PM CHECK LIST FOR ANALYSER-CONDUCTIVITY

BCIN-007
---------------------------------------------------------------
TASK DESCRIPTION FRQUENCY Q
---------------------------------------------------------------
1.0 REFERENCE DOCUMENT & MATERIALS
1.1 REFER RELEVANT DRAWINGS AND DOCUMENT.
1.2 COLLECT REQUIRED DATA FROM DATA SHEET.
1.3 PREPARE TEST EQUIPMENT AND TOOLS REQUIRED FOR
CHECKS/TESTS.
----------------------------------------------------------------
2.0 SAFETY & PERMIT
2.1 OBTAIN WORK PERMIT FROM DESIGNATED AUTHORITY.
2.2 OVERRIDE ANY S/D ACTION AS REQUIRED AND RECORD AS
APPLICABLE
2.3 ENSURE NECESSARY EQUIPMENT ISOLATION IN PLACE.
2.4 DISPLAY CAUTION AND WARNING NOTICES.
2.5 ADVISE CONTROL ROOM OPERATOR TO CHECK INDICATION AND
ACKNOWLEDGE ALARM AS NECESSARY.
2.6 AS APLLICABLE TAKE THE LOOP ON MANUAL & PUT THE LOOP
BACK TO AUTO AFTER JOB COMPLETION
2.7 OBSERVE ALL SAFETY PRECAUTIONS.
----------------------------------------------------------------
3.0 WORK SUMMARY
3.1 CHECK FOR POWER SOURCE AND ITS CONNECTIONS
3.2 ISOLATE PROCESS SAMPLE LINE
3.3 REMOVE ELECTRODE AND CHECK THE CONDITION
3.4 CLEAN THE ELECTRODE WITH PROPER SOLVENT
3.5 CHECK THE RESISTANCE OF THE TEMPERATURE COMPENSATOR
3.6 PERFORM TEMPERATURE CALIBRATION
3.7 PERFORM CALIBRATION WITH KNOWN SOLUTION.
3.8 VERIFY PRESENT READING W/TESTED SAMPLE LAB RESULT
3.9 ENSURE SENSOR IS PROPERLY MOUNTED IN TEE ASSY
3.10 ENSURE CORRECT LOOP POLARITY AND TIGHTEN CONNECTIONS
3.11 ENSURE WIRING NOT SHORTED OR OPEN
3.12 RESTORE PROCESS SAMPLE LINE
3.13 ENSURE FREE FLOW OF PROCESS FLUID AND SENSOR FULLY IMMERSED
3.14 CHECK THE FUNCTION OF ANALYSER
3.15 LINE UP AND CONFIRM LOOP HEALTHY
3.16 CHECK THE CONDITION OF LOCAL DISPLAY METER
3.17 VISUALLY CHECKS ALL FIELD WIRING /CABLES TIDE-UP AS
NECESSARY.
3.18 RAISE SEPARATE WR FOR CORRECTIVE ACTION AS REQUIRED.
----------------------------------------------------------------
4.0 JOB COMPLETION.
4.1 CLEAN UP THE WORK PLACE
4.2 ENSURE INTERLOCKS ARE NORMALISED
4.3 SIGN OFF WORK PERMIT.
4.4 REMOVE CAUTION AND WARNING NOTICES.
4.5 LOG IN TEST RESULTS AND “AS FOUND” VALUES.
----------------------------------------------------------