Professional Documents
Culture Documents
Polarizing Microscope
ECLIPSE 50i POL
Instructions
Preface
Thank you for purchasing this Nikon product.
This instruction manual is intended for users of the Nikons Polarizing Microscope ECLIPSE 50i POL.
To ensure correct use, please read this manual carefully before operating the product.
No part of this manual may be reproduced or transmitted in any form without the prior consent of Nikon.
The contents of this manual are subject to change without notice.
Every effort has been made to ensure the accuracy of this manual. If you find that any portion of this
manual is unclear or incorrect, please contact your nearest Nikon representative.
This manual may describe products not included in the set purchased.
Also be sure to read the manuals for any other products that you are using with this system (the super
high-pressure mercury lamp power supply, high-intensity light source, and such).
Symbol
Meaning
Disregarding instructions accompanying this symbol may lead to serious injury or
WARNING
death.
Disregarding instructions accompanying this symbol may lead to injury or property
CAUTION
damage.
Preface
Warning: If a specimen is spilled onto the microscope body, it may cause the danger of
biohazard.
To avoid biohazard contamination, do not touch the contaminated portion with your bare
hands.
Decontaminate the contaminated portion according to the standard procedure of your facility.
Caution for heat
This marking on the lamphouse of the Eclipse 50i POL calls your attention on the following;
The lamp and its surroundings (including the lamphouse) become very hot during and
immediately after the illumination.
Do not touch the lamp and its surroundings during and immediately after the illumination to
prevent the risk of burns.
Make sure that the lamp and its surroundings are sufficiently cool before the lamp
replacement.
Safety Precautions
Take note of the following points to use this product safety and correctly.
WARNING
1.
2.
Do not disassemble
Disassembling this product may result in electric shock or malfunctions. Damage or injury that may
occur due to mishandling is unwarranted.
Never attempt to disassemble any part other than the parts described in this manual. If you experience
problems with the product, contact your nearest Nikon representative.
3.
4.
5.
6.
Hazardous specimen
This microscope is intended primarily for microscopy of stones, rocks, minerals, high-polymer
materials, and biomedical materials using a polarized light illumination.
Check to determine whether the specimen is hazardous before handling. If the specimen is hazardous,
follow your standard facility procedures. If the specimen is a biomedical material and potentially
infectious, wear rubber gloves and avoid touching specimens. If the specimen is spilled onto this
product, decontaminate the portion according to the standard procedure of your facility.
Safety Precautions
CAUTION
1.
Isolate the products from the power source during assembly, connection/disconnection of cords,
lamp replacement, and maintenance
To prevent electric shock and/or malfunctions, always turn off the power switches of the product (press
to the O position) and unplug the power cord from the wall outlet before assembly, connecting or
disconnecting of cords, lamp replacement, and cleaning of the product and the objective.
2.
3.
4.
5.
6.
Safety Precautions
CAUTION
7.
8.
9.
2.
3.
4.
5.
Installation location
This product is a precision instrument. The usage or storage in an inappropriate environment may result
in malfunctions or poor performance. Consider the following factors when selecting an installation
location:
Select a vibration-free location. Install the product on a level surface.
Install the product at least 10 cm away from walls.
Choose a location less exposed to hazards in the event of collisions, earthquakes, or other potential
disasters. To keep the product from falling, use strong wire or other means if necessary to secure it
to the working desk or to another heavy, stable item.
Avoid locations exposed to direct sunlight, locations immediately under room lights, and other bright
locations.
Select a dust-free location.
To avoid splashes, do not use the product near water.
Make sure the ambient temperature is 0 to 40C and humidity is 85% or less. And, to transport or to
store the product, the ambient temperature must be -20 to 60C and humidity is 90% or less (with
no condensation). Using or storing the product in hot, humid locations may result in mold formation
or condensation on lenses, performance degradation, or malfunction.
Do not install the product in a locker or cabinet.
Select a layout that allows easy removal of the power cord from the product's AC inlet in the event of
an emergency.
Room lights just above the product may enter the objective as extraneous light. If possible, switch
off room lights directly above the product when making observations.
Do not use on a desk mat or the like.
6.
Focusing knobs
Never turn the focus knobs on the left and right sides of the product in opposite directions at the
same time. Doing so may damage the product.
Turning the coarse focus knob past its farthest point will damage the product. Never use excessive
force when turning the knob.
7.
Protect the ports from dust and extraneous light (when the trinocular eyepiece tube is attached).
To keep out extraneous light and dust, always attach the supplied cap to any port not currently in use.
Name of device
Expressions
50i POL
P swing-out condenser
P swing-out condenser
PT polarizer
Dia polarizer
YM-PO polarizer
Epi polarizer
Lamphouse
Camera head
DS-L1
DS camera cable
Camera cable
Contents
Contents
Preface ................................................................................................................ 1
WARNING and CAUTION Symbols ................................................................................... 1
Meaning of symbols used on the equipment ..................................................................... 2
Part Names.................................................................................................. 11
1.1
2.
3.
1.1.2
1.2
1.3
1.3.2
Microscopy .................................................................................................. 16
2.1
2.2
2.3
2.4
2.5
Photomicroscopy............................................................................................... 26
3.2
3.3
3.4
3.1.2
Brightness Adjustment....................................................................................... 30
3.2.1
3.2.2
3.2.3
3.2.4
3.2.5
3.2.6
3.2.7
Camera Adjustment (Adjusting the Brightness of the Image on the Monitor) ..... 32
3.3.2
3.3.3
Contents
3.4.2
3.4.3
3.4.4
3.4.5
3.5
Stage Rotation.................................................................................................. 36
3.5.1
3.5.2
3.6
3.7
3.8
3.9
3.9.2
3.10
3.11
3.12
3.13
3.14
3.15
3.16
3.16.1
3.16.2
3.16.3
3.16.4
3.16.5
3.17
5.
3.17.1
3.17.2
3.17.3
3.17.4
3.18
4.
3.18.1
3.18.2
3.18.3
3.18.4
3.18.5
3.18.6
Assembly..................................................................................................... 18
4.1
4.2
10
Contents
6.
7.
8.
Troubleshooting............................................................................................ 27
6.1
6.2
Mechanical System............................................................................................ 67
6.3
7.2
7.3
7.4
Storage ........................................................................................................... 69
7.5
Specifications............................................................................................... 32
8.1
Specifications ................................................................................................... 70
11
Part Names
1.1
1.1.1
Binocular part
Analyzer slider
P-CL 1/4 & tint plate *1
Centering nosepiece
Vernier
Condenser
(P swing-out condenser)
Dia polarizer
(bottom of the condenser)
Power switch
Orientation plate
*1 It can be changed to the optional P-CS Senarmont compensator or P-CQ quartz wedge.
12
Chapter 1
1.1
1.1.2
Part Names
Eyepiece
Diopter adjustment
i
Condenser aperture
diaphragm ring
*2 They are removed when the optional attachable mechanical stage is used.
13
Chapter 1
1.2
1.2
Part Names
Lamphouse *3
Filter slider
Dummy slider
Aperture diaphragm
open/close lever
Field diaphragm
centering screws (both sides)
Field diaphragm
open/close lever
Polarizer slider or
dummy slider
14
Chapter 1
1.3
Part Names
Rear Side
1.3
Rear Side
1.3.1
Handle
AC inlet
Tool
1.3.2
Lamphouse cover
Halogen lamp
15
Microscopy
2.1
16
Remove the
analyzer from
the optical path.
Chapter 2
2.1
Microscopy
Remove the
Bertrand lens
from the optical
path.
Set the
specimen with
the specimen
holders.
diaphragm with
the aperture
diaphragm
knob.
10
17
Chapter 2
2.1
11
Microscopy
12
13
14
18
Chapter 2
2.1
15
Microscopy
Switch to any
desired objective
Condenser aperture
diaphragm knob
Field diaphragm
control
Focus on the specimen using
the coarse focus knob and the
fine focus knob.
16
19
Chapter 2
2.2
2.2
Microscopy
Orthoscopic Observation
Orthoscopic Observation
This section describes the orthoscopic observation procedure. This is the characteristic
observation method of polarizing microscopes. In this method, the specimen is observed with
the polarizer and the analyzer placed in the optical path.
The shape of the specimen in the direction of the optical axis and its optical properties in the
direction of the thickness can be observed. The vibration direction of the light and the property
of the refraction can be measured with observations of light extinction or interference colors of
specimens and rotations of the stage.
1
2
3
4
10X or higher
IN
4X or lower
OUT
20
Chapter 2
2.3
2.3
Microscopy
Conoscopic Observation
Conoscopic Observation
This section describes the conoscopic observation procedure. This is the characteristic
observation method of polarizing microscopes. In this method, the specimen is observed with
the polarizer, the analyzer, and the Bertrand lens placed in the optical path.
The specimen can be observed from various angles with diascopic light in the form of a single
image. However, the shape of the specimen itself is not visible with this observation. You can
distinguish the property of the specimen between uniaxial and biaxial and observe the optical
axial angle and optical characteristics of the specimen.
1
2
Move the
Bertrand lens
into the optical
path.
Center the Bertrand
lens using the two
screws.
P.48
Focus the
Bertrand lens.
21
Chapter 2
2.4
2.4
Microscopy
22
Remove the
analyzer from the
optical path.
Chapter 2
2.4
Microscopy
Push in the
illumination
selection lever (BF).
23
Chapter 2
2.4
9
10
Microscopy
11
12
24
Chapter 2
2.4
13
Microscopy
Adjust the
brightness
with the
ND filters.
Switch to
any desired
objective.
P.52
Focus on the specimen again using the fine focus
knob or the coarse focus knob.
P.53
Use the ND filters on the epi illuminator to adjust
the brightness.
To use an oil immersion type objective or a water
immersion type objective, perform the oil
immersion or water immersion.
14
15
16
25
Chapter 2
2.5
2.5
Microscopy
Photomicroscopy
Photomicroscopy
For detailed discussions of the camera, photomicroscopic software, and PC, refer to the
operating manuals provided with the respective products. The following instructions assume a
DS camera head DS-5M and DS-L1 camera control unit.
(B)
(A)
4
5
26
Chapter 2
2.5
Microscopy
Photomicroscopy
27
Individual Operations
Item
Title
Operating sections
3.1
Power ON/OFF
Power switch
3.2
Brightness adjustment
3.3
eyepiece tube)
3.4
3.5
Stage rotation
3.6
Diopter adjustment
3.7
Eyepiece sleeve
3.8
3.9
3.10
Selecting a condenser
Condenser
3.11
3.12
3.13
3.14
3.15
3.16
Polarization microscopy
3.17
Episcopic microscopy
Epi illuminator
3.18
Image capturing
Camera
28
Chapter 3
3.1
Individual Operations
Power ON/OFF
3.1
Power ON/OFF
3.1.1
Power switch
Power indicator
3.1.2
29
Power indicator
Power switch
Chapter 3
3.2
Individual Operations
Brightness Adjustment
Brightness Adjustment
3.2
Operating controls
body)
Preset switch, preset brightness
Section
3.2.1
3.2.2
volume-control
Setting ND filters (free from color
3.2.3
3.2.4
3.2.5
3.2.6
3.2.7
temperature shifts)
Episcopic image
temperature shifts)
Extincting episcopic images
(Monitor image)
Camera adjustment
3.2.1
Image brightness
Clockwise rotation
Becomes brighter.
Counterclockwise rotation
Becomes darker.
Brightness
control knob
30
Chapter 3
3.2
3.2.2
Individual Operations
Brightness Adjustment
Preset switch
Preset brightness
volume-control
3.2.3
3.2.4
Image brightness
Rotate clockwise
Rotate counterclockwise
31
Chapter 3
3.2
3.2.5
Individual Operations
Brightness Adjustment
ND4
ND16
Out
Out
1/4
In
Out
1/16
Out
In
1/64
In
In
3.2.6
3.2.7
Camera Adjustment
(Adjusting the Brightness of the Image on the Monitor)
When observing images captured by the camera and displayed on the monitor, you can adjust
brightness by varying camera adjustment parameters, such as display mode, exposure mode,
metering mode, exposure compensation, and image level adjustment.
For detailed information, refer to the operating manual provided with the camera or camera
control software.
32
Chapter 3
3.3
Individual Operations
3.3
3.3.1
Light Distribution
With the trinocular eyepiece tube, the optical path
selection lever allows the light distribution to the
Optical path
selection lever
Optical path
selection lever
position
3.3.2
Vertical tube
section
section
Pushed in
100
20
80
100
TV vertical tube
adapter
3.3.3
33
Chapter 3
3.4
Individual Operations
3.4
3.4.1
Prohibited Actions
Avoid the following actions, which can cause equipment malfunctions. Never do them at all.
3.4.2
Rotating the right and left coarse/fine focus knobs in opposite directions.
3.4.3
Approx. 13.8 mm
Approx. 0.1 mm
1 m
The vertical motion range (coarse/fine focus stroke) of the stage is from 2 mm above the focal
point (reference position) to approximately 28 mm below the focal point.
34
Chapter 3
3.4
3.4.4
Individual Operations
3.4.5
direction of arrow
turn
turn
Coarse focus
stopper ring
Example usage
With the sample in focus, turn the coarse focus
stopper ring as far as it goes in the direction of the
arrow (labeled CLAMP). When changing the
sample, lower the stage by turning only the coarse
focus knob. After changing the sample, gently raise
the stage by turning only the coarse focus knob as
far as it goes. The sample should be roughly in
focus.
Rotate the ring in the direction of
35
Chapter 3
3.5
3.5
Stage Rotation
3.5.1
Stage Rotation
Individual Operations
Stage Rotation
Stage rotation
clamp screw
3.5.2
Clamp screw
36
Positioning pins
Chapter 3
3.6
3.6
Individual Operations
Diopter Adjustment
Diopter Adjustment
Diopter adjustment compensates for differences in visual acuity between the right and left eyes,
improving binocular observation. It also minimizes focal deviations when switching objectives.
In the case of a polarizing microscope, since an eyepiece containing crosshairs is used for the
right eye, the procedure for adjusting the diopter differs from that of an ordinary microscope.
(1)
(2)
(3)
3.7
37
View field
Converge until
the right and left
view fields
coincide.
Chapter 3
3.8
3.8
Individual Operations
(2)
(3)
(4)
(5)
(6)
(7)
Condenser
focus knob
Condenser
centering screw
Eyepiece viewfield
Eyepiece viewfield
38
Chapter 3
3.9
Individual Operations
3.9
The aperture diaphragm for the diascopic microscopy is adjusted with the condenser aperture
diaphragm knob.
For the diaphragms of the epi illuminator, refer to 3.17 Episcopic microscopy.
3.9.1
Objective
40X/ 0.65 P
/0.17 WD 0.65
40x
40X/0.65 P
/0.17 WD 0.65
Condenser
aperture
diaphragm
knob
Adjusting the size of the aperture diaphragm according to the condenser scale
Since the condenser scale indicates the numerical aperture, adjust the aperture diaphragm ring
according to the scale. (Normally, the index on the aperture diaphragm ring should be aligned
with the scale line corresponding to 70% to 80% of the numerical aperture of the objective.)
Adjusting the size of the aperture diaphragm using the Bertrand lens
Insert the Bertrand lens into the optical path (by placing in position B). Turn the diaphragm
control ring to stop down the aperture diaphragm to its minimum setting. Turn the Bertrand lens
focus ring to focus on the aperture diaphragm image. Turn the diaphragm control ring to adjust
the aperture diaphragm. (This is normally adjusted to 70-80% of the view field.)
39
Chapter 3
3.10
3.9.2
Individual Operations
Selecting a Condenser
3.10
Selecting a Condenser
To perform the polarization microscopy, the P swing-out condenser must be used.
magnification
Objective
P swing-out condenser
(: suitable, -: not suitable)
1x
2x
Note 1
4x
10x to 100x
Note 1:
Depending on the type of objective, the indicated numerical aperture of the objective may not be
achieved.
For example, when an objective with an N.A. of 1.4 is used, the maximum aperture of the P
swing-out condenser will be only about 65% of the objective's N.A., even when the condenser
aperture diaphragm is fully open.
Top lens
40
Swing-out knob
Chapter 3
3.11
3.11
Individual Operations
3.12
Application
ND 16 (transmittance: 6%)
GIF
41
Chapter 3
3.13
3.13
Individual Operations
(2)
(3)
(4)
(5)
(6)
180
rotation
Target
42
Middle point
of the
movement
Chapter 3
3.14
Individual Operations
3.14
Turn the nosepiece slightly to move the oil-immersed objective back and forth once or
twice. (In the case of the condenser, gently turn the condenser focus knob to move the
condenser up and down slightly.)
Use as little oil as possible (just enough to fill the space between the tip of the objective and the
specimen, or between the tip of the condenser and the specimen). If too much oil is applied, the
excess oil will flow onto the stage or around the condenser.
CAUTION
When using petroleum benzine or absolute alcohol, always follow the
instructions provided by the manufacturer. These liquids are highly
flammable and must be kept away from flames and sparks.
43
Chapter 3
3.15
3.15
Individual Operations
44
Chapter 3
3.16
Individual Operations
Polarization Microscopy
3.16
Polarization Microscopy
3.16.1
Dia polarizer
Epi
polarizer
Lateral
direction
Vertical
direction
Polarizer orientation
(Polarizer rotating ring)
45
Chapter 3
3.16
3.16.2
Individual Operations
Polarization Microscopy
Analyzer
slider
Slider
Analyzer
Pulled out
IN
Push in
OUT
0
140 13
12
100 90
11
vernier scales.
150
OUT
IN
20
16
30
40
50
0.
60
70
Analyzer
rotating dial
10
10
0
17
Vernier
Analyzer rotation
clamp screw
The intermediate tube also has a de-polarizer. So, you can use the
photomicrography devices independently of the orientation of the polarizer.
46
Chapter 3
3.16
3.16.3
Individual Operations
Polarization Microscopy
(2)
(3)
Dia polarizer
(7)
80
0
140 13
15 0
40
0.
OUT
IN
50
30
60
70
0
12
(6)
16
0
100 90
(5)
Orientation
plate
0
11
(4)
20
10
10
17
Analyzer
scale: 0
Dark cross
It should be noted that the X direction is explained as that of the analyzer and Y
direction as that of the polarizer in some commercially available technical
manuals and reference books.
(1)
Push in the analyzer setting knob to move the analyzer out of the optical path.
(2)
Place a dummy specimen. It must have high reflectance with optical isotropy, for
example a mirror. And then, focus on the specimen.
(3)
Pull out the analyzer setting knob to move the analyzer into the optical path.
(4)
Turn the analyzer rotation dial and align at the 0 position on the scale.
(5)
Push in the polarizer slider on the epi illuminator to place the epi polarizer into the
optical path.
(6)
Move the Bertrand lens into the optical path. Turn the polarizer rotating ring on the
epi polarizer and adjust so that the dark cross image appears in the pupil as shown in
the figure.
47
Chapter 3
3.16
3.16.4
Individual Operations
Polarization Microscopy
(2)
(3)
Bertrand lens
centering screws
(5)
(6)
48
Bertrand lens
focus ring
Chapter 3
3.16
Individual Operations
Polarization Microscopy
3.16.5
The intermediate tube has a slot for the P-CL 1/4 & tint plate. It is used not only for the plate
but also for the optional P-CS Senarmont compensator or P-CQ quartz wedge to perform the
retardation measurement.
To use an objective of 10X or higher magnification in standard observation, place the top
lens of the P swing-out condenser into the optical path or pull out the slider of the slide
condenser.
To perform the retardation measurement or to perform evaluation by interference color, the
illumination light must be as parallel to the optical axis as possible. So, the condenser
aperture diaphragm must be stopped down or the top lens of the P swing-out condenser
must be swung out (the aperture diaphragm is fully opened) even if an objective of 10X or
higher is used.
Measurement
Place the GIF filter on the field lens and replace the P-CL 1/4 & tint plate with the P-CS
Snarmont compensator. Rotate the analyzer so that the section of the specimen to be
measured becomes darkest. When the rotation angle of the analyzer at that time is taken to
be theta () degrees, then retardation (R) (nm) is determined with the following formula:
R=
: wavelength
180
The value of when using the GIF filter is 546 nm.
49
Chapter 3
3.16
Individual Operations
Polarization Microscopy
Measurement
Move the section of the specimen to be measured
to the center of the crosshairs of the eyepiece.
Next, slide the P-CQ quartz wedge along the slot
and observe that the interference color sequentially
changes. Stop sliding the quartz wedge where the
dark stripe covers the section of the specimen to
be measured. Reading the value from the quartz
wedge scale at that time can make a rough
measurement of retardation.Retardation can be
measured even more accurately by using the P-CS
Snarmont compensator in combination with the
P-CQ quartz wedge.
50
Chapter 3
3.17
3.17
Individual Operations
Episcopic Microscopy
Episcopic Microscopy
To perform episcopic microscopy, attach the LV-UEPI universal epi illuminator to the microscope.
3.17.1
illumination.
Lever position
3.17.2
Illumination method
Push in (BF)
Field diaphragm
open/close lever
Field diaphragm
centering screw
(both side)
Field diaphragm
image
51
Chapter 3
3.17
Individual Operations
Episcopic Microscopy
Perform steps in 2.4 Episcopic Microscopy (with the Epi Illuminator Option) and focus on
the specimen using the 10X objective under the episcopic illumination.
(2)
(3)
Lower the field diaphragm open/close lever to stop down the field diaphragm.
Turn the field diaphragm centering screws (on both sides) so that the field diaphragm
image is positioned in the center of the view field.
(4)
Adjust the field diaphragm image with the field diaphragm open/close lever and centering
screws so that it inscribes the view field.
(5)
To observe the specimen, raise the field diaphragm open/close lever so that the field
diaphragm image circumscribes the view field.
Field diaphragm image
3.17.3
70~80
100
52
Chapter 3
3.17
3.17.4
Individual Operations
Episcopic Microscopy
Filter sliders
Filters
Usage
Contrast adjustment
IF (interference filter)
For interference
53
Chapter 3
3.18
Individual Operations
Image Capturing
Image Capturing
3.18
Microscopy images can be captured by attaching a camera head to the trinocular eyepiece tube.
For detailed information, refer to the operating manual provided with the camera head or
camera control software.
Proper adjustment of light intensity and focus on the microscope are important for obtaining
clear images. Listed below are key considerations in capturing clear images.
3.18.1
Lamp voltage:
Filter:
3.18.2
For normal operations, set the diaphragm aperture to 70 to 80% of the N.A. of the
objective.
3.18.3
3.18.4
54
Chapter 3
3.18
3.18.5
Individual Operations
Image Capturing
Stop down the diaphragm to a setting just slightly wider than the
area shown on the monitor.
Eyepiece:
3.18.6
Anti-vibration Measures
If the exposure is less than 1/8 of a second, reduce light intensity with ND filters to make
exposures longer than 1/8 of a second. (If accurate color reproduction is not important, you can
use the brightness control knob to reduce light intensity.)
55
Assembly
System Configuration
4.1
TV vertical
tube adapter
TV vertical
tube adapter
0.55X
or
Eyepieces
Eyepieces
Binocular
eyepiece tube
Trinocular
eyepiece tube
P-CL 1/4 & tint plate
P-CS Senarmont compensator
Quintuple
centering
nosepiece
DIN slot
Filters
Lamphouse
Objectives
Attachable
mechanical stage
Epi illuminator
12V 50W
halogen lamp
Specimen
holders (x2)
Lamp cable
40
50
60
70
80
0
10
20
30
or
6V 30W
halogen lamp
POWER
MIN.
MAX.
Circular
graduated stage
0
Power cord
10
90
80
20
70
30
0.4
JAPAN
0.2
60
0.90
0.6
50
0.8
40
P swing-out
condenser
Lamp power
supply
Dia polarizer
56
Lamp cover
Power cord
Chapter 4
4.2
Assembly
Assembly Procedure
Checking the input voltage
Make sure that the input voltage indicated on the back
panel of the microscope is the same as the voltage
provided in your area.
CAUTION
Input
voltage
indication
Specimen holder
Specimen holder
(3)
(4)
Clamp screw
for the stage
(2)
(3)
(4)
57
Chapter 4
Assembly
(2)
Attaching objectives
Lower the stage completely. Screw objectives into the nosepiece so that the magnification
increases with the clockwise rotation (as viewed from above the microscope) of the nosepiece.
(2)
(3)
(5)
58
Chapter 4
Assembly
(1)
lamphouse
screwdriver.
(2)
(3)
(4)
59
To power
supply
Chapter 4
Assembly
Loosen sufficiently the clamp screw for the intermediate tube (or the illuminator) on the
microscope arm.
Use the tool provided with the microscope to loosen the clamp screw for the epi illuminator.
Plarizing intermediate
tube
Remove any looseness between the positioning pin and groove by pushing in the
intermediate tube while rotating in the clockwise direction.
(3)
10
Protrusion
Notch
60
Chapter 4
11
Assembly
Mount the TV vertical tube adapter or the TV vertical tube adapter 0.55X onto the vertical
tube section of the trinocular eyepiece tube.
(2)
Attach the suitable adapter for the camera onto the tip of the TV vertical tube adapter.
(3)
(4)
12
(2)
Insert the power cord into the AC inlet at the back of the microscope.
(3)
Insert the other end of the power cord into a wall outlet.
AC inlet
Power switch for
the microscope
The microscope assembly is now completed.
61
5
5.1
Replacing Consumables
Replacing the lamp
CAUTION
Be careful to avoid burns:
Wait until the lamp and nearby parts have cooled before attempting to
replace the lamp.
Be careful to avoid electrical shock:
Turn off the power switch and unplug the power cord from the outlet.
Be careful to avoid abnormal heat generation:
Use only the lamp specified.
Be careful to avoid soiling:
Avoid touching the bare lamp bulb with bare hands. Soiling will reduce the
service life of the lamp.
(2)
Cover hook
(3)
Lamp
Lamphouse cover
CAUTION
Be sure to attach the lamphouse cover. Failure to do so may result in burns or
fire from the heat generated by the lamp.
62
Chapter 5
5.1
Replacing Consumables
Replacing the lamp
CAUTION
To prevent electrical shock and damage to the microscope, always turn off
the power switch (flip it to the O side) and unplug the power cord from
the outlet before attaching or detaching the lamphouse.
To prevent burn injury, allow the lamp and the lamphouse to cool down
sufficiently (for at least 30 minutes after the lamp is turned off), before
replacing the lamp.
Use the Nikon LV-HL50W 12V 50W LONGLIFE halogen lamp or non-Nikon
12V 50W SHORTLIFE halogen lamp (model OSRAM HLX 64610, OSRAM HLX
64611, or PHILIPS 7027) for the lamp. If you wish to buy these lamps,
please contact your nearest Nikon representative.
Do not touch the glass surface of the lamp with bare hands. Fingerprints or
grease on the bulb surface will reduce the illumination intensity of the
lamp. Wipe clean any fingerprints or grease attached to the surface.
Securely attach the lamphouse cover to the lamphouse after replacing the
lamp. Never light the lamp with the lamphouse cover removed.
When you dispose of the replaced lamp, do not break it up. Instead,
dispose of the used lamp as special industrial waste or dispose of it
according to the local regulations and rules.
63
Chapter 5
5.1
Replacing Consumables
Replacing the lamp
(1)
Loosen the lamphouse cover clamp screw using the hexagonal wrench.
(2)
(3)
Push down the lamp clamp lever and remove the old lamp.
(4)
With the lamp clamp lever held down, insert the electrodes of a new lamp into the pin
holes of the socket. Press the lamp as far as it goes, and then release the lamp clamp
lever to secure the lamp.
Be careful not to touch the glass surface with bare hands.
When releasing the lamp clamp lever, use care so that the lamp does not tilt.
(5)
Close the lamphouse cover and secure it by tightening the clamp screw.
CAUTION
Be sure to attach the lamphouse cover. Failure to do so may result in burns or
fire from the heat generated by the lamp.
64
Troubleshooting
When the product does not function properly, take appropriate action as described below. If the
problem is still not resolved after referring to "Troubleshooting," please contact your nearest Nikon
representative.
6.1
Optical System
Problem
Cause
Countermeasure
65
Chapter 6 Troubleshooting
6.1 Optical System
Problem
The brightness is
insufficient.
Cause
Countermeasure
66
Chapter 6 Troubleshooting
6.2 Mechanical System
6.2
Mechanical System
Problem
6.3
Possible causes
Remedy
Electrical System
Problem
Possible causes
Connect it properly.
Connect it properly.
67
Remedy
7
7.1
Either brush away dust with a soft brush, or else gently wipe it off with a piece of gauze.
Only if there are fingerprints or grease on a lens, dampen lightly a piece of soft, clean
cotton cloth, lens tissue, or gauze with absolute alcohol (ethyl or methyl) and gently wipe
off the dirt.
When removing the immersion oil off from the objective, use only petroleum benzine. For
optimum results, we recommend following up petroleum benzine with absolute alcohol
(ethyl or methyl alcohol). If petroleum benzine is unavailable, use methyl alcohol alone.
However, methyl alcohol does not clean as well as petroleum benzine, it will be necessary
to wipe the surface repeatedly. (Usually, three or four times are sufficient to clean the
lenses.)
Never use petroleum benzine to clean the entrance lens of the eyepiece tube or prism
surface of the eyepiece tube.
Absolute alcohol and petroleum benzine are highly flammable. Be careful when handling
it, when around open flames, when turning the power switch on/off, etc.
When using petroleum benzine or absolute alcohol, always follow the instructions
provided by the manufacturer.
7.2
7.3
We recommend that you use a piece of silicon cloth to clean this product.
For persistent dirt, dampen a piece of gauze with neutral detergent and wipe gently.
We recommend that you use 70% medical alcohol for normal disinfection of this product.
In case of spillage of a specimen onto this product, determine whether the specimen is
hazardous. If the specimen is hazardous, follow your standard facility procedures.
68
Chapter 7
7.4
Storage
Storage
Store this product in a dry place where mold is not likely to form.
Store the objectives and eyepieces in a desiccator or similar container with a drying
agent.
Put the dust-proof cover over this product to protect it from dust.
Before putting on the dust-proof cover, turn off the power switch of the microscope (flip it
to the O position) and wait until the lamphouse gets cool sufficiently
7.5
69
Specifications
8.1
Specifications
Optical system
CFI60
Field number 22 (for binocular eyepiece tube)
Field number 25 (the trinocular eyepiece tube)
Focusing mechanism
Nosepiece:
Driving type:
Stroke:
Stage
Illumination
100 hours
Input ratings
Power cord
Operating conditions
Temperature:
0C to +40C
Humidity:
Altitude:
2000 m max.
Pollution degree:
Degree 2
Installation category:
Category II
Class I
Temperature:
-20C to +60C
Humidity:
70
Chapter 8
8.1
Specifications
Specifications
External dimensions:
Safety standard
Approx. 11 kg
71