Vieworks – How to calibrate a panel

1. Log in to the VIVIX Setup


Utility

2. Make sure to have all


connected hardware
highlighted

3. Once highlighted, click


Next >
4. Locate the detector you
will be calibrating and click
Calibration
5. Select the Offset
Calibration tab

6. Select Download Offset

7. Browse to the location


where the calibration file is
located. (14x17 panels will
be found on the calibration
disk, 17x17 panels will be
found on the panel)

8. Click save when file is


located

9. Click OK when window


pops up
10. Select the Defect
Calibration Tab

11. Click on Download


Defect Map
12. Browse to the location
where the calibration file is
located. (14x17 panels will
be found on the calibration
disk, 17x17 panels will be
found on the panel)

13. Click Save

14. Click OK when window


pops up
15. Select the Gain
Calibration Tab

16. Take an exposure and


try to get within 10 %
threshold of the Target
Value

17. Once you get your


current value within 10 %
of your Target Value, click
on the Get button
18. Your stage will progress
from 0/10 to 10/10 as
shots are acquired

19. A save calibration


screen will then appear.

20. Save Gain calibration


21. Click OK when window
pops up

22. Calibration of the panel


is now complete. If
multiple panels are
connected to the system,
repeat previous steps.
If you have any issue feel free to give UMG Technical
Support a call 1-800-261-9808