You are on page 1of 6

Technical articles

Sample preparation for X-ray fluorescence analysis


IV.Fusion bead methodpart 1 basic principals

Mitsuru Watanabe

1.Introduction beads, various applications such as for ferroalloy, silicon


For the analysis of powders by XRF, sample carbide and copper concentrate samples will be reported.
inhomogeneity due to segregation, grain size and
mineralogical effects inuence X-ray intensity and can 2.Preparation method
cause analysis errors. It is therefore recommended Figure 2 shows the general procedure for preparing a
to analyze powder samples after ne pulverizing fusion bead.
as described in Sample Preparation for X-ray
Fluorescence Analysis II. Pulverizing methods of 2.1.Weighing
powder samples(1). However, when inhomogeneity Dry powder sample, ux, oxidizing agent are weighed
can not be sufciently removed by pulverization and to 0.1 mg precision.
more accurate analysis is required, fusion bead (Fig. 1)
method is advisable. 2.2.Mixing
The fusion bead method rst established in the Transfer weighed sample and ux into a crucible.
1950s(2), has since progressed such that it is not only Be sure that the specimen is thoroughly mixed prior to
applicable to powders as oxides but also to non-oxides fusion and add releasing agent if fusion bead is difcult
such as metals, carbides, suldes which had previously to remove from the mold. Add oxidizing agent if
been considered to be difcult(3),(4). Characteristics of oxidation of sample is required.
the fusion bead method are that (a) analysis error
due to grain size and mineralogical effects can be 2.3.Oxidization/Calcination
removed, (b) matrix effect is reduced due to dilution, (c) Samples containing metals, carbides, suldes are
standard samples can be prepared by mixing of reagents. mixed with oxidizing agent and oxidized at around
Characteristics of fusion bead and pressed powder 600800C. In cement raw meal and limestone, CaO
methods are compared in Table 1. in sample can actually be present as CaCO3. For such
In this article, general preparation methods, equipment, samples, sudden increase in temperature can cause
reagents and other important considerations for powders signicant foaming due to release of CO2 caused by
with typical grain size and drying conditions are the chemical reaction CaCO3CaO+CO2. Calcination
described. If analysis must be performed in accordance at around 800C before fusion reduces risk of sample
to a specic standard test method, adhere to its overow and air being trapped in the nished fusion
prescribed conditions. In a following issue on fusion bead.

Fig. 1. Fusion beads.

SBU WDX, X-ray Instrument Division, Rigaku Corporation.

Rigaku Journal, 31(2), 2015 12


Sample preparation for X-ray fluorescence analysis IV. Fusion bead methodpart 1 basic principals

Table 1. Comparison between fusion bead method and pressed powder method.

Fusion bead method Pressed powder method

Grain size effect Not affected Affected

Mineralogical effect Not affected Affected

Matrix effect Reduced due to dilution Affected by elemental composition

Standard sample Possible to prepare by mixing reagents Need to be similar to test sample

Grain size of sample 106 m (140 mesh) or less 46 m (330 mesh) or less

X-ray intensity Reduced due to dilution No change

Storage In desiccator to avoid deliquescence In desiccator

Handling Easy since in glass form Care required not to break

Preparation time 1530 min. including weighing Several minutes

Fig. 2. General fusion bead preparation method.

2.4.Fusion
Samples are fused at 10001200C for a certain
period of time. Keep the crucible stationary until the
sample and the ux melt, and then begin to swing to
homogenize and remove air bubbles. The higher the
temperature applied, the higher the uidity during
fusion, but can increase volatilization of ux and
analytes, and decrease crucible lifetime. By keeping
fusing time xed, ux volatilization is consistent
and therefore its effect can be ignored. However, to
minimize variation in sample volatilization due to
differences in sample composition, it is advisable to fuse
at lower temperatures.
When the fusion is completed, stop swinging and
keep the crucible horizontal for fusion/casting combined
vessels, or pour the fused sample into a mold. Fig. 3. Sample cool down timeline.

2.5.Cooling sample, ux type, dilution ratio. Figure 3 illustrates


It is important to cool down the fused sample while sample cool down timeline.
maintaining its property as a glass. Rapid cooling
can cause the sample to crack, while cooling too 3.Instruments
slowly can cause crystallization. Because of this, initial 3.1.Fusion types
passive slow cooling followed by active cooling is There are three types of fusion machines for preparing
recommended. Cooling time will vary according to fusion beads, namely high frequency (HF) induction,

Rigaku Journal, 31(2), 2015 13


Sample preparation for X-ray fluorescence analysis IV. Fusion bead methodpart 1 basic principals

electric furnace and a gas burner types. Fusion (b) Crucible/mold separate type
temperature range slightly varies depending on type Sample is fused in the crucible and then transferred
but is typically around 10001250C and have a swing/ into a mold to form the fusion bead. Preheating the mold
cooling mechanism. Room ventilation is recommended to around 800C before pouring into the mold prevents
especially when many fusion bead samples are prepared cracking of the fusion bead.
with oxidizers or releasing agent.
Most vessels are made of platinum and gold (95 and
(a) HF induction type 5 mass%) to assure release of the formed fusion bead. To
Due to direct heating by high frequency coil, heating improve hardness and prevent deformation, some have
is efcient and rapid(5). added rhodium and consist of reinforced platinum or
(b) Electric furnace type platinum alloy. Whatever the material, the atness of the
There are mufe and non-mufe type electrical molds surface in contact with the sample is critical for
furnaces. It is necessary to turn power on before use and sample detachment and should therefore also be mirror
time is required to reach high enough temperatures. polished.
(c) Gas burner type
Temperature is controlled by adjusting gas ow 4.Analysis sample and reagents
rate. Since combustible gases such as propane is used, 4.1.Analysis sample
environment needs to be carefully monitored. Analysis sample is pulverized to smaller than 106 m
particle size (140 mesh), dried for over two hours in
3.2.Vessels (Crucible, Mold) 1105C air atmosphere, then cooled down and stored
Regarding use of vessels, there are two different in a desiccator.
approaches for forming the fusion bead namely crucible/
mold combined and separate types. The fusion instrument 4.2.Flux
mentioned above uses combined type vessels (Fig. 4). Flux is available as ne powders or in granular and
globular forms, and some already contain releasing
(a) Crucible/mold combined agents. In any case, it should not contain any water
Sample is fused and cooled down in the crucible, and depending on analysis requirement, the appropriate
where the crucible is also used as a mold. purity grade should be chosen.
Impurity levels can vary from lot to lot, so it is
desirable to have a substantial amount of the same lot to
minimize analysis error. Flux is a hygroscopic substance
and therefore it should be dried before use. Place the
ux in a platinum tray and heat in an electric furnace
at 200250C below its melting point for 4 hours, and
then cool down and store in a desiccator (for example,
lithium tetraborate is dried at 650700C for 4 hours).
Fusing temperature is typically 200250C higher than
its melting point to improve uidity. Table 2 shows the
physical properties of various types of ux(6),(7).

(a) Lithium tetraborate Li2B4O7=Li2O2B2O3


Lithium tetraborate is the most commonly used ux
Fig. 4. Crucible/mold in one type.

Table 2. Physical properties of various uxes.

Lithium tetraborate Lithium metaborate Mixed ux Sodium tetraborate

Formula Li2B4O7 LiBO2 Li2B4O7, LiBO2 Na2B4O7

Melting 930C 845C 875C (66 : 34) 741C


point 870C (50 : 50)
825C (12 : 22)
840C (20 : 80)

Acid/Base Acidic Basic Weak acidWeak base Acidic

Suitable Lime stone Silicate, Rock Silicate, Rock Metal, Mineral ore
sample Cement Refractory Refractory

Remarks High melting point Crystalizes when cooled Low melting point Lowest melting point
High solubility for High solubility for High deliquescence
basic oxides acidic oxides Na analysis not possible

Rigaku Journal, 31(2), 2015 14


Sample preparation for X-ray fluorescence analysis IV. Fusion bead methodpart 1 basic principals

in XRF analysis. Since it is a relatively more acidic bubbles from the sample due to lower viscosity. Typical
compound compared to lithium metaborate (LiBO2), iodides are LiI, NaI, KI and NH4I, and bromides are
it is suitable for fusing samples such as lime stone and LiBr, NaBr and KBr.
cement which contain basic oxides (CaO, MgO, Na2O, Bromides compared to iodides act as stronger
K2O, etc.). However, its 930C melting point is the releasing agents and therefore can be effective in
highest among the various ux types and consequently smaller quantities, but tend to remain in the fusion bead.
fusion temperature must be relatively high. Therefore, Halogens remaining in the sample can interference
this ux requires special attention to sample and ux analysis lines. Table 3 shows analysis and interference
volatilization as well as damage to the vessel. lines due to the releasing agent.
(b) Lithium metaborate LiBO2=Li2OB2O3 Releasing agent can be added before heating or during
Lithium metaborate is a relatively more basic fusion. Consistent addition of releasing agent quantity
compound compared with lithium tetraborate, and is can be difcult since typical amounts are less than 1 mg.
therefore suitable for fusing samples such as silicates, By preparing 550% (w/v) solutions in advance and
rocks and refractories which contain acidic oxides (SiO2, addition by a micropipette can improve consistency. The
ZrO2, TiO2, etc.). It has a low melting point of 845C iodide solution should be kept in a light-resistant bottle
and tends to cause crystallization instead of vitrication as many iodide materials tend to liberate iodine due to
when cooled. It is therefore not used by itself but rather air oxidation or light. Moreover, as described below,
in combination with lithium tetraborate as a mixed ux. lithium uoride in addition to acting as a releasing agent
(c) Mixed ux (Lithium tetraborate Li2B4O7, Lithium also lowers the viscosity during fusion.
metaborate LiBO2) When releasing agent quantity is insufcient, the
As this is a mixture of lithium tetraborate and lithium samples surface area in contact with molds internal
metaborate, its acidity/basicity can be adjusted by the wall is relatively large and become difcult to remove.
mixing ratio. It also has the advantage of lowering the On the other hand, when too much is added, surface
melting point for easier fusion compared to pure lithium tension becomes too high such that sample does not
tetraborate. fully cover the bottom of the vessel resulting in a
Lithium tetraborate to metaborate ratios are typically crescent or ball shaped fusion bead. Effect of releasing
66% to 34%, 50% to 50%, 35.3% to 64.7% (12 : 22 agent quantity on sample surface tension is shown in
ux), 20% to 80%, respectively. The melting point of Fig. 5 above. Adequate amount of the releasing agent
12 : 22 ux is 825C which is the lowest among the depends on type of sample, dilution ratio, and surface
mixed types. condition of the mold.
(d) Sodium tetraborate Na2B4O7
This is an acid compound and has the lowest melting
point of 741C and samples can therefore easy be fused.
However, its high deliquescence makes it unsuitable for
long term use, and the Na in the ux restricts analysis of
Na in sample.

4.3.Releasing agent Fig. 5. Effect of releasing agent quantity on sample surface


Halides such as iodides or bromides are used as tension.
releasing agents to facilitate removal of the cooled
fusion bead from the mold. Addition of releasing agent
Table 3. Example of interference line from releasing agent.
increases surface tension of sample during fusion since
halogen remains on the surface and therefore surface Releasing agent Analysis line Interference line
area of sample in contact with the vessel is reduced. As a Bromide (Br) Al-K (=0.8340 nm) Br-L (=0.8375 nm)
result, the cooled fusion bead becomes easier to remove
Iodide (I) Ti-K (=0.2750 nm) I-L2 (=0.2752 nm)
from the mold. In addition, it improves removal of air

Table 4. Physical properties of various oxidizing agents.

Lithium Sodium Potassium Strontium Ammonium Lithium Sodium Vanadium


Oxidizing agent
nitrate nitrate nitrate nitrate nitrate carbonate carbonate oxide

Formula LiNO3 NaNO3 KNO3 Sr(NO3)2 NH4NO3 Li2CO3 Na2CO3 V 2O 5

Melting point 264C 306.8C 339C 570C 169.6C 618C 851C 690C

Decomposition temperature 600C 380C 400C 570C 210C 618C 851C 1750C

Deliquescence Yes Yes Yes Yes Yes No No No

NO2 NO2 NO2 NO2 NH3, NO2 CO2 CO2 Oxidation


Remarks
originated originated originated originated originated originated originated catalyst

Rigaku Journal, 31(2), 2015 15


Sample preparation for X-ray fluorescence analysis IV. Fusion bead methodpart 1 basic principals

4.4.Oxidizing agent mixing sample and ux before fusion is advisable in


Samples which include metals, carbon and sulfur this case. When unfused residue is observed, change
are fused with oxidizing agent since they can react the ux type considering the acid-base, or fuse it again.
and form an alloy with the platinum in the vessel and Pulverization of samples can also reduce residuals since
cause irreversible damage. It is possible to oxidize the coarse powders require longer fusion time.
sample using a strong oxidant such as nitric acid and
then dried prior to fusion. However, for some samples 6.Vessel maintenance
it may be easier to perform oxidization during the Fragments of fusion bead attached to the vessel
pre-heating stage prior to fusion. Typical oxidizing and difcult to remove are soluble in 30% (w/v) citric
agents are nitrates such as LiNO3, NaNO3, KNO3 and acid solutions. Heated citric acid solutions or diluted
Sr(NO3)2 which cause oxidation at high temperatures. hydrochloric acid can speed up removal.
Since decomposition temperature varies depending on Thermal stress from repeated use of the vessel can
oxidizer, often times a mixture of several different cause mosaic-like patterns to appear on the surface due
nitrates are used. to sample, oxidizer or releasing agent residue in the
Other than ammonium nitrate, elements in the micro-cracks. This not only can lead to the fusion bead
oxidizer remain in the fusion bead and therefore a to crack during cooling and make releasing sample
consistent amount should be added. Generally, since difcult, but also can be a cause of analysis error. In
nitrates easily dissolve in water, preparation of oxides as such cases, polish the vessel surface with a cloth or
water solution in advance and addition by a micropipette brous buff with alumina paste or diamond paste with
improves consistent addition of a homogeneous mixture. particle size less than 1 m.
Carbonate is occasionally used as another type of Polishing allows repeat use of the vessel, but
oxidizing agent. Since carbonate has higher melting eventually the bottom surface begins to deform making
point and decomposition temperature compared with preparation of at samples difcult. When the damage to
nitride, it is mostly used for oxidization of ferroalloy or the vessel surface increases, problems such as cracking,
metal sample. Vanadium oxide works as an oxidation air being trapped and releasing become more serious. In
catalyst for the oxidization of sample by oxygen in the such cases, recasting of vessel may be required.
air. Table 4 is a compilation of the physical properties of
various oxidizing agents(8). 7.Standard
Following standards regarding XRF analysis by
4.5.Other reagents fusion bead method are published, and detailed sample
(a) Lithium uoride preparation methods are also described(10).
Lithium uoride works not only as a releasing
agent but also acts to lower the viscosity and melting JIS M 8205:2000 Iron oresX-ray uorescence
temperature. For mixed ux of lithium tetraborate 90% spectrometric analysis
and lithium uoride 10%, the melting point is as low as JIS R 2216:2005 Methods for X-ray uorescence
780C, and therefore volatilization of sample and ux spectrometric analysis of refractory products
can be reduced drastically. Adhesion to the vessel is low JIS R 5204:2002 Chemical analysis method of cement
as its uidity is very high. by x-ray uorescence
(b) Oxide of heavy element ISO 4503:1978 Hardmetals Determination of
In case of wide elemental concentration range such contents of metallic elements by X-ray uorescence
for geological samples, addition of heavy element based Fusion method
oxides (La2O3, CeO2, etc.) further reduces absorption/ ISO 9516-1:2003 Iron oresDetermination of various
excitation effects due to matrix, and the linearity of the elements by X-ray uorescence spectrometryPart
calibration line is improved (Heavy element dilution 1: Comprehensive procedure
effect)(9). ISO 12677:2011 Chemical analysis of refractory
products by X-ray uorescence (XRF) Fused cast-
5.Other sample preparation considerations bead method
5.1.Trapped air ISO 29581-2:2010 CementTest methodsPart 2:
As mentioned above, samples having high Chemical analysis by X-ray uorescence
concentration of Ca such as limestone, cement raw meal
etc. cause signicant foaming when fused, and they air 8.Summary
may be trapped in the fusion bead. Calcination at lower This article describes general principles of fusion
temperature prior to fusion can reduce the air being bead preparation regarding operation, instruments,
trapped in the fusion bead. reagents and precaution. In a following issue regarding
fusion beads, various fusion bead preparation methods
5.2.Residual sample unfused with practical application examples will be discussed.
Sample with high quartz concentration such as high
silica materials are difcult to fuse completely, and References
the unfused residual causes analysis error. Thoroughly (1) A. Morikawa: Rigaku Journal (English version), 30 (2014), No.

Rigaku Journal, 31(2), 2015 16


Sample preparation for X-ray fluorescence analysis IV. Fusion bead methodpart 1 basic principals

2, 2327. and Allied Materials, John Wiley and Sons Ltd., (1992).
(2) F. Claisse: Norelco Reporter JanuaryFebruary, 3 (1957), No. (7) F. Claisse and J. S. Blanchette: Physics and Chemistry of Borate
1, 37. FusionFor X-ray Fluorescence Spectroscopists, Fernand
(3) M. Watanabe, H. Inoue and Y. Yamada, M. Feeney, L Oelofse Claisse Inc., (2004).
and Y. Kataoka: Adv. in X-ray Anal., 56 (2013), 177184. (8) Kagaku binran kisohen I, Ed. The Chemical Society of Japan,
(4) M. Watanabe, Y. Yamada, H. Inoue and Y. Kataoka: Adv. in Maruzen, Tokyo, (2004), 113363 (in Japanese).
X-ray Chem. Anal. Japan, 44 (2013), 253259. (9) K. Norrish and J. T. Hutton: Geochim. Cosmochim. Acta, 33
(5) Rigaku Corporation: Guide for X-ray uorescence spectroscopy, (1969), 431453.
(2008), 101103 (in Japanese). (10) H. Homma: Keiko Xsen bunseki no jissai, Ed. I. Nakai, Asakura
(6) H. Bennett and G. Oliver: XRF Analysis of Ceramics, Minerals Shoten, Tokyo, (2005), 7071 (in Japanese).

Rigaku Journal, 31(2), 2015 17

You might also like