Professional Documents
Culture Documents
SCIENTIFIC
2015 E d iti on
DESIGN
2
TRAINING
TESTING
CERTIFICATION
Our Philosophy
Copley Scientific is a strong, stable Continuous improvement is a core
company with a track record of element of this approach and we
success, but equally important we strive to exceed the expectations of
are agile and forward-looking, with a the industry, not only by enhancing
philosphy closely aligned to the needs equipment performance but also
of the market we serve. through unrivalled service.
3
CONTENTS
4
CONTENTS
Description of
Page 8 Page 9 Page 10 Page 11
Drug Delivery Device
Sampling Apparatus (Page 24) Sampling Apparatus (Page 28) DUSA for Nebulisers (Page 32) N/A
Waste Shot Collector (Page 31) Waste Shot Collector (Page 31) Breath Simulator (Page 73) Direct into Vial
Breath Actuation Controller Critical Flow Controller Adapter for above (Page 33)
(Page 84) - USP only (Page 78) Mouthpiece Adapter
Delivered Dose Flow Meter (Page 90) Flow Meter (Page 90) (Page 92)
Uniformity Adapter for above (Page 91) Adapter for above (Page 71)
Mouthpiece Adapter Mouthpiece Adapter
(Page 92) (Page 92)
Pump (Page 93) Pump (Page 93)
DUSA Shaker (Page 106) DUSA Shaker (Page 106)
Introduction (117-122)
Qualification IQ / OQ Qualification Documentation and Tools (Pages 123-125)
6
Inhaled drug products
Introduction The devices used for inhaled and nasal However, their relatively high cost and
drug delivery are collectively referred reliance on inhalation strength and
to as orally inhaled and nasal drug duration are potential drawbacks.
products (OINDPs).
Aqueous droplet inhalers are a
The range of products available is new generation of devices that
broad, encompassing inhalers (metered- deliver a pre-metered dose of liquid
dose, dry powder and aqueous droplet), formulation without using a propellant.
nebulisers (jet, ultrasonic and vibrating They actively aerosolise the liquid,
mesh) and nasal (aqueous based, dry producing a soft mist of fine
powder and propellant based). particles. These inhalers generally
deliver a higher fine particle fraction to
Pressurised metered-dose inhalers
the lungs than MDIs or DPIs. As with
(pMDIs or simply MDIs) use a
any multi-dose liquid system, microbial
propellant to deliver a fixed volume
contamination can be a problem.
of liquid solution or suspension to the
patient in the form of an aerosol. They Nebulisers, like aqueous droplet
are small, inexpensive, convenient for inhalers, actively aerosolise a liquid
the user and suitable for a wide range formulation. Nebulisers, however,
of drugs. At the same time, the use normally operate continuously once
of MDIs requires good coordination loaded. They are widely used at home
and technique, and the actuation force and in hospital and demand little
needed means they are not always or no coordination for effective use.
suitable for elderly or paediatric users. The disadvantage is that they tend to
Spacers (or valved holding chambers) be cumbersome and require either
and new breath-actuated MDIs can compressed air or an electrical supply.
resolve these problems. New vibrating mesh technology is an
improvement, delivering portable,
Dry Powder Inhalers (DPIs) are
silent battery-operated devices.
an attractive option to an industry
well used to powder formulations. Historically used to treat allergic
Typically, the active drug is mixed with or seasonal rhinitis, there is now
an excipient containing much larger increasing interest in the use of Nasal
particles, for example lactose, to which Sprays for systemic drug delivery. Like
it attaches. During aerosolisation the inhalers, nasal sprays can be liquid or
active is stripped from the carrier and powder based. Aqueous systems can
inhaled whilst the carrier particles be manually actuated or propellant
impact on the mouth and throat and are driven. They are commonly multi-dose
ingested. DPIs synchronise drug delivery although unit dose devices are popular
with inhalation. for delivering vaccines and pain relief.
7
Inhaled drug products
8
Inhaled drug products
Historically, the DPI was limited to The passive DPI can be sub-divided 3. Aqueous Droplet Inhalers
single dose capsule systems and the into two categories: (Solution Metering Inhalers)
inhaler market was dominated by the
a) pre-metered (single or multi-dose) Both Metered-Dose (MDIs) and Dry
chlorofluorocarbon (CFC) propelled
where the dose is pre-measured Powder Inhalers (DPIs) suffer from
MDI.
during manufacture as, for the same two inherent problems: low
When, in 1997, the Montreal Protocol example, blisters, capsules or lung deposits (typically 5-20%) and
effectively banned the ozone similar cavities dose variability (often due to patient
depleting CFCs, the pharmaceutical b) device metered where the drug is difficulties in coordination or inspiration).
industry was faced with the option of: contained in a reservoir within the
Aqueous Droplet Inhalers (often known
device which pre-measures each
a) finding an alternative propellant as Solution Metering or Soft Mist
dose on actuation
(as in the HFA propelled MDI), or Inhalers), are a new generation of inhaler
b) developing new ways of delivering Some DPIs have devices for assisting designed to overcome these problems.
drug to the lungs the patients inspiration whilst
The Aqueous Droplet Inhaler is
simultaneously improving the accuracy
It is the latter that has provided a effectively a propellant-free Metered-
and reproducibility of the delivered
resurgence of interest in the DPI. As Dose Liquid Inhalers (MDLIs) sharing the
dose.
the name suggests, in the DPI the following common characteristics:
medication comes in the form of a dry Such devices are normally termed
Precision-dosed liquid-based
powder. active DPIs and are particularly useful
metering (normally water or
where the patients own inspiration
The majority of DPIs are passive ethanol)
capability is compromised. Assistance
devices, that is to say they rely on the Active aerosol generation
normally comes in the form of
patients inspiration to operate. There (mechanical or electromechanical)
pressurised/compressed air or through
is no need to coordinate breathing Patient-independent, accurate and
vibrations generated by
with the activation the reproducible dosing
a piezoelectric
patient simply inhales High fine particle fraction
transducer.
deeply to access the drug. Require sterile production and
addition of bacteriostatic agents to
prevent microbial contamination in
the case of multi-dose solution
reservoirs
The Human Respiratory Methods of aerosol generation vary: (a)
System
forcing liquid through nozzles,
(b) electrospraying, (c) thermal
generation, and (d) vibration mesh all
being typical examples.
9
Inhaled drug products
4. Nebulisers
Jet Nebuliser
10
Inhaled drug products
11
Title and their roles
organisations
Organisations and
their roles
Introduction b) in the case of OINDPs, by the and Q12 published by the International
International Standards Conference on Harmonisation of
Organisation (ISO) whose function Technical Requirements for Registration
The ultimate responsibility for the
is to define the standards and of Pharmaceuticals for Human Use
safety, quality and efficacy of
methods relating to the medical (ICH), extends this philosophy to all
medicines and medical devices lies
device, e.g. inhaler, nebuliser, etc., parts of the product cycle from product
with the various national regulatory
concerned. development, transfer through to
bodies designated to safeguard
manufacturing, manufacturing and
public health. In 2002 the FDA launched a new
finally product end.
initiative Pharmaceutical cGMPs
In Europe and in the USA this
for the 21st Century in which it ICH Q8 Pharmaceutical Development
function is performed by the
proposed a new risk-based approach describes the suggested contents of
European Medicines Agency
to pharmaceutical manufacturing. a regulatory submission based on the
(EMA) and by the Food and Drug
QbD format.
Administration (FDA) respectively. This initiative gave birth to Process
Analytical Technology (PAT), a ICH Q9 details a systematic approach
The regulatory authorities are
framework for understanding and to quality risk management whilst
supported in this role by:
improving the processes involved ICH Q10 describes a new quality
a) the Pharmacopoeias whose job is in Pharmaceutical Development, management system based on the
to define the standards with which Manufacturing and Quality Control, complete product lifecycle and referred
the drug formulation shall comply described in FDAs Guidance of to as the Pharmaceutical Quality
and the methods by which September 2004. System.
compliance will be adjudged, and
PAT operates on the premise that ICH Q11 provides a Guideline to
quality cannot be tested into products; the Development and Manufacture
rather, it should be built-in or should of Drug Substances including the
be by design. type and extent of information to be
submitted in regulatory dossiers.
The goal is to ensure final product
quality by understanding and Finally, ICH Q12 (final concept stage)
controlling the processes involved in is intended to work with ICH Q8-Q11
the manufacturing operation. guidelines to provide a framework to
facilitate the management of the entire
The Quality by Design (QbD)
Pharmaceutical Product Lifecycle
approach agreed and recently
focusing particularly on the Commercial
recommended for adoption by the
Manufacturing phase.
EMA, FDA and the Japanese MWHL
in the form of the five quality related
guidelines, ICH Q8, Q9, Q10, Q11
12
organisations and their roles
Guideline on the Requirements for Clinical Documentation for Orally Inhaled Products (OIP)
EMA including the Requirements for Demonstration of Therapeutic Equivalence between Two
Guidelines Inhaled products for use in the Treatment of Asthma and Chronic Obstructive Pulmonary
Disease (COPD) in Adults and for the use of Asthma in Children and Adolescents (2009)
Drug Efficacy
European Pharmacopoeia Preparations for Inhalations (Dosage Forms 0671) Nasal Preparations Preparations for
2014 Nebulisation
(8th Edition) Aerodynamic Assessment of Fine Particles (Chapter 2.9.18) (Dosage Forms 0676) (Chapter 2.9.44)
Inhalation & Nasal Drug Products - General Information & Product Quality Tests <5> Products for
Aerosols, Nasal Sprays, Metered-Dose Inhalers and Dry Powder Inhalers <601>
Nebulization
US Pharmacopeia Uniformity of Dosage Units <905>
Pharmaceutical Dosage Forms (Aerosols - Inhalations) <1151> <1601>
2015
(USP 38)
Spacers & VHCs
<Draft 1602>
Device Efficacy
International Standards Aerosol Drug Delivery Devices - Requirements and test methods Nebulizing Systems
Organisation (ISO 20072: 2013) (ISO 27427: 2013)
Expert Groups
European Pharmaceutical EPAG
Aerosol Group (EPAG) European based industry expert group involved in orally inhaled and nasal drug products
13
organisations and their roles
The parameters are limited to those Q4 - Q4B Pharmacopoeias Q10 - Pharmaceutical Quality System
that specifically relate to inhaled Q5A - Q5E Quality of Q11 - Development and Manufacture
products and which are critical to the Biotechnological Products of Drug Substances
safety, quality and efficacy of the final
Q6A - Q6B Specifications Q12 - Lifecycle Management
product.
14
organisations and their roles
2. INTERNATIONAL REGULATION
AND HARMONISATION
15
organisations and their roles
Current Ph.Eur.
16
organisations and their roles
Current USP
Site of Release Typical Dose Forms Product Tests for Product Tests for
ROUTE OF ADMINSTRATION QUALITY PERFORMANCE
<601>, <602>
Inhalation and Nasal Drug Products Aerosols, sprays, <603>, <604>
Lung, nasal cavity powders <5>
Chapter <5> <1601>, <1602>
17
organisations and their roles
5. EXPERT GROUPS
Pharmacopeial Forum
In addition to the above, there are a
number of industry and quasi-industry
expert groups whose role is to assist
the regulatory bodies in establishing
best practice in their thinking and
guidance.
4. DEVICE SAFETY,
QUALITY AND EFFICACY European Pharmaceutical
- ISO Aerosol Group (EPAG)
Most orally inhaled and nasal A group of 20 member companies
drug products (OINDPs) are active in the OINDP market
unique dosage forms in so within Europe formed to establish
far as that they comprise scientifically based best practice,
two components: provide consensus comment to
industry and government agencies
(a) The drug formulation
on safety and quality issues and
Like Ph.Eur., USP produce a bi-monthly (b) The medical device delivering that recommend harmonised standards
publication which contains discussion formulation to the patient and methodology.
documents relating to new and/or
The responsibility of defining the
amended chapters and monographs.
standards relating to the medical
Pharmacopeial Forum features items
device resides with the International
relating to In-Process Revision,
Standards Organisation (ISO).
Harmonisation and Stimuli to the
Revision Process, e.g. The relevant standards are ISO International Pharmaceutical
Verification of Operating the 20072 Aerosol drug delivery device Consortium on Regulation and
Andersen Cascade Impactor at verification Requirements and test Science (IPAC-RS)
Different Flow Rates Vol. 22(2), methods for Inhalers and ISO
A group of 19 international companies
p. 2211-2215 (1996) 27427 Anaesthetic and respiratory
including suppliers, that is committed
ACI: Procedure for Powder Inhalers: equipment Nebulising Systems and
to advancing consensus-based,
Modified Configuration Vol. 28(2) Components for nebulisers.
scientifically driven standards and
p. 601-603 (2002)
regulations for orally inhaled and nasal
drug products worldwide.
18
organisations and their roles
As can be seen from the preceding Apart from the mandatory testing for 6A. DELIVERED DOSE
pages, there are many advantages to leachables, extractables and microbial
The Sampling Apparatus used
using inhaled drugs for targetting the contaminants, two of the main factors
for determining the amount and
lungs or nasal mucosa as a means of largely recognised as Critical Quality
uniformity of the delivered dose for
providing local or systemic therapy. Attributes (CQAs) in the testing of
Metered-Dose Inhalers (MDIs) was
OINDPs (in both pharmaceutical
These advantages have led to a originally designed by Charles Thiel
development and batch release) are:
growing number of new types of who was then at 3M Laboratories,
device designed to provide the Delivered Dose (Emitted Dose) Minneapolis, USA.
accuracy and sophistication necessary
The total amount of drug emitted from The design has subsequently been
to deliver the drugs concerned.
the drug device and hence available amended to replace the original screw
Other challenges have presented to the user and fittings with easier-to-use bayonet
themselves over the last few years fittings, whilst maintaining the critical
Particle Size (Aerodynamic Size
involving drug delivery device internal dimensions of the original
Distribution)
changes, not least the ban on ozone design.
depleting chlorofluorocarbons (CFCs) The size of the particles or droplets
A second and larger sampling
used as propellants in MDIs and the that make up the emitted aerosol
apparatus, the Sampling Apparatus
incorporation of dose counters into cloud. Particle size determines the
for Dry Powder Inhalers, has been
multi-dose devices. percentage of the total emitted dose
introduced for DPIs based on a similar
that actually reaches the lungs or nasal
The regulatory bodies (EMA, FDA, design by Byron and Hindle, Virginia
mucosa during inhalation and is thus,
MHWL, ICH and others) are constantly Commonwealth University.
therapeutically effective.
evolving their requirements (both in
Both types of apparatus appear in
terms of pharmaceutical development These two physical tests form the basis
the European Pharmacopoeia under
and manufacture) to meet the of many of the parameters used by the
Dosage Forms - Preparations for
challenges of these new technologies regulators to characterise inhalation
Inhalation <0671> and in the US
and to ensure their safety, quality and and nasal products.
Pharmacopeia under Section <601>.
efficacy in the global marketplace.
19
organisations and their roles
6B. PARTICLE SIZE (AERODYNAMIC Ph.Eur. Apparatus C: Multi-Stage impactors appear in both Ph.Eur. and
SIZE DISTRIBUTION) Liquid Impinger (MSLI) USP:
The cascade impactor is the instrument Ph.Eur. Apparatus D: Andersen Multi-Stage Liquid Impinger (MSLI)
of choice for both regulators and Cascade Impactor (ACI)
Andersen Cascade Impactor (ACI)
Pharmacopoeias when measuring the
Ph.Eur. Apparatus E: Next
aerodynamic size distribution (particle Next Generation Impactor (NGI)
Generation Impactor (NGI)
size) of inhaled products.
Both Pharmacopoeias specify test
Procedures for Apparatus E - NGI
The aerodynamic size distribution methods for all three impactors for use
(Chapter 2.9.44) are also specified for
of an aerosol cloud defines where with DPIs and for the NGI for nebulisers
nebulisers.
the particles in that cloud are likely - see Ph.Eur. Chapter 2.9.44 and USP
to deposit following inhalation. It is The United States Pharmacopeia Chapter <1601>.
generally accepted, for example, that to (USP) Test Chapter <601> specifies
In the case of USP however, the use
be therapeutically effective the particles six impactors suitable for aerodynamic
of the MSLI is restricted to DPIs only,
should be in the range of 1 to 5 microns size distribution:
which leaves just the Andersen Cascade
in order to deposit in the lungs. The
USP Apparatus 1 for MDIs: Andersen Impactor (ACI) and Next Generation
particle mass below 5 microns is normally
Cascade Impactor (ACI) Impactor (NGI) as suitable candidates
described as the fine particle mass.
for testing both DPIs and MDIs if both
USP Apparatus 2 for DPIs: Marple
Particles having an aerodynamic size in Pharmacopoeial standards are to be
Miller Impactor (MMI)
excess of 5 microns will generally impact satisfied.
in the oropharynx and be swallowed USP Apparatus 3 for DPIs: Andersen
whereas below 1 micron the possibility Cascade Impactor (ACI) +
exists that the particles will remain Preseparator
entrained in the air stream and be
USP Apparatus 4 for DPIs: Multi-
exhaled.
Stage Liquid Impinger (MSLI)
The European Pharmacopoeia (Ph.
USP Apparatus 5 for DPIs: Next
Eur.) Method Chapter 2.9.18 currently
Generation Impactor (NGI) +
specifies one twin and three multi-
Preseparator
stage impactors for the aerodynamic
assessment of fine particles in both MDIs USP Apparatus 6 for MDIs: Next
and DPIs: Generation Impactor (NGI)
20
delivered dose uniformity
Delivered Dose
Uniformity
The devices used for inhaled and nasal
Introduction Pressurised Metered-Dose Inhalers
drug delivery are collectively referred (pMDIs) are relatively insensitive
to as orally inhaled and nasal drug to changes in flow rate because
products (OINDPs). the aerosolisation and dispersion
mechanism is dependent on the force
The safety, quality and efficacy of
generated by the propellant.
OINDPs is dependent on four critical
quality attributes:
Therefore, for MDIs, the air flow rate is
the delivered dose (the amount of fixed at an arbitrary rate of 28.3 L/min
drug that the user actually receives) equivalent to 1 cubic foot per minute.
the aerodynamic particle size
For Dry Powder Inhalers (DPIs), the
distribution of that dose (or more
test regime is more complex. The
precisely the fraction of the
aerosolisation of DPIs depends on
delivered dose of the appropriate
the strength and duration of a single
size to reach the target site)
inhalation on the part of the user.
the presence and possible inhalation
of leachables or microbial When inhaling in this manner, the
contaminants typical adult produces a pressure drop
in some instances, the spray pattern over the device of approximately
or plume geometry of the device 4 kPa. Depending on the device flow
under test resistance this will yield a flow rate,
typical of the mean patient inhalation
The Delivered Dose is measured by
flow rate, that is then used for all the
firing the test device into a sampling
required testing of that device.
apparatus containing a filter.
Similarly, the duration of the test is set
The dose is captured, the active drug
on the basis of the total air volume
is dissolved in solvent and an aliquot
typically inhaled in one adult breath,
is then analysed, normally using High
adjudged to be 4 litres in the case of
Pressure Liquid Chromatography
the Ph.Eur. and 2 litres in the case of
(HPLC).
the FDA and USP.
During testing, air is
In the case of nebulisers and MDIs
drawn through the
with spacers or VHCs, the user inhales
sampling apparatus
the drug as part of tidal breathing at
to broadly simulate
rest. In this instance, the breathing
DUSA for MDIs inhalation. The manner in
cycle in vitro is replicated by means of
(Aluminium) which the air is drawn through
a breath simulator.
the apparatus is dependent on the
device under test.
21
delivered dose uniformity
The Delivered Dose measures the A further study is required in the case If 2 or 3 values lie outside the
mass of the drug that is emitted from of DPIs, DDU over patient flow rate 75% - 125% limits then the test must
the mouthpiece of an inhaler when the range. This is because DPIs rely on the be repeated for 2 more inhalers
device is actuated according to the patients inspiration for their therapeutic whereupon not more than 3 of the 30
manufacturers instructions. effectiveness. values lie outside the 75% -125% band
and no value lies outside the
The tests described under DDU are As far as manufacture is concerned,
65% - 135% band.
designed to demonstrate: tests are required to determine both
the Delivered Dose Uniformity and If the inhaler contains more than one
1. The consistency of drug emitted
the Mean Delivered Dose. active, then a separate test should be
from a number of inhalers within
carried out for each individual drug.
a specified batch. The Mean Delivered Dose is the
amount of drug in one actuation and is Checks are also required to ensure
2. In the case of multi-dose inhalers,
determined by calculating the mean of that the number of deliveries from
the consistency of drug emitted from
the DDU test results. Limits of +/-15% the device are within the stated label
various actuations throughout the
of the label claim apply. claim.
life of a specified inhaler.
Regarding the method to be employed, Note: For pre-metered systems,
3. That the number of deliveries per
the EMA simply states that the DDU collect and analyse 10 individual
inhaler is equal to or or greater
test should be conducted according to doses.
than the labelled amount.
an accepted pharmacopoeial method,
4. In the case of DPIs, that the effect or a suitably validated alternative. FOOD & DRUG ADMINISTRATION
of varying flow rates as (FDA)
demonstrated by various patients EUROPEAN PHARMACOPOEIA
The FDA guidelines on MDIs and DPIs
has been taken into account. (PH.EUR. 8.0)
is contained in the Draft Guidance of
The sampling procedure and The references in the European that name published in 1998. FDA
acceptance criteria for delivered Pharmacopoeia to the Uniformity suggests two tests applicable to
dose uniformity of orally inhaled of delivered dose, the Number of both MDIs and DPIs, namely, Dose
products (OIPs) varies according to the deliveries per inhaler and, in the Content Uniformity and Dose
Regulatory Authority concerned (see case of DPIs, the Number of deliveries Content Uniformity through container
below). per inhaler for multidose inhalers life.
are to found under Preparations for
The Dose Content Uniformity test
EUROPEAN MEDICINES AGENCY Inhalation (0671) in the chapter on
recommends one sample be taken
(EMA) Dosage Forms.
from 10 separate inhalers. To comply,
The EMA guidance for OINDPs is In the case of DDU, the same sampling 9 out of the 10 results must lie
contained in the 2006 publication procedure applies to both pressurised between 80% and 120% of the label
Guideline on the Pharmaceutical and non-pressurised MDIs and DPIs. claim, all are between 85% and 115%
Quality of Inhalation and Nasal of the claim and the mean does not lie
Ph.Eur. specifies that a total of 10 doses
products and is divided into two outside 85% -115% of the label claim.
are to be collected in order to obtain
sections, one relating to pharmaceutical
a representative sample over the life If 2 or 3 values lie outside the
development and a second relating to
time of the inhaler, basically 3 at the 80% - 120% limits ( all other criteria
product manufacture.
beginning, 4 in the middle and 3 at the being met ), an additional 20 inhalers
In the case of DDU, it applies to all MDI end. should be sampled. To comply to this
(pressurised and non-pressurised) and second tier of testing, 3 out of the
To comply, 9 out of the 10 results must
DPI products. 30 results must lie between 80% and
lie between 75% and 125% of the
120% of the label claim ( all other
The main study applicable to average value and all between 65% and
criteria being met ).
pharmaceutical development relates to 135%.
the DDU through container life.
22
Title
delivered dose uniformity
The Dose Content Uniformity If 2 or 3 values lie outside the 80% - uniformity over the entire unit life)
through container life recommends 120% limits (all other criteria being are to be found in Chapter <601>.
the collection of 9 samples throughout met), an additional 6 inhalers (1
The USP specifies that samples
the life of 3 individual inhalers, 1 at beginning, 1 middle, 1 end) should be
should be taken from 10 separate
the beginning, 1 in the middle and 1 sampled. To comply to this second tier
inhalers and in the case of dose
at the end. of testing, not more than 3 out of the
uniformity over the entire unit life,
total of the 27 results must lie between
To comply, 8 out of the 9 results must a sample at the beginning and end
80% and 120% of the label claim (all
lie between 80% and 120% of the of each inhaler making a total of 20
other criteria being met).
label claim, all are between 85% and determinations.
115% of the claim and the means
UNITED STATES PHARMACOPEIA As at today, USP 38 no longer states
for the beginning, middle and end
(USP 38) specfic acceptance criteria for orally
samples do not lie outside 85% -115%
inhaled products.
of the label claim. The references in USP 38 to Delivered-
Dose Uniformity (including dose
1st Test Tier 1st Test Tier 2nd Test Tier 2nd Test Tier
REGULATORY AUTHORITY No. of Inhalers Criteria No. of Inhalers Criteria
EMA 2006
Delivery Dose Uniformity 1 Inhaler / 10 doses 3 Inhalers / 30 doses
9/10 doses to be 75-125% 27/30 doses to be 75-125%
DDU through container life 1 Inhaler / 10 doses of Average Value 3 Inhalers / 30 doses of Average Value
DDU over patient flow rate range As appropriate As appropriate As appropriate As appropiate
Number of deliveries per inhaler 1 Inhaler > Label Amount N/A N/A
FDA 1998
Dose Content Uniformity 9/10 doses to be 80-120% 27/30 doses to be 80-120%
10 Inhalers / 1 dose of Label Claim 30 Inhalers / 1 dose of Label Claim
DCU through container life 3 Inhalers / 9 doses 8/9 doses to be 9 Inhalers / 27 doses 24/27 doses to be
80-120% of Label Claim 80-120% of Label Claim
Effect of Varying Flow Rates (as appropriate)
USP 38 <601>
Delivered Dose Uniformity 10 Inhalers / 1 dose
Not applicable Not applicable Not applicable
DDU over the entire unit life 10 Inhalers / 2 doses
23
delivered dose uniformity
Component Parts
(and example MDI)
INTRODUCTION
It is used to perform
those tests specified in
the Pharmacopoeias
relating to delivered or
emitted dose, namely
24
delivered dose uniformity
DESCRIPTION
The Dosage Unit Sampling Apparatus Alternative materials are available The combination of the clamp to
(DUSA) for MDIs consists of one on request, e.g. aluminium or 316 secure the filter support cap and
collection tube, two rinsing caps, one stainless steel. All tubes and caps the boss head to alter its angle
filter support cap, one flow meter are laser numbered to assist in allows collection tubes to be quickly
adapter and a starter pack of filters traceability. connected to the vacuum system prior
supplied in a handy carrying case. to testing and removed once the test
The sample collection tube is fitted
is complete.
Spare collection tubes and caps are with a 25 mm glass fibre filter having a
available. typical aerosol retention of A further base plate option includes
0.3 microns. mounting fixtures for the Waste Shot
The standard collection tube itself and
Collector (see Page 31) and Switching
rinsing caps are manufactured from It has a volume of approx. 50 mL
Valve.
TecaPro MT, an FDA approved inert which equates to that of the human
polypropylene specifically formulated oropharynx. Using a Waste Shot Collector and a
for medical and pharmaceutical suitable Switching Valve mounted on
The standard unit comes with
applications. the Base Plate used as the stand for
silicone rubber seals. LDPE seals are
the DUSA, in conjunction with two
available as an option, in the event
mouthpiece adapters (one for the
of extractables being an issue with
DUSA and one for the WSC2) and a
silicone rubber.
number (say, 10) of spare collection
A stand comprising base plate, boss tubes, can provide substantial gains in
head and clamp for supporting the terms of throughput.
DUSA during use is available as an
option.
BAC 2000
Timer
25
delivered dose uniformity
System for Testing the Dose Uniformity of MDIs (incl. Waste Shot Collector and Switching Valve)
The minimum set-up for delivered In addition to the specifications laid This allows the time that the test flow
dose testing as specified by Ph.Eur. down in Ph.Eur., the FDA recommends, is applied to the inhaler to be adjusted
comprises a sample collection tube, and USP 38 specifies, that the volume to a specific volume, for example, the
fitted at one end with a suitable of air sampled should not exceed 2 litres required by USP.
mouthpiece adapter to accept the 2 litres, this being the volume of air
Operation can be triggered via the
inhaler under test and connected adjudged to be typical of the average
instrument front panel, foot switch or
at the other end to a vacuum pump patient.
RS 232 interface.
capable of continuously drawing
This additional criterion can be met by
28.3 L/min through the assembled The BAC 2000 can also be used for
positioning an electrically operated,
system (including the filter and the testing of Breath Actuated (or
timer controlled, two-way solenoid
inhaler). Breath Operated) MDIs.
valve, such as that incorporated in the
A flow meter should be used to adjust Breath Actuation Controller BAC In this case, the BAC 2000 is used to
the flow at the inlet to the correct rate 2000 (see Page 84), in the line between initiate the flow and hence trigger the
prior to testing, using the flow meter the collection tube and the vacuum breath actuated inhaler simultaneously.
adapter (see Pages 90-91). pump to control the air flow supply to
An optional DUSA Shaker for holding
the inhaler.
Once the device has been shaken, up to 21 DUSA for MDI collection
primed and actuated and the test is The BAC 2000 provides near tubes is available as an option (see
complete, the collection tube together instantaneous starting and stopping of Page 106).
with the filter is removed. Solvent is the air flow during testing and has both
then added and the tube capped and delay and inhaled time functions.
agitated to assist in drug dissolution
prior to recovery and analysis.
DUSA for MDIs with Breath Actuation Controller and Pump
testing a Breath Actuated MDI
26
delivered dose uniformity
It is important to note that the The following ancillaries are Breath Actuation Controller
British Pharmacopoeia has its own recommended to complete a fully (see Page 84)
unique apparatus for determining the operating test system for the delivered Flow Meter (see Page 90)
Content of Active Ingredient delivered dose testing of MDIs: Waste Shot Collector (see
by actuation of the valve (see below), Page 31)
Mouthpiece Adapter (see Page 92)
likely retained for historical reasons. DUSA Shaker for DUSA for MDIs
Vacuum Pump (see Page 93)
(see Page 106)
This comprises a stainless steel base
plate having three legs and a central
hole to accept the actuator stem in a
small vessel (to which solvent is added)
Cat. No. Description
suitable for shaking. 8201 Dosage Unit Sampling Apparatus for MDIs (Silicone Rubber Seals)
8201A Dosage Unit Sampling Apparatus for MDIs (LDPE Seals)
Accessories
Spare Parts
27
delivered dose uniformity
System for Testing the Dose Uniformity of DPIs (incl. Waste Shot Collector and Switching Valve)
INTRODUCTION
A second and larger version of As with the system suggested for testing Instruments such as the Critical
the Sampling Apparatus for MDIs, MDIs according to USP 38 <601>, an Flow Controller Model TPK 2000
capable of sampling at a variety electrically operated, timer controlled, (see Page 82) interposed between
of flow rates up to 100 L/min, is two-way solenoid valve is positioned in DUSA and vacuum pump simplify
available for use with Dry Powder the line between the collection tube and set-up in accordance with these
Inhalers (DPIs). the vacuum pump to control the air flow pharmacopoeial recommendations,
supply to the inhaler. measuring and recording all the
The Dosage Unit Sampling
parameters required for testing
Apparatus (DUSA) for DPIs is used In the case of DPIs this is mandatory
and controlling flow conditions and
to perform those tests specified because, unlike MDIs, the majority
ensuring critical (sonic) flow conditions
by the Pharmacopoeias that relate of these devices are passive breath-
during testing.
to delivered or emitted dose, actuated devices which rely on the
namely Uniformity of Delivered patients inspiration rather than a They also allow the time that the test
Dose, Dose Content Uniformity propellant for dose emission. flow is applied to the inhaler to be
and DDU or DCU through container adjusted to a specific volume, for
The testing of DPIs is further complicated
life. example 2 or 4 litres, thus equating
by the fact that different inhalers provide
to the inhaled volume of a typical
varying degrees of resistance to flow, i.e.
patient.
some require more effort to inhale than
others.
Schematic of Sampling Apparatus for DPIs with Pump and Critical Flow Controller
28
delivered dose uniformity
DESCRIPTION
The Dose Unit Sampling Apparatus The collection tube also differs from
(DUSA) for DPIs utilises the same that employed for MDIs in having
materials of construction as the a pressure tap (P1) in its wall that is
unit for MDIs. However, alternative used in conjunction with the Critical
materials are available on request, e.g. Flow Controller to measure the
aluminium or 316 stainless steel. pressure drop across the device.
Stand for 10
Collection Tubes
29
delivered dose uniformity
PROCEDURE
The minimum start-up requirement for 7. Replace the inhaler and discharge ANCILLARIES
DPI delivered dose testing is the same the dose into the collection
The following ancillaries are
as that for MDI testing described in tube by activating the timer on the
recommended to complete a fully
the preceding section, namely DUSA, Critical Flow Controller controlling
operating test system for the delivered
mouthpiece adapter, pump and flow the solenoid valve. Repeat as
dose testing of DPIs:
meter plus the addition of the Critical necessary to achieve the desired
Flow Controller (e.g. TPK) to measure number of doses. Mouthpiece Adapter (see Page 92)
the pressure drop across the device Vacuum Pump (see Page 93)
8. Add solvent to the collection tube,
and control the flow conditions during Critical Flow Controller
apply rinsing caps and then shake
testing accordingly. (see Page 78)
the tube vigorously before assaying
Flow Meter (see Page 90)
Proceed as follows: the contents.
Waste Shot Collector (see Page 31)
1. Assemble the system as per the An optional DUSA Shaker for holding Shaker for DUSA for DPIs (see Page
schematic of the DUSA for DPIs. up to 12 DUSA collection tubes is 106)
available (see Page 106).
2. Connect the inhaler to the
Collection Tube using a suitable
mouthpiece adapter. Cat. No. Description
3. Connect the tube marked P1 on 8601 Dosage Unit Sampling Apparatus for DPIs (Silicone Rubber Seals)
the Critical Flow Controller to the 8601A Dosage Unit Sampling Apparatus for DPIs (LDPE Seals)
pressure tap on the Collection
Tube.
Accessories
30
delivered dose uniformity
31
delivered dose uniformity
32
delivered dose uniformity
DESCRIPTION
Exhalation Filter (example; not included)
The Sampling Apparatus for Nebulisers
consists of a breath simulator to generate Inhalation Filter
Mouthpiece Adapter
the specified breathing profile, a filter (optional)
(optional)
holder containing the filter to capture the
active drug and a suitable mouthpiece
adapter to connect the filter holder to
the nebuliser under test. Breath
Inlet/Outlet Simulator
An angle adapter can be provided as an
optional extra where required to adjust Nebuliser (example; Angle Adapter
the angle of the nebuliser mouthpiece to not included) (optional)
Copley Scientific supply a range of 4. Set the breath simulator to ceases, i.e. the reservoir is empty.
Breath Simulators specifically designed generate the specified breathing
8. Using a suitable method,
to meet the requirements of the tests pattern.
determine the amount of active
concerned (see Pages 73-77).
5. Start both nebuliser and simulator on each filter.
and run for 60 seconds (or for such
PROCEDURE Determine the active substance
time that sufficient active is
delivery rate by dividing the mass of
Use a suitable Breath Simulator to collected on the filter for analysis).
active collected on the first filter by
generate the breathing pattern required
6. Pause both units and remove the the time taken to collect it.
in conjunction with the Filter Holder and
filter from the holder.
Adapter, Angle Adapter and a suitable Determine the total active substance
Mouthpiece Adapter to perform these 7. Place a fresh filter in the holder and delivered by summing the mass
two tests. continue the test until nebulisation collected on both filters.
Proceed as follows:
33
delivered dose uniformity
Spacer -
Bidirectional Open exit (no valve)
INTRODUCTION A Valved Holding Chamber (VHC) is The additional dead volume provided
similar but normally incorporates a one by the reservoir not only provides a
Pressurised metered dose inhalers
way valve close to the mouthpiece or reservoir for aerosol expansion, but
(pMDIs) are an inexpensive and
facemask. This opens to release the also particle impaction, settling and/
convenient means of treating asthma
aerosol cloud once the patient starts or electrostatic deposition within the
and other pulmonary diseases.
to inhale but prevents emptying the chamber itself, all of which can change
However, patient coordination holding chamber during exhalation as the emitted dose ahead of inhalation.
of actuation with inhalation can in the case of a simple spacer device.
As the use of add-on devices has
be a problem when using pMDIs Reverse Firing Spacers and VHCs
become more widespread, the
particularly in the young, old or are designed with an integral actuator
regulatory authorities responsible for
chronically ill. to accept an inhaler canister directly.
the safety and efficacy of OIPs have
In this instance, the pMDI is actuated
Add-on devices such as Spacers, become increasingly aware of the
into a bag in a direction pointing away
Valved Holding Chambers and need to test add-on devices as distinct
from the body (hence the description,
Reverse Firing VHCs, which reduce from pMDIs used on their own.
Reverse Firing) and then the patient
or eliminate the need for coordination
inhales slowly from the bag. As a result, USP has released a new
between actuation and inhalation and
draft chapter for testing Spacers and
also the cold freon effect associated All three Add-on devices result in
Valved Holding Chambers used with
with them, are widely used in the patient inhaling the drug from a
Inhalation Aerosols <1602>.
conjunction with pMDIs to overcome reservoir of aerosolised particles, not
this problem. dissimilar to a nebuliser, rather than The tests set out in the new draft
directly from the pMDI. chapter are based on experience
A Spacer is an open tube placed
in Canada gained over a ten year
between the inhaler and the
IN VITRO ASSESSMENT period culminating in the release of a
mouthpiece, or in some instances
new standard Spacers and Holding
facemask of the patient. In practical When a patient uses a pMDI without
Chambers for use with Metered-Dose
terms, they extend the distance an add-on device, the emitted
Inhalers by the Canadian Standard
between the inhaler and patient and dose and hence the drug particles
Association in 2011.
thus provide additional volume for the contained within it are inhaled almost
aerosol plume to develop. instantaneously as the formulation is The new methods reflect that, as
aerosolised. with a nebuliser, the amount of drug
received by the patient employing
In contrast, when an add-on device is
an add-on device with a pMDI will be
used, the patient inhales drug from
directly influenced by the inhalation
a reservoir of aerosolised
profile of the user concerned.
particles.
For that reason, the tests in the new
chapter call for the application of
specific breathing profiles to reflect
the physiology of the intended user
(see Page 35).
Proceed as follows:
35
delivered dose uniformity
DESCRIPTION - TEST PART 3 The FMA System comprises two key 2. The Face Model Support accepts
(Facemask based products) elements: three different models: infant, child
and adult. All models are fitted with
The purpose of this test is to confirm 1. The Device Securing Fixture
replaceable face skins representative
that the emitted dose from a spacer/ which secures the spacer/VHC and its
of real life tissue.
VHC used in conjunction with a associated facemask into position prior
facemask is comparable to that to testing. The device securing fixture The support also can be adjusted in
obtained in the fully coordinated has been designed to accommodate both axes to tilt the head from front to
simulation with the facemask various sizes of spacer/VHC. The back or orientate it from side to side.
removed. fixture is adjustable in two axes:
In this instance, the filter holder is
A critical component of the test x (horizontal) and located in a cavity behind the face
apparatus is the face model y (height) models lips.
employed. This should be appropriate by means of handwheels.
As with the mouthpiece based
to the age group for which the
An in-built digital gauge (Range 0 - 2.5 product, the specified breathing
spacer/VHC is intended, e.g. infant,
kg) measures the force applied to the profile is provided by a Breath
child or adult.
face model in Newtons or kg (e.g. 1.6 Simulator, details of which can be
The face model should be such as to: kg) as suggested in USP <1602>. found on Page 73.
36
delivered dose uniformity
Proceed as follows:
37
Aerodynamic Particle size
Aerodynamic
Particle Size
Introduction Together with delivered dose, the The Pharmacopoeias recommend
Aerodynamic Particle Size Distribution several commercially available impactors
(APSD) is widely recognised as a for the routine testing of OINDPs
Critical Quality Attribute (CGA) in the including the Andersen Cascade
in vitro characterisation of OINDPs Impactor (ACI) and Next Generation
since it is the APSD of an aerosol Impactor (NGI) both of which are used
cloud that defines where the particles globally for the testing of MDIs, DPIs
in that cloud are deposited following and ADIs (Aqueous Droplet Inhalers).
inhalation.
Special versions are available for the
It is generally accepted, for example, testing of nasal delivery systems,
that to be therapeutically effective, nebuliser systems and the add-on
the particles should be in the range of devices such as spacers and valved
1 to 5 microns. Particles in excess of holding chambers sometimes used with
5 microns will generally impact in the pMDIs.
oropharynx and be swallowed whereas
Following on from FDAs Guidance on
below 1 micron the possibility exists
Process Analytical Technology (PAT) in
that the particles will remain entrained
2004, in the last few years considerable
in the air stream and be exhaled.
interest has focused on the Quality
The preferred instrument of choice by Design (QbD) approach to
for measuring the APSD of inhaled pharmaceutical development and
products for both regulators and manufacture.
pharmacopoeias alike is the cascade
Because of the amount of APSD data
impactor.
demanded by these new initiatives,
This is because: attention has once again turned to faster
1. Cascade impactors measure methods of APSD determination and in
aerodynamic particle size (APSD). particular to the concept of Abbreviated
2. Cascade impactors measure active Impactor Measurement (AIM).
pharmaceutical ingredient (API).
In order to meet these demands and to
3. Cascade impactors measure the
provide a basis for the proof-of-concept
entire dose.
work to validate them, Copley Scientific
Cascade impactors are precision has introduced a number of different
engineered instruments that separate a versions of Abbreviated Impactors for
sample on the basis of particle inertia use in a QbD environment. These are
(which is a function of velocity and based on reduced stage versions of the
aerodynamic particle size) without the ACI and NGI respectively (see Page 60).
need to know either particle density or
shape.
38
Aerodynamic Particle size
INTRODUCTION
39
Aerodynamic Particle size
Inter-Stage Losses
Terminology
collect particles
using HPLC to determine the amount greater than 0.7
of drug actually present. a certain size 0.6
as shown in
By analysing the amount of drug this graph of 0.5
deposited on the various stages in this aerodynamic 0.4
manner, it is then possible to calculate diameter vs.
collection
0.3
the Fine Particle Dose (FPD) and Fine Actual collection
efficiency. The 0.2
Particle Fraction (FPF) and, following Undersized particle collected
stage cut-off
further manipulation, the Mass Median 0.1
diameter is
Aerodynamic Distribution (MMAD) defined as the 0
and Geometric Standard Deviation midway point on 0.1 1 10
the curve (D50). d50diameter
(GSD).
Aerodynamic Diameter (mm)
40
Aerodynamic Particle size
Andersen Cascade
Impactor (ACI)
(Aluminium, 316 Stainless Steel and Titanium)
INTRODUCTION
41
Aerodynamic Particle size
Next Generation Impactor (NGI) Main features of the Next Generation Advantages include:
Impactor (NGI) include:
In 1997, a group of prominent 4 Stages between 1.7 and 13 microns
pharmaceutical companies involved Designed by the pharmaceutical Operates between 30 and 100 L/min
in the development and manufacture industry for the pharmaceutical Virtually no inter-stage losses
of inhalers formed a consortium to industry Eliminates particle bounce and
develop a new impactor specifically Operates between 15 and 100 L/min hence re-entrainment problems
designed for testing pharmaceutical 7 stages (5 out of the 7 always Choice of aluminum, 316 stainless
inhalers using the latest design theory. between 0.54 and 6.12 microns) or titanium construction
Easy drug recovery with low inter- Easy and quick drug recovery
The result, the Next Generation
stage losses
Impactor (NGI), was launched in
High stage efficiency: all stages: Other Impactors
2000. Both design and subsequent
500 < Re < 3000
archival calibration are documented to The following impactors are also
3-part construction lends itself to
pharmaceutical standards. worthy of mention and are described
semi and full automation
in more detail later in the brochure:
The NGI is a high performance, Documented and published design
precision, particle classifying cascade and archival calibration Glass Impinger
impactor having seven stages plus a
Marple-Miller Cascade Impactor
micro-orifice collector (MOC). Multi-Stage Liquid Impinger (MSLI)
(MMI)
In practice, its flexibility of use and The Multi-Stage Liquid Impinger (MSLI)
high productivity are making the NGI was the first cascade impactor/impinger
the new workhorse within many specifically designed for inhaler testing.
inhaler research laboratories.
Whilst the 4-Stage MSLI does not
This trend will no doubt continue as offer the number of stages of the ACI
reproducibility and productivity are or NGI, it does, by definition, have
improved with the addition of new no inter-stage losses and is suitable
accessories designed to automate the throughout the range 30-100 L/min.
particle sizing process (see Page 105).
Unlike the ACI and NGI, the
Correlation studies between ACI and collection stages of the MSLI
NGI show good agreement between are kept moist which eliminates
particle size distributions although this the problem of particle bounce
does not necessarily mean they are associated with conventional
interchangeable for all DPIs. impactors.
Multi-Stage Liquid Impinger (MSLI) in Aluminium, 316 Stainless Steel and Titanium
42
Aerodynamic Particle size
The Andersen Cascade Impactor (ACI) particular stage collection plate, whilst The appropriate flow rate, Q, to
manufactured by Copley Scientific smaller particles with insufficient inertia give a pressure drop of 4 kPa
is an 8-stage cascade impactor that will remain entrained in the air stream The duration of simulated
has been designed for measuring the and pass to the next impaction stage. inspiration to give a volume of
aerodynamic particle size distribution 4 litres
By analysing the amount of drug
(APSD) generated by MDIs and DPIs. Flow rate stability in terms of critical
deposited on the various stages, it
(sonic) flow
It complies with the specifications is then possible to calculate the Fine
laid down in USP Chapter <601>, Particle Dose (FPD) and Fine Particle These factors require the use of the
Ph.Eur. 2.9.18 and the latest proposals Fraction (FPF) and following further General Control Equipment for
aimed at harmonising the respective manipulation, the Mass Median DPIs specified in USP <601> and
Pharmacopoeias. Aerodynamic Distribution (MMAD) and Experimental Set Up for testing DPIs
Geometric Standard Deviation (GSD) of in Ph.Eur. 2.9.18 which take all of these
IMPACTOR USE (METERED-DOSE the active drug particles collected. factors into account.
INHALERS)
These specifications form the basis
IMPACTOR USE (DRY POWDER
The standard Andersen Cascade of the Critical Flow Controllers (see
INHALERS)
Impactor is designed for use at Page 78) which incorporate all of
28.3 L/min (which is equivalent to 1 The same impactor can be used for the equipment required into a single
cubic foot/min). determining the particle size of Dry integrated system.
Powder Inhalers (DPIs).
The 8 stages have the following
particle size collection bands: In this instance, however, a
preseparator is interposed between
Stage 0 9.0 + microns
the induction port and stage 0 of the Diagram of the Human
Stage 1 5.8 9.0 microns Respiratory System
impactor in order to collect the large
Stage 2 4.7 5.8 microns
mass of non-inhalable powder boluses
Stage 3 3.3 4.7 microns
typically emitted from a DPI prior to
Stage 4 2.1 3.3 microns
their entry into the impactor.
Stage 5 1.1 2.1 microns
Stage 6 0.7 1.1 microns In the case of Dry Powder Inhalers
Stage 7 0.4 0.7 microns (DPIs), a number of additional
factors must be taken into
The Andersen Cascade Impactor, like
account when testing:
other cascade impactors, is designed
such that as the aerosol stream passes The pressure drop generated
through each stage, particles having by the air drawn through the
sufficient inertia will impact upon that inhaler during inspiration
43
Aerodynamic Particle size
Interchange
MODIFIED CONFIGURATIONS FOR questionable, the further the test flow QUALITY
USE AT 60 AND 90 L/MIN rate deviates from 28.3 L/min.
A number of papers published in
In many cases (particularly with low In order to help address these problems, the late 1990s highlighted concerns
resistance DPIs), it is necessary to two modified configurations of ACI are relating to the manufacture and
operate at flow rates greater than available for operating at flow rates of performance of the Andersen Cascade
28.3 L/min, if a pressure drop over the 60 and 90 L/min. These are described Impactor manufactured by Graseby-
inhaler of 4 kPa is to be achieved. in USP Pharmacopoeial Forum Volume Andersen between 1992 and 1998.
28, Number 2, 2002, p. 601-603 and are
Whilst the ACI can be operated at These focused on the choice of
now enshrined in USP 38.
flow rates greater than 28.3 L/min, it material used in their design, their
is important to consider the change In the 60 L/min version, stages 0 and construction, ease of use, accuracy,
in cut-points that will occur for each 7 are removed and replaced with two calibration and the ability to suitably
stage. An empirical equation can additional stages, -0 and -1. Similarly, qualify the impactors prior to use.
be used to calculate these cut-point in the 90 L/min version, stages 0, 6
Because of these criticisms, Copley
changes over the range of and 7 are removed and replaced with
Scientific commenced manufacturing
28.3 100 L/min. However, the three additional stages, -0, -1 and -2.
the Andersen Cascade Impactor using
user should be aware that reduced Changes are also made to the
the latest state-of-the-art production
discrimination between the cut- configuration of the collection plates
techniques .
points will occur as the flow rate is (with and without centre holes).
increased. Furthermore, the validity These techniques ensure that 100%
This results in a new set of cut-points
of the empirical equation becomes of the jets of every stage of every
as per the table below.
Copley impactor conform to the
published critical dimensions for the
Cut-off Diameters at 28.3 60 90 L/min
ACI stated in USP Chapter <601> and
Stage -2 ---- ---- 8.0 microns Ph.Eur. Chapter 2.9.18.
Stage -1 ---- 8.6 6.5 microns
Stage -0 ---- 6.5 5.2 microns The validity of this data is guaranteed
Stage 0 9.0 ---- ---- microns by dimensional verification using
Stage 1 5.8 4.4 3.5 microns the very latest vision inspection
Stage 2 4.7 3.2 2.6 microns technology having a demonstrated
Stage 3 3.3 1.9 1.7 microns optical reproducibility of 1 micron (to a
Stage 4 2.1 1.2 1.0 microns 99% confidence interval).
Stage 5 1.1 0.55 0.22 microns
Stage 6 0.7 0.26 ---- microns
Stage 7 0.4 ---- ---- microns
44
Aerodynamic Particle size
45
Aerodynamic Particle size
MENSURATION, QUALIFICATION For this reason, all cascade impactors The following ancillaries are required
AND SYSTEM SUITABILITY (including induction ports and in addition to the ACI to complete
preseparators) manufactured by a fully operating test system for
Every impactor manufactured by
Copley Scientific are checked at every determining the aerodynamic particle
Copley Scientific is machined to the
stage of manufacture using the very size distribution of MDIs:
same precision tolerances in order to
latest in metrology equipment and
guarantee reproducibility between Mouthpiece Adapter (see
are provided with a mensuration
impactors and to ensure stage Page 92)
certificate and leak test certificate
mensuration.
prior to release. Induction Port (see Page 47)
Stage mensuration replaces the
Vacuum Pump (see Page 93)
need for repetitive calibration using SUMMARY
standardised aerosols and ensures Flow Meter (see Page 90)
Andersen Cascade Impactors
that only impactors conforming to
manufactured by Copley Scientific are: Data Analysis Software (see
specification are used in testing.
Page 86)
Available in aluminium, 316 stainless
In practice, this means that every jet
steel or titanium
on every stage of every impactor must plus the following in order to test DPIs:
Capable of operation at 28.3, 60 or
be individually inspected to ensure
90 L/min Preseparator (see Page 47)
compliance.
Manufactured to USP and Ph.Eur.
Critical Flow Controller (see
critical dimensions
Page 78)
Supplied with full stage mensuration
certificate, certificate of conformity Options:
to USP/Ph.Eur. and leak test
Automation (see Page 105)
certificate
Modified 28.3 and 60 L/min Preseparator Lids (Cat. No. 8421/8422) and Inlet
Cone (Cat. No. 8366) for use with NGI Induction Port Cat. No. 8366
(Prices available on request)
46
Aerodynamic Particle size
Induction Ports
Options
Spare Parts
47
Aerodynamic Particle size
INTRODUCTION
Before the introduction of the NGI, The result, the NGI, was an impactor Supplied with full stage mensuration
the Andersen Cascade Impactor having the following features: report (system suitability)
was the main impactor used by the
Designed by the pharmaceutical Low inter-stage wall losses ensure
pharmaceutical industry.
industry for inhaler testing good drug recovery (mass balance)
Although originally designed for
Meets and exceeds all Ph.Eur. and User friendly design for maximum
microbial sampling, the ACI is a
USP specifications throughput and easy automation
well-established instrument that has
served the industry well. It remains Particle size range: 0.24 11.7 Electrically conductive; unaffected
in widespread use and is expected microns (dependent on flow rate) by static
to do so in the foreseeable future.
Seven stages; five with cut-offs Design and archival calibration
Because of its air sampling origins
between 0.54 and 6.12 microns formally documented and published
however, the ACI does suffer from
at flow rates from 30 to 100 L/min
certain drawbacks and it is not easy to
DESCRIPTION
automate. Excellent stage efficiency, accuracy
and reproducibility The initial design considerations
In developing the Next Generation
concentrated on the number of stages
Impactor, the consortium involved Archivally calibrated flow rate range:
and basic layout. Seven stages were
drew on their extensive experience 30 100 L/min
finally specified to give five with
to come up with a list of musts and
Additional calibration at 15 L/min cut-off diameters in the 0.5 - 5 micron
wants for the new impactor.
for nebuliser applications range and a horizontal planar layout
adopted for ease of operation and
Cut-off diameters at 15 30 60 100 L/min automation.
Stage 3 5.39 3.99 2.82 2.18 microns The air flow passes through the
Stage 4 3.30 2.30 1.66 1.31 microns impactor in a saw tooth pattern.
Particle separation and sizing is
Stage 5 2.08 1.36 0.94 0.72 microns achieved by successively increasing
Stage 6 1.36 0.83 0.55 0.40 microns the velocity of the airstream as it
passes through each by forcing
Stage 7 0.98 0.54 0.34 0.24 microns
it through a series of nozzles
MOC 0.70 0.36 0.14 0.07 microns containing progressively reducing
jet diameters.
48
Aerodynamic Particle size
49
Aerodynamic
Title Particle size
Preseparator
(internal view)
Micro-orifice
Stage1 Stage 3 Stage 5 Stage 7 collector (MOC)
1hole 24 holes 152 holes 630 holes
Location pin
Location pin
Schematic of Seal Body showing orientation of the
recess
various stages
Bottom frame with
cup tray in place
50
Aerodynamic Particle size
Spare Parts
5208 Collection Cup Tray
5209 Set of 8 Collection Cups (2 Large, 6 Small)
5245 Welded Cup Tray Manifold
51
Aerodynamic Particle size
52
Aerodynamic Particle size
53
Aerodynamic Particle size
The value of the Glass Twin Impinger, in position and is supplied with a the particle cut-off is 6.4 microns.
particularly with respect to routine quality mensuration certificate confirming Particles smaller than 6.4 microns pass
control applications, is recognised by that the critical dimensions conform to into the lower impingement chamber.
its retention as Apparatus A in Ph.Eur. those stated in Ph.Eur.
Prior to testing, 7 mL of solvent is
2.9.18.
It operates on the principle of typically dispensed into the upper
Its usage is restricted to the assessment liquid impingement to divide the impingement chamber and 30 mL to
of nebulisers, MDIs and such DPIs where dose emitted from the inhaler into the lower impingement chamber. After
it can be demonstrated that a flow rate of respirable and non-respirable portions. the test is complete, the active drug
60 (+/- 5) L/min is suitable. collected in the lower impingement
The non-respirable dose impacts on
chamber is assayed and expressed as
Developed at GSKs laboratories in Ware, the oropharynx and is subsequently
a respirable fraction (or percentage) of
UK, the Glass Twin Impinger is relatively swallowed. This is considered as the
the delivered dose.
simple and easy to use, and assemble. back of the glass throat and the upper
impingement chamber (collectively The Glass Impinger requires a special
The major advantage is that it is
described as Stage 1). The remaining mouthpiece adapter, a vacuum pump
manufactured solely from glass so that it
respirable dose penetrating the lungs and a flow meter to complete the
is not prone to corrosion in the same way
is collected in the lower impingement system.
as conventional metallic impactors.
chamber (Stage 2).
A special modification for the
The Glass Twin Impinger comes
The upper impingement chamber measurement of the particle size of
complete with stainless steel base plate,
is designed such that at a flow rate nasal sprays according to Aaiche and
stand, clamp and boss head in addition
of 60 L/min through the impinger, Beyssac is also available as an option.
to plastic clips to retain the glass parts
Spare Parts
54
Aerodynamic Particle size
egnahcretnI
egnahcretnI
used in conjunction with pressurised ACI based System (with Pump and Breath Actuation Controller
metered dose inhalers (pMDIs) to Model BAC 2000)
evlaV lortnoC wolF
yaw-2/trop-2
overcome the problems of poor evlav dionelos
3-way valve
2-port/2-way solenoid valve 2-port/2-way soleno
55
Aerodynamic Particle size
Traditionally, nasal preparations have At the same time however, most sprays
been used for the local administration deliver a proportion (typically <5%) of
of anti-histamines, decongestants, and fine droplets in the <10 micron range.
steroids in order to alleviate cold or
It is important to quantify this fine is preferred. In the case of nasal
allergy symptoms and nasal congestion.
particle dose since it can penetrate aerosols, a 1 litre chamber is used to
More recently, attention has focused on beyond the nasal tract and into the maximise drug deposition below the
two other areas: lower respiratory tract or lungs, which top stage of the impactor.
may prove undesirable.
a) The potential rapid drug absorption Each of the chambers contains an
into the systemic circulation provided In its Draft Guidance Bioavailability entry port at approx. 30 degrees
by the turbinates and lymphoid tissues and Bioequivalence Studies for to the outlet port for insertion of
located at the back of the nasal cavity. Nasal Aerosols and Nasal Sprays for the nasal spray or aerosol. Special
This is already in use in a number of Local Action of April 2003, the FDA nosepiece adapters are available
areas, e.g. migraine and pain relief, recognises the nature of this problem for the entry port to accommodate
osteoporosis, vaccines, etc., and and recommends the use of a cascade powder, spray and aerosol based
impactor in conjunction with a high devices.
b) The potential of the Nose to Brain
volume expansion chamber to measure
entry to the central nervous system Adapters are also available to connect
the amount of drug in small particles or
presented by the olfactory region at the the outlet port of the expansion
droplets in respect of nasal sprays
top of the nasal cavity for the treatment, chamber to the inlet cone of the
and the particle/droplet size
for example, of diseases of aging such as Andersen Cascade Impactor (ACI).
distribution in the case of nasal
Alzheimers Disease, etc.
aerosols. The adapters are available in
Conventional nasal technologies fall into aluminium, 316 stainless steel or
The purpose of this exercise is to
three main categories: titanium and have internal dimensions
quantify the amount of drug present
similar to those at the outlet of the
Metered Spray Pumps (Aqueous in the form of particles or droplets that
Universal Induction Port typically
based) are less than 10 microns, with a view to
used for orally administered inhaled
Propellant based Nasal Aerosols (MDIs) predicting their possible deposition
products.
Powder based Nasal Devices in the lungs.
Each adapter is supplied with a
Nasal sprays typically produce droplets In accordance with the draft guidance,
clamping device which allows the
in the range 20-200 microns which is Copley Scientific now offers a range
glass expansion chamber to be easily
outside the effective range of inertial of glass expansion chambers to meet
removed from the impactor for assay.
impactors. For this reason, the droplet these requirements.
size distribution of nasal sprays and
In the case of nasal sprays, a 2 litre
aerosols is normally determined by
or larger (5 litre) expansion chamber
means of laser diffraction.
56
Aerodynamic Particle size
During use, the clamp provides an As a general rule, the potential areas
airtight seal between the expansion of interest may be divided into three
chamber and the adapter through groups:
the use of an FDA approved silicone
1. Those particles >10 microns and
rubber O-ring incorporated into the
hence retained in the intranasal
neck of the adapter.
passageways.
A special adapter and clamp are also 2. Those particles between 5
available for the Next Generation and 10 microns destined for the
Impactor. gastrointestinal tract.
3. Those particles <5 microns
The majority of nasal products are
potentially capable of depositing The following ancillaries are required
designed to generate droplets/
in the lungs. in addition to the items below to
particles having a mass median
complete a fully operating test system
aerodynamic diameter (MMAD) of After validation, it may therefore be
for determining the aerodynamic
greater than 10 to 20 microns. This appropriate to use a reduced impactor
particle size distribution of nasal sprays
is to increase nasal deposition and stack (e.g. Stage 0 = >9 microns, Stage
and aerosols:
minimise deposition in the lungs. 2 = 4.7 to 9 microns, Stage F = 0.4
4.7 microns of an Andersen Cascade Cascade Impactor (see Page 43
Cascade Impactors, on the other
Impactor at 28.3 L/min). for ACI, Page 48 for NGI)
hand, are designed to capture
particles in the range 0 to 10 microns. In these cases, we would recommend Vacuum Pump (see Page 93)
It follows that the majority of particles the use of the Quick Clamp (see Page
Flow Meter (see Page 90)
discharged from a nasal product will 47) which is designed to allow clamping
be deposited on the upper stages of of the ACI with a reduced number of
the impactor concerned. stages, or the special versions of the
ACI and NGI such as the FSA or FSI
classified under AIM (see Page 62).
Nasal Adapters
8957 Nasal Aerosol Nosepiece Adapter for Expansion Chamber Inlet
8958 Tooling Charge for above (per nasal aerosol device)
8959 Nasal Spray Nosepiece Adapter for Expansion Chamber Inlet
8960 Tooling Charge for above (per nasal spray device)
8956 Expansion Chamber to Flow Meter Adapter
* Please specify Aluminium (A), 316 Stainless Steel (S) or
Metered Nasal Propellent based Titanium (T) when placing your order.
Spray Pump Nasal Aerosol
57
Aerodynamic Particle size
NGI Cooler
In 2006, the European Medicines The recommended flow rate of It is believed that for devices such as
Agency (EMA) issued a new 15 L/min employed in the APSD testing nebulisers, which deliver the active as
Guideline on the Pharmaceutical of nebulisers is lower than that of other an aerosolised solution, evaporation
Quality of Inhalation and Nasal OINDPs in order to better simulate caused by heat from the impactor can
Products in which they included the normal tidal breathing conditions be a problem.
regulatory guidance on the drug employed in their in vivo use.
The ensuing loss of solvent reduces
aspects of nebulisers on the
For this reason, an EPAG (European droplet size, producing artificially low
grounds that the safety and efficacy
Pharmaceutical Aerosol Group) led particle size measurements and thus
of nebulisers was dependent on the
initiative was launched in 2002 to compromising the integrity of the
nebuliser/drug combination and not
provide an extension to the archival resulting data. Cooling the impactor
just on the nebuliser alone.
calibration of the Next Generation to approximately 5 degrees Celsius
As a result of the EMA initiative Impactor (NGI) to 15 L/min. is the recommended method for
and recognising the lack of suitable overcoming this problem.
The results published in 2004 indicated
test methods for nebulisers, the
that the NGI could be used to meet the The NGI Cooler comfortably
Pharmacopoeias have in turn
requirements of the future standard, accommodates the NGI, either
introduced a new Chapter on
albeit without the preseparator and by closed or open, allowing testing in a
Preparations for Nebulisation:
using the internal filter holder to collect temperature controlled environment.
Characterisation (see Ph.Eur.
any fine droplets less than 0.98 microns. Rapid cooling ensures that test
Chapter 2.9.44 and USP Chapter
Cup coating is not normally required. temperatures, user adjustable as low
<1601>).
as 5 degrees C, are reached in less
This produces an impactor with seven
It is these proposals that form the than 5 minutes; temperature stability
stages having cut-off diameters at 14.1,
basis for the tests specified in is to within +/- 1.5 degrees C. Large
8.61, 5.39, 3.30, 2.08, 1.36 and 0.98
Annex C of the new ISO front and rear opening doors allow for
microns respectively at 15 L/min.
27427:2013 requirements (based easy access with special access ports
on the European Standard EN to accommodate the nebuliser and
13544-1:2007) for the safety, vacuum pump connections.
performance and testing for
general purpose nebulising systems Internal
Filter 3-Way Valve
intended for continuous or breath-
Holder
actuated delivery of liquids in an
aerosol form, to humans through the
respiratory system, and the tests
and equipment outlined below.
58
Aerodynamic Particle size
PROCEDURE This provides near instantaneous 7. Dismantle the impactor and, using
starting and stopping of the air flow a suitable method, determine
Determine the sampling time
during testing and has both delay the mass of active collected in the
(To) by balancing stage overload
and inhaled time functions without induction port, on each stage
against analytical sensitivity. The
compromising the integrity of the and on the final filter.
time chosen should be sufficient to
induction port and/or test set-up. 8. Collect and present the data as
ensure adequate sample is collected
described in the monograph.
for analysis without overloading the Proceed as follows:
collection cups concerned. The following ancillaries are required
1. Prepare the nebuliser for operation
in addition to the items below to
Set the flow rate to 15 L/min by in the normal manner.
complete a fully operating test system
proceeding as follows: 2. Switch on the vacuum pump, the
for determining the aerodynamic
nebuliser and the cooler (if
1. Assemble the test system as particle size distribution of nebulisers:
required) and allow to stabilise.
instructed.
3. Ensure that the environmental Vacuum Pump (see Page 93)
2. Attach a suitable flow meter to the
conditions are as stated.
inlet of the induction port. Breath Actuation Controller (see
4. Open the 3-way valve** so that
3. Switch on the vacuum pump. Page 84)**
the flow passes through the
4. Open the 3-way valve** connecting
impactor. Flow Meter (see Page 90)
the vacuum pump to the impactor.
5. Sample for the predetermined
5. Adjust the flow control valve on the Data Analysis
time (To).
vacuum pump until the flow rate Software (see Page 86)
6. Close the 3-way valve and switch
through the system is steady at
off the nebuliser and vacuum pump.
15 L/min +/- 5%.
6. Switch off the pump, close the
3-way valve** and remove the flow Cat. No. Description
meter and adapter. Impactor
* Note: If method development 5201 Next Generation Impactor (NGI)
indicates that the nebulised aerosol is 5203 NGI Induction Port
significantly affected by evaporation 5206 Internal Filter Holder
then use the NGI Cooler to cool the 5240 Box of 100 Filters (for Internal/External Filter Holder above)
impactor before and during testing.
5003 Mouthpiece Adapter
** Alternatively, use a timer controlled 5004 Tooling Charge for above
two way solenoid valve such as the 5006 3-Way Valve**
Breath Actuation Controller Model NGI Cooler
BAC 2000 between the impactor and
the pump (see description on Page 5009 NGI Cooler (option)
84). 5011 NGI Cooler Qualification Documentation
5012 NGI Cooler Qualificaton Tools
5013 Re-calibration of NGI Cooler Qualification Tools
59
ABBREVIATED IMPACTORS (AIM)
60
ABBREVIATED IMPACTORS (AIM)
AIM IN THE QC ENVIRONMENT Typically the APSDs of inhaled Although EDA can be applied to
products are in the form of a full-resolution impactor testing, its
The concept of Abbreviated Impactor
Normal (or Gaussian) Distribution true value comes from combining
Measurement (AIM), as typified by the
centred around the Mass Median it with AIM, since only a reduced
Glass Impinger on Page 54 (still available
Aerodynamic Diameter (MMAD). It number of impactor stages are
as Apparatus A in Ph.Eur.) and the Fisons
is therefore possible to determine required, speeding up throughput
Single Stage Metal Impactor described
even subtle changes in the APSD by and further reducing analytical error.
in earlier versions of the European
measuring the following: Full-resolution impactor testing is
Pharmacopoeia (until 2002) is not new.
then reserved for Out-of-Specification
However, the initiative in recent years 1. Impactor Sized Mass (ISM)
(OOS) investigations.
started with abbreviated versions of the which is considered the sum of the
Andersen Cascade Impactor (ACI). drug mass deposited on the filter In the diagram below, the AIM-QC
and all impactor stages except the model shows how abbreviating the
The concept is founded on the basis that
uppermost. This metric indicates ACI to just 2 stages and a filter, with
once the full Aerodynamic Particle Size
any shift in the amplitude of the the central stage (Stage X) selected
Distribution (APSD) profile of the product
APSD. to have a cut-off diameter close to
has been established in development
the product MMAD, allows the EDA
using a full-resolution cascade impactor 2. Ratio of Large Particle Mass to
metrics of ISM and LPM/SPM to be
(and the process validated) then for Small Particle Mass (LPM/SPM)
easily determined.
product batch release testing and QC which is considered to be the ISM
applications, it is possible to use simpler split into two fractions on either side The table on Page 44 indicates which
but highly sensitive metrics, solely of the MMAD: LPM greater than the ACI stage can be used for Stage X
to determine if the product is fit for MMAD and SPM smaller than the depending on the test flow rate and
purpose. This is known as Efficient Data MMAD. This ratio indicates any shift product MMAD (as determined from
Analysis (EDA). in the central tendency of the APSD. full-resolution impactor testing).
EPM
ECDs at Size ISM LPM/
28.3 I.P. Bands SPM I.P. I.P.
>5.0 microns
L/min (microns)
CPM
<1.0 microns
EPM
Note: Stage X is that stage having a cut-point closest to the MMAD of the aerosol,
as measured by a full resolution impactor
Adapted from: Mitchell, J.P. et al. Relative Precision of Inhaler Aerodynamic Particle Size Distributon (APSD) Metrics by Full Resolution and Abbreviated Andersen
Cascade Impactors (ACIs): Part 1., AAPS PharmSciTechnol., 2010, 11(2): 843-851
61
ABBREVIATED IMPACTORS (AIM)
AIM IN THE R&D ENVIRONMENT 1. Coarse Particle Mass (CPM) In this case, the AIM-HRT model
That portion of the aerosol considered shown in the diagram on Page 61
Abbreviated Impactor Measurement
to be too large to be inhaled (usually shows how abbreviating the ACI to 2
(AIM) has also been suggested as a
considered to be >5 microns). stages, plus a filter and a spacer can
possible useful tool in R&D for the
be used to determine the CPM, FPM
fast screening of new formulations in 2. Fine Particle Mass (FPM)
and EPM.
product development. That portion between 5 and 1 micron,
usually considered likely to deposit The selected stages have cut-off
Abbreviated impactors have
deep into the lung and hence be diameters equal or close to 5 and
three main advantages over their
therapeutically effective. 1 microns. The Spacer provides
more conventional multi-stage
additional dead space prior to the
counterparts: 3. Extra-fine Particle Mass (EPM)
first impaction stage, equivalent to a
That portion below 1 micron, usually
a) speed of throughput (this allows full resolution impactor. This has been
considered to be too small to deposit
more samples to be measured shown to be significant for improving
in the lung and therefore exhaled.
in a given time frame) the equivalence between AIM and
b) less complicated so less prone to full-resolution measurements for
method and analyst error and ethanol based
c) far easier to automate. MDIs.
62
ABBREVIATED IMPACTORS (AIM)
ABBREVIATED IMPACTOR
MEASUREMENT (AIM) Cat. No. Description
FSA-QC with Stage X cut-off diameter close to product MMAD
FAST SCREENING ANDERSEN
(FSA) 8341 FSA-QC - 28.3 L/min (Stages 0, X and F)*
8342 FSA-QC - 60.0 L/min (Stages -1, X and F)*
The Fast Screening Andersen (FSA) 8343 FSA-QC - 90.0 L/min (Stages -2A, X and F)*
is an AIM version of the standard
Andersen Cascade Impactor (ACI) FSA-HRT with cut-off diameters of 5.0 and 1.0 or 4.7 and 1.1 microns
suitably modified to provide a reduced
stack plus filter (F) suitable for either: 8344 FSA-HRT - 28.3 L/min (Spacer, Stages 5.0 and 1.0 micron, and F)*
8345 FSA-HRT - 28.3 L/min (Spacer, Stages 2, 5 and F)*
a) Quality Control (FSA-QC) or 8346 FSA-HRT - 60.0 L/min (Spacer, Stages 1, 4 and F)*
b) Product Development (FSA-HRT). 8347 FSA-HRT - 90.0 L/min (Spacer, Stages -0, 3 and F)*
The principles of each type are
described on Pages 61 and 62. Induction Ports
In the FSA-QC, Stages 0 (or -1, or -2A) 8501 Universal Induction Port (Standard)*
and F are used in conjunction with a 8510 Universal Induction Port (One-piece 316 Stainless Steel)
Stage X, having a cut-off diameter as
close as possible to the Mass Median Preseparators for testing DPIs
Aerodynamic Diameter (MMAD) of 8401 28.3 L/min Preseparator*
the aerosol, as determined during full 8420 60 L/min Preseparator*
resolution cascade impactor testing. 8420-90 90 L/min Preseparator*
In the FSA-HRT stages with cut-off
diameters are available at 5.0 and 1.0 Spare Parts
microns for MDI applications 8367-I Stage 5.0 micron cut-off @ 28.3 L/min*
at 28.3 L/min. Also, for this flow 8368 Stage 1.0 micron cut-off @ 28.3 L/min*
rate and higher flow rates (60 and 8371 FSA Spacer Stage*
90 L/min) stages having traditional 8334 Complete Set of 7 FSA Silicone Rubber O-Rings
ACI cut-points of 4.7 and 1.1 microns 8335 Set of 2 Stainless Steel Collection Plates (28.3 L/min)
are available, primarily for DPI 8336 Set of 2 Stainless Steel Collection Plates (60 or 90 L/min)
applications. 8316 Box of 100 Glass Fibre Filters (81 mm)
Preseparators can be used, where 8308A Set of 3 Shortened Spring Clamps - 4 Stage
appropriate, to remove large powder 8308B Set of 3 Shortened Spring Clamps - 3 Stage
boluses, as in the case of most DPIs. * Please specify Aluminium (A), 316 Stainless Steel (S)
or Titanium (T) when placing your order.
63
ABBREVIATED IMPACTORS (AIM)
Stage Cut-off Diameters for the Next Generation Impactor at Different Flow Rates
Flow Rate (L/min)
Stage 15 30 40 50 60 70 80 90 100
1 14.10 11.72 10.03 8.89 8.06 7.42 6.90 6.48 6.12
2 8.61 6.40 5.51 4.90 4.46 4.12 3.84 3.61 3.42
3 5.39 3.99 3.45 3.09 2.82 2.61 2.44 2.30 2.18
4 3.30 2.30 2.01 1.81 1.66 1.54 1.45 1.37 1.31
5 2.08 1.36 1.17 1.04 0.94 0.87 0.81 0.76 0.72
6 1.36 0.83 0.70 0.61 0.55 0.50 0.46 0.43 0.40
7 0.98 0.54 0.45 0.38 0.34 0.31 0.28 0.26 0.24
Determined from: Virgil A. Marple et al. Next Generation Impactor (A New Impactor for Pharmaceutical Inhaler Testing).
Part II: Archival Calibration: Journal of Aerosol Medicine; Volume 16, Number 3, 2003; 301-324 and
Part III: Extension of Archival Calibration to 15 L/min: Journal of Aerosol Medicine; Volume 17, Number 4, 2004; 335-343
64
ABBREVIATED ImPACTORS (AIM)
Interchangeable Inserts
Filter Holder
ABBREVIATED IMPACTOR
MEASUREMENT (AIM)
Fine Fraction Collector for users that already have NGI Preseparator
65
IMPROVED IVIV CORRELATION
IVIVC System for MDIs with VHC with Alberta Idealised Throat (Child Version), Mixing Inlet, NGI and Breathing Simulator BRS 3000
INTRODUCTION A child for example with chronic asthma 1. Replacing the existing
will exhibit a vastly different breathing Ph.Eur./USP Induction Port with
Accelerating time to market and
profile from an otherwise healthy adult an age-appropriate mouth/throat
adopting better practice in the
using the device for systemic purposes. model having a more realistic
development, manufacture and quality
human-like geometry.
assurance of medicines are ongoing One strategy for improving the
goals for the pharmaceutical industry. significance of cascade impaction data The Ph.Eur./USP Induction Port
is to modify the test set-up in order to (Throat) normally used to interface the
Better in vitro - in vivo correlation
mimic the in vivo drug delivery process device with the impactor has a simple,
(IVIVC) has long been an industry aim,
more closely. well defined geometry developed with
but the current climate clearly adds
testing standardisation in mind.
impetus to the desire for progress. Two factors that have been identified
as being critical to the improvement It is easy to manufacture and gives
Inhaled product development in
process are: consistent performance, essential
particular presents some unique
for QC testing. However, it is widely
challenges in this respect.
accepted that this port does not
NGI with Alberta Idealised Throat
The difficulty of precisely correlating provide an accurate representation of
(Child Version) and MDI with Valved
drug deposition behaviour with clinical Holding Chamber (VHC) what happens in the upper respiratory
efficacy, the impact of patient-to- tract in vivo in that it consistently
patient variability and the complex under-predicts the amount of active
interaction between formulation captured in this area.
and device, all complicate the
development process.
66
IMPROVED IVIV CORRELATION
67
IMPROVED IVIV CORRELATION
Alberta Idealised
Throat (open)
One way to accurately simulate the Developed with testing standardisation The Alberta Idealised Throat
deposition of orally inhaled drug in mind, it has a simple well defined (AIT) was developed as a result of
products (OIDPs) in the throat is to use geometry that lends itself to high extensive research into typical patient
an anatomically correct human throat precision manufacture and the populations including information
cast. consistent performance demanded in provided by CT and MRI scans, direct
product QC testing. visual observation of living subjects
The major drawback is that the
and data in the archival literature.
geometry represented by such a cast is Unfortunately, these benefits come at
The throat has a standardised, highly
that of a single human subject. the cost of in vivo correlation. Indeed,
reproducible, human like geometry
whilst the induction port is ideal for QC
Experimental work has shown offering robust performance
applications, in practice, it has been
significant differences in deposition independent of flow rate.
found to significantly underestimate
behaviour between various throat casts,
the actual amount of active found in Its smooth, more uniform internal
attributable to inter-subject variability in
the throat in vivo. geometry has been specifically
the geometry of the mouth and throat.
designed to make drug recovery
One method of improving in vitro -
Arguably, the Ph.Eur./USP induction quick and simple in comparison with a
in vivo correlation is to replace the
port routinely used in testing represents human throat cast. Quick release clips
standard Ph.Eur./USP Induction Port
the opposite approach in inlet design make for easy internal access.
(Throat) normally used in the testing
to that of the cast.
of inhalers with a throat having more Two versions are available
human-like characteristics. corresponding to adult and child
(6-14 years old range)
For more than a decade, researchers
geometries respectively.
at the Aerosol Research
Laboratory at the
University of Alberta,
Canada, have been working to
develop a more suitable
representation of the mouth-throat
for routine cascade impactor testing.
68
IMPROVED IVIV CORRELATION
IVIVC System for DPIs with Alberta Idealised Throat, Mixing Inlet, NGI and Breathing Simulator BRS 3000
OPTIMISATION
The test set-up shown above illustrates Finally, the AIT addresses widely
how new equipment for in vitro recognised limitations of the standard
testing is being exploited to optimise USP induction port, which does not
data gathering for demonstrating provide a particularly accurate in
bioequivalence in a DPI. vitro realisation of aerosol transport
through the upper respiratory tract.
There are three pieces of equipment
Part way between a human throat cast
present that are routinely absent from
and the simple right angled tubular
the standard set-up: a breathing
design of the USP induction port, the
simulator; an Alberta Idealised
AIT produces data that are
Throat (AIT) (in place of the standard
more representative of
USP induction port) and a Mixing
measured in vivo
Inlet.
behaviour, thereby
It is worth looking in detail at exactly supporting the robust
what each element contributes. demonstration of bioequivalence.
Furthermore it ensures that
The Mixing Inlet decouples the flow
the APSD measurement obtained
profile applied across the device from
via cascade impaction only occurs on
the flow conditions applied in the
the portion of the aerosol that would
cascade impactor.
likely enter the lungs. Alberta Idealised Throat (Child Version)
It allows the application of a patient-
relevant breathing profile across the
Cat. No. Description
DPI while at the same time enabling
the cascade impactor to work at the 8511 Alberta Idealised Throat (AIT) in Aluminium (Adult Version)
constant flow rate required for 8514 Flow Meter To Alberta Idealised Throat Adapter (Adult Version)
accurate APSD measurement. 8515 Mouthpiece Adapter for Alberta Idealised Throat (Adult Version)
5004 Tooling Charge for above (Adult Version)
The Breathing Simulator enables
8516 Spare Silicone Seal for Alberta Idealised Throat (Adult Version)
exploration of the impact of different
8518 Leak Test Inlet Cap and Outlet Adapter (Adult Version)
breathing profiles. In bioequivalance
testing it therefore allows the robust 8530 Alberta Idealised Throat (AIT) in Aluminium (Child Version)
demonstration of equivalent drug 8531 Flow Meter to Alberta Idealised Throat Adapter (Child Version)
delivery performance across a range of 5003 Mouthpiece Adapter for Alberta Idealised Throat (Child Version)
conditions that represent the variability 5004 Tooling Charge for above (Child Version)
associated with a target user group. 8532 Spare Silicone Seal for AIT (Child Version)
8533 Leak Test Inlet Cap and Outlet Adapter (Child Version)
The flexibility to fully scope variability
is far greater than with the standard 8512 Alberta Idealised Throat to ACI/FSA Adapter (Adult & Child)
pharmacopoeial test set-up. 8513 Alberta Idealised Throat to NGI/FSI Adapter (Adult & Child)
69
IMPROVED IVIV CORRELATION
MIXING INLET
The cut-off diameters generated by For this reason, over the years,
each stage of any cascade impactor are there have been various attempts
dependent on a steady, fixed flow of air to link cascade impactors directly
passing through it. to breathing simulators in order
to reproduce the actual clinical
In contrast, the in vivo inhalation profiles NGI Mixing Inlet
condition more closely.
of breathing cycles generated by
patients produce a continually varying Any such system must be capable DESCRIPTION
flow rate far removed from the fixed, of varying the flow rate through The Mixing Inlet fits between the USP
steady-state flow rates employed in in the inhaler whilst ensuring that the Induction Port (or Alberta Idealised
vitro testing. aerosol generated is sampled at a Throat) and the inlet of the impactor
fixed rate through the impactor. used to carry out the test.
Another problem can be when It is designed to permit the cascade
the test flow rate applied to the impactor to be operated at a constant
inhaler is lower than the minimum flow rate (e.g. 100 L/min), whilst
calibrated flow rate of the impactor. allowing a lower fixed or variable rate,
This is common, for example, in such as a breathing pattern generated
paediatric studies where the user by a breath simulator, to pass through
wishes to simulate the flow rate the inhaler itself.
typical in vivo of a minor, say Supplementary (or make-up) air
10 L/min, whereas the impactor is provided from a compressed air
requires a flow rate of 28.3 L/min in source to the inlet port on the side
order to operate within its calibrated of the Mixing Inlet using a Flow
range. Control Manifold with 6 mm or 1/4
compressed air fittings. A flow control
valve contained within the manifold
is used to match the steady-state
Next Generation Impactor (NGI)
flow rate exiting the impactor. This
fitted with Mixing Inlet
balances the flow and ensures that the
flow rate at the inlet to the Induction
Port is zero.
70
IMPROVED IVIV CORRELATION
Mixing Inlet
Suitable breathing simulators can 8323 Set of 5 O-Rings for Mixing Inlet
be found on Pages 73-77. 8338 Flow Control Manifold for Mixing Inlet (6 mm)
8339 Flow Control Manifold for Mixing Inlet (1/4)
71
ancillaries
Ancillaries
Introduction The Critical Flow Controller is copy format. It will accept data from
designed to generate a standardised the ACI, NGI, MSLI and/or MMI.
square-wave breath profile suitable Provision is made to customise the
for the routine testing of passive data options to individual needs.
breath activated devices such as
Flow rate is a critical parameter in the
Dry Powder Inhalers, where the
in vitro testing of OINDPs. Copley
delivered and fine particle dose
Scientific offers two Flow Meters
of the device is dependent on
with the required range and accuracy
the strength and duration of the
to perform this task, one based on
patients inspiration.
differential pressure and the other on
thermal mass measurement methods.
The Breath Actuation
Both units will give the same results
Controller is an electrically
providing they are calibrated and
operated timer controlled
operated correctly.
two-way valve specifically
designed for testing Breath Actuated No inhaler testing system would be
This section describes the ancillaries (or Breath Operated) MDIs, Spacers complete without the Mouthpiece
required in addition to the Dosage and VHCs used with MDIs and Adapters, Tubing and Quick
Unit Sampling Apparatus (DUSA) and Nebulisers to USP <1601> and Ph.Eur. Release Connectors designed to
Cascade Impactor to make up a fully 2.9.44. link the various components of the
operating test system for determining system together.
The Data Analysis function is
the Delivered Dose Uniformity and provided by CITDAS (Copley Inhaler Finally, at the heart of every inhaler
Aerodynamic Particle Size Distribution Testing Data Analysis Software), a testing system is the Pump. The
of orally inhaled and nasal drug proven third generation software Pharmacopoeias are careful to point
products (OINDPs). program designed specifically for out that a vacuum pump with excess
Breathing simulators are increasingly the simple and rapid processing capacity must be selected in order to
used in testing OINDPs to replace of impactor drug deposition data draw air, at the designated volumetric
existing constant flow rate conditions according to pharmacopoeial flow rate through the system and,
with breathing profiles more requirements. It has an existing in the case of Dry Powder Inhalers to
representative of conditions in vivo. database of more than 200 users. generate sonic flow.
Copley Scientific offer a choice of three Based on over 12 years of experience, Copley Scientific offers a choice
Breath Simulators covering the range CITDAS can be installed and up and of three pumps dependent on the
of breathing patterns to be found running in minutes it requires no set-up concerned and the capacity
in neonatal, infant, child and adult specialist IT knowledge to install required.
physiologies. and does not require 21 CFR 11
because the data output is in hard
72
ancillaries
Breathing Simulators
BRS 2000 and 3000
BREATHING SIMULATORS
Breathing Simulators, instruments regulatory testing with a unit capable simulators to measure the effects
that generate an inhalation and/or of producing breath profiles more of different profiles, flow rates and
exhalation profile that micmics that of a representative of conditions in vivo. breathing techniques during their
human subject, have become a routine development.
A pMDI, when used without a
feature of orally inhaled product (OIP) Such an approach is supported by
spacer or VHC, actively delivers the
testing. Their use is two fold: the QbD (Quality by Design) strategy
drug directly to the patient using
1. Regulatory a propellant. With these devices, outlined in ICH Q8 which relies on
inhalation must be coordinated with scoping the potential impact of
To measure the delivered dose of drug any variability that may arise from,
the actuation to ensure success, but
(DDU) from: for example, differences in patient
the shape and characteristics of the
a) Nebulisers as per USP <1601> and breathing profile employed by the physiology or technique.
Ph.Eur. 2.9.44 and patient in their use are unlikely to
Copley Scientific produce a range
have much effect on the aerodynamic
b) Pressurised metered dose inhalers of breathing simulators from the
particle size distribution (APSD) of
(pMDIs) when used in conjunction with BRS 1100, an inexpensive unit
the delivered aerosol and/or the
spacers and valved holding chambers producing sinusoidal wave forms
effectiveness of delivery.
as per the proposed USP Chapter and designed specifically to meet
<1602> This is not the case for dry powder the regulatory requirements of USP
inhalers (DPIs), nebulisers or pMDIs and Ph.Eur., to the BRS 3000, a
in accordance with the appropriate versatile, fully computer controlled
used with spacers / valve holding
regulations. Breathing Simulator designed for
chambers. Here the breathing profile
2. Better in vitro - in vivo of the patient directly influences the both regulatory and research and
correlation (IVIVC) efficiency of drug delivery. development applications, using more
varied, patient derived, breathing
To replace the fixed flow rate vacuum For this reason, more laboratories
profiles.
pump normally employed for are turning to the use of breathing
Volume (manually adjust): 0 to 800 mL Volume (computer controlled): 0 to 900 mL Volume (computer controlled): 0 to 5000 mL
Frequency: 12 - 40 bpm Frequency: 0+ (upper defined by acceleration limit) Frequency: 0+ (upper defined by acceleration limit)
I:E Ratio: 1:1, 1:2 or 1:3 I:E Ratio: variable I:E Ratio: variable
Waveforms: Sinusoidal Waveforms: Sinusoidal, Square, Triangular, User Waveforms: Sinusoidal, Square, Triangular, User
defined (flow vs time) defined (flow vs time)
Tidal breathing: inhalation & exhalation Tidal breathing profiles: inhalation and exhalation Tidal breathing profiles: Inhalation and exhalation
Select start on inhalation or exhalation Inhalation only profiles Inhalation only profiles
stroke User defined profiles (flow vs time) User defined profiles (flow vs time)
User Interface: Keypad & 4-line display User Interface: Embedded computer (Windows XP) User Interface: Embedded computer (Windows XP)
Uses: Uses: Uses:
Testing Nebulisers (Ph.Eur. 2.9.44 and Testing Nebulisers (Ph.Eur. 2.9.44 and USP <1601>) Limited testing of Nebulisers (Ph.Eur. 2.9.44 and
USP <1601>) Testing MDIs with Spacers/VHCs (USP <1602>) USP <1601>) and MDIs with Spacers/VHCs-
Testing MDIs with Spacers/VHCs (USP (USP <1602>)
<1602>) Improving IVIVCs for MDIs and DPIs:
- With DUSA for MDI/DPI (Dose Uniformity)
- With Impactor and Mixing Inlet (APSD)
73
ancillaries
The Breathing Simulator Model European Pharmacopoeia Chapter Tidal volume of 0 - 800 mL
BRS 1100 is a microprocessor 2.9.44 Preparations for Nebulisation: (155 to 770 mL certified)
controlled instrument which was Characterisation and USP <1601>
Frequency adjustable between 12
designed specifically for generating Products for Nebulization:
and 40 breaths per minute
the neonatal, infant, child and adult Characterization.
breathing patterns required for the Sinusoidal waveform
It can also be used to generate the
dose uniformity testing of nebulisers,
profiles required in the new USP Inhalation/Exhalation Ratio (I:E
in accordance with ISO 27427:2010
proposed Chapter <1602> for testing Ratio) of 1:1, 1:2 or 1:3
Anaesthetic & Respiratory Equipment -
Spacers and Valve Holding Chambers
Nebulising Systems and components, Selectable start position (inhalation
used with Inhalation Aerosols.
or exhalation) for testing spacers/
The BRS 1100 has the following features: VHCs
74
ancillaries
Qualification KIt
c) the I/E Ratio, and I:E Ratio 1:2 1:2 1:1 1:2 1:3 1:3
d) the start position (inhalation
or exhalation) and then select Run USP <1601> - - 3 3 3 3
Method and OK.
75
ancillaries
The Breathing Simulator Model Standard breathing patterns can be Selecting Start activates the breathing
BRS 2000 is an advanced computer defined by editing the following cycle programme. During operation,
controlled breathing simulator, with parameters: the current position within the cycle
up to 900 mL volume, suitable for is indicated on screen by a moving
Selected Pattern: square, sinusoidal
the testing of Nebulisers and Spacers red dot located on the inhalation/
or triangular
and Valved Holding Chambers (VHCs) exhalation curve.
Tidal Volume: 0 - 900 mL
used with Metered-Dose Inhalers
(155 to 770 mL certified) The BRS 2000 compensates for test
(MDIs)
Duration of inhalation (in seconds) equipment induced flow resistance
The BRS 2000 has been specifically Delay after inhalation (in seconds) experienced at the inlet, by adjusting
designed to generate all of the Duration of exhalation (in seconds) power to the motor controlling the
breathing profiles used in measuring Delay after exhalation (in seconds) piston/cylinder arrangement. If the flow
the drug delivery rate and total drug Number of Breathing Cycles line becomes blocked, the BRS 2000
delivered of Nebulisers according to will automatically abort the test.
The in-built software automatically
Ph.Eur. 2.9.44 and USP <1601> (see
calculates the: The inlet for connection to the test
Page 24), namely neonate, infant,
apparatus concerned is fully adjustable
child and adult. Duration of the test
in terms of height and angle.
Breathing Frequency (bpm)
It will also generate the neonate,
Inhalation / Exhalation (I:E) Ratio (%) The BRS 2000 measures 750 mm (W)
infant, child, adult 1 and adult 2
and displays all of the parameters x 350 mm (D) x 700 mm (H).
breath profiles proposed in the new
on screen together with a graphic
USP Chapter <1602> for the in vitro
display of the pattern generated
assessment of Spacers and Valved BRS 2000 Set-up for the testing of
Holding Chambers used with MDIs. Alternatively, the user can generate Spacers and VHCs with
Facemasks
their own Flow versus Time profiles
The BRS 2000 is also suitable for
in the form of text files containing
generating other wave forms used in
tabulated data points. Up to 1000
developmental studies of nebulisers
data points can be entered, with
and other inhaled products requiring
time intervals as small as 20
an inhalation volume of < 900 mL.
milliseconds, allowing the creation
The control function is provided in of high-resolution breathing profiles,
the form of an embedded computer (e.g. as measured in clinic).
running Windows XP used in
Breathing patterns, which can consist
conjunction with a colour monitor,
of single or multiple breaths, with
keyboard and mouse. An ethernet
or without exhalation phases,
connection socket is provided for
can be saved and loaded into the
network printing.
software, as required.
76
ancillaries
The Breathing Simulator Model BRS 3000 As already mentioned in the section on In the case of APSD measurements, a
is similar in design and operation to the in vitro - in vivo correlation (see Pages Mixing Inlet (See Page 70) is required
BRS 2000 except that it has a volume of 66-71), two of the main factors that to decouple the variable flow through
500 mL - 5 Litres (certified). have been identified as critical to IVIVC the inhaler (generated by the Breath
are: Simulator) from the steady-state flow
It also features a maximum flow rate of
rate through the cascade impactor.
240 L/min (free flow) and a maximum 1. Replacing the conventional
acceleration of 25 L/s2 (free flow) making Induction Port with a mouth/ Therefore in order to generate a test
it the ideal unit for the testing of MDIs throat model having a more realistic system for the measurement of APSD,
and DPIs for improved IVIVCs. human geometry (such as the using realistic patient profiles, the
Alberta Idealised Throat described following items are required:
Delivered Dose Uniformity (DDU) and
on Page 68).
Aerodynamic Particle Size Distribution 1. The conventional Ph.Eur. / USP
(APSD) continue to be subjects of 2. Replacing the existing flow rate Induction Port or Alberta Idealised
close scrutiny as the concept of Quality conditions employed in testing with Throat (Adult or Child).
by Design (QbD) becomes more breathing profiles more typical of
2. Mixing Inlet with BRS 2000/3000
widespread. The emphasis is now on conditions in vivo (see the Mixing
Flow Control Valve and Adapter
method development that uses design Inlet described on Page 70).
for connection to a compressed air
of experiments (DoE) to identify the
The Breathing Simulator Model BRS supply (see ordering information
most significant factors, the critical
3000 has been specifically designed to below).
quality attributes (CQAs), relevant to
provide the latter.
the product concerned. 3. Cascade Impactor (ACI, NGI, MSLI,
In the case of DDU, the BRS 3000 (and FSA or FSI).
For this reason, laboratories are devoting
BRS 2000) can be connected directly
more resources to method development The BRS 3000 measures 800 mm (W) x
to the Dose Unit Sampling Apparatus
in an attempt to try to establish in vitro - 400 mm (D) x 850 mm (H).
using a suitable adapter (see Pages
in vivo relationships at an early stage in
24-30, 67 and ordering information
the product design.
below).
Accessories for use with Mixing Inlet and Cascade Impactors (IVIVCs)
9123 BRS 2000/3000 Flow Control Manifold for Mixing Inlet (6 mm) BRS 1100/2000/3000 Qualification Kit
9124 BRS 2000/3000 Flow Control Manifold for Mixing Inlet (1/4)
77
ancillaries
The vast majority of Dry Powder In the case of the in vitro testing of In the in vitro case, the in vivo
Inhalers (DPIs) are classified as DPIs, the pharmacopoeias specify that strength and duration of the users
passive breath activated devices; the duration of a single inhalation cycle inspiration is replicated by the flow
that is to say, they rely solely on the (equivalent to that of a typical user when rate used and the time for which the
patients inspiration to operate. inhaling the drug) be achieved through solenoid valve concerned remains
the use of a 2-way switching valve open.
There is no necessity to co-ordinate
connected to a vacuum pump.
breathing with the activation the To establish the correct flow rate
patient simply inhales deeply to The operation of the switching valve, to be used, it is first necessary to
access the drug. and hence the duration of the breathing establish the flow rate required to
cycle, is controlled by means of a timer. produce a pressure drop comparable
It follows that both the delivered
with that found at the mouth of the
and fine particle dose of DPIs are One side of the valve is connected to
user in vivo when using the particular
dependent on the strength and either the sampling apparatus (in the case
inhaler being studied.
duration of the patients inspiration, of delivered dose) or a cascade impactor
a critical quality attribute (CQA) (in the case of particle size determination) Both European and US
which must be simulated during and the other to a vacuum pump. Pharmacopoeias suggest a pressure
the course of in vitro testing. drop over the inhaler of 4 kPa as
In pre-test mode, the switching valve is
being broadly representative of the
The testing of DPIs is further in the closed position such that no flow
pressure drop generated during
complicated by the fact that passes through the test apparatus.
inhalation by patients using DPIs.
different inhalers provide varying
On initiation of the test, the 2-way valve
degrees of resistance to flow i.e. The pressure drop created by the
switches such that flow now passes
some require more effort to inhale air drawn through an inhaler can be
through the test apparatus and hence the
than others (see graph below). measured directly by measuring the
inhaler under test. On expiration of the
absolute pressure downstream of the
pre-set time, the solenoid closes again
inhaler mouthpiece and comparing
and the inhalation cycle is complete.
this directly with atmospheric
pressure.
The relationship between pressure drop and flow rate for a Using a flow control valve, it is then
range of commercially available DPIs a simple matter to adjust the flow
Assi, K.H. and Chrystyn, H, et al. The device resistance of recently introduced rate from the vacuum pump to
dry-powder inhalers. Journal of Pharmacy and Pharmacology, 52 (Suppl): 58, 2000.
produce the required pressure drop
of 4 kPa and then, by replacing the
10 Easyhaler
inhaler with a suitable flow meter, to
Twisthaler measure the flow rate, Q, required to
Pressure drop (cmH20)0.5
78
ancillaries
The relationship between flow rate and test time for DPI testing.
4 L at 100 L/min
150
4 L at 60 L/min
79
ancillaries
In summary, a number of factors can The main features of the TPK are as 2. Aerodynamic Particle Size
have a significant effect on the testing follows: Distribution
of Dry Powder Inhalers:
Differential pressure meter for Use a Cascade Impactor (ACI, NGI
The resistance to flow posed by measuring pressure drop, P1 or MSLI) together with the system
the inhaler under test (Range: 0 to 100 kPa Resolution: components relevant to that particular
The appropriate flow rate, Q, 0.1 kPa) impactor to perform this test.
required to generate a 4 kPa
Flow control valve to adjust flow rate
pressure drop over the inhaler Proceed as follows:
concerned Timer-controlled solenoid valve to
1. Assemble the appropriate system
The duration of inspiration required adjust duration of flow (Range: 0
as shown in the schematic for
to give the specified test volume to 999.9 seconds, Resolution: 0.1
testing the APSD of DPIs (see
The stability of the flow rate in terms seconds)
below).
of critical (sonic) flow
Absolute pressure meter for
2. Attach a suitable flow meter (see
The Apparatus suitable for measuring sonic flow, P2 and P3
Page 90) to the inlet of the
measuring the uniformity of delivered (Range: 0 to 100 kPa, Resolution
induction port 3. Switch on the
dose for powder inhalers and the 0.1 kPa)
pump and open the two-way
Experimental set-up for testing
Membrane keypad control solenoid valve.
powder inhalers described in
Ph.Eur. in Chapters 0671 and 2.9.18 LED display of differential and 3. Adjust the flow control valve until
respectively and Apparatus B: absolute pressure values the flow rate, as measured by
Sampling apparatus for dry powder the flow meter, is equal to Q (as
LCD back-lit display of set and
inhalers and the Apparatus 2,3, measured either during DDU see
elapsed actuation times
4 or 5: General control equipment 1 above or by employing the
described in USP Chapter 601 take all Induction Port P1 Measurement
PROCEDURE
of these factors into account. Adapter depicted below and
1. Delivered Dose Uniformity described on Page 96).
These specifications form the basis
of the Critical Flow Controller Model Follow the procedure described under
TPK which incorporates all of the Delivered Dose Uniformity Dosage
equipment required into a single Unit Sampling Apparatus (DUSA) for Induction Port P1 Measurement Adapter
integrated system. DPIs (see Page 30). for measuring Device Flow Resistance in
absence of DUSA for DPIs
Schematic of NGI System for
Interchange
testing DPIs including Critical Flow
Controller (Inlet/Outlet Reversed
option) and Pump
80
ancillaries
81
ancillaries
CRITICAL FLOW CONTROLLER USB printer port and RS232 port for The TPK 2000 is fitted with two external
MODEL TPK 2000 data output pressure ports. These can be used
External temperature/humidity singularly or together to perform a
The Critical Flow Controller Model
sensor (option) variety of functions.
TPK 2000 is designed to control and
* For measuring environmental test
document all the critical parameters The first port can be connected to the
conditions as recommended by
associated with the delivered dose Dose Unit Sampling Apparatus (DUSA)
FDA
testing and aerodynamic particle for DPIs or to the Induction Port P1
size distribution measurement of Dry It also includes a series of extra Measurement Adapter depicted on
Powder Inhalers (DPIs). functions including those for: Page 80 and is used to determine
Auto print-out or download of device resistance by measuring the
Its predecessor, the Critical Flow
relevant calibration and test pressure drop over the inhaler (P1).
Controller Model TPK has already
become an international standard in the parameters The second of the two ports is used
field of DPI testing. Facility to measure impactor leak for the measurement of atmospheric
rates and total pressure drop pressure during the test set-up process.
The more advanced Critical Flow User calibration function (with
Controller Model TPK 2000 retains the optional calibration kit) In combination, the two pressure ports
same critical specifications laid down Storage of calibration time/date can also be used for determining
in Ph.Eur. and USP as the earlier model Spacer testing delay function impactor leak rates and for ACI and
but incorporates a number of additional NGI Delta-P measurements.
features namely: A series of menus guide the user
through the operation of the Delta-P measurements can be
4-Line menu-driven LCD display of instrument. Interaction with the unit extremely useful in monitoring
all parameters is via a touch sensitive membrane day-to-day performance of the
Improved differential pressure meter keypad. impactor nozzles and can be used as
for measuring pressure drop, an early warning of any critical
P1 (range: 0 to 15 kPa, resolution: A highly durable, illuminated test nozzle wear or occlusion (see
0.01 kPa) button independent from the keypad Page 124).
Automatic calculation and display of allows high speed, repeat actuations
sonic (critical) flow for multiple actuation testing. The
TPK 2000 Firmware Version
Automatic test set-up function facility for TTL trigger inputs is also
4.04+ now provides for the Leak
Setting and counting of number of provided to allow actuations to be
Testing of cascade impactors as
test actuations performed remotely.
part of the routine test set-up,
Illuminated test actuation button Multiple ports/sockets allow without the need for additional
indicates readiness to actuate connection to external devices leak test equipment.
Foot switch or external TTL trigger such as a PC, a printer, a foot switch, a
facility for remote actuation This encourages leak testing
flow meter, and a temperature/relative
RS232 connection for flow meter prior to each and every analysis,
humidity sensor for the monitoring of
(for recording flow rate during thereby safeguarding data quality.
environmental conditions.
set-up)
82
ancillaries
PROCEDURE
The TPK 2000 menu system Prior to running the actual test, the During the test, the current actuation
automatically guides the user through TPK 2000 prompts the user to enter number (e.g. 3 of 20) is indicated on the
the correct test set-up procedure. the duration of the test (the inspiration display together with the test duration,
time) along with the number of time delay (if used) and elapsed time.
During initial delivered dose uniformity
actuations for the test phase.
testing, the user is instructed to set Actuations are normally triggered
the pressure drop over the inhaler (P1) Provision is also made to programme by depressing the illuminated RUN
to the desired value (typically 4 kPa) a time delay before commencing button on the TPK 2000 front panel.
using the flow control valve located on the flow (i.e. opening of the solenoid This button is illuminated once the
the left side of the unit. valve). test parameters have been entered,
indicating that the unit is ready for the
If the flow rate (as measured by a This has a number of advantages since
first actuation and repeated for each
flow meter positioned in place of the user can reproducibly control the
further actuation.
the inhaler) required to generate the time between priming the inhaler
pressure drop is more than 100 L/min, and starting the flow. This facility can If desired, test actuations can be
then provision is made to skip the P1 be used to control both inspiration triggered remotely using the optional
setting and re-adjust the control valve delay times and effective inspiratory foot switch, RS232 input or external
to achieve a flow rate of exactly volume, essential when testing the TTL trigger (for automation) via sockets
100 L/min. This also applies during effectiveness of spacers and valved located on the rear panel.
impactor particle sizing when the holding chambers (VHCs).
test flow rate has already been
predetermined during the DDU phase.
Cat. No. Description
The TPK 2000 now automatically
measures and calculates the test 8760 Critical Flow Controller Model TPK 2000
set-up parameters: P1, P2, P3, P3/2 8760-R Critical Flow Controller Model TPK 2000-R (Inlet/Outlet
ratio, flow rate, temperature, relative Reversed)
humidity, atmospheric pressure 8769 Temperature and Relative Humidity Sensor (option)
together with test set-up time, TPK 8761 Foot Switch (option)
2000 and flow meter serial numbers 8766 Printer (option)
and calibration data. 8763 TPK 2000 Re-calibration Certificate
8752 Flow Time Verification Kit
All of this data can be printed out or
8753 Re-calibration of Flow Time Verification Kit
output to an external PC immediately
8502 Induction Port P1 Measurement Adapter
after the test set-up process is
complete. Note: See Page 90 for details of Flow Meters suitable for connection
to the TPK 2000
83
ANCILLARIES
BREATH ACTUATION CONTROLLER 1) Breath Actuated (or Breath 2) Spacers and Valved Holding
MODEL BAC 2000 Operated) Metered Dose inhalers Chambers (VHCs) used with MDIs
(MDIs)
A cut-down version of the TPK 2000, For the testing of the Spacers and
the Breath Actuation Controller is an As the name implies, the Breath Valved Holding Chambers (VHCs)
electrically operated, timer controlled Actuation Controller Model BAC 2000 commonly employed with MDIs in
two-way solenoid valve. is a key element in determining the accordance with the specifications
Delivered Dose Uniformity and as laid down in the new USP draft
In practice, it is positioned in the line
Particle Size Distribution of Breath Chapter <1602>.
between the DUSA collection tube/
Actuated or Breath Operated MDIs.
cascade impactor and the vacuum Spacers and VHCs are add-on devices
pump in order to control the air flow In this instance, the initiation of the which are used in conjunction with
supply to the inhaler under test. flow triggers the inhaler such that pMDIs to overcome the problems
sampling from the MDI occurs only at associated with poor uncoordinated
The solenoid valve employed provides
the predetermined flow rate. inhalation technique on the part of
near instantaneous starting and
the user. This usually occurs when the
stopping of the air flow during testing The volume of air sampled (the breath)
patient delays inhalation rather than
and has both delay and inhaled time is the product of the flow rate (typically
breathing in on actuation.
functions. 28.3 or 30 L/min) and the time that the
valve is open. The draft chapter specifies two tests
User interface is via a membrane
to determine aerodynamic particle
keypad with a 4-line LCD and menu The BAC 2000 can also be used to
size, Test Part 1A to measure the
driven operation. restrict the volume of air sampled to
APSD under optimal conditions i.e.,
the 2 litres or less as recommended
An independent illuminated test button on actuation and Test Part 1B under
by the FDA and specified in USP 38
allows high speed, repeat actuation and worst case conditions that is to say
when carrying out Delivered Dose
recording for multi-shot testing. with a delay of 2 or more seconds
Uniformity testing on conventional
between inhaler actuation and
Facility for TTL trigger inputs (via a DIN MDIs.
sampling onset.
socket) and RS232 communication are
See Pages 25-26 for further details.
also provided to allow actuations to be
performed remotely, e.g. via an optional
footswitch or external triggering system.
System for Breath Actuated MDIs
The BAC 2000 is a microprocessor
controlled instrument designed
specifically for three test applications: BAC 2000
Timer
84
ANCILLARIES
egnahcretnI
egnahcretnI
valve automatically opens to draw the Breath Actuation Controller Model BAC 2000
aerosol into the the cascade impactor. Electronic timer for control of solenoid valve
2-way/2-port solenoid valve
See Page 55 for further details.
4-line LCD display of all parameters
Illuminated mechanical start button and keypad
3) Nebulisers to USP <1601> and Timer range 0-999.9s, resolution 0.1s
Ph.Eur. 2.9.44 Solenoid valve opening/closing time 25/25 ms
Actuation setting/counting function
For the testing of Nebulisers according
Foot switch (or TTL input) interface for remote actuation
to USP <1601> and Ph.Eur. 2.9.44.
RS232 (serial) interface for and remote actuation
In this case, the BAC 2000 is used as Spacer/VHC testing delay function
a substitute for the manually operated
3-way valve to set a test operating
time (typically 60 seconds) for which System for Nebulisers
the nebulised aerosol is drawn into the
Next Generation Impactor (NGI).
85
ancillaries
INTRODUCTION CITDAS Version 3.10 has been to the impactor being used (i.e.
designed to run on Microsoft impactor stage ECDs). The software
USP Chapter <601> and Ph.Eur.
Windows Vista, XP, 7 and 8 operating defaults to the stage effective cut-off
Chapter 2.9.18 specify various types of
systems. Installation is particularly quick diameters (ECDs) recommended by
multi-stage cascade impactor that can
and easy and does not require special the manufacturer of the impactor at
be used for measuring the particle size
IT knowledge. that flow rate selected by the operator.
distribution of inhalers together with
This includes NGI calibration data for
suggestions as to how the resulting
SIMPLE OPERATION 15 to 30 L/min such that the software
data should be analysed. Hitherto,
can now be used for nebulisers as well
this data analysis has largely been On entering the program, the operator
as MDIs and DPIs.
performed using a variety of techniques is prompted to select the source of
with little attention being paid to the data to be analysed, whether new, Alternatively, specific calibration
standardisation and validation. stored or imported (1) and the type of data, as for example relating to that
impactor being used (2). One of the particular impactor, can be entered if
Copley Inhaler Testing Data Analysis
powerful new features incorporated desired and saved as a template.
Software (CITDAS) Version 3.10
into Version 3.10 of the software is the
is a standardised approach to the The Deposition Data Screen provides
Import CSV file facility which allows
entry, analysis and reporting of the fields for the entry of the drug
Comma Separated Variables data
Aerodynamic Size Distribution of drug deposition (5) recovered from the
input into CITDAS to be streamlined
output from MDIs, DPIs and Nebulisers impactor stages in addition to:
without the need for manual entry.
in accordance with USP and Ph.Eur.
Dose No. - The number of the dose
The system then defaults to the User
Fully validated, the software will accept sampled from the inhaler
Set-up Screen where the user is
data from the Andersen Cascade Device - The drug deposition on the
requested to fill in various information
Impactor, the Multi-Stage Liquid actuator/inhaler
fields to identify themselves, the
Impinger, the Marple-Miller Impactor Flow Rate - The flow rate employed
impactor being used and the inhaler(s)
and the Next Generation Impactor. during operation (L/min)
under test.
Doses to Device - The number of
The majority of the doses sampled by the device
fields are free-form Delivered Dose - As determined
3 4 and the field names during Delivered Dose Uniformity
can be customised in Number of Runs - Number of runs to
Maintenance Mode be processed (1-12)
(there are two levels
This screen also allows the user to
of access: operator
select the number of runs to be
and maintenance) to
processed (1-12) and the units of
meet individual user
measurement in terms of g or mg (6).
requirements.
Provision is also made for omitting the
The same screen also deposition data from the preseparator
contains the stages where it is not deployed and also,
(3) and calibration Stages 6 and/or 7 when using the ACI
3: User Set-up (NGI) data (4) applicable at flow rates greater than 60 L/min.
86
ancillaries
5 6
The number of runs to be processed can size. This includes the preseparator. weights expressed in terms of the
be adjusted between 1 and 12. Three Mean and Standard Deviation for each
This facility is accessed through the
runs are displayed per screen - further particular run (11).
Group button (7) on the Deposition Data
runs are accessed by means of the scroll
screen which in turn reveals the Group The following summarised data (12) is
bar.
Specification Screen (8). listed for each run:
The same screen allows the user to
The results of the various tests are Total Dose per Shot [g] or [mg]
specify the criteria to be applied when
calculated automatically and displayed Delivered Dose [g] or [mg]
calculating the Fine Particle Dose (FPD)
on the Shot Weight Input & Results Fine Particle Dose [g] or [mg]
and Geometric Standard Deviation
Summary Screen together with graphical Fine Particle Fraction [%]
(GSD). In the case of the FPD, this can
representations (9) of the data for each Mass Medium Aerodynamic
be expressed in terms of either impactor
run selected from the following options: Diameter (MMAD) [m]
stage or selected aerodynamic particle
Geometric Standard Deviation
size (e.g. 5 microns). Log Probability Graphs of
(GSD)
Percentage of Mass Less than
One key feature of Version 3.10 that Particle Undersize log-probit graph
Stated Aerodynamic Diameter
will be welcomed by users is the ability (probit values 4-6)
against Log Aerodynamic Diameter
to define up to five fine particle dose/ Regression Coefficient (R^2)
Histograms of Drug Mass against
fraction groupings (7). Each group being Device Sampling Flow Rate [L/min]
Drug Distribution
defined by a range of either impactor
Cumulative Graphs of Percentage Version 3.10 also provides the facility
stage or aerodynamic particle diameter
Drug Distribution against particle to display, print or output Mean/
(by interpolation).
size dose, SD/dose and %RSD/dose
This means that in addition to reporting data at the end of the 12 runs. This
The drug deposition bar chart (10) can
fine particle dose values, it is now data can be accessed by scrolling right
be viewed (and subsequently printed)
possible to routinely subdivide the at data entry.
with or without throat deposition. The
reported delivered dose into up to five
same screen allows the user to input shot The Group Results Screen (overleaf)
groupings based on stage or particle
can be accessed through the Group
Results button positioned below the
6: Short Input
summarised data fields.
and Results
Summary It gives up to 5 stage groupings (13),
11 between two stages or between two
particle sizes in each case. The group
results can be printed on a separate
printout.
10
12
87
ancillaries
13
14
7: Group Results Screen
8: Printing
PRINTOUT FORMATS SUMMARY OF KEY FEATURES Full printouts in both Ph.Eur. and
USP formats, incl. graphical, tabular
CITDAS now has five printout types Standardised approach to the
and group summaries
allowing the user to present data in five analysis of impactor data
Fully updated to reflect the
different formats: Accepts data from ACI, MSLI, MMI
definitive archival calibrationof the
European Pharmacopoeia format and NGI
NGI at 15 L/ min so as to include
United States Pharmacopeia format Instant comparison of up to 12 runs
Nebulisers
Graphical Summary Auto-correction of results to allow
Facility to import/export data as
Tabular Summary for impactor calibration differences
CSV files for manipulation in
Group Results and/or differing flow rates
Microsoft Excel or similar software
Fine Particle Dose (FPD) selection
All of the printout formats are located packages
based on either impactor stage or
on a special Print Screen (15) which is Stage groupings
aerodynamic particle size
accessed by pressing the Print Button
(e.g. 5 m) CITDAS is supplied as standard
(14) on the Shot Weight Input and
Shot weight report option with full supporting validation
Results Summary Screen.
Automatic calculation of FPD, FPF, documentation, which provides
Examples of the various printouts may MMAD and GSD verification of the correct storage of
be found on Page 89. In response On-screen graphs in either input parameters and details of the
to customer feedback, Version 3.10 histogram, log probability or algorithms, methods and conclusions
includes three significant improvements cumulative formats employed to calculate the results.
on earlier versions:
88
ancillaries
Printout formats
89
ancillaries
Interchange
INTRODUCTION tolerances and there are no inherent breath-operated devices trigger only
leaks in the system, it can be seen that when the flow rate through them
The Delivered Dose Uniformity
the cut-off diameter (the aerodynamic exceeds a certain value.
(DDU) and Aerodynamic Particle Size
diameter of particles that accumulate
Distribution (APSD) are two of the main DETERMINING THE PROPER TEST
on any given collection surface) of any
Critical Quality Attributes (CQAs) in FLOW RATE
given stage is directly related to the
measuring the therapeutic efficacy of
volumetric flow rate of the inlet air Although patient inspiration subjects
orally inhaled and nasal drug products
passing through it. A change in the inhalers to varying flow rates, cascade
(OINDPs).
flow rate results in a change in the impaction requires a constant
The data produced by both of these aerodynamic particle size characteristics volumetric air flow: within this
tests can be severely compromised of the stage or stages concerned. constraint, the flow rate must, as far
if the inlet flow rate (the flow rate at possible, reflect the conditions of use.
Indeed, a simplified version of Stokes
the entrance to the induction port or
law, which describes the relationship For propellant or pump-driven
DUSA) used during testing is inaccurate
between stage cut-off diameter, delivery, particle aerosolisation is
and/or inconsistent, generating
nozzle diameter and air flow rate, generally insensitive to test flow rate.
discrepancies with regard to its effects
demonstrates that a 5% deviation in MDIs and the majority of nasal sprays
on both the cascade impactor itself and
flow rate changes the stage cut-off are therefore normally tested at
the inhaler under test.
diameter by approximately 2.5%. 28.3 L/min equivalent to 1 cubic foot/
Cascade impaction, the method used At a flow rate of 60 L/min, min. The NGI, however, was calibrated
to measure APSD, divides the aerosol Stage 1 of an Andersen at 30 L/min and should be operated at
cloud into various size fractions on Cascade Impactor should that rate for this type of device.
the basis of particle inertia which is a give a theoretical cut-off
function of aerodynamic particle size of 4.4 microns reduce that Flow Meter Model DFM3
and velocity. flow rate to 57 L/min and
cut-off is effectively increased
In this technique, particle-laden air
to 4.5 microns.
is drawn through a series of stages,
each of which has a predetermined The volumetric air flow can
number of nozzles of defined diameter. not only affect the ability
Providing that the volumetric flow rate of the cascade impactor
of the air stream remains constant, the to function correctly but
air velocity increases from stage to also compromises the actual
stage. performance of the inhaler itself.
Particles with sufficient inertia impact For many inhaled products, for
on the collection surface at a set example, the air flow drives the
distance beneath the nozzles while aerosolisation of the formulation and
smaller particles are retained in the air can therefore have a marked impact on
stream and carried to the next stage. how the dose disperses and hence on
The result is a series of size fractions, the resulting aerodynamic particle size
typically between 0 and 10 microns. determination.
90
ancillaries
Interchange
91
ancillaries
Example of Mouthpiece
Adapter Engraving
Example of Mouthpiece
Adapter Tool
MOUTHPIECE ADAPTERS
The mouthpiece adapters supplied be required of that design, in the adapter is transferable between all of
by Copley Scientific are specially future. these instrument types.
moulded from high quality silicone
The Induction Port used with the ACI, The Glass Twin Impinger however,
rubber in order to guarantee an
NGI, MMI and MSLI together with the because of its glass construction,
airtight seal between the inhaler under
Sampling Apparatus for both MDIs and the Fluticasone Propionate (FP)
test and the test apparatus.
and DPIs and the WSC2 all have the Induction Port, do have different
There is no standard mouthpiece same external dimensions at the inlet external dimensions at the inlet and so
adapter per se as each inhaler design and hence the same mouthpiece require their own mouthpiece adapter.
is different. Adapters are available
however for the more common
devices on the market (see ordering Cat. No. Description
information).
5003 Custom Mouthpiece Adapter (for Induction Port and/or DUSA
For other unlisted inhalers, we require for MDIs/DPIs)
a sample of the inhaler to be tested, 5237 Custom Mouthpiece Adapter (for Glass Twin Impinger and FP
so that we can make a cast of the Induction Port)
mouthpiece concerned and produce 5004 Tooling Charge for 5003 & 5237 (per inhaler design)
an appropriate moulding tool. R
5003C Easyhaler Mouthpiece Adapter
R
This moulding tool is used to mould 5003D Cyclohaler Mouthpiece Adapter
R
the mouthpiece adapter(s) to that 5003E Handihaler Mouthpiece Adapter
R
particular inhaler design. 5003F Diskus Mouthpiece Adapter
5003G Novolizer R Mouthpiece Adapter
The tool is then supplied along with R
5003H Rotahaler Mouthpiece Adapter
the mouthpiece adapter(s) to the end R
5003I Turbuhaler Symbicort Mouthpiece Adapter
user so that it can be reused should R
5003J Diskhaler Mouthpiece Adapter
any additional mouthpiece adapters R
5003K Respimat Mouthpiece Adapter
R
5003L Evohaler Mouthpiece Adapter
R
5003M Pari LC Plus Mouthpiece Adapter
5003N Trudell AeroChamber R Plus Mouthpiece Adapter
R
5003O Tobi Podhaler Mouthpiece Adapter
R
5003X Inhaler Support Accessory for Mouthpiece Adapter
Induction Port with Diskus 5003Y Mouthpiece Adapter Engraving (each)
Mouthpiece Adapter and
Inhaler Support Accessory
5238 Mouthpiece Adapter (Induction Port to DFM 2000)
92
ancillaries
PUMPS
High Capacity Pump Model HCP5 (left) and Low Capacity Pump Model LCP5 (right)
INTRODUCTION
The Copley Scientific Low and High Both pumps, for example, come with For this reason, the Pharmacopoeias
Capacity Pumps Models LCP5 and the appropriate fittings to connect insist that, in the case of dry powder
HCP5 are vacuum sources that have to any inhaler testing system and to inhalers, critical (sonic) flow conditions
been specifically designed for use in allow the user to position the pump are achieved within the system prior
the testing of MDIs, DPIs, nebulisers to either right or left of that system to testing, to ensure that the vacuum
and nasal sprays in accordance with depending on the available space pump employed is of sufficient
the specifications laid down in the on the laboratory bench. Provision is capacity for the task.
European and US Pharmacopoeias. also made to vent the exhaust to an
To meet these considerations the
extraction system.
The units represent the very latest in Copley LCP5 and HCP5 Pumps have
high performance, low maintenance, It should be noted that the resistance been carefully designed to cover a
oil-free rotary-vane vacuum pump to flow imposed by the test apparatus, range of free flow conditions between
technology. in conjunction with the requirement 133 and 833 L/min.
to achieve sonic flow in the case
Foremost in the design specification Both pumps are fully encased with an
of DPIs, can reduce the effective
were those features that you, the user, internal acoustic lining and vibration
capacity of the pump to less than 20%
identified as being essential to inhaler damping to reduce noise levels to less
of the flow rate measured in free flow
testing, namely that the pump should: than or equal to 60 dB (A).
(unrestricted) conditions.
Be equipped with the correct Being oil-free, neither pump requires
In practice, this means that in order to
fittings to link to the test system oil changes or regular replacement of
achieve 60 L/min sonic flow through
concerned oil filters.
the system a vacuum pump having a
Have sufficient capacity to free flow of 300 L/min must be used. Self-sealing compound carbon vanes
provide the required test flow Even Metered-Dose Inhaler (MDI) continually adjust so that the pump
and in the case of DPIs to ensure systems provide significant resistance effectively performs with as new
critical (sonic) flow to flow typically in the region of 50% efficiency throughout its service life.
of free flow conditions.
Have low noise levels, suitable for
a laboratory environment Stable flow control is critical to good
impactor measurement practice. This
Be low maintenance
is because the aerodynamic particle
sizing ability of inertial impactors is
dependent on the velocity of the
entrained particles as they pass
through each stage and that velocity is
directly related to air flow.
93
ancillaries
1000
900
800
700
500
400
PUMPS 300
200
LOW CAPACITY PUMP MODEL LCP5 Backs of High
Capacity 100
The Low Capacity Pump Model LCP5 Pump Model
is a small footprint vacuum pump HCP5 (right) 0
90 80 70 60 50 40 30 20 10
designed for optimal operation at low and Low
Absolute Pressure (kPa)
Capacity
flow rates.
Pump Model
Performance Chart Key
This makes it ideal for Metered-Dose LCP5
Inhalers (MDIs) and Nasal Sprays HCP5 (50 Hz) LCP5 (50 Hz)
which are tested at 28.3 or 30 L/min HCP5 (60 Hz) LCP5 (60 Hz)
2 x HCP5 (50 Hz)
and Nebulisers which are typically
2 x HCP5 (60 Hz)
tested at 15 L/min. These devices
do not generally require the use of a
Critical Flow Controller. HIGH CAPACITY PUMP MODEL HCP5
In this instance, the test apparatus The High Capacity Pump Model HCP5 The unregulated inlet bypasses
(the dose unit sampling apparatus is a well established high capacity pump the flow control valve and is used
(DUSA) in the case of delivered dose for the higher, sonic flow rate testing in conjunction with the Critical
testing and the cascade impactor in the requirements of Dry Powder Inhalers Flow Controller for dry powder
case of particle size determination) is (DPIs), although it can equally well applications. It provides a maximum
connected directly to the pump. be used for MDIs, Nasal Sprays and flow of 416 L/min.
Nebulisers.
The unit contains a 0.35 kW motor In instances where this flow rate
capable of generating a maximum of Like the LCP5, an on/off switch and flow is still not adequate, it is possible
133 L/min free flow (at 50 Hz mains control valve are located on the front to connect a Secondary HCP5 in
frequency). panel of the unit. Two sets of vacuum parallel to the primary pump to give
inlets on either side of the pump allow a maximum unregulated flow rate of
The flow rate can be adjusted between
the user to choose the location of the 833 L/min. This is typically required
0 and 133 L/min free flow using the
pump in relation to the test apparatus. when testing DPIs under sonic flow
flow control valve mounted on the front
Each set of vacuum inlets consists of a conditions with the NGI, at high
panel.
regulated and unregulated inlet. flow rates. Special hose fittings are
The unit is provided with two vacuum supplied with the secondary pump to
The regulated inlet is connected to the
inlets such that the user can decide connect it to the primary unit.
pump via the flow control valve and is
whether to place the pump on the
used to regulate flow between 0 and The pump measures 320 x 560 x 390
right or left side of the test system
250 L/min for MDI, nasal spray and mm (w x d x h) and weighs 45 kg.
dependent on available bench space.
nebuliser applications.
The LCP5 has an in-built cooling fan
located on the top side of the pump
and a ventilation grill on the bottom Cat. No. Description
of the front panel. Two handles are 7903 Low Capacity Pump Model LCP5
located on the top of the pump for 7904 Overhaul Kit for LCP5
lifting and positioning.
7901 High Capacity Pump Model HCP5
The pump measures 270 x 310 x 300 7902 High Capacity Pump Model HCP5 (with additional hose
mm (w x d x h) and weighs 18 kg. fittings for use as a secondary pump)
7905 Overhaul Kit for HCP5
94
ancillaries
1000
900
800
700
Flow Rate (L/min)
600
500
400
300
200
100
0
90 80 70 60 50 40 30 20 10
Absolute Pressure (kPa)
Performance Chart Key
The Super Capacity Pump Model used to regulate flow between 0 and that there is no oil vapour in the
SCP5 is a bench-top vacuum pump for 280 L/min for MDI, nasal spray and exhaust air, making it suitable for use
the laboratory capable of generating nebuliser applications. A maximum in a laboratory environment.
sonic (P3/P2 0.5) flow rates through unregulated flow of 683 L/min is
The pump measures 420 x 600 x 450
the Next Generation Impactor (NGI) available for DPI applications.
mm (w x d x h) and weighs 45 kg.
up to 100 L/min. The vacuum is provided by an oil
Important Note:
It is designed to provide a viable lubricated rotary vane pump having a
Special electrical supply requirements
alternative to the use of two HCP5 1.5 kW motor and exceptionally low
are necessary for UK and US
Pumps to achieve these conditions. noise levels for its size.
applications. Please check details prior
The flow rate is controlled by means of The SCP5 is fitted with two access to placing your order.
a valve on the front panel of the unit. panels to allow easy access for
Two sets of vacuum inlets on either replenishing oil and changing the oil
side of the pump allow the user to filter. A dual filtration process, ensures
choose the location of the pump in
relation to the test apparatus. Each Pump Model (50 Hz Version) LCP5 HCP5 2 x HCP5 SCP5
set of vacuum inlets consists of a Type Rotary Vane Rotary Vane Rotary Vane Rotary Vane
regulated and unregulated inlet. Lubrication Type Dry Dry Dry Oil
Max. Flow in L/min (unrestricted) 120 416 833 683
The regulated inlet is connected to the Max. Sonic Flow through NGI N/A 80 100 100
pump via the flow control valve and is Max. Vacuum Level < 15 kPa <15 kPa <15 kPa <0.1 kPa
Applications: Nasal Yes Yes Yes Yes
Nebulisers Yes Yes Yes Yes
MDIs Yes Yes Yes Yes
DPIs No Yes Yes Yes
Noise in dB (A) 55 60 70 65
Routine Maintenance None None None Oil/Filter Change
Dimensions (w x d x h) 27 x 31 x 30 cm 32 x 56 x 39 cm 74 x 56 x 39 cm 42 x 60 x 45 cm
Weight (kg) 18 45 90 70
95
Title
ANCILLARIES
SUNDRIES
A simple device that is placed between the inhaler and the induction port and is
used in conjunction with a Critical Flow Controller to measure the flow resistance
of the inhaler under test.
RINSING CAPS
Silicone rubber and 316 stainless steel rinsing caps are available for capping off
the open ends of the ACI and NGI induction ports and the NGI preseparator
during manual drug recovery. Simple rubber stoppers are also available.
5227 Set of 2 Stainless Steel Rinsing Caps for NGI Induction Port
5228 Set of 2 Stainless Steel Rinsing Caps for NGI Preseparator
5232 Set of 2 Silicone Rubber Stoppers for NGI I.P./Preseparator Rinsing Caps
96
specialTitle
applications
Special Applications
97
special applications
Dissolution Tester
Model DIS 8000
98
special applications
DISSOLUTION TESTING
DOSE COLLECTION
Watchglass/PTFE
DESCRIPTION Assembly for use with ACI
99
special applications
100
special applications
Indeed, the Ph.Eur. and USP are not This volume is generally considered to
fully harmonised with respect to orally be representative of a typical patient
inhaled and nasal drug products in with asthma or COPD.
general.
101
special applications
Andersen Cascade
Impactor for
FP/Salmeterol
Inlet Cone for FP/
Powders
Salmeterol Aerosols
102
specialTitle
applications
1. Set Distance from Impaction Plate 2. Actuate the Device 3. Display Impaction Force
INTRODUCTION The sensation is due to the high SPRAY FORCE TESTER SFT 1000
velocity blast and subsequent
Spray pattern and plume geometry The amount of aerosol deposited on
evaporation of liquid propellant
are common techniques employed the throat is largely dependent on
remaining in droplets that impact on
in the pharmaceutical industry to inertial impaction i.e. the velocity and
the back of the users throat.
characterise and quantify the shape of APSD of the aerosol cloud concerned.
the emitted spray from Metered-Dose The ability to produce a softer, warmer
It follows that since velocity is directly
Inhalers (MDIs) and Nasal Sprays. aerosol plume by:
related to impaction force the latter
However, possibly of as much concern a) changing the formulation (drug, should be a good indicator of local
as either of these two parameters propellant, co-solvent) or delivery equivalence for an inhaled
is the reaction of the user to the b) changing the device (metering drug.
impaction force of the MDI or spray on volume, actuator orifice diameter) or
The Spray Force Tester Model SFT
the throat or nasal passages. c) using add-on devices such as spacers
1000 is a high precision instrument for
or holding chambers
The cold freon effect (the initial measuring the maximum impaction
reaction to the cold blast of MDI is an important tool in product force of MDIs and nasal sprays over
propellant on the back of the throat) development to ensure better patient the duration of the spray plume.
can often result in the patient aborting compliance.
It has a range of 0 to 2500 mN and an
the inhalation process resulting in
Studies into various CFC and HFA accuracy of +/- 2.5 mN.
inconsistent delivery to the lungs.
based suspensions and solutions, for
The distance of the inhaler relative to
Similar reactions can be generated by example, together with different orifice
the impactor can be adjusted between
nasal sprays. geometries have shown that plume
0 and 200 mm +/- 0.03 mm using the
temperature can be widely affected by
The cold freon effect is a function precision digital gauge.
these parameters.
of aerosol spray force and
The device is held in a quick release
plume temperature.
clamp to ensure rigidity during
This effect is widely actuation.
recognised amongst the inhaler
A sample of the inhaler to be
community. Indeed, the cold freon
tested is required at the
effect is specifically mentioned as one
time of placing an order
of the criteria required to substantiate
so that the special clamp
therapeutic equivalence in EMAs
can be made.
Guideline on the requirements for
clinical documentation for orally
inhaled products (OIP) including
the requirements for demonstration
of therapeutic equivalence between
two inhaled products for use in the
treatment of asthma and chronic
obstructive pulmonary disease (COPD)
in adults and for use in the treatment
of asthma in children and adolescents Spray Force Tester Model SFT 1000
published in January 2009.
103
special applications
The main features of the SFT 1000 PLUME TEMPERATURE TESTER The outside diameter of the inlet of
include: MODEL PTT 1000 the sampling tube is reduced such that
it is the same as the induction port
High sensitivity digital load cell The Plume Temperature Tester PTT
in order to accommodate a similar
Range: 0 to 2500 mN +/- 2.5 mN 1000 consists of a polypropylene
mouthpiece adapter (and therefore
Circular impaction plate easily sampling tube, 130 mm long x
MDI or nasal spray).
removed for cleaning 96 mm o.d. having the same internal
High visibility load cell display dimensions as the horizontal section The outlet of the Plume Temperature
Menu-driven controls of the Ph.Eur./USP Induction Port. Tester is normally connected to a
RS 232 output to computer or Waste Shot Collector (see photo on
The temperature profile of the plume
printer this page and Page 31) and vacuum
is measured by 4 centrally aligned
Memory capability for up to 100 pump to capture the measured doses
thermocouples mounted vertically
spray force measurements at the relevant flow rate.
at 25 mm, 50 mm, 75 mm and 100
Pass/Fail alarms for user-
mm from the inlet and linked to a It can, however, easily be connected
programmable limits (for QC)
data acquisition system under the directly to a DUSA tube or Ph.Eur./USP
Precision slides for positioning of
control of a PC. The thermocouples Induction Port if preferred, since the
inhaler relative to impaction plate
are easily removed for cleaning and outside diameter of all of these three
Quick-release device clamp ensures
calibration. accessories is identical.
rigid inhaler support
Special rubber feet eliminate
vibration transmission to load cell Cat. No. Description
Battery or Mains powered
9000 Spray Force Tester Model SFT 1000
Compact dimensions: 580 mm (L),
9001 Additional Device Clamp
200 mm (W), 80 mm (H)
9002 Re-calibration of Spray Force Gauge
Supplied complete with calibration
9003 Re-calibration of Distance Gauge
certificates for load cell and gauge
9004 Spare Impaction Plate
Digital load cell and gauge easily
9005 Compact Printer (Force Gauge)
removed for re-calibration
9006 IQ/OQ Documentation for SFT 1000
Load cell calibration verification
easily performed by user 9010 Plume Temperature Tester Model PTT 1000 (incl. Software)
5001 Waste Shot Collector WSC2
8060 Flow Meter Adapter (PTT 1000 to Flow Meter)
9011 IQ/OQ Documentation for PTT 1000
9012 Re-calibration of 4 Thermocouples
9013 Shortened Mouthpiece Adapter
5004 Tooling Charge for above
104
aUTOMATION
Automation
105
aUTOMATION
DUSA SHAKER This manual shaking process: The Shaker accepts DUSAs for both
MDIs and DPIs.
Both Ph.Eur. and USP state that dose is time consuming
uniformity tests should be carried out can be highly variable (both inter- The rinsing action is achieved by a
on a minimum of 10 doses (from 1 and intra-analyst) due to combination of lateral (side-to-side)
inhaler in the case of Ph.Eur. and 10 inconsistent and incomplete wetting shaking whilst simultaneously rolling
inhalers in the case of USP). of internal surfaces and the sealed collection tubes.
can lead to Repetitive Strain Injury
If then the inhaler fails to meet the Tier The resultant multi-directional mixing
(RSI)
1 dose uniformity criteria concerned, action ensures that all internal surfaces
it may be necessary to repeat the test are wetted and that agitation is
DESCRIPTION
involving the collection of a further 20 performed with a consistent, smooth
doses. The DUSA Shaker has been designed but vigorous action.
to automate the internal rinsing of the
In addition, in the case of USP, The rubber coated rollers are
DUSA collection tubes to ensure full,
further tests have to be carried out specifically designed to grip the
fast and repeatable drug delivery from
throughout the life of the inhaler i.e. collection tubes during processing
all internal surfaces whilst freeing up
Dose Uniformity through Container whilst reducing noise to a minimum.
analysts to perform other tasks and
Life which involves capturing a further
reducing analyst exposure to RSI. This eliminates the necessity to use
10 doses from a single inhaler.
clamps or other fixtures to hold the
All of these tests require the tubes in position during mixing
collection and drug recovery of and permits tubes to be added
individual doses into the collection or removed at any time.
tube of a Dosage Unit Sampling
Apparatus (DUSA) appropriate to its
type (MDI or DPI) prior to assay.
106
aUTOMATION
Control Panel
MAIN FEATURES Designed with a small footprint of 570 Alternatively DPI Collection tubes
mm (W) x 610 mm (D) , the shaker fits without the P1 port are available
The Shaker is designed to accept up
comfortably onto a laboratory bench. as Cat. No. 8608A Collection Tube
to a maximum of 21 MDI Collection
without P1 Port.
Tubes or 12 DPI Collection Tubes (or Full supporting IQ/OQ documentation
a combination of both). is available. Spare Rinsing Caps are available with
either Silicone Rubber (Cat. No. 8607)
Partial loads are quite acceptable. The Note: In order to allow rotation, the
or LDPE Seals (Cat. No. 8607A).
rollers do not have to be fully filled DUSA Shaker is only compatible with
as the rubber coating on the rollers DPI Collection Tubes that have the P1 See Page 30 for further details.
provides sufficient friction to prevent port blanking plug fitted.
lateral movement of the DUSA tubes
Reciproca
during operation. ting
107
aUTOMATION
SPU 2000 with 2 x NGI Preseparator Fixtures SPU 2000 with NGI Preseparator and ACI/NGI Induction Port Fixtures
SAMPLE PREPARATION UNIT On initialisation, the Sample During processing, both the nominal
MODEL SPU 2000 Preparation Unit automatically adjusts duration and remaining duration to
the orientation of the two fixtures to the end of the cycle are displayed on
A significant number of the
the loading position prior to starting screen in terms of either rpm, or time
procedures performed during
the rinsing process. together with the selected speed.
inhaler testing are highly repetitive
The bi-directional process cycle (50%
but not technically complex and The SPU 2000 has variable speed
clockwise, 50%
do not necessarily justify full control between 20 and 60 rpm (+/- 1
anti-clockwise)
automation. rpm). This allows the user to adjust the
ensures thorough
mixing intensity and consequently the
Relatively simple and inexpensive wetting of
dissolution conditions applicable to
sample preparation tools can help all the internal
that particular formulation.
reduce the unwanted effects of surfaces.
repetitive strain injury (RSI), alleviate Similarly, the duration of the rinsing
bottlenecks and ensure that testing cycle can be selected in either
is carried out in a consistent, revolutions of the fixture (0-9999) or
reproducible and controlled manner. time in hours, minutes and seconds
(up to 8 hours).
DESCRIPTION
Control of the unit is provided by a
The Copley Sample Preparation membrane keypad linked to a 4-line
Unit Model SPU 2000 is designed 20 character back-lit LCD screen.
to provide an inexpensive means
of recovering active drug from the
induction ports employed on the Cat. No. Description
ACI and NGI and the preseparator 9202 Sample Preparation Unit Model SPU 2000 (without Fixtures)
of the NGI.
9216 Fixture for ACI/NGI Induction Port (each)
The fixtures feature custom made 8504 Set of 2 Silicone Rubber Rinsing Caps for ACI Induction Port
brackets and use simple silicone 5265 Set of 2 Silicone Rubber Rinsing Caps for NGI Induction Port
rubber end caps to secure and seal
the equipment during operation. 9217 Fixture for NGI Preseparator (each)
5266 Set of 2 Silicone Rubber Rinsing Caps for NGI Preseparator
The SPU 2000 is designed to accept
two fixtures at any one time. 9212 IQ/OQ Documentation for SPU 2000
9213 SPU 2000 Qualification Tools
9214 Re-calibration of SPU 2000 Qualification Tools
108
aUTOMATION
GENTLE ROCKER These sample recovery tools can be The unit comprises a pivoting platform
divided into manual or semi-automated specifically designed to accept the
One of the main objectives of the NGI
systems dependent on the degree of NGI cup collection tray linked to a
Consortium when designing the NGI
automation provided. synchronous motor drive unit (20 or 40
was that the unit should be easy to use
rpm models) and controller.
and automate. A significant number of procedures
performed during inhaler testing In operation, the Gentle Rocker
Crucial to this objective and one of
are highly repetitive and their rocks back and forth about a central
the unique features of the NGI is the
performance can lead to bottlenecks longitudinal axis. The resulting
design of the collection cup tray.
which compromise overall laboratory constant motion helps to dissolve the
During the test, the size-fractionated operations and efficiency. drug in a controlled manner freeing up
particles are deposited in a series of analyst time for other tasks.
Relatively simple and inexpensive
eight cups located in a removable cup
sample recovery tools have been Approximately 10-15 mL of solvent
tray in the base of the impactor. This
designed to alleviate such problems in the small cups and 20-25 mL in
allows all eight cups to be removed in
and to ensure testing is carried out the large cups is normally sufficient
a single movement. It is then a simple
in a consistent, reproducible and to provide good coverage of the cup
matter to insert a new tray containing
controlled fashion. surface whilst avoiding spills during
eight clean cups into the NGI to
operation.
perform a further test. DESCRIPTION
The run time can be set by the
Once a test has been performed, the The Gentle Rocker, for example, is a
user, dependent on the nature of
analyst is required to dissolve the simple, economical device for gently
the dissolution. An evaporation-
active drug present in each sample agitating the contents of the NGI
eliminating cover is available as an
by adding a small amount of solvent collection cups in order to dissolve
optional extra in order to minimise any
to each cup and then agitating the the active drug in the solvent prior to
solvent loss during operation.
cup to dissolve the active drug in analysis.
the cup prior to analysis. A similar
technique must be employed with the
Cat. No. Description
mouthpiece adapter, induction port
and preseparator (if used). 5220 Gentle Rocker (complete with dust cover and 20 rpm motor)
5221 Gentle Rocker (complete with dust cover and 40 rpm motor)
Whilst the NGI itself has been
5223 Evaporation Cover (with seals and clips to prevent solvent loss)
designed to increase productivity in a
5255 Dust Cover (Spare)
standalone form using conventional
5224 Storage Cabinet for 6 NGI cup trays
wash and assay methods, the design
5225 IQ/OQ Documentation
has also led to further improvements
5235 Verification of Gentle Rocker
in productivity through the use of a
5256 Gentle Rocker Qualification Tools
number of specially designed labour
5257 Re-calibration of Gentle Rocker Qualification Tools
saving devices.
109
aUTOMATION
Particle bounce and re-entrainment can For this reason, it is important to assess The amount, uniformity and method
be a particular problem when using the likely effect of particle bounce and of application are critical parameters
cascade impactors to measure the subsequent re-entrainment of the within the coating process.
aerodynamic particle size distribution of particles down-stream to lower stages,
The NGI Cup Coater has been
OINDPs. at an early point in development with a
designed as a standardised approach
view to taking corrective action.
Particle bounce is a phenomenon to this problem and eliminates many
whereby the particle impacts against The normal method of addressing of the variables inherent in this
the collection surface appropriate to this problem is to coat the collection process.
its size but rather than adhering to that surfaces with a tacky viscous material
surface bounces back into the air such as, for example, glycerol or DESCRIPTION
stream, whereupon it is re-entrained silicone oil.
According to the Pharmacopoeias,
and passes to a lower stage than that
Another solution is to use an impinger, To ensure efficient particle capture,
intended.
such as the Glass Twin Impinger or coat the particle collection surface
This effect is particularly noticeable Multi-Stage Liquid Impinger (MSLI) in of each stage with glycerol, silicone
when the collection surface is solid, as which the collection surfaces are liquid, oil, or other suitable liquid typically
in the case of the Andersen Cascade as opposed to an impactor in which the deposited from a volatile solvent
Impactor (ACI) and Next Generation collection surfaces are solid. unless, in the case of USP, it has been
Impactor (NGI), and where the particles demonstrated to be unnecessary.
If a surface coating is employed, then
are hard and dry as in the case of dry
the amount, its uniformity, the method A wide variety of methods are
powder inhalers (DPIs).
in which it is applied and its potential currently in use for coating impactor
It may also be prevalent in some to affect the drug recovery process collection surfaces to meet this
formulations dispensed by Metered- (if applicable) should all be carefully requirement.
Dose Inhalers (MDIs) particularly where assessed during method development,
The NGI Cup Coater is unique in
only a few actuations are delivered to as these all could impact on the final
providing a reproducible method
the impactor. results.
of applying coatings directly to NGI
The result is to over-estimate the There is no one solution for all Collection Cups whilst in situ
amount of active available in the form inhaler devices each drug/device in the Collection Cup Tray, thus
of the fine particle dose and hence bias combination must be assessed as a eliminating the problem of
the measured size distribution data to separate entity. inter-analyst variability.
finer sizes.
110
aUTOMATION
The Coating Station measures 600 5900 NGI Cup Coater (excl. NGI Cup Tray & Cups)
mm x 170 mm x 230 mm and the 5901 500 mL Solvent Reservoir complete with 9-way Cap
Dispenser 150 mm x 220 mm x 130 5902 1000 mL Solvent Reservoir complete with 9-way Cap
mm (w x d x h).
111
Automation
ACI Carrying Rack in Ultrasonic ACI Carrying Rack in Rinse Bath ACI Carrying Rack in Drying Oven
Cleaning Bath
112
Automation
113
Automation
SEMI-AUTOMATION
ANDERSEN CASCADE
IMPACTOR (ACI)
The manual recovery of samples from During operation, all of the impactor The whole operation takes
such impactors can be tedious, time parts are held rigidly in position by a approx. 15 minutes. Once the
consuming and prone to human error. rinsing head to create a sealed system. run is complete, the mixing head
This allows a fixed volume of solvent to retracts into the ceiling of the
Systems that automate the sample
be introduced into each of the various sample processing area to allow the
recovery process can be a valuable
components in order to flush out the operator full access to the sample
asset to the pharmaceutical laboratory.
contents and circulate them through a preparation area and the system is
The Andersen Sample Recovery rinsing loop. flushed out ready for the next run.
System (A-SRS) has been specifically
Once the rinsing process is complete, The whole system is controlled by
designed to provide an accurate,
an aliquot from each of the 10 lines is means of a separate PC. Remarkably
reproducible and efficient means
extracted from the loop and injected simple to use, the Windows based
of recovering samples from the ACI
into a septum vial located in a storage software stores individual test
following testing thereby increasing
rack which can then be subsequently methods and provides an on-screen
throughput and reducing analyst-
removed for analysis. The storage rack display of progress throughout the
related variability.
has provision for a back-up run, i.e. run. It is 21 CFR Part 11 compliant.
Once the test is complete, the operator 20 vials in total. The HPLC vial can be
Emergency stop safety features
simply disassembles the ACI and sealed or open according to preference.
and ventilation extraction systems
places the various components (the
The Sample Collection and Waste are provided as standard. The
mouthpiece adapter, the induction port
Modules are located on the right of the system comes complete with all
and the various stages and collection
main processing area and are fitted with the necessary documentation and
plates) into their respective holders on
their own shatterproof transparent safety includes Factory Acceptance Testing
the sample recovery bed.
window. The system also incorporates (FAT), IQ/OQ documentation and
Each of the 10 components has its own an analytical balance for calibrating the on-site installation and completion
dedicated recovery sample channel in volumetric dispensing process. of the IQ/OQ.
the form of a closed liquid rinsing loop.
(Note: The Inlet Cone and Stage 0 are
treated as one component). Cat. No. Description
Typically between 30 and 50 mL 5420 Andersen Sample Recovery System (A-SRS)
of solvent is required to create an 5428 Quadruple Sample Capability
effective wash cycle. Up to 4 different
solvents can be employed during any 5426 A-SRS Spare Parts Kit
one test. 5427 A-SRS On-site Preventative Maintenance Service
114
Automation
INDUCTION PORT AND Under normal operating conditions, the The operator then places up to three
PRESEPARATOR RINSERS control of both rinsers is achieved by collection trays on the location lugs
a 3-position switch off / clamp / run. provided and adds a dust cover
The Induction Port and Preseparator
Safety interlocks and an emergency (or an optional sealed evaporation
Rinsers are two similar devices
stop button are provided as standard. cover) to prevent solvent loss. An
designed to automate the process of
x-y-z robotic pipette is then used to
rinsing the NGI Induction Port and
NGI ASSISTANT add a precise volume of solvent to
Preseparator.
each collection cup whereupon the
The NGI Assistant is a labour saving
Both units are bench standing and can rocking motion starts to dissolve the
device that places a known quantity
accommodate three induction ports or active in the solvent.
of solvent in each cup of three NGI
preseparators respectively.
cup trays, gently agitates the contents After dissolution, the robot then
Operation is extremely simple. A in order to dissolve the active drug aspirates an aliquot of each of
known volume of solvent is added in the solvent and then places a the dissolved samples into HPLC
to each induction port (ca. 20 mL) or representative sample of solution from vials for further analysis. Software
preseparator (ca. 30 mL) and the ends each of the cups into HPLC vials for controlled, the system allows pre-
secured using the end caps provided. subsequent analysis. defined methods to be saved and
recalled as necessary.
The induction ports or preseparators It takes approx. 45 60 minutes to
are then mounted on special fixtures process 3 trays. All of the above systems come
and clamped in position using sets of complete with all the necessary
By removing two of the three cup trays,
solenoid valve operated plungers. A documentation and include Factory
the NGI Assistant can also be used to
programmable logic controller is used Acceptance Testing (FAT), IQ/
dispense known quantities of solvent
to control the rinse time (normally 10 OQ documentation and on-site
into either three Induction Ports or
minutes). installation and completion of IQ/
Preseparators simultaneously prior to
OQ.
The actual rinsing action takes the rinsing.
form of a tumbling end-over-end
motion, combined in the case of
the preseparator rinser with an Cat. No. Description
axial turning motion such that the
preseparator rotates simultaneously 5407 NGI Induction Port Rinser
about two axes (35 rpm in the case of 5227 Set of 2 Stainless Steel Rinsing Caps for Induction Port (spare)
the Induction Port; 25 rpm in the case 5411 NGI Preseparator Rinser
of the Preseparator). 5228 Set of 2 Stainless Steel Rinsing Caps for Preseparator (spare)
This action together with the special
rinsing caps employed maximises the 5415 NGI Assistant (3-Tray)
potential for solvent to reach the cup 5223 Evaporation Cover to prevent solvent loss*
in the centre of the insert and hence * Note: 3 required for NGI Assistant
reduce rinsing times.
115
Automation
NGI SAMPLE RECOVERY SYSTEM Typically between 15 and 50 mL The sampling area will accommodate
(N-SRS) of solvent is required to create an four Agilent or Waters HPLC vial racks
effective wash cycle although smaller each capable of accepting 48 vials in an
The NGI Sample Recovery System
volumes can be used if required. 8 x 6 configuration. Other racks can be
(N-SRS) has been specifically
Up to 4 different solvents can be accommodated on request.
designed to provide an accurate,
employed during any one test.
reproducible and efficient means of The whole operation takes approx. 12-15
recovering samples from the NGI During operation, the cup tray is held minutes. Once the run is complete, the
following testing thereby increasing rigidly in position by a mixing head mixing head and pneumatic actuators
throughput and freeing up analysts to (in a similar manner to the NGI Seal for the induction port retract to allow full
pursue other duties. Body) to create a sealed system. This access to the sample preparation area
allows a fixed volume of solvent to and the system is flushed out ready for
Once the test is complete, the
be introduced into each of the cups the next run.
operator simply disassembles the
in order to flush out the contents and
impactor and places the various The whole system is controlled by means
circulate them through a mixing loop.
components, the mouthpiece of an in-built computer via a touch
adapter, the induction port and the Two pneumatic actuators perform the sensitive screen. Remarkably simple to
cup tray containing the collection same function on the induction port. use, the Windows based software stores
cups into their respective holders on individual test methods and provides an
Once the mixing process is complete,
the sample recovery bed. on-screen display of progress throughout
an aliquot from each of the 9 lines is
the run. It is 21 CFR Part 11 compliant.
Each component has its own extracted from the loop and injected
dedicated sample recovery channel into a sealed septum vial located Emergency stop safety features
in the form of a closed liquid rinsing in a storage rack which can then be and ventilation extraction systems
loop. In the case of Cups 1 to 7 subsequently removed for analysis. are provided as standard. The system
and the micro-orifice collector, this comes complete with all the necessary
The Sample Collection and Waste
comprises a syringe pump, a bi- documentation and includes Factory
Modules are located on the right of
directional metering pump, a solvent Acceptance Testing (FAT), IQ/OQ
the main processing area and are
retention bottle, 3-way solenoid documentation and on-site installation
fitted with their own shatterproof
valve, a fixture for mounting the and completion of IQ/OQ.
transparent safety window.
component, a check valve and a
dispensing needle. The ninth line
is shared between the mouthpiece Cat. No. Description
adapter and the induction port and
5417 NGI Sample Recovery System (N-SRS)
has two of every component with the
5434 Quadruple Sample Capability
exception of only one solvent bottle
and one dispensing needle.
5423 N-SRS Spare Parts Kit
5429 N-SRS On-site Preventative Maintenance Service
116
qualification
Qualification
117
qualification
Qualification Documentation
The performance of cascade impactors Inhaler Related Errors 3. Air Leaks into System:
and the methods associated with them Air leaking into impactor through
Inhaler related errors can also be
can be influenced by factors other than the inter-stage seals
controlled in a similar manner
the impactor itself. Poor seal between induction port/
by using (a) properly qualified
preseparator/impactor interfaces
Nevertheless, in broad terms, most of instrumentation and (b) validated
Improper seal between inhaler
the errors identified in APSD testing analytical procedures to use them.
mouthpiece and induction port
emanate from two sources:
PQRI list the following impactor
4. Collection Surface Preparation
Analytical (human error) related issues as potential causes of
Problems in coating collection
Instrument (errors in instrument error in the day-to-day measurement
surfaces (ACI, NGI, etc.)
and/or ancillary equipment) of particle size distributions:
Ensuring collection surfaces are
If these sources of error can be 1. Instrument Qualification moist (MSLI)
eliminated then it is fair to assume that Instrument not qualified prior to or
5. Flow Related Factors
any anomalies in results are a product during use
Incorrect flow rate setting
of the device/formulation combination Inter-stage losses (in excess of 5%)
Incorrect timer operation of
itself. Inadequate cleaning between tests
solenoid valve (DPIs)
Worn/corroded/clogged nozzles
These analytical and instrument
Improper sample recovery 6. Inhaler Actuations
factors have been systematically
Insufficient or excessive number of
reviewed by the PQRI (see Page 18) 2. Impactor Assembly
inhalations
Particle Size Distribution Mass Balance All collection plates or cups present
Working Group in an article entitled All collection plates or cups in the 7. Environmental Factors
Considerations for the Development correct order Barometric pressure (failure to
and Practice of Cascade Impaction All collection plates or cups seated account for)
Testing including a Mass Balance correctly Temperature and humidity (failure
Failure Investigation Tree, J.Aerosol. Final filter present and located to account for)
Med., 2003; 16(3):235-247. correctly Electrostatic effects
Inappropriate liquid levels (MSLI/
Analytical Errors Preseparators)
118
qualification
IMPACTOR QUALIFICATION
119
qualification
STAGE MENSURATION
INTRODUCTION
120
qualification
STAGE MENSURATION
CLEANING
121
qualification
Pinning Kit
with close-
Pinning various stages up of Pin
of the ACI
STAGE MENSURATION
RESTORING IMPACTOR Extreme care should be exercised in Kit is supplied with 14 pins, 2 pins to
PERFORMANCE performing this function to avoid nozzle each of the 7 stages concerned.
damage, particularly for aluminium
If stage mensuration results in an The NGI Pinning Kit is supplied with 12
impactors.
Effective Diameter in excess of an upper pins commensurate with Stages 1 to 6.
limit, then the stage must be replaced. The pins are precision manufactured in In the case of Stage 7, it is not practical
This is a sign that the nozzles have worn, a range of sizes corresponding to the to use a pin due to the small size of
either as a result of corrosion from the nozzle dimensions of the impactor stages the nozzles. A special brush with fine
solvents used to dissolve the active concerned. For each stage, there are two bristles is supplied instead.
drug or erosion from the constant rapid pin diameters provided, one pin having a
passage of particles through the nozzles slightly smaller diameter than the other. ADVANTAGES OF REGULAR
concerned. In this case there is no further The smaller diameter pin in each case MENSURATION
option available as it is not practical to can be used as an initial probe in cases,
A paper by Cooper and Blatchford
reapply metal to impactor nozzles. for example, where the holes in the stage
(3M Drug Delivery Systems) entitled
are heavily occluded and the larger pin
This situation is, however, rare, as the Statistical analysis of Andersen
cannot be inserted into the nozzle. In this
vast majority of impactors tend to drift Cascade Impaction mensuration data
instance, the pinning becomes a two-
out of specification because Effective presented at Drug Delivery to the
stage process.
Diameter decreases below the lower Lungs 19, December 10-12, 2008,
limit for the stage. The pins are supplied in wooden boxes. demonstrates the advantages of
Small pins are supplied in protective regular mensuration. Download the
This can be caused by a build-up of
tubes with sealing corks. The ACI Pinning paper at: www.copleyscientific.com.
hardened particulates or, more likely,
because the corrosion process produces
metal salts that occlude the nozzle. The Cat. No. Description
formation of oxidised impurities at the
nozzle exit is a commonly encountered 8590 Induction Port Mensuration
8390 ACI Stage Mensuration
cause of occlusion, particularly for
8391 ACI Returns Box
aluminium impactors, which is why
8990 60 L/min Conversion Kit Mensuration
materials such as stainless steel and
5236 90 L/min Conversion Kit Mensuration
titanium are now also used.
8490 ACI Preseparator Mensuration
If the Effective Diameter is too small, 5430 ACI Pinning Service (per stage)
performance can sometimes be 5431 ACI Pinning Kit
improved or restored. Ultrasonics 8311 ACI Stage Mensuration Histogram (per stage)
(see Page 112) can be used to clean 8890 MSLI Stage Mensuration and Leak Test
all impactors prior to mensuration. A 5290 NGI Stage Mensuration
combination of this and other rigorous 5291 NGI Preseparator Mensuration
cleaning is often sufficient to remove 5292 NGI Seal Body Returns Box
5432 NGI Pinning Service (per stage)
deposits and restore performance.
5433 NGI Pinning Kit
Otherwise, stage pinning is a secondary 8312 NGI Stage Mensuration Histogram (per stage)
option. Pushing stainless steel go pins 8591 Alberta Idealised Throat Mensuration
with a diameter between the nominal 8340 FSA Stage Mensuration
diameter and the lower tolerance limit 5270 FSI Insert Mensuration
for the stage through each nozzle can 8917 Glass Twin Impinger Mensuration
serve to clear accumulated debris.
122
qualification
QUALIFICATION DOCUMENTATION
According to USP Chapter <1058>, The IQ/OQ Documentation is written the qualification concept and the
Analytical Instrument Qualification to GxP standards and is designed for documentation structure to be used
is the collection of documented qualifying the complete test system during the qualification work.
evidence that an instrument performs including the test apparatus (e.g.
The second section comprises the
suitably for its intended purpose Dosage Unit Sampling Apparatus
Qualification Report Summary which
or Cascade Impactor), vacuum
It is important to note that the stage gives details and a summary of the
pump, critical flow controller (where
mensuration process described test results together with a list of users
applicable), flow meter and/or any
on previous pages is intended to trained in the system use.
other accessories that form part of an
replace the need for repetitive
inhaler testing system. 2. Installation Qualification
impactor calibration based on
This 16-page Section outlines the
standard aerosols. It ensures that The IQ/OQ Documentation is divided
test plan, the standard operating
only impactors that conform to into three chapters:
procedures and test protocols
specification are used in testing.
1. The Master Plan necessary to perform the IQ for
Whilst mensuration or calibration is The purpose of this 33-page Master the system concerned. It includes a
an important part of the qualification Plan is to define the aim and scope of general description of the system,
process, it does not in itself qualify the the qualification. delivery check, utilities/facility/
whole inhaler testing system for use. environmental factors, system
The first section of the Master
This is a separate process. installation and verification.
Plan describes in detail the various
The Installation Qualification/ constituents that normally make up an 3. Operation Qualification
Operation Qualification inhaler testing system and provides an This 44-page Section outlines the
Documentation (IQ/OQ) analysis of the likely risks associated test plan and the standard operation
Documentation provided by Copley with the parameters required to test procedures and test protocols to
Scientific guides the them. It goes on to describe the perform the OQ of the system
user through this various responsibilities assigned concerned. It includes the information
important process to the various parties necessary to carry out both fixed
and confirms that undertaking the parameter and functionality testing of
the system is qualification, the system.
fully qualified
for use.
123
qualification
124
qualification
QUALIFICATION TOOLS
The build-up of static electricity on A Digital Static Meter is also of standard aerosols. This is achieved
plastic, non-conductive surfaces such available which shows both intensity by passing standardised particles,
as those found in inhaler actuators and polarity of static charge in the of known dimensions, through the
and/or spacers can present specific range 0 to +/- 20 kV. This is a useful impactor at a precisely controlled flow
problems when working with inhalers, tool for ensuring that the static levels rate. The deposition of such particles
particularly dry powder devices. around equipment are not excessive within the impactor is then measured
and are controlled. to determine the effective cut-off
Generally speaking, metal objects
diameters applicable to each stage.
such as impactors do not present such
QUALIFICATION SERVICES
a problem. Indeed, most problems Whilst mensuration replaces the
of this nature emanate from the Leak test and Delta-P data can also need for repetitive calibration using
movement of the analyst around the be provided as an addition to our standard aerosols, cut-point calibration
laboratory prior to handling static normal mensuration service. Many may be of interest where specific cut-
sensitive equipment. users take advantage of this service points are required at flow rates that
and incorporate the resulting data are less commonly used or specified
These problems can be exacerbated
together with the mensuration data by the manufacturer.
by the low humidity levels often found
into their performance qualification
in air conditioned laboratories (<40%
files, to determine the impactors
RH) and also if the analyst is seated
continuing suitability for its intended
on a stool/chair and thereby isolated
use.
from the ground.
Antistatic
Such data is particularly relevant if Grounding
Irrespective of the source, it is
a stage or micro-orifice collector Kit
preferable to take action to reduce
has been replaced as a result of the
static to a minimum on the grounds
mensuration process.
that it is one less variable capable of
affecting the results. We are also able to offer a Cut-
Point Particle Calibration Service for
The Antistatic Grounding Kit
individual impactors based on the use
comprises a user wrist-band and
bench mat, both grounded through
the earth of an electrical socket. Cat. No. Description
This dissipates any static when
handling the impactor/inhaler and all 5233 ACI or NGI Leak Test Certificate
parts coming into contact with the 5234 ACI or NGI Delta-P Certificate
laboratory bench during preparation. 5251 LTK or NGI Leak Tester Re-calibration Certificate
5443 Re-calibration of Additional Tools (Cat. No. 5439)
The Electrostatic Eliminator is an 5442 ACI Cut-Point Particle Calibration Certificate
efficient ioniser with variable speed 9300 Antistatic Grounding Kit
fan and wide angle diffuser capable 9301 Electrostatic Eliminator
of eliminating static over a lab bench 9302 Digital Static Meter
area of 2 m x 0.6 m.
125
Design and Servicing
DESIGN
126
Training
TRAINING
127
INDEX
128
INDEX
G M P
129
INDEX
S V
130
COPLEY
SCIENTIFIC
Austria, France, Germany, UK, Ireland & International Sales: USA, Canada & Mexico:
Italy & Switzerland:
Tel: +41 (0)61 725 25 35 Tel: +44 (0)115 961 6229 Tel: +1 651 287 8100
Fax: +41 (0)61 721 31 87 Fax: +44 (0)115 961 7637 Fax: +1 651 287 8140