You are on page 1of 6

SERIAL #

CONTROL #

Testing Bipolar Output - 100


POWER
10 W

30 W

95 W

MODE
Low
Standard
Macro
Low
Standard
Macro
Low
Standard
Macro

Output current range (in mA)


291 to 339
291 to 339
291 to 339
504 to 587
504 to 587
504 to 587
900 to 1044
900 to 1044
900 to 1044

Testing Cut Output - 300 (2 button pencil)


POWER

SYS. OUTPUT
Pure

Output range (in mA)


146 to 218

Blend
Pure

146 to 218
461 to 536

Blend
Pure
Blend

461 to 536
922 to 1072
753 to 876

10 W

75 W
300 W
300 W

Testing Cut Output - 300 (3 button pencil)


POWER

SYS. OUTPUT
Pure

Output range (in mA)


146 to 218

Blend
Pure

146 to 218
461 to 536

Blend

461 to 536

Blend

753 to 876

10 W

75 W

200 W

SERIAL #

CONTROL #

Testing Coag Output - 500 (2 button pencil)


POWER

SYS. OUTPUT
Fulgurate

Output range (in mA)


20 to 100

Spray
Fulgurate

20 to 100
226 to 263

Spray
Fulgurate

226 to 263
452 to 525

Spray

452 to 525

1W

30 W

120 W

Testing LigaSure Output - 50


Debug / Ligasure Test

Ligasure 1

SYS. OUTPUT
1 amp
2.7 amps
5.5 amps

Output range (in mA)


288 to 355
761 to 885
1540 to 1790

Bipolar High-Frequency Leakage Current 200 - Debug Mode


Mode- Debug / Bipolar

Activate Bipolar at
maximum setting

SYS. OUTPUT
Low
Standard
Macro

Max. Output range (in mA)


69 mA
69 mA
69 mA

LigaSure High-Frequency Leakage Current 200 - Debug Mode


Mode- Deb / LigaTest

Activate LigaSure at
maximum output

SYS. OUTPUT
Low
Standard
Macro

Max. Output range (in mA)


130 mA
130 mA
130 mA

SERIAL #

CONTROL #

Patient High-Frequency Leakage Current 200 - Debug Mode


Mode-Debug / Monocut

Activate LigaSure at
maximum output

SYS. OUTPUT
Cut
Blend
HDW / Valley Lab
Fulgurate
Spray

Max. Output range (in mA)


100 mA
100 mA
100 mA
100 mA
100 mA

Tech:
Date:

Actual current output

n pencil)
Actual output

n pencil)
Actual output

Tech:
Date:

on pencil)
Actual output

Actual output

0 - Debug Mode
Actual output

00 - Debug Mode
Actual output

Tech:
Date:

0 - Debug Mode
Actual output