You are on page 1of 10

Med Biol Eng Comput (2013) 51:11671175

DOI 10.1007/s11517-013-1064-3

ORIGINAL ARTICLE

Monitoring intracellular, interstitial, and intravascular volume


changes during fluid management procedures
Leslie D. Montgomery Wayne A. Gerth Richard W. Montgomery

Susie Q. Lew Michael M. Klein Julian M. Stewart


Manuel T. Velasquez

Received: 10 August 2012 / Accepted: 17 March 2013 / Published online: 3 April 2013
International Federation for Medical and Biological Engineering 2013

Abstract The bioimpedance spectroscopic (BIS) analyti- between the BIS and CritLine hematocrits. BlandAltman
cal algorithm described in this report allows for the non- analysis show that the BIS algorithm can be used inter-
invasive measurement of intravascular, interstitial, and changeably with the CritLine monitor for the measurement
intracellular volume changes during various fluid manage- of hematocrit. The present study demonstrates for the first
ment procedures. The purpose of this study was to test time that BIS can provide real-time continuous measure-
clinical use feasibility and to demonstrate the validity of the ments of compartmental intravascular, interstitial and
BIS algorithm in computing compartmental volume shifts in intracellular fluid volume changes during fluid management
human subjects undergoing fluid management treatment. procedures when used in conjunction with this new
Validation was performed using volume changes recorded algorithm.
from 20 end stage renal disease patients. The validation
procedure involved mathematically deriving post hoc Keywords Bioimpedance spectroscopy  Fluid
hematocrit profiles from the BIS data-generated fluid redistribution  Hematocrit  Hemodialysis  Refill
redistribution time profiles. These derived hematocrit pro-
files were then compared to serial hematocrit values mea-
sured simultaneously by a CritLine monitor during 60 1 Introduction
routine hemodialysis sessions. Regression and BlandAlt-
man analyses confirm that the BIS algorithm can be used to Hemodynamic redistribution of fluids between body seg-
reliably derive the continuous and real-time rates of change ments and between the fluid compartments within those
of the compartmental fluid volumes. Regression results segments are of central importance to various fluid man-
yielded a R2 [ 0.99 between the two measures of hematocrit agement procedures [4, 25] and disease states [20, 2730,
at different times during dialysis. The slopes of the regres- 34]. Such fluid redistributions affect the cardiovascular
sion equations at the different times were nearly identi- function, water balance and perhaps skeletal muscle func-
cal, demonstrating an almost one-to-one correspondence tion through physiological mechanisms that may be bet-
ter understood with simultaneous characterization of
both the extent and the rate of the inter-compartment
redistributions.
Bioelectric impedance spectroscopy (BIS) has proved
useful for non-invasive monitoring of inter-compartmental
L. D. Montgomery (&)  W. A. Gerth  R. W. Montgomery
fluid shifts [2, 5, 8, 19, 23, 26, 33, 35, 38], among others.
LDM Associates, 1764 Emory Street, San Jose, CA 95126, USA
e-mail: pmontgomery@telis.org To date, however, this approach has been limited to two
compartments: intra- and extracellular volumes. The pres-
S. Q. Lew  M. T. Velasquez ent paper demonstrates that, using a new analytical algo-
George Washington University, Washington, DC, USA
rithm, it is possible to derive the fluid shifts that take place
M. M. Klein  J. M. Stewart between the intracellular, interstitial, and vascular com-
New York Medical College, New York, NY, USA partments from the current BIS output data.

123
1168 Med Biol Eng Comput (2013) 51:11671175

2 Methods parallel with the intracellular and interstitial compartments


of the other soft tissue, as illustrated in Fig. 1a.
2.1 Background The complex admittance Y of the circuit shown in
Fig. 1a is given as a function of the applied current fre-
Various resistance/capacitance (R/C) models [1, 6, 21, 24] quency f [16] by:
are used to quantify segmental volumes from recorded Y 1=Z 1=Rblood 1=Rbone 1=Re 1=Ri
impedance measurements. The R/C model that is used by 1
1=Ri =1 Ri Cm jx1  a:
most currently available impedance systems represents the
tissue as two parallel conductance paths, one through an where Z is a series of complex impedances, j = square root
extracellular compartment having average resistance (Re) of -1, x = 2pf is the angular frequency, Re, Ri, a and Cm
and the other through an intracellular compartment having are obtained from resistance and reactance values recorded
average resistance (Ri) and capacitance (Cm). This model, during each sweep of the frequency range used by the BIS.
which is limited to the determination of intracellular and An additional component of our analytical procedure is
extracellular resistances (Ri and Re, respectively), can the development and application of the extravascular soft
therefore only be used to estimate the corresponding tissue cellular model [16, 17, 32] illustrated in Fig. 1b. This
intracellular, extracellular, and overall volumes of a mon- model can be used to calculate the interstitial and intra-
itored body segment. Unfortunately, an issue of principal vascular compartment volumes that comprise the previ-
concern in many clinical contexts is how the extracellular ously calculated extracellular volume.
volume is distributed between its interstitial and intra- The extravascular soft tissue compartment of the mon-
vascular compartments; an issue that the two component itored segment is modeled as a cylindrical homogeneous
R/C models cannot address. suspension of identical oblate spheroidal cells with long
Since both blood and bone are nearly wholly resistive axes, a, oriented parallel to the applied field, I, as shown in
over the frequency range used in current BIS instruments Fig. 1b. Cells are allowed to change the volume only
[2, 37], they should affect only the passage of current in the through variation in minor axis, b. The cell volume fraction
extracellular volume compartment. This configuration can and the intracellular and extracellular conductivities of this
be approximated by considering blood and bone to be compartment are related to the compartments Re, Ri,
additional fluid compartments that are each electrically in and Cm in Fig. 1a with theory developed by Hugo Fricke
[1115].
The model shown in Fig. 1b is based on an assumed
physical structure of the monitored segment, particularly its
A i Cm
extravascular compartment (Fig. 1a), which governs the
values and frequency dependence of the resistive and
e
reactive components of the segmental impedance. The
model is fit to each measured spectrum by an iterative
blood
numerical process in which values of the model parameters
are found that bring model-prescribed spectra into closest
possible agreement with the measured spectrum. Because
bone
the model is a function of more parameters that are
uniquely determined by the information in an impedance
spectrum; selected model parameters must be assigned
fixed values in the fitting process. Detailed descriptions
B of this analytical approach and the solution of Eq. 1 can be
I found in Sasser et al. [32], Gerth et al. [17] and Gerth and
Watke [16].

2.2 Validation

b The previous sections describe how swept frequency bio-


impedance monitoring provides a convenient, minimally
a invasive means of monitoring intracellular, interstitial, and
vascular volume changes. The question is whether these
Fig. 1 a Equivalent circuit diagram used in BIS analysis. b Extra-
vascular soft tissue compartment model used in calculation of recordings reliably correspond to the actual volume chan-
intravascular and interstitial volumes ges in these three compartments. In the absence of invasive

123
1168 Med Biol Eng Comput (2013) 51:11671175

2 Methods parallel with the intracellular and interstitial compartments


of the other soft tissue, as illustrated in Fig. 1a.
2.1 Background The complex admittance Y of the circuit shown in
Fig. 1a is given as a function of the applied current fre-
Various resistance/capacitance (R/C) models [1, 6, 21, 24] quency f [16] by:
are used to quantify segmental volumes from recorded Y 1=Z 1=Rblood 1=Rbone 1=Re 1=Ri
impedance measurements. The R/C model that is used by 1
1=Ri =1 Ri Cm jx1  a:
most currently available impedance systems represents the
tissue as two parallel conductance paths, one through an where Z is a series of complex impedances, j = square root
extracellular compartment having average resistance (Re) of -1, x = 2pf is the angular frequency, Re, Ri, a and Cm
and the other through an intracellular compartment having are obtained from resistance and reactance values recorded
average resistance (Ri) and capacitance (Cm). This model, during each sweep of the frequency range used by the BIS.
which is limited to the determination of intracellular and An additional component of our analytical procedure is
extracellular resistances (Ri and Re, respectively), can the development and application of the extravascular soft
therefore only be used to estimate the corresponding tissue cellular model [16, 17, 32] illustrated in Fig. 1b. This
intracellular, extracellular, and overall volumes of a mon- model can be used to calculate the interstitial and intra-
itored body segment. Unfortunately, an issue of principal vascular compartment volumes that comprise the previ-
concern in many clinical contexts is how the extracellular ously calculated extracellular volume.
volume is distributed between its interstitial and intra- The extravascular soft tissue compartment of the mon-
vascular compartments; an issue that the two component itored segment is modeled as a cylindrical homogeneous
R/C models cannot address. suspension of identical oblate spheroidal cells with long
Since both blood and bone are nearly wholly resistive axes, a, oriented parallel to the applied field, I, as shown in
over the frequency range used in current BIS instruments Fig. 1b. Cells are allowed to change the volume only
[2, 37], they should affect only the passage of current in the through variation in minor axis, b. The cell volume fraction
extracellular volume compartment. This configuration can and the intracellular and extracellular conductivities of this
be approximated by considering blood and bone to be compartment are related to the compartments Re, Ri,
additional fluid compartments that are each electrically in and Cm in Fig. 1a with theory developed by Hugo Fricke
[1115].
The model shown in Fig. 1b is based on an assumed
physical structure of the monitored segment, particularly its
A i Cm
extravascular compartment (Fig. 1a), which governs the
values and frequency dependence of the resistive and
e
reactive components of the segmental impedance. The
model is fit to each measured spectrum by an iterative
blood
numerical process in which values of the model parameters
are found that bring model-prescribed spectra into closest
possible agreement with the measured spectrum. Because
bone
the model is a function of more parameters that are
uniquely determined by the information in an impedance
spectrum; selected model parameters must be assigned
fixed values in the fitting process. Detailed descriptions
B of this analytical approach and the solution of Eq. 1 can be
I found in Sasser et al. [32], Gerth et al. [17] and Gerth and
Watke [16].

2.2 Validation

b The previous sections describe how swept frequency bio-


impedance monitoring provides a convenient, minimally
a invasive means of monitoring intracellular, interstitial, and
vascular volume changes. The question is whether these
Fig. 1 a Equivalent circuit diagram used in BIS analysis. b Extra-
vascular soft tissue compartment model used in calculation of recordings reliably correspond to the actual volume chan-
intravascular and interstitial volumes ges in these three compartments. In the absence of invasive

123
Med Biol Eng Comput (2013) 51:11671175 1169

direct biochemical analysis of repeated fluid samples, it 2.4 Instrument


would be impossible to answer this question with absolute
certainty. Fortuitously, however, HD treatment sessions A tetrapolar BIS (UFI Inc., Morro Bay, CA, USA) was
provide ideal opportunities to monitor exogenously used to measure the tissue resistance and reactance at 40
induced inter-compartmental fluid shifts with minimal discrete frequencies once a minute during this study. The
additional procedures. BIS electrodes were attached to the subjects dominant
If the BIS net volume changes are correct, our compu- lower leg as shown in Fig. 2.
tations provide a way to track three compartment fluid The source electrodes were placed just above the knee
shifts, then such BIS-recorded fluid shifts should correlate and just above the lateral malleolus; the sampling elec-
with those of an independent physiologic measure of fluid trodes were placed just below the knee and above
volume changes. The HD patients hematocrit (HCT), the ankle. After a 30-min semi-reclined acclimatization/
expressed as %, rises (due to hemoconcentration) as excess instrumentation period, the measurements listed above
fluids are removed by ultrafiltration (UF)and falls or were recorded for a period of 30 min prior to HD. These
rises more slowly (due to hemodilution) as extravascular measurements served as the control values for the intra-HD
fluid moves into the vascular system as a result of com- and post-HD recovery period values. Systolic and diastolic
pensatory osmotic and hydrodynamic pressure changes. blood pressure (brachial cuff measures), and heart rate
Both changes in HCT and changes in the rate of vascular were measured every 30 min during the dialysis procedure
refill can be measured non-invasively during routine HD as per dialysis facility protocol. Body weights were mea-
therapy sessions. Changes in HCT can be measured by a sured before and after each HD session. Possible signs and
CritLine optical monitor (HemaMetrics, Kaysville, UT, symptoms of hypovolemia, such as breathlessness, dizzi-
USA). The rate of refill can be calculated from the BIS ness, presyncope, nausea, vomiting, thirst, fatigue, and
volume data. cramps were monitored and recorded by the attending
medical officer. Mid-session dialysis prescription altera-
2.3 Subject population tions, such as changes in dialysate and blood flow rates,
and net fluid removal were recorded continuously by the
Data used to validate our analytical procedures were col- dialysis machine.
lected from a series of dialysis treatments conducted at the A CritLine optical monitor served as the gold stan-
New York Medical Center (NYMC). The validation pro- dard for validation and was used simultaneously to detect
tocol was fully approved by the appropriate Institutional on-line changes in hematocrit. The CritLine measures the
Review Board prior to study initiation. Twenty chronic HD
patients over the age 21 who were capable of giving
informed consent underwent testing. Each patient was
tested during a nephrologist prescribed treatment using
Polyflux Series (8L, 170H or 210H) high-flux dialyzers
(Hechingen, Germany) for 3 to 4 h, and was tested three
times on subsequent days during a given week. In this way, Source 2
60 tests were conducted at NYMC. All testing was con-
ducted in a temperature-controlled room (2426 C).
Eleven male and nine female patients were included in
the test group. Even though both male and female patients Sampling 2
were tested, gender response differences were not the focus
of this work. Females, who have lower hematocrits [22]
were included to provide a wider range of hematocrit
Sampling 1
values.
The mean age of the 20 subjects was 65.3 11.3
(years). Their pre- and post-dialysis HCT (%) were
Source 1
34.84 2.50 and 38.47 3.60, respectively. Their pre-
and post-dialysis weights (kg) were 78.78 16.17 and
76.53 15.91, respectively. The HCT and weight changes
that took place during dialysis were 3.63 1.90 (%) and
2.25 1.13 (kg), respectively. A total mean of 2413
776.2 (ml) of fluid was removed during each of the 60
dialysis treatments. Fig. 2 BIS electrode placement

123
1170 Med Biol Eng Comput (2013) 51:11671175

optical absorption and scattering properties of red blood Recalling that fluid movement (ML/min) here is defined
cells as they pass through the HD circuit. Since red blood as the movement into the indicated compartment, a
cell mass and hemoglobin do not usually change during negative value of Mbi is a net movement of interstitial
treatment, the relative changes in hematocrit should par- (or extravascular) fluid into the vascular compartment.
allel the changes in blood fluid content. The CritLine Hereafter this will be referred to as vascular refill, and
instrument is well validated [36, 37], FDA-approved such refill will be assumed to offset some of the effect of
[9, 10] and routinely used for patient management. ultrafiltration (UF) on hematocrit. That is, as UF tends to
raise the value of HCT, a negative value of Mi tends
2.5 Data analysis to decrease HCT via hemodilution.
This relationship derived from the BIS data is then
Three calf compartment and total calf compartment net compared with minute-by-minute changes in HCT as
fluid volume changes are displayed in real time by the BIS recorded independently by the CritLine optical device.
monitor we have developed. This is shown as Fig. 3. These two variables are hereafter labeled D REFILL and D
The recorded minute-to-minute changes in each of these CLHCT, and the comparison consists of a simple least-
time series are used to calculate fluid shifts across com- squares regression: D CLHCT = a ? b (D REFILL),
partments. The calculation takes advantage of the fact that where a is the Y axis intercept, and b is the slope coefficient
intracellular fluid can only move into (or come from) the of the regression equation.
interstitial compartment; there is no other adjacent com-
partment. This implies that any change in intracellular fluid 2.6 Statistics
volume (Vc) must reflect a movement of fluid between the
intracellular and interstitial compartments. Thus, using the Statistical analysis was performed using MedCalc Ver.
notation Mic for fluid movement into cells from the 10.0.0.0. A Students t test for paired observations was
interstitial compartment, (ML/min): used to compare pre- and post-hemodialysis results. Tests
Mic DVc 2 of significant differences between the BISHCT and
CLHCT values at the specified times were performed by
The sign of Mic indicates the direction of movement. A non-paired Students t test. In addition, Pearsons correla-
negative value indicates an outflow from the intracellular tion and BlandAltman [3] analyses were used to compare
compartment into the interstitial compartment. the BISHCT and CLHCT results at each time. Results are
Proceeding in this manner, the movement of fluid from reported as mean SD. The variance in the results given
the vascular (blood) compartment (Mbi) into the interstitial for the analyses described in Sect. 3.1 and listed in Table 1
compartment and the movement of fluid from other seg- are reported as the SE of the estimate.
ments (Mob) into the vascular compartment are calculated First difference analyses were used in the calculation of
via a block triangular linear transformation of the com- BISHCT values from the impedance data to mitigate con-
partment volume changes: cern about spurious correlation due to autocorrelation in
Mbi DVc DVi 3
Mob DVc DVi DVb 4 Table 1 Results for regression changes of CritLine hematocrit on
BIS refill data

Calf compartment fluid volumes Subject BIS a b t value SE of


100 Observations est.
Intracellular
50 S1 152 0.186 -0.344 -12.361 0.005057
S2 154 -0.761 -0.195 -8.014 0.006674
Fluid volume (ML)

0
S3 137 0.407 -0.176 -4.037 0.011771
-50 Interstitial S4 167 -0.809 -0.058 -2.422 0.015844
-100 S5 152 0.160 -0.231 -2.679 0.021884
-150
S6 169 0.030 -0.166 -9.191 0.011536
Vascular S7 192 1.280 0.192 16.130 0.010501
-200
Calf total S8 156 -0.428 -0.313 -10.638 0.010638
-250 S9 141 0.251 -0.097 -12.802 0.006613
0 30 60 90 120 150 180 210 240
Elapsed time (minutes) S10 165 -0.058 -0.103 -5.219 0.018701
S11 164 0.065 -0.424 -9.083 0.014421
Fig. 3 Kalman-smoothed fluid compartment volumes relative to S12 145 0.344 -0.084 -5.412 0.011966
initial values

123
Med Biol Eng Comput (2013) 51:11671175 1171

un-differenced time series. DurbinWatson [7] and 38.0

Augmented DickeyFuller [31] tests were employed 37.5


37.0
nonetheless to assure that first-differenced series was not
36.5
autocorrelated and stationary, but still co-integrated. Sta-

HCT ratio
36.0
tistical tests also confirmed that D REFILL Granger-
35.5
causes [18] D CLHCT. 35.0
34.5
34.0 HCT
Prediction
3 Results 33.5
33.0
0 20 40 60 80 100 120 140 160 180 200
3.1 Controlled sample results
Elasped time (minutes)

Of the 60 sessions completed, 12 were best suited for val- Fig. 4 Comparison of CritLine HCT with BIS-predicted HCT
idation procedures as they involved no changes in the UF-
rate, no patient complications, and no interventions (such as between the two series for all sessions was evaluated by
therapeutic infusions). Table 2 presents the regression sampling all of the charts at a specified elapsed time,
results for this subset of HD sessions. The observations thereby obtaining 60 pairs of values (i.e., of CLHCT and its
indicated are the number of minutes of elapsed time during predicted value). This sampling procedure was repeated for
which the regression data were recorded. several intervals selected on the basis of the following
The t values for the estimate of the slope coefficient (b) considerations.
indicate that the null hypothesis of no relationship between All dialysis sessions included in this analysis lasted at
D REFILL and D CLHCT can safely be rejected (p  0.01 least 120 min. Shortly after 120 min, some of the dialysis
in every case). The standard errors (SE) are very small, sessions were terminated for different reasons: patients
ranging 0.0050570.021884. For example, assuming that experiencing syncope, cramps, completion of the required
nominal hematocrit values are approximately 36 during dialysis, etc. Mean calculated BIS and observed CritLine
dialysis, SE = 0.00507 (as in the first session above) hematocrit (SD) values are presented in Table 2 for
implies that 99 % (3 SD) of any errors in predicting elapsed dialysis times of 30, 60, 90, 120, and END. The
changes in CLHCT from the BIS data will be within END values that are given in Table 2 were taken from the
36 0.01521 (i.e., within four thousandths of one percent last minute before cessation of dialysis. The average (SD)
of the HCT value). time for the END data is 183.4 13.1 min from the start
The remarkably close relationship between the changes of dialysis. The CritLine hematocrit values measured by
in vascular refill calculated from the BIS data and the optical sensor in Table 3 and the following tables and
changes in CLHCT implies that the (un-differenced) tra- figures are designated CLHCT. The values calculated from
jectories of these two variables will be very close. This can the BIS measurements are designated by BISHCT. No
easily be illustrated by time integrating both series, and significant differences were found between the BISHCT
setting the constant of integration equal to the actual and the CLHCT values at any of the elapsed times listed in
HCT value as an initial condition. This is illustrated in Table 2.
Fig. 4 using the estimated coefficients for the first session Regression and BlandAltman [3] analyses were per-
shown in Table 1. formed to compare the BIS hematocrit values to the
This same procedure was used to produce graphs like observed CritLine values at each of the elapsed times
Fig. 4 for all 60 HD sessions. The strength of correlation listed in Table 3. Graphical results of the regression and
BlandAltman analyses at the specified elapsed times are
given in Fig. 5. Regression charts of BISHCT vs. CLHCT
Table 2 BIS and CritLine Hematocrit values SD at specified at elapsed times of 30, 60, 90, 120 min and during the last
elapsed times during dialysis minute of dialysis are given in Fig. 5a, c, e, h, and i,
ELAPSED N BISHCT BISHCT CLHCT CLHCT respectively. The solid trace in each of these charts rep-
TIME-min (%) SD (%) SD resents the corresponding regression equation and the
dashed traces show the 95 % confidence limits of the
30 60 35.92 2.9 35.93 2.8
regression. The regression equations and R2 values for each
60 60 36.35 3.1 36.37 3.1
of these comparisons are given in the appropriate panels
90 60 36.86 3.3 36.88 3.3
within Fig. 5.
120 60 37.34 3.4 37.39 3.4
The differences between BISHCT and CLHCT are
End 60 38.15 3.4 38.22 3.5
represented in BlandAltman graphs [3] for the same

123
1172 Med Biol Eng Comput (2013) 51:11671175

Table 3 Differences SD, limits of agreement, and independent The BlandAltman analysis (Table 3) indicates that
t test probabilities of significance between BISHCT and CLHCT there are no significant differences between the BISHCT
values at specified elapsed times during dialysis
and CLHCT values at any of the elapsed times. The mean
TIME- Difference with Limits of agreement p (by non- difference between BISHCT and CLHCT was 0.06066 at
min CLHCT (mean 1.96 SD) pair t test) the end of dialysis. This does indicate a slight overesti-
30 -0.00295 0.1331 -0.2638 0.3180 0.97 mation of the BISHCT as the time of treatment continues.
60 -0.00864 0.2021 -0.4048 0.4789 0.96 However, this difference is negligible when the two actual
90 0.01207 0.2263 -0.5338 0.5580 0.95 hematocrit values are compared. Given the mean hemato-
120 0.05661 0.2564 -0.5618 0.6750 0.92 crit of 38.22 for the CLHCT at the end of dialysis from
END 0.06066 0.2514 -0.5456 0.6669 0.91
Table 2 and the corresponding mean difference for that
time from Table 3 of 0.06066, the actual error is 0.15 % of
the CLHCT.
elapsed times in Fig. 5b, d, f, h, and j, respectively. The The ability to differentiate between volume changes in
mean difference at each time is shown by the solid hori- the three fluid compartments during fluid management
zontal line, the limits of agreement is shown by the large therapy will fill a critical need in many health care situa-
dashed traces, and the 95 % confidence interval is illus- tions. For example, in critically ill patients who are hypo-
trated by the small dashed lines in Fig. 5b, d, f, h, and j. tensive, such as in septic shock, ICU physicians often give
Table 3 lists the corresponding BlandAltman results. large amounts of fluids (as much as 10 or more liters a day)
to maintain blood pressure and hemodynamic stability.
Ideally, the replacement fluid should stay intravascular to
4 Discussion raise blood pressure. Unfortunately, if this fluid moves into
the interstitial space, this will lead to massive edema and if
This study describes how a new algorithm can be applied to it goes into lung tissue, it leads to acute respiratory distress
derive the intracellular, interstitial, and intravascular fluid syndrome (ARDS) which is associated with high mortality.
volume changes from the data provided by current BIS Treatment of this complication requires fluid removal by
instruments. Intra-dialysis blood volume changes were diuretic therapy in patients with normal kidney function
measured by a CritLine and these measurements were and renal replacement procedures in those with kidney
used to demonstrate and confirm the BIS algorithm results. failure. Usually these patients require invasive monitoring,
The results presented in Table 1 represent the subset of such as intra-arterial catheters to measure BP, and place-
treatments during which patients required no intervention ment of central venous catheters to measure wedge pres-
for hemodynamic change. These results demonstrate that sures, pulmonary arterial pressure, and cardiac output.
a simple linear model can be used to compare the BISHCT Continuous calculation and display of the intracellular,
values derived from the intracellular, interstitial, and interstitial, and intravascular volume changes would pro-
intravascular compartmental volume changes to the CLHCT vide a better guide in fluid management in such patients, so
measurements. that fluid administration could be adjusted only after
Regression and BlandAltman analyses of the data from intravascular repletion was obtained and before the volume
the entire cohort of 60 dialysis sessions were then com- overload occurred.
pleted to determine if the linear model is applicable to Our BIS analytical procedure will also be useful for
dialysis sessions during which patients require clinical intravenous fluid management in other acute settings, such
intervention, such as changes to ultrafiltration or flow rates. as patients suffering from septic or cardiogenic shock,
This data included results from dialysis sessions with extensive burns, lymphedema, or perhaps critical patients
varying UF rates and changes in patient posture, and ses- in emergency or operating rooms. Our non-invasive pro-
sions during which fluid or medications were administered cedure allows bedside monitoring of fluid shifts and, in
or food consumed. doing so, informs individualized, immediate clinical deci-
The R2 values obtained from the regression analyses sion making regarding fluid resuscitation.
shown in Fig. 5 are all greater than 0.9, ranging from 0.993 The general algorithm described in this work may be
to 0.994, thus, demonstrating a very close relationship used in conjunction with the output data from other BIS
between BISHCT and CLHCT. In addition, the slope instruments. However, most of our work to date has been in
coefficients of the various regression equations given in applications to data from the human calf, and we expect
Table 3, are all nearly identical confirming that there is that its use in application to whole legs or arms will require
almost a one-to-one correspondence between BISHCT and some modifications to better represent the cells in those
CLHCT at each of the elapsed times. body segments.

123
Med Biol Eng Comput (2013) 51:11671175 1173

Fig. 5 Linear regression and 50 1.0


BlandAltman analyses A Y = 0.4823 + 0.9863 X B

BISHCT30 - CLHCT30
comparing the BISHCT and 45 R2 = 0.997
0.5
CLHCT results at specified

BISHCT30
+1.96 SD
40
times during dialysis 0.26
Mean
0.0
-0.00
35 -1.96 SD
-0.27
-0.5
30

25 -1.0
25 30 35 40 45 50 30 35 40 45 50
CLHCT30 AVERAGE of BISHCT30 and CLHCT30
50 1.5
C Y = 1.1610 + 0.9676 X D

BISHCT60 - CLHCT60
BISHCT60 45 R2 = 0.993 1.0

0.5 +1.96 SD
40
0.39
Mean
0.0
35 -0.01
-1.96 SD
-0.5 -0.40
30
-1.0

25 -1.5
25 30 35 40 45 50 30 35 40 45 50
CLHCT60 AVERAGE of BISHCT60 and CLHCT60

50 1.5
E Y = 0.4599 + 0.9870 X F

BISHCT90 - CLHCT90
1.0
45 R2 = 0.995
BISHCT90

0.5 +1.96 SD
40 0.46
Mean
0.0
0.01
35 -1.96 SD
-0.5 -0.43
30
-1.0

25 -1.5
25 30 35 40 45 50 30 35 40 45 50
CLHCT90 AVERAGE of BISHCT90 and CLHCT90

50 1.5
BISHCT120 - CLHCT120

G Y = 0.3052 + 0.9903 X
1.0
H
45 R2 = 0.995
+1.96 SD
BISHCT120

0.5 0.56
40
Mean
0.0 0.06
35 -1.96 SD
-0.5 -0.45
30 -1.0

25 -1.5
25 30 35 40 45 50 30 35 40 45 50
CLHCT120 AVERAGE of BISHCT120 and CLHCT120

50 1.5
BISHCTEND - CLHCTEND

I Y = 0.6186 + 0.9819 X
1.0
J
45 R2 = 0.994
BISHCTEND

+1.96 SD
0.5 0.55
40
Mean
0.0 0.06
35 -1.96 SD
-0.5 -0.43
30
-1.0

25 -1.5
25 30 35 40 45 50 30 35 40 45 50
CLHCTEND AVERAGE of BISHCTEND and CLHCTEND

123
1174 Med Biol Eng Comput (2013) 51:11671175

In addition, specification of appropriate parameter val- 9. Federal Drug Administration (2002) K011741 Premarket
ues and of the initial conditions used in the application of Approval Crit-Line IIITQA URR Monitor
10. Federal Drug Administration (2010) K093834 Premarket
the algorithm to calculate interstitial volumes may be dif- Approval Crit-Line Anemia Management Monitor
ferent for different body segments that cannot be approx- 11. Fricke H (1924) A mathematical treatment of electric conduc-
imated as a cylinder or contain nonconductive voids, such tivity and capacity of disperse systems. I. The electric conduc-
as the head and torso. Each body segment may require a tivity of a suspension of homogeneous spheroids. Phys Rev
24:575587
different set of cellular parameters for a given application. 12. Fricke H (1925) A mathematical treatment of electric conduc-
The present study demonstrates that a new BIS algo- tivity and capacity of disperse systems. II. The capacity of a
rithm can be used to provide real-time continuous mea- suspension of conducting spheroids surrounded by a noncon-
surements of intracellular, interstitial, and intravascular ducting membrane for a current of low frequency. Phys Rev
26:678681
fluid volume changes during fluid management procedures. 13. Fricke H (1925) The electric capacity of suspensions of red
Such information may prove valuable in the diagnosis and corpuscles of a dog. Phys Rev 26:682687
management of rapid changes in body fluid balance. 14. Fricke H (1953) The electric permittivity of a dilute suspension of
membrane-covered ellipsoids. J Appl Phys 24:644646
Acknowledgments This work was funded, in part, by the National 15. Fricke H, Morse S (1925) The electric resistance and capacity of
Heart, Lung and Blood Institute of the National Institutes of Health blood for frequencies between 800 and 4.5 million cycles. J Gen
through SBIR Grants 1 R43 HL074524-01 and 2 R44 HL074524- Physiol 9:153167
02A2 entitled, Intra/Extracellular Volume and Hemodynamics 16. Gerth WA, Watke CM (1993) Electrical impedance spectroscopic
between 1 September 2003 and 31 July 2008. The authors wish to monitoring of body compartment volume changes. J Clin Eng
thank Ms. Sharon Hanish, Brian Scholfield, and Marty Loughry of 18(3):253260
UFI, Inc., Morro Bay, CA for their technical support and encour- 17. Gerth WA, Montgomery LD, Wu YC (1990) A computer-based
agement during this project. We, especially thank Dr. Jennifer E. bioelectrical impedance spectroscopic system for noninvasive
Flythe of Brigham and Womens Hospital, Harvard Medical School assessment of compartmental fluid redistribution. In: Proceedings
for her assistance in editing this manuscript. of third annual IEEE symposium on computer-based medical
systems. IEEE Computer Society Press, Chapel Hill, NC, pp 446
Conflict of interest The authors from LDM Associates were the 453
recipients of the National Heart, Lung and Blood Institute of the 18. Granger CWJ (1969) Investigating causal relations by econometric
National Institutes of Health through SBIR Grants 1 R43 HL074524- models and cross-spectral methods. Econometrica 37(3):424438
01 and 2 R44 HL074524-02A2 entitled, Intra/Extracellular Volume 19. Hannan WJ, Cowen SJ, Plester CE, Fearon KCH, DeBeau A
and Hemodynamics 1 September 2003 and 31 July 2008. None of (1995) Comparison of bio-impedance spectroscopy and multi-
the authors has received or will receive any compensation or mone- frequency bio-impedance analysis for the assessment of extra-
tary benefit from the publication of this article. cellular and total body water in surgical patients. Clin Sci
89:651658
20. Jaeger JQ, Mehta RL (1999) Assessment of dry weight in he-
modialysis: an overview. JASN 10:392403
21. Jindal GD (1986) Impedance plethysmography for screening
vascular disorders. J Postgrad Med 32:13
References 22. Kameneva MV, Watach MJ, Borovetz HS (1999) Gender dif-
ference in rheologic properties of blood and risk of cardiovascular
1. Ackman JJ, Seitz MA (1984) Methods of complex impedance diseases. Clin Hemorheol Microcircul 21:357363
measurement in biologic tissue. CRC Crit Rev Biomed Eng 23. Kamimura MA, Dos Santos NSJ, Avesani CM, Canziani MEF,
11:281311 Draibe SA, Cuppari L (2003) Comparison of three methods for
2. Bartok C, Schoeller DA (2004) Estimation of segmental muscle the determination of body fat in patients on long-term hemodi-
volume by bioelectrical impedance spectroscopy. J Appl Physiol alysis therapy. J Am Diet Assoc 103:195199
96(1):161166 24. Kanai H, Katsuyuki S, Haeno M (1983) Electrical measurement
3. Bland J, Altman DG (1996) Statistical methods for assessing of fluid distribution in human legs: estimation of extra- and
agreement between two methods of clinical measurement. Lancet intracellular fluid volume. J Microwave Power 18:233243
1:307313 25. Maisch B, Dristic A (2003) Practical aspects of the management
4. Boyd JH, Forbes J, Nakada T, Walley KR, Russell JA (2011) of pericardial disease. Heart 89:10961103
Fluid resuscitation in septic shock: a positive fluid balance and 26. Mendley SR, Majkowski NL, Schoeller DA (2005) Validation of
elevated central venous pressure are associated with increased estimates of total body water in pediatric dialysis patients by
mortality. Crit Care Med 39(2):259265 deuterium dilution. Kidney Int 67:20562062
5. Chertow GM, Lowrie EG, Wilmore DW, Gonzalez J, Lew NL, 27. Montgomery LD, Hanish HM, Marker RA (1989) An impedance
Ling J, Leboff MS, Gottlieb MN, Huang W, Zebrowski B (1995) device for study of multisegmental hemodynamic changes during
Nutritional assessment with bioelectrical impedance analysis in orthostatic stress. Aviat Space Environ Med 60:116122
maintenance HD patients. J Am Soc Nephrol 6(1):7581 28. Montgomery LD, Dietrich MS, Armer JM, Stewart BR, Ridner
6. Cole KS, Cole RH (1941) Dispersion and absorption in dielec- SH (2011) Segmental blood flow and hemodynamic state of
trics. J Chem Phys 9:341352 lymphedematous and non-lymphedematous arms. Lymphatic Res
7. Durbin J, Watson GS (1950) Testing for serial correlation in least Biol 9(1):3142
squares regression, I. Biometrika 37:409428 29. Nanovic L (2005) Electrolytes and fluid management in he-
8. Ellis KJ, Wong WW (1998) Human hydrometry: comparison of modialysis and peritoneal dialysis. Nutr Clin Pract 20(2):192201
multifrequency bioelectrical impedance with 2H2O and bromine 30. Ridner SH, Montgomery LD, Hepworth JT, Stewart BR, Armer
dilution. J Appl Physiol 85(3):10561062 JM (2007) Comparison of upper limb volume measurement

123
Med Biol Eng Comput (2013) 51:11671175 1175

techniques and arm symptoms between healthy volunteers and analysis compared to isotope dilution, dual energy X-ray
individuals with known lymphedema. Lymphology 40(1):3546 absorptiometry, and anthropometry. J Am Soc Nephrol 10:1067
31. Said SE, Dickey DA (1984) Testing for unit roots in autore- 1079
gressive-moving average models of unknown order. Biometrika 36. van der Sande FM, Kooman JP, Barendregt JNM, Nieman FHM,
71:599607 Leunissen KML (1999) Effect of intravenous saline, albumin, or
32. Sasser DC, Gerth WA, Wu YC (1993) Monitoring of segmental hydroxyethylstarch on blood volume during combined ultrafil-
intra- and extracellular volume changes using electrical imped- tration and hemodialysis. J Am Soc Nephrol 10:13031308
ance spectroscopy. J Appl Physiol 74:21802187 37. van der Sande FM, Luik AJ, Kooman JP, Verstappen V, Leun-
33. Shulman T, Heidenheim AP, Kianfar C, Shulman SM, Lindsay issen KML (2000) Effect of intravenous fluids on blood pressure
RM (2001) Preserving central blood volume: changes in body course during hemodialysis in hypotensive-prone patients. J Am
fluid compartments during hemodialysis. ASAIO J 47:615618 Soc Nephrol 11:550555
34. Stewart JM, Medow MS, Glover JL, Montgomery LD (2006) 38. Van Marken Lichtenbelt WD, Snel YEM, Brummer RJM, Ko-
Persistent splanchnic hyperemia during upright tilt in postural ppeschaar HPF (2005) Deuterium and bromide dilution, and
tachycardia syndrome. Am J Physiol Heart Circ Physiol 290: bioimpedance spectrometry independently show that growth
H665H673 hormone-deficient adults have an enlarged extracellular water
35. Van Den Ham ECH, Kooman JP, Christiaans MHL, Nieman compartment related to intracellular water. J Clin Endocrinol
FHM, Van Kreel BK, Heidendal GAK, Van Hooff JP (1999) Metab 82(3):907911
Body composition in renal transplant patients: bioimpedance

123

You might also like