1 is not lit, the instrument without any 1. The power plug connection is bad reaction . 2. Fuse blown
SOLUTION 1. Check the switch Check the power plug
2. Replace the fuse
3. Check the power supply output is 3. The switching power supply for protection short-circuited
HC, CO, CO2 large fluctuations in value
Gas analysis platform issues . Sample flow is low, slow response . Gas 1. sampling tube or probe blockage line Jam symbol appears . 2. The filter clogging
4 Oxygen show a dramatic swing .
5 Low measurement results
1. The oxygen sensor connector poor
contact . 2. The oxygen sensor failure . . 1. The sampling system or pipeline leak probe Insertion depth is too shallow . 2. Calibration is incorrect.
HC display value instability , HC display Filter and piping system has been value contaminated We can not return to zero , HC drift.
Back to the factory maintenance
1. Check the hose and sampling probe
2. Replace the filter element 1. Check the connections. 2. Replace the oxygen sensor . 1. Check and remove leaks. 2. recalibration .
1. Replace the pre-filter , dust filter
Element . 2. Clean or replace the contaminated tubes 3. The sampling tube with compressed air cleaning 4. Clean the air chamber .
Measurements are performed
continuously repeatedly forced drain.
1. Inside car exhaust pipe water.
2. The proximity switch detects the head attachment moisture. 3. proximity switch sensitivity is too high . 4. The proximity switch is damaged .
1. Adjust the placement of the sample
tube. 2. Wipe the front of the water close to the switch . 3. Adjust the sensitivity of the proximity switch . 4. Replace proximity switch and adjust the sensitivity Degree , if not replace proximity switch to play Unplug the drive to open the instrument cover plate J14 plug can also be temporarily normal measurement ( please do Fast Replace proximity switch ) .
In addition to the water level inside the
water heater in addition to increased 8 Metal water height within the filter , still Undrained
1. The proximity switch sensitivity is too low.
2. The drain line blockage .
3. The proximity switch is damaged .
9 Leak check failure
1. The internal air passage , trachea.
2. dust filter. 3. The pump assembly .
1. Adjust the sensitivity of the proximity
switch . 2. Detection drain line . 3. Replace proximity switch and adjust the sensitivity degree. 1. Check the internal pipe is loose or burst. 2. Detection of dust filter holder or cover is No burst. 3. Detection of the pump or whether aging occurs crack.