From Wikipedia, the free encyclopedia

Network isolators are installed as part of a copper Ethernet system, as galvanic isolators.

Network data continues to be transmitted across an electrically non-conducting barrier, through the applied
principle of electromagnetic induction, whereby high frequency AC voltages conveying data are induced across an
isolating gap.

The network isolator is therefore a passive device, and functions without any requirement of an external power
supply.

Contents
1 Applications
1.1 Medical
1.2 Other applications
2 Technical requirements
3 Models
4 See also
5 Sources
6 External links

Applications
Medical

Network isolators are used in the medical industry to protect patients against leakage currents.

Network connections between medical devices and Ethernet networks (and other non-medical devices, such as
personal computers) must be in accordance with the IEC standard IEC 60601-1 (3rd edition). This standard
specifically deals with medical electronic and electrical equipment and systems, and classifies non-medical devices
as potential hazard sources. A specific hazard stems from possible differences in ground potentials between
network components, which when not properly isolated can result in a leakage current that can flow through the
patient, which is dangerous, and potentially lethal. Such voltage differences can also arise through incorrect
installation and wiring of network systems, electrical shorting within damaged cables and cabling, or shorting
between damaged network cables and other voltage sources.

Network isolators work to remove this hazard, by electrically disconnecting medical devices from a network.
Isolators may be used as network accessories, built into medical devices, or installed within a medical network
system. Networks isolators should be installed as close as possible to the medical device in question.

en.wikipedia.org/wiki/Network_isolator 1/3
11/7/2009 Network isolator - Wikipedia, the free e…
As they serve no therapeutic or diagnostic purpose, network isolators themselves are not classed as medical
electronic equipment according to the standard IEC 60601-1, nor do they fall within the scope of the Medical
Devices Directive 93/42/EEC.

They are often installed in the medical field in conjunction with isolation transformers, which serve to protect the
patient from electrical faults.

Other applications

Network isolators have many other applications in overcoming the problems of differing ground potentials across
networks, or between network components. For example:

Electrical testing facilities, where measurement and monitoring equipment are connected to a central control
centre through an Ethernet;
For redundant server systems, which are physically connected via a copper network;
For private or commercial networks, where inherent potential differences within a building, or between
buildings, become problematic, and a fibre-optic solution is not economically viable;
General usage as filters for shield currents in Ethernet connections.

Technical requirements
The international standard IEC 60601-1 Medical Electrical Equipment (3rd edition) specifies stringent criteria on
the safety and isolation of medical devices. These requirements are of course much more stringent than those of a
typical consumer product standard (i.e. IEC 60950 Safety of Information Technology Equipment), as the patient
may be unconscious, anaesthetised, or otherwise unable to move, and the need for patient protection is therefore
paramount.

The IEC 60601-1 standard requires that two independent precautionary measures (Means of Patient Protection, or
MOPPs) are set in place, to protect the patient from electrical shock in the medical system. A network isolator can
be constructed to provide one or both required MOPPs. In the case where only a single MOPP is present in a
system, a second measure of protection for the medical device must be made. This can be achieved, for example,
by installing a fixed, low-resistance earthing cable, bar, or strap, referred to as a potential equalisation conductor.
Conformation to the requirements of applicable standards and guidelines, in particular the IEC 60601-1 standard,
involves manufacturers themselves declaring CE compliance and/or gaining certification from an approved
independent testing laboratory.

The number of MOPPs a network isolator provides will be stated in the certification of the device. These are
achieved by a combination of clearance and creepage distances, and the breakdown voltage of the insulating
materials. Although requirements on breakdown voltage levels are 1.5 kilovolts for one MOPP, and 4 kilovolts for
two MOPPs, the number of MOPPs actually provided cannot be derived from the breakdown voltage alone.

Network connections made with unshielded cables are occasionally used as a form of isolation, in that no shield
connection exists between the device and the network. However, this fails to provide sufficient protection in the
scope of IEC 60601-1, as the data lines themselves are not isolated, and may be floating at dangerous voltage
potentials.

Network cards and other active network components which are designed to meet IEC 60950 (Safety of
Information Technology Equipment), will usually not comply with the more demanding isolation requirements of IEC
en.wikipedia.org/wiki/Network_isolator 2/3
11/7/2009 Network isolator - Wikipedia, the free e…
60601-1 (Medical Electrical Equipment), and therefore require additional isolation before being used in medical
applications.

The transmission quality of a network isolator can be determined by measuring the industry standard parameters or
characteristics, for example, insertion loss, return loss, and near end crosstalk.

By definition, a network isolator completely eliminates each and every direct electrical connection between devices
connected through a network connection - specifically all the data lines, and the cable shield.

Models
Network isolators are commercially available in various designs, to suit customers’ requirements:

stand-alone devices, ready for immediate and simple installation;


isolating components, suitable for OEM manufacturers;
integrated into network wall sockets, for permanent installations.

Voltage ratings (rated in kilovolts) and maximum transmission speeds (e.g. 1000BASE-T) vary between models.

See also
Ethernet over twisted pair
Ethernet physical layer
Medical devices

Sources
Krankenhaus IT Journal Ausgabe 05/2008
DeviceMed Ausgabe 10/2008

External links
EMO Systems range of network isolators (http://www.emosystems.de/en/products/network-isolator)
International Electrotechnical Commission (http://www.iec.ch/)
Retrieved from "http://en.wikipedia.org/wiki/Network_isolator"
Categories: Networking hardware

This page was last modified on 17 October 2009 at 21:46.


Text is available under the Creative Commons Attribution-ShareAlike License; additional terms may apply.
See Terms of Use for details.
Wikipedia® is a registered trademark of the Wikimedia Foundation, Inc., a non-profit organization.

en.wikipedia.org/wiki/Network_isolator 3/3

You might also like