1000
Onboard Dilutions
Load Peagenl Wedges and Diluenl Wedges onlo lhe Peagenl Carousel.
Load each onboard dilulion in lhe ollowing sequence.
1) Place a Sample Cup Holder conlaining undiluled sample on lhe Load Plalorm.
2) Place a Dilulion Cup Holder conlaining an emply Sample Cup in lhe nexl posilion on
lhe Load Chain. (Dilulion Cup Holders can be dislinguished rom Sample Cup Holders
by a yellow ring below lhe barcode.)
3) Leave an emply posilion immedialely ollowing lhe Dilulion Cup Holder.
4) Place a maximum o our Tesl Unils lhal use lhe same dilulion aclor and diluenl be-
hind lhe emply posilion. The irsl Tesl Unil delermines lhe dilulion aclor and lype
o diluenl lo be used.
IMMULITE 1000
Quick Reference Guide
Tray o Dilulion
Cup Holders
Diluenl Wedge Undiluled
palienl sample
in Sample Cup
Holder
Lmply Sample
Cup in Dilulion
Cup Holder
Lmply posilion Up lo our Tesl
Unils lhal use
lhe same dilulion
aclor and diluenl