FOFEN FGA-4100 BASIC TROUBLE SHOOTING

ERROR

POSSIBLE REASON

Turn on the power switch , display full It


1 is not lit, the instrument without any
1. The power plug connection is bad
reaction .
2. Fuse blown

SOLUTION
1. Check the switch Check the power
plug

2. Replace the fuse


3. Check the power supply output is
3. The switching power supply for protection
short-circuited

HC, CO, CO2 large fluctuations in value


Gas analysis platform issues
.
Sample flow is low, slow response . Gas
1. sampling tube or probe blockage
line Jam symbol appears .
2. The filter clogging

4 Oxygen show a dramatic swing .

5 Low measurement results

1. The oxygen sensor connector poor


contact .
2. The oxygen sensor failure . .
1. The sampling system or pipeline leak
probe Insertion depth is too shallow .
2. Calibration is incorrect.

HC display value instability , HC display Filter and piping system has been
value
contaminated
We can not return to zero , HC drift.

Back to the factory maintenance

1. Check the hose and sampling probe


2. Replace the filter element
1. Check the connections.
2. Replace the oxygen sensor .
1. Check and remove leaks.
2. recalibration .

1. Replace the pre-filter , dust filter


Element .
2. Clean or replace the contaminated
tubes
3. The sampling tube with compressed
air cleaning
4. Clean the air chamber .

Measurements are performed


continuously repeatedly forced drain.

1. Inside car exhaust pipe water.


2. The proximity switch detects the head
attachment moisture.
3. proximity switch sensitivity is too high .
4. The proximity switch is damaged .

1. Adjust the placement of the sample


tube.
2. Wipe the front of the water close to
the switch .
3. Adjust the sensitivity of the proximity
switch .
4. Replace proximity switch and adjust
the sensitivity
Degree , if not replace proximity switch
to play
Unplug the drive to open the instrument
cover plate
J14 plug can also be temporarily normal
measurement ( please do Fast Replace
proximity switch ) .

In addition to the water level inside the


water heater in addition to increased
8
Metal water height within the filter , still
Undrained

1. The proximity switch sensitivity is too low.

2. The drain line blockage .


3. The proximity switch is damaged .

9 Leak check failure

1. The internal air passage , trachea.


2. dust filter.
3. The pump assembly .

1. Adjust the sensitivity of the proximity


switch .
2. Detection drain line .
3. Replace proximity switch and adjust
the sensitivity degree.
1. Check the internal pipe is loose or
burst.
2. Detection of dust filter holder or cover
is No burst.
3. Detection of the pump or whether
aging occurs crack.