Safety - Electrical Test


System
G20 Electrical Safety Test Limits

08647831
08647815

© Siemens AG 2004
The reproduction, transmission or use
of this document or its contents is not
permitted without express written
Siemens Medical
Ultrasound
lne DivisionSolutions
authority. Offenders will be liable for
damages. All rights, including rights
created by patent grant or registration
of a utility model or design, are
reserved.

Print No.: US06-101.860.01.02.02 English


Replaces: US06-101.860.01.01.02 Doc. Gen. Date: 04.05
n.a.

Part No.: n.a.


sd
2 Revision / Disclaimer
1Revision / Disclaimer

Copyright
Copyright © 2004 by Siemens Corporation. All rights reserved. No part of this publication
may be reproduced, transmitted, transcribed, stored in retrieval systems, or translated into
any language or computer language, in any form or by any means, electronic, mechanical,
magnetic, optical, chemical, manual, or otherwise, without the prior written permission of
Siemens Corporation. Siemens Corporation reserves the right to change its products and
services at any time. In addition, this manual is subject to change without notice. Siemens
Corporation welcomes customer input on corrections and suggestions for improvements
to this manual.Although Siemens Corporation has attempted to ensure accuracy through-
out this manual, Siemens Corporation assumes no liability for any errors or omissions, nor
for any damages resulting from the application or use of this information.
Trademarks
ACUSON, Sequoia, Cypress, AEGIS, Aspen, 128XP, XP, AcuNav, CV70, SONOLINE,
Adara, Antares, G20, G40, G50, G60S, Omnia and KinetDx are trademarks of Siemens
Corporation registered in the U.S. Patent and Trademark Office. CWS3000, DS3000,
WS3000, and WebPro are trademarks of Siemens Corporation.
Windows XP, 2000 and NT are registered trademarks of Microsoft Corporation. Internet
Explorer is a trademark of Microsoft Corporation. Netscape and Netscape Navigator are
registered trademarks of Netscape Communications Corporation. Java and the JavaScript
name are registered trademarks of Sun Microsystems, Inc. Adobe Acrobat is a registered
trademark of Adobe Systems, Inc.
Other products and brand names are trademarks of their respective owners.
Disclaimer
The service of equipment described herein is to be performed by qualified personnel who
are employed by Siemens or one of its affiliates or who are otherwise authorized by Sie-
mens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated
with or authorized by Siemens or one of its affiliates are directed to contact one of the
local offices of Siemens or one of its affiliates before attempting installation or service pro-
cedures.
License Agreement
All computer programs copyright 1990-2004 by Siemens Corporation or its suppliers. Pro-
grams are licensed under the following agreement:
Siemens or its suppliers retain(s) ownership of and title to any computer program supplied
with the equipment and to the trade secrets embodied in such computer programs. Sub-
ject to the Buyer’s acceptance and fulfillment of the obligations in this paragraph, Siemens
grants the Buyer a personal, non-transferable, perpetual, non-exclusive license to use any
computer program supplied with the Equipment that is necessary to operate the Equip-
ment solely on the medium in which such program is delivered for the purpose of operat-
ing the equipment in accordance with the instructions set forth in the operator’s manuals
supplied with the Equipment and for no other purpose whatsoever. Buyer may not
reverse-assemble, reverse-compile or otherwise reverse-engineer such computer pro-
grams nor may Buyer make a copy of such program or apply any techniques to derive the
trade secrets embodied therein. In the event of a failure by Buyer to comply with the terms

SONOLINE G20 US06-101.860.01.02.02 Page 2 of 10 Siemens


04.05 sd Medical Solutions USA, Inc.
Revision / Disclaimer 3

of this license, the license granted by this paragraph shall terminate. Further, because
unauthorized use of such computer programs will leave Siemens without an adequate
remedy at law, Buyer agrees that injunctive or other equitable relief will be appropriate to
restrain such use, threatened or actual. Buyer further agrees that (i) any of Siemens"s
suppliers of software is a direct and intended beneficiary of this end-user sublicense and
may enforce it directly against Buyer with respect to software supplied by such supplier,
and (ii) No supplier of Siemens shall be liable to buyer for any general, special, direct, indi-
rect, consequential, incidental or other damages arising out of the sublicense of the com-
puter programs supplied with the equipment.

Siemens US06-101.860.01.02.02 Page 3 of 10 SONOLINE G20


Medical Solutions USA, Inc. 04.05 sd
4 Table of Contents
1- 0Table of Contents

1 _______ G20 Electrical Safety Test Limits ___________________________________ 5

Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Test Limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Part 1: Visual Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Part 2: Protective Ground Continuity Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Part 3: Ground Leakage Current Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Part 4: Chassis Leakage Current Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Part 5: ECG Leakage Current Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Part 6: Transducer Leakage Current Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Part 7: Equivalent Leakage Current Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Part 8: Leakage Current Test for Specialty Transducers . . . . . . . . . . . . . . . . . . . . . 10

SONOLINE G20 US06-101.860.01.02.02 Page 4 of 10 Siemens


04.05 sd Medical Solutions USA, Inc.
G20 Electrical Safety Test Limits 5
2-
Overview
1G20 Electrical Safety Test Limits

The Electrical Safety Test Limits document provides the test limits for the electrical safety
test, which must be performed on the SONOLINE G20 ultrasound imaging systems during
planned maintenance, after servicing the system if mechanical or electrical safety may
have been affected, or if a safety check is specified in a service procedure.
For detailed information about performing the electrical safety test in the field, refer to the
(US88-100.888.01 / ACUSON/SONOLINE Electrical Safety Test Procedure).

NOTE In the data tables, Table 1 through Table 6, the switch numbers S1,
S2, and S3 refer to the analyzer switches designated in the draw-
ings of the related procedures contained in (US88-100.888.01 /
ACUSON/SONOLINE Electrical Safety Test Procedure).

CAUTION Improper use of the safety analyzer can affect the validity of the
test or cause the safety analyzer fuse to open.
¹ Review the operating instructions carefully before using
the electrical safety analyzer.

CAUTION When performing safety tests with the electrical safety analyzer,
always verify that the UUT has had sufficient time to re-initialize,
after a change of the test mode, before performing the next test.
¹ Failure to do so may result in inaccurate measurements.

CAUTION When performing safety tests, ensure the power up and power
down sequences are complete and the system is fully de-ener-
gized prior to changing any tester switch settings.
¹ Failure to do so may cause software corruption.

Siemens US06-101.860.01.02.02 Page 5 of 10 SONOLINE G20


Medical Solutions USA, Inc. 04.05 sd
6 G20 Electrical Safety Test Limits

Test Limits 0

NOTE The values listed in this Electrical Safety Test Limits document for
the G20, are factory maximum allowable limits for the G20 system,
or SMS Ultrasound maximum allowable per the Compliance SOP.
They are not nominal or average values.

Part 1: Visual Inspection 0

Inspect the system for internal and external electrical faults. Use the following criteria to
pass or fail the visual inspection:
• Pass—if no internal or external electrical faults are found.
• Pass—if internal or external electrical faults are found and they can be corrected.
• Fail—if internal or external electrical faults are found and they cannot be corrected.

Part 2: Protective Ground Continuity Test 0

The protective ground continuity test passes if:


• The measured values do not exceed the values specified below and if the measured
values do not exceed 1.5 times the initial factory values, if the factory data is available.

Test Location Test Limit


From the power cord ground pin to the equipotential connector that ≤100 mΩ
is typically located near where the power cord enters the chassis, or
to another metal component of the main chassis
From the power cord ground pin to an exposed metal component of ≤ 200 mΩ
the monitor, or to the mounting hardware of the system monitor
From the power cord ground pin to an exposed metal component of ≤ 200 mΩ
the control panel, or to the mounting hardware of the control panel
From the power cord ground pin to a metal component of an ≤ 200 mΩ
on-board OEM chassis

Part 3: Ground Leakage Current Test 0

The system passes the ground leakage current test if:


• The measured value for each test configuration does not exceed the values specified in
(Tab. 1 / p. 7) or (Tab. 2 / p. 7) by 1.5 times.

SONOLINE G20 US06-101.860.01.02.02 Page 6 of 10 Siemens


04.05 sd Medical Solutions USA, Inc.
G20 Electrical Safety Test Limits 7

• The measured value for each test configuration does not exceed the values by a total of
10 µA above the value (whichever limit is greater).

NOTE Different tolerances apply to different system nominal voltages.

Tab. 1 Ground Leakage Current Criteria for 100/115 V Systems

Configuration Ground Leakage


Maximum Maximum
Setup Polarity S2 Ground S3 Neutral S1 Minimum (at 50 Hz) (at 60 Hz)

1 Normal Normal1 Normal 0 µA 5 µA 5 µA


2 Normal Normal1 Open 0 µA 5 µA 5 µA
3 Normal Open Normal 10 µA 70 µA 84 µA
4 Normal Open Open 50 µA 135 µA 162 µA
5 Reversed Normal1 Normal 0 µA 5 µA 5 µA
6 Reversed Normal1 Open 0 µA 5 µA 5 µA
7 Reversed Open Normal 10 µA 70 µA 84 µA
8 Reversed Open Open 50 µA 135 µA 162 µA

1. When using the Fluke 505 Pro to perform measurements, the ground switch is not functional, there-
fore a ground reading is not possible.

Tab. 2 Ground Leakage Current Criteria for 230 V Systems

Configuration Ground Leakage


Maximum Maximum
Setup Polarity S2 Ground S3 Neutral S1 Minimum (at 50 Hz) (at 60 Hz)

1 Normal Normal1 Normal 0 µA 8 µA 8 µA


2 Normal Normal1 Open 0 µA 5 µA 5 µA
3 Normal Open Normal 10 µA 170 µA 200 µA
4 Normal Open Open 70 µA 315 µA 375 µA
5 Reversed Normal1 Normal 0 µA 8 µA 8 µA
6 Reversed Normal1 Open 0 µA 5 µA 5 µA
7 Reversed Open Normal 10 µA 170 µA 200 µA
8 Reversed Open Open 70 µA 315 µA 375 µA

Siemens US06-101.860.01.02.02 Page 7 of 10 SONOLINE G20


Medical Solutions USA, Inc. 04.05 sd
8 G20 Electrical Safety Test Limits

1. When using the Fluke 505 Pro to perform measurements, the ground switch is not functional, there-
fore a ground reading is not possible.

Part 4: Chassis Leakage Current Test 0

The system passes the chassis leakage current test:


• If the measured value for each test configuration does not exceed the values specified
in (Tab. 3 / p. 8) or (Tab. 4 / p. 8) by 1.5 times.
• If the measured value for each test configuration does not exceed the values by a total
of 10 µA above the value listed (which ever is greater).

NOTE Different tolerances apply to different system nominal voltages.

Tab. 3 Chassis Leakage Current Criteria for 100/115 V Systems

Configuration Chassis Leakage


Maximum Maximum
Setup Polarity S2 Ground S3 Neutral S1 Minimum (at 50 Hz) (at 60 Hz)

1 Normal Normal Normal 0 µA 5 µA 5 µA


2 Normal Normal Open 0 µA 5 µA 5 µA
3 Normal Open Normal 10 µA 70 µA 84 µA
4 Normal Open Open 50 µA 135 µA 162 µA
5 Reversed Normal Normal 0 µA 5 µA 5 µA
6 Reversed Normal Open 0 µA 5 µA 5 µA
7 Reversed Open Normal 10 µA 70 µA 84 µA
8 Reversed Open Open 50 µA 135 µA 162 µA

Tab. 4 Chassis Leakage Current Criteria for 230 V Systems

Configuration Chassis Leakage


Maximum Maximum
Setup Polarity S2 Ground S3 Neutral S1 Minimum (at 50 Hz) (at 60 Hz)

1 Normal Normal Normal 0 µA 8 µA 8 µA


2 Normal Normal Open 0 µA 5 µA 5 µA
3 Normal Open Normal 10 µA 170 µA 200 µA

SONOLINE G20 US06-101.860.01.02.02 Page 8 of 10 Siemens


04.05 sd Medical Solutions USA, Inc.
G20 Electrical Safety Test Limits 9

Configuration Chassis Leakage


Maximum Maximum
Setup Polarity S2 Ground S3 Neutral S1 Minimum (at 50 Hz) (at 60 Hz)

4 Normal Open Open 70 µA 315 µA 375 µA


5 Reversed Normal Normal 0 µA 8 µA 8 µA
6 Reversed Normal Open 0 µA 5 µA 5 µA
7 Reversed Open Normal 10 µA 170 µA 200 µA
8 Reversed Open Open 70 µA 315 µA 375 µA

Part 5: ECG Leakage Current Test 0

Not required.

Part 6: Transducer Leakage Current Test 0

The system and the transducers pass the transducer leakage current test if:
• The measured value for each transducer does not exceed the corresponding value
specified in (Tab. 5 / p. 9) by 10 µA or 1.5 times the value listed, whichever is greater.
• If the measured value does not exceed the initial factory value (if the data is available)
by 10 µA or 1.5 times the factory value, whichever is greater.
Tab. 5 Transducer Leakage Current Limits

Leakage Current Limit


Transducer
115V/50Hz (µA) 115V/60 Hz (µA) 230V/50 Hz (µA) 230V/60Hz (µA)
C5-2 ≤ 9 µA ≤ 10 µA ≤ 17 µA ≤ 20 µA
ENDO-V II ≤ 9 µA ≤ 10 µA ≤ 17 µA ≤ 20 µA
EC9-4 ≤ 9 µA ≤ 10 µA ≤ 17 µA ≤ 20 µA
EV9-4 ≤ 9 µA ≤ 10 µA ≤ 17 µA ≤ 20 µA
L10-5 ≤ 9 µA ≤ 10 µA ≤ 17 µA ≤ 20 µA
7.5L75S ≤ 9 µA ≤ 10 µA ≤ 17 µA ≤ 20 µA

Siemens US06-101.860.01.02.02 Page 9 of 10 SONOLINE G20


Medical Solutions USA, Inc. 04.05 sd
10 G20 Electrical Safety Test Limits

Part 7: Equivalent Leakage Current Test 0

For regions using the IEC 60601-1 standard, equivalent leakage current measurements
involve comparing the measured value to the value of an initial measurement taken previ-
ously. The value of the current measurement must not exceed the value of the initial mea-
surement by 1.5 times, or by a total of 10 µA above the initial value (whichever limit is
higher).

NOTE Perform this test only if the other leakage current tests included in
(US88-100.888.01 / ACUSON/SONOLINE Electrical Safety Test Pro-
cedure) cannot be performed. Consult ARTD-002.731.17.01.02
“Safety Regulations for Installation and Maintenance” for informa-
tion on this test.

Part 8: Leakage Current Test for Specialty Transducers 0

Not required.