Objectives
11Equation Section (Next)To calibrate an orifice meter for flow measurements in small
tubing, to become familiar with computer data acquisition and to explore cavitation.
Discussion
In engineering practice, it is rarely possible to measure the rate of flow of a fluid by a
direct volumetric or gravimetric determination. Frequently, the metering is accomplished
by the observation of some measurable quantity related to the rate of discharge. Orifices,
nozzles, Pitot and Venturi tubes produce a differential pressure related to the flow
velocity. The differential pressure can be measured with a manometer, pressure gage, or
pressure sensor. Pressure sensors are widely used because they provide a voltage output
that can be monitored easily by computer.
Pressure Sensors
Volumetric
Detector
Pressure
reducing
valve
Shut off
valve
Flow
control
valve
200 kPa
pressure sensor
Orifice
7 kPa pressure
sensor
Spring 2005
Orifice Calibration
The change in piezometric head resulting from the flow of a fluid through an orifice is
obtained by application of the energy equation between a cross section upstream and a
cross section at the vena contracta. Several different locations are commonly used for the
pressure taps and the pressures measured are strongly dependent on the pressure tap
location. Some kinetic energy is converted back to potential energy downstream from the
vena contracta. The pressure tap locations in the laboratory orifice meter are configured
as shown in Figure 2.
The orifice equation contains an empirically determined coefficient, Korifice, that is a
function of the orifice Reynolds number (based on the orifice diameter).
8D
2.5 D
h
D
Figure 2. Location of the pressure taps on the laboratory orifice meter.
Q K orifice Aorifice 2 gh
212\* MERGEFORMAT .
The orifice coefficient is a weak function of Reynolds number.
Scope of Experiment
In this laboratory you will calibrate one orifice plate and determine the cavitation
potential of the orifice.
Procedure
Pressure Sensor Scaling and Monitoring
1)
2)
3)
4)
Spring 2005
Orifice Calibration
1 Connect the 200-kPa pressure sensor to the orifice meter. The high-pressure port
should be connected to the upstream pressure tap (2.5 D upstream from orifice plate).
The bore of the orifice plate is 1.59 mm.
2 Connect the high pressure port of the 7-kPa pressure sensor to the port at the base of
the volumetric detector.
1) Connect the orifice meter to the supply and drain lines.
2) Insert the drain line into the small tube at the top of the volumetric detector
3) Open the flow control valve very slowly! Water hammer from a surge of water
through an air filled orifice tube can damage the pressure sensor.
4) Alternately hold sections of the supply line including the pressure reducing valve
vertically so that air is purged from the line.
5) After the tubing has filled with water and there is a small amount of water in the
volumetric detector turn off the water and then subtract the offset voltage from both
sensors by pressing
. Note that there is a zero function that zeros all sensors
as well as zero functions that zero individual sensors.
6) Turn on data logging to file
Spring 2005
Report
Before leaving the lab complete #2 below and show your results to the TA. Make sure
you will be able to answer all of the questions below!
Write a discussion and conclusion for the results of this experiment, incorporating
answers to the following items. Submit your report electronically as a well formatted
Word document and include your Excel spreadsheet as an appendix.
1) Explain how you obtain the actual flow rate for each run from the output of the
volumetric detector.
2) The 200-kPa pressure sensor measured the difference in pressure (a drop in the HGL)
but the desired term is really a head loss (a drop in the EGL). Why can the pressure
sensor measurements be used directly to obtain the head loss?
3) Create a graph showing the relationship between the orifice coefficient and Reynolds
number.
4) Identify the major sources of error for this experiment and comment on which errors
are potentially most significant.
5) The pressure sensor accuracy is approximately 1% of full scale. What is the flow rate
that corresponds to 1% of the full-scale voltage? How does this affect your accuracy
at a flow rate of 4 mL/s? (Determine the pressure drop that corresponds to 4 mL/s and
then increase the pressure drop by 1% of the full range for the pressure sensor and
calculate a new flow rate.)
6) Calculate the velocity in the vena contracta of the orifice at the highest flow rate
achieved. Estimate the absolute pressure in the small orifice at the highest flow rate
achieved. State your assumptions and show your work. Create a scale drawing of the
HGL and EGL for your system from upstream of the orifice to the outlet. Was the
orifice in danger of cavitation? If the orifice wasnt in danger of cavitation, how could
you explain your laboratory observations? Dont forget the distinction between gage
and absolute pressure!
Lab Setup
Use 100 mV range for the data acquisition ports.
Spring 2005