
5
4
3
2
1
21/05/2002
TTL-TX
TTL-RX
N_PLT
AUX_9
AUX_8
NU_PIN
R26
2
3
4
5
6
7
8
9
10
C25
uC Section (sheet 2 of 4)
GND
LED_CUT
LED_COG
EN_RF
Z_SLCT
C24
NU_PIN
C23
CUTH_COMD
ALSATOM SU MPC
C22
CKWDT
AN_G12
C21
CUTP_COMD
34
EN_MON
35
36
37 AUDIO
38 EN_BIP
17
18
19
20
21
22
23
24
25
INT0/P5.0
26
INT1/P5.1
27
INT2/P5.2
28
INT3/P5.3
29
INT4/P5.4
30
ADTRG/INT5/P5.5
31
TMIB/INT6/P5.6
32
INT7/P5.7
P1.0/TMOW
P1.4/PWM
P1.5/IRQ1
P1.6/IRQ2
P1.7/IRQ3/TRGV
IRQ0
PB0/AN0
PB1/AN1
PB2/AN2
PB3/AN3
PB4/AN4
PB5/AN5
PB6/AN6
PB7/AN7
3
AVss
58
AVcc
52
51
50
TEST
SCLK
47
TTL-RX 48
TTL-TX 49
53
54
55
56
57
16
C5
1
2 FVPP
6
X1
5
X2
KZ938
MODLT
N_PLT
AUX_8
NU_PIN
R5
10
VPP
J2
2
1
64
63
62
61
60
59
C10
XT2
C9
C4
R3
+5VuP
MOD_MEAN
AN1
AN2
AN3
AN4
AN5
AN6
AN7
DOUT
MOD_MEAN
C8
XT1
C7
RES
J9-J10: cortocircuitare
dopo la taratura di Vref
J7
+5VF
R9
R8
+5VF
VSS
VDD
AN7
AN2
+5VF
+5VF
R7
C3
R6
C1
U4
OUT_DAC 2
7
OUT
REF
CS_DA
3
6
CS
LDAC
4
5
DIN SCLK
OUT_DAC
DOUT
OUT_DAC
AN_G12
CTP2_AN
AN_G1
CTP1_AN
AN_G12
CGP2_AN
12 AN_G1
AN1
J1
1
8
9
16
CKWDT
R1
1
2
GNDVDD
8
11
FVPP
TEST
4
33
VCC
VSS
7
R22
VREF
R65
+5VF
+5VF
GND
+15V
+15V
TPD
R32
R33
C33
R31
10
R40
R37
C35
+15V
R34
R36
R35
OUT_DAC
R39
C36
R38
C34
C32
+Vr
D1
R41
U8C
C30
C37
R42
E/A OUT
INV. INP
N.I. INP
RAMP
RT
VREF
CLK
CT
+15V
OUT_DAC
16
U9
U2B
15
+VIN
SS
8
R29
C29
R28
13
C38
R43
10
14
12
11
R44
D15
GND
OUT B
C.L.
PWR GND
Q1
C31
R30
OUT A
VC
D4
R64
TPE
R51
D3
D2
R50
R48
U8B
C39
5 2
4
11
R47
R46
I_PEAK
I_LIMIT
OUT_B
J12
I_PEAK
I_LIMIT
AUX_3
OUT_B
1
2
3
4
11
NEU_P
C42
R52
U2E
TPC
13 KZ938
MOD_MEAN
U8D
OUT_A
MOD_MEAN
OUT_A
R45
C40
R49
U2A
MODLT
12 EN_RF
10
NEU_P
R53
71 KHz
SCLK
KZ938
EN_RF
MODLT
N_PLT
8 OSC_NP
N_PLT
OSC_NP
Drawn by:
Rev.:3.0
VSS
VSS
VDD
Approval:
21/05/2002
U2C
U2D
SCLK
TPB
ALSATOM SU MPC
U8A
code: 801337-46
C43
R54
C41 +5VF
14
7
C44
GND
+5VF
J13
1
2
3
4
5
+5VF
SCLK
CS_DSPL
DOUT
GND
+5VF
+5VF
R55
13
12
18
GND
CLK
LOAD(CS)
DIN
ISET
SEG A
SEG B
SEG C
SEG D
SEG E
SEG F
SEG G
DP
DIG 0
DIG 1
DIG 2
DIG 3
DIG 4
DIG 5
DIG 6
DIG 7
U10
+5VF
C45
C46
19
VCC
GND
14
16
20
23
21
15
17
22
2
11
6
7
3
10
5
8
SEG A
SEG B
SEG C
SEG D
SEG E
SEG F
SEG G
DP
DIG 0
DIG 1
DIG 2
DIG 3
DIG 4
DIG 5
SEG [A..G]
dp
dp
dp
COAG SECTION
1's
SEG [A..G]
com
D5
dp
R56
com
R57
CUT
D11
D6
Q2
R58
+5VF
3
8
DIG 1
com
3
8
DIG 0
LED_COG
D7
dp
com
3
8
R59
D12
Q3
R61
+5VF
D8
R60
COAG
DIG 5
com
D13
R62
+5VF
ALSATOM SU MPC
Drawn by:
Rev. 3.0
Approval:
D14 ON/OFF
R63
+5VF
21/05/2002
N_PLT
N_PLT
D10
dp
D9
code: 801337-46
com
3
8
DIG 4
3
8
DIG 3
3
8
DIG 2
LED_CUT
SEG A
SEG B
SEG C
SEG D
SEG E
SEG F
SEG G
DP
10
9
7
5
4
2
1
6
10's
SEG A
SEG B
SEG C
SEG D
SEG E
SEG F
SEG G
DP
10
9
7
5
4
2
1
6
100's
SEG A
SEG B
SEG C
SEG D
SEG E
SEG F
SEG G
DP
10
9
7
5
4
2
1
6
1's
SEG A
SEG B
SEG C
SEG D
SEG E
SEG F
SEG G
DP
10
9
7
5
4
2
1
6
10's
CUT SECTION
SEG A
SEG B
SEG C
SEG D
SEG E
SEG F
SEG G
DP
10
9
7
5
4
2
1
6
100's
SEG A
SEG B
SEG C
SEG D
SEG E
SEG F
SEG G
DP
10
9
7
5
4
2
1
6
_______________________________________________________________________________
- alsa apparecchi medicali s.r.l. b-1
_______________________________________________________________________________
- alsa apparecchi medicali s.r.l. b-2
_______________________________________________________________________________
- alsa apparecchi medicali s.r.l. b-3
_______________________________________________________________________________
- alsa apparecchi medicali s.r.l. b-4
alsa code
_______________________________________________________________________________
R1
R2
R3
R4
R5
R6
R7
R8
R9
R10
R11
R12
R13
R14
R15
R16
R17
R18
R19
R20
R21
R22
R23
R24
R25
R26
R27
R28
R29
R30
R31
R32
R33
R34
R35
R36
R37
R38
R39
R40
R41
R42
R43
R44
R45
R46
R47
R48
R49
R50
R51
R52
R53
R54
R55
R56
R57
R58
R59
10K_SMD
10K_SMD
15K_SMD
4K7_SMD
1M_SMD
100_SMD
100_SMD
3K3/1%_SMD
3K3/1%_SMD
100_SMD
100_SMD
100_SMD
100_SMD
33K/1%_SMD
330_SMD
330_SMD
330_SMD
330_SMD
2K7/1%_SMD
1KT/6mm
3K3/1%_SMD
100_SMD
X_RNET10K
10K_SMD
100_SMD
10K_SMD
X_RNET10K
680_SMD
22_SMD
22_SMD
5KT/6mm
8K2_SMD
47_SMD
1K8_SMD
12K_SMD
56K_SMD
1K_SMD
10K_SMD
20KT/6mm
56K/1%_SMD
22K/1%_SMD
1K_SMD
10M
470_SMD
10KT/6mm
47K_SMD
27K_SMD
2K2_SMD
2K2_SMD
10_SMD
10_SMD
15K_SMD
20KT/6mm
100_SMD
10K/1%_SMD
4K7_SMD
10K_SMD
560_SMD
4K7_SMD
430513/S
430513/S
430516/S
430511/S
430524/S
430496/S
430496/S
430509/S
430509/S
430496/S
430496/S
430496/S
430496/S
430519/S
430498/S
430498/S
430498/S
430498/S
430507/S
403111/S
430509/S
430496/S
430484
430513/S
430496/S
430513/S
430484
430501/S
430494/S
430494/S
403113/S
430526/S
430495/S
430505/S
430515/S
430522/S
430502/S
430513/S
403115/S
430523/S
430525/S
430502/S
430431
430499/S
403114/S
430520/S
430518/S
430506/S
430506/S
430493/S
430493/S
430516/S
403115/S
430496/S
430514/S
430511/S
430513/S
430500/S
430511/S
______________________________________________________________________________________
alsa apparecchi medicali s.r.l. c-1
alsa code
_______________________________________________________________________________
R60
R61
R63
R64
R65
C1
C2
C3
C4
C5
C6
C7
C8
C9
C10
C11
C12
C13
C14
C15
C16
C17
C18
C19
C20
C21
C22
C23
C24
C25
C26
C27
C28
C29
C30
C31
C32
C33
C34
C35
C36
C37
C38
C39
C40
C41
C42
C43
C44
C45
C46
D1
D2
D3
D4
D5
D6
D7
D8
10K_SMD
560_SMD
560_SMD
100_SMD
1K5/1%_SMD
100N_SMD
100N_SMD
100N_SMD
1U/T/35V
100N_SMD
10U/T/35V
22P_SMD
22P_SMD
12P_SMD
12P_SMD
100N_SMD
1U/T/35V
100N_SMD
100N_SMD
100N_SMD
100N_SMD
220U/E/16V
10U/T/35V
100N_SMD
100N_SMD
100N_SMD
100N_SMD
100N_SMD
100N_SMD
100N_SMD
100N_SMD
100N_SMD
100N_SMD
470P_SMD
10U/T/35V
100N_SMD
100N_SMD
220P/PP/100V
100P/NPO
100N_SMD
100N_SMD
10N_SMD
10N_SMD
100N_SMD
1U/T/35V
100N_SMD
1N/P/100V
100N_SMD
100N_SMD
100N_SMD
10U/T/35V
1N4148_SMD
11DQ06
11DQ06
BAT83
TDSL3160
TDSL3160
TDSL3160
TDSL3160
430513/S
430500/S
430500/S
430496/S
430504/S
400320/S
400320/S
400320/S
400173
400320/S
400134
400315/S
400315/S
400314/S
400314/S
400320/S
400173
400320/S
400320/S
400320/S
400320/S
400274
400134
400320/S
400320/S
400320/S
400320/S
400320/S
400320/S
400320/S
400320/S
400320/S
400320/S
400317/S
400134
400320/S
400320/S
400310
400260
400320/S
400320/S
400319/S
400319/S
400320/S
400173
400320/S
400312
400320/S
400320/S
400320/S
400134
420098/S
420054
420054
420070
420096
420096
420096
420096
______________________________________________________________________________________
alsa apparecchi medicali s.r.l. c-2
alsa code
_______________________________________________________________________________
D9
D10
D11
D12
D13
D14
D15
Q1
Q2
Q3
U1
U2
U3
U4
U5
U6
U7
U8
U9
U10
XT1
XT2
J1
J2
J3
J4
J5
J6
J7
J8
J9
J10
J11
J12
J13
TDSL3160
TDSL3160
GL5HY8
GL5HB8
GL5HR8
GL5HG8
1N4148_SMD
VN10KM
BC817
BC817
TC32M
74HCT14_SMD
ADG712
NU
AD5300
H8/3644(FP64)
LM431
74HCT00_SMD
UC3825_SMD
MAX7219
32KHZ
15MHZ/HC18
CONN_2PM
CONN_2PM
407032
PICO_6
PICO_26
CONN_2PM
PAD_CI
CONN_2PM
CONN_2PM
399028
CONN_1PM
PICO_4
CONN_5PF
______________________________________________________________________________________
alsa apparecchi medicali s.r.l. c-3
_______________________________________________________________________________________
alsa apparecchi medicali s.r.l. d-4
______________________________________________________________________________________
alsa apparecchi medicali s.r.l. e-3
ALSATOM SU 140/BD
ALSATOM SU 100
7
8
9b
1)
2)
3)
4)
5)
6)
7)
DETAILS OF COMPONENTS
9a) Active socket
7) Bipolar socket
701364: M2 nut
496019: M10x0.75 nut
325021: Nut
10) Knob
494100: Cap
494096: Body
494097: Indicator