Professional Documents
Culture Documents
instructions.
Changing Batteries
1. Open battery INC
compartment
++
lid (with
attached
lanyard loop).
2. Remove old
batteries;
replace
with fresh DEC
ones. Note
Diagram 1
polarity as
shown in battery compartment. See Diagram 1.
3. Recalibrate after battery change.
Warranty
Eutech Instruments warrants its meter free from
manufacturing defects for 2 years and electrode
module for 6 months. If repair, adjustment or
replacement is necessary and has not been the result
of abuse or misuse within the time period, please
return the tester – freight prepaid – and correction
will be made without charge. Out of warranty products
will be repaired on a charge basis.
Return Of Items
Authorization must be obtained from Eutech’s
distributor before returning items for any reason.
When applying for authorization, please include
information regarding the reason the item(s) are to • Waterproof & Floats
be returned.
Note: We reserve the right to make improvements in
• Replaceable Electrode
design, construction and appearance of products • Auto-off
without notice. Prices are subject to change without
• Automatic Temperature
notice.
Compensation (ATC)
Waterproof ECScan High Instructions
+
lid (with
attached
lanyard loop).
2. Remove old
batteries;
replace with
fresh ones. DEC
Note polarity
Diagram 1
as shown in
battery compartment. See Diagram 1.
3. Recalibrate after battery change.
Warranty
Eutech Instruments warrants its meter free from
manufacturing defects for 2 years and electrode
module for 6 months. If repair, adjustment or
replacement is necessary and has not been the result
of abuse or misuse within the time period, please
return the tester – freight prepaid – and correction
will be made without charge. Out of warranty products
will be repaired on a charge basis.
Return Of Items
Authorization must be obtained from Eutech’s
distributor before returning items for any reason.
When applying for authorization, please include
information regarding the reason the item(s) are to
• Waterproof & Floats
be returned.
Note: We reserve the right to make improvements in • Replaceable Electrode
design, construction and appearance of products • Auto-off
without notice. Prices are subject to change without
notice. • Automatic Temperature
Compensation (ATC)
Waterproof ECScan Low Instructions
++
attached
lanyard loop).
2. Remove old
batteries;
replace with
fresh ones.
Note polarity DEC
as shown
in battery Diagram 1
compartment. See Diagram 1.
3. Recalibrate after battery change.
Tester Maintenance
• To improve performance, clean the electrode
by rinsing them in alcohol for 10 to 15 minutes.
• Replace all batteries if low battery indicator
appears, or if readings are faint or unstable.
• If you experience drift, periodically let electrode
fully dry.
When you need a new electrode, see “Electrode
Replacement” on insert in back of box.
Warranty
Eutech Instruments warrants its meter free from
manufacturing defects for 2 years and electrode
module for 6 months. If repair, adjustment or
replacement is necessary and has not been the result
of abuse or misuse within the time period, please
return the tester – freight prepaid – and correction
will be made without charge. Out of warranty products • Waterproof & Floats
will be repaired on a charge basis. • Replaceable Electrode
Return Of Items • Auto-off
Authorization must be obtained from Eutech’s • Adjustable TDS Factor
distributor before returning items for any reason.
When applying for authorization, please include • Automatic Temperature
information regarding the reason the item(s) are to
be returned.
Compensation (ATC)
Note: We reserve the right to make improvements in
design, construction and appearance of products
without notice. Prices are subject to change without
notice.
WP TDScan High Instructions
+
attached
lanyard loop).
2. Remove old
batteries;
replace with
fresh ones.
Note polarity DEC
as shown
in battery Diagram 1
compartment. See Diagram 1.
3. Recalibrate after battery change.
Tester Maintenance
• To improve performance, clean the electrodes
by rinsing them in alcohol for 10 to 15 minutes.
• Replace all batteries if low battery indicator
appears, or if readings are faint or unstable.
• If you experience drift, periodically let electrode
fully dry.
When you need a new electrode, see “Electrode
Replacement” on insert in back of box.
Warranty
Eutech Instruments warrants its meter free from
manufacturing defects for 2 years and electrode
module for 6 months. If repair, adjustment or
replacement is necessary and has not been the result
of abuse or misuse within the time period, please
return the tester – freight prepaid – and correction
will be made without charge. Out of warranty products • Waterproof & Floats
will be repaired on a charge basis. • Replaceable Electrode
Return Of Items • Auto-off
Authorization must be obtained from Eutech’s • Adjustable TDS Factor
distributor before returning items for any reason.
When applying for authorization, please include • Automatic Temperature
information regarding the reason the item(s) are to Compensation (ATC)
be returned.
Note: We reserve the right to make improvements in
design, construction and appearance of products
without notice. Prices are subject to change without
notice.
WP TDScan Low Instructions
6. After 3 seconds without a key 3. After 3 seconds without a key press, the display
press, the display flashes 3 times, flashes 3 times, then shows ‘ENT’. Tester accepts
then shows ‘ENT’. The tester accepts TDS factor and returns to measurement mode.
calibration value; returns to
4. Replace battery cap.
measurement mode.
7. Replace battery cap.
Self-Diagnostic Messages
bAt Weak batteries–replace with fresh ones specified
by manufacturer.
Err Wrong keypress.
Or, Ur Over range/Under range signal, or electrode
is not in contact with solution, or electrode is failing.
---- Calibration mode activated but not performed.
Lanyard
connection
(Picture B)
REPLACE THE ELECTRODE
Replaceable B
Sensor
• Push-button calibration
• Fast, stable, repeatable readings
• Available through authorized distributors
SPECIFICATIONS
Tester TDScan TDScan ECScan ECScan Electrode Replacement
Low High Low High
Range 0 to 0 to 0 to 0 to You can replace the electrode module at
1990 10.00 1990 19.90 the fraction of the cost of a new tester.
ppm ppt µS/cm mS/cm When the tester fails to calibrate or gives
Resolution 10 0.10 10 0.10 fluctuating readings in calibration
ppm ppt µS/cm mS/cm standards, you need to change the
Accuracy ±1% Full Scale electrode.
TDS factor 0.4 to 1.0 selectable - 1. With dry hands, grip the ribbed tester
Calibration 200 to 2 .00 to 200 to 2.00 to collar with electrode facing you. Twist the
Standard 1990 10.00 1990 19.90 collar counter clockwise (see picture A).
Range ppm ppt µS/cm mS/cm Save the ribbed tester collar and O-ring
Calibration 1 point (calibration range is ±30% for later use.
of factory default parameter) 2. Pull the old electrode module away from
ATC 0 to 50 °C the tester.
Temp 2% per °C 3. Align the four tabs on the new module
Coefficient so that they match the four slots on the
Operating 0 to 50 °C tester (see picture B).
Temp 4. Gently push the module onto the slots
Power Four X 1.5V to sit it in position. Push the smaller O-
Battery (Type: A76, LR 44 or V136A); ring fully onto the new electrode module.
> 140 hrs continuous use Push the collar over the module and thread
Dimension; Tester: 16.5 cm x 3.8 cm; 90 g it into place by firmly twisting clockwise.
Weight Boxed: 22 cm x6 cm x 5 cm; 170 g
Applications
Manufactured By:
Eutech Instruments Pte Ltd Water quality testing • water and
Blk 55, Ayer Rajah Crescent, Wastewater treatment • hydroponics
#04-14/24, Singapore 139949
Tel: (65)778 6876 Fax: (65)773 0836 • labs • boilers • factories
E-Mail: marketing@eutechinst.com • ecology studies • and more!
Web-Site: http://www.eutechinst.com
Made in Singapore
68X247828 11/00 Rev 0