Rapid Communications
This paper reports on some preliminary experimental results of a work in progress regarding a problem involving the
quantitative analysis of hydrogen isotopes by mass spectrometry of low resolution: the triatomic (trimer) ions interferences
with the isotopic hydrogen species having the same mass/charge. These results indicate that, in complex mixtures of
hydrogen isotopes, trimer ions are strongly affected by the presence of other species, and a new approach that takes into
account the destruction mechanism of trimer ions is necessary for a proper determination of their contributions.
(Received September 20, 2012; Accepted October 1, 2012; Published November 10, 2012)
Introduction
One of the main issues in the quantitative analysis of hydrogen
isotopes gas mixtures by mass spectrometry of low resolution is
the interferences of triatomic (also named trimer) ions with ions
of isotopic species of molecular hydrogen having the same
mass/charge (e.g. H3+/HD+, H2D+/[D2++HT+], HD2+/DT+, and so
forth). The contributions of these trimer ions to the total
ion-currents measured at the m/z ratios of interest have to be
properly determined and subtracted as a condition for obtaining
reliable analysis results for hydrogen isotopic composition. The
pathway for the production of trimer ions in the electron impact
ion source is the reaction between molecular hydrogen species
and its ions (e.g. reaction H2+ + H2 H3+ + H).1 In hydrogen
isotope ratio mass spectrometry (IRMS, conventional or modern),
the main problem is the interference of H3+ with HD+ at mass
3ion-current.24 According to the above reaction, the production
of H3+ is proportional to both the abundance of H2+ and the H2
partial pressure. Under linear operation conditions of the mass
spectrometer, the abundance of H2+ is proportional to the partial
pressure of H2 gas in the ion source. Thus, the production of
H3+ is proportional to the second power of the H2+ abundance.
The result is that for natural (or close to natural) molecular
hydrogen the total current measured at mass 3 is:35
I3 = IH + IHD = KI22 + R*I2,
+
(1)
E-mail: nicoale.bidica@icsi.ro
Experimental
The experimental results reported here were obtained with a
commercial Quadrupole Mass Spectrometer (QMS) of unit
resolution, Hiden Analytical HPR20, intended to replace the
current gas-chromatograph used to monitor hydrogen isotope
species in the process gases to the water detritiation facility
from ICSI Rm. Valcea, Romania. The inlet system utilises a
one-metre long capillary sampling line that is connected directly
to the ion source, and is bypass pumped. Bypass pumping
provides rapid evacuation of the inlet system volume; it also,
with the aim of a micrometric control valve, allows the user to
control the pressure in the ion source. A cold cathode (Penning)
gauge is fitted to monitor the UHV housing pressure. The
operation parameters of the (gas-tight) ion source were: cage
voltage, 5 V; focus voltage, 110 V; electron energy, 70 V; and
emission current, 1 mA.
The samples used were chemically pure hydrogen (with
natural content of deuterium) and deuterium (with about 0.2%
content of hydrogen), and laboratory-prepared mixtures (by
partial pressure method of mixing) of hydrogen and deuterium
with nitrogen gas. The sample IDs and the corresponding
approximate proportions are as follows: HD-1 (50% H2, 50%
D2), HN-1 (75% H2, 25% N2), HN-2 (50% H2, 50% N2), HN-3
(30% H2, 70% N2), HN-4 (10% H2, 90% N2), DN-1 (80% D2,
20% N2), DN-2 (50% D2, 50% N2), DN-3 (30% D2, 70% N2),
and DN-4 (10% D2, 90% N2).
1034
ANALYTICAL SCIENCES
I6 = ID = KI42.
+
(2)
References
1. T. R. Hogness and E. G. Lunn, Phys. Rev., 1925, 26, 44.
2. I. T. Platzner, K. Habfast, A. J. Walder, and A. Goetz, in
Fig. 2Variation of the ion-current ratios, I3/I2 and I6/I4, on I28 for the
two sets of samples analysed, respectively, with I2 and I4 established at
4 1011 A.
Fig. 1Ion currents measured at different values for the total pressure in UHV housing: (a) I3 as
function of I2, for H2/N2 mixtures, (b) I6 as function of I4, for D2/H2 and D2/N2 mixtures.
1035
Fig. 3Two mass spectra recorded for samples D2 and DN-4 for the same value of the ion-current
measured at m/z 4 (about 4 1011 A).