Professional Documents
Culture Documents
V Meter Manual PDF
V Meter Manual PDF
TABLE OF CONTENTS
SECTION 1
PRINCIPALS OF TESTING BY ULTRASONIC PULSE VELOCITY MEASUREMENT
1.1
1.2
1.3
1.4
1.5
1.6
Introduction
Velocity of Longitudinal pulses in Elastic Solids
Effect of Size and Shape of Specimen Tested
Frequency of Pulse Vibrations
Method of Testing
Application of Pulse Velocity Testing
1
1
2
2
2
4
SECTION II
SPECIFICATION OF V-METER MK II
2.1
2.2
2.3
2.4
2.5
2.6
The V-Meter MK II
The V-Meter Field Kit C-7901
Description
Construction of the V-Meter MK II
Transducers and Leads
Summary of Specifications
6
6
6
8
8
9
SECTION III
OPERATING INSTRUCTIONS
3.1
3.2
3.3
3.4
3.5
3.6
3.7
3.8
3.9
3.10
3.11
Introduction
General Description
Initialization Items Explained
Menu Items - Main Menu
Menu Items - Run Menu
Familiarization
Summary
Addenda
A.C Operation
Battery Operation
Full Charging the Internal Battery
11
11
15
17
20
20
24
24
24
25
25
SECTION IV
TAKING PULSE VELOCITY MEASUREMENTS
4.1
4.2
4.3
4.4
4.5
Transducers
Offset (Zero Set)
Couplant
Placing the Transducers
Pulse Velocity
26
26
26
26
27
SECTION V
SYSTEM DESCRIPTION
5.1
5.2
5.3
5.4
Pulse Generator
CPU
Receiver Amplifier
Master Clock ADC and Display
29
29
29
29
SECTION VI
TESTING CONCRETE
6.1
6.2
6.3
6.4
6.5
6.6
6.7
6.8
6.9
6.10
6.11
6.12
6.13
6.14
Applications
Accuracy
Coupling the Transducers with Concrete Surface
Choice of Transducer Arrangement
Influence of Test Conditions
Homogeneity of the Concrete
Detection of Defects
Detection of Large Voids or Cavities
Estimating the Depth of Surface Cracks
Monitoring Changes in Concrete with Time
Estimation of Strength After Fire Damage
Estimation of Strength
Estimation of Elastic Modulus
Determination of the Dynamic Modulus of Elasticity
And Dynamic Poisson's Ratio
30
30
31
31
35
39
40
40
40
44
44
48
51
53
SECTION VII
ADDITIONAL ACCESSORIES FOR USE WITH THE V-METER MK II
7.1
7.2
7.3
7.4
7.5
7.6
56
56
56
56
58
58
References
59
SECTION I
PRINCIPALS OF TESTING BY ULTRASONIC PULSE VELOCITY MEASUREMENT
1.1
INTRODUCTION
The velocity of ultrasonic pulses traveling in a solid material depends on the density and elastic
properties of that material. The quality of some materials is sometimes related to their elastic
stiffness so that measurement of ultrasonic pulse velocity in such materials can often be used to
indicate their quality as well as to determine their elastic properties.
Materials which can be assessed in this way include, in particular, concrete and timber but
exclude metals.
When ultrasonic testing is applied to metals its object is to detect internal flaws which send
echoes back in the direction of the incident beam and these are picked up by a receiving
transducer. The measurement of time taken for the pulse to travel from a surface to a flaw and
back again enables the position of the flaw to be located.
Such a technique cannot be applied to heterogeneous materials like concrete or timber since
echoes are generated at the numerous boundaries of the different phases within these materials
resulting in a general scattering of pulse energy in al directions.
1.2
It can be shown that the velocity of a pulse of longitudinal ultrasonic vibrations traveling
in an elastic solid is given by:
1.3
The above equation may be considered to apply to the transmission of longitudinal pulses
through a solid of any shape or size provided the least lateral dimension (i.e the dimension
measured perpendicular to the path traveled by the pulse) is not less than the wavelength of the
pulse vibrations.
The pulse velocity is not affected by the frequency of the pulse so that the wavelength of the
pulse vibrations is inversely proportional to its frequency.
Thus the pulse velocity will generally depend only on the properties of the materials and the
measurement of this velocity enables an assessment to be made of the condition of the materials.
1.4
The pulse frequency used for testing concrete or timber is much lower than that used in metal
testing. The higher the frequency, the narrower the beam of pulse propagation but the greater the
attenuation (or damping out) of the pulse vibrations.
Metal testing requires high frequency pulses to provide a narrow beam of energy but such
frequencies are unsuitable for use with heterogeneous and coarse grain materials because of the
considerable amount of attenuation which pulses undergo when they pass through these
materials.
The frequencies suitable for these materials range from about 20 kHz to 250 kHz, which 50 kHz
being appropriate for the field testing of concrete. These frequencies correspond to wavelength
ranging from about 8 inches ( for the lower frequency) to about .6 inches at the higher frequency.
1.5
METHOD OF TESTING
For assessing the quality of materials from ultrasonic pulse velocity measurement, it is necessary
for this measurement to be of a high order of accuracy. This is done using an apparatus which
generates suitable pulses and accurately measures the time of their transmission (i.e transit time)
through the material tested. The distance which the pulse travel in the material (i.e the path
length) must also be measured to enable the velocity to be determined from:
Pulse velocity =
Path length
____________
Transit Time
Path lengths and transit time should each be measured to an accuracy of about + 1%.
The instrument indicates the time taken for the earliest part of the pulse to reach the receiving
transducer measures from the time it leaves the transmitting transducer when these transducers
are placed at suitable points on the surface of the material.
Figure 1 shows how the transducer may be arranged on the surface of the specimen tested, the
transmission being either direct, semi direct or indirect.
The direct transmission arrangement is the most satisfactory since the longitudinal pulses leaving
the transmitter are propagated mainly in the direction normal to the transducer face. The indirect
arrangement is possible because the ultrasonic beam of energy is scattered by discontinuities
within the material tested but the strength of the pulse detected in this case is only about 1 or 2%
of the detected for the same path length when the direct transmission arrangement is used.
Pulses are not transmitted through large air voids in a material and, if such a void lies directly in
a pulse path, the instrument will indicate the tome taken by the pulse which circumvents the void
by the quickest route. It is thus possible to detect large voids when a grid of pulse velocity
measurements is made over a region in which these voids are located.
1.6
This method of testing was originally developed for use on concrete and the published accounts
of its application are concerned predominately with this material
A considerable volume of literature has been published over the past 40 years describing the
results of research on the use of ultrasonic testing for concrete and for fuller details of this
application the reader is referred to the selected list given at the end of this Section.
The method was first developed in Canada by Leslie and Cheesman between 1945 and 1949 and
also independently in Britain at about the same time by Jones and Gatfield. The apparatus
developed at that time made use of a cathode-ray oscilloscope for the measurement of transit
times and modified forms of this equipment have been widely used in many countries. The
equipment was particularly useful in the laboratory but was less easy to use under field
conditions.
The apparatus described in this Manual has been designed particularly for field testing being
light, portable and simple to use. It can be operated independently of the main power supply
when used in the field and directly from the A.C. supply for laboratory use.
There has been a steadily growing interest in no-destructive testing of concrete in several
countries and, a Symposium held in Canada in 1984 and recently in Britain enabled reviews and
accounts of current research on the subject to be discussed and recorded.
ASTM has the specification ASTMC-597 for the use of this method since 1967 and the British
Standards Institution has issued Recommendations for measurement of velocity of ultrasonic
pulses in concrete. B.S. 1881: Part 203. 1986.
Ultrasonic testing is now widely used throughout the world and it is clear that the advantages of
this method over traditional methods of testing are likely to increase further its application. In
particular its ability to examine the state of concrete in depth is unrivaled.
The pulse velocity method has been shown to provide a reliable means of estimating the strength
of timber and has been used to test various kinds of timber products. It is in use for the detection
of rot in utility poles and provides a very economic method of inspecting these poles while in
service.
The same equipment can be used to test rock strata and to provide useful data for geological
survey work. The method has also been used for testing graphite, ceramics and any coarse grain
materials and it is likely that it will prove useful for testing other non-metallic materials.
SECTION II
SPECIFICATION OF V-METER MK II
2.1
THE V-METER MK II
The V-Meter MK II has been designed with site testing particularly in ind so as to be fully
portable, simple to operate and with a high order of accuracy and stability. It generates low
frequency ultrasonic pulses and measures the time taken for them to pass from one transducer to
the other through the material interposed between them.
2.2
The V-Meter MK II
Two transducers (54 kHz)
Two transducer leads
Carrying case
V-Meter MK II manual
Can of couplant
AC / charger unit.
DESCRIPTION
The V-Meter Mk II gives a direct reading of the time of transmission of an ultrasonic pulse
passing from a transmitting to a receiving transducer, up to 6400 micro seconds, with resolutions
of 0.1 micro seconds which represents approximately 100 ft. of concrete.
If the transmitted pulse is not received, or when the transducers are removed from the test piece,
the LCD will automatically show the maximum number, approximately 6400 micro seconds.
No reference calibration bar is needed because the built in micro computer remembers the delay
due to the different transducers and cables when first initialized and then subtracts from the
following measurement automatically.
The pulse generator may be operated at a high voltage of either 1,200 V or 500V as selected by a
switch on the top panel. Generally, for concrete testing and for long path lengths the generator is
operated at 1,200V but, if fine racks are being investigated it may be advantageous to reduce the
high voltage to 500V. Pulse repetition frequencies of 1, 3 and 10 may be selected by a menu
driven initialization procedure. The LCD read out is updated once, twice or five times per second
for both PRFs depending on the pulse repetition rate.
The receiving amplifier has a high input impedance enabling the instrument to be used with
piezo-electric and ferro-electric transducers over the frequency range 5kHz to 1MHz.
For field use an internal Nickel Cadmium battery will, when fully charged, supply power for
about 8 hours continuous use. The battery voltage is monitored by a micro-power sensor which,
when the voltage is near to the discharge end point, will cause the low battery LED on the front
panel to light up.
When the instrument is used on the A.C. mains a built in constant current charger will
continuously trickle charge the battery. A fully discharged batter should be charged as soon as
possible by connecting the instrument to the A.C. mains, for a 16 hour overnight charge.
A cathode ray oscilloscope may be coupled to the CRO and Time Base Trigger BNC sockets
mounted on the panels.
Both transit time, pulse waveform, and attenuation observations and measurements may be made
by using the V-Meter Mk II in conjunction with the CRO.
MEMORY OPTION
See Section III 4.5 and 8.
2.4
The electronic circuit of V-Meter Mk II uses high speed CMOS (74HC) integrated circuits, and a
micro computer chip with E-prom.
The assembly is designed for ease of access for servicing the component boards.
2.5
The transducers consist of lead zirconate titanate (PZT4) ceramic piezo electric elements
mounted in stainless steel cases. The elements are very tightly held on to the inside face of the
case to provide highly efficient acoustic transmission. The transducer assembly is very robust
and able to withstand reasonably rough conditions as normally encountered under conditions of
industrial usage.
The shock excitation of the pulse generator causes the transducer to oscillate mechanically at its
own natural frequency, this frequency depending on the size and stiffness of the whole
transducer assembly. Different sizes of piezo electric element and case enable different pulse
frequencies to be obtained.
Each transducer is fitted with a cable socket to enable it to be detached from the cable and to
allow cables of different length to be used.
Short co-axial cables for connecting the transducers to the V-Meter Mk II are supplied with the
instrument. Longer cables may be used when access to the areas to be tested is restricted.
Although long cables reduce the size of both the transmitted pulse and the receiving signal, these
are generally both of adequate magnitude for testing even when the cables are 25 ft. long.
Note: The instrument should be re-zeroed when changing to long cables. A delay of about 0.8
microsecs. Is introduced when using 100 ft. cables on both the transmitting and receiving
transducers. To minimize attenuation when using long cables, a signal amplifier, with a low
output impedance, can be inserted between the receiving transducer and the cable.
2.6
SUMMARY OF SPECIFICATION
Accuracy
Lost signal
INPUT SENSITIVITY:
Signal
Impedance
Approximately 2MOHMS.
TRANSMITTER:
Energizing pulse
Discharge time
PRF
POWER SUPPLY:
Battery low
Battery Charger
Display
Temp. Range
0C to 40 C.
OPTION
SECTION III
OPERATING INSTRUCTIONS
3.1
INTRODUCTION:
The Pulse Velocity Tester (James Instruments V-METER Mk II) is an instrument designed to
determine the quality of concrete structures. Like previous instruments, it uses bursts of
ultrasonic waves to determine the velocity of sound through the medium. Unlike others, it
incorporates an advanced microprocessor and state of the art electronics, making possible a new
level of convenience in use and reliability of data. In addition, this system is equipped to analyze
S-wave response with relation to P-wave response, calculating Poisson's Ratio to a high level of
accuracy.
Powered by the A.C. line or by rechargeable battery, the unit is equally useful in the lab and in
the field. Measurement of time is accomplished fro a ten megahertz clock, yielding a
measurement capacity from 0 to 6.2 milliseconds with resolution of 100 nanoseconds. The
optional hand held terminal and liquid crystal display allow convenient operator-computer
interaction, and an oscilloscope like display makes the use of an external oscilloscope
unnecessary.
3.2
GENERAL DESCRIPTION
10
11
12
depressed for about one second for it to be recognized. This characteristic may be used to
advantage when it is desired to hold readings and plots for study ... holding down on any
uncommitted key (not a menu item or ESC) will stop the action, and releasing it will allow the
run to continue.
3.3
1)
In order to calibrate for a given set of transducers and cables, it is necessary to determine the
minimum time for the pulse to be detected by the receiver and to be honored by the electronics.
The operator is instructed to hold the transducers together, then depress "enter." The computer
then takes a reading and saves it as offset, to be subtracted from all future readings (Be sure to
use the couplant).
2)
Trace:
The bottom three lines of the display are reserved, in the RUN mode, for an oscilloscope like
display. The TRACE selection gives the operator three choices:
0)
1)
Trace on. In this mode, the waveform displayed is the envelope of the actual signal, not
the signal as it would be seen on an oscilloscope. The time at which the electronics
detects the signal is marked by a vertical line through the envelope. The time displayed is
auto scaled, and can be as great as 6.2 milliseconds or as small as 100 nanoseconds (.1
microsecond.)
NOTE: In this mode, the pulse rate is approximately one pulse/second, regardless of the pulse
rate choice.
2)
Expanded trace. In this mode, the waveform displayed is the actual waveform, for a
period of about 200 microseconds starting about 600 nanoseconds before the signal
detection. The actual pulse rate is limited to approximately two pulses/second. In this
mode the marker is held at the constant position of 800 nanoseconds from the beginning
of the trace and marks the first detection of the signal. The exception is the case in which
the signal is detected in a total time of less than 800 nanoseconds, in which case the
marker is at the beginning of the trace.
NOTE: In all cases of the trace mode, the marker serves only as a pointer ... the actual numbers
are being displayed on the screen.
3)
Density:
Density is used in the calculation of Young's modulus.
5)
Display Mode:
The V-METER mode shows time in large letters, with (Poisson's Ratio) and E
(Young's Modulus) where appropriate. When a Trace mode is selected, the plotting
occurs at the bottom of the screen. The DETAIL mode uses normal size letters and
therefore can display more information. Distance, Time and Velocity are displayed, along
with the Run menu. As in the V-METER mode, , E, and trace are displayed when
appropriate.
60
Rate:
Pulse rates are 1, 3, or 10 pulses per second. Two cases exist:
A)
B)
Trace on. In this case, the trace plotting time interferes with the specified rate.
Trace and display are updated for every pulse when rate=1, every second pulse
when rate=3, and every fifth pulse when rate=10.
3.4
1)
Wave type:
Allows selection of Compression (P) or Transverse (S) waves. The system has no way of
knowing which type of wave is being monitored. It is necessary, therefore, for the
operator to specify the type, in order for the functions (Poison's Ratio) and E (Young's
modulus of Elasticity) to be determined.
2)
English (Metric):
This selection enables the choice of entry display units. Internally, the data are stored in
English units.
3)
4)
Simple E: a of .1 is assumed.
Derived u: u is calculated after each reading.
Arbitrary : is set by the operator, and held as a constant.
Young's modulus can be calculated by two methods, the simpler of the two being an
approximation to the other.
Method I:
E=Vc^2 * d/144g
where # is Young's modulus of elasticity,
Vc^2 is the square of Compression wave (P-wave) velocity,
d is the density of the medium,
g is acceleration due to gravity.
Method II:
E=(d/144g)*(Vc^2 (+)(1-2))/(1-)
where d is density
g is acceleration due to gravity
Vc^2 is the square of the velocity of the P wave
is Poisson's ratio:
= (Vc^2 -2Vt^2) / 2*(Vc^2 - Vt^2)
Vc is the velocity of the compression (P) wave
Vt is the velocity of the transverse (S) wave
Method II requires that the system be equipped with transducers for both P and S type waves.
When only P types are used, the simpler form (I) is calculated and displayed as E. Note that the
wave type must be specified by menu time 1. The default type is P.
When both P and S types can be taken, the Derived u option (B) may be used, and should yield
best accuracy. A particular sequence must be followed:
A) With the P type transducer connected, P type is selected from the menu. Then with
each reading, the velocity will be saved in a temporary location.
B) The system is then stopped, (ESC key), and the S type transducer is connected. From
the menu, S type is selected. Then the system, when running S type, checks the
temporary location to determine the last P-wave velocity along with the current S-wave
velocity to calculate . is then used to calculate E.
In general, if Poisson's ration () is not displayed, the modulus is of the simple or the arbitrary
types. If u is being displayed, the modulus has been found from the more involved calculation,
16
Upload to IBM:
When the RS232 connector is in place and attached to an IBM PC, the program
UPLOAD in the PC works with this entry to send the stack data to the PC RAM, which
can be saved as a disk file by menu command. Note that the data are in English units. The
English-Metric information is sent in order that the analyst can know as much as possible
about the conditions under which the data were taken.
This system uploads the data stack in V-Meter Mk II to an IBM compatible computer,
program diskette supplied. Two modes are available: With the program UPLOAD.EXE,
the entire stack is transmitted on command. With the program ONLINEEXE, the same
method may e used or, if desired, the individual data samples may be uploaded as they
are taken. Of course this requires that the IBM computer be online.
When readings are being taken with the V-Meter Mk II system, they are saved using key
#3. This gives the options:
0)
1)
2)
With the battery backup RAM, when the system is turned on, old data remains in the
stack. If a new set of data is to be taken, the stack must be initialized by the operator
(code 3).
To save data to the stack, use code 3, then 1. The current reading is saved. To upload the
entire stack, stop the V-Meter MK II, (ESC key), and run the program UPLOAD.EXE or
ONLINE.EXE on the IBM. Selecting code 2 (upload) on the IBM will give the prompt,
which tells you to select UPLOAD on the V-Meter MK II. Then a carriage return to the
IBM starts the upload process.
17
To upload the data sample by sample, run the program ONLINE. The V-Meter MK II
must be started (or reset) AFTER the ONLINE program has been started.
Procedure: Run ONLINE.EXE
Select code 6. You will be prompted to start the pulse system. Start the V-Meter MK II
and get a satisfactory reading saved. Use code 3 to initialize the tack, if desired.
NOTE: That the un-initialized stack may contain misleading or garbage information.
Use code 3 to save the stack. The data will be saved as usual, ans also transmitted to the
IBM. With each save the data will be sent to the IBM, and stored there.
Use of the IBM programs UPLOAD.EXE and ONLINE.EXE:
These programs are identical, with the exception that UPLOAD does not have the sample
by sample uploading capability. There are two methods of saving to a disc file: When
UPLOAD is selected (either method) you will be asked if you wish to save to a disc file.
If you answer yes, you will be asked for a file name. Just invent one. If you answer no,
the data will be stored in an array, and lost when the IBM is witched off. However, the
menu gives the option of saving the array, which has the same effect as the other type of
save.
It is also possible to review the upload data on the IBM screen. The user will probably
find that the best method is to upload the data without using a disc file, then inspecting
the data by menu command. If the data is then found to be OK, the SAVE menu site can
be selected, and the data saved on disc.
FORMAT:
Data are saved in ASCII files. This is inefficient from the stand point of the disc space,
but best for easy transfer to another program.
The menu looks like this:
V-METER K II
MENU:
1.
Read a disc file
2.
Upload Data
3.
DOS Call
4.
Display data array
5.
Save this data to disc.
6.
Online upload
7.
Quit
The program UPLOAD.EXE is the same except for item 6, which is omitted.
18
6)
7)
Check Stack: This selection allows the operator to scan through the information saved
under the RUN menu.
8)
Run: Selection of this item starts the system. ESCAPE stops and returns to the main
menu.
3.5
Run:
2)
3)
Save/Clear:
Used to store a reading. An option given is to clear the storage space (stack) for a
new start. The data saved are:
A)
B)
C)
D)
E)
Distance
Velocity
Time
Wave type
System
(In inches)
(In inches/second)
(In microseconds)
(P (1) or S (0))
(English (0) or Metric (1))
NOTE: The numbers saved are always based on English units. If taken in Metric mode,
the conversion has already been accomplished. Memory space is adequate for the storage
of 1000 readings.
4)
3.6
Pulse Rate:
Enables the choice of pulse rates. Options are 1, 3, and 10 pulses/second. Any
other choice causes the rate to be approximately 10 pulses/sec.
FAMILIARIZATION:
A few practice sessions with various cases should remove any mysteries.:
CASE 1: NO HAND HELD TERMINAL
Disconnect keypad, turn on power.
1)
Displays "Contact transducers, then enter." If you wish to re-initialize offset (zero set),
hold the transducers together with couplants
and press the "ENTER" switch. If you wish
to accept the default value, press "ESC."
19
2)
3)
2)
3)
4)
5)
6)
The system should now show the main menu. You may now RUN (8), or change the
variables by menu.
20
Press 2
2)
3)
4)
Press 2.
Press 3
2)
3)
4)
5)
System uses the value entered, takes a time reading, and calculates the distance. Note that
a known value for either velocity or distance could have been entered, and the system
would have calculated the other for initialization.
System now displays the current
values and "Any key to continue."
21
1)
2)
Press 1 (Derived ).
3)
Press 8 (run)
4)
5)
6)
7)
8)
22
3.7
SUMMARY
The operation of this system is primarily a matter of observing the interaction. With each menu
selection the system asks the appropriate questions, and the operator answers them. If the CASE
5 example has been done successfully, it seems unlikely that the operator will need a separate
example of reach menu item. Rather, the best way to gain familiarity with the system is to use it,
or simply practice with it, exercising all menu items and observing the action.
3.8
AGENDA
A.C. OPERATION
The instrument may be operated from the A.C. mains by plugging the AC/charge cable into the
plug on the rear panel and switching the front panel switch to ON. Whenever the instrument is
used on the A.C. mains the internal battery will be trickle charged with a constant current. This
feature ensures that the battery is maintained in a ready to use state.
3.10
BATTERY OPERATION
When the instrument is disconnected fro the mains supply it is only necessary to switch the front
panel switch to ON to operate from the battery. When fully charged the battery will operate the
instrument for 9 hours. Just prior to the battery discharge end point the front panel red light
lights up to indicate the battery low state. The instrument should be turned OFF as overdischarge will damage the battery.
3.11
The discharged battery should be charged at the full rate for 16 hours. Connect the instrument to
the mains supply and leave it on for the required time.
23
SECTION IV
TAKING PULSE VELOCITY MEASUREMENTS
4.1
TRANSDUCERS
Lead Zirconate Titanate ceramics have a relatively low leakage loss and can retain a high voltage
charge for a considerable time. A charge can also build up in a ceramic over a period of time due
to the crystals being subjected to vibrations during transport. Care should be exercised when
handling the coaxial plug prior to connecting to the instrument so as to avoid a shock from a
charged transducer.
4.2
Apply a smear of grease to the transducer faces and press the transducers firmly face to face. The
V-Meter Mk II is highly stable and it is not necessary to make frequent checks. Press the switch
marked "ENT."
4.3
COUPLANT
It is essential in all ultrasonic tests to use some form of couplant between the faces of the
transducers and the material under test. Failure to do so will result in a loss of signal due to
inadequate acoustical coupling.
Silicone grease, medium bearing grease or liquid soap provide good coupling when used on
concrete or other materials having smooth surfaces. For rougher surfaces, water pump grease or
thick petroleum jelly is recommended.
4.4
b.
c.
Method a.
Is the most sensitive method as the receiving transducer will receive maximum
energy from the transmitted pulse.
Method b.
Is the next preferred method and c. should be used only when it is impossible to
get to two faces of the material being tested. The received amplitude of the C
24
method, for the same path length, is only about 2% of that received when using
the method a.
4.5
PULSE VELOCITY
Having determined the method of transducer placement, make careful measurements of the path
length L. Apply couplant to the faces of the transducers and press hard onto the material under
test. Do not move the transducers while a reading is being taken as this can generate noise
signals and errors in measurement.
The pulse velocity is given by: v = L/T where T is the transit time.
It is advisable, when using very long leads, to prevent the two leads from coming into close
contact with each other when transit time measurement are being made. If this is not done it is
possible for the receiver lead to pick up unwanted signals from the transmitter lead resulting in
an incorrect display of transit time. Such incorrect displays are readily detected by their
instability and the fault may be remedied by separating the leads.
25
SECTION V
SYSTEM DESCRIPTION
The V-Meter Mk II generates an ultrasonic pulse in the transmitting transducer and measures the
transmission time taken by the pulse to pass from the transmitter to the receiving transducer.
26
PULSE GENERATOR
The Pulse Generator comprises a high voltage power unit and a high voltage transistor.
The capacitance of the transmitting transducer is charged to a potential of 1.2kV or 500V as
selected by the switch on the top panel. This capacitance is then rapidly discharged through a
transistor triggered by the pulse derived from the pulse selected by the PRF menu. The repetition
frequencies are controlled by CPU.
Discharging the capacitance causes the transmitter to be shock excited and so produce a train of
longitudinal vibrations at its own natural frequency.
5.2
CPU
The central processing unit controls all the functions, calculations, E-prom RAM and memory.
5.3
RECEIVER AMPLIFIER
After transmission through the material under test the ultrasonic pulse is converted to an
electrical signal in the receiving transducer. The received signal is amplified and shaped to
produce a steeply rising "STOP" pulse coincident with the onset of the leading edge of the
received signal waveform.
5.4
A 10 MHz quartz crystal oscillator module generates the timing pulses for the 0.1 micro sec unit.
The 10 MHz pulses are also applied to dividers to produce timing pulses as required. Analog
signals are applied to ADC to convert to digital to put into the CPU and CPU output to the
display.
27
SECTION IV
TESTING CONCRETE
As stated in Section 1 of this Manual, the measurement of the velocity of ultrasonic pulses as a
means of testing materials was originally developed for assessing the quality and condition of
concrete and the V-Meter Mk II will undoubtedly be used predominately for this purpose. This
Section is therefore included to provide the user with some guidance in this field of use.
6.1
APPLICATIONS
The pulse velocity method of testing may be applied to the testing of plain, reinforced and
prestressed concrete whether it is precast or cast in site. The measurement of pulse velocity may
be used to determine:
6.2
a)
b)
c)
changes in the concrete which may occur with time (i.e. due to the cement
hydration) or through the action of fire, frost r chemical attack,
d)
ACCURACY
In most of the applications it is necessary to measure the pulse velocity to a high degree of
accuracy since relatively small changes in pulse velocity usually reflect relatively large changes
in the condition of the concrete. For this reason it is important that care be taken to obtain the
highest possible accuracy of both the transit time and the path length measurements since the
pulse velocity measurement depends on both of these.
It is desirable to measure pulse velocity to within an accuracy of 1%.
When such accuracy of path length measurement is difficult or impossible, an estimate of the
limits of accuracy of the actual measurements should be recorded with the results so that the
reliability of the pulse velocity measurements can be assessed.
6.3
Accuracy of transit time measurement can only be assured if good acoustic coupling between the
transducer face and the concrete surface can be achieved.
28
For a concrete surface formed by casting against steel or smooth timber forms, good coupling
can be obtained if the surface is free from dust and grit and covered with a light or medium
grease or other suitable couplant. A wet surface presents no problem.
If the surface is moderately rough, a stiffer grease should be used but very rough surfaces require
more elaborate preparation. In such cases the surface should be ground flat over an area large
enough to accommodate the transducer face or this area may be filled to a level smooth surface
with a minimum thickness of a suitable material such as Plaster or Paris, cement mortar or epoxy
resin, a suitable time being allowed to elapse for the filling material to harden.
If the value of the transit time displayed remains constant to within 1% when the transducers
are applied and reapplied to the concrete surface, it is a good indication that satisfactory coupling
has been achieved.
6.4
Figure 1 shows three alternative arrangements for the transducers when testing concrete.
Whenever possible, the direct transmission arrangement should be used. This will give
maximum sensitivity and provide a well defined path length.
It is, however, sometimes required to examine the concrete by using diagonal paths and semidirect arrangements are suitable for these.
29
center to center of the transducers. Instead, the method shown in Figure 7 should be adopted to
determine the effective path length.
In this method, the transmitting transducer is placed on a suitable point on the surface and the
receiving transducer is placed on the surface at successive positions along a line and the center to
center distance is plotted against the transit tie. The slope of the straight line drawn through these
points gives the mean pulse velocity at the surface.
In general, it will be found that the pulse velocity determined by the indirect method of testing
will be lower than that using the direct method. If it is possible to employ both methods of
measurement than a relationship may be established between them and a correction factor
derived. When it is not possible to use the direct method an approximate value for Vo may be
obtained as follows:
VD = 1.05 V1
where VD is the pulse velocity obtained using the direct method
V1 is the pulse velocity obtained using the indirect method.
If the points do not lie in a straight line, it is an indication either that the concrete near the
surface is of variable quality or that a crack exists in the concrete within the line of the tests
position (see 6.9)
A change of slope (as shown in Fig. 8) in the plot could indicate that the pulse velocity near the
surface is much lower than it is deeper down in the concrete. This layer of inferior quality could
arise as a result of damage by fire, frost, sulphate attack, etc.
31
32
where Vd
6.5
Vs
Xo
path length,
b)
c)
d)
The influence of path length will be negligible provided it is not less than 4 inches when .75
inches size aggregate is used or not less than 6 inches for 1.5 inches size aggregate.
Pulse velocity will not be influenced by the shape of the specimen provided its least lateral
dimension (i.e. its dimension measured at right angles to the pulse path) is not less than the
wavelength of the pulse vibrations. For pulses of 50 kHz frequency, this corresponds to a least
lateral dimension of about 3 inches. Otherwise the pulse velocity may be reduced and the results
of pulse velocity measurements should be used with caution (see Table 1).
33
Fig. 9: Influence of steel reinforcement on pulse velocity. Bars at right angles to path.
34
The temperature of the concrete has been found to have no significant effect on pulse velocity
over the range from 5 to 30C so that, except for abnormally extreme temperatures, influence
may be disregarded (see Table 2).
Table 1. Effect of specimen
dimensions on pulse transmission
Transducer
frequency
kHz
24
54
82
150
Inches
6
2.6
1.8
1.0
Inches
6.8
3
2.0
1.1
Inches
7.5
3.3
2.2
1.2
C
60
40
20
0
-4
The velocity of pulses in a steel bar is generally higher than they are in concrete. For this reason,
pulse velocity measurements made in the vicinity of reinforcing steel may be high and not
representative of the concrete since the V-Meter Mk II indicates the tie for the first pulse to reach
the receiving transducer.
The influence of the reinforcement is generally very small if the bars run in the direction at right
angles to the pulse path and the quantity of steel is small in relations to the path length.
Figure 9 shows how this influence may be allowed for when the bar diameter lies directly along
the pulse path. If the ration Ls/L is known, the measured pulse velocity may be corrected by
multiplying it by the correction factor corresponding to the ratio and the quality of the concrete.
It is however preferable to avoid such a path arrangement and to choose a path which is not in a
direct line with the bar diameters.
35
36
When the steel bars lie in a direction parallel to the pulse path, the influence of the steel may be
more difficult to avoid as can be seen from Figure 10. It is, however, not easy to make reliable
corrections for the influence of the steel and the correction factors given in Figure 10 should be
regarded as approximate only. It is generally found that these values represent an upper limit of
the steel influence. Again, it is advisable to choose pulse paths which avoid the influence of the
steel as far as possible.
The moisture content of concrete can have a small but significant influence on the pulse velocity.
In general, the velocity is increased with increased moisture content, the influence being more
marked for lower quality concrete. The pulse velocity of saturated concrete may be up to 2%
higher than in dry concrete of the same composition and quality, although this figure is likely to
be lower for high strength concrete.
When pulse velocity measurements are made on concrete as a quality check, a contractor may be
encouraged to keep the concrete wet for as long as possible in order to achieve an enhanced
value of pulse velocity. This is generally an advantage since it provides an incentive for good
curing practice.
6.6
Measurement of pulse velocities at points on a regular grid on the surface of a concrete structure
provides a reliable method of assessing the homogeneity of the concrete. The size of the grid
chosen will depend on the size of the structure and the amount of variability encountered.
It is useful to plot a diagram of pulse velocity contours from the results obtained since this gives
a clear picture of the extent of variations. It should be appreciated that the path length can
influence the extent of the variations recorded because the pulse velocity measurements
correspond to the average quality of the concrete along the line of the pulse path and the size of
concrete sample tested at each measurement is directly related to the path length.
6.7
DETECTION OF DEFECTS
When an ultrasonic pulse traveling through concrete meets a concrete-air interface, there is a
negligible transmission of energy across this interface so that any air filled crack or void lying
directly between the transducers will obstruct the direct beam of ultrasound when the void has a
projected area larger than the area of the transducer faces. The first pulse to arrive at the
receiving transducer will have been diffracted around the periphery of the defect and the transit
time will be longer than in similar concrete with no defect.
It is sometimes possible to make use of this effect for locating flaws, etc. but it should be
appreciated that small defects often have little or no effect on transmission times.
37
6.8
A large cavity may be detected by measuring the transit times of pulses passing between the
transducers when they are placed in suitable positions so that the cavity lies in the direct path
between them. The size and position of such cavities may be estimated by assuming that the
pulses pass along the shortest path between the transducers and around the cavity. Such
estimates are more reliable if the cavity has a well defined boundary surrounded by uniformly
dense concrete.
If the projected area of the cavity is smaller than the diameter of the transducers, the cavity
cannot be detected by transit time measurement alone.
6.9
An estimate of the depth of a crack visible at the surface can be obtained by measuring the transit
times across the crack for two different arrangements of the transducers placed on the surface.
One suitable arrangement is shown in Fig. 11 (a) in which the transmitting and receiving
transducers are placed on opposite sides of the crack and equidistant from it. Two values of x are
chosen, one being twice that of the other, and the transit times corresponding to these are
measured.
The equation given in Fig. 11 (a) is derived by assuming that the plane of the crack is
perpendicular to the concrete surface and that the concrete in the vicinity of the crack is of
reasonably uniform quality.
a)
38
b)
39
40
A check may be made to assess whether the crack is lying in a plane perpendicular to the surface
by placing both transducers near to the crack and moving one of them as in Fig. 11 (b)
It is important that the distance x be measured accurately and that very good coupling is
developed between the transducers and the concrete surface. The method is valid provided the
crack is not filled with water.
6.10
Changes occurring in the structure of concrete with time caused by either hydration (which
increases strength) or by an aggressive environment, such as frost, or sulphates, may be
determined by repeated measurements of pulse velocity at different times. Changes in pulse
velocity are indicative of changes in strength and their measurement can be made over
progressive periods of time on the same test piece or concrete product.
This facility is particularly useful for following the hardening process during the first two days
after casting and it is sometimes possible to take measurement through formwork before it is
removed at very early ages. Figure 12 shows some typical experimental results of pulse velocity
measurements at early ages. This has a useful application for determining when formwork can be
removed or when prestressing operations can proceed. (See also Section VI, Figure 18.)
Figure 12: Typical curves of change of pulse velocity with age.
6.11
Pulse velocity measurements may be used to assess the extent of damages to concrete after a fire.
Figures 13 and 14 shows some typical results as obtained by Watkeys who showed that a good
correlation existed between the maximum temperature reached by the concrete and the
percentage reduction in pulse velocity due to heating. He also showed that a useful correlation
could be obtained to estimate the residual crushing strength of the concrete after heating from
pulse velocity tests.
In Figure 13, the two curves are for two sets of concrete specimens which had been heated and
cooled down either by spraying with water (quenched) or by loss of heat slowly in air
(unquenched). Figure 14 shoes that, for a given residual strength, the pulse velocity was
apparently less than that for undamaged concrete. These results were for concrete made with
gravel aggregate and are typical of normal concrete although no information is available
regarding the effect of different types of aggregate on the correlations.
42
43
Fig. 13: Effect of Maximum Firing Temperature on Ultrasonic Pulse Velocity in concrete (after
Watkeys).
44
Fig. 14: Curve relating residual strength with pulse velocity for concrete damaged by fire.
6.12
ESTIMATION OF STRENGTH
Concrete quality is generally assessed by measuring its cylinder (or cube) crushing strength. It
has been found that there is no simple correlation between cylinder strength and pulse velocity
but the correlation is affected by:
type of aggregate,
aggregate/cement ratio
age of concrete
45
46
The Correlation graph below was obtained by measuring pulse velocity and equivalent cube
strength of beams as shown above.
47
Fig. 16: Typical correlation curves obtained for 1:6 gravel aggregate concrete, cured in water.
48
Although such correlations can be obtained from tests on cylinders, it is preferred in Europe to
use beams such as those used for testing the modulus of rupture of concrete. These beams are
500 mm long and more accurate value of pulse velocity is obtained by using the long axis as the
pulse path. After testing ultrasonically, the beams are tested in flexure to determine the modules
of rupture and the broken halves tested by crushing to measure the equivalent cube strength.
Figure 15 shows a typical curve obtained for a concrete made with river gravel aggregate and
calibrated by using beams.
When testing the concrete in a structure, it would be unreasonable to expect the value of the
cylinder strength estimated from pulse velocity measurements to be the same as that specified for
the control cylinders made on the site since the design of concrete structures takes into account
the fact that cylinder are likely to be of higher strength than the concrete in the structure which it
represents. A suitable tolerance is therefore required to allow for this. This subject is discussed
more fully in other reference.
Figure 16 shows further typical correlation curves including one for modulus of rupture and
these are for a 1:6 mix using gravel aggregate concrete.
Instead of expressing the strength in terms of cylinder strength, it is preferable to obtain a direct
correlation between the strength of a structural member and the pulse velocity whenever this is
possible. Such correlation can often be readily applied to precast units and it is possible to obtain
a curve relating pulse velocity with the appropriate mechanical test (such as bending) for the
unit.
6.13
The estimating of elastic modulus is less complicated than the estimation of strength and it has
been found that a single curve may be used to relate pulse velocity to elastic modulus for a wide
range of different aggregates, including concrete made with lightweight aggregate (see Table 3).
49
Fig, 17: Curves relating Pulse Velocity with Static and Dynamic Elastic Modulus
50
6.14
The relationship between elastic constants and the velocity of an ultrasonic compressional wave
pulse in an isotropic elastic medium of infinite dimension is given by:
Ed = V2 (1+)(1-2)/(1-)
where
Also, the dynamic modulus of elasticity Ed, found from the longitudinal resonant frequency test
(see ASTMC 215), is given by:
Ed = 4fL 2L2
51
Fig. 18: Dynamic Poisson's Ratio derived from longitudinal resonant frequency and longitudinal
52
pulse velocity.
where fL
L
p
0.45
0.40
0.35
0.30
0.25
0.20
0.15
0.10
0.05
53
SECTION VII
ADDITIONAL ACCESSORIES FOR USE WITH THE V-METER MK II
7.1
Several different frequency transducers are available for specific applications other than the
standard 54 Khz and 150 kHz i.e. 24 kHz, 37 kHz, 84 kHz, 500 kHz etc.
7.2.
Shear wave transducers with approximately 100 kHz frequency are available to measure and
calculate Poisson's ratio.
7.3
Transducers with different shapes are available other than standard flat faced cylindrical
configurations for example, exponential transducers for very rough surfaces or wheel transducers
for fast scanning of flat surface.
7.4
Voltage gain
Frequency response
x4/x7
100 Hz-240kHz "with 200 ft. cable on output - 100
Hz - 60kHz"
Amplifier delay
Amplifier noise
Protection
Life
Switch
3 positions
x 4 Amplifier
x 7 Amp.
D/Off Direct. Transducer connected directly to
output socket.
TO CHANGE BATTERIES
1.
2.
3.
4.
5.
Remove lid.
Remove old battery.
Insert new battery.
Red (+) Back (-)
Replace lid.
OPERATION
1.
The C-4896 is intended for use with the V-Meter Mk II when making U.P.V.
measurements on very early age concrete, or long path lengths.
The input to the amplifier is via a BNC plug which enables the amplifier to be plugged
directly into the socket on conventional transducers.
3.
The V-Meter Mk II should be zeroed with pre-amp in circuit and switched on for long
path lengths.
Thee is a short delay, typically 0.2 microsecond, in the amplifier, but as the C-4896 is
intended for use only at the start of a test when the transit time can be several hundred
microsecond the delay can be considered negligible. After about 3 hours from casting the
C-4896 can be switched to the "D" position as the signal strength will have increased
considerably.
S the amplifier will be switched to "D" before the end of recording, the V-Meter Mk II
and recorder should be zeroed at the commencement of test with the amplifier switch to
"D."
Because of the increased sensitivity of the system it is essential that the transducers and
cables be perfectly stationary when reading are taken.
55
7.5
Functionally the same as C-4896 but can be screwed into 2" diameter transducers to form an
integral part of the receiving transducer.
7.6
This hand held terminal enables the operator to interact with the computer to get more features of
the V-Meter Mk II other than transit time and graphs. See 3.6 for details.
56
REFERENCES
F-6063 Ultrasonic Testing of Concrete. The use of the V-Meter. General Paper. James
Instruments, Inc.
V-103 Pulse Velocities Recorded on Sectioned Pieces of Wooden Utility Poles. R.A. Muenow,
1966.
V-107 Non-Destructive Testing of Construction Materials. James Electronics.
V-108 Report on Pulse Velocity Measurement for the Garage Structure - First National
Building, Detroit, Michigan. R.A. Muenow, 1965.
V-117 The Ultrasonic Nondestructive Evaluation of Wood Utility Poles, Wood and Concrete
Railway Ties and Trestles, and Other Similar Structures. R.A. Muenow, 1966.
V-118 Pulse Velocity - A method of Evaluating and Inspecting Large and Small Hydro-Electric
Structures. R.A. Muenow, 1966.
V-119 The Evaluation of a Reinforced Concrete Frame Building by Pulse Velocity Methods.
R.A. Muenow, 1966.
V-122 The Determination of fracture Healing by Measurement of Sound Velocity Across the
Fracture Site. Siegel/Anast/Fields, 1958.
V-124 Fire Resistance of a Reinforced Concrete Building. Observations Recorded After the
First and Repair of Damage. Brice/Chefdeville, 1959.
V-127 Evaluation of Wooden Elements Taken from Various Railroad Structures. James, 1966.
V-129 The Use of Velocity Measurements on Living Trees. James, 1966.
V-130 Caisson Investigations. James, 1966.
V-131 The Correlation of Compressive Strength with the Velocity of Sound Propagation
Through Concrete. R.A. Muenow, 1967.
V-138 Caissons Inspection at Hot Strip Mill for Youngstown Sheet & Tube Company,
Hammond, Indiana. R.A. Muenow, 1966.
V-143 Evaluation of Prestressed Concrete Beam. Richard A. Muenow.
V-144 Inspection of Solid Fuel Elements. R.A. Muenow, 1967.
57
58
59
V-189 Report on the Failure of Roof Beams at Sir John Cass' Foundation and Red Coat Church
of England Secondary School, Stepney. S.C. Bate, 1974.
V-190 Further Notes on Ultrasonic Techniques. W.R. Davis, 1974.
V-191 Ultrasonic Control.
V-194 Further Investigations Into the Strength of Concrete in Structures. S.G. Davis.
V-195 Nondestructive Testing of Particleboard. R. Pellerin & Chas. Morschauser, 1973.
V-196 Ultrasonic Testing of Ceramics and Refractories with James V-Meter. K. Choi, 1978.
V-197 Utility Pole Testing, Report on test program with V-Meter.
V-198 Past and Future of Concrete Quality Evaluation. Journal of The Construction Division. V.
Ramakrishnan.
V-199 Nondestructive Method for Measuring the Electric Anisotropy of Wood using an
Ultrasonic Pulse Technique. I.D.G. Lee
V-200 Nondestructive testing of concrete and timber. Institute of Civil Engineers and the British
Natl. Com. For nondestructive testing.
V-201 Lumber Defect Detection by Ultrasonics. Forest Service, Madison, Wisconsin
V-202 Nondestructive Testing of Wood. College of Engineering Research, Washington State
University. Wm. L. Galligan.
V-204 The Ultrasonic Pulse Velocity Method of Test for Concrete in Structures. Cement and
concrete Association. H.N. Tomset, Oct. 1977.
V-205 Testing of Wood and Wood Products with Low Frequency Ultrasonics. A Review, .
Choi, 1978.
V-207 Assessment Techniques for Graphite Electrodes. John R. Lakin, Fuel, 1978
V-208 An Appraisal of the Ultrasonic Pulse Technique for Detecting Voids in Concrete. H.W.
Chung, University of Hong Kong.
V-209 The In Situ Evaluation of Concrete Using Pulse Velocity Differences. H.N. Tomsett,
1978 American Concrete Institute Fall Convention in Houston, Texas.
V-210 "The Cylinder Test" Reliable Informer or False Prophet. James M. Shilstone, Sr., FACI
60
Sonic Testing of Refractories. T.H. Hawisher & C.E. Semler, American Ceramic Society
80th Annual Meeting, Detroit, 1978.
2.
Improved Reliability of Fused S1O2 Pouring Tubes for Continuous Casting with
Nondestructive Inspection. M.S. Judd, C.G. Hammersmith, R.L. Wessel & D.M. Scott,
American Ceramic Society 81st Annual Meetings, Cincinnati, 1979.
3.
4.
Pulse Ultrasonics in Process Control. Bruce Dunsworth and Don Smith. Refractory
Division Fall Meeting, Bedford Spring, PA, 1979.
5.
Nondestructive Ultrasonic Testing of fireclay Refractories. L.B. Lawler, R.H. Ross and
E. Ruh, American Ceramic Society 82nd Annual Meeting, Chicago, 1980.
6.
Refractory Evaluation with Pulse Ultrasonic. W. Miller, American Ceramic Society 82nd
Annual Meeting, Chicago, 1980.
7.
Cordierite Slab. W.C. Mohr, B.E. Dunworth, D.B. McCuen and M.W. Morris, American
Ceramic Society 82nd Annual Meeting, Chicago, 1980.
8.
Fracture Energy Testing: Single Bar Determination of WOF and NBT. Merrill Wood,
General Refractories Company, U.S. Refractories Division - Research Center, Baltimore,
MD.
9.
Sonic Velocity Quality Control of Steel Plant Refractories. Robert O. Russell & Gary D.
Morrow, American Ceramic Society, Ceramic Bulletin, Vol. 63 no. 7, 1984.
61