You are on page 1of 2

New Product Guide

Fully Automated Clinical


Chemistry Analyzer
CA-
CA-180

 Throughput: 180 tests/hour (Up to 450 tests/hour w/ISE)


 Environment-friendly reusable Pyrex cuvette
 Reagent and sample barcode ID recognition available
 Optional built-in ISE unit available
Specifications
Type Desktop fully automated random access analyzer
Usage General chemistry as photometric assay
Immunology as photometric assay (Latex reagents available)
Assay type 1 point end, 2 point end, 1 point rate, 2 point rate and electrolytes
Through put 180 tests per hour (Up to 450 tests per hour with ISE)
Incubation time Five minutes after applying sample for one reagent assay
Five minutes after applying second reagent for two reagents assay
Sample type Serum, Plasma and Urine
Sample input system Type: Removable tray with sample tube holder on a turntable
Barcode sample ID detection with optional barcode reader
Capacity: Up to 40 samples (handling of STAT available)
Sampling pipette Type: Micro-pipette with liquid level detector
Rinsing: Inside and outside with purified water
Sampling pump Type: Micro syringe
Volume: 2 to 35 ul (increment by 0.1ul)
Reagent system Type: Removable tray with reagent bottle holder on a turntable
Barcode reagent ID detection with barcode reader
Capacity: 40 reagent positions on a tray
(Twenty 100ml and twenty 20ml bottle positions)
Inventory: Calculation of remaining reagent volume available
Cooling: Cooling with Peltier element. (8 to 15 degrees Celsius)
Reagent pipette Numbers of pipettors: One
Type: Micro-pipette with liquid level detector
Rinsing: Inside and outside with purified water
Number of reagents: Up to two reagents
Inventory: Calculation of remaining reagent volume available
Reagent pump Type: Micro syringe
Volume: 20 to 400 ul (increment by 1ul)
Cuvette system Number of cuvettes 45 cuvettes on reaction line
Type: Reusable with onboard washing system
Material: Pyrex
Volume for measurement: Minimum 180 ul
Maximum volume: 500 ul in total
Reaction system Type: Direct heating system
Temperature: 37 degrees Celsius
Detector Method: Direct measurement of absorbance in cuvette
Bi-chromatic or Mono-chromatic
Wavelength: 8 selectable from 340 nm to 800 nm
Light source: Halogen tungsten lamp
Stirring system Type: Stick type rotating stirrer
Dimensions 770 mm (W) x 620 mm (D) x 505 mm (H) (Analyzer only)
Weight 135 Kgs
Option ISE unit, Sample Barcode Reader, Liquid Level Sensors

Notes: Specifications subject to change without notice


Details of specifications to be determined based on discussion and agreements of further meetings.

FURUNO ELECTRIC CO., LTD.


System Products Division
2-20, Nishinomiyahama, Nishinomiya-City, 662-0934, Japan
Tel: +81 798-33-7555 Fax: +81 798-33-7511
URL: http://www.furuno.co.jp