Laboratory

Matrix Lab, 230-231, 2nd Floor, Heena Arcade, G.I.D.C., Char Rasta, Vapi,
Gujarat

Accreditation Standard ISO/IEC 17025:2005


Discipline

Electro-Technical Calibration

Issue Date

26.11.2013

Certificate Number

C-0986

Valid Until

25.11.2015

Last Amended on

27.01.2014

Page

1 of 3

Quantity Measured/
Instrument

Range / Frequency

*Calibration Measurement
Capability ()

Remarks

SOURCE#
1.

DC VOLTAGE

1mV to 2V
2V to 200V
200V to 900V

0.86% to 0.5 %
0.5% to 1.2%
1.2% to 2.2%

Using 5 Multi
Function Calibrator
By Direct Method

2.

AC VOLTAGE

50Hz
10mV to 2V
2V to 200V
200V to 900V

0.8% to 0.6%
0.6 % to 0.55%
0.55 to 0.65 %

Using 5 Multi
Function Calibrator
By Direct Method

3.

DC CURRENT

1mA to 20 mA
20 mA to 1A
1 A to 9 A
100A to 800A

0.8% to 0.2 %
0.2 % to 0.1.15%
1.15 % to 1.8 %
3.8 % to 5.8 %

Using 5 Multi
Function Calibrator
By Direct Method
With Current Coil for
(100 A to 800 A)

4.

AC CURRENT

50Hz
1mA to 20 mA
20 mA to 1mA
1 A to 9 A
100A to 800A

0.40% to 0.55%
0.55 % to 1.52 %
1.52 % to 0.85 %
2.6% to 4.2%

Using 5 Multi
Function Calibrator /
Current Coil
By Direct Method

45Hz to 950 Hz

0.82% to 0.31%

Using 5 Multi
Function Calibrator
By Direct Method

1 to 100
100 to 100 k
100 k to1000 M

2.45 % to 2.35 %
2.35 % to 2.35 %
2.35 % to 2.8 %

Using Decade
Resistance Box
Vaiseshika/ 8400
By Direct Method

5.

FREQUENCY

6.

DC RESISTANCE

Laboratory

Matrix Lab, 230-231, 2nd Floor, Heena Arcade, G.I.D.C., Char Rasta, Vapi,
Gujarat

Accreditation Standard ISO/IEC 17025:2005


Discipline

Electro-Technical Calibration

Issue Date

26.11.2013

Certificate Number

C-0986

Valid Until

25.11.2015

Last Amended on

27.01.2014

Page

2 of 3

Quantity Measured/
Instrument

7.

TEMPERATURE
SIMULATION
(Controller, Recorder,
Indicator)
RTD Type
PT-100
Thermocouple
J Type
K Type
T Type
R Type
S Type

Range / Frequency

*Calibration Measurement
Capability ()

-20 to 850C

0.15C to 1.2C

-200C to 1200C
-200C to 1370C
-200C to 400C
0 to 1750C
0 to 1750C

1.3C to 3.55C
1.3C to 4.1C
1.7C
2.0C to 5.2C
1.8C to 5.2C

Remarks

Using Masibus
Universal Calibrator
UNICAL 3001M
By Direct Method

MEASURE$
1.

DC VOLTAGE

10mV to 1V
1V to 100V
100V to 1000V

0.23% to 0.4 %
0.4% to 0.1 %
0.1 % to 0.12 %

Using Fluke 8846A 6


Digital Multimeter
By Direct Method

2.

AC VOLTAGE

50Hz
10mV to 1V
1V to 100V
100V to 1000V

0.62% to 0.14 %
0.14% to 0.86 %
0.86 % to 0.4 %

Using Fluke 8846A 6


Digital Multimeter
By Direct Method

AC HIGH VOLTAGE

1 kV to 10 kV

6.53%

Using High Voltage


Probe with DMM
By Direct Method

DC CURRENT

1mA to 10 mA
10 mA to 1A
1 A to 10 A

0.4% to 0.1%
0.1% to 0.84%
0.84% to 0.65%

Using Fluke 8846A 6


Digital Multimeter
By Direct Method

3.

Laboratory

Matrix Lab, 230-231, 2nd Floor, Heena Arcade, G.I.D.C., Char Rasta, Vapi,
Gujarat

Accreditation Standard ISO/IEC 17025:2005


Discipline

Electro-Technical Calibration

Issue Date

26.11.2013

Certificate Number

C-0986

Valid Until

25.11.2015

Last Amended on

27.01.2014

Page

3 of 3

Quantity Measured/
Instrument

4.

AC CURRENT

5.

DC RESISTANCE

Range / Frequency

*Calibration Measurement
Capability ()

Remarks

50Hz
1mA to 100 mA
100 mA to 1A
1 A to 10 A

0.4% to 0.55%
0.5 % to 1.52 %
1.52 % to 0.82 %

Using Fluke 8846A 6


Digital Multimeter
By Direct Method

1 to 100
100 to 100 k
100 k to1000 M

1.6 % to 0.15 %
0.15 % to 0.16 %
0.16 % to 2.8 %

Using Fluke 8846A 6


Digital Multimeter
By Direct Method

* Measurement Capability is expressed as an uncertainty () at a confidence probability of 95%


$

Only in Permanent Laboratory


The laboratory is also capable for site calibration however, the uncertainty at site depends on the prevailing actual
environmental conditions and master equipment used.
#