You are on page 1of 5

Founded 1964

LND, INC.
Designers & Manufacturers of Nuclear Radiation Detectors

3230 LAWSON BLVD., OCEANSIDE, NEW YORK 11572 E-mail: info@lndinc.com Web Site: http://www.lndinc.com 1-516-678-6141 Fax: 1-516-678-6704

7311 Pancake Mica Window-Alpha-Beta-Gamma Detector

GENERAL SPECIFICATIONS
Gas filling Cathode material Maximum length (inch/mm) Effective depth (inch/mm) Maximum diameter (inch/mm) Effective diameter (inch/mm) Connector Operating temperature range C Ne +Halogen 446 Stainless Steel 3.00/76.1 0.5/12.7 2.11/53.6 1.75/44.5 Grid Cap -55 to +75

WINDOW SPECIFICATIONS
Areal density (mg/cm) Effective diameter (inch/mm) Material 2.0 1.75/44.5 Mica

ELECTRICAL SPECIFICATIONS
Recommended anode resistor (meg ohm) Maximum starting voltage (volts) Recommended operating voltage (volts) Operating voltage range (volts) Maximum plateau slope (%/100 volts) Minimum dead time (micro sec) Gamma sensitivity Co60 (cps/mr/hr) Tube capacitance (pf) Weight (grams) Maximum background shielded 50mm Pb + 3mm Al (cpm) 3.3 750 900 850-1000 10 20 60 3 125 30

You might also like